3B Scientific ESCOLA2 User Manual

3B SCIENTIFIC
Multimeter Escola 2 1006811
01/13 ALF
®
PHYSICS
1 Meter display with mirror scale
2 Slotted screw for zero calibra-
tion
3 Adjustment knob for zero cali-
bration
4 Measurement range dial
5 Operating mode switch
6 Safety measurement socket
7 Safety ground socket
1. Safety instructions
The Escola 2 multimeter conforms to the safety requirements for electrical equipment for meas­urement, control and laboratory use in DIN EN 61010 part 1. Safe operation of the apparatus is guaranteed with correct handling. However, safety is not guaranteed if the apparatus is han­dled improperly or carelessly.
Read this manual carefully before using the
multimeter and follow the instructions!
Before using the meter, check the case and
test leads for any damage. In the event of any malfunction/operational defect or visible damage, do not use the meter. Pay particu­lar attention to the insulation surrounding the measurement sockets.
The limit of the measurement range must
not be exceeded. If the values of the meas­urand are unknown, always switch from a
higher measurement range to a lower one.
Do not conduct measurements in a humid
environment. Work area, hands, shoes and floor must be dry.
When measuring current, turn off circuit
power before connecting the meter in the circuit.
Connect the common test lead before you
connect the live test lead. When you discon­nect test leads, disconnect the live test lead first.
Before the case is opened, the meter has to
be switched off and the leads must be dis­connected from the meter.
When disposing empty batteries follow the
local regulations. Never dispose of them in the regular household garbage.
1
2. Symbol legend
Read instruction sheet
V Voltage A Current
Moving coil galvanometer
Apparatus with electronic amplifier
DC quantities accuracy class 2
AC quantities accuracy class 3
Use in horizontal position
2
Test voltage
Use in vertical position
DC quantities
AC quantities
Needle position zero centre
“OFF” position
3. Description
The Escola 2 is a meter for use by students to measure to measure voltages and currents (both DC and AC quantities) in low voltage ranges.
All measurement ranges are selected by means of a rotary dial. Two linearised mirror scales, graded in a 1:3 ratio, guarantee easy readability of the measured values.
The trimmer for setting the zero point in the centre allows the zero point to be set precisely when it drifts electrically.
AC quantities are calibrated for a sinusoidal waveform. Conversion and linearisation of the measurement is achieved by half-wave rectifica­tion (positive half-wave) using an op-amp.
The use of a robust moving coil galvanometer and a mechanically resilient chassis also allows for rough operating conditions.
The overcurrent protection of the Escola 2 automatically limits the power when overloaded. After a brief cooling phase, the apparatus auto­matically switches on again.
4. Technical data
Voltage ranges: 0.3; 1; 3; 10; 30 V AC/DC Source impedance: 10 kOhm/V Current ranges: 1; 10; 100; 1000;
3000 mA AC/DC
Voltage drop for current measurements: 100 mV AC/DC approx.
Accuracy: DC class 2 AC class 3
Electrical zerocalibration: For all DC ranges Accuracy for zero centre: Class 5 Scale length: 80 mm Power supply: 1x 1.5 V, IEC R6 with
test function
Overload protection: Up to ± 50 V AC/DC
peak Frequency range: 20 Hz…50Hz Dimensions: 98x148x49 mm
…20 kHz
3
approx.
Weight: 300 g approx.
5. Operation
Switch the device on by select the desired
operating mode,
To turn off the multimeter, set the mode
switch to the off position
To test the battery, set the measurement
range dial to the position
or
.
.
.
5.1 Current and voltage measurements
Conduct all normal current and voltage
measurements with the operating mode switch in the position
. This way, setting the measurement range with the measure­ment range dial automatically takes care of AC/DC switchovers.
Connect the lower measurement potential to
the left-hand socket. For DC quantities, the positive polarity should always be connected to the right-hand socket.
5.2 Measurements with the zero position of the needle centred
This mode of measurement only works for the DC current and DC voltage ranges.
For measurements requiring the zero posi-
tion of the needle position to be centred, set the operating mode switch to the position
.
Before connecting the external quantity to be
measured, use the adjustment knob to accu­rately centre the zero position on the scale.
2