3M 718 Instruction Manual

Page 1
Instruction Manual
3
718 Static Sensor
Safety Information
Intended Use
The 3M™ 718 Static Sensor is a portable, handheld instrument designed for measuring voltages associated with electrostatic charge. Its intended use is for measuring the amount of voltage, in a range from 0-20 kilovolts, associated with an electrostatic charge buildup on a surface. Any deviation from this intended use could impair the instrument’s effectiveness and possibly lead to an unsafe operating condition.
The 3M™ 718A Air Ionizer Test Kit is a set of accessories for use in conjunction with the 3M 718 Static Sensor. Their intended use is for verifying the operation of air ionizers by measuring the neutralization time for a static charge placed on a stationary metal plate. Any deviation from this intended use could impair the instrument’s effectiveness and possibly lead to an unsafe operating condition.
Caution Statements
• The 718 Static Sensor and 718A Charger use DC 9V power supplied by a 9V alkaline battery. Usage of any other power source may cause damage to the instruments.
• The 718 Static Sensor and 718A Air Ionizer Test Kit have no user-serviceable parts. Do not disassemble the products for any reason. UNAUTHORIZED SERVICE WILL VOID THE WARRANTY.
• The 718 and 718A are NOT designed for usage in hazardous environments where the possibility of explosion or fire exists.
Read and understand all safety information before installing and operating this equipment.
Page 2
Page 3
1.0 Description
The 3M™ 718 Static Sensor is a portable handheld instrument used for locating and measuring electrostatic charges. It can be used to locate ESD trouble-areas, and is a valuable tool for the ESD-control engineer. Used in conjunction with the 3M™ 718A Air Ionizer Test Kit (available separately), it can be used for verification and auditing of air ionizers. The 718 Static Sensor is battery-powered and has several measurement features: Range: measurements can be taken in a 0­2 kV or 0-20 kV range. Automatic Zero: pushbutton feature allows easy adjustment to zero. No screws or dials to turn. HOLD function: allows the User to "freeze" a displayed measurement, for later evaluation. Automatic shutoff: conserves battery power by shutting off the instrument after 20 minutes of inactivity.
2.0 Power Requirements and
Battery Installation
2.1 Both the 718 and 718A Charger us
DC 9V power supplied by a 9V alkaline battery.
2.2 To install the battery on either unit:
2.2.1 Remove the battery cover, located in
the lower back of the unit. To do this, press down on the cover and slide it downward.
2.2.2 Pull the battery connector out of the
housing, and align the male/female ends of the connector with the proper terminals on the battery.
2.2.3 Connect the two, and place the connected battery into the housing by inserting the connector end first, then following up with the other end of the battery.
2.2.4 Replace the cover.
2.3 The 718 Static Sensor has a Low
Battery indicator. Once the battery is depleted to approximately 6.5 volts, the instrument will show BAT in the display. At this time, the 718 will not produce accurate results and the battery should be replaced. A fresh alkaline battery should last greater than 50 hours, when the instrument is in continuous usage.
2.4 The 718A Charger also has a low-
battery indicator. This is an LED located at the left-hand side of the unit. When the battery voltage drops below operating level, the LED will light up. At this time, the user should
Page 4
replace the battery. Usage of the 3M™ 718A Charger under low­battery conditions would lead to insufficient voltage levels being generated.
3.0 Operation of the 718 Static Sensor
Supplemental notes: For accurate measurements during usage, it is recommended that the outside housing of the 3M™ 718 Static Sensor be connected to an electrical ground. This can be accomplished by having the user holding the instrument connected to ground through either a static control wrist strap, or while wearing static control footwear. The enclosure of the 718 is made of conductive plastic and is, therefore, electrically connected to whomever is holding the instrument. In addition, ground for the instrument can be provided through the Voltage Monitor Output on the front of the case.
The 718 Static Sensor is a precision electronic instrument. Improper use or rough treatment can damage the unit, and render it incapable of providing accurate measurements.
3.1 Turning the unit ON and OFF: To turn on the 718 Static Sensor, momentarily press down on the membrane switch labeled POWER. Determining that the power is on can be verified by seeing that the liquid­crystal display (LCD) is on, and that the red light-emitting diodes in the front of the sensor are lit. To shut off the 718 Static Sensor, momentarily press down on the membrane switch
labeled POWER. Determining that the power is off can be verified by seeing that the liquid-crystal display (LCD) is off.
3.2 Making Electrostatic Voltage Measurements: hold the instrument 1­inch (2.54 cm) away from the object being measured. The display will update with the voltage measurement in kilovolts. If the measured voltage is greater than the measurement range of the instrument, a –1. will be displayed. At this time, switch to
a greater range. If over-ranging occurs even with the 20 kV range activated, the static charge on the object cannot be measured with the 718 Static Sensor.
3.3 Measurement Range: all measurements are in kilovolts (kV) as stated on the front label of the unit. The 718 has two measurement ranges: 0-2 kV and 0-20 kV. The unit’s current measurement range mode can be verified by checking the display. Three digits following the decimal point indicate that the unit is in 0-2 kV range. Two digits following the decimal point indicate that the sensor is in 0-20 kV range. To change between measurement ranges, press the RANGE/HOLD button once, momentarily.
3.4 HOLD Function: in the event that the user wishes to freeze the current measurement, the HOLD function of the 718 Static Sensor may be used. Simply press the HOLD switch momentarily and the currently displayed voltage will be frozen.
Page 5
A HOLD notice will also be displayed to alert the user that the instrument is currently in HOLD status. To unfreeze the display and return to floating measurement, momentarily press the HOLD switch once again. Please note that during HOLD condition the distance­indicating LED’s are turned off.
3.5 Zero adjustment: the 3M™ 718 Static Sensor has a zero adjustment function, which sets a zero reference point for all subsequent measurements. This zero reference can be set by pointing the instrument at a known zero-voltage surface, and holding down the RANGE/ZERO button for longer than 3 seconds. After 3 seconds, the display will flash and adjust to zero. Repeat this step for both the 2 kV and the 20 kV ranges. The zero adjustment should be performed every time the unit is turned on.
3.6 Measurement Accuracy:
3.6.1 Distance Indicator: the 718 Static
Sensor is factory calibrated to give accurate measurements when it is placed one inch (2.54 cm) away from the object to be measured. To assist the user in gauging this distance, two light-emitting diodes (LED’s) are present on the front face of the instrument. These LED’s emit two red, bullseye targets on the surface of the object being measured. As the instrument gets closer to the one inch measurement distance, the bullseyes begin to converge. When they converge and become one, the instrument is approximately one inch away, and the measurement can be made. For more accurate measurements, it is recommended that the user manually measure the distance between the front housing of the instrument and the object being measured.
3.6.2 Accuracy and Size of Object to be Measured: the minimum surface area on an electrostatically charged object which can be accurately measured is a 5 square inch (32.3 cm
2
) area.
3.6.3 Measurements from Greater than One Inch (2.54 cm) Away: in the event that a one inch separation between object-to-be measured and the 718 cannot be achieved, it is possible to get approximate readings. This can be accomplished by holding the instrument at multiples of one inch (two inches, three inches, etc.) away from the object. The measurement reading is then adjusted according to the following equation:
Measurement distance (in inches) x displayed voltage (in kilovolts) = approximate voltage in kilovolts
Page 6
3.7 Continuous Output: an output jack is provided on the front of the 3M™ 718 Static Sensor. This output can be used to feed a continuous signal into a data storage device for continuous monitoring of measured voltages. Please use a 3/32 inch (2.5 mm) mono-phone plug to connect into the output jack. The output is designed to drive a digital voltmeter with an input impedance of greater than 50 kΩ. The output signal is dependent on the measurement range currently selected. For the low (2 kV) range, the output signal is 1/1000 of the measured electrostatic voltage. For the high (20 kV) range, the output signal is 1/10,000 of the measured voltage.
3.8 Automatic Shut-Off: the 718 Static Sensor will automatically shut-off 20 minutes after the last switch activity. This is done in order to conserve battery power. In the event that the user needs to have the unit stay ON continuously, when turning the unit on depress the POWER AND RANGE switches simultaneously. This deactivates the Automatic Shut­Off feature. The BAT indicator will then flash three times to indicate that the automatic shut-off features has been disabled. The Automatic Shut­Off feature will reset itself the next time the instrument is turned on.
4.0 Operation and Use of the 718A Air Ionizer Test Kit
It is recommended that the user be familiar with ionizer test standards ANSI/ESD S3.1 and draft standard ESD DSP3.3 if the 3M™ 718A Air Ionizer Test Kit is used to perform verification testing on ionizer performance.
4.1 Assembly: slide the charge plate over the 718 Static sensor until it stops. The charge plate slides onto the lower groove, on the sides of the 718.
4.2 Charging the plate: holding the 718 Static Sensor (with charge plate attached) in one hand, use the other hand to touch the probe of the 718A Charger to the charge plate. Press either the
+ button (for a positive
voltage) or the
button (for a negative
voltage), then remove the probe from the charge plate. Be sure to keep the button pressed while removing the probe from the charge plate. The display on the 3M™ 718 Static Sensor will indicate a positive or negative charging voltage (1.1kV minimum). If a voltage of less than ±1.1kV is displayed, check to see if the low battery indicator on the 718A Charger is illuminated. If illuminated, replace the battery in the charger. If the unit continues to supply an incorrect voltage to the charge plate, please contact 3M for additional instructions.
Page 7
4.3 Testing ionizer discharge time: after charging the plate, hold the 718 approximately one foot (30.5 cm) away from the ionizer. Monitor the display to see how quickly the 1.1 kV charge is dissipated to 0.1 kV. The speed at which this occurs (the discharge time) indicates how well the ionizer is operating. Please refer to the specific ionizer’s operating manual or consult with the ionizer manufacturer to determine what this discharge time should be. Repeat this procedure for both a positively and a negatively charged plate.
4.4 Testing ionizer offset balance: zero the charge plate by touching it with a grounded object. This can either be the finger of a grounded person or some other item which is connected to electrical ground. In either case, zeroing the charge plate should make the display on the 718 read zero. Hold the 718 approximately one foot (30.5 cm) in front of the ionizer.
Monitor the display. The value displayed is the offset balance of the ionizer, which is the difference between the number of positive and negative ions being emitted. Please refer to the specific ionizer’s operating manual or consult with the ionizer manufacturer to determine what this offset balance should be.
5.0 Service/Calibration
5.1 Service and Repair: in the event that you believe the 718 Static Sensor or the 3M™ 718A Air Ionizer Test Kit is in need of repair, please contact your local 3M representative for troubleshooting help, and, as needed, repair information. There are no user-serviceable parts on either product.
5.2 Calibration: the 718 and 718A products are supplied by the factory pre-calibrated. 3M does not specify a minimum calibration cycle for the 718 or 718A products. The user, usually according to internal Quality procedures, determines calibration cycles. 3M does offer calibration services on these instruments. Please contact your local 3M representative for information on this service. In the event that the user wishes to perform a self-calibration, the following steps should be followed for the 718 Static Sensor (user-calibration not possible on the 718A Air Ionizer Test kit).
5.3 Equipment Needed:
Test fixture High-Voltage Power Supply,
capable of supplying voltages up to 10,000 V Voltmeter, with > 50 kΩ input impedance, capable of measuring voltages down to the µV range. Cable with a 3/32 inch (2.5mm) mono plug and secondary connector to interface with voltmeter
Page 8
5.4 Test fixture:
Metal plate of at least 5 square inches area (38.7 cm
2
) area.
Metal stand capable of supporting 718, and holding it one inch (2.54 cm) away from the metal plate, centered. Connectors on the plate with which it can interface with the high voltage power supply. Connections on the metal stand with which it can be connected to electrical ground.
5.5 Procedure:
5.5.1 Place the 3M™ 718 Static Sensor on the metal stand. Verify that it is exactly one inch (2.54 cm) away from the metal plate, and that its position is centered relative to the plate.
5.5.2 Connect the stand to ground.
5.5.3 Turn on the 718 Static Sensor and set it to the 2 kV range.
5.5.4 Ground the plate. Zero the display.
5.5.5 Remove the ground from the plate, and connect it to the high voltage power supply. Apply a 1 kV charge to the plate.
5.5.6 The instrument should now be reading 1.000. If it is not, remove the battery door and use a screwdriver to turn the small screw located inside the battery compartment. Use a small screwdriver to turn the small screw, located on the front right side of the instrument. This screw should adjust the reading on the display. Once the display has been adjusted to read 1.000, the low range of the 718 is now calibrated.
5.5.7 Repeat procedures 5.5.2 - 5.5.6 for the 20 kV range of the meter, using a test voltage of 5,000 volts.
Page 9
6.0 Physical Characteristics (All values typical)
(at 1 inch (2.54 cm) distance from sensor to target)
3M718 Static Sensor
Dimensions 0.85" (H) x 2.4" (W) x 4.15" (L)
2.2 cm (H) x 6.1 cm (W) x 10.5 cm (L) Weight 4.5 oz. (128 g) with battery Operating Conditions 41º F - 95º F (5º C - 35º C)
up to 80% RH, non-condensing Indoor use only For use at altitudes below 42, 300 ft. (2000 m) Pollution Degree II
Class III Enclosure Conductive Polycarbonate Power Requirements One 9-volt alkaline battery Measurement Ranges 0 – 2 kV
0 – 20 kV Voltage Display 3 1/2 digit liquid crystal display Display Resolution 1 V/inch (0.39 V/cm) @ low range
10 V/inch (3.9 V/cm) @ high range
+/- 10 counts Display Update Rate 3 Hz Voltage Output Jack 3/32 in. (2.5 mm) monophone
Tip: signal
Sleeve: ground Voltage Output 1/1000 of measured voltage @ low range
1/10,000 of measured voltage @ high range Automatic Shut-off 20 minutes after last switch activity Operating Time for Battery Greater than 50 hours, with new battery, @ 21º C
Continuous usage Distance Indicator LED targets. Aligned targets indicate 1 in.
(2.54 cm) measurement distance Measurement Accuracy within 5% of measured voltage Measurement Stability ± 10 counts Certifications UL, C-UL, CE, CB-scheme, NOM
3M718A Air Ionizer Test Kit Charge Plate Assembly Per ESD Association Draft Standard D3.3
Charge Plate Bracket Dimensions 1" (H) x 3.1" (W) x 4.25" (L)
Charge Plate Area 3.25" (W) x 1.25" (L)
Charge Plate Assembly Weight 2.5 oz (70 g) Charger Dimensions 0.85" (H) x 2.4" (W) x 5.0" (L)
Charger Weight 6 oz. (170 g) with battery Charger Power Requirements One 9 volt alkaline battery Charger Output 1.1kV minimum for positive or negative voltage (using 718 with charge plate) Certifications UL, C-UL, CE, CB-scheme, NOM
Aluminum bracket
Bare stainless steel plate
PTFE spacers isolate plate from bracket
2.5 cm (H) x 7.9 cm (W) x 10.8 cm (L)
8.3 cm (W) x 3.2 cm (L)
2.2 cm (H) x 6.1 cm (W) x 12.7 cm (L)
Page 10
7.0 Parts Listing
3M™ 718 Static Sensor
1 ea. 718 Static Sensor 1 ea. 718/718A Operator’s Manual 1 ea. Certificate of Conformance
3M™ 718A Air Ionizer Test Kit
1 ea. 718A Charge Plate Assembly 1 ea. 718A Charger 1 ea. 718/718A Operator’s Manual 1 ea. Certificate of Conformance
8.0 Contact Information:
Customer and Technical Service
Within the U.S. : Customer service and technical support can be obtained by calling the 3M Electronic Handling & Protection Division
Customer Service: (800) 328-1368 Technical Support: (512) 984-3200
Outside of the U.S.: For customer service and technical support, please contact your local representative of the 3M Electronic Handling & Protection Division.
Page 11
Page 12
Electronic and Interconnect Solutions Division
6801 River Place Blvd. Austin, TX 78726-9000 www.3m.com/eisd
3
Important Notice:
All statements, technical information, and recommendations related to 3M’s products are based on information believed to be reliable, but the accuracy or completeness is not guaranteed. Before using this product, you must evaluate it and determine if it is suitable for your intended application.You assume all risks and liability associated with such use. Any statements related to the product which is not contained in 3M’s current publications, or any contrary statements contained on your purchase order shall have no force or effect unless expressly agreed upon, in writing, by an authorized officer of 3M.
Warranty; Limited Remedy; Limited Liability.
This product will be free from defects in material and manufacture for a period of one year from the time of purchase.
3M MAKES NO OTHER WARRANTIES INCLUDING, BUT NOT LIMITED TO, ANY IMPLIED WARRANTY OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE.
If this product is defective within the warranty period stated above, your exclusive remedy shall be, at 3M’s option, to replace or repair the 3M product or refund the purchase price of the 3M product. Except where prohibited by
law, 3M will not be liable for any indirect, special, incidental or consequential loss or damage arising from this 3M product, regardless of the legal theory asserted.
Litho in USA
© 3M 2002
34-7051-6227-8
Recycled paper 40% pre-consumer waste paper 10% post-consumer waste paper
Loading...