3B Scientific Digital Multimeter E User Manual

Page 1
3B SCIENTIFIC
Digital Multimeter E 1006809
05/12 ALF
®
PHYSICS
1 Measuring probe 1a Finger guards 2 Measurement socket “20 A“for
current measurement 20 A range (plus)
3 Measurement socket “A” for
current measurement for up to 2 A (plus)
4 Measurement socket “COM“
(minus)
5 Measurement socket “V/“ for
voltage and resistance meas-
urement (plus) 6 Transistor test socket 7 Measurement range dial 8 Button Hold function 9 On/Off switch 10 Display 11 Unlock button
!
1. Safety instructions
The Digital Multimeter E conforms to the safety requirements for electrical equipment for meas­urement, control and laboratory use in DIN EN 61010 part 1. Safe operation of the apparatus is guaranteed with correct handling. However, safety is not guaranteed if the apparatus is han­dled improperly or carelessly.
Read this manual carefully before using the
digital multimeter and follow the instructions!
Before using the meter, check the case and
test leads for any damage. In the event of any malfunction/operational defect or visible damage, do not use the meter. Pay particu­lar attention to the insulation surrounding the measurement sockets.
Use with caution when working above 30 V
ACrms, or 60 V DC. Such voltages pose a shock hazard.
The limit of the measurement range must
not be exceeded. If the values of the meas­urand are unknown, always switch from a higher measurement range to a lower one.
When measuring current, turn off circuit
power before connecting the meter in the circuit.
Connect the common test lead before you
connect the live test lead. When you discon­nect test leads, disconnect the live test lead first.
When using the measuring probes, keep
your fingers behind the finger guards on the probes.
Do not operate the meter around explosive
gas, vapour, or dust.
Do not conduct measurements in a humid
environment. Work area, hands, shoes and floor must be dry.
To avoid false readings, which could lead to
possible electric shock or personal injury, replace the batteries as soon as the low bat-
tery indicator (
Before the case is opened, the meter has to
be switched off and the leads must be dis­connected from the meter.
Never use the meter when the case is open.
When disposing empty batteries follow the
local regulations. Never dispose of them in the regular household garbage.
+ -
) appears.
1
Page 2
2. Symbol legend
!
Read instruction sheet
Dangerous voltages
V DC voltage
A DC current
V AC voltage
A AC current
Diode and continuity test
hFE Current amplifying factor of a transistor
Resistance
+ -
Empty battery
3. Description
The Digital Multimeter E is a robust, battery operated multimeter with a 3½-digit LCD display for measuring voltage, current and resistance as well as for diode and hFE gain testing.
All measurement ranges are selected by means of a rotary dial. All measurement ranges are protected against overload except the 20 A range.
The meter is equipped with a hold function, negative polarity indication, over range indica­tion, low battery indication and automatic switch off after 15 minutes. After the power is automati­cally switched off it needs to be turned off and turned on again to continue the power.
The digital display is folding for ease of reading and on the backside there is a fold out prop for standing the device on a table.
4. Equipment supplied
1 Digital multimeter 1 Pair of measuring probes 1 Battery 1 Instruction manual
5. Technical data
General specifications
Display: 3½- display LCD, 24 mm, max. 1999
Operating voltage: 9 V battery 6F22 Fuse: F2A/250 V Measurement rate: 2 - 3 / sec Operating temperature: 0°C - 40°C, 0 - 75%
R.H.
Storage temperature: -10°C - 50°C, 0 - 75%
R.H. Safety classification: CAT II Dimensions: 85x185x35 mm
3
approx.
Weight: 310 g approx. (including
battery)
Electrical specifications
V
Measuring range Accuracy
200 mV ±0.5 % ± 3 digits
2 V, 20 V, 200 V ±0.8 % ± 2 digits
1000 V ±1.0 % ± 2 digits
DC voltage Input impedance: 10 M
V
Measuring range Accuracy
200 mV ±1.2 % ± 5 digits
2 V, 20 V, 200 V ±1.0 % ± 5 digits
750 V ±1.2 % ± 5 digits
AC voltage Input impedance: 10 M Frequency range: 40 – 400 Hz
A
Measuring range Accuracy
20 µA ±1.8 % ± 2 digits
200 µA, 2 mA, 2 mA 20 mA, 200 mA
2 A, 20 A ±2.0 % ± 10 digits
DC current Measuring voltage drop: 200 mV
±2.0 % ± 2 digits
2
Page 3
!
!
!
A
Measuring range Accuracy
20 µA, 200 µA, 2 mA 20 mA
±2.0 % ± 3 digits
200 mA ±2.0 % ± 5 digits
2 A, 20 A ±2.5 % ± 10 digits
AC current Measuring voltage drop: 200 mV Frequency range: 40 – 400 Hz
Measuring range Accuracy
200 ±1.0 % ± 10 digits
2 k, 20 K, 200 k 2 M
20 M ±1.0 % ± 10 digits
Accuracy is given for 1 year after calibration at 23°C ±5°C, RH<75%.
6.1 Method of measurement
Warning! Dangerous voltages may be present at the Input terminals and may not be displayed.
6.1.1 Voltage measurement
Set the measurement range dial at the re-
quired position
Connect the black test lead to the meas-
urement socket "COM" and the red test lead to the "V/" socket. The meter is connected parallel to the measuring point. The polarity of the red lead connection will be indicated at the same time as the voltage.
Note
lf the voltage to be tested is unknown be-
forehand, set the measurement range dial to the highest range and work down.
When only the figure "1" is displayed, over
range is being indicated and the measure­ment range dial has be set to a higher range.
Never measure voltages higher than 1000 V.
6.1.2 Current measurement
Set the measurement range dial at the re-
quired position
Connect the black test lead to the meas-
urement socket "COM" and the red test lead to the "A" socket for measurements up to
±1.0 % ± 4 digits
6. Operation
or .
or .
2 A. For measurements over 2 A connect it to the socket “20A”. The meter is connected in series to the measuring object. The polar­ity of the red lead connection will be indi­cated at the same time as the current.
Note
lf the current to be tested is unknown be-
forehand, set the measurement range dial to the highest range and work down.
When only the figure "1" is displayed, over
range is being indicated and the measure­ment range dial has be set to a higher range.
Limit measurements inthe 20A range to
max. 15 s.
6.1.3 Resistance measurement
Warning! To avoid electrical shock or damage to the meter when measuring resis­tance in a circuit, make sure the power to the circuit is turned off and all capacitors are dis­charged.
Set the measurement range dial to the Ω
range.
Connect the black test lead to the meas-
urement socket "COM" and the red test lead to the "V/" socket. Measurement is done parallel to the resistor
Note
lf the resistance to be tested is unknown
beforehand, set the measurement range dial to the highest range and work down.
When only the figure "1" is displayed, over
range is being indicated and the measure­ment range dial has be set to a higher range.
When the input is not connected, i.e. at open circuit, the figure "1" will be displayed for the over range conditlon.
6.1.4 Diode test
Set the measurement range dial to .
Connect the black test lead to the meas-
urement socket "COM" and to the cathode of the diode. Connect the red test lead to the "V/" socket and the anode of the diode.
Note
When the input is not connected, i.e. at open circuit, the figure "1" will be displayed
The meter displays the forward voltage drop and displays figure "1" for overload when the diode is reversed.
6.1.5 Continuity test
Warning! To avoid electrical shock or damage to the meter when measuring continuity in a circuit, make sure the power to the circuit is turned off and all capacitors are discharged.
Set the measurement range dial to .
3
Page 4
Connect the black test lead to the meas­urement socket "COM" and the red test lead to the "V/" socket.
A built-In buzzer sounds if the resistance is less than 30 ± 10 Ω.
6.1.6 Transistor hFE test
Set the measurement range dial to hFE.
Make sure the transistor is "NPN" or "PNP"
type. Insert the transistor correctly into the corresponding transistor test socket.
Display reading is approx. transistor hFE value. Base current approx. 10 µA, V
approx.2.8 V.
CE
6.2 LCD Display panel angle selection
LCD display panel is locked in lie down position in normal operating condition and storage.
To change the display panel angle, push
down the button which is above the top case, and release lock.
Rotate the display panel to the best angle.
6.3 Battery and fuse replacement
Battery and fuse replacement should only be
done after the test leads have been discon­nected and power is off.
Loosen screws with suitable screwdriver and
remove case bottom.
Replace the battery resp. the fuse.
Replace the case bottom and reinstall the
three screws. Never operate the meter unless the case bottom is fully closed.
7. Maintenance
Beyond replacing batteries and fuses, do not attempt to repair or service your meter unless you are qualified to do so and have the relevant calibration, performance test, and service in­structions.
Turn off the meter and remove the test leads
before you service or clean the device.
Periodically wipe the case with a damp cloth
and mild detergent.
Do not use abrasives or solvents.
Dirt or moisture in the measurement sockets can affect readings.
Shake out any dirt that may be in the meas-
urement sockets.
Soak a new swab with isopropyl alcohol and
work around the inside of each measure­ment socket.
8. Disposal
The packaging should be disposed of at
local recycling points.
Do not dispose of the battery in the regular
household garbage. Follow the local regula­tions.
Should you need to dispose of the equip-
ment itself, never throw it away in normal domestic waste. Local regulations for the disposal of electrical equipment will apply.
Elwe Didactic GmbH Steinfelsstr. 5 08248 Klingenthal Germany www.elwedidactic.com
3B Scientific GmbH Rudorffweg 8 21031 Hamburg Germany www.3bscientific.com
Subject to technical amendments
© Copyright 2012 3B Scientific GmbH
Loading...