3B Scientific Electron Diffraction Tube D User Manual

3B SCIENTIFIC
Electron Diffraction Tube D 1013885
07/13 ALF
®
PHYSICS
1 4-mm sockets for connecting
heater supply
2 2-mm socket for connecting
cathode 3 Internal resistor 4 Filament 5 Cathode 6 Anode 7 4-mm plug for connecting
anode 8 Focussing electrode 9 Polycrystalline graphite
grating 10 Boss 11 Fluorescent screen
1547
1. Safety instructions
Hot cathode tubes are thin-walled, highly evacuated glass tubes. Treat them carefully as there is a risk of implosion.
Do not subject the tube to mechanical
stresses.
Do not subject the connection leads to any
tension.
The tube may only be used with tube holder
D (1008507).
If voltage or current is too high or the cathode is at the wrong temperature, it can lead to the tube becoming destroyed.
Do not exceed the stated operating parame-
ters.
Only change circuit with power supply
equipment switched off.
Only exchange tubes with power supply
equipment switched off.
When the tube is in operation, the stock of the tube may get hot.
If necessary, allow the tube to cool before
dismantling.
The compliance with the EC directive on elec­tromagnetic compatibility is only guaranteed when using the recommended power supplies.
62R3
89
10
11
2. Description
The electron diffraction tube illustrates the wave nature of electrons by allowing observation of interference caused by a beam of electrons passing through a polycrystalline graphite target on a fluorescent screen (Debye-Scherrer diffrac­tion). The wavelength of the electrons can be calculated for various anode voltages from the radius of the diffracted rings and the distance between the crystal layers in the graphite. The tube also confirms the de Broglie hypothesis.
The electron diffraction tube is a highly evacu­ated tube with an electron gun consisting of a pure tungsten heater filament and a cylindrical anode all contained in a clear glass bulb. The electrons emitted by the heated cathode are constrained to a narrow beam by an aperture and are then focussed by means of an electron­optical system. The resulting tight, monochro­matic beam then passes through a micro-mesh nickel grating situated at the aperture of the gun. Onto this grid, a thin layer of polycrystalline graphitised carbon has been deposited by va­porisation. This layer affects the electrons in the beam much like a diffraction grating. The result of this diffraction is seen in the form of an image comprising two concentric rings that become visible on the fluorescent screen. A spot result­ing from the undeflected electron beam contin­ues to be visible at the centre of the rings.
1
=
λ
A magnet is also supplied with the tube. This allows the direction of the electron beam to be changed, which may be necessary if the graph­ite target has slight damage as a result of the manufacturing process or due to later overheat­ing.
3. Technical data
Filament voltage: 7.0 V AC/DC Anode voltage: 0 – 5000 V DC Anode current: typ. 0.15 mA
at 4000 V DC Lattice constant of graphite:
d d
= 0.213 nm
10
= 0.123 nm
11
Distance from graphite target to fluorescent screen: 125 ± 2 mm approx.
Fluorescent screen: 100 mm dia. approx. Glass bulb: 130 mm dia. approx. Total length: 260 mm dia. approx.
4. Operation
To perform experiments using the electron dif­fraction tube, the following equipment is also required:
1 Tube holder D 1008507 1 High voltage power supply 5 kV (115 V, 50/60 Hz)
1003309 or 1 High voltage power supply 5 kV (230 V, 50/60 Hz)
1003310 2 Pair of Experiment Leads, 75 cm 1002850 1 Experiment Lead, Plug and Socket 1002838
Additionally recommended: 1 Protective Adapter, 3-Pole 1009960
2 Pair of Safety Experiment Leads, 75 cm 1002849 1 Experiment Lead, Safety Plug/Socket 1002839
4.1 Setting up the tube in the tube holder
The tube should not be mounted or removed
unless all power supplies are disconnected.
Push the jaw clamp sliders on the stanchion
of the tube holder right back so that the jaws open.
Push the bosses of the tube into the jaws.
Push the jaw clamps forward on the stan-
chions to secure the tube within the jaws.
If necessary plug the protective adapter onto
the connector sockets for the tube.
4.3 General instructions
The graphite foil on the diffraction grating is only a few layers of molecules thick and any current greater 0.2 mA can cause its destruction.
The internal resistor is there to prevent damage to the graphite foil.
The graphite target itself should be monitored throughout the experiment. If the graphite target starts to glow, the anode must immediately be disconnected from its power supply
If the diffraction rings are not satisfactorily visi­ble, the electron beam can be redirected by a magnet so that it passes through an undamaged region of the target.
5. Example experiment
Set u the experiment as in Fig. 2. Connect
the negative pole of the anode supply via the 2-mm socket.
Apply the heater voltage and wait about 1
minute for the heater temperature to achieve thermal stability
Apply an anode voltage of 4 kV.
Determine the diameter D of the diffraction
rings.
Two diffraction rings appear on the fluorescent screen centred on the undeflected beam in the middle. The two rings correspond to Bragg re­flections from atoms in the layers of the graphite crystal lattice.
Changing the anode voltage causes the rings to change in diameter. Reducing the voltage makes the rings wider. This supports de Broglie's postulate that the wavelength in­creases as momentum is reduced.
a) Bragg equation:
λ = wavelength of the electrones ϑ = glancing angle of the diffraction ring
d = lattice plane spacing in graphite L = distance between sample and screen D = diameter D of the diffraction ring R = radius of the diffraction ring
D
=ϑ22tan
L
b) de-Broglie equation:
h = Planck’s constant p = momentum of the electrones
2
p
2
m
=
Ue
m = electron mas, e = electron charge
4.2 Removing tube from the tube holder
To remove the tube, push the jaw clamps
right back again and take the tube out of the jaws.
ϑ
sin2 d
R
d =λ
L
h
=λ
p
h
Uem
⋅=λ2
2
Loading...
+ 2 hidden pages