3M 245, 247 User Manual

Check the 3M
TM
RangerTM system website to ensure you have the most recent version of this document. www.rangerfluidwarming.com reorder #202524A
3
Ranger
Blood/Fluid Warming Unit,
Model 245
Service Manual
For I.V. Use
3
Ranger
Blood/Fluid Warming Unit,
Service Manual
Not for I.V. Use
1 Model 245 • Blood/Fluid Warming Unit
• For I.V. Use
3 Modell 245 • Blut-/Infusionserwärmungseinheit
•für intravenöse Anwendungen
2 Modèle245 • Appareil de réchauffement de
sang/soluté
• Pour perfusion I.V
1 Model 247 • Irrigation Fluid Warming Unit
• Not for I.V. Use
3 Modell 247 • Spülüssigkeitserwärmungseinheit
• Nicht für intravenöse Anwendungen
2 Modèle247 • Appareil de réchauffement pour
liquide d'irrigation
• Ne pas utiliser pour une perfusion I.V
Service Manual 2
Wartungshandbuch 23
Manuel d`entretien 47
Check the 3M
TM
RangerTM system website to ensure you have the most recent version of this document. www.rangerfluidwarming.com reorder #202524A

Table of Contents

Technical Service and Order Placement ................................................................................................3
In-warranty Repair and Exchange ............................................................................................... 3
When You Call for Technical Support ........................................................................................3
Initial Equipment Check ..........................................................................................................................3
Introduction ............................................................................................................................................... 4
Model 245 Blood/Fluid Warming System ...................................................................................4
Model 247 Irrigation Fluid Warming System ............................................................................4
Indications for Use ..........................................................................................................................4
Denition of Symbols ..................................................................................................................... 4
Explanation of Signal Word Consequences ................................................................................ 6
Warnings and Cautions ................................................................................................................. 6
Before Servicing Equipment ..........................................................................................................7
Product Description .................................................................................................................................8
Ranger Warming Unit ...................................................................................................................8
Ranger Models 245 and 247 ....................................................................................................................9
Maintenance and Storage ......................................................................................................................10
Blood/Fluid Warming System
Preventative Maintenance Checklist ..................................................................................................12
Ranger Blood/Fluid Warming Unit, Model 245 Maintenance Log ................................................15
Ranger Irrigation Fluid Warming System, Model 247 Preventative Maintenance Checklist .....16
Ranger Irrigation Fluid Warming, Model 247 Maintenance Log ................................................... 18
Replacing the Power Fuse ......................................................................................................................19
Troubleshooting ......................................................................................................................................20
Specications ...........................................................................................................................................22
Check the 3M
TM
RangerTM system website to ensure you have the most recent version of this document. www.rangerfluidwarming.com reorder #202524ACheck the 3M
TM
3M
RangerTM Blood/Fluid and Irrigation Warming Systems

Technical Service and Order Placement

USA
TEL: +1-952-947-1200 +1-800-733-7775 FAX: +1-952-947-1400
Outside of the USA
Contact your local 3M Patient Warming representative.

In-warranty repaIr and exchange

To return a device to 3M Patient Warming for service, rst obtain a Return Authorization (RA) number from a customer service representative. Please use the (RA) number on all correspondence when returning a device for service. A shipping carton will be delivered to you at no charge, if needed. We will service and ship your device within ve (5) working days of our receipt. Call your local supplier or sales representative to inquire about a loaner device while your device is being serviced.

when you call for technIcal Support

Remember, we will need to know the serial number of your unit when you call us. e serial number label is located on the bottom of the warming unit.
English

Initial Equipment Check

• Make sure the 3M
• When checking for grounding, clamp to the screw threads on the unit’s I.V. pole clamp or the
equipotential stud on the rear of the unit. Clamp should be tightened securely.
• Sometime during shipping, this unit may have been exposed to extreme temperatures. e unit may need a period of 20 minutes or more to return to normal operating temperature. Store the Ranger warming unit at room temperature, allowing it to equilibrate, before using or testing the unit.
TM
RangerTM warming unit’s power cord is plugged into the unit.
202524A
3
Check the 3M
TM
RangerTM system website to ensure you have the most recent version of this document. www.rangerfluidwarming.com reorder #202524A
English
TM
3M
RangerTM Blood/Fluid and Irrigation Warming Systems

Introduction

e Ranger warming system includes a warming unit and a disposable warming set. e Ranger warming unit is designed to be mounted to an I.V. pole. A handle located on the top of the unit makes transport easy. When mounted to the I.V. pole, the unit ts easily above the 3MTM Bair HuggerTM 500 or 700 series warming unit. For additional information on Ranger warming sets visit us online at rangeruidwarming.com.

Model 245 Blood/fluId warMIng SySteM

e Ranger blood/uid warming system is designed to warm blood, blood products, and liquids and deliver these at ow rates from KVO to 500 mL/min. At these ow rates, the device maintains uid output temperatures ranging from 33°C to 41°C (Note: is is for room temperature uids only). It takes less than two minutes to warm up to the 41°C ±1 set point temperature. e alert points on the Model 245 are 43°C ±1 and 46°C [+3, -2°C].
Disposable warming sets are available for standard ow, high ow, and pediatric applications. Warming sets are sterile, latex-free, single-use-only items and are designed to be used with the reusable warming unit. Do not use other warming sets with the Model 245, including those designed for the Model 247 irrigation uid warming system.

Model 247 IrrIgatIon fluId warMIng SySteM

e Ranger irrigation uid warming system is designed to warm irrigation uids and deliver these at ow rates from 0 to 865 mL/min when the bag is hung 100 cm above the scope. At these ow rates, the device maintains uid output temperatures ranging from 33°C to 41°C (Note: is is for room temperature uids only). It takes less than two minutes to warm up to the 41°C ±1 set point temperature. e alert points on the Model 247 are 48°C (±2) and 50°C (±2).
e Ranger irrigation disposable uid warming set is sterile, latex-free, and is designed to be used with the Ranger irrigation uid warming unit. Do not use other warming sets with the Model 247, including those designed for the Model 245 blood/uid warming unit.

IndIcatIonS for uSe

Blood/Fluid Warming SyStem
e Ranger blood/uid warming system is intended to warm blood, blood products, and liquids.
irrigation Fluid Warming SyStem
e Ranger irrigation uid warming system is intended to warm irrigation uids.

defInItIon of SyMBolS

ON
OFF
Fuse
CAU T ION
Type B Applied Part
4
Check the 3M
TM
RangerTM system website to ensure you have the most recent version of this document. www.rangerfluidwarming.com reorder #202524ACheck the 3M
TM
3M
RangerTM Blood/Fluid and Irrigation Warming Systems
Voltage, Alternating Current (AC)
An equipotentiality plug (grounded) conductor other than a protective earth
conductor or a neutral conductor, providing a direct connection between electrical equipment and the potential equalization busbar of the electrical installation. Please consult IEC 6060-1; 2005 for requirements.
is system is subject to European WEEE Directive 2002/96/EC.
is product contains electrical and electronic components and must not be disposed of using standard refuse collection. Please consult local directives for disposal of electrical and electronic equipment.
Protective earth ground
Consult instructions for use
Follow instructions for use
English
Recycle to avoid environmental contamination. is product contains recyclable
parts. For information on recycling - please contact your nearest 3M Service Center for advice.
Date of manufacture
Manufacturer
Sterile, ethylene oxide
Store at room temperature
DEHP-free
Container quantity
Single use
202524A
Latex free
Keep dry
Temperature limits
5
Check the 3M
TM
RangerTM system website to ensure you have the most recent version of this document. www.rangerfluidwarming.com reorder #202524A
English
TM
3M
RangerTM Blood/Fluid and Irrigation Warming Systems

explanatIon of SIgnal word conSequenceS

WARNING:
Indicates a hazardous situation which, if not avoided, could result in death or serious injury.
CAUTION:
Indicates a hazardous situation which, if not avoided, could result in minor or moderate injury.
NOTICE:
Indicates a situation which, if not avoided, could result in property damage only.

warnIngS and cautIonS

WARNINGS:
1. To reduce the risks associated with hazardous voltage and re and thermal energy hazards:
• Do not substitute other devices (i.e. Model 245) for the Ranger irrigation uid warming unit or Ranger irrigation warming sets.
2. To reduce the risks associated with hazardous voltage and re:
• Do not modify this device.
• Connect power cord to receptacles marked “Hospital Only,” “Hospital Grade,” or a reliably
grounded outlet.
• Use only the power cord specied for the product and certied for the country of use.
• Do not allow the power cord to get wet.
• Do not use the warming unit when it appears the Ranger warming unit, power cord, or
warming set is damaged. Use only 3M specied replacement parts.
• Keep power cord visible and accessible at all times. e plug on the power cord serves as the disconnect device. e wall socket outlet shall be as close as practical and shall be easily accessible.
CAUTION:
1. To reduce the risks associated with cross-contamination:
• e cleaning tool provides only supercial cleaning, it does not disinfect or sterilize the interior of the unit.
2. To reduce the risks associated with impact and facility medical device damage:
• Clamp the Ranger warming unit to an I.V. pole with a minimum 14" (35.6 cm) radius wheelbase and at a height no higher than 44" (112 cm).
3. To reduce the risks associated with environmental contamination:
• Follow applicable regulations when disposing of this device or any of its electronic components.
4. Do not use the Ranger blood/uid warming system 245 Model for direct cardiac application. When using the Model 245 with a central venous catheter (CVC), ensure that the catheter tip does not have direct contact with the heart and ensure that all electrical devices connected to or near the patient have an appropriate leakage current rating for the application. If a CVC tip is found to be in direct contact with a patient’s heart, the Model 245 should be immediately disconnected from the CVC until the CVC is safely repositioned. Failure to follow these precautions may cause cardiac disturbance and/or patient injury.
6
Check the 3M
TM
RangerTM system website to ensure you have the most recent version of this document. www.rangerfluidwarming.com reorder #202524ACheck the 3M
TM
3M
RangerTM Blood/Fluid and Irrigation Warming Systems
NOTICE:
1. Federal law (USA) restricts this device to sale by or on the order of a licensed healthcare professional.
2. To avoid device damage:
• Do not clean the warming unit with solvents. Damage to the case, label, and internal components may result.
• Do not immerse the warming unit in cleaning or sterilizing solutions. e unit is not liquid proof.
• Do not insert metallic instruments in the warming unit.
• Do not use abrasive materials or solutions to clean the heater plates.
• Do not allow spills to dry inside the unit, as this may make it more dicult to clean the unit.
3. e Ranger blood/uid warming unit and irrigation uid warming unit meet medical electronic interference requirements. If radio frequency interference with other equipment should occur, connect the unit to a dierent power source.
English

Before ServIcIng equIpMent

All repair, calibration, and servicing of this equipment must be performed by 3M Patient Warming or an authorized service technician. 3M Patient Warming assumes no responsibility for the reliability, performance, or safety of the equipment if:
• Modications or repairs are performed by unauthorized personnel.
• e equipment is used in a manner other than that described in the operator’s manual.
• e equipment is installed in an environment that does not meet the appropriate electrical and
grounding requirements.
• Do not modify this equipment without authorization of the manufacturer.
WARNING:
• To reduce the risk associated with exposure to biohazards always perform the Cleaning the Ranger
warming unit procedure on page 10 prior to beginning service procedures, returning the warming
unit for service, and prior to disposal.
202524A
7
Check the 3M
ALERT
TM
RangerTM system website to ensure you have the most recent version of this document. www.rangerfluidwarming.com reorder #202524A
English
TM
3M
RangerTM Blood/Fluid and Irrigation Warming Systems

Product Description

e Ranger blood/uid, and irrigation uid warming systems consist of a warming unit and a sterile disposable uid warming set.
Slot for warming cassette
Alphanumeric display
Alert indicator light
model 245
I.V. pole clamp
Bubble trap holder
Power switch
Slot for warming cassette
Alphanumeric display
Alert indicator light
ALERT
model 247
I.V. pole clamp
Power switch

ranger warMIng unIt

e warming unit is a compact, lightweight, liquid-resistant device with a clamp located on the side for attachment to an I.V. pole. A carrying handle on the top of the unit makes it easy to transport.
Located on the front panel you will nd:
• Alphanumeric display that indicates the heater temperature during normal operation. In an over-temperature condition, the display alternately ashes a temperature of 43°C for the Model 245 (48°C for the Model 247) or higher and the word “HI.” An audible alert also sounds. In an under­temperature condition, the display alternately ashes a temperature of 33°C or lower and the word “LO.”
• Alert indicator light that comes on when either an over- or under-temperature condition occurs.
8
Check the 3M
TM
RangerTM system website to ensure you have the most recent version of this document. www.rangerfluidwarming.com reorder #202524ACheck the 3M

Ranger Models 245 and 247

17.
7.
8.
TM
3M
RangerTM Blood/Fluid and Irrigation Warming Systems
18.
1.
3.
5.
11.
English
6.
4.
16.
21.
1. Upper Enclosure
2. Lower Enclosure
3. Upper Heater
15.
2.
8. ermistor Sensor
9. Pole Clamp
10. Pole Clamp Knob
19.
13.
9.
10.
20.
12.
14.
15. Mic ro s w it c h
16. Ala rm Board
17. Controller
202524A
4. Lower Heater
5. Upper Heater Plate
6. Lower Heater Plate
7. RTD Sensor
11. Bubble Trap Holder (Model 245)
12. Power Entry Module
13. Cord Retai ner
14. Solid State Relay
18. Ha nd le
19. Cable Cla mp
20. Ground Plane
21. Heat Sin k
9
Check the 3M
TM
RangerTM system website to ensure you have the most recent version of this document. www.rangerfluidwarming.com reorder #202524A
English
TM
3M
RangerTM Blood/Fluid and Irrigation Warming Systems

Maintenance and Storage

CleAning the RAngeR WARming Unit
Clean the Ranger warming unit on an as-needed basis.
CAUTION
1. To reduce the risks associated with cross-contamination:
• e cleaning tool provides only supercial cleaning, it does not disinfect or sterilize the interior of the unit.
NOTICE
1. To avoid device damage:
• Do not immerse the warming unit in cleaning or sterilizing solutions. e unit is not liquid proof.
• Do not clean the warming unit with solvents. Damage to the case, label, and internal components may result.
• Do not insert metallic instruments in the warming unit.
• Do not use abrasive materials or solutions to clean the heater plates.
• Do not allow spills to dry inside the unit, as this may make it more dicult to clean the unit.
to CleAn the exteRioR of the WARming Unit:
1. Disconnect the Ranger warming unit from the power source.
2. Wipe the outside of the unit with warm, soapy water, nonabrasive cleaning solutions, dilute bleach, or cold sterilants. Do not use abrasive materials.
3. Wipe with a dry, so cloth.
to CleAn the heAting plAteS:
e Ranger warming unit cleaning tool is intended to clean both heating plates of the warming unit. It is not necessary to disassemble the warming unit to use the tool.
10
Check the 3M
TM
RangerTM system website to ensure you have the most recent version of this document. www.rangerfluidwarming.com reorder #202524ACheck the 3M
TM
3M
RangerTM Blood/Fluid and Irrigation Warming Systems
method
1. Unplug the warming unit.
2. Unfold the cleaning tool. Wet the foam pads with a nonabrasive solution, such as AlconoxTM brand detergent.
3. Insert the tool from the back of the unit and pull the tool all the way out from the front.
4. Rinse the tool with water and repeat 3 times. Discard the tool according to institutional protocol.
5. Wipe o the unit to remove excess uid.
to clean reSiStant, dried-on FluidS:
1. Spray a nonabrasive solution inside the slot of the warming unit and let sit for 15-20 minutes.
2. Clean the unit by using the cleaning tool.
NOTE: You may use a nonmetal instrument, such as a cotton swab, to clean the upper channels. If you are unable to adequately clean the unit, call 3M Patient Warming technical support.
English
StoRAge
Store all components in a cool, dry place when not in use.
Call 3M Patient Warming technical support at 800-733-7775 or 952-947-1200 for service information.
Alconox is a trademark of Alconox, Inc.
202524A
11
Check the 3M
TM
RangerTM system website to ensure you have the most recent version of this document. www.rangerfluidwarming.com reorder #202524A
English
Model 245
TM
3M
RangerTM Blood/Fluid and Irrigation Warming Systems

Blood/Fluid Warming System Preventative Maintenance Checklist

caution
Temperature adjustments cannot be made to the Ranger warming unit. If the test shows the temperature is out of specication, please call 3M Patient Warming technical support at 1-800-733-7775.
toolS and equipment
• ermocouple: 0.005" lead or smaller (part #90020)
• Calibrated thermocouple temperature monitor
• Aluminum tape
• Ranger standard ow blood/uid warming set
• Phillips screwdriver
• Saline/water
Inlet tube
Insert cassette
before priming.
from cassette to remove.
Drain 2mL or more
Aluminum tape
0.75"
1.80"
Thermocouple
Fig. 1: Ranger standard ow blood/uid warming cassette
12
Check the 3M
TM
RangerTM system website to ensure you have the most recent version of this document. www.rangerfluidwarming.com reorder #202524ACheck the 3M
3M
1. Use aluminum tape to ax a thermocouple
to the top surface of the Ranger warming as shown in Figure 1.
2. Slide the cassette into the slot of the
Ranger warming unit.
3. Using room temperature saline/water,
prime the cassette per instructions included with it.
note: To achieve correct results, uids
must be administered into the inlet tube, NOT into the outlet tube.
4. Close patient line with white clamp to
stop ow.
5. Turn the Ranger warming unit ON and
allow it to warm up for 5 minutes.
6. e Ranger warming unit display should
read 41°C±1. Note the temperature on the thermocouple temperature monitor. If the temperature reading is not 41°C±1, call 3M Patient Warming technical support.
TM
RangerTM Blood/Fluid and Irrigation Warming Systems
Cassette
Blue
clamp
Inlet tube
Fluid
source
Bubble
trap
Roller clamp
Outlet
tube
White clamp
Exit
English
Model 245
note: Most handheld temperature meters
have a ±1°C tolerance.
Fig. 2: Ranger standard ow blood/uid warming set
202524A
13
Check the 3M
TM
RangerTM system website to ensure you have the most recent version of this document. www.rangerfluidwarming.com reorder #202524A
English
Model 245
TM
3M
RangerTM Blood/Fluid and Irrigation Warming Systems
teSting the oveR-tempeRAtURe AleRt pointS (43°C [±1] And 46°C [+3, -2])
e over-temperature alert test causes the Ranger warming unit to go into an over-temperature test mode by manually overriding the primary control system and engaging the heaters. e Ranger warming system is very responsive to heater input; therefore the test procedures are technique sensitive. Read instructions thoroughly before beginning these tests.
e rst alert point at 43°C (±1) noties user to the rise in temperature (see step 7). At the 46°C (+3, -2) secondary alert set point, the unit cuts power to the heaters (see step 8). Please note that due to the heaters’ rapid response, you may see temperature readings dri within the range of 44­49°C.
7. Loosen the over-temperature screw on the bottom of the Ranger warming unit (see Fig. 3). Listen for a slight click as the microswitch opens. Loosen ½ turn more. is unit is now in over-temperature test mode, and the temperature output of the unit should begin to rise. When the Ranger display reads 43°C, verify that an audible alert sounds, the alert light illuminates, and the display alternately ashes “HI” and the plate temperature.
Fig. 3: Ranger warming unit over-temperature screw
Over-temperature
screw
8. Listen for the power relay to click at 46°C (+3, -2) (you can feel the click when holding the lower le corner of the unit as you face the front display). e click signies the secondary alert trip point and signals that power to the heaters has been shut o. e over-temperature alert sounds, the alarm light illuminates, and the alphanumeric display alternately reads “HI” and the plate temperature. Watch the thermocouple temperature monitor to determine the unit’s peak temperature output (less than 1 minute). e temperature reading on the monitor should peak at 46°C (+3, -2).
note: If the temperature displayed on the thermocouple monitor exceeds 49°C during the test, tighten
the over-temperature test screw on the bottom of the Ranger warming unit, unplug it, and run cold uids through the cassette. is returns the unit to normal operating temperature. Call 3M Patient Warming technical support.
9. e secondary alert is designed to latch. Verify this feature by momentarily turning o the power switch, then turning it on again. e Ranger warming unit should continue alerting throughout this sequence.
10. Securely tighten the over-temperature screw, turn o the Ranger warming unit, and unplug the unit. Once power is cut to the Ranger warming unit, the alert will cease. e unit is ready to return to service aer it cools to normal operating temperature (approximately 20-30 minutes).
14
Check the 3M
TM
RangerTM system website to ensure you have the most recent version of this document. www.rangerfluidwarming.com reorder #202524ACheck the 3M
TM
3M
RangerTM Blood/Fluid and Irrigation Warming Systems

Ranger Blood/Fluid Warming Unit, Model 245 Maintenance Log

The operating temperature and alerts should be tested at least every 6 months or according to institutional protocol.
mAintenAnCe ACtion peRfoRmed model ___________ Sn ________
Date _________ Operating temperature: (Specication = 41°C±1) ______°C
Over-temperature alert points:
Audible (43°C [±1] on Ranger unit display) ______°C
Secondary (46°C [+3, -2]) ______°C
Date _________ Operating temperature: (Specication = 41°C±1) ______°C
Over-temperature alert points:
English
Model 245
Audible (43°C [±1] on Ranger unit display) ______°C
Secondary (46°C [+3, -2]) ______°C
Date _________ Operating temperature: (Specication = 41°C±1) ______°C
Over-temperature alert points:
Audible (43°C [±1] on Ranger unit display) ______°C
Secondary (46°C [+3, -2]) ______°C
Date _________ Operating temperature: (Specication = 41°C±1) ______°C
Over-temperature alert points:
Audible (43°C [±1] on Ranger unit display) ______°C
Secondary (46°C [+3, -2]) ______°C
Date _________ Operating temperature: (Specication = 41°C±1) ______°C
Over-temperature alert points:
Audible (43°C [±1] on Ranger unit display) ______°C
Secondary (46°C [+3, -2]) ______°C
202524A
Date _________ Operating temperature: (Specication = 41°C±1) ______°C
Over-temperature alert points:
Audible (43°C [±1] on Ranger unit display) ______°C
Secondary (46°C [+3, -2]) ______°C
15
Check the 3M
TM
RangerTM system website to ensure you have the most recent version of this document. www.rangerfluidwarming.com reorder #202524A
English
TM
3M
RangerTM Blood/Fluid and Irrigation Warming Systems

Ranger Irrigation Fluid Warming System, Model 247 Preventative Maintenance Checklist

caution
Temperature adjustments cannot be made to the Ranger irrigation uid warming unit. If the test shows the temperature is out of specication, please call 3M Patient Warming technical support at 1-800-733-7775.
toolS and equipment
• ermocouple: 0.005" lead or smaller (part #90020)
• Calibrated thermocouple temperature monitor
• Aluminum tape
• Ranger irrigation uid warming set
• Phillips screwdriver
• Saline/water
1. Use aluminum tape to ax a thermocouple to the top surface of the Ranger irrigation uid warming cassette as shown in Fig. 1.
Inlet tube
before priming.
Insert cassette
Irrigation Fluid Warming
Model 247
from cassette to remove.
Drain 2mL or more
Aluminum tape
0.75"
Fig. 1: Ranger irrigation uid warming cassette
Thermocouple
8.50"
Model 247
2. Slide the warming cassette into the slot of the Ranger irrigation uid warming unit.
3. Using room temperature saline/ water, prime the cassette per instructions included with it.
note: To achieve correct results,
uids must be administered into the inlet tube, NOT into the outlet tube.
4. Close patient line with white clamp to stop ow.
5. Turn the Ranger irrigation warming unit ON and allow it to warm up for 5 minutes.
6. e Ranger irrigation warming unit display should read 41°C±1. Note the temperature on the thermocouple temperature monitor. If the temperature reading is not 41°C±1, call 3M Patient Warming technical support.
note: Most handheld temperature
meters have a ±1°C tolerance.
Fluid source
Inlet tube
Blue clamp
Fig. 2: Ranger irrigation uid warming set
Cassette
White clamp
Outlet tube
Exit
16
Check the 3M
TM
RangerTM system website to ensure you have the most recent version of this document. www.rangerfluidwarming.com reorder #202524ACheck the 3M
TM
3M
RangerTM Blood/Fluid and Irrigation Warming Systems
teSting the oveR-tempeRAtURe AleRt pointS (48°C [±2] And 50°C [±2])
e over-temperature alert test causes the Ranger irrigation warming unit to go into an over-temperature test mode by manually overriding the primary control system and engaging the heaters. e Ranger irrigation warming system is very responsive to heater input; therefore the test procedures are technique sensitive. Read instructions thoroughly before beginning these tests.
e rst alert point at 48°C (±2) noties user to the rise in temperature (see step 7). At the 50°C (±2) secondary alert set point, the unit cuts power to the heaters (see step 8).
7. Loosen the over-temperature screw on the bottom of the Ranger irrigation unit (see Fig. 3). Listen for a slight click as the microswitch opens. Loosen ½ turn more. is unit is now in over-temperature test mode, and the temperature output of the unit should begin to rise. When the display reads 48°C, verify that an audible alert sounds, the alert light illuminates, and the display alternately ashes “HI” and the plate temperature.
Fig. 3: Ranger irrigation uid warming unit over-temperature screw
Over-temperature
screw
English
Model 247
8. Listen for the power relay to click at 50°C (±2) (you can feel the click when holding the lower le corner of the unit as you face the front display). e click signies the secondary alert trip point and signals that power to the heaters has been shut o. e over-temperature alert sounds, the alert light illuminates, and the alphanumeric display alternately reads “HI” and the plate temperature. Watch the thermocouple temperature monitor to determine the unit’s peak temperature output (less than 1 minute). e temperature reading on the monitor should peak at 50°C (±2).
note: If the temperature displayed on the thermocouple monitor exceeds 52°C during the test, tighten
the over-temperature test screw on the bottom of the Ranger irrigation warming unit, unplug it, and run cold uids through the cassette. is returns the unit to normal operating temperature. Call 3M Patient Warming technical support.
9. e secondary alert is designed to latch. Verify this feature by momentarily turning o the power switch, then turning it on again. e Ranger irrigation warming unit should continue alerting throughout this sequence.
10. Securely tighten the over-temperature screw, turn o the Ranger irrigation warming unit, and unplug the unit. Once power is cut to the Ranger irrigation warming unit, the alert will cease. e unit is ready to return to service aer it cools to normal operating temperature (approximately 20-30 mi nutes).
202524A
17
Check the 3M
TM
RangerTM system website to ensure you have the most recent version of this document. www.rangerfluidwarming.com reorder #202524A
English
TM
3M
RangerTM Blood/Fluid and Irrigation Warming Systems

Ranger Irrigation Fluid Warming, Model 247 Maintenance Log

The operating temperature and alerts should be tested at least every 6 months or according to institutional protocol.
mAintenAnCe ACtion peRfoRmed model ___________ Sn _______
Date ________ Operating temperature: (Specication = 41°C±1) ______°C
Over-temperature alert points:
Audible (48°C [±2] on Ranger unit display) ______°C
Secondary (50°C [±2]) ______°C
Date ________ Operating temperature: (Specication = 41°C±1) ______°C
Over-temperature alert points:
Model 247
Audible (48°C [±2] on Ranger unit display) ______°C
Secondary (50°C [±2]) ______°C
Date ________ Operating temperature: (Specication = 41°C±1) ______°C
Over-temperature alert points:
Audible (48°C [±2] on Ranger unit display) ______°C
Secondary (50°C [±2]) ______°C
Date ________ Operating temperature: (Specication = 41°C±1) ______°C
Over-temperature alert points:
Audible (48°C [±2] on Ranger unit display) ______°C
Secondary (50°C [±2]) ______°C
Date ________ Operating temperature: (Specication = 41°C±1) ______°C
Over-temperature alert points:
Audible (48°C [±2] on Ranger unit display) ______°C
Secondary (50°C [±2]) ______°C
Date ________ Operating temperature: (Specication = 41°C±1) ______°C
Over-temperature alert points:
Audible (48°C [±2] on Ranger unit display) ______°C
Secondary (50°C [±2]) ______°C
18
Check the 3M
TM
RangerTM system website to ensure you have the most recent version of this document. www.rangerfluidwarming.com reorder #202524ACheck the 3M
TM
3M
RangerTM Blood/Fluid and Irrigation Warming Systems

Replacing the Power Fuse

e power fuses are located in the power entry module.
toolS And eqUipment
Small slotted screwdriver
method
1. Disconnect the warming unit from the power source.
2. Remove the power cord.
3. Locate the fuse carrier in the power entry module.
4. Use the small screwdriver to remove the fuse carrier from the power entry module.
5. Remove the blown fuse from the fuse carrier and replace with a new fuse (as marked).
English
6. Place the fuse carrier back into the power entry module.
7. Reattach the power cord.
8. Reconnect the warming unit to the power source.
9. Turn the unit on to verify normal operation.
10. Record the maintenance action taken.
202524A
19
Check the 3M
TM
RangerTM system website to ensure you have the most recent version of this document. www.rangerfluidwarming.com reorder #202524A
English
TM
3M
RangerTM Blood/Fluid and Irrigation Warming Systems

Troubleshooting

CONDITION CAUSE SOLUTION
Nothing illuminates on the warming unit panel.
CONDITION CAUSE SOLUTION
Alert indicator illuminates and alert sounds, alphanumeric
display alternately ashes a temperature of 43˚C for the Model 245 (48°C for the Model 247) or higher and the word “HI.”
•Unitisnotturnedon,plugged
in, or power cord is not plugged into an appropriate outlet.
•Unitfailure.
Temporary over-temperature condition because:
•Anextremechangeinowrates
occurred (e.g., from 500 mL/min to stop ow).
•Turnuniton.Makesurethepower
cord is plugged into the power entry module of the warming unit. Make sure the warming unit is plugged into a properly grounded outlet.
•Checkfuses.
•Call3MPatientWarmingtechnical
support.
•Open ow to reduce temperature. Alerts will stop when the display reads 41°C. e unit is ready to use.
•Unitwasturnedonandreached
set point temperature before
•Alertswillstopwhenthedisplay
reads 41°C. e unit is ready to use.
cassette was inserted.
•Fluidswereprewarmedtoabove
42°C before being run through the warming unit.
•Turnounit.Discontinueinfusion
of uids. Do not warm uids before infusing them through the Ranger warming unit.
CONDITION CAUSE SOLUTION
Alert sounds, alphanumeric display and alert indicator light
go dark.
Primary controller failure. Unit will no longer operate.
Power to heating plates will shut o if temperature rises to 44°C for the Model 245 (50°C for the Model
247). Turn unit o and unplug it.
Discontinue use of unit. Discard warming set. Alert will continue
to sound if you do not unplug unit. Call 3M Patient Warming technical support.
20
Check the 3M
TM
RangerTM system website to ensure you have the most recent version of this document. www.rangerfluidwarming.com reorder #202524ACheck the 3M
TM
3M
RangerTM Blood/Fluid and Irrigation Warming Systems
CONDITION
•Unitalerts soon aer plugging it in (unit does not have to be turned on for this condition to occur).
CAUSE
Over-temperature screw on bottom of unit is loose or missing.
SOLUTION
Make sure over-temperature screw is completely tightened. If it is missing, turn unit o and unplug it. Call 3M Patient Warming technical support.
•Heatertemperaturerisesto44°Cfor
Model 245 (50°C for Model 247) and unit shuts down soon aer plugging it in (unit does not have to be turned on for this condition to occur).
CONDITION CAUSE SOLUTION
Alert sounds but unit has been turned o.
•Independentbackupsafety
system has been activated.
•Over-temperaturescrew
loose.
•Unplugunit.Call3MPatient
Warming technical support.
•Makesureover-temperaturescrewis
completely tightened. If it is missing, turn unit o and unplug it. Call 3M Patient Warming technical support.
English
CONDITION CAUSE SOLUTION
Cannot remove cassette from unit. •Cassetteistoofull,uids
are still being infused, or clamp is open proximal to the cassette.
•Makesureuidisdrainedfrom
cassette before sliding out the cassette, that uids are no longer infusing, and that clamp is closed proximal to the warming cassette.
•Warmingunitisbelow
•Raiseunitabovepatientlevel.
patient level, creating excessive back pressure.
CONDITION CAUSE SOLUTION
Alert indicator light illuminates and alert sounds, alphanumeric display
alternately ashes a temperature of 33°C or below and the word “LO.”
CONDITION CAUSE SOLUTION
Alphanumeric display reads “Er 4” or “Open.”
CONDITION CAUSE SOLUTION
Under-temperature condition caused by very high ow using very cold uid, or defective heater/ relay.
Open wire on temperature sensor.
Alert should stop when temperature rises above 33°C. If alert continues,
turn o unit, unplug unit and discontinue use. Call 3M Patient Warming technical support.
Do not use unit. Call 3M Patient Warming technical support.
202524A
Alphanumeric display reads “Er 5” or “Open.”
Electrical interference. Remove the unit. Refer to biomedical
technician or call 3M Patient Warming technical support.
21
Check the 3M
TM
RangerTM system website to ensure you have the most recent version of this document. www.rangerfluidwarming.com reorder #202524A
English
TM
3M
RangerTM Blood/Fluid and Irrigation Warming Systems

Specifications

phySiCAl ChARACteRiStiCS
Warming unit
4.5 in. (11 cm) high x 7.5 in. (19 cm) wide x 10 in. (25 cm) long; wt.: 7 lb. 7 oz. (3.4 kg)
certiFicationS
IEC/EN 60601-1; CAN/CSA-C22.2, No. 601.1
claSSiFication
MEDICAL — GENERAL MEDICAL EQUIPMENT AS TO ELECTRICAL SHOCK, FIRE AND MECHANICAL HAZARDS ONLY IN ACCORDANCE WITH UL 60601-1; CAN/CSA-C22.2, No.601.1; ANSI/AAMI ES60601-1:2005 CSA-C22.2 No. 60601-1:08; Control No. 4HZ8
Classied under IEC 60601-1 Guidelines (and other national versions of the Guidelines) as Class I, Type B, Ordinary equipment, Continuous operation. Classied by Underwriters Laboratories Inc. with respect to electric shock, re and mechanical hazards only, in accordance with IEC/EN2 60601-1 and in accordance with Canadian/CSA C22.2, No. 601.1. Classied under the Medical Device Directive as a Class IIb device.
eleCtRiCAl ChARACteRiStiCS tempeRAtURe ChARACteRiStiCS
device rating
100-120 VAC, 50/60 Hz 220-240 VAC, 50/60 Hz
maximum heating poWer
900 W
Set point temperature
41°C
over-temperature alert
Model 245: 43°C
Model 247: 48°C
FuSe
2 x T10A-H (250V) for 100-120 VAC
under-temperature alert
33°C
2 x T6.3A-H (250V) for 220-240 VAC
FuSe type
Time delay, high breaking
over-temperature cutoFF
Model 245: 44°C (warming units Rev N and newer) 46°C (warming units Rev A to M)
Model 247: 50°C
leAkAge CURRent
Meets leakage current requirements in accordance with UL/IEC 60601-1.
enviRonmentAl ConditionS
operating temperature range
15° to 40°C (59° to 104°F)
Storage temperature range
-20° to 45°C (-4° to 113°F)
operating humidity
10 to 85% RH, noncondensing
atmoSpheric preSSure range
50 kPa to 106 kPa
22
Made in the USA by 3M Health Care.
3M is a trademark of 3M Company, used under license in Canada. RANGER, BAIR HUGGER, and the BAIR HUGGER logo are trademarks of Arizant Healthcare Inc., used under license in Canada. ©2013 Arizant Healthcare Inc. All rights reserved.
3M Deutschland GmbH, Health Care Business Carl-Schurz-Str. 1, 41453 Neuss, Germany
3M Health Care, 2510 Conway Ave., St. Paul, MN 55144 USA
TEL 800-228-3957 | www.rangeruidwarming.com
202524A 07/13
Loading...