ProSolo ODO/CT
OPTICAL DISSOLVED OXYGEN, CONDUCTIVITY, AND TEMPERATURE METER
DON’T GO IT ALONE. GO SOLO.
The ProSolo DIGITAL handheld meter with ODO/CT offers exceptional affordability with unmatched durability. The user-friendly display delivers not only the industry’s best DO, temperature, and conductivity data, but also calculated parameters such as salinity, specific conductance, and seawater density.
What do you get if you Go Solo?
3-year warranty on handheld, 2-year warranty on probe/cable assembly and ODO sensor cap
Highly-accurate ODO® sensor technology with no stirring
Real-time salinity compensation of
DO measurements
Rechargeable lithium-ion battery that supports a full day of sampling
ProSolo OPTICAL DISSOLVED OXYGEN, CONDUCTIVITY, AND TEMPERATURE METER
ProSolo Specifications
Dimensions
Power
Operating Temperature
Storage Temperature
Display
Memory
Barometer
Units: mmHg, inHg, mbar, psi, kPa, atm
Sites and Data ID
Calibration Records
Languages
Certifications
Warranty
8.3 cm width x 21.6 cm length x 5.6 cm depth; 567 g (with battery)
Rechargeable lithium-ion battery pack provides ~48 hours with the handheld only; Battery recharge time is ~ 9 hours with the AC power adapter;
The instrument can also be powered via AC or external power through the USB port
0 to 50°C
0 to 45°C with battery installed;
0 to 60°C without battery installed
Color, LCD graphic display; 3.9 cm width x 6.5 cm height
>100,000 data sets
Range: 375 to 825 mmHg;
Accuracy: ±1.5 mmHg from 0 to 50°C; Resolution: 0.1 mmHg
100 user-defined sites and 100 user-defined data ID tags; Site pictures can be sent to the handheld via KorDSS Software
400 detailed calibration records can be stored and are available to view, download, and print (printing only available via KorDSS Software)
English, Spanish, German, French, Italian, Norwegian, Portuguese, Japanese, Chinese (Simplified & Traditional), Korean, Thai
CEC, CE; RoHS; IP-67; WEEE; FCC; UN Part III, Section 38.3, Test methods for lithium-ion batteries (Class 9)
3 years on handheld
ODO/CT Specifications
Size |
2.46 cm diameter, cable options for 1, 4, 10, 20, 30, 50, and 100 m lengths |
|
Range: 0 to 500%, 0 to 50 mg/L; |
Dissolved Oxygen |
Accuracy: 0 to 200%: ±1% of reading or 1% saturation, whichever is greater; |
(Optical luminescence) |
200 to 500%: ±8% of reading; |
Units: % saturation, % saturation |
0 to 20 mg/L: ±0.1 mg/L or 1% of reading, whichever is greater; |
local, mg/L, ppm |
20 to 50 mg/L: ±8% of reading |
|
Resolution: 0.01 mg/L and 0.1%, or 0.1 mg/L and 1% (auto-adjusts based on range) |
Temperature
(Thermistor) Units: °C, °F, K
Conductivity
(Four nickel electrode cell) Units: §S/cm, mS/cm
Salinity
(See YSI.com for additional calculated parameters) Units: §S/cm, mS/cm
Warranty
YSI, a Xylem brand 1725 Brannum Lane
Yellow Springs, OH 45387 Tel +1.800.897.4151
Fax +1.123.456.7890 www.xylem.com
Range: -5 to 70°C (temperature compensation range for DO mg/L measurement: -5 to 50°C ); Accuracy: ±0.2°C;
Resolution: 0.1°C or 0.1°F (auto-adjusts based on range)
Range: 0 to 200 mS/cm;
Accuracy: 0 to 100 mS/cm ±0.5% of reading or .001 mS/cm, whichever is greater; 100 to 200 mS/cm ±1.0% of reading;
Resolution: 0.001, 0.01, or 0.1 §S (range dependent)
Range: 0 to 70 ppt
Accuracy: ±1.0% of reading or ±0.1 ppt, whichever is greater
Resolution: 0.01 ppt
2 years on cable, probe, and ODO sensor cap
SPECIFICATIONS W145-01
ODO, ProODO, and ProOBOD are trademarks of Xylem Inc. or one of its subsidiaries. |
YSI.com/ProSolo-ODOCT |
|
|
© 2018 Xylem W145-01 0918 |
|