Suction Hose Connection, Positioning, and Length 14
Instrument Setup and Calibration 16
Installing the Peristaltic Pump Tube 16
Turning the ProSample On 17
Pump Calibration 17
Calibration Containers 17
Calibration Procedure 17
Calibration Check 18
After Calibration and/or Calibration Check 18
Installing Sample Bottle(s) 19
Remove the Top of the Housing 19
Prepare Sample Bottles and Install 19
Reattach the Top of the Housing 20
Installing a Distributor Arm 21
Configure Key Instrument Settings 22
3
Page 4
Wiring and Connecting Analog, Digital I/O, and SDI-12 Devices 24
Flow Signal Cable Wiring 24
Calibrating the Analog Signal 25
SDI-12 Cable Wiring 26
SDI-12 with Flying Leads 26
SDI-12 with EXO3 Connector 26
Connection of the Flow Signal Cable and/or SDI-12 Cable 26
Programming 27
Sampling Scenario List and Descriptions 27
Stand-Alone Sampling 28
Analog Input Trigger 32
Analog Input Volume Modulation 38
Digital Pulse Input Trigger 42
Data Pulse Input Event 45
SDI-12 Parameter-Driven Sampling 48
Maintenance and Cleaning 54
Draining the Bottom Sampler Housing 54
Desiccant Replacement 54
Peristaltic Pump Tube Replacement 55
Cleaning 55
Changing the Fuse(s) 56
Long-Term Instrument Storage 56
Factory Reset 57
YSIConnect PC Software 58
Replacement Parts and Accessories 59
Warranty and Liability 66
Technical Support 66
Service Information 66
Appendix A: ProSample Menu Tree 67
4
Page 5
THIS IS AN INTERACTIVE DOCUMENT
When viewing this document as an AdobeTM PDF, hovering
your cursor over certain phrases will bring up the finger-point
icon. Clicking elements of the Table of Contents, website URLs,
or references to certain sections will take you automatically
to those locations.
The information contained in this manual is subject to change without notice.
Effort has been made to make the information in this manual complete, accurate, and current.
The manufacturer shall not be held responsible for errors or omissions in this manual.
Consult YSI.com for the most up-to-date version of this manual.
Safety Information
Please read this entire manual before unpacking, setting up or operating this equipment. Pay attention to all
precautionary statements. Failure to do so could result in serious injury to the operator or damage to the equipment.
Make sure that the protection provided by this equipment is not impaired. Do not use or install this equipment in any
manner other than that specified in this manual.
NOTICE: The manufacturer is not responsible for any damages due to misapplication or misuse of this
product including, without limitation, direct, incidental and consequential damages, and disclaims
such damages to the full extent permitted under applicable law. The user is solely responsible to
identify critical application risks and install appropriate mechanisms to protect processes during a
possible equipment malfunction.
Precautionary Symbols in the Manual
NOTE: Information that requires special emphasis
NOTICE: Indicates a situation which, if not avoided, may cause damage to the instrument
WARNING: Indicates a potentially hazardous situation which could result in injury
Precautionary Symbols on the Instrument
Reference relevant section of this manual for operation and/or safety
information
Appears on the outside of the peristaltic pump cover. This cover
should not be open while the instrument is on, as the user can be
seriously injured if they are exposed to moving parts in the peristaltic
pump assembly.
5
Page 6
Intended Use and General Overview
Thank you for purchasing a YSI portable automatic sampler. Samplers in the ProSample series are designed to collect
field samples and to provide temporary storage before they are analyzed in a laboratory.
The ProSample portable automatic samplers are designed for use in surface water, stormwater, and wastewater
applications. These samplers can be used in a variety of scenarios, most of which are described in this manual. If
you have a question regarding the use of a sampler in your project, please contact YSI (YSI.com/contact-us) or an
authorized YSI distributor.
There are four samplers in the ProSample series. Content in this user manual applies to all ProSample models unless
otherwise noted. ProSample models include:
• ProSample P: full-size portable sampler
• ProSample P-12: full-size portable sampler with SDI-12 connectivity
• ProSample PM: mini portable sampler
• ProSample PM-12: mini portable sampler with SDI-12 connectivity
Composite sampling can be completed with all ProSample models. In this type of sampling, a single container is used
to collect multiple samples, resulting in a mixture of all samples collected.
Discrete sampling can be completed with the ProSample P and ProSample P-12. In this type of sampling, samples are
placed in different bottles. There are several bottle configurations available for the purpose of discrete sampling.
All ProSample models have the following features:
• Robust PE, double-walled, insulated housing for temperature control.
• Easy tube replacement for minimal downtime.
• Spring-loaded roller bearings in the peristaltic pump providing long tubing life.
• Highly accurate sample volume using two capacitive sensors in the peristaltic pump for volume control.
• Simple programming and calibration.
• Long battery life.
• Non-volatile data memory for up to 5 years, so you never have to worry about losing your data if power is lost.
6
Intended Use and General Overview
Page 7
Keypad Overview
ProSample Keypad
Select the desired menu.
Confirmation of entered values and selections. A confirmed
selection is indicated with .
Move within the menu.
Move to the next menu.
Entry/change of values.
Selection within the menu.
Scrolling within the data memory or bottle memory.
Entry/change of values.
Return to higher menu level.
The activity is canceled.
Terminate sleep mode.
Enter Key
Arrow Keys
Arrow Keys
Back Key
Back Key
Press for at least 5
sec
Intended Use and General Overview
7
Page 8
LED Functionality
There is an LED on the front of each ProSample that functions as follows:
• OFF: Device power is off, battery is dead, or no program is running
• Solid Green: Program is running and there are no errors
• Blinking Green: Program is running, but the battery is getting low
• Solid Red: An error has occurred. The display should be viewed for more information regarding the cause
of the error
• Blinking Red: Battery capacity is critically low
Package Contents
Included with each ProSample:
• *5 meter suction hose with screw connection and sinker weight
• 10L PE container (only the ProSample PM and the ProSample PM-12)
• Battery
• USB cable
• Two peristaltic pump tubes
• Printed copy of the ProSample Getting Started Guide
*Up to a 30 meter suction hose can be used with any ProSample model.
Please note the following items are not included with any ProSample. These items must be purchased separately.
• Bottle sets and distributor arms for the ProSample P and ProSample P-12
• SDI-12 cable for the ProSample P-12 and ProSample PM-12
• Flow signal cable
• Battery charger
• Other accessories such as strainers
Please see the Replacement Parts and Accessories section on page 60 for more details and item numbers, including
longer suction hose lengths.
The ProSample User Manual, installation file for YSIConnect, and the ProSample Ordering Guide can be found at
YSI.com/ProSample.
8
Intended Use and General Overview / Package Contents
Page 9
Unpacking
Carefully unpack the unit and accessories, and inspect for shipping damages. Compare received parts with materials
listed in the Package Contentssection. Notify YSI immediately of any damage or missing parts. Save all packing
materials until satisfactory operation is confirmed.
Unpacking
Removing the top tray (1) and the ProSample (2) from the packaging
9
Page 10
Charging the Battery
Initial Charge
The battery must be completely charged prior to first use. This should take approximately 14 to 16 hours.
NOTICE: A total discharge of the battery can result in damage to the battery. A protective mechanism built-in
which automatically switches off the device when the voltage is too low (9V or less). However, when
storing the battery long-term, regularly connect the battery to the charger to top off the charge, as this
will ensure the battery does not completely discharge.
NOTE: Please include the battery if the ProSample is returned to YSI for service — see the Technical Support
section on page 66 and the Service Information section on page 66.
To charge, connect a ProSample battery charger to the battery. Once the red light on the power supply turns to green,
the battery is fully charged and can be installed in the ProSample unit.
NOTE: There are three different battery chargers available for the ProSample. Please see the Replacement Parts
and Accessories section on page 60 for a list of available chargers.
10
Connection of the ProSample battery to the battery charger
Charging the Battery
Page 11
Battery Installation
The battery pack is not installed in the ProSample during shipment.
NOTICE: It is recommended to disconnect the battery from the instrument when transporting the ProSample.
To install the battery.
1. Open the ProSample cover.
2. Insert the battery pack into the battery tray.
3. Secure it with the strap.
4. Connect the battery plug to the battery.
1
2
3
4
Battery installation
Charging the Battery
11
Page 12
Subsequent Charges
NOTE: When charging a new battery, it is recommended to use the steps in the Initial Charge section
on page 10.
The battery must always be installed in the ProSample when the sampler is in use, as the battery is the only way to
power the sampler. The ProSample will shut off when the battery drops below 9V.
There are two options for charging the battery that depend on access to mains (AC) power.
No Mains (AC) Power at Sampling Site
The battery can be periodically charged by removing it from the instrument and charging it with a ProSample battery
charger whenever a recharge is needed (see the Initial Charge section). This is most the common method of powering
the instrument when there is no mains (AC) power source located near the ProSample unit.
Mains (AC) Power Available at Sampling Site
It is also possible to keep the battery continuously connected to mains (AC) power. In this scenario, the battery is kept
installed in the instrument and an optional Y-cable (item # 0069810) is used to simultaneously connect the battery to
the battery charger and the instrument. This is the most common method of powering the instrument where there is
a mains (AC) power source located near the ProSample unit. Since the battery charger will typically be outdoors and/
or exposed to water, an IP-67 battery charger is recommended (see the Replacement Parts and Accessories section on
page 60).
NOTE: The battery must be completely charged prior to first use. It is recommended to initially charge the battery
by connecting the battery directly to the battery charger (i.e. the ‘Y’ cable is not used) – see theInitial
Charge section on page 10. The Y-cable can be used once the battery is initially charged.
For continuous connection of the battery to AC power via the Y-cable:
1. Open the instrument cover.
2. Disconnect the battery from the instrument.
3. Connect the Y-cable to the instrument.
4. Connect the Y-cable to the battery.
5. Connect the Y-cable to the battery charger.
12
Charging the Battery
Page 13
12
3
4
Using the Y-cable to keep the battery continuously connected to AC power
Charging the Battery
13
Page 14
Installing at a Sampling Site
The ProSample unit must be placed on a steady, level surface.
YSI recommends carefully planning out the installation location before drilling any holes, cutting any tubing, etc.
The ProSample should be installed in a location where the sample temperature is between 0 °C and 40 °C. The
sampler should not be exposed to explosive substances or an environment with an explosive atmosphere.
WARNING: If the unit is to be continuously connected to AC power, ensure the electrical supply is sufficiently
protected against short circuits.
Suction Hose Connection, Positioning, and Length
The suction hose is the hose that will transport water from the sampling site to the sampler. The other key consideration
when installing a ProSample unit is the correct positioning of the suction hose. One end of the suction hose connects
to the intermediate tube, while the other end of the suction hose is placed in the environment where samples are
collected.
INTERMEDIATE TUBE
SUCTION HOSE CONNECTION
SUCTION HOSE
Suction hose connection to the intermediate tube
NOTICE: The connectors on the ProSample intermediate tube and suction hose are ¾ inch-14BSPP. If not using
a YSI suction hose (5, 10, and 20 meter available), ensure the suction hose being used has a ¾ inch14BSPP connection. A ¾ inch-14BSPP female connection (item # 630127) is available if the suction
hose being used does not have a ¾ inch-14BSPP connection.
There is a small gasket zip-tied to the suction hose upon delivery. Place this gasket in the suction hose connector
before connecting the suction hose to the intermediate tube. This gasket prevents air from getting in the line, thus
ensuring accurate dispensing of solution.
14
TO INTERMEDIATE TUBE
Installing at a Sampling Site
Page 15
The suction hose cannot be longer than 30 meters. In addition, it should be positioned so it is descending in
elevation from the sampler to the sampling point. Locations marked with a red ‘X’ on the figure below are examples
where this requirement is not satisfied. Placing the hose in a PVC pipe can help ensure this requirement is met.
The hose should not be coiled, as this can allow water to collect in the excess hose and impede the travel of water
through the hose. If the suction hose is longer than needed, remove the sinker weight at the end of the suction hose,
cut the suction hose to the desired length, then reinstall the sinker weight.
Installing at a Sampling Site
Positioning of the ProSample unit and suction hose
15
Page 16
Instrument Setup and Calibration
Installing the Peristaltic Pump Tube
The peristaltic pump tube does not come pre-installed in the ProSample. There are two peristaltic tubes included with
each ProSample — one can be found under the lid of the ProSample, while the other is in the top tray of the ProSample
packaging. To install the peristaltic pump tube:
1. Turn the ProSample OFF and disconnect the battery.
WARNING: Do NOT attempt to install a peristaltic pump tube or service the pump while the ProSample is ON.
The moving components of the pump can cause serious injury.
2. Use a flathead screw driver to loosen the bolt on the outside of the peristaltic pump cover.
3. There are two tube connectors the peristaltic pump tube will connect to — one connector is on the intermediate
tube while the other is on the composite sampling tube. Slide the connectors out of the side of the pump
housing.
4. Connect the peristaltic pump tube to each tube connector and slide the connectors back into the side of the
pump housing.
5. Place the peristaltic pump tube on the outside of the rollers of the peristaltic pump. You may need to turn the
roller assembly in order to get the pump tube to properly seat in the pump housing. Refer to the image below.
6. Close the battery cover to the peristaltic pump and tighten the bolt using a flathead screwdriver. Reconnect the
battery.
16
Installed pump tube
Instrument Setup and Calibration
Page 17
Turning the ProSample On
Press the ON/OFF button once to turn the ProSample on. Press and
release the button again to turn the unit off.
Turning on the ProSample
Pump Calibration
The peristaltic pump must be calibrated at each sampling site, as the length and gradient of the suction hose impact
the calibration result.
NOTE: Calibration in a lab setting can be completed to verify the instrument is functioning properly, but it is still
highly recommended to recalibrate the instrument once at the sampling site.
Calibration Containers
The ProSample is calibrated to an upper and lower volume, resulting in a 2-point calibration. The upper calibration
will dispense a large volume of solution, typically over 1000 mL, whereas the lower calibration will dispense a small
volume, around 250 mL.
The volume dispensed during calibration will need to be accurately measured. It is recommended to obtain a large,
plastic graduated cylinder and a small, plastic graduated cylinder when measuring the amount of solution dispensed
during calibration.
Calibration Procedure
The following steps describe how to calibrate the ProSample peristaltic pump:
1. Turn the ProSample OFF.
2. Remove the top of the ProSample housing and flip it over.
3. Remove the plastic ‘elbow’ piece from the end of the composite
sampling tube. There is a plastic bracket that keeps the
composite sampling tube in place during sampling – remove
the tube from this bracket.
4. Reinstall the top of the ProSample housing.
5. Pull the composite sampling tube out of the ProSample
housing. This will allow solution to be poured directly into a
graduated cylinder.
6. Place a graduated cylinder with a volume of at least 2000 mL
under the composite sampling tube.
7. Turn the ProSample ON.
Calibration with large graduated cylinder
8. From the Main Menu, navigate to Setup g System Settings
g Calibration Vol g Volume Calibration g Start Upper
Volume. After pressing the Enter key, the ProSample will purge
the line and then dispense a large amount of solution into the
graduated cylinder three separate times.
Instrument Setup and Calibration
17
Page 18
9. Do NOT remove the graduate cylinder from underneath the
sampling tube until Enter Actual Value is displayed on the
screen. Enter the amount of solution that was dispensed and
press the Enter key.
10. Start Lower Volume should appear on the screen. Place a
graduated cylinder with a volume of at least 250 mL under
the composite sampling tube. After pressing the Enter key,
the ProSample will purge the line and then dispense a small
amount of solution into the graduated cylinder three separate
times.
11. Do NOT remove the graduated cylinder from underneath the
sampling tube until Enter Actual Value is displayed on the
screen. Enter the amount of solution that was dispensed and
press the Enter key.
Calibration Check
The ProSample can dispense a user-specified amount of solution, allowing the pump calibration to be verified. To
complete a calibration check:
1. From the Main Menu, navigate to Manual Sample
Bottle.
2. Place a suitable graduated cylinder underneath the sampling
tube.
3. Enter the amount of Sample Volume to be dispensed. It is
recommended to enter the amount of solution that will typically
be dispensed when sampling. Press Enter.
4. Observe the amount of solution dispensed and recalibrate if
needed.
After Calibration and/or Calibration Check
g In Present
1. Turn OFF the ProSample.
2. Insert the composite sampling tube into the hole in the
ProSample housing.
3. Remove the top of the ProSample housing and flip it over.
4. Reinstall the plastic ‘elbow’ piece on the end of the composite
sampling tube. There is a plastic bracket that keeps the
composite sampling tube in place during sampling — place the
tube in this bracket.
5. Reinstall the top of the ProSample housing.
18
Instrument Setup and Calibration
Page 19
Installing Sample Bottle(s)
There are several bottle configurations available for the ProSample P and ProSample P-12. Plastic (PE) and glass bottle
sets are available. Please see the Replacement Parts and Accessories section on page 60 for a complete listing of
available bottle sets.
If multiple bottles will be installed in a ProSample P or ProSample P-12, a distributor arm must be installed. Please see
Installing a Distributor Armon page 21.
A single 10 liter PE bottle can be placed in the ProSample PM and ProSample PM-12, but no other bottles can be used,
as a distributor arm cannot be installed in the mini samplers.
Remove the Top of the Housing
Disconnect the clips that hold the top part of the housing to the rest of the instrument. Lift the top part of the housing
off the instrument.
2
1
Removal of the top part of the ProSample housing
Prepare Sample Bottles and Install
Ensure the sample bottles are clean and suitable for use before installing in the bottom ProSample housing.
NOTE: A fixing plate is included with some bottle sets. If a fixing plate is included with your bottle set, it should
also be installed in the bottom ProSample housing.
Instrument Setup and Calibration
19
Page 20
Installation of plastic (PE) bottles
Installation of glass bottles
Bottle #1 position and filling direction on the ProSample P and ProSample P-12 housing
Reattach the Top of the Housing
After installing the sample bottle(s), reattach the top part of the housing by lining it up with the bottom part of the
housing. Use the clips on the side of the instrument to secure the top part of the housing to the bottom.
20
Instrument Setup and Calibration
Page 21
Installing a Distributor Arm
Multiple bottles can be placed in the ProSample P and ProSample P-12, allowing for the collection of discrete sample
(see Installing Sample Bottle(s) on page 19). If multiple bottles are to be used, a distributor arm must be installed.
There are two distributor arms available for the ProSample P and ProSample P-12. Each distributor arm includes a preinstalled distributor tube.
• Item # 630112 is for the following bottle sets: 24 x 1 L PE, 8 x 2L glass, 12 x 950 mL glass, and 24 x 350 mL glass.
• Item # 630113 if for the following bottles sets: 4 x 4L PE and 8 x 2 L PE.
To install a distributor arm:
1. Disconnect the composite sampling tube from the tube connector on the peristaltic pump housing.
2. Flip over the top of the ProSample housing and remove the composite sampling tube.
3. Remove the bolt with a red top. This covers the distributor arm axle.
4. Place the distributor arm on the axle by lining up the pin on the axle with the hole on the distributor arm. Install
the bolt with the red top and hand-tighten.
2
1
5. There is a plastic bracket on the bottom of the ProSample housing that keeps the distributor tube in place
during sampling — place the distribution tube into this plastic bracket.
6. Slide the end of the distributor tube into the hole near the peristaltic pump.
7. Reinstall the top of the ProSample housing.
8. Attach the distributor tube to the tube connector on the peristaltic pump housing.
Instrument Setup and Calibration
21
Page 22
Configure Key Instrument Settings
There are seven key instrument settings that are used for each program created. From the Main Menu, navigate to the
following screens to modify these key settings:
1. Setup g Date/Time
2. Setup g System Settings g Language
3. Setup g System Settings g Distributor (ProSample P and
ProSample P-12 Only)
On the ProSample P and P-12, select the number of bottles (1,
4, 8, 12 or 24 ) that will be used (see Installing Sample Bottle(s)
on page 19).
4. Setup g System Settings g Max. Suction Time
The maximum suction time determines how long the
ProSample will wait to display an error and shut off the pump
when no water is detected by the instrument’s sensors.
22
Instrument Setup and Calibration
Page 23
5. Setup g System Settings g Pre purge
In order to clear the line of moisture and remove debris that
may have collected around the suction hose opening, the
peristaltic pump will push air through the suction hose before
a new sample is collected. Pre purge identifies the amount of
time air will be pushed through the suction hose.
The recommended Pre purge time depends on the length of
the suction hose. To identify the exact amount of Pre purge
time needed, set Pre purge to a relatively high value (e.g. 10
seconds). Once a Pre purge cycle starts, use a stopwatch to
determine how long it takes for air to be blown out of the
suction hose. This amount of time can then be entered for Pre
purge.
NOTE: The ProSample can be manually purged by going to
Diagnostics/Test
right arrow key to run the pump backwards for purging.
6. Setup g System Settings g Post purge
Post purge is the amount of time the peristaltic pump will
push air through the line after a sample is collected. It is
recommended to set the Post purge to the same value as Pre
purge.
gComponent Test gPump. Press the
7. Setup
Rinse Before Sampl. is the number of times the suction hose will
be rinsed with sample solution before a sample is collected. Up
to 3 rinses can be completed.
g System Settings g Rinse Before Sampl.
Instrument Setup and Calibration
23
Page 24
Wiring and Connecting Analog, Digital I/O,
and SDI-12 Devices
The ProSample P and ProSample PM can be configured to sample based on measurements/signals from analog and
digital I/O devices, while the ProSample P-12 and ProSample PM-12 can additionally be configured to sample based
on measurements from SDI-12 devices (e.g. EXO3 sonde).
Flow Signal Cable Wiring
The flow signal cable (item # 630147) can be used to send a digital, analog, or event signal to the ProSample. One
end of this cable has flying leads so it can be connected to an analog and/or digital device, while the other end can be
connected to the Signal Input on the ProSample. Please refer to the wire colors below when connecting the flow signal
cable to the digital and/or analog device.
Wire ColorReference NumberInput SignalPurpose
Brown1Analog +Connection of an analog device
White2Analog -Connection of an analog device
Grey3ComConnection of a digital device
Yellow4DigitalConnection of a digital device
Green5EventEvent
NOTE: The above table is only relevant if item #630147 is used.
NOTE: An analog and digital device can be connected to the ProSample using the flow signal cable.
NOTE: The green and common (grey) can be connected to a relay connection. When the relay closes, an event is
triggered and sampling will start.
24
Wiring and Connecting Analog, Digital I/O, and SDI-12 Devices
Page 25
Calibrating the Analog Signal
The analog signal coming to the ProSample from an analog device must be calibrated before configuring any analog
program. To calibrate the analog signal:
1. From the Main Menu, select Setup
Analog Signal g Calibration
2. Analog 4-20 mA and Analog 0-20 mA are available options.
Highlight the appropriate option for your device and press the
Enter key.
g System Settings g
3. On your analog device, choose to output a mA value that
corresponds to the low value of the range selected in step #3.
As an example, if 4-20 mA was selected in step #3, output a 4
mA signal from your analog device.
4. The Measured Value on the ProSample display should equal
(or nearly equal) the desired mA value. Press the Enter key to
confirm.
As an example, if 4 mA is the signal sent from the analog
device, the Measured Value might be equal to 3.9 mA. Please
note the screenshot below was taken before the mA device
sent a 4 mA signal to the ProSample.
5. On your analog device, choose to output 20 mA.
6. The Measured Value on the ProSample display should equal
(or nearly equal) 20 mA. Press the Enter key to confirm.
7. The ProSample will display Calibration Done. Press the Enter
key to return to the System Settings menu.
Wiring and Connecting Analog, Digital I/O, and SDI-12 Devices
25
Page 26
SDI-12 Cable Wiring
There are two SDI-12 cables available from YSI for use with the ProSample P-12 and ProSample PM-12.
SDI-12 Cable with Flying Leads
One end of item # 630146 has flying leads for connecting to an SDI-12 device or a Signal Output Adapter for the SDI12 device (e.g. YSI EXO1 and YSI EXO2). The other end of this cable is connected to the SDI-12 input on the ProSample
P-12 or ProSample PM-12.
Please refer to the wire colors below when connecting the flow signal cable to the digital and/or analog device, as well
as the manual for your SDI-12 device .
Wire ColorReference NumberDescription
Brown1SDI-12 DATA
White2SDI-12 GND
Grey312 VDC
Yellow4GND
Green5Error Signal
Blue612 VDC
Pink7Prog. Input
SDI-12 Cable with EXO3 Connector
620145 cables have all of the connections necessary to directly connect a YSI EXO3 directly to the ProSample P-12 and
ProSample PM-12. Therefore, no wiring is needed. These cables are offered in lengths of 2, 4, 10, 15, and 33 meters.
Connection of the Flow Signal Cable and/or SDI-12 Cable
There are two connectors on the side of the ProSample near the peristaltic pump. The top connector is for the flow
signal cable, while the bottom connector (ProSample P-12 and ProSample PM-12 only) is for connection of an SDI-12
cable.
26
Wiring and Connecting Analog, Digital I/O, and SDI-12 Devices
Page 27
Programming
The ProSample can be used in a variety of scenarios that range from simple to complex. This programming section
reviews the most common sampling scenarios. Please contact YSI Technical Support if you would like to use the
ProSample in a sampling scenario that is not described in this manual.
NOTE: A complete review of the ProSample menu can be found in Appendix A: ProSample Menu Tree.
ProSample units with an SDI-12 input (ProSample P-12 and ProSample PM-12) can be used in all of the scenarios
described below, while ProSample units without an SDI-12 input (ProSample P and ProSample PM) can be used for all
scenarios expect SDI-12 parameter-driven sampling.
NOTE: The settings modified in the section Configure Key Instrument Settings on page 22 will be applied to each
and every program, so it is important these settings are carefully set before starting any program.
Sampling Scenario List and Descriptions
1. Stand-Alone Sampling: The ProSample will collect samples at a user-defined interval. All ProSample models can
be used for this sampling scenario. See page 28.
2. Analog Input Trigger: The ProSample will begin collecting samples when a 4-20 mA device (e.g. YSI Amazon
Bubbler or YSI IQSN sensor) meets user-defined sampling criteria. All ProSample models can be used for this
sampling scenario. See page 32.
3. Analog Input Volume-Modulation: The ProSample will begin collecting samples when a 4-20 mA device (e.g.
YSI Amazon Bubbler or YSI IQSN sensor) meets user-defined sampling criteria. The amount of sample collected
depends upon the 4-20 mA signal. All ProSample models can be used for this sampling scenario. See page 38.
4. Digital Pulse Input Trigger: The ProSample will trigger a program to start collecting samples when a digital pulse
is sent to the ProSample from a digital device, such as a data logger or digital flow meter. The ProSample P-12
and ProSample PM-12 can be used for this sampling scenario. See page 42.
5. Digital Pulse Input Event: The ProSample will take a sample each time it receives a digital pulse from a digital
device, such as a data logger or digital flow meter. All ProSample models can be used for this sampling
scenario. See page 45.
6. SDI-12 Parameter-Driven Sampling: Samples will be collected when parameter(s) from an SDI-12 device (e.g.
YSI EXO3 sonde) meet user-defined sampling criteria. The ProSample P-12 and ProSample PM-12 can be used
for this sampling scenario. See page 48.
Programming
27
Page 28
Stand-Alone Sampling
The ProSample will collect samples at a user-defined interval. All ProSample models can be used for this sampling
scenario. A flow signal cable and/or SDI-12 cable is not required to run this program.
NOTE: The settings modified in the section Configure Key Instrument Settings on page 22 will be applied to each
and every program, so it is important these settings are carefully set before starting any program.
NOTE: Screenshots demonstrate an example setup. Settings should be adjusted for your specific application.
1. From the Main Menu, select Program
program you would like to modify by using the left and right
arrow keys. In this example, we will modify Program # 1. Press
the Enter key to begin modifying the program.
2. Highlight Time and press the Enter key.
3. Enter a Sampling Interval. This determines how often a sample
will be collected and is determined by the user based on the
application and sampling requirements. Press the Enter key to
begin modifying the program.
g Change. Choose the
28
4. The Sample Volume is displayed. This is the amount of sample
that will be collected at each Sampling Interval. The Sample
Volume is determined by the user based on the application and
sampling requirements. Press the Enter key to begin modifying
the program.
Programming: Stand-Alone Sampling
Page 29
5. The Bottle Fill Time is displayed.
If using a composite bottle, Bottle Fill Time determines how
long the program will run. As an example, if Sampling Interval
is set to 15 minutes, Sample Volume is set to 100 mL, and Bottle
Fill Time is set to 2 hours, then the ProSample will collect 100
mL of sample every 15 minutes for 2 hours, resulting in a total
of 800 mL of sample collected.
If using discrete sampling bottles (i.e. more than one bottle is
used), Bottle Fill Time determines how many times each bottle
in the program will be filled. As an example, if Sampling Interval
is set to 15 minutes, Sample Volume is set to 100 mL, and
Bottle Fill Time is set to 15 minutes (i.e. the same as Sampling
Interval), then each bottle is only filled once and each bottle will
contain 100 mL of sample. The entire length of the program
is determined by the number of bottles that will be filled (see
steps # 5-7). Therefore, every 15 minutes a different bottle will
be filled. If there are 6 bottles, the entire program length will be
90 minutes.
As another example of discrete sampling, if Sampling Interval
is set to 15 minutes, Sample Volume is set to 100 mL, and
Bottle Fill Time is set to 45 minutes (i.e. three times longer than
Sampling Time), then each bottle is filled three times and each
bottle will contain 300 mL of sample
6. The Info screen is displayed.
If there is only one bottle installed (composite sampling):
• Select Settings Complete and the ProSample will return to
the Programs menu. Skip to step # 10.
If there is a distributor arm and multiple bottles installed
(discrete sampling):
• Select More Settings g Bottle Selection.
t If you would like to fill all bottles in the sampler, select All Bottles. Return to the More Settings menu by pressing the
Back key once, and then select Programming OK. Skip to
step #10.
t If you only want to fill certain bottles, select Bottle Selection then continue with step #7 below.
Programming: Stand-Alone Sampling
29
Page 30
7. First Bottle is displayed if the ProSample has more than one
bottle installed. Enter the bottle number in which the first
sample should be placed. Press the Enter key.
8. Last Bottle is displayed. Enter the bottle number in which the
last sample should be placed. Press the Enter key.
9. The More Settings menu is displayed. Highlight Programming
OK and press the Enter key. The Programs menu is displayed.
10. To run the program, highlight Start and press the Enter key.
NOTE: Ensure the instrument is calibrated before running any
programs. Please see the Pump Calibration section (pg.
17).
11. Use the right arrow key to select the program to run. Press the
Enter key.
30
Programming: Stand-Alone Sampling
Page 31
12. Choose when the program should start. Press the Enter key.
NOTE: Ensure the date and time of the unit is correct. See
Configure Key Instrument Settings on page 22.
13. Choose when the program should end. Press the Enter key.
Programming: Stand-Alone Sampling
31
Page 32
Analog Input Trigger
In this type of program, sampling will only occur when the trigger criteria are met. If the trigger criteria are no longer
met while the program is active, sampling will pause. Sampling will continue if and when the trigger criteria are met
again and will continue until the program is complete.
All ProSample models can be used for this sampling scenario. A flow signal cable (item # 630147) must be connected
to the analog device and the ProSample. As is shown in the Flow Signal Cable Wiring section on page 24, the brown
and white wires should be used when connecting the flow signal cable to the analog device.
NOTE: The settings modified in the section Configure Key Instrument Settings on page 22 will be applied to each
and every program, so it is important these settings are carefully set before starting any program.
NOTE: Screenshots demonstrate an example setup. Settings should be adjusted for your specific application.
NOTE: The analog signal should be calibrated before running an analog program. See Calibrating the Analog
Signal on page 25.
1. From the Main Menu, select Setup g System Settings g
Measurement Inputs
2. Trigger (CH) is displayed. Use the right arrow key to select the
trigger channel to configure. Press the Enter key to configure
this channel.
3. Highlight Analog 1 and press the Enter key.
4. The Unit does not ultimately impact the measurement result,
but a unit does need to be used. In this example, m will be used
as the ‘dummy’ unit. After highlighting a unit, press the Enter
key.
32
Programming: Analog Input Trigger
Page 33
5. Input a 0/4 mA value. In this example, 4 will be entered. After
entering a value, press the Enter key.
6. Input a 20 mA value. In this example, 20 will be entered. After
entering a value, press the Enter key.
7. Select the type of Limit Value to be used.
If a Low Setpoint is selected, any measurement value lower
than the Low Setpoint will trigger samples to be taken. As an
example, if a Low Setpoint of 6 mA is entered, then sampling
will be triggered once the mA value lowers to 6.
If a High Setpoint is selected, any measurement value higher
than the High Setpoint will trigger samples to be taken. As an
example, if a High Setpoint of 10 mA is entered, then sampling
will be triggered once the mA value increases to 10.
Programming: Analog Input Trigger
8. Enter the Setpoint value and press the Enter key.
33
Page 34
9. The ProSample will request a Deadband be entered.
If a High Setpoint is entered, Deadband is the value which the
mA signal must drop below before sampling no longer occurs.
As an example, if 10 mA is the High Setpoint and a Deadband
of 0.5 is entered, the mA signal must drop below 9.5 mA before
sampling no longer occurs.
If a Low Setpoint is entered, Deadband is the value which the
mA signal must rise above before sampling no longer occurs.
As an example, if 5 mA is the Low Setpoint and a Deadband
of 0.5 is entered, the mA signal must rise above 5.5 mA before
sampling no longer occurs.
10. Trigger Channel 1 will be displayed. Highlight Event and press
the Enter key.
11. Go back to the Main Menu by pressing the Back key a few
times.
12. Select Programs g Change. Choose the program you would
like to modify by using the left and right arrow keys. In this
example, we will modify Program # 1. Press the Enter key to
begin modifying the program.
13. Highlight Event Time and press the Enter key.
34
Programming: Analog Input Trigger
Page 35
14. Enter a Sampling Interval. This determines how often a sample
will be collected and is determined by the user based on the
application and sampling requirements. Press the Enter key to
begin modifying the program.
15. The Sample Volume is displayed. This is the amount of sample
that will be collected at each Sampling Interval. The Sample
Volume is determined by the user based on the application and
sampling requirements. Press the Enter key to begin modifying
the program.
16. The Bottle Fill Time is displayed.
If using a composite bottle, Bottle Fill Time determines how
long the program will run. As an example, if Sampling Interval
is set to 15 minutes, Sample Volume is set to 100 mL, and Bottle
Fill Time is set to 2 hours, then the ProSample will collect 100
mL of sample every 15 minutes for 2 hours, resulting in a total
of 800 mL of sample collected.
If using discrete sampling bottles (i.e. more than one bottle is
used), Bottle Fill Time determines how many times each bottle
in the program will be filled. As an example, if Sampling Interval
is set to 15 minutes, Sample Volume is set to 100 mL, and
Bottle Fill Time is set to 15 minutes (i.e. the same as Sampling
Interval), then each bottle is only filled once and each bottle will
contain 100 mL of sample. The entire length of the program
is determined by the number of bottles that will be filled (see
steps # 5-7). Therefore, every 15 minutes a different bottle will
be filled. If there are 6 bottles, the entire program length will be
90 minutes.
As another example of discrete sampling, if Sampling Interval
is set to 15 minutes, Sample Volume is set to 100 mL, and
Bottle Fill Time is set to 45 minutes (i.e. three times longer than
Sampling Time), then each bottle is filled three times and each
bottle will contain 300 mL of sample.
Programming: Analog Input Trigger
35
Page 36
17. The Info screen is displayed.
If there is only one bottle installed (composite sampling):
• Select Settings Complete and the ProSample will return to
the Programs menu. Skip to step #21.
If there is a distributor arm and multiple bottles installed
(discrete sampling):
• Select More Settings g Bottle Selection.
t If you would like to fill all bottles in the sampler, select All Bottles. Return to the More Settings menu by pressing the
Back key once, and then select Programming OK. Skip to
step #21.
t If you only want to fill certain bottles, select Bottle Selection
and continue with step #18.
18. First Bottle is displayed if the ProSample has more than one
bottle installed. Enter the bottle number in which the first
sample should be placed. Press the Enter key.
19. Last Bottle is displayed. Enter the bottle number in which the
last sample should be placed. Press the Enter key.
20. The More Settings menu is displayed. Highlight Programming
OK and press the Enter key. The Programs menu is displayed.
21. To run the program, highlight Start and press the Enter key.
NOTE: Ensure the instrument is calibrated before running any
programs. Please see the Pump Calibration on page 17.
36
Programming: Analog Input Trigger
Page 37
22. Use the arrows to select the program to run. Press the Enter key.
23. Choose when the program should start. Press the Enter key.
NOTE: Ensure the date and time of the unit is correct. See
Configure Key Instrument Settings on page 22.
24. Choose when the program should end. Press the Enter key.
Programming: Analog Input Trigger
25. The program will start sampling once the Trigger criteria have
been met. As an example, if the High Setpoint is 10 m and
Deadband is 0.5, the ProSample will not take a sample until
the mA signal reaches a value of at least 10 and will not stop
sampling unless the mA value drops below 9.5 or all bottles
have been filled. If the mA value drops below 9.5 and then
raises to a value higher than 10, sampling will continue until all
bottles have been filled.
37
Page 38
Analog Input Volume Modulation
In this type of program, the ProSample will begin collecting samples when a 4-20 mA device (e.g. YSI Amazon Bubbler
or YSI IQSN sensor) meet user-defined sampling criteria. The amount of sample collected depends upon the 4-20 mA
signal.
All ProSample models can be used for this sampling scenario. A flow signal cable (item # 630147) must be connected
to the analog device and the ProSample. As is shown in the Flow Signal Cable Wiring section on page 24, the brown
and white wires should be used when connecting the flow signal cable to the analog device.
NOTE: The settings modified in the section Configure Key Instrument Settings on page 22 will be applied to each
and every program, so it is important these settings are carefully set before starting any program.
NOTE: Screenshots demonstrate an example setup. Settings should be adjusted for your specific application.
NOTE: The analog signal should be calibrated before running an analog program. See Calibrating the Analog
Signal on page 25.
1. From the Main Menu, select Programs g Change. Choose the
program you would like to modify by using the left and right
arrow keys. In this example, we will modify Program # 1. Press
the Enter key to begin modifying the program.
2. Highlight Variable Time and press the enter key.
3. Var Unit is displayed. This ultimately does not impact the
measurement result, but a unit does need to be used. In this
example, l/s will be used as the ‘dummy’ unit. After highlighting
a unit, press the Enter key.
4. Var Set Flow at 20 mA is displayed. Although the value
entered here ultimately does not impact the measurement
result, it can be used to document the measurement value (e.g.
flow) that corresponds to a mA signal of 20 mA. After entering a
value, press the Enter key.
38
Programming: Analog Input Volume Modulation
Page 39
5. Enter a Var Sample Interval. This determines how often a
sample will be collected and is determined by the user based
on the application and sampling requirements.
6. The Var Sample Volume 20 mA is displayed. This is the
amount of solution that should be collected when the mA
signal is 20 mA. Enter a value and press the Enter key.
NOTE: The Var Sample Volume should be lower than or equal
to the Max Sample Volume set under Setup g System
Settings g Max. Sample Volume.
7. Var Fill Mode is displayed. Time Related is the only option, so
the Enter key can be pressed.
8. Enter a Var Fill Time.
If using a composite bottle, Var Fill Time determines how long
the program will run. As an example, if Sample Interval is set to
15 minutes and Var Fill Time is set to 2 hours, the ProSample
will collect a total of 8 samples in the composite bottle. The
amount collected will vary based on the mA signal at the time
of sampling (e.g. 500 mL collected at 20 mA and 0 mL collected
at 4 mA).
If using discrete sampling bottles (i.e. more than one bottle is
used), Var Fill Time determines how many times each bottle in
the program will be filled. As an example, if Sample Interval is
set to 15 minutes and Var Fill Time is set to 15 minutes (i.e. the
same as Sampling Interval), then each bottle is only filled once
and each bottle will contain a volume of sample that varies
based on the mA output at the time of sample collection. The
entire length of the program is determined by the number of
bottles that will be filled (see steps # 5-7). Therefore, every 15
minutes a different bottle will be filled. If there are 6 bottles, the
entire program length will be 90 minutes.
Programming: Analog Input Volume Modulation
39
Page 40
9. The Var Max. Sample Volume is displayed. Enter a value that
is larger than the Var Sample Volume 20 mA (see step #6) but
smaller than or equal to the maximum sample volume of the
containers being used (e.g. 1000 mL when using 1000 mL
bottles).
10. Highlight a Var Function and press the Enter key. This
determines what will happen if there is more solution that will
be placed in a bottle than the bottle can actually hold (see Var
Max. Sample Volume). It is recommended to highlight Drop
Samples and press the Enter key.
11. The Info screen is displayed.
If there is only one bottle installed (composite sampling):
• Select Settings Complete and the ProSample will return to
the Programs menu. Skip to step #15.
If there is a distributor arm and multiple bottles installed
(discrete sampling):
• Select More Settings g Bottle Selection.
t If you would like to fill all bottles in the sampler, select All Bottles. Return to the More Settings menu by pressing the
Back key once, and then select Programming OK. Skip to
step #15.
t If you want only to fill certain bottles, select Bottle
Selection and continue with step #12.
40
12. First Bottle is displayed if the ProSample has more than one
bottle installed. Enter the bottle number in which the first
sample should be placed. Press the Enter key.
Programming: Analog Input Volume Modulation
Page 41
13. Last Bottle is displayed. Enter the bottle number in which the
last sample should be placed. Press the Enter key.
14. The More Settings menu is displayed. Highlight Programming
OK and press the Enter key. The Programs menu is displayed.
15. To run the program, highlight Start and press the Enter key.
NOTE: Ensure the instrument is calibrated before running any
programs. Please see the Pump Calibration on page 17.
16. Use the arrows to select the program to run. Press the Enter key.
17. Choose when the program should start. Press the Enter key.
NOTE: Ensure the date and time of the unit is correct. See
18. Choose when the program should end. Press the Enter key.
19. The amount of sample collected will vary based on the mA
signal from the analog device.
Programming: Analog Input Volume Modulation
Configure Key Instrument Settings on page 22.
41
Page 42
Digital Pulse Input Trigger
In this sampling scenario, a sampling program will automatically start when a pulse is received from a digital
device. The program will continue until it is complete, even if a pulse is only sent once from the digital device to the
ProSample. Therefore, a digital pulse essentially triggers a stand-alone sampling program to start.
This sampling scenario can only be completed with the ProSample P-12. The SDI-12 cable with flying leads (item
#630146) must be used. Please refer to the SDI-12 Cable with Flying Leads section on page 26 or contact YSI for
assistance when setting up this sampling scenario.
NOTE: The settings modified in the section Configure Key Instrument Settings on page 22 will be applied to each
and every program, so it is important these settings are carefully set before starting any program.
NOTE: Screenshots demonstrate an example setup. Settings should be adjusted for your specific application.
1. From the Main Menu, select Setup g System Settings g
Program Inputs g Program Input.
2. Highlight Prg. Start Pulse and press the Enter key.
3. Select a Program Input. If no other programs will be
configured (e.g. SDI-12 or analog programs), Program 1 Start
can be selected. Otherwise, select an appropriate Program
Input given the other programs that will be configured.
4. Press the Enter key. This will add a checkmark next to the
appropriate Program Input.
5. Go back to the Main Menu by pressing the Back key a few
times.
42
6. Select Programs g Change. Choose the program you would
like to modify by using the left and right arrow keys. The
program number selected must match the program number set
in steps #3 and 4.
Programming: Digital Pulse Input Trigger
Page 43
7. Highlight Time and press the Enter key.
8. Enter a Sampling Interval. This determines how often a sample
will be collected and is determined by the user based on the
application and sampling requirements. Press the Enter key to
begin modifying the program.
9. The Sample Volume is displayed. This is the amount of sample
that will be collected at each Sampling Interval. The Sample
Volume is determined by the user based on the application and
sampling requirements. Press the Enter key to begin modifying
the program.
Programming: Digital Pulse Input Trigger
10. The Bottle Fill Time is displayed.
If using a composite bottle, Bottle Fill Time determines how
long the program will run. As an example, if Sampling Interval
is set to 15 minutes, Sample Volume is set to 100 mL, and Bottle
Fill Time is set to 2 hours, then the ProSample will collect 100
mL of sample every 15 minutes for 2 hours, resulting in a total
of 800 mL of sample collected.
If using discrete sampling bottles (i.e. more than one bottle is
used), Bottle Fill Time determines how many times each bottle
in the program will be filled. As an example, if Sampling Interval
is set to 15 minutes, Sample Volume is set to 100 mL, and
Bottle Fill Time is set to 15 minutes (i.e. the same as Sampling
Interval), then each bottle is only filled once and each bottle will
contain 100 mL of sample. The entire length of the program
is determined by the number of bottles that will be filled (see
steps # 5-7). Therefore, every 15 minutes a different bottle will
be filled. If there are 6 bottles, the entire program length will be
90 minutes.
43
Page 44
As another example of discrete sampling, if Sampling Interval
is set to 15 minutes, Sample Volume is set to 100 mL, and
Bottle Fill Time is set to 45 minutes (i.e. three times longer than
Sampling Time), then each bottle is filled three times and each
bottle will contain 300 mL of sample.
11. The Info screen is displayed.
If there is only one bottle installed (composite sampling):
• Select Settings Complete and the ProSample will return to
the Programs menu. Skip to step # 14.
If there is a distributor arm and multiple bottles installed
(discrete sampling):
• Select More Settings g Bottle Selection.
t If you would like to fill all bottles in the sampler, select All
Bottles. Return to the More Settings menu by pressing the
Back key once, and then select Programming OK. Skip to
step #14.
t If you only want to fill certain bottles, select Bottle
Selection and continue with step #12.
12. First Bottle is displayed if the ProSample has more than one
bottle installed. Enter the bottle number in which the first
sample should be placed. Press the Enter key.
13. Last Bottle is displayed. Enter the bottle number in which the
last sample should be placed. Press the Enter key. The More Settings menu is displayed. Highlight Programming OK and
press the Enter key.
14. The program will automatically start when a digital pulse is
sent to the ProSample. The program will continue until it is
complete, even if a digital pulse is only sent once.
44
Programming: Digital Pulse Input Trigger
Page 45
Digital Pulse Input Event
In this program, sampling will occur each time the ProSample receives a pulse from a digital device (e.g. a data logger).
Please refer to the user manual of your device to ensure it is programmed as desired.
All ProSample models can be used for this sampling scenario. A flow signal cable (item # 630147) must be connected
to the digital device and the ProSample. As is shown in the Flow Signal Cable Wiring section on page 24, the grey and
yellow wires should be used when connecting the flow signal cable to the digital device.
NOTE: The settings modified in the section Configure Key Instrument Settings on page 22 will be applied to each
and every program, so it is important these settings are carefully set before starting any program.
NOTE: Screenshots demonstrate an example setup. Settings should be adjusted for your specific application.
1. From the Main Menu, select Programs g Change. Choose the
program you would like to modify by using the left and right
arrow keys. In this example, we will modify Program # 1. Press
the Enter key to begin modifying the program.
2. Highlight Flow Digital and press the Enter key.
3. Enter 0001 for Sample After. The data logger will send a pulse
to the ProSample when a sample needs to be taken.
Programming: Digital Pulse Input Event
45
Page 46
4. The Sample Volume is displayed. This is the amount of sample
that will be collected at each Sampling Interval. The Sample
Volume is determined by the user based on the application and
sampling requirements. Press the Enter key to begin modifying
the program.
5. Fill Mode is displayed. Highlight Sample Related and press
the Enter key.
6. Samples/Bottle identifies the number of samples that will be
placed in each bottle.
7. The Info screen is displayed.
If there is only one bottle installed (composite sampling):
• Select Settings Complete and the ProSample will return to
the Programs menu. Skip to step #11.
If there is a distributor arm and multiple bottles installed
(discrete sampling):
• Select More Settings g Bottle Selection.
t If you would like to fill all bottles in the sampler, select All Bottles. Return to the More Settings menu by pressing the
Back key once, and then select Programming OK. Skip to
step #11.
t If you only want to fill certain bottles, select Bottle
Selection and continue with step #8.
46
8. First Bottle is displayed if the ProSample has more than one
bottle installed. Enter the bottle number in which the first
sample should be placed. Press the Enter key.
Programming: Digital Pulse Input Event
Page 47
9. Last Bottle is displayed. Enter the bottle number in which the
last sample should be placed. Press the Enter key.
10. The More Settings menu is displayed. Highlight Programming
OK and press the Enter key. The Programs menu is displayed.
11. To run the program, highlight Start and press the Enter key.
NOTE: Ensure the instrument is calibrated before running any
programs. Please see the Pump Calibration section on
pg. 17.
12. Use the right arrow key to select the program to run. Press the
Enter key.
Programming: Digital Pulse Input Event
13. Choose when the program should start. Press the Enter key.
NOTE: Ensure the date and time of the unit is correct. See
Configure Key Instrument Settings on page 22.
14. Choose when the program should end. Press the Enter key.
15. Once the data loggers is connected and the sample criteria
(defined during setup of the digital device) are met, the digital
device will send a pulse to the ProSample to collect a sample.
47
Page 48
SDI-12 Parameter-Driven Sampling
In this sampling scenario, the ProSample will collect samples based upon measurements from an SDI-12 device (e.g.
YSI EXO3 sonde). This sampling scenario can only be completed with the ProSample P-12 and ProSample PM-12.
Please refer to the SDI-12 Cable Wiring section on page 26 when connecting the SDI-12 device to the ProSample.
In order to use an SDI-12 device, the user must configure it properly before programming the ProSample P-12 or
ProSample PM-12. As an example, if pH is a trigger parameter, the pH channel on the sonde and the ProSample must
be the same (e.g. channel 1). Please consult the user manual of your SDI-12 device regarding proper configuration.
Up to 4 SDI-12 parameters can be configured to trigger sampling. If there is a need for an AND scenario with multiple
parameters to trigger sampling (e.g. solution must have pH > 8 AND turbidity >124 FNU for sampling to begin), then a
data logger must be used to trigger sampling (see Digital Pulse Input Event on page 45).
There are two different methods of programming the sampler to stop sampling – continuous method and event
method. Steps # 9 and 12 in the following procedure determine which method is utilized.
Continuous method: If the trigger criteria are met, sampling will occur until the program is complete, even if the
trigger criteria are no longer met while the program is active. If more than one SDI-12 parameter is set to trigger
sampling, only the continuous method can be used.
Event method: Sampling will only occur when the trigger criteria are met. If the trigger criteria are no longer met while
the program is active, sampling will pause. Sampling will continue if and when the trigger criteria are met again and
will continue until the program is complete.
NOTE: The settings modified in the section Configure Key Instrument Settings on page 22 will be applied to each
and every program, so it is important these settings are carefully set before starting any program.
NOTE: Screenshots demonstrate an example setup. Settings should be adjusted for your specific application.
The following procedure should be followed for each SDI-12 parameter (up to 4) that will be used to trigger sampling.
NOTE: If using the continuous method, it is recommended to program the sampler (i.e. follow the procedure
below), place the SDI-12 device in the sampling environment, and then connect the SDI-12 device to the
ProSample P-12 or ProSample PM-12. This is recommended because the program is automatically started
if the trigger criteria are met (i.e. the program is not started by going to Program g Start).
1. From the Main Menu, select Setup g System Settings g
Measurement Input.
2. Trigger (CH) is displayed. Use the right arrow key to select the
trigger channel to configure. Press the Enter key to configure
this channel.
48
NOTE: Select a different trigger channel for each SDI-12
parameter that will be used to trigger sampling.
Programming: SDI-12 Parameter Driven Sampling
Page 49
3. Highlight an SDI-12 CHANNEL to configure and press the
Enter key.
NOTE: Select a different input channel for each SDI-12
parameter that will be used to trigger sampling. The
input channel selected for each parameter should match
the channel configured for that parameter on the SDI-12
device.
4. Select the appropriate Unit for the channel that is being
configured and press the Enter key. In this example, pH is
configured to be the Channel 1 unit, as pH is configured to be
the channel 1 unit on the EXO3 sonde.
5. Select the Limit Channel to be used.
If a Low Setpoint is selected, any measurement value lower
than the Low Setpoint will trigger samples to be taken. As an
example, if a Low Setpoint of 6 is entered, then sampling will be
triggered once pH lowers to 6.
If a High Setpoint is selected, any measurement value higher
than the High Setpoint will trigger samples to be taken. As an
example, if a High Setpoint of 8 is entered, then sampling will
be triggered once pH increases to 8.
6. Enter the Setpoint value and press the Enter key.
7. The ProSample will request a Deadband be entered.
Deadband is important when using the Event method of
sampling, but Deadband is not relevant with the Continuous
method (any value can be entered since ProSample will
continue to run the program until completion).
If a High Setpoint is entered, Deadband is a value the selected
parameter must drop below before sampling no longer occurs
(Event method only). As an example, if pH 8 is the High Setpoint
and a Deadband of 0.5 is entered, the pH must drop below 7.5
before sampling no longer occurs.
If a Low Setpoint is entered, Deadband is the value the selected
parameter must rise above before sampling no longer occurs
(Event method only). As an example, if pH 6 is the Low Setpoint
and a Deadband of 0.5 is entered, the pH must rise above 6.5
before sampling no longer occurs.
Programming: SDI-12 Parameter Driven Sampling
49
Page 50
8. Trigger Channel is displayed.
To collect samples using the Event method, highlight Event,
then press the Enter key.
To collect samples using the Continuous method, highlight
Start Prg. 1, then press the Enter key.
NOTE: If more than one SDI-12 parameter is set to trigger
sampling, only the continuous method can be used.
9. Press the Back key a few times to return to the Main Menu.
Select Programs g Change. Choose the program you would
like to modify by using the left and right arrow keys. A different
channel must be configured for each SDI-12 parameter that
will be used, but the trigger channel number and the program
number do not need to be the same (e.g. the trigger/input
channel could be #3, but this could be used for program #1).
Press the Enter key to begin modifying the program.
10. If using the Event method (see page 48), highlight Event Time
and press the Enter key.
If using the Continuous method (see page 48), highlight Time
and press the Enter key.
NOTE: If more than one SDI-12 parameter is set to trigger
sampling, only the continuous method can be used.
11. Enter a Sampling Interval. This determines how often a sample
will be collected and is determined by the user based on the
application and sampling requirements. Press the Enter key to
begin modifying the program.
12. The Sample Volume is displayed. This is the amount of sample
that will be collected at each Sampling Interval. The Sample
Volume is determined by the user based on the application and
sampling requirements. Press the Enter key to begin modifying
the program.
50
Programming: SDI-12 Parameter Driven Sampling
Page 51
13. The Bottle Fill Time is displayed.
If using a composite bottle, Bottle Fill Time determines how
long the program will run. As an example, if Sampling Interval
is set to 15 minutes, Sample Volume is set to 100 mL, and Bottle
Fill Time is set to 2 hours, then the ProSample will collect 100
mL of sample every 15 minutes for 2 hours, resulting in a total
of 800 mL of sample collected.
If using discrete sampling bottles (i.e. more than one bottle is
used), Bottle Fill Time determines how many times each bottle
in the program will be filled. As an example, if Sampling Interval
is set to 15 minutes, Sample Volume is set to 100 mL, and
Bottle Fill Time is set to 15 minutes (i.e. the same as Sampling
Interval), then each bottle is only filled once and each bottle will
contain 100 mL of sample. The entire length of the program
is determined by the number of bottles that will be filled (see
steps # 14-16). Therefore, every 15 minutes a different bottle
will be filled. If there are 6 bottles, the entire program length
will be 90 minutes.
As another example of discrete sampling, if Sampling Interval
is set to 15 minutes, Sample Volume is set to 100 mL, and
Bottle Fill Time is set to 45 minutes (i.e. three times longer than
Sampling Time), then each bottle is filled three times and each
bottle will contain 300 mL of sample
14. The Info screen is displayed.
If there is only one bottle installed (composite sampling):
• Select Settings Complete and the ProSample will return to
the Programs menu. Skip to step #18.
If there is a distributor arm and multiple bottles installed
(discrete sampling):
• Select More Settings g Bottle Selection.
t If you would like to fill all bottles in the sampler, select All Bottles. Return to the More Settings menu by pressing the
Back key once, and then select Programming OK. Skip to
step #18.
t If you only want to fill certain bottles, select Bottle Selection and continue with step #15.
Programming: SDI-12 Parameter Driven Sampling
51
Page 52
15. First Bottle is displayed if the ProSample has more than one
bottle installed. Enter the bottle number in which the first
sample should be placed. Press the Enter key.
16. Last Bottle is displayed. Enter the bottle number in which the
last sample should be placed. Press the Enter key.
17. The More Settings menu is displayed. Highlight Programming
OK and press the Enter key. The Programs menu is displayed.
If using the continuous method, the program is automatically
started if the trigger criteria are met (i.e. the program is not
started by going to Program g Start), so there is no need
to continue with step # 18. Up to 3 other SDI-12 trigger
parameters can be configured by following steps # 1-17.
If using the event method, continue with step #18 below.
18. To run the program, highlight Start and press the Enter key.
NOTE: Ensure the instrument is calibrated before running any
programs. Please see the Pump Calibration on page 17.
19. Use the arrows to select the program to run. Press the Enter key.
52
Programming: SDI-12 Parameter Driven Sampling
Page 53
20. Choose when the program should start. Press the Enter key.
NOTE: Ensure the date and time of the unit is correct. See
Configure Key Instrument Settings on page 22.
21. Choose when the program should end. Press the Enter key.
22. The program will start sampling once the Trigger criteria have
been met.
Programming: SDI-12 Parameter Driven Sampling
53
Page 54
Maintenance and Cleaning
Draining the Bottom Sampler Housing
On the ProSample P and ProSample P-12, there is a drain plug in the center of the bottom sampler housing. This drain
can be removed when cleaning the sampler housing, removing any spilled solution, or when draining any ice used to
keep collected samples cold.
Draining the ProSample P and ProSample P-12
Desiccant Replacement
A desiccant cartridge (40 % rel. humidity; item # 0060589) is located inside the ProSample controller to absorb
moisture and prevent corrosion. Over time the desiccant will become saturated with moisture and should be replaced.
Monitor the desiccant color through the clear plastic window. The color will change from blue to pink when the
desiccant is saturated.
54
Desiccant replacement
Maintenance and Cleaning
Page 55
Peristaltic Pump Tube Replacement
Inspect and clean the pump tubing and rollers on a regular basis. Replace the pump tube (item # 0901062) when it has
become dirty or worn out. Cleaning and/or replacing the pump tube may also prove helpful if the ProSample displays
an error message during a program.
YSI does not have a recommended replacement interval for the peristaltic pump tube, as the lifetime of the tube
varies based on application. As an example, if the sampler is used at the outlet of a plant, the sampling environment is
likely relatively clean with little/no solids. However, if samples are collected at the inlet, the tube may have a very short
lifetime due to the amount of solids in solution. YSI recommends always having a spare peristaltic pump tube available.
NOTE: The inside of the pump tube can be cleaned with a sponge. Cut a 12x12 mm piece of sponge and go
to DIAGNOSTICS / TEST g COMPONENT TEST g PUMP. You can choose to run the pump forward
(suction) and backward (purging). Moisten the sponge, hold it to a tube end and let it “suck” through the
tube. It works in both directions. Repeat until the hose is clean again.
NOTICE: Use of tubing other than that supplied by the manufacturer may cause excessive wear on mechanical
parts and/or poor pump performance.
There is an extra pump tube included with each ProSample unit. It can be found inside the ProSample lid. Please refer
to Installing the Peristaltic Pump Tubeon page 16 when replacing the peristaltic pump tube.
WARNING: Do NOT attempt to replace the peristaltic pump tube or service the pump while the ProSample
is ON. The moving components of the pump can cause serious injury.
Cleaning
The ProSample should be kept clean, especially the parts exposed to the sample such as the dosing unit, water
sensors, distributor, bottles and suction hose. Failure to do so could result in damage not covered by the warranty.
Clean the interior and exterior of the housing with a damp, lint-free cloth. Commercial household cleaner in water can
be used if needed.
Clean the unit around the distributor arm as required.
NOTICE: Do not manually rotate the distributor arm, as it can cause permanent damage to the drive. If you
need to move the arm, go to Diagnostics/Test g Component Test g Distributor. Enter the bottle
position you would like the arm to move to.
In addition to the peristaltic pump tube, also ensure the suction hose, distributor tube, and the intermediate tube are
clean. Replace if necessary.
Maintenance and Cleaning
55
Page 56
Changing the Fuse(s)
If the ProSample is not operating as expected, it may be necessary to change one of the two fuses installed in the
ProSample – 8A and 2A. To remove the fuse, rotate as outlined in the picture below and pull out. Insert the replacement
fuse and tighten.
Removal of ProSample fuses
Long-Term Instrument Storage
To prepare the ProSample for long-term storage (e.g. over the winter):
1. Close all active programs.
2. Purge the peristaltic pump tube, intermediate tube, and suction hose. From the Main Menu, go to Diagnostics/Test g
Component Test g Pump. Press the right arrow key to run the
pump backwards for purging. Keep the button pressed until all
tubes are empty.
3. Remove all liquids and, if necessary, solid matter from all tubes
and bottle compartments. Clean as needed.
4. Remove the peristaltic pump tube.
5. Remove the battery.
6. Store all items in a cool, dry place.
56
Maintenance and Cleaning
Page 57
Factory Reset
A factory reset can reset all settings to the factory default. Please note that all setting and data will be deleted when
completing a factory reset. To complete a factory reset:
1. Turn OFF the ProSample.
2. Hold the power key and the Back key at the same time.
3. When the instrument beeps, release the power key but keep the Back key pressed.
4. Once the screen initializes and shows the “Load Factory Settings” page, release the Back key and allow the Pro-
Sample to boot.
Factory Reset
57
Page 58
YSIConnect PC Software
YSIConnect is the PC software for the ProSample series of automatic samplers. This software can be used to:
• Control the sampler
• Download data
• Update instrument firmware
The sampler is connected to a PC with the USB cable included with each ProSample.
58
USB port cover on the front of the ProSample
YSIConnect PC Software
Page 59
Replacement Parts and Accessories
Ordering
Telephone: 800 897 4151 (USA)
+1 937 767 7241 (Globally) Monday through Friday, 8:00 AM to 5:00 ET
Fax: +1 937 767 9353 (orders)
Email: info@ysi.com
Mail: YSI Incorporated 1725 Brannum Lane Yellow Springs, OH 45387 USA
Internet: ysi.com
When placing an order please have the following available:
1. YSI account number (if available)
2. Name and phone number
3. Purchase Order or Credit Card number
4. Model Number or brief description
5. Billing and shipping addresses
6. Quantity
Replacement Parts and Accessories
59
Page 60
Replacement Parts and Accessories
ProSample Portable Automatic Samplers
YSI Item
Number
630141
630111
630140
630110
DescriptionNotes
ProSample P portable sampler with plastic housing, peristaltic pump,
microprocessor control, battery operated 12 V/7.5 Ah. Includes battery,
two peristaltic pump tubes, USB cable, Getting Started Guide, and a 5
meter suction hose with screw connection and sinker weight.
ProSample P-12 portable sampler with plastic housing, peristaltic
pump, microprocessor control, battery operated 12 V/7.5 Ah, and
SDI-12 connectivity. Includes battery, two peristaltic pump tubes, USB
cable, Getting Started Guide, and a 5 meter suction hose with screw
connection and sinker weight.
ProSample PM mini portable sampler with plastic housing, peristaltic
pump, microprocessor control, battery operated 12 V/7.5 Ah. Includes
10 L PE composite sampling bottle with lid, battery, two peristaltic pump
tubes, USB cable, Getting Started Guide, and a 5 meter suction hose
with screw connection and sinker weight.
ProSample PM-12 mini portable sampler with plastic housing,
peristaltic pump, microprocessor control, battery operated 12 V/7.5 Ah,
and SDI-12 connectivity. Includes 10 L PE composite sampling bottle
with lid, battery, two peristaltic pump tubes, USB cable, Getting Started
Guide, and a 5 meter suction hose with screw connection and sinker
weight.
Battery charger not
included, optional sample
bottle sets with distributor
arm sold separately
Battery charger not
included, optional sample
bottle sets with distributor
arm sold separately
Battery charger not
included
Battery charger not
included
Sample Bottles and Distributor Arm for ProSample P and ProSample P-12
YSI Item
Number
630134Set of (24) 1 L bottles (PE) with lids
630135Set of (8) 2 L bottles (PE) with lids and fixing plate
630136Set of (4) 4 L bottles (PE) with lids and fixing plate
630152Set of (24) 350 mL bottles (glass) with lids and fixing plate
630142Set of (12) 1 L bottles (glass) with lids and fixing plate
630143Set of (8) 2 L bottles (glass) with lids and fixing plate
63011826 L (PE) composite bottle with lid
630112
630113
DescriptionUsed With
ProSample P and
ProSample P-12
ProSample P and
ProSample P-12
ProSample P and
ProSample P-12
ProSample P and
ProSample P-12
ProSample P and
ProSample P-12
ProSample P and
ProSample P-12
ProSample P and
ProSample P-12
Distributor arm for 24 x 1 L PE, 8 x 2 L glass, 12 x 950 mL glass, 24 x
350 mL glass; distributor tube (0901064) is pre-installed
Distributor arm for 4 x 4 L PE and 8 x 2 L PE; distributor tube (0901064)
is pre-installed
Bottle sets 630134,
630143, 630142, and
630152
Bottle sets 630135 and
630136
60
Replacement Parts and Accessories
Page 61
Replacement Parts and Accessories
Flow Signal and SDI-12 Cables
YSI Item
Number
630147
630146
630145-02
630145-04
630145-10
630145-15
630145-33
DescriptionUsed With
10 meter flow signal cable with open cable end for wiring to a digital or
analog device, as well as non-SDI-12 event connections (i.e. a relay)
10 meter SDI-12 cable with open cable end for wiring to an SDI-12
device (e.g. EXO1 and EXO2 with signal output adapter)
2 meter SDI-12 cable with 6-pin female wet-mate connector for direct
connection to YSI EXO3 sonde
4 meter SDI-12 cable with 6-pin female wet-mate connector for direct
connection to YSI EXO3 sonde
10 meter SDI-12 cable with 6-pin female wet-mate connector for
direct connection to YSI EXO3 sonde
15 meter SDI-12 cable with 6-pin female wet-mate connector for
direct connection to YSI EXO3 sonde
33 meter SDI-12 cable with 6-pin female wet-mate connector for
direct connection to YSI EXO3 sonde
Battery Chargers
YSI Item
Number
630137
630122
630144
630153International adapter kit
DescriptionUsed With
Battery charger, IP-67 (waterproof) rated, for US customers, CEC
compliant
Battery charger, IP-67 (waterproof) rated, for non-US customers, does
not include international adapters
Battery charger, not IP-67 (waterproof) rated, for non-US customers,
does not include international adapters
All ProSample models
ProSample P-12 and
ProSample PM-12
ProSample P-12 and
ProSample PM-12
ProSample P-12 and
ProSample PM-12
ProSample P-12 and
ProSample PM-12
ProSample P-12 and
ProSample PM-12
ProSample P-12 and
ProSample PM-12
All ProSample models
All ProSample models
All ProSample models
630122 and 630144
chargers
Replacement Parts and Accessories
61
Page 62
Replacement Parts and Accessories
Accessories
YSI Item
Number
0030051Strainer 8 x 8 mm
0901025Strainer 2 x 2 mm
0069810
0901074Lockable latches - Qty of 3
0901072
0901073
0900045Suspension Bar for harness
0901097Transport Trolley
DescriptionUsed WithNotes
Y-Cable for continuous connection of the
ProSample battery to AC power. Only
recommended if AC power source near location
of ProSample. This allows the battery to receive
continuous charge.
Suspension harness for ProSample PM and
ProSample PM-12
Suspension harness for ProSample P and
ProSample P-12
All ProSample
models
All ProSample
models
All ProSample
models
All ProSample
models
ProSample PM and
ProSample PM-12
ProSample P and
ProSample P-12
All ProSample
models and
suspension
harnesses
ProSample P and
ProSample P-12
Only recommended if
there is an AC power
source near the location of
the ProSample. This allows
the ProSample to always
be connected to AC power.
One lock per latch
Ideal for use in sewers
Replacement Parts
YSI Item
Number
630132-05
630132-10
630132-20
630125
0050598Stainless steel sinker weight, length 180 mm
0069793USB cable - USB2.0 to USB Mini
DescriptionUsed WithNotes
5 meter suction hose with screw connection and
sinker weight
10 meter suction hose with screw connection
and sinker weight
20 meter suction hose with screw connection
and sinker weight
SDI-12 connector to ProSample, no cable
included
All ProSample
models
All ProSample
models
All ProSample
models
ProSample P-12
and ProSample
PM-12
All ProSample
models
All ProSample
models
Included with all
ProSample models
Same as 630146 but
without the cable
Pre-installed in all YSI
suction hoses
Included with all
ProSample models
62
Replacement Parts and Accessories
Page 63
Replacement Parts and Accessories
Replacement Parts
YSI Item
Number
0901055
0060589Desiccant cartridge
630119Fuse T 8A 5x20
630120Fuse 2A
DescriptionUsed WithNotes
Battery pack, 7.5 Ah with connection cable and
fuse holder
All ProSample
models
All ProSample
models
All ProSample
models
All ProSample
models
Included with all
ProSample models
Replace the desiccant
cartridge installed inside
the ProSample controller
when it changes from
blue to pink.
There are two fuses
installed in the
ProSample. Please refer
to the Changing the
Fuse(s) section on page
56 to ensure the correct
replacement fuse is
ordered.
There are two fuses
installed in the
ProSample. Please refer
to the Changing the
Fuse(s) section on page
56 to ensure the correct
replacement fuse is
ordered.
0901101Battery retaining band
0901062Peristaltic pump tube
Intermediate tube; does not include the 630126
0901063
adapter that allows connection of the suction
hose
Replacement Parts and Accessories
All ProSample
models
All ProSample
models
All ProSample
models
Replacement for the
retaining band already
installed on all ProSample
units.
An extra tube is supplied
with each ProSample.
Replace the pump tube
when it has become dirty
or worn out.
This tube is located
between the sensors on
the peristaltic pump.
63
Page 64
Replacement Parts and Accessories
Replacement Parts
YSI Item
Number
0901064Distributor tube
630126
630127
DescriptionUsed WithNotes
Adapter, male 3/4 inch-14BSPP to barbed 10
mm ID; this adapter is installed at the end of
intermediate tube and allows for connection of
the intermediate tube to the suction hose
Adapter, female 3/4 inch-14BSPP to barbed 3/8
inch ID; this adapter is installed at the end of
the suction hose and allows for connection of
the suction hose to the intermediate tube
All ProSample
models
All ProSample
models
All ProSample
models
Pre-installed in the
630112 and 630113
distributor arms; connects
to the peristaltic pump;
for the ProSample PM
and ProSample PM-12,
this tube can be used as
a replacement composite
tube, although the user
may want to cut it to the
appropriate length.
This adapter is preinstalled in the
intermediate tube
included with each
ProSample unit but it is
NOT included with the
replacement intermediate
tube (0901063).
This adapter is preinstalled in each suction
hose. This adapter is
helpful if the adapter on
the suction hose breaks
or if using a suction hose
that is not sold by YSI.
There are two tube
connectors located on
the outside of the pump
housing. ProSample
units ship with the V2A
connectors.
There are two tube
connectors located on
the outside of the pump
housing. ProSample
units ship with the V2A
connectors.
Page 65
Replacement Parts and Accessories
Replacement Parts
YSI Item
Number
0060584Replacement bottle, 1 L, PE
630150Replacement cap for 1 L bottle
0060636Replacement bottle, 2 L, PE
0060634Replacement bottle, 4 L, PE
0030052Replacement bottle, 350 mL, Glass
0060628Replacement cap for 350 mL bottle, PE white
0030054Replacement bottle, 950 mL, Glass
0060640Replacement cap for 950 mL bottle, PE white
0030013Replacement bottle for 2 L, Glass
0060161Replacement cap for 2 L, PE white
Description
630151Replacement 10 L (PE) bottle with lid
Replacement Parts and Accessories
65
Page 66
Warranty and Liability
The manufacturer warrants that the product supplied is free of material and manufacturing defects and undertakes the
obligation to repair or replace any defective parts at zero cost.
The warranty period is 12 months from shipment date. Consumables and damage caused by improper handling, poor
installation or incorrect use are excluded from this warranty.
With the exclusion of the further claims, the supplier is liable for defects including the lack of assured properties
as follows: all those parts that, within the warranty period calculated from the day of the transfer of risk, can be
demonstrated to have become unusable or that can only be used with significant limitations due to a situation present
prior to the transfer of risk, in particular due to incorrect design, poor materials or inadequate finish will be improved
or replaced, at the supplier’s discretion. The identification of such defects must be notified to the supplier in writing
without delay, however at the latest 7 days after the identification of the fault. If the customer fails to notify the supplier,
the product is considered approved despite the defect. Further liability for any direct or indirect damages is not
accepted.
If instrument-specific maintenance and servicing work defined by the supplier is to be performed within the warranty
period by the customer (maintenance) or by the supplier (servicing) and these requirements are not met, claims for
damages due to the failure to comply with the requirements are rendered void.
Any further claims, in particular claims for consequential damages cannot be made.
Consumables and damage caused by improper handling, poor installation or incorrect use are excluded from this
warranty.
Technical Support
Telephone: 800 897 4151 (USA)
+1 937 767 2762 (Globally)
Monday through Friday, 8 AM to 5 PM ET
Email: info@ysi.com
Service Information
When returning the ProSample for service, include the YSI Product Return Form with cleaning certification. The
form must be completely filled out for the YSI Service Center to accept the instrument for service. The form may be
downloaded from YSI.com.
NOTE: When sending a ProSample to YSI for service, it is recommended to include the battery.
66
Warranty and Liability / Technical Support / Service Information
Page 67
Appendix A: ProSample Menu Tree
PROGRAMS
InfoDisplay of program details
Status / Stop
Start
ChangeProgram No [xx]
PauseInterruption of the running program (max. 120 min)
StopStop the current program or all programs
ImmediatelyStart program immediately
Date / TimeStart program with date / time (dd:mm:yyyy hh:mm)
Weekday / TimeStart program with weekday / time (day; hh:mm)
Change operating modes:
• TIME
• FLOW DIGITAL
• FLOW ANALOG
• EVENT TIME
• EVENT DIGITAL
• EVENT ANALOG
• VARIABLE TIME
• BATCH SAMPLING
MANUAL SAMPLE
In Present BottleSample extraction into PRESENT bottle
Into Bottle XSample extraction into selectable bottle X
DATA MEMORY
Sampling DataLog of sampler data (e.g. bottle changes and limit values)
Temperature DataTemp. sampling compartment. Temp. Ambient, PT1000 Temp.
Bottle ProtocolData of each bottle (e.g. requested/taken samples)
Bottle ArchiveArchive of the last 50 Bottle Protocols for the last 50 program cycles
Appendix A: ProSample Menu Tree
67
Page 68
Appendix A: ProSample Menu Tree
DIAGNOSTICS / TEST
Pump
Component Test
Digital Inputs
Analog Inputs
Version Info
Perform a functional check of componentsDistributor
Digital Outputs
Display of
(DI - digital input):
Flow digital: 0 Event: 0
DI3 DI4 DI5
DI6 DI7 DI8
Display of:
• Analog 1
• Water Sensor 1
• Water Sensor 2
• PT 1000 Sensor (Option)
• Operating Voltage
• Analog 2
• Flow: xxxx l/s (m3/h)
Display of:
• Firmware version
• Serial No. PLC
SETUP
Date / TimeSettings of date / time
Trigger ChannelConfigure analog and SDI-12 input channels
Language
Distributor
System Settings
Max. Suction Time
Pre purge
Post purge
Setting of the language.
Default is English.
Selection of the distributor type.
Default is single bottle.
Setting of the maximum suction time (0-600 sec.).
Default is 60 seconds.
Pre purge = purge of suction hose PRIOR to the sample extraction
(0-99.99 sec.). Default is 5 seconds.
Post purge = purge of the suction hose AFTER the sample extraction
(0-99.99 sec.). Default is 5 seconds.
68
Appendix A: ProSample Menu Tree
Page 69
Appendix A: ProSample Menu Tree
System Settings
(continued)
Rinse Before Sampling
Calibration VolVolume calibration for Peristaltic Pump
Log Entries
Internal Temperature
Fault SignalPossibility to invert the output
Analog Signal
Display
Option to rinse intake line with source liquid prior to each sample,
up to 3 rinses. Default is no rinses.
Set of the Log entries. Log interval for Temp. board and PT1000 can
be adjusted 1-60 min
• Via NTC
• Via PT1000
• Limit value (1-20°C)
• Delay time (1-60 min.)
(Example: limit value 7°C, delay time 10 min. An alarm message is
sent if the limit value is exceeded for 10 min.)
Selection:
• 0-20 mA
• 4-20 mA
Calibration (adjustment with the signal of plant)
• Always switched on
• Switch off after certain time (0-999 sec.)
• Contrast
• Max. brightness
• Min. brightness
Sleep Mode
Status LEDEnable or disable the LED indicator on the front of the ProSample.
Program can be interrupted for 10-120 min. for example for
Pause Duration
Prog. Inputs
Min. Sample Volume
Max. Sample Volume
Active
InactiveSleep mode is deactivated
cleaning. After expiry of the time entered the program is
automatically resumed.
Programmable inputs: input signal to start a program e.g. via an
external pulse.
Four inputs are available.
Setting of the min. sample volume which shall be taken.
Default is 1 mL.
Setting of the max. sample volume which shall be taken.
Default is 250 mL.
If the sleep mode has been activated and the program is to be
started in 20 min. at the earliest, the message “Attention device
switches to sleep mode” is displayed for 30 sec.
Thereafter the display is switched off and only activated again 2 min.
prior to the program start.
Appendix A: ProSample Menu Tree
69
Page 70
Appendix A: ProSample Menu Tree
Change PasswordIn general the password can be changed
Change SettingsA password for settings can be entered
Password
Change ProgramsA password for changing programs can be entered
Stop ProgramsA password for stopping programs can be entered
Service
Setting of base parameters (only to be done by a service technician)
(Password protected)
Item# 630123REF
Revision A
June, 2018
70
Appendix A: ProSample Menu Tree
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.