General Maintenance ....................................................................... 58
THIS PAGE LEFT INTENTIONALLY BLANK
WAR R A N T Y
The YSI Professional Plus Instrument (Pro Plus) is warranted for three (3) years
from date of purchase by the end user against defects in materials and workmanship,
exclusive of batteries and any damage caused by defective batteries. Pro Plus field
cables are warra nted for two (2) years from date of purchase by the end user against
defects in material and workmanship (6 months for non-field rugged cables*). Pro
Plus sensors (pH, ORP, pH/ORP combo, Polarographic DO) are warranted for
one (1) year from date of purchase by the end user against defects in material and
workmanship (6 months for ammonium**, nitrate**, chloride**, and Galvanic DO).
Pro Plus systems (instrument, cables & sensors) are warranted for 90 days from
date of purchase by the end user against defects in materia l and workmanship when
purchased by rental agencies for rental purposes. Within the warranty period, YSI
will repair or replace, at its sole discretion, free of charge, any product that YSI
determines to be covered by this warranty.
To exercise this warranty, call your local YSI representative, or contact YSI Customer
Service in Yellow Springs, Ohio at +1 937 767-7241, 800-897-4151 or visit www.
YSI.com (Support tab) for a Product Return Form. Send the product and proof
of purchase, transportation prepaid, to the Authorized Service Center selected by
YSI. Repair or replacement w ill be made and the produc t returned, transportation
prepaid. Repaired or replaced products are warranted for the balance of the origina l
warranty period, or at least 90 days from date of repair or replacement.
LIMITATION OF WARRANTY
This Warranty does not apply to any YSI product damage or failure caused by:
failure to instal l, operate or use the product in accordance with YSI’s written 1.
instructions;
abuse or misuse of the product; 2.
failure to maintain the product in accordance with YSI’s written instructions or 3.
standard industry procedure;
any improper repairs to the product; 4.
use by you of defective or improper components or parts in servicing or 5.
repairing the product;
modification of the product in any way not expressly authorized by YSI.6.
THIS WARRANTY IS IN LIEU OF ALL OTHER WARRANTIES, EXPRESSED OR
IMPLIED, INCLUDING ANY WARRANTY OF MERCH ANTABILITY OR FITNESS
FOR A PARTICULAR PURPOSE. YSI’s LIABILITY UNDER THIS WARRANTY
IS LIMITED TO REPAIR OR REPLACEMENT OF THE PRODUCT, AND THIS
SHALL BE YOUR SOLE AND EXCLUSIVE REM EDY FOR A NY DEFECTIVE
PRODUCT COVERED BY THIS WAR RANTY. IN NO EVENT SHALL YSI BE
LIABLE FOR ANY SPECIAL, INDIRECT, INCIDENTAL OR CONSEQUENTIAL
DAMAGES RESULTING FROM ANY DEFECTIVE PRODUCT COVERED BY
THIS WARRANTY.
i
ii
1
* The warranty period for the non-field rugged cables (605107, 605177, 605108,
605178, 605109, 605179) is listed as 6 months. However, the true “working life” of
these sensors may be 3-9 months depending on storage and usage in solutions other
than clean aqueous samples.
** The warrant y for the ammonium, nitrate, and chloride sensors (605104, 605105,
605106) is listed as 6 months. However, the true “working life” of these sensors may
be 3-9 months depending on storage and usage in solutions other than clean aqueous
samples.
INTRODUCTION
Thank you for purchasing the YSI Professional Plus (Pro Plus). The Pro Plus
features a waterproof (IP-67) case, backlit display and keypad, user-selectable
cable options, USB connectivity, large memory with extensive site list capabilities,
and a rugged, rubber over-molded case.
Reading the entire manual before use is recommended for an overall
understanding of the instrument’s features.
GETTING STARTED
InItIal InspectIon
Carefully unpack the instrument and accessories and inspect for damage.
Compare received parts with items on the packing list. If any parts or materials
are damaged, contact YSI Customer Service at 800-897-4151 (+1 937 767-7241)
or the authorized YSI distributor from whom the instrument was purchased.
Battery InstallatIon
The Pro Plus requires (2) alkaline C-cell batteries which are included with the
purchase of a new instrument. Battery life depends on parameters and usage.
Under normal conditions, battery life is approximately 80 hours for continuous
use at room temperature. To install or replace the batteries:
Turn the instrument over to view the battery cover on the back. 1.
Unscrew the four captive battery cover screws. 2.
Remove the battery cover and install the new batteries, ensuring correct 3.
polarity alignment on the instrument or the removed cover. (Figure 1)
Replace the battery cover on the back of the instrument and tighten the four 4.
screws. Do NOT over-tighten.
Figure 1. Pro Plus with battery cover removed. Notice
battery symbols indicating polarities.
2
3
Batteries must be installed in the instrument even if powering
the unit via the USB connection. This will retain the correct
i
NOTE - On subsequent battery changes you will have approximately 2 minutes
to change the batteries before the clock resets. If the clock resets, the instrument
will automatically bring up the Date/Time menu the next time it is powered on
in order to update this information. This is important, especially if you intend
to log data!
date and time if the PC is turned off. If the USB power is
disconnected and there are no batteries in the instrument, the
date and time will need to be reset upon subsequent power on.
setup
The Pro Plus instrument has several compatible field-rugged cable/sensor
options, each with temperature:
Cable: Available Sensors:
Cable number 60520-x DO/temp (605780 for lab BOD)
Cable number 60530-x Conductivity/temp
Cable number 60510-x ISE*/temp
Cable number 6051010-x ISE*/ISE*/temp
Cable number 6051020-x ISE*/DO/temp
Cable number 6051030-x ISE*/conductivity/temp
Cable number 6052030-x DO/conductivity/temp
Cable number 605790-x DO/conductivity/ISE*/ISE*/temp (Quatro**)
*ISE (Ion Selective Electrode) notates a port that can accept pH, ORP, Ammonium,
Nitrate, Chloride, and, in some cases, a pH/ORP combination sensor.
recommended for lab use or controlled conditions where a more rugged, field
cable is not necessary. These cables include:
Cable number 605107 1-meter cable; single-junction pH sensor
Cable number 605177 4-meter cable; single-junction pH sensor
Cable number 605108 1-meter cable; single-junction ORP sensor
Cable number 605178 4-meter cable; single-junction ORP sensor
Cable number 605109 1-meter cable; single-junction pH/ORP sensors
Cable number 605179 4-meter cable; single-junction pH/ORP sensors
STANDARD PRO SERIES SENSOR INSTALLATION
Throughout the manual, the term “sensor” refers to the removable portion or
electrode sensing portion of the cable assembly. For example, the DO sensor or
pH sensor is the part that can be removed from a field cable and replaced with a
new sensor. The conductivity sensor is not removable from a non-Quatro cable
but still refers to the “sensing” portion and will be referred to as a sensor. This
section covers most of the sensor installations on a Professional Series cable
bulkhead including the following sensors:
2003 - Polarographic
DO (black)
2002 - Galvanic
DO (gray)
See the next section of this manual for installation instructions for the Quatro
cable’s Conductivity/Temperature sensor.
1001 - pH1003 - pH/ORP 1005 - Chloride
1002 - ORP1004 - Ammonium 1006 - Nitrate
**Cable 605790 will be referred to as a Quatro cable throughout this manual.
All cables come in standard lengths of 1, 4, 10, 20, and 30-meters (3.28, 13,
32.8, 65.6, and 98.4-feet) with options for special order lengths up to 100-meters
(328-feet) on the 60520-x cables. Contact YSI or your local representative for
additional information.
In addition there are several cable options with built in sensors for the
measurement of pH and ORP that are not considered field-rugged (nonreplaceable sensors, less rugged single-junction sensors). These cables are
Dual sensor bulkhead ports are numbered 1
and 2, see figure to the left. Please refer to the
following tables to determine correct sensor
installation into each port of a two port cable.
4
5
Port 1 OptionsPort 2 Options
pHpH
ORPORP
1010 dual cable
* If using a 6051010 cable, a sensor must be installed in port 1 for correct
operation. If you install a pH/ORP combo sensor into a 6051010 cable, ORP
will not be measured. It is not recommended to use a pH/ORP combo sensor
on a 6051010 cable.
1020 dual cable
If using a 1020 cable, install a pH, ORP, pH/ORP, Ammonium, Nitrate or
Chloride sensor in port 1 and a DO sensor in port 2.
pH or pH/ORP*pH or pH/ORP*
ammoniumammonium
chloridechloride
nitratenitrate
none (port plug)
Port 1 OptionsPort 2 Options
pHPolarographic DO
ORPGalvanic DO
pH or pH/ORPnone (port plug)
ammonium
chloride
nitrate
none (port plug)
Port 1
Options
pHpHPolarographic DO5560
Quatro
Cable (pn:
605790)
* If using a Quatro cable, a sensor must be installed in port 1 for correct
operation of port 2. If you install a pH/ORP combo sensor into a Quatro
cable, ORP will not be measured. It is not recommended to use a pH/ORP
combo sensor on a Quatro cable.
i
ORPORPGalvanic DO
pH or pH/
ORP*
ammoniumammonium
chloridechloride
nitratenitrate
Before installing either dissolved oxygen sensor, the instrument
must be configured for the sensor being installed. See
the Setup - Dissolved Oxygen section of this manual for
instrument configuration instructions. Failure to do this
may result in damage not covered under warranty.
Port 2
Options
pH or pH/
ORP*
none (port
plug)
DO Port
Options
none (port
plug)
CT Port Options
Conductivity/
Temperature
sensor only
i
If using a 605103 pH/ORP combination probe on a
6051020 or 6051030 cable you can report both pH
and ORP. However, it is recommended to set ISE1 as
pH and ISE2 as ORP in the Sensor Setup menu.
The Quatro cable bulkheads are labeled
1, 2, DO, and CT, see figure to the left. All
sensors except the Conductivity/Temperature
sensor can be installed following the Standard
Pro Series Sensor Installation instructions.
Conductivity/Temperature sensor installation
is described in the next section. For ease of
installation, YSI recommends that you install
a sensor into port 1 first; followed by DO
installation, then port 2, and lastly C/T.
First, ensure both the sensor connector and sensor port on the cable are
clean and dry. To connect the sensor, grasp the sensor with one hand and the
sensor connection end of the cable (bulkhead) in the other. Push the sensor
into the connector on the cable until it is properly seated and only one o-ring
is visible. Failure to properly seat the probe may result in damage. Twist the
sensor clockwise to engage threads and finger tighten (Figure 2). Do not use a
tool. This connection is waterproof. Please refer to the sensor installation sheet
that is included with each sensor for detailed instructions.
6
7
Figure 2. The image on the left shows a clean, dry sensor being
aligned with the bulkhead. On the right, the sensor has been
pushed into the bulkhead and is being screwed into place.
Figure 3. The sensor(s) will install directly
in the cable bulkhead. Once installed, the
sensor guard will protect the sensor during
sampling (DO cap membrane not shown).
Align the connectors of the new sensor and the port. With connectors 3.
aligned, push the sensor in towards the bulkhead until you feel the sensor
seat in its port. You will experience some resistance as you push the sensor
inward, this is normal
Once you feel the sensor seat into the port, gently rotate the stainless steel 4.
sensor nut clockwise with your fingers, Do not use the tool.
The nut must be screwed in by hand. If the nut is difficult to turn, STOP, as 5.
this may indicate cross threading. If you feel resistance or cross threading
at any point, unscrew the nut and try again until you are able to screw the
nut down completely without feeling any resistance. Damage to your cable/
sensor may occur if you force the parts together.
Once completely installed, the nut will seat flat against the bulkhead. At 6.
this point, use the tool that was included with the sensor to turn the nut an
additional ¼ to ½ turn so it cannot come loose (figure 4). DO NOT over
tighten.
Do not cross thread the sensor nut. Seat nut on
i
face of bulkhead. Do not over tighten.
Please refer to the sensor installation sheet that is included with the conductivity/
temperature sensor for detailed instructions.
Sensor Guard
Sensor
Cable Bulkhead
CONDUCTIVITY/TEMPERATURE SENSOR INSTALLATION IN A
QUATRO CABLE
As mentioned, the installation of the Conductivity/Temperature sensor (model
5560) in a Quatro cable is different from all other Pro Series sensor installations.
Follow these instructions when installing a conductivity/temperature sensor in
a Quatro cable:
Locate the C/T port and, if replacing, remove the old sensor using the 1.
installation tool to loosen the stainless steel retaining nut. Once the stainless
steel retaining nut has been completely unscrewed from the bulkhead,
remove the old sensor from the bulkhead by pulling the sensor straight out
of the bulkhead.
Apply a thin coat of o-ring lubricant (supplied with the sensor) to the o-rings 2.
on the connector side of the new sensor.
i
Visually inspect the port for moisture. If moisture is found,
it must be completely dried prior to sensor installation.
Figure 4. Installation tool used to
tighten stainless steel retaining nut of
5560 conductivity/temperature sensor.
INSTALLING PORT PLUGS IN UNUSED PORTS
As necessary, install a port plug into any port that does not have an installed
sensor. This will protect the bulkhead from water damage. Port plugs and a
tube of o-ring lubricant are included with all Quatro cables. These items can be
ordered separately if needed. To install a port plug, apply a thin coat of o-ring
lubricant to the two o-rings on the port plug. After application, there should
be a thin coat of o-ring lubricant on the o-rings. Remove any excess o-ring
lubricant from the o-ring and/or port plug with a lens cleaning tissue. Next,
insert the plug into an empty port on the bulkhead and press firmly until seated.
Then, turn the plug clockwise to engage the threads and finger-tighten until the
plug is installed completely. Do not use a tool to tighten the plug.
8
9
CONNECTING THE CABLE TO AN INSTRUMENT
To connect a cable, align the keys on the cable connector to the slots on the
instrument connector. Push together firmly, then twist the outer ring until it
locks into place (figure 5). This connection is water-proof.
Keypad
2
3
Figure 5. Note the keyed connector. The cable and instrument
connectors can only be mated once the keys are properly aligned.
When a sensor is not installed, the sensor and cable sensor
connectors are NOT water-proof. Do not submerge the cable
without a sensor or port plug installed in all available ports.
i
When the cable is disconnected, the cable’s instrument connector
and the connector on the instrument maintain an IP-67 rating.
SENSOR STORAGE
The cable assembly is supplied with a storage container, or sleeve, that installs
on to the cable. The container is used for short-term storage (less than 30 days).
Be sure to keep a small amount of moisture (tap water) in the container during
storage. This is done to maintain a 100% saturated air environment which is
ideal for short-term sensor storage (see Care, Maintenance, and Storage for more
detailed information). Do not submerge the sensors in an aqueous solution.
The intent is to create a humid air storage environment.
1
13
12
11
10
9
NumberKeyDescription
1
Opens System Menu from any screen.
2
Opens Sensor Menu from any screen.
Use to enable sensors and display units.
3
Opens Calibrate Menu from any screen.
Use to calibrate all parameters except temperature.
4
Use to view data and GLP files, set up
5
Press to turn the instrument backlight on and
off and to adjust the display contrast when
pressed with the left or right arrow key.
4
5
6
7
8
System
Use to adjust system settings.
Sensor
Calibrate
File
Opens File Menu from any screen.
site and folder lists, and delete data.
Backlight
10
11
NumberKeyDescription
6
Use to navigate right in alpha/numeric entry
screens. Can be pressed simultaneously with
Backlight key to increase display contrast.
7
Use to navigate through menus and to navigate
down in alpha/numeric entry screens.
8
P re s s t o tu rn th e i n st r u me nt on .
Press and hold for 3 seconds to turn off.
9
Press to receive hints & tips during operation.
10Enter
Press to confirm selections, including
alpha/numeric key selections.
11
Use to navigate left in alpha/numeric entry screens.
Press to return to previous menu in all
screens except alpha/numeric entry.
Can be pressed simultaneously with Backlight
key to decrease display contrast.
12
When in alpha/numeric entry screen,
13
Use to navigate through menus and to navigate
up in alpha/numeric entry screens.
Right Arrow
Down Arrow
Exits back to Run Screen.
escapes to previous menu.
Power
Help
Left Arrow
Exit/Escape
Up Arrow
MaIn dIsplay
Press the Power key to turn the instrument on. The instrument will briefly
display the splash screen with the YSI logo then go directly to the main run
screen. The first time the instrument is powered up or if the instrument
has had a battery change (with batteries removed for more than 2 minutes),
you will need to set the date and time. Follow the instructions under
System Menu | Date/Time.
The display at the left shows the run mode (main
display) with temperature in °C, barometer
in mmHg, DO in % and mg/L, and pH as the
reported parameters. The date, time and battery
level are indicated at the bottom of the screen.
The logging preference of Log One Sample at a
time is indicated at the top of the screen.
This screen also shows the message line towards
the bottom of the display above the date and
time. In this case, it doesn’t show a message
but messages will appear frequently to indicate
calibration steps, set date and time, etc.
A USB symbol
bottom of the display when connected to a PC
through USB with the communications saddle. The instrument will display full
battery power when it is receiving power through the USB connection.
i
Contrast – the contrast adjustment can be accomplished by pressing
the backlight key and the left or right arrow key at the same time.
will show up on the
Menu layout
Press Esc at anytime in the menus to escape back to the Run screen. The
left arrow
alpha/numeric entry screens. You must use Esc to get out of the alpha/numeric
screens if you want to exit before finishing or without saving changes. Functions
that are enabled appear as a circle with a dot or a box with a check mark Q.
Disabled functions appear as a circle only or an empty P.
can be used to go back to the previous menu in all screens except
12
13
alpha/nuMerIc entry
The numeric screens will display numbers only (shown on the left). Alpha/
numeric screens will display numbers across the top and letters along the
bottom rows (shown on the right). Letters appear as a common keyboard
arrangement.
When an alpha or numeric character is required, the screen will show the alpha/
numeric entry screen. To select a character, highlight it by using the arrow keys
to move the highlight box over the desired selection. Then, press Enter on the
keypad to confirm the selection. After confirming the selection, it will appear
in the line at the top of the display.
For capital letters or lower case entry, highlight “SHIFT” and press Enter on the
keypad to change the characters from upper to lower case.
To delete the entire line of the current entry, highlight ) and press Enter on the
keypad. The ! symbol functions as a backspace key in the alpha/numeric entry
screens by deleting one character at a time. Use the “SPACE” function to add a
space between characters.
When you have finished entering the correct information (16 character max),
highlight <<<ENTER>>> at the bottom of the screen and press Enter
the keypad to confirm.
key cannot be used to escape to the previous menu from
key to go
i
The
an alpha/numeric entry screen. Instead, use the
back to the previous menu when in alpha/numeric entry screens.
on
SYSTEM MENU
date/tIMe
The date and time will need to be reset if a battery
i
change takes longer than 2 minutes. When this occurs,
the Date/Time menu will automatically appear upon
power up and require you to set the date and time.
Press System to access any of the following
menu items.
The System menu will allow you to access the
setup options of the instrument including; Date/
Time, GLP, Language, Radix Point, Logging,
Auto-Shutoff, Backlight, SW (Software)
Ve rs i on , Serial #, and Unit ID. Any item with
[brackets] shows the current setting inside the
brackets. For instance, in the example at the
left, Radix Point is currently set to [Decimal].
The brackets will also give a quick visual clue as
to what items can be changed.
Highlight Date/Time from the System menu.
Press enter to select.
Date Format – Highlight and press enter to
open a sub menu for selecting the preferred
date format: YY/MM/DD, MM/DD/YY,
DD/MM/YY, or YY/DD/MM.
Date – Highlight and press enter to use the
numeric entry screen to set the correct date.
Time Format – Highlight and press enter to
open a submenu to select the preferred time
format from 12-hour or 24-hour.
Time – Highlight and press enter to use the
numeric entry screen to set the correct time.
14
15
Glp
The GLP or ‘Good Laboratory Practice’ file saves detailed information
about calibrations. It also includes diagnostic information about the sensors.
Calibrations are logged into a file, the GLP, for later review as needed. A single
GLP file is utilized to store all calibration records and is capable of storing 500
records. Once the GLP file is full, the instrument will begin to overwrite the
oldest record with each new calibration record.
In order to keep all of your GLP records, periodically download
the GLP to Data Manager and export it to another program.
Otherwise, the unit will overwrite the oldest record once
i
Several calibration parameters are saved for each calibration record including
optional ones that can be enabled by the user. Standard parameters include
date/time stamp, calibration method, and sensor information. Optional, user
selectable parameters include User ID, Probe ID, and User Fields 1 and 2.
The sensor specific information that is saved with each calibration point is
different for each sensor. The sensor specific values saved are:
the memory is full. Also, since Data Manager saves GLP
files under the Unit ID, you must periodically export and
rename the GLP file on your PC or it will be overwritten
each time you upload the GLP file from the instrument.
DO
Method (%, mg/L)
Cal Value
Sensor Value (Sensor Current)
Sensor Type (Polarographic/Galvanic)
Membrane Type (Teflon Black, PE Yellow, PE Blue)
Salinity Mode (user entered value if in Manual Salinity Mode)
Temperature
Barometer
Calibrate Status
pH (up to 6 calibration points)
Buffer Value
Sensor Value (mV)
Temperature
Slope (mV/pH)
Slope (% of ideal)
Calibrate Status
ORP
Cal Solution Value
Sensor Value
Temperature
Calibrate Status
Conductivity
Method (Spec Cond, Cond, Salinity)
Cal Value (value of calibration solution)
Sensor Value (Cell Constant)
Temperature Reference (User selected in Sensor Setup menu)
Temperature Compensation Coefficient %/°C (User selected in Sensor Setup
menu)
TDS Constant (User selected in Sensor Setup menu)
Temperature
Cal Cell Constant
Calibrate Status
Ammonium
Buffer Value
Sensor Value (mV)
Temperature
Calibrate Status
Chloride
Buffer Value
Sensor Value (mV)
Temperature
Calibrate Status
Nitrate
Buffer Value
Sensor Value (mV)
Temperature
Calibrate Status
16
17
An example of a GLP record
(Operation performed is single point % DO Calibration)
*** Calibrate – DO% ***
Date 02/03/09 MM/DD/YY
Time 12:14:57PM 12-hour
User ID: Tech 1
Probe ID 08D
Method DO Air Calibrate
Cal Value: 100.00%
Sensor Value: 5.175155uA
Sensor Type Polarographic
Membrane Type 1.25 PE Yellow
Salinity Mode 5.175165 Auto
Temperature 23.9 °C
Barometer 731.4 mmHg
Calibrate Status Calibrated
GLP SETTINGS
In the System menu, highlight GLP and press
enter to view and modify the GLP settings.
assembly from another, typically by serial number. Highlight Include Probe ID
and press enter to turn this function on (Q) and off (P). Highlight Probe ID
and press enter to add, view, edit, delete, or select a Probe ID. The Probe ID
may also be changed in the Calibration menu during the calibration process.
The character limit is 16 characters.
User Fields 1 and 2 are stored with the calibration record and may be used
to enter other parameters pertinent to the user, such as weather conditions,
elevation, etc. Highlight Include User Field 1 or Include User Field 2 and press
enter to turn this function on and off. Highlight User Field 1 or User Field 2
and press enter to add, delete, view, edit, or select a User Field. The character
limit is 16 characters. When enabled, a prompt for selecting a User Defined
Field will appear during the calibration process.
Re-Cal Prompt may be used to remind the
user to perform a calibration. To set a time
interval, highlight the parameter you wish to
be reminded about and press enter to access the
numeric entry screen. Enter a value in days and
press enter to confirm the reminder time. To
turn off the Re-cal prompt, set the reminder to
zero (0) days (this is the default).
Highlight Options and press enter to access
User ID, Probe ID, User Defined Fields, and
Re-Cal Prompt.
User ID may be used to identify the person
calibrating the instrument. Highlight User ID
and press enter to select, edit, or delete a User
ID from a list of previously entered IDs. Or,
highlight Add New and press enter to create
a new User ID using the alpha/numeric entry
screen. The User ID may also be changed in
the Calibration menu during the calibration
process. The selected User ID will be stored in
the GLP file with each calibration record. A
User ID could be a person’s initials or badge
number. The character limit is 16 characters.
Probe ID is stored with the calibration record
and may be used to distinguish one cable/probe
The Security section of the GLP menu is a password protected area. This area
includes options to set a new password and to lock access to the calibration
menu. When first viewing the security menu, you will be required to enter a
password. Use the “shift” on the alpha/numeric screen to switch to lower case if
necessary and enter “ysi123”. This is the default password.
Protect Cal can be enabled (Q) or disabled (P). When enabled, the user must
know and enter the instrument’s password to enter the calibration menu option.
Highlight Protect Cal and press enter to enable or disable this feature.
Set Password allows a user to set the security password. Highlight Set Password,
press enter, and use the alpha/numeric entry screen to set the new password.
The password can have up to 16 characters.
Contact YSI Technical Support at environmental@ysi.com or +1 937 767-7241 if
you forget or misplace your password.
Once a password is set, and the GLP security screen
i
exited, a password must be entered to make changes
under GLP security. Keep passwords in a safe place.
18
19
lanGuaGe
The Pro Plus can be configured to display all
text in English, Spanish, French, German,
Portuguese, Italian, Norwegian, Simplified
Chinese, Traditional Chinese, or Japanese. From
the factory, the instrument includes English,
Spanish, and French language options. The
other language options can be downloaded from
www.ysi.com/support.
Check the box for Continuous Mode if you want to log samples continuously
at a specific time interval. To set the length of time between logged samples,
highlight Interval and press enter. Enter the interval as HH:MM:SS. This
interval will display at the top of the screen when you select the Start Logging
option in run mode.
To log one sample at a time, uncheck Continuous Mode. When Continuous
Mode is unchecked, Log One Sample will appear at the top of the run screen.
auto shutoff
radIx poInt
loGGInG
Once the appropriate language file is in the
instrument, press System
Language, and press enter. Highlight the desired
language and press enter to confirm.
Radix Point allows the user the option to
choose between a comma or a decimal in
numeric displays. For example, 1.00 becomes
1,00 when Use Comma is selected. Highlight
Use Decimal or Use Comma and press enter to
make your selection.
From the System menu, highlight Logging
and press enter to view or change the logging
options. Logging options include Use Site
List, Use Folder List, Continuous Mode, and
Interval.
, highlight
Auto Shutoff powers the instrument off after a user specified time period.
Highlight Auto Shutoff and press enter. Using the alpha/numeric entry screen,
enter a value between 0 and 360 minutes. To disable auto shutoff, set the value
to 0 (zero).
BacKlIGht
Backlight can be set to Automatic or Manual.
Automatic turns the backlight on when you
turn the instrument on and when you press any
key. Manual allows you to turn the backlight
on or off with the backlight key
in Automatic mode, the instrument will turn
the backlight off 60 seconds after the last key
press. The instrument will “reset” the 60 second
time period every time a key is pressed. The
lighted keypad will turn off after approximately
20 seconds.
. When
sW VersIon (softWare VersIon)
SW Version shows the instrument’s software version. The instrument’s software
can be updated via www.ysi.com/support. There you will find the new software
files and instructions on how to update the instrument. There is no need to send
the instrument back to the factory for upgrades.
Use Site List and Use Folder List are optional
ways of filing or ‘tagging’ your logged data
points. If these settings are enabled, you will be
prompted to select a Site and/or Folder to ‘tag’ to the logged data point. See the
File and Site Lists section of this manual for information on creating Site and
Folder Lists.
serIal #
Serial # shows the instrument’s serial number and allows you to match it with
the number engraved on the back of the instrument’s case.
20
21
unIt Id
Unit ID is used to identify instruments in the Data Manager software program
that was included with your instrument. It is also used to identify GLP files, Site
Lists, Configuration files, and Data files transferred from the instrument to the
PC. The default Unit ID is the instrument’s serial number. To modify the Unit
ID, highlight Unit ID, press enter and then use the alpha/numeric entry screen.
The character limit is 16 characters.
temperature units of °F, °C, or K and press enter to confirm the selection. Only
one temperature unit may be displayed at a time. You may also choose not to
display temperature. If you choose not to display temperature, other parameters
that require a temperature reading will still be temperature compensated.
dIssolVed oxyGen (do)
DO sensors can be used on 60520-X, 6051020-X, 6052030-X, and Quatro
cables.
PA R A M E T E R S : S E T U P, D I S P L A Y ,
AUTO STABLE, AND CALIBRATION
The following section is separated by parameter and will discuss sensor setup,
display options, auto stable features, and calibration procedures for each
parameter. The sections are separated by parameter due to the versatility of the
Pro Plus. You may focus solely on the parameters of your choice.
For the highest accuracy, calibrate or verify each sensor regularly. For your
convenience, YSI offers 5580 Confidence Solution® which allows you to check
the accuracy of pH, conductivity, and ORP readings to help determine if a sensor
calibration is necessary.
If you receive an error message during a calibration that indicates questionable
results, you have the option to either accept or decline the calibration. YSI
recommends that you decline a questionable calibration since accepting it may
result in erroneous data. After declining a questionable calibration, ensure the
sensor is clean, the calibration solution is good, the calibration vessel is clean,
and that you are entering the correct calibration value if entering manually.
Then, try to recalibrate the sensor. If you continue to have problems, see the
Troubleshooting section of this manual.
teMperature
All probe/cable assemblies, except the Quatro,
have a built-in temperature sensor. The Quatro
cable ships with a Conductivity/Temperature
sensor that must be installed on the cable.
Temperature calibration is not required nor is
it available.
PREPARING THE DO SENSOR FOR THE FIRST TIME
The dissolved oxygen sensor is shipped with a dry, protective red cap that will
need to be removed before using. It is very important to put a new membrane
with electrolyte solution on the sensor after removing the red cap.
Prepare the membrane solution according to the instructions on the bottle. After
mixing, allow the solution to sit for 1 hour. This will help prevent air bubbles
from later developing under the membrane. Ensure you are using the correct
electrolyte solution for the correct sensor. Galvanic sensors utilize electrolyte
with a light blue label and Polarographic sensors utilize electrolyte with a white
label. The dissolved oxygen sensor is supplied with cap membranes specific
to the sensor type ordered (Polarographic or Galvanic). 5912, 5913, and 5914
membrane kits are for Galvanic sensors and the 5906, 5908, and 5909 membrane
kits are for Polarographic sensors. See the Setup - Dissolved Oxygen section of
this manual for more information on the different types of membranes available
from YSI.
Remove the red cap by pulling it straight off the sensor tip. Discard or save for
later use during long term storage. Thoroughly rinse the sensor tip with distilled
or deionized water. Fill the cap membrane 3⁄4 full of electrolyte solution, then
tap the cap with a finger to release any trapped air. Be careful not to touch
the membrane portion of the cap. Thread the membrane cap onto the sensor,
moderately tight. Do not use a tool. It’s typical for some of the electrolyte
solution to spill over. For detailed instructions on changing a membrane cap,
see the Care, Maintenance, and Storage section of this manual.
To set the units, press Sensor
Display and press enter. Highlight Temperature
and press enter. Highlight the desired
, highlight
22
23
SETUP DISSOLVED OXYGEN
Press Sensor , highlight Setup and press enter. Next, highlight DO and press
enter.
Enabled allows you to enable or disable the
Dissolved Oxygen function. Highlight Enabled
and press enter to activate(
dissolved oxygen. Disable dissolved oxygen
if you do not have a dissolved oxygen sensor
connected to the instrument.
) or deactivate(P)
Q
IMPORTANT – The instrument default setting is Galvanic.
Please change the Sensor Type to match the correct sensor. If
you observe readings very close to 0 or extremely high readings
u
Membrane sets the type of membrane used on the dissolved oxygen sensor.
Highlight Membrane and press enter. Highlight the correct membrane type
installed on the sensor and press enter to confirm. The DO sensor is supplied
with membranes specific to the sensor type ordered and are color coded as
described in the following tables.
(i.e. 600%), your Sensor Type setting (Polarographic or
Galvanic) may be set incorrectly and you should immediately
ensure it matches the sensor installed on your cable.
If a sensor is Enabled that isn’t connected to the
i
Sensor Type sets the type of oxygen sensor being used: either Polarographic
(black) or Galvanic (grey). Highlight Sensor Type and press enter. Highlight the
correct sensor type installed on the cable and press enter to confirm.
If using a ProBOD sensor/cable assembly, the sensor type should be set to
polarographic.
The Pro Plus has two compatible sensors for use with a field cable:
Polarographic – This sensor has a black sensor body and is engraved with the
model number 2003.
Galvanic – This sensor has a grey sensor body and is engraved with the model
number 2002.
In terms of physical configuration, membrane material, and general performance,
YSI Professional Series Galvanic dissolved oxygen sensors are exactly like the
Professional Series Polarographic sensors. The advantage of using Galvanic
sensors is convenience. Galvanic sensors provide for an instant-on sensor without
the need for warm-up time but this affects the life of the sensor. Polarographic
sensors last longer and have a longer warranty but require a 5-15 minute warmup time before use or calibration.
instrument, the display will show an unstable, false
reading, ?????, or ----- next to the units.
Galvanic membrane kits:
ItemColorMaterialDescription
5912Black1 mil Teflon®Traditional membrane material
5913Ye l l o w1.25 mil polyethylene Improved response time and less
flow dependence than Teflon®
Ships standard with the sensor.
5914Blue2 mil polyethyleneLess flow dependence than 1.25 mil
but somewhat slower response
Polarographic membrane kits:
ItemColorMaterialDescription
5906Black1 mil Teflon®Traditional membrane material
5908Ye l l o w1.25 mil polyethylene Improved response time and less
flow dependence than Teflon®
Ships standard with the sensor.
5909Blue2 mil polyethyleneLess flow dependence than 1.25 mil
but somewhat slower response
Selecting a Dissolved Oxygen Membrane:
Typical Response
Membrane TypeFlow Dependence After 4 Minutes
5912, 5906 - Black 60%18 seconds
5913, 5908 - Yellow25%8 seconds
5914, 5909 - Blue18%17 seconds
Time - 95%
24
25
Local DO allows for localized DO% measurements. This sets the calibration
value to 100% regardless of the altitude or barometric pressure. Highlight Local
DO and press enter to enable (Q) or disable (P) this function. Local DO is
a method for the Pro Plus to factor in the barometric pressure on each DO
measurement. In essence, if the barometric pressure changes you wouldn’t notice
the difference in the DO% readings in air-saturated water or water-saturated air.
Local DO is ideal for EU compliance. When Local DO is enabled, an L will
appear next to DO% on the run screen. DO mg/L readings are unaffected by the
selection of DO Local.
is achieved) and press enter to confirm. The Auto Stable Sensitivity can be
decreased or increased. Highlight Sensitivity and use the left and right arrow
keys to slide the bar. The more sensitive you make it (larger black bar) the
harder it is to achieve stability in a changing environment.
The Auto Stable system works by examining the previous 5 readings, computing
the percent change in the data and comparing that change against a % threshold
value. The % threshold value is determined by the Sensitivity bar setting. The
following chart can be used as a guide when setting the Sensitivity bar.
LDS (Last Digit Suppression) rounds the DO value to the nearest tenth; i.e.
8.27 mg/L becomes 8.3 mg/L. Highlight LDS and press enter to enable (Q) or
disable (
) this function.
P
DISPLAY DISSOLVED OXYGEN
Press Sensor , highlight Display and press enter. Highlight DO and press
enter. All DO units can be displayed simultaneously. Highlight the unit(s)
and press enter to activate (Q) or deactivate (P) units from the run screen.
Note - You will not be able to display dissolved oxygen unless it is Enabled in the
Sensor Setup menu first, see previous section.
DO % will show DO readings in a percent scale
from 0 to 500%.
DO mg/L will show DO readings in milligrams
per liter (equivalent to ppm) on a scale from 0
to 50 mg/L.
DO ppm will show DO readings in parts per million (equivalent to mg/L) on a
scale from 0 to 50 ppm.
AUTO STABLE DISSOLVED OXYGEN
Sensitivity selected by User% Data Variance Threshold
The instrument is programmed to determine the minimum and maximum
data value over the previous 5 samples, and to compute the percent difference
between those values. In this example, that gives a percent change of:
0.05%
2.5%
Auto Stable indicates when a reading is stable.When Auto Stable is enabled, AS
will blink next to the parameter until it is stable. Once the parameter is stable,
AS will stop blinking.
To enable Auto Stable, press Sensor
highlight Auto Stable and press enter. Highlight DO and press enter.
Highlight Enabled and/or Audio Enabled
(instrument will beep when the stability
In this example, if the Sensitivity bar is set to the far right, the Auto Stable
requirement would not be met and AS would continue to blink. However,
if the sensitivity bar is set to the median threshold (1.275%), the Auto Stable
requirement would be met and AS would display steadily on the display.
Within the Auto Stable menu, you can also choose to Hold All Readings for as
many parameters as you set for Auto Stable. For instance, if DO and pH have
26
27
Auto Stable and Hold All Readings enabled,
then the display will hold the readings once DO
and pH have both reached their Auto Stable
settings. You must press the Esc key to “release”
the held display in order to take subsequent
readings Hold All Readings must be reactivated
after each use!
SALINITY CORRECTION
The last feature in the Sensor menu is the Salinity
correction value which is used to calculate the
dissolved oxygen mg/L and ammonia readings
when a conductivity sensor is not in use. Press
Sensor
Then, use the numeric entry screen to enter the
Salinity value of the water you will be testing
from 0 to 70 ppt.
If using a cable with a conductivity sensor, the salinity
measured by the conductivity sensor will be used in the DO and
ammonia mg/L calculations and ‘As Measured’ will be displayed next to Salinity
in the Sensor menu.
, highlight Salinity, and press enter.
CALIBRATION DISSOLVED OXYGEN
The Pro Plus offers several options for calibrating dissolved oxygen: DO% in
water saturated air, DO mg/L and DO ppm in a solution of known dissolved
oxygen determined by a Winkler Titration, and a Zero point. If performing a
zero point calibration, you must also perform a %, mg/L, or ppm calibration
following the zero calibration. For both ease of use and accuracy, YSI recommends
performing the following 1-point DO % water saturated air calibration:
It is not necessary to calibrate in both % and mg/L or ppm.
i
Calibrating in % will simultaneously calibrate
mg/L and ppm and vice versa.
Calibrating DO % in Water Saturated Air:
1-Point Calibration
The supplied sensor storage container (a grey sleeve for a single port cable or a
screw on plastic cup for the dual-port and Quatro cables) can be used for DO
calibration purposes.
Moisten the sponge in the storage sleeve or plastic cup with a small amount of
clean water. The sponge should be clean since bacterial growth may consume
oxygen and interfere with the calibration. If using the cup and you no longer
have the sponge, place a small amount of clean water (1/8 inch) in the plastic
storage cup instead.
As the salinity of water increases, its ability to dissolve oxygen decreases. For
example, fully oxygenated 20 °C water at sea level with zero salinity will hold
9.092 mg/L of dissolved oxygen. If that same sample had a salinity value of 9
ppt, then it would hold 8.621 mg/L of dissolved oxygen. Therefore, to obtain
accurate mg/L readings, it is important to know the salinity of the water you
will be testing and to input that value into the instrument. The salinity of fresh
water is typically 0-0.5 ppt and seawater is typically 35 ppt. You will also have
the opportunity to enter or modify the Salinity correction value during DO
calibration.
Make sure there are no water droplets on the DO membrane or temperature
sensor. Then install the storage sleeve or cup over the sensors. The storage
sleeve ensures venting to the atmosphere. If using the cup, screw it on the cable
and then disengage one or two threads to ensure atmospheric venting. Make
sure the DO and temperature sensors are not immersed in water. Turn the
instrument on and wait approximately 5 to 15 minutes for the storage container
to become completely saturated and to allow the sensors to stabilize.
Press Cal
if you wish to add, select, edit, or delete an ID.
Probe ID must be enabled in the System GLP
menu to appear in the Calibrate menu. User
ID will appear automatically. Select ‘None’
if you do not want a User ID stored with the
calibration. When enabled, these IDs are stored
with each calibration record in the GLP file.
. Highlight Probe ID or User ID
28
29
After selecting your User ID and/or Probe ID if appropriate, highlight DO and
press enter.
Calibrating DO% in Water Saturated Air:
2-Point Calibration with Zero Solution
Highlight DO % and press enter to confirm.
The instrument will use the internal barometer
during calibration and will display this value
in brackets at the top of the display. Highlight
Barometer and press enter to adjust it if needed.
If the barometer reading is incorrect, it is
recommended that you calibrate the barometer.
Note - the barometer should be reading “true”
barometric pressure (see Barometer section
for more information on “true” barometric
pressure). If the value is acceptable, there is
no need to change it or perform a barometer
calibration.
The Salinity value displayed near the top of the
screen is either the salinity correction value
entered in the Sensor menu or the Salinity
value as measured by the conductivity sensor
in use and enabled. If you are not using a
conductivity sensor, the Salinity correction
value should be the salinity of the water you
will be testing. Highlight Salinity and press
enter to modify this setting if necessary. See the
Salinity Correction section of this manual for
more information.
Place the sensor in a solution of zero DO.
A zero DO solution can be made by dissolving approximately8 - 10 gramsof
sodium sulfite (Na2SO3) into 500 mL tap water or DI water. Mix the solution
thoroughly. It may take the solution 60 minutes to be oxygen-free.
Press Cal
delete an ID. Probe ID must be enabled in the System GLP menu to appear in
the Calibrate menu. When enabled, these IDs are stored with each calibration
record in the GLP file.
After selecting the Probe ID and/or User ID if appropriate, highlight DO and
press enter. Highlight Zero and press enter. Wait for the temperature and
DO% values under “Actual Readings” to stabilize, then press enter to Accept Calibration. If User Field 1 or 2 are enabled, you will be prompted to select the
fields and then press Cal
prompt for a follow-up second point calibration.
Highlight DO% and press enter to continue with the next calibration point.
Rinse the sensor of any zero oxygen solution using clean water. Then follow the
steps under Calibrating DO % in Water Saturated Air to complete the second
point.
. Highlight Probe ID or User ID if you wish to add, select, edit, or
to complete the calibration. The screen will then
Calibrating in mg/L or ppm as a Titration:
1-Point Calibration
Wait for the temperature and DO% values under “Actual Readings” to stabilize,
then highlight Accept Calibration and press enter to calibrate. Or, press Esc
to cancel the calibration. If User Field 1 or 2 are enabled in the GLP menu, you
will be prompted to select these inputs and then press Cal
calibration. The message line at the bottom of the screen will display “Calibrating
Channel...” and then “Saving Configuration...”.
to complete the
Place the sensor into an adequately stirred sample that has been titrated to
determine the dissolved oxygen concentration. Allow the sensor to stabilize.
Press Cal
delete an ID. Probe ID must be enabled in the System GLP menu to appear in
the Calibrate menu. When enabled, these IDs are stored with each calibration
record in the GLP file.
After selecting the Probe ID and/or User ID if appropriate, highlight DO and
press enter. Highlight DO mg/L or ppm and press enter.
. Highlight Probe ID or User ID if you wish to add, select, edit, or
30
31
Highlight Calibration value and press enter to
manually input the sample’s dissolved oxygen
value. Highlight Accept Calibration and press
enter once the temperature and Dissolved
Oxygen readings stabilize. Or, press Esc
cancel the calibration. If User Field 1 or 2 are
enabled in the GLP menu, you will be prompted
to select the fields after selecting Accept Calibration. After making your selection,
press Cal
completing the calibration, the message line
will display “Calibrating Channel...” and then
“Saving Configuration...”.
to complete the calibration. After
to
Calibrating in mg/L or ppm as a Titration:
2-Point Calibration with Zero Solution
Place the sensor in a solution of zero DO.
A zero DO solution can be made by dissolving approximately 8 - 10 grams of
sodium sulfite (Na2SO3) into 500 mL tap water. Mix the solution thoroughly. It
may take the solution 60 minutes to be oxygen-free.
Press Cal
delete an ID. Probe ID must be enabled in the System GLP menu to appear in
the Calibrate menu. When enabled, these IDs are stored with each calibration
record in the GLP file.
. Highlight Probe ID or User ID if you wish to add, select, edit, or
BaroMeter
All Professional Plus instruments contain an internal barometer.
DISPLAY BAROMETER
Press Sensor , highlight Display and press enter. Highlight Barometer and
press enter. The measurement unit options are: mmHg, inHg, mBar, PSI, kPa,
or Atm. Only one unit can be displayed at a time. Select None if you do not
want to display a barometric pressure reading.
Whether or not you choose to display the barometer reading, the barometric
pressure will still be used for calibrating DO% and for compensating for pressure
changes if Local DO is enabled.
CALIBRATION BAROMETER
The barometer in the instrument is calibrated
at the factory. If the barometer requires
calibration, press Cal
or User ID if you wish to add, select, edit, or
delete an ID. Probe ID must be enabled in the
System GLP menu to appear in the Calibrate
menu. When enabled, these IDs are stored with
each calibration record in the GLP file.
After selecting the Probe ID and/or User ID
if appropriate, highlight Barometer and press
enter.
. Highlight Probe ID
After selecting the Probe ID and/or User ID if appropriate, highlight DO and
press enter. Highlight Zero and press enter. Wait for the temperature and
DO% values under “Actual Readings” to stabilize, then press enter to Accept Calibration. If User Field 1 or 2 are enabled, you will be prompted to select the
fields and then Press Cal
prompt for a follow-up second point calibration.
Highlight the desired calibration units (mg/L or ppm) and press enter
to continue with the next point. Rinse the sensor of any zero oxygen
solution using clean water. To complete the second calibration point,
follow the steps under Calibrating in mg/L or ppm as a Titration:
1-Point Calibration.
to complete the calibration. The screen will then
Highlight the desired unit and press enter.
32
33
Highlight Calibration Value and press enter to
manually enter the correct “true” barometric
pressure. Next, highlight Accept Calibration,
and press enter. If User Field 1 or 2 are enabled,
you will be prompted to select the fields and
then press Cal
or press Esc
Laboratory barometer readings are usually “true”
(uncorrected) values of air pressure and can be used “as is”
for barometer calibration. Weather service readings are
i
usually not “true”, i.e., they are corrected to sea level, and
therefore cannot be used until they are “uncorrected”. An
approximate formula for this “uncorrection” is below:
True BP = [Corrected BP] – [2.5 * (Local Altitude in ft. above sea level/100)]
to complete the calibration
to cancel the calibration.
conductIVIty
Conductivity sensors are supplied with 60530-X, 6051030-X, 6052030-X , and
Quatro cables. Conductivity sensors are built into the 60530-X, 6051030-X, and
6052030-X cables and are not replaceable. Conductivity/Temperature sensors
are shipped with the Quatro cable, must be installed, and are replaceable.
temperature all Specific Conductance values are compensated to. The default
is 25 °C. To change the Reference Temperature, highlight Temp Ref and press
enter. Use the numeric entry screen to enter a new value between 15.00 and
25.00 °C. Next, highlight <<<ENTER>>> at the bottom of the screen and press
enter on the keypad to confirm.
%/°C (Percent per Degree Celsius) is the temperature coefficient used to
calculate temperature compensated Specific Conductance. The default is
1.91% which is based on KCl standards. To change the temperature coefficient,
highlight %/°C and press enter. Use the numeric entry screen to enter a new
value between 0 and 4%. Next, highlight <<<ENTER>>> at the bottom of the
screen and press Enter on the keypad to confirm.
TDS Constant is a multiplier used to calculate an estimated TDS (Total
Dissolved Solids) value from conductivity. The multiplier is used to convert
Specific Conductance in mS/cm to TDS in g/L. The default value is 0.65. This
multiplier is highly dependent on the nature of the ionic species present in the
water sample. To be assured of moderate accuracy for the conversion, you must
determine a multiplier for the water at your sampling site. Use the following
procedure to determine the multiplier for a specific sample:
Determine the specific conductance of a water sample from the site;1.
Filter a portion of water from the site;2.
Completely evaporate the water from a carefully measured volume of the 3.
filtered sample to yield a dry solid;
Accurately weigh the remaining solid;4.
Divide the weight of the solid (in grams) by the volume of water used (in 5.
liters) to yield the TDS value in g/L for this site; Divide the TDS value in g/L
by the specific conductance of the water in mS/cm to yield the conversion
multiplier. Be certain to use the correct units.
Enabled allows you to enable or disable the
conductivity measurement. Highlight Enabled
and press enter to activate (Q) or deactivate (P)
conductivity. Disable conductivity if you do
not have a conductivity sensor connected to the
instrument.
i
Temp Ref (Temperature Reference) is the reference temperature used for
calculating temperature compensated Specific Conductance. This will be the
If a sensor is Enabled that isn’t connected to the instrument, the
display will show an unstable, false reading next to the units.
If the nature of the ionic species at the site changes between
i
To change the multiplier, highlight TDS Constant and press enter. Use the
numeric entry screen to enter a new value between 0 and 0.99. Highlight
<<<ENTER>>> at the bottom of the screen and press Enter on the keypad to
confirm.
sampling studies, the TDS values will be in error. TDS cannot
be calculated accurately from specific conductance unless the
make-up of the chemical species in the water remains constant.
34
35
DISPLAY CONDUCTIVITY
Press Sensor , highlight Display and press enter. Highlight Conductivity and
press enter. Highlight Sp. Conductance (Specific Conductance), Conductivity, Salinity, TDS, or Resistivity, and press enter to select the reporting units for
each parameter. One reporting unit per parameter may be enabled. To disable a
parameter, select None. You will not be able to display any of these parameters
unless the Conductivity sensor is Enabled in the Sensor Setup menu first.
Sp. Conductance can be displayed in us/cm or
ms/cm. Specific conductance is temperature
compensated conductivity.
Conductivity can be displayed in uS/cm or mS/
cm. Conductivity is the measure of a solution’s
ability to conduct an electrical current.
Unlike specific conductance, conductivity
is a direct reading without any temperature
compensation.
Salinity can be displayed in ppt (parts per
thousand) or PSU (practical salinity units).
The units are equivalent as both use the Practical
Salinity Scale for calculation.
TDS can be displayed in mg/L (milligrams per
liter), g/L (grams per liter), or kg/L (kilograms per liter).
Resistivity can be displayed in ohm-cm (ohms per centimeter), kohm-cm (kilo
ohms per centimeter), or Mohm-cm (mega ohms per centimeter).
AUTO STABLE CONDUCTIVITY
Press Sensor , highlight Auto Stable and press enter. Highlight Conductivity
and press enter.
The Auto Stable Sensitivity can be decreased or increased. Highlight Sensitivity
and use the left and right arrow keys to slide the bar. The more sensitive you
make it (larger black bar) the harder it is to achieve stability in a changing
environment.
The Auto Stable system works by examining the previous 5 readings, computing
the percent change in the data and comparing that change against a % threshold
value. The % threshold value is determined by the Sensitivity bar setting. The
following chart can be used as a guide when setting the Sensitivity bar.
Sensitivity selected by User% Data Variance Threshold
100 - Most Sensitive, Sensitivity
bar is set to the far right
750.39375%
500.7625%
251.13125%
0 - Least Sensitive, Sensitivity
bar is set to the far left
Within the Auto Stable menu, you can also
choose to Hold All Readings for as many
parameters as you set for Auto Stable. For
instance, if conductivity and DO have Auto
Stable and Hold All Readings enabled, then
the display will hold the readings once
conductivity and DO have both reached their
Auto Stable settings. You must press the Esc
key to “release” the held display in order to take
subsequent readings. Hold All Readings must
be reactivated after each use!
0.025%
1.5%
Auto Stable indicates when a reading is stable.
Highlight Enabled and/or Audio Enabled
(instrument will beep when the stability is
achieved) and press enter enable (Q) or disable
(P). When Auto Stable is enabled, AS will blink
next to the parameter until it is stable. Once the
parameter is stable, AS will stop blinking.
36
37
CALIBRATION CONDUCTIVITY
The 6051030 ISE/conductivity cable has a specialized calibration
i
container that resembles a large test tube. This calibration chamber
can be used to calibrate the conductivity sensor with an ISE sensor
installed. A ring-stand should be used to support this chamber.
Choose the units in either SPC-us/cm, C-us/cm
or SPC-ms/cm, C-ms/cm and press enter.
Press Cal
if you wish to add, select, edit, or delete an ID.
Probe ID must be enabled in the System GLP
menu to appear in the Calibrate menu. User
ID will appear automatically. Select ‘None’
if you do not want a User ID stored with the
calibration. When enabled, these IDs are stored
with each calibration record in the GLP file.
After selecting the User ID and/or Probe ID if
appropriate, highlight Conductivity and press
enter.
Highlight the desired calibration method; Sp. Conductance, Conductivity, or Salinity and
press enter. YSI recommends calibrating in
specific conductance for greatest ease.
. Highlight Probe ID or User ID
Highlight Calibration value and press enter
to input the value of the calibration standard.
Then, once the temperature and conductivity
readings stabilize, highlight Accept Calibration
and press enter. Or, press Esc
calibration. If User Field 1 or 2 are enabled in
the GLP menu, you will be prompted to select
the fields and then press Cal
calibration. After completing the calibration,
the message line at the bottom of the screen
will display “Calibrating Channel...” and then
“Saving Configuration...”.
Calibrating in Salinity
Place the sensor into a salinity calibration solution. The solution must cover the
holes of the conductivity sensor that are closest to the cable. Ensure the entire
conductivity sensor is submerged in the solution or the instrument will read
approximately of half the expected value!
Select SAL ppt or SAL PSU and press enter.
to cancel the
to complete the
Calibrating in Specic (Sp.) Conductance or Conductivity
Place the sensor into a fresh, traceable conductivity calibration solution. The
solution must cover the holes of the conductivity sensor that are closest to the
cable. Ensure the entire conductivity sensor is submerged in the solution or the
instrument will read approximately of half the expected value!
Highlight Calibration value and press enter
to input the value of the calibration standard.
Then, once the temperature and conductivity
readings stabilize, highlight Accept Calibration
and press enter. Or, press Esc
calibration. If User Field 1 or 2 are enabled, you
will be prompted to select the fields and then
press Cal
to complete the calibration.
to cancel the
38
39
pH
DISPLAY H
pH sensors can be used on 60510-X, 6051020-X, 6051030-X, 6051010-X, and
Quatro cables.
If using a 605103 pH/ORP combination sensor on a 6051020 or 6051030 cable
you can report both pH and ORP by configuring ISE1 as pH and ISE2 as ORP
in the Sensor Setup menu.
The 605103 pH/ORP combination sensor is not recommended for use on a
6051010 or Quatro cable. If used on one of these cable, only pH will be reported
and ORP will not be measured.
SETUP H
Press Sensor , highlight Setup, press enter. Highlight ISE1 if using a 60510,
6051020, or 6051030 cable. If using a 6051010 or Quatro cable, highlight ISE1 if
the pH sensor is installed in port 1 or highlight ISE2 if the pH sensor is installed
in port 2(a sensor must be installed in port 1 for port 2 to operate). Press enter.
Enabled allows you to enable or disable the
ISE function and select which ISE sensor is
installed on the cable. Highlight Enabled and
press enter to enable (Q) or disable (P) the ISE
you selected previously (either ISE1 or ISE2).
Disable the ISE function(s) if you do not have a
ISE sensor connected to the instrument.
Press Sensor , highlight Display and press
enter.
Highlight ISE (pH) and press enter. You will
not be able to Display the sensor unless it is
Enabled in the Sensor Setup menu.
Highlight pH and/or pH mV, press enter to
enable (Q) or disable (P). Both can be reported
at the same time.
AUTO STABLE H
Press Sensor , highlight Auto Stable and press enter. Highlight ISE (pH)
and press enter.
Auto Stable indicates when a reading is stable.
Highlight Enabled and/or Audio Enabled
(instrument will beep when the stability is
achieved) and press enter enable (Q) or disable
(P). When Auto Stable is enabled, AS will blink
next to the parameter until it is stable. Once the
parameter is stable, AS will stop blinking.
After enabling the ISE function, ensure that it
is set to pH as shown in the left screen shot. If
necessary, highlight pH and press enter to set
the ISE to pH.
Highlighting pH[USA] and pressing enter
will also allow you to select the values for
auto buffer recognition which are used during
calibration. The buffer options are USA (4, 7,
10), NIST (4.01, 6.86, 9.18), and User-Defined. The selected option will be
displayed in [brackets].
i
If a sensor is Enabled that isn’t connected to the
instrument, the display will show an unstable false
reading, ?????, or ----- next to the units.
The Auto Stable Sensitivity can be decreased or increased. Highlight Sensitivity
and use the left and right arrow keys to slide the bar. The more sensitive you
make it (larger black bar) the harder it is to achieve stability in a changing
environment.
The Auto Stable system works by examining the previous 5 readings, computing
the percent change in the data and comparing that change against a % threshold
value. The % threshold value is determined by the Sensitivity bar setting. The
following chart can be used as a guide when setting the Sensitivity bar.
40
41
Sensitivity selected by User% Data Variance Threshold
100 - Most Sensitive, Sensitivity
bar is set to the far right
750.39375%
501.5%
251.13125%
0 - Least Sensitive, Sensitivity
bar is set to the far left
Within the Auto Stable menu, you can also
choose to Hold All Readings for as many
parameters as you set for Auto Stable. For
instance, if ORP and pH have Auto Stable
enabled and Hold All Readings is enabled, then
the display will hold the readings once ORP and
pH have both reached their Auto Stable settings.
You must press the Esc key to “release” the held
display in order to take subsequent readings.
Hold All Readings must be reactivated after each use!
0.025%
0.15%
CALIBRATION H
Calibration can be accomplished in any buffer order. pH
i
7 buffer should be used regardless of how many calibration
points you use but it does not have to be used first.
in the Sensor Setup menu may be incorrect. If
necessary, highlight the Calibration Value and
press enter to input the correct buffer value.
Once the pH and temperature readings stabilize,
highlight Accept Calibration and press enter to
accept the first calibration point. The message
line will then display “Ready for point 2”.
If you do not wish to perform a second point,
press Cal
press Esc to cancel the calibration. If User
Field 1 or 2 are enabled, you will be prompted
to select these fields and then press Cal
finalize the calibration.
To continue with the 2nd point, place the sensor
in the second buffer solution. The instrument
should automatically recognize the second
buffer value and display it at the top of the
screen. If necessary, highlight the Calibration
Va lu e and press enter to input the correct buffer
value. Once the pH and temperature readings
stabilize, highlight Accept Calibration and
press enter to confirm the second calibration
point. The message line will then display ‘Ready
for point 3” and you can continue with the 3rd
calibration point if desired.
to finalize the calibration. Or,
to
Press Cal
if you wish to add, select, edit, or delete an ID.
Probe ID must be enabled in the System GLP
menu to appear in the Calibrate menu. User
ID will appear automatically. Select ‘None’
if you do not want a User ID stored with the
calibration. When enabled, these IDs are stored
with each calibration record in the GLP file.
After selecting your User ID and/or Probe ID if appropriate, highlight ISE (pH)
and press enter. The message line will show the instrument is “Ready for point
1”. The pH calibration allows up to six calibration points.
Place the sensor in a traceable pH buffer solution. The instrument should
automatically recognize the buffer value and display it at the top of the calibration
screen. If the calibration value is incorrect, the auto buffer recognition setting
. Highlight Probe ID or User ID
If you do not wish to perform a 3rd calibration
point, press Cal
If User Field 1 or 2 are enabled, you will be prompted to select these fields and
then press Cal
Continue in this fashion until the desired number of calibration points is
achieved (up to six).
u
to finalize the calibration.
Once you’ve achieved the desired number of cal points you
must press Cal
instrument to update the pH offset and slope. The instrument will
not take these cal values into account until Cal has been pressed.
to finalize the calibration and to allow the
to complete the calibration.
42
43
The actual readings displayed during the calibration will
NOT reflect the updated calibration information. These
i
values will not change until Cal is pressed to finalize
the calibration and to update the instrument.
orp
ORP sensors can be used on 60510-X, 6051020-X, 6051030-X, 6051010-X, and
Quatro cables.
If using a 605103 pH/ORP combination sensor on a 6051020 or 6051030 cable
you can report both pH and ORP by configuring ISE1 as pH and ISE2 as ORP
in the Sensor Setup menu.
The 605103 pH/ORP combination sensor is not recommended for use on a
6051010 or Quatro cable. If used on one of these cable, only pH will be reported
and ORP will not be measured.
SETUP ORP
Press Sensor , highlight Setup, press enter.
Highlight ISE1 if using a 605102 (ORP sensor)
on a 60510, 6051020, or 6051030 cable. Highlight
ISE2 is using a 605103 (pH/ORP sensor) on a
60510, 6051020, or 6051030 cable. If using a
6051010 or Quatro cable, highlight ISE1 if the
ORP sensor is installed in port 1 or highlight
ISE2 if the ORP sensor is installed in port 2 (a
sensor must be installed in port 1 for port 2 to
operate). Press enter.
If a sensor is Enabled that isn’t connected to the
i
instrument, the display will show an unstable false
reading, ?????, or ----- next to the units.
DISPLAY ORP
AUTO STABLE ORP
Press Sensor , highlight Display and press
enter.
Highlight ISE (ORP) and press enter. You will
not be able to Display the sensor unless it is
Enabled in the Sensor Setup menu.
Press enter to enable (Q) or disable (P) ORP
m V.
Press Sensor , highlight Auto Stable and
press enter. Highlight ISE (ORP) and press
enter.
Auto Stable indicates when a reading is stable.
Highlight Enabled and/or Audio Enabled
(instrument will beep when the stability is
achieved) and press enter enable (
(P). When Auto Stable is enabled, AS will blink
next to the parameter until it is stable. Once the
parameter is stable, AS will stop blinking.
) or disable
Q
Enabled allows you to enable or disable the ISE
function and select which ISE sensor is installed
on the cable. Highlight Enabled and press enter
to enable (Q) or disable (P) the ISE you selected
previously (either ISE1 or ISE2).
After enabling the ISE function, ensure ORP is
selected as the ISE sensor as shown in screen
shot to the left. If necessary, highlight ORP and
press enter to set the selected ISE to ORP.
The Auto Stable Sensitivity can be decreased or
increased. Highlight Sensitivity and use the left and right arrow keys to slide
the bar. The more sensitive you make it (larger black bar) the harder it is to
achieve stability in a changing environment.
The Auto Stable system works by examining the previous 5 readings, computing
the percent change in the data and comparing that change against a % threshold
value. The % threshold value is determined by the Sensitivity bar setting. The
following chart can be used as a guide when setting the Sensitivity bar.
44
45
Sensitivity selected by User% Data Variance Threshold
100 - Most Sensitive, Sensitivity
bar is set to the far right
750.62525%
501.275%
251.8875%
0 - Least Sensitive, Sensitivity
bar is set to the far left
Within the Auto Stable menu, you can also
choose to Hold All Readings for as many
parameters as you set for Auto Stable. For
instance, if ORP and pH have Auto Stable
enabled and Hold All Readings is enabled,
then the display will hold the readings once
ORP and pH have both reached their Auto
Stable settings. You must press the Esc key
to “release” the held display in order to take
subsequent readings. Hold All Readings must
be reactivated after each use!
0.05%
2.5%
CALIBRATION ORP
Highlight Calibration value and press enter to
input the value of the ORP calibration standard.
If using the YSI Zobell calibration solution,
the Pro Plus will automatically determine the
calibration value. However, the calibration value
should be verified against the chart on the side
of the Zobell bottle. Next, once the temperature
and ORP readings stabilize, highlight Accept
Calibration and press enter to calibrate. Or,
press Esc
Field 1 or 2 are enabled, you will be prompted
to select the fields and then press Cal
complete the calibration.
to cancel the calibration. If User
to
aMMonIuM, nItrate, chlorIde
Ammonium, Nitrate, and Chloride sensors can be used on 60510-X, 6051020-X,
6051030-X, 6051010-X, and Quatro cables. These cables also accommodate pH
and ORP sensors so instrument setup is important.
WARNING: Ammonium, Nitrate, and Chloride sensors
should only be used at DEPTHS OF LESS THAN 55 FEET
i
(17 METERS). Use of the sensors at greater depths is
likely to permanently damage the sensor membrane.
Press Cal . Highlight Probe ID or User ID
if you wish to add, select, edit, or delete an ID.
Probe ID must be enabled in the System GLP
menu to appear in the Calibrate menu. User
ID will appear automatically. Select ‘None’
if you do not want a User ID stored with the
calibration. When enabled, these IDs are stored
with each calibration record in the GLP file.
After selecting your User ID and/or Probe
ID if appropriate, highlight ISE (ORP) and
press enter. The message line will show the
instrument is “Ready for point”.
Place the sensor in a solution of known ORP
and wait for the readings to stabilize.
i
WARNING: Ammonium, Nitrate, and Chloride
sensors should only be used in FRESHWATER.
SETUP AMMONIUM, NITRATE, CHLORIDE
Install the Ammonium, Nitrate, or Chloride sensor in Port 2 if using in
conjunction with pH or ORP sensor on a 6051010 or Quatro cable. See the
Getting Started Setup section of this manual for a complete list of cable/sensor
configurations.
Press Sensor
ammonium, nitrate, or chloride sensor on a 60510, 6051020, or 6051030 cable.
If using a 6051010 or Quatro cable highlight ISE1 if the sensor is installed in
Port 1 or highlight ISE2 if the sensor is installed in Port 2. Press enter.
, highlight Setup, press enter. Highlight ISE1 if using an
46
47
Enabled allows you to enable or disable the ISE
function and select which ISE sensor is installed
on the cable.
Highlight Enabled and press enter to enable (Q)
or disable (P) the ISE you selected previously
(either ISE1 or ISE2).
After enabling the ISE function, choose the
parameter you want enabled for that ISE. In
this example, NH4 is selected.
Cl - Chloride
NH4 - Ammonium
NO3 - Nitrate
i
If a sensor is Enabled that isn’t connected to the instrument, the
display will show an unstable, false reading next to the units.
DISPLAY AMMONIUM, NITRATE, CHLORIDE
Press Sensor , highlight Display, press enter. Highlight ISE2(NH4), press
enter. You will not be able to Display the sensor unless it is Enabled.
Highlight the value you wish to display and
press enter to enable (Q) . Ammonium can
be displayed as NH4-N mg/L (Ammonium),
NH3-N (Ammonia) and/or NH4 mV (sensor
signal).
The same steps would be followed to display
nitrate or chloride.
Ammonia is calculated from the pH, salinity, and temperature readings. If a
pH sensor is not in use, the instrument will assume the sample is neutral (pH
7) for the calculation. If a conductivity sensor (Salinity) is not in use, the
instrument will use the salinity correction value entered in the Sensor Menu
for the calculation (see Salinity Correction within the Dissolved Oxygen Setup
section of this manual for more information).
AUTO STABLE AMMONIUM, NITRATE, CHLORIDE
Auto Stable indicates when a reading is stable. When Auto Stable is enabled, AS
will blink next to the parameter until it is stable. Once the parameter is stable,
AS will stop blinking.
To enable Auto Stable, press Sensor
highlight Auto Stable and press enter. Highlight ISE1 or ISE2 and press enter.
Highlight Enabled and/or Audio Enabled
(instrument will beep when the stability is
achieved) and press enter to confirm. The Auto
Stable Sensitivity can be decreased or increased.
Highlight Sensitivity and use the left and right arrow keys to slide the bar. The
more sensitive you make it (larger black bar) the harder it is to achieve stability
in a changing environment.
The Auto Stable system works by examining the previous 5 readings, computing
the percent change in the data and comparing that change against a % threshold
value. The % threshold value is determined by the Sensitivity bar setting. The
following chart can be used as a guide when setting the Sensitivity bar.
Sensitivity selected by User% Data Variance Threshold
100 - Most Sensitive, Sensitivity
bar is set to the far right
750.62525%
501.275%
251.8875%
0 - Least Sensitive, Sensitivity
bar is set to the far left
Within the Auto Stable menu, you can also
choose to Hold All Readings for as many
parameters as you set for Auto Stable. For
instance, if pH and Ammonium have Auto
Stable enabled and Hold All Readings is also
enabled, then the display will hold the readings
once pH and Ammonium have both reached
their Auto Stable settings. You must press the
Esc key to “release” the held display in order to
take subsequent readings. Hold All Readings
must be reactivated after each use!
0.05%
2.5%
,
48
49
CALIBRATION AMMONIUM, NITRATE, CHLORIDE
The 6051030 ISE/conductivity cable has a specialized calibration container that
resembles a large test tube. This calibration chamber can be used to calibrate
the ISE sensors with the conductivity sensor. A ring-stand should be used to
support this chamber.
The ISE sensors can be calibrated at 1, 2, or 3-points.
A 2-point calibration without chilling a third calibration
i
solution is extremely accurate and is the preferred method.
Greatest accuracy is achieved if the actual samples to be
measured are within 10 ˚C of the calibration solutions.
balance, 1000 mL volumetric flask, an accurate 10 mL measuring devices, and
1000 mL glass or plastic storage vessels.
1000 mg/L Standard: Accurately weigh 1.655 grams of anhydrous sodium
chloride and transfer into a 1000 mL volumetric flask. Add 0.5 grams of
anhydrous magnesium sulfate to the flask. Add 500 mL of water to the flask,
swirl to dissolve all of the reagents , and then dilute to the volumetric mark with
water. Mix well by repeated inversion and then transfer the 1000 mg/L standard
to a storage bottle. Rinse the flask extensively with water prior to its use in the
preparation of the 10 mg/L standard. Alternatively, simply add 0.5 grams of
magnesium sulfate to a liter of a 1000 mg/L chloride standard from a certified
supplier.
CALIBRATION TIP: Exposure to the high ionic content of pH buffers can
cause a significant, but temporary, drift in the ammonium, nitrate, and chloride
ISE sensors. Therefore, when calibrating the pH sensor, YSI recommends that
you use one of the following methods to minimize errors in the subsequent
readings:
When calibrating pH, remove ISE sensors from the cable bulkhead and •
plug the ports. After pH calibration is complete, replace the ISE sensors and
proceed with their calibration with no stabilization delay.
Calibrate pH first, immersing all of the sensors in the pH buffers. After •
calibrating pH, place the sensors in 100 mg/L nitrate or ammonium standard
or 1000 mg/L chloride standard depending on the sensor in use and monitor
the reading. Usually, the reading starts low and may take awhile to reach
a stable value. When it does, proceed with the calibration. This may take
several hours.
Preparing Chloride Standards
The following recipes are provided for preparation of 10 and 1000 mg/L chloride
reagents. Nitrate and Ammonium standards can be purchased from YSI or other
laboratory supply companies.
It is important to note that some of the chemicals required for these solutions
could be hazardous under some conditions. It is the responsibility of the user
to obtain and study the MSDS for each chemical and to follow the required
instructions with regard to handling and disposal of these chemicals.
You will need: Solid sodium chloride or a certified 1000 mg/L chloride solution
from a supplier, magnesium sulfate, high purity water, a good quality analytical
10 mg/L Standard: Accurately measure 10 mL of the above 1000 mg/L standard
solution into a 1000 mL volumetric flask. Add 0.5 grams of anhydrous magnesium
sulfate to the flask. Add 500 mL of water, swirl to dissolve the solid reagents, and
then dilute to the volumetric mark with water. Mix well by repeated inversion
and then transfer the 10 mg/L standard to a storage bottle.
AMMONIUM (NH4+) , NITRATE (NO3-), AND CHLORIDE CL-
2-POINT
The calibration procedures for ammonium, nitrate, or chloride are similar to
pH. The only differences are the calibration solutions. Recommended values
for calibration solutions and the order of calibration are as follows:
Sensor1st Point2
Ammonium-nitrogen (NH4-N)1 mg/L100 mg/L
Nitrate-nitrogen (NO3-N)1 mg/L100 mg/L
Chloride (Cl-)10 mg/L1000 mg/L
Place the proper amount of 1 mg/L standard for Ammonium or Nitrate (10
mg/l for Chloride) into a clean, dry or pre-rinsed calibration cup. Carefully
immerse the sensor into the solution. Allow at least 1 minute for temperature
equilibration before proceeding.
Press Cal
or delete an ID. Probe ID must be enabled in the System GLP menu to appear
in the Calibrate menu. User ID will appear automatically. Select ‘None’ if you
do not want a User ID stored with the calibration. When enabled, these IDs are
stored with each calibration record in the GLP file.
. Highlight Probe ID or User ID if you wish to add, select, edit,
nd
Point
50
51
After selecting your User ID and/or Probe ID if
appropriate, highlight Ammonium, Nitrate, or
Chloride to access the appropriate calibration,
and press enter. The parameter you want
to calibrate may appear under ISE1 or ISE2
depending on your cable type and setup. The
message line will show the instrument is ready
for the 1st calibration point.
The instrument will display the calibration value
at the top of the screen. If necessary, highlight
the Calibration value and press enter to input the correct value.
Once the readings stabilize, highlight Accept Calibration and press enter to
accept the first calibration point. The message line will then display “Ready for
point 2”.
If you do not wish to perform a second point, press Cal
calibration. If User Field 1 or 2 are enabled, you will be prompted to select these
fields and then press Cal
press Esc
to cancel the calibration.
to finalize the calibration. Alternatively, you may
to finalize the
high concentration solution and one of the low concentration solutions should
be at ambient temperature. The other low concentration solution should be
chilled to less than 10 °C prior to calibration.
WARNING: The chilled calibration solution MUST BE CHILLED
u
Follow the procedure for a 2-point cal. After the second calibration point is
complete, the message line with state ‘Ready for point 3”. Place the proper
amount of chilled 1 mg/L standard (10 mg/L for the chloride) into a clean, dry
or pre-rinsed calibration cup. Carefully immerse the sensor into the solution.
Allow for temperature equilibration. If necessary, highlight Calibration value
and press enter to manually enter the 3rd buffer value. Once the readings are
stable, highlight Accept Calibration and press enter to confirm. Press Cal
to complete the calibration. If User Field 1 or 2 are enabled, you will be
prompted to select these fields and then press Cal
Alternatively, press Esc
TO AT LEAST 5 °C COOLER THAN THE 1ST CALIBRATION
POINT, otherwise the 1st point will be OVERRIDDEN.
to finalize the calibration.
to cancel the calibration.
TAKING MEASUREMENTS
To continue with the 2nd point, rinse the sensor with clean water, then dry it
before placing it in the second calibration standard. Allow at least 1 minute
for temperature equilibration before proceeding. The instrument will display
the second calibration value at the top of the screen. If necessary, highlight the
Calibration value and press enter to input the correct buffer value. Once the
readings stabilize, highlight Accept Calibration and press enter to confirm the
second calibration point. The message line will then display “Ready for point 3”
and you can continue with the 3rd calibration point if desired.
If you do not wish to perform a 3rd calibration point, press Cal
the calibration. If User Field 1 or 2 are enabled, you will be prompted to select
these fields and then press Cal
may press Esc
to cancel the calibration.
to finalize the calibration. Alternatively, you
to complete
AMMONIUM (NH4+) , NITRATE (NO3-), AND CHLORIDE CL-
3-POINT
A 2-point calibration without chilling a third calibration solution is extremely
accurate and is the preferred method. If you must perform a 3-point calibration,
the following procedure requires one portion of the high concentration calibration
solution and two portions of the low concentration calibration solution. The
To obtain the most accurate readings, be sure the instrument is calibrated before
taking measurements.
dIssolVed oxyGen
Turn the instrument on and wait 5-15 minutes if using a polarographic sensor.
If using a field cable/sensor, install the sensor guard to protect the sensor and
membrane. Place the probe in the sample to be measured and give the probe
a quick shake to release any air bubbles. Allow the temperature readings to
stabilize. Next, stir the probe in the sample to overcome the stirring dependence
of the dissolved oxygen sensor. You must provide at least 3 inches per second
for 2.0 PE membranes, 6 inches per second for 1.25 PE membranes, and 12
inches per second for Teflon® membranes. Once the values plateau and stabilize,
you may record the measurement and/or log the data set. The dissolved oxygen
reading will drop over time if stirring is ceased.
If placing the DO sensor into a stream or fast flowing waters it is best to place it
perpendicular to the flow and NOT facing into the flow.
If using the DO sensor in an aeration tank/basin, it is helpful to make sure
bubbles do not burst on the membrane since this may cause unstable readings.
You should be able to prevent this by pointing the sensor upwards so it’s facing
52
53
the sky and then twist tying, zip tying, or rubber banding the bulkhead to the
cable. Making a simple curve to the cable without bending or breaking the cable
will allow you to lower the sensor into the aeration tank while the sensor points
skyward so the bubbles are no longer bursting on the membrane surface.
conductIVIty
The conductivity sensor will provide quick readings as long as the entire sensor
is submerged and no air bubbles are trapped in the sensor area. Immerse the
probe into the sample so the sensors are completely submerged and then shake
the probe to release any air bubbles. Occasional cleaning of the sensor may be
necessary to maintain accuracy and increase the responsiveness. To clean the
sensor, use the conductivity cleaning brush with a mild detergent.
ph/orp
pH and ORP readings are typically quick and accurate. However, it may take the
sensors a little longer to stabilize if they become coated or fouled. To improve
the response time of a sensor, follow the cleaning steps in the Maintenance
section of this manual.
aMMonIuM, nItrate, and chlorIde
During a continuous log, the Start Logging dialog box on the Run screen will
change to Stop Logging. Press Enter to stop continuous logging.
FILES AND SITE LISTS
fIle MeMory
To view the file memory, press File.
The Data Memory shows a percentage
indicating the amount of memory available.
If the file memory is near 0%, files should be
downloaded to a PC and/or deleted to free up
memory.
These sensors may take a little longer to stabilize if the tips are dirty or fouled. If
installed with a pH sensor, always maintain a clean pH sensor for a more rapid
sensor stabilization.
These sensors can only be used in freshwater.
loGGInG data
Log One Sample is already highlighted in Run mode. Press enter to open a
submenu. If Use Site List and or Use Folder List are enabled in the Logging
Setup menu, you will have to option to select these two items before the data
point is logged. If necessary, use the keypad to create a new Site or Folder name.
If Site List and Folder List are disabled in the System menu, you will not see these
options when logging a sample. Once the Site and/or Folder name is selected,
highlight Log Now and press Enter. The instrument will confirm that the data
point was successfully logged.
If you would like to log at a specific interval vs. logging one sample at a time
or vice versa, press SystemContinuous Mode and adjust the time Interval if necessary. On the Run screen,
the option to log will change from Log One Sample to Start Logging based on
the time interval entered in the Logging Menu.
, then highlight Logging and press enter. Select
VIeWInG saVed data
Press File, highlight View Data and press enter.
Configuring your data view:
Site: will allow you to view data from one
particular site or all sites. Highlight Site, press
enter, and select the site you wish to view data
from or select All Sites to view data from all
sites.
Folder: will allow you to view data from one
particular folder or all folders. Highlight
Folder, press enter, and select the file you wish
to view data from or select All Folders to view
data from all folders.
Begin Date, Begin Time, End Date, and End
Time: will allow you to view data collected
between a specific time period. Highlight the
54
55
time qualifier you would like to set, press enter,
and use the numeric entry screen to select the
date/time you wish to view.
D A TA M A N AG E R
DESKTOP SOFTWARE
sIte lIst
folder
After making your selections in the Data Filter,
highlight Show Data and press enter. The data
will have date and time stamps. You will likely
have to scroll up and down and side to side using
the arrow keys to completely view the data file.
No more than 100 data records can be viewed
at one time.
To modify the Site List, press File, highlight Site List, and press enter. Enter new site names
or edit existing sites with the alpha/numeric
entry screen. Site lists can also be created and
edited on your PC with Data Manager and then
downloaded to the instrument.
Data Manager is provided with the purchase of a Pro Plus Instrument. Data
Manager is a powerful Windows® based software that will allow you to easily
manage logged data, set up instruments, and conduct real time studies.
Minimum PC system requirements for Data Manager are Windows® 2000 with
SP4 (minimum) or Windows® XP with SP2 (minimum) Operating System, 300
MHz or higher Pentium®-compatible CPU, 128 MB of RAM or higher, 80 MB or
more of free hard-disk space, USB 2.0, and Microsoft® .NET.
Data Manager needs to be installed on a PC before use and before you try to
connect a Pro Plus to your PC. First install Data Manger, then connect the
communications saddle to the PC and, lastly, connect the saddle to your Pro
Plus. Data Manager will identify the connected instruments by their Unit
ID.Refer to the Data Manager Readme file for detailed installation instructions.
Data Manager will then recognize the attached instruments.
From the ‘home’ screen of Data Manager, see below, you can select one of the
following functions: Retrieve Instrument Data, Real Time Instrument Data,
Instrument Configuration, or View Saved File/Data.
To modify the Folder List, press File, highlight Folder List, and press enter.
Enter new Folder names or edit existing folders with the alpha/numeric entry
screen.
delete data
Press File, highlight Delete Data, and press enter. Enter the criteria for the
data you wish to delete in the Delete Data Filter, then highlight Delete Data and
press enter.
56
57
usInG the coMMunIcatIons saddle
ManaGe loGGed data
WARNING: DO NOT connect the Communications Saddle to
your PC before installing Data Manager. The Communication
Saddle drivers MUST be installed prior to connecting it to your
PC. The drivers will install automatically during the Data
u
A PC will recognize the Communications Saddle (saddle) as a YSI water quality
instrument with or without the Pro Plus installed in the saddle.
To connect the saddle to a Pro Plus, simply align the saddle to the oval section on
top of the instrument and push it down to snap it in place (Figure 6).
Figure 6. Locate the oval alignment groove at the top of the instrument
and inside the saddle. Insert the saddle into this oval groove. Press the
Manager installation. The first time the saddle is connected to
the PC, you may have to walk through a couple of installation
wizards. For detailed instruction, please refer to the Readme file
located on the CD that was included with your instrument.
saddle towards the back of the instrument until it snaps into place.
Data that has been logged to the Professional Plus can easily be uploaded to the
PC via the provided USB saddle. You can upload sensor data, GLP files, site lists,
and instrument configuration files individually or all at once. After connecting
the instrument to the PC via the USB saddle and cable and launching Data
Manager, click the Retrieve Instrument Data tab. Click on the Instrument’s
Unit ID you would like to retrieve data from, then select the files you would like
to retrieve and click Start.
Once the sensor data is uploaded to the PC, you can graph and view tabular data
by instrument Unit ID, date/time, site name, and/or folder name. This allows
you to configure the report according to your needs. You can choose to view
all data from all instruments, or select a certain date/time range for only a few
specific instruments, there are multiple ways to view the data. Once the report
has been defined, you will be able to print the graph and/or export the table.
Data Manager takes information management one step further and allows
you to delete specific points instead of entire files. This allows you to clean up
data that is no longer needed or that may have been collected erroneously, for
example, when the sensor was out of the water. If you can not delete data due
to regulation and compliance purposes, Data Manager has the solution. While
viewing logged data or real time data, you have the ability to ‘tag’ individual data
points with comments.
In addition to sensor data, you will be able to view GLP files, site lists, and
configuration files that have been uploaded from the instrument. These can be
printed and exported as well.
real tIMe studIes
Connect the USB cable to the top of the saddle and to a USB port on the PC.
Once Data Manager is launched, the program will recognize all saddles with
instruments connected to the PC.
The instrument will be powered through the saddle and USB connection when
connected to the PC. However, the instrument must still have batteries installed
in order to keep the date and time correct when powering the PC off at night.
Make sure the instrument is turned off first, then turn off the PC to keep the
instrument from running all night on the batteries. If you power it off and
power off the PC the instrument will keep the correct date and time if it has
batteries installed. If batteries are not installed, the instrument’s date and time
will not remain correct and will need to be reset each morning.
Data Manager allows you to view real time data on the PC.
After selecting your instrument, click the Real Time Instrument Data tab.
Next, input your sample interval, site/folder name, select the parameters you
wish to view and click OK. You must click Start on the next screen to begin
your real time study. Choose to hide the table or graph by unchecking the
box next to these options. Click Stop, then Edit Setup to change the Y-scale
min/max of the graph, to select different colors, or to name your graph. Add a
comment to a data point by clicking in the comment field of the table next to
the data point. You may also Print the graph and Export the data for viewing
in another program.
58
59
confIGure InstruMents
Data Manager allows for easy and quick configuration of single or multiple
instruments. Once you have uploaded a site list or configuration file, you can edit
it as needed, save it, and download it to other instruments. You no longer need
to configure each instrument individually. By using the same configuration file
for all instruments, you can rest assured that all instruments will have identical
settings.
New site lists and configuration files can be created in Data Manager as well.
These lists and files can be downloaded to one or multiple instruments. Save
time by creating these files on your PC and downloading them to the instrument
as opposed to creating them on the instrument.
C A R E , M A I N T E N A N C E ,
AND STORAGE
This section describes the proper procedures for care, maintenance and storage
of the sensors. The goal is to maximize their lifetime and minimize down-time
associated with improper sensor usage.
updatInG InstruMent fIrMWare
The instrument’s firmware can be updated via www.ysi.com. There you will find
the new firmware file and instructions on how to update the instrument. There
is no need to send the instrument back to the factory for upgrades.
General MaIntenance
GENERAL MAINTENANCE ORINGS
The instrument utilizes o-rings as seals to prevent water from entering the battery
compartment and sensor ports. Following the recommended procedures will
help keep your instrument functioning properly.
The same inspection should be made of the o-rings associated with the sensor
connectors when they are removed. If no dirt or damage to the o-rings is
evident, then they should be lightly greased without removal from their groove.
However, if there is any indication of damage, the o-ring should be replaced
with an identical o-ring. At the time of o-ring replacement, the entire o-ring
assembly should be cleaned.
To remove the o-rings:
Use a small, flat-bladed screwdriver or similar blunt-tipped tool to remove the
o-ring from its groove. Check the o-ring and the groove for any excess grease or
contamination. If contamination is evident, clean the o-ring and nearby plastic
parts with lens cleaning tissue or equivalent lint-free cloth. Alcohol can be used
to clean the plastic parts, but use only water and mild detergent on the o-ring
itself. Also, inspect the o-rings for nicks and imperfections.
Using alcohol on o-rings may cause a loss of elasticity and may
u
Before re-installing the o-rings, make sure to use a clean workspace, clean hands,
and avoid contact with anything that may leave fibers on the o-ring or grooves.
Even a very small amount of contamination (hair, grit, etc.) may cause a leak.
To re-install the o-rings:
Place a small amount of o-ring grease between your thumb and index finger.
(More grease is NOT BETTER!)
Draw the o-ring through the grease while pressing the fingers together to place
a very light covering of grease to the o-ring. Place the o-ring into its groove
making sure that it does not twist or roll.
Use your grease-coated finger to once again lightly go over the mating surface
of the o-ring.
promote cracking. Do not use a sharp object to remove the
o-rings. Damage to the o-ring or the groove may result.
If the o-rings and sealing surfaces are not maintained properly, it is possible that
water can enter the battery compartment and/or sensor ports of the instrument.
If water enters these areas, it can severely damage the battery terminals or sensor
ports causing loss of battery power, false readings, and corrosion to the sensors
or battery terminals. Therefore, when the battery compartment lid is removed,
the o-ring that provides the seal should be carefully inspected for contamination
(e.g. debris, grit, etc.) and cleaned if necessary.
u
Do not over-grease the o-rings. The excess grease may collect
grit particles that can compromise the seal. Excess grease
can also cause the waterproofing capabilities of the o-ring
to diminish, potentially causing leaks. If excess grease is
present, remove it using a lens cloth or lint-free cloth.
60
61
GENERAL MAINTENANCE SENSOR PORTS
It is important that the entire sensor connector end be dry when installing,
removing or replacing. This will prevent water from entering the port. Once
a sensor is removed, examine the connector inside the port. If any moisture is
present, use compressed air to completely dry the connector or place directly in
front of a steady flow of fresh air. If the connector is corroded, return the cable
to your dealer or directly to an YSI Repair Center.
Thread the membrane cap onto the sensor. It is normal for
a small amount of electrolyte to overflow.
i
Remove sensors upside down (facing the ground) to help
prevent water from entering the port upon removal.
sensor MaIntenance
SENSOR MAINTENANCE DISSOLVED OXYGEN
Membrane Cap Installation
The DO sensor (Polarographic and Galvanic) is shipped with a dry, protective
red cap that will need to be removed before using. Remove the protective cap
or used membrane cap and replace it with a new membrane cap following these
instructions:
Remove the sensor guard to access the sensor tip.
Unscrew and remove any old membrane cap by holding the
sensor when unscrewing the membrane cap and discard.
Thoroughly rinse the sensor tip with distilled or DI water.
Fill a new membrane cap with O2 sensor electrolyte
solution that has been prepared according to the
directions on the bottle. Be very careful not to
touch the membrane surface. Lightly tap the side
of the membrane cap to release bubbles that may
be trapped.
Polarographic Sensors - Model # 605203
The KCl (potassium chloride) solution and the membrane cap should be changed
at least once every 30 days during regular use. In addition, the KCl solution and
membrane should be changed if (a) bubbles are visible under the membrane; (b)
significant deposits of dried electrolyte are visible on the membrane; and (c) if
the sensor shows unstable readings or other sensor-related symptoms.
During membrane changes, examine the gold cathode at the tip of the sensor
and the silver anode along the shaft of the sensor. If either the silver anode is
black in color or the gold cathode is dull, the sensor may need resurfaced using
the fine sanding disks included in the membrane kit. Do not sand the electrode
every membrane change as this is not routine maintenance. In fact, visually,
the anode may appear tarnished and operate just fine. YSI recommends using
the 400 grit wet/dry sanding disks to resurface the electrodes if the sensor has
difficulty stabilizing or calibrating after a membrane change.
To resurface the sensor using the fine sanding disk, follow the instructions
belo w.
Gold Cathode:
For correct sensor operation, the gold cathode must be textured properly. It can
become tarnished or plated with silver after extended use. Never use chemicals
or abrasives not recommended or supplied by YSI.
First dry the sensor tip completely with lens cleaning tissue. Wet a sanding disk
with a small amount of clean water and place it face up in the palm of your hand.
Next, with your free hand, hold the sensor in a vertical position, tip down. Place
the sensor tip directly down on the sanding disk and twist it in a circular motion
to sand the gold cathode. The goal is to sand off any build-up and to lightly
scratch the cathode to provide a larger surface area for the O2 solution under
the membrane. Usually, 3 to 4 twists of the sanding disk are sufficient to remove
deposits and for the gold to appear to have a matte finish. Rinse thoroughly and
wipe the gold cathode with a wet paper towel before putting on a new membrane
62
63
cap. If the cathode remains tarnished, contact YSI Technical Support or the
Authorized dealer where you purchased the instrument.
Silver Anode
After extended use, a thick layer of Silver Chloride (AgCl) builds up on the
silver anode reducing the sensitivity of the sensor. The anode must be cleaned to
remove this layer and restore proper performance. The cleaning can be chemical
or mechanical:
Galvanic Sensors – Model # 605202
We recommend that the Sodium Chloride (NaCl) solution and the membrane
cap be changed at least once every 60 days during regular use. In addition, the
NaCl solution and membrane should be changed if (a) bubbles are visible under
the membrane; (b) significant deposits of dried electrolyte are visible around the
membrane; and (c) if the sensor shows unstable readings or other sensor-related
symptoms.
Chemical cleaning: Remove the membrane cap and rinse the electrodes with
deionized or distilled water. Soak the sensing anode section of the sensor in
a 14% ammonium hydroxide solution for 2 to 3 minutes or in a 3% ammonia
solution overnight for 8-12 hours (most household ammonia cleaners are
typically around 3%). Rinse heavily in cool tap water followed by a thorough
rinsing with distilled or deionized water. The anode should then be thoroughly
wiped with a wet paper towel to remove the residual layer from the anode. You
can smell the tip of the sensor to help ensure all the ammonia has been rinsed off.
Trapping residual ammonia under the new membrane cap can quickly tarnish
the electrode and/or give false readings.
Chemical cleaning should be performed as infrequently as possible.
First attempt a membrane change and recalibrate. If a new
i
membrane does not resolve the problem, then proceed with cleaning.
Mechanical cleaning: In order to sand the silver anode along the shaft of the
sensor, simply hold the sensor in a vertical position. Wet the sanding disk with
a small amount of clean water then gently wrap it around the sensor shaft and
twist it a few times to lightly sand the anode (the goal is to simply sand off any
build-up without scratching or removing layers of the anode itself ). Usually, 3
to 4 twists of the sanding disk are sufficient to remove deposits. However, in
extreme cases, more sanding may be required to regenerate the original silver
surface.
After completing the sanding procedure, repeatedly rinse the electrode with
clean water and wipe with lens cleaning tissue to remove any grit left by the
sanding disk. Thoroughly rinse the entire tip of the sensor with distilled or
deionized water and install a new membrane.
IMPORTANT: Be sure to: (1) Use only the fine sanding disks
provided and (2) Sand as mentioned in the above procedures.
Not adhering to either of these instructions can damage the
u
electrodes. If this procedure is unsuccessful, as indicated by
improper electrode performance, contact YSI Technical Support
or the Authorized dealer where you purchased the instrument.
The Galvanic dissolved oxygen sensor is continuously reducing oxygen even
when the display of the instrument is not active. This factor allows the sensor
to be used with no warm-up period as soon as the instrument is powered on
(instant on DO). However, because the sensor is “on” all the time, some solid
from the oxidation of the zinc anode will form in the electrolyte within 1-2 weeks
of activation. Small amounts of the solid will generally cause no performance
problems, but excessive amounts may result in jumpy dissolved oxygen readings.
The rate of solid formation is dependent on the type of membrane installed. The
formation of solids based on membrane type typically form more rapidly with
the 5912 (1 mil Teflon), less rapid with 5913 (1.25 mil PE), and least rapid with
5914 (2 mil PE).
The Galvanic DO sensor solution will appear milky white
after use but will NOT affect the accuracy of the sensor unless
i
there is excessive build up. The color change is acceptable
and normal as long as DO readings remain stable.
At the time the membrane cap is changed, YSI recommends that you rinse the
anode (silver shaft of the sensor) with purified water and wipe with a clean paper
towel. If white deposits are evident on the anode after cleaning, YSI recommends
that you remove this material by sanding the anode with the sandpaper disk
enclosed in your membrane kit. Follow the “Mechanical Cleaning” instructions
under the Polarographic Silver Anode section.
IMPORTANT: Be sure to: (1) Use only the fine sanding disks
provided and (2) Sand as mentioned in the above procedures. Not
adhering to either of these instructions can damage the electrodes.
u
WARNING: DO NOT PERFORM THE POLAROGRAPHIC
CHEMICAL CLEANING ON A GALVANIC SENSOR. If
this procedure is unsuccessful, as indicated by improper
electrode performance, contact YSI Technical Support or the
Authorized dealer where you purchased the instrument.
64
65
SENSOR MAINTENANCE CONDUCTIVITY
The openings that allow sample access to the conductivity electrodes should be
cleaned regularly. The small cleaning brush included in the Maintenance Kit is
ideal for this purpose. Dip the brush in clean water and insert it into each hole 10
to 12 times. In the event that deposits have formed on the electrodes, it may be
necessary to use a mild detergent (laboratory grade soap or bathroom foaming
tile cleaner) with the brush. Rinse thoroughly with clean water, then check the
response and accuracy of the conductivity cell with a calibration standard.
If this procedure is unsuccessful, as indicated by improper
i
electrode performance, contact YSI Technical Support or the
Authorized dealer where you purchased the instrument.
SENSOR MAINTENANCE TEMPERATURE
You must keep the temperature portion of the sensor free of build up. Other
than that, the sensor requires no maintenance. The conductivity cleaning brush
can be used to scrub the temperature sensor if needed. Alternatively, you can
use a toothbrush to clean the sensor.
If good pH and/or ORP response is not restored, perform the following additional
procedure:
Soak the sensor for 10-15 minutes in clean water containing a few 1.
drops of commercial dishwashing liquid.
GENTLY clean the glass bulb and platinum button by rubbing with 2.
a cotton swab soaked in the cleaning solution.
Rinse the sensor in clean water, wipe with a cotton swab saturated 3.
with clean water, and then rerinse with clean water.
If good pH and/or ORP response is still not restored, perform the following
additional procedure:
Soak the sensor for 30-60 minutes in one molar (1 M) hydrochloric 1.
acid (HCl). This reagent can be purchased from most lab supply
distributors. Be sure to follow the safety instructions included with
the acid.
Rinse the sensor in clean water, wipe with a cotton swab saturated 2.
with clean water (not DI water), and then rerinse with clean water.
To be certain that all traces of the acid are removed from the sensor
crevices, soak the sensor in clean water for about an hour with
occasional stirring.
SENSOR MAINTENANCE
H, ORP AND COMBINATION H/ORP
Typical working life for pH and ORP sensors is approximately 12-24
i
Cleaning is required whenever deposits or contaminants appear on the glass
and/or platinum surfaces or when the sensor’s response slows. The cleaning can
be chemical and/or mechanical.
Removing the sensor from the cable may make cleaning easier. Initially, use
clean water and a soft clean cloth, lens cleaning tissue, or cotton swab to remove
all foreign material from the glass bulb and/or platinum button. Then use a
moistened cotton swab to carefully remove any material that may be blocking
the reference electrode junction of the sensor.
months depending on usage, storage, and maintenance. Proper
storage and maintenance generally extends the sensor’s working life.
CAUTION: When using a cotton swab, be careful NOT to
wedge the swab between the guard and the glass sensor. If
u
necessary, remove cotton from the swab tip, so that the cotton
can reach all parts of the sensor tip without stress. You can
also use a pipe cleaner for this operation if more convenient.
If biological contamination of the reference junction is suspected or if good
response is not restored by the above procedures, perform the following
additional cleaning step:
Soak the sensor for approximately 1 hour in a 1:1 dilution of 1.
commercially-available chlorine bleach.
Rinse the sensor with clean water and then soak for at least 1 hour 2.
in clean water with occasional stirring to remove residual bleach
from the junction. (If possible, soak the sensor for a period of time
longer than 1 hour in order to be certain that all traces of chlorine
bleach are removed.) Then rerinse the sensor with clean water and
retest.
i
Dry the port and sensor connector with compressed air and apply a
very thin coat of o-ring lubricant to all o-rings before reinstallation.
SENSOR MAINTENANCE CHLORIDE
Typical working life for chloride sensors is approximately 3-6
i
The chloride sensor is considered a pellet membrane ISE. As always, when
handling sensors, care should be taken to avoid damaging the membrane. This
months depending on usage, storage, and maintenance. Proper
storage and maintenance generally extends the sensor’s working life.
66
67
sensor can be regenerated by washing with alcohol and/or gently polishing with
fine emery paper in a circular motion to remove any deposits or discoloration,
then thoroughly washing with deionized water to remove any debris. The
sensor may require soaking in the high standard chloride calibration solution to
recover its performance.
SENSOR MAINTENANCE AMMONIUM AND NITRATE
Long-term Storage - Conductivity
No special storage is required. Sensors can be stored dry or wet as long as
solutions in contact with conductivity electrodes are not corrosive (for example,
chlorine bleach). However, it is recommended that the sensor be cleaned with
the provided brush prior to and after long term storage.
Long-term Storage Temperature: -5 to 70°C (23 to 158°F)
Typical working life for ammonium and nitrate sensors is
approximately 3-6 months depending on usage, storage and
i
The ammonium and nitrate sensors are PVC membranes. As always, when
handling a sensor, care should be taken to avoid damaging the membrane. After
extensive use the membranes may become coated with a deposit or scoured
with fine scratches which may cause a slow or reduced response (low slope)
or unstable readings. Deposits may be removed with a fine jet of deionized
water or rinsing in alcohol followed by soaking in the high standard calibration
solution. Gently dab dry with a lint-free tissue before taking measurements.
maintenance. Proper storage and maintenance generally
extends the sensor’s working life.
sensor storaGe
SHORTTERM STORAGE
The cable assembly is supplied with a sensor storage container, or sleeve, that
attaches to the cable. The container is used for short-term storage (less than 30
days). Be sure to keep a small amount of moisture (tap water) in the container
during storage. This is done to maintain a 100% saturated air environment which
is ideal for short-term sensor storage. The sensors should not be submersed in
water. The intent is to create a humid air storage environment.
LONGTERM STORAGE
Long-term Storage - Temperature
No special storage is required. The temperature sensor can be stored dry or
wet as long as solutions in contact with the thermistor are not corrosive (for
example, chlorine bleach).
Long-term Storage Temperature: -5 to 70°C (23 to 158°F)
Long-term Storage - Dissolved Oxygen
Dissolved oxygen sensors (Polarographic and Galvanic) should be stored in a
dry state for long term storage First, remove the membrane cap and thoroughly
rinse the sensor with clean water. Next, either blow it dry with compressed air
or allow to air dry completely. Install a clean, dry new membrane cap over the
sensor to keep it dry and to protect the electrodes.
After storing the sensor for a long period of time, it is necessary to “condition” the
sensor by putting a new membrane with electrolyte solution on the sensor and
then turning the instrument on to allow the sensor sufficient time to stabilize.
Long-term Storage Temperature: -5 to 70°C (23 to 158°F)
Long-term Storage - pH
The key to pH sensor storage, short or long-term, is to make certain that the
sensor does not dry out. Sensors which have been allowed to dry out due to
improper storage procedures may be irreparably damaged by the dehydration
and will require replacement. You can try to rehydrate the sensor by soaking it
(preferably overnight) in a potassium chloride solution or a pH 4 buffer before
attempting to calibrate.
To store the sensor, remove it from the cable and seal the vacant port with a port
plug. Fill the original shipping/storage vessel (plastic boot or bottle) with buffer
4 solution and then submerge the sensor into the solution. The sensor should
remain submerged in the solution during the storage period; therefore, make
certain that the vessel is sealed to prevent evaporation and periodically check the
vessel to ensure the sensor does not dry out.
Long-term Storage Temperature: 0 to 30°C (32 to 86°F)
i
It is important not to store the pH sensor in distilled or deionized
water as the glass sensor may be damaged by exposure to this
medium.
68
69
Long-term Storage - ORP
error MessaGes
To store, remove the sensor from the cable and seal the vacant port with the
provided plug. Fill the original shipping/storage vessel (plastic boot or bottle)
with buffer 4 solution and then submerge the sensor into the solution. The sensor
should remain submerged in the solution during the storage period; therefore,
make certain that the vessel is sealed to prevent evaporation and periodically
check the vessel to ensure the sensor does not dry out.
Long-term Storage Temperature: 0 to 30°C (32 to 86°F)
Long-term Storage - Ammonium, Nitrate, and Chloride
The key to ISE sensor storage, short or long-term, is to make certain that the
sensor does not dry out. Sensor junctions that have been allowed to dry out due
to improper storage procedures may be irreparably damaged by the dehydration
and will require replacement. You can attempt to rehydrate the sensor by
soaking it (preferably overnight) in the sensor’s high calibration solution before
attempting to calibrate.
The recommended storage of these sensors is in moist air. Remove the sensor
from the cable and seal the vacant port with the provided plug. Place the sensor
in its original shipping storage vessel (plastic boot or bottle) with a small amount
of tap water or its high calibration standard. The vessel should remain a saturated
air environment. The sensor only needs to be kept in moist air, not submerged.
Make certain that the vessel is sealed to prevent evaporation.
Long-term Storage Temperature: 0 to 30°C (32 to 86°F)
If readings for a certain parameter are over range you will see a series of +++++
and if the readings are under range you will see a series of ----- plus the error
message along the bottom of the screen. If you see a series of ????? that will
indicate that a certain parameter can not be calculated. The following are
potential error messages:
Probe Temp over range
Probe Temp under range
Case Temp over range
Case Temp under range
pH over range
pH under range
ORP over range
ORP under range
Cl over range
Cl under range
NH4 over range
NH4 under range
NO3 over range
NO3 under range
DO over range
DO under range
Conductivity over range
Conductivity under range
Barometer over range
Barometer under range
TROUBLESHOOTING
Illegal Value may appear during alpha/numeric entry on the message line. This
only appears if the values entered do not match the formatting. This will also
appear in GLP security area if the password is incorrect.
If you forget the GLP Security Password please contact YSI Tech Support at
environmental@ysi.com, 800-897-4151, or +1 937 767-7241.
help
During use of the Professional Plus instrument, press Question from any
screen to view help messages directly on the display.
Error messages for the sensors typically indicate a need to properly clean the
sensor. First verify the sensor is properly setup in the Sensor menu, then conduct
the recommended cleaning and attempt to calibrate the sensor. If this does not
work, it may indicate the useful life of the sensor has been reached and may need
to be replaced. You may also contact Technical Support to help determine the
next step.
70
71
dIssolVed oxyGen
The dissolved oxygen sensors will use Probe Current (DO uA) and Probe Slope
(%/uA) as part of their GLP file records. The following information indicates
the acceptable values for each of these readings:
ACCESSORIES / PART NUMBERS
Cable Part Number*Description
6050000Professional Plus Instrument
Polarographic DO at 25 ºC, 100% saturated air environment at 760 mmHg
Probe Current
1.25 mil PE membrane
Average 6.15 uA (min. 4.31 uA, max. 8.00 uA)
2.0 mil PE membrane
Average 3.38 uA (min. 2.37 uA, max. 4.40 uA)
1 mil Teflon® membrane
Average 16.29 uA (min. 11.40 uA, max. 21.18 uA)
Probe Slope
1.25 mil PE membrane
Average 16.26 % sat/uA (min. 12.51 uA, max. 23.23
uA)
2.0 mil PE membrane
Average 29.56 % sat/uA (min. 22.74 uA, max. 42.23
uA)
1 mil Teflon® membrane Average 6.14 % sat/uA (min. 4.72 uA, max. 8.77 uA)
RESTORE DEFAULT
C A L I B R A T I O N VA L U E S
Occasionally, the instrument may need to have the factory calibration default
values restored. In order to accomplish this press Calibrate
Restore Default Cal and press enter. Highlight the parameter you wish to restore
to default and press enter. Next you will be asked to confirm the operation.
Highlight Ye s and press enter to confirm.
, highlight
60510-1, -4, -10, -20, or -301, 4, 10, 20, or 30-meter cable for ISE/temp
60520-1, -4, -10, -20, or -30**1, 4, 10, 20, or 30-meter cable for DO/temp
60530-1, 4, -10, -20, or -301, 4, 10, 20 or 30-meter cable for Cond/temp
6051010-1, 4, -10, -20, or -301, 4, 10, 20, or 30-meter cable
for ISE/ISE/temp
6051020-1, -4, -10, -20, or -30 1, 4, 10, 20, or 30-meter cable for ISE/DO/
temp
6051030-1, 4, -10, -20, or -301, 4, 10, 20, or 30-meter cable for ISE/Cond/
temp
6052030-1, -4, -10, -20, or -30 1, 4, 10, 20 or 30-meter cable for DO/Cond/
temp
605790-1, -4, -10, -20, or -301, 4, 10, 20 or 30-meter Quatro cable for DO/
Cond/temp/ISE/ISE
6051071-meter pH/temp single junction lab-grade
combo electrode
6051774-meter pH/temp single junction lab-grade
combo electrode
6051081-meter ORP/temp single junction lab-grade
combo electrode
6051784-meter ORP/temp single junction lab-grade
combo electrode
6051091-meter pH/ORP/temp single junction lab-
grade combo electrode
6051794-meter pH/ORP/temp single junction lab-
grade combo electrode
Sensor Part NumberDescription
605202Galvanic DO sensor
605203Polarographic DO sensor
605101pH (ISE)
605102ORP (ISE)
605103***pH/ORP Combination (ISE)
605104****Ammonium (ISE)
72
73
Sensor Part NumberDescription
605105****Chloride (ISE)
605106****Nitrate (ISE)
605780Self-Stirring BOD sensor
005560Conductivity/Temperature sensor for Quatro
cable
* All cables include temperature.
Cables with conductivity include sensor
(no need to order separate conductivity sensor).
** Special order cables up to 100-meters are available with 60520 cables.
*** Not compatible with 6051010-X or Quatro cables.
**** Freshwater only
Accessory Part NumberDescription
603059Flow cell, standard, 203 mL (for two-port sensors)
Accessory Part NumberDescription
605990Flow cell kit for Quatro cable assemblies.
603056Flow cell mounting spike
605604Communications saddle kit
605515Data Manager desktop software
603075Carrying case, soft-sided
603077Flow cell kit, 1 or 2 port sensor (includes 603059 flow
cell for two-port sensors with the 603078 adapter for
one-port sensors)
603078Flow cell adapter, single port (use with 603059 flow
The undersigned hereby declares on behalf of the named manufacturer under our
sole responsibility that the listed product conforms to the requirements for the listed
European Council Directive(s) and carries the CE mark accordingly.
Manufacturer:
Product Name:Professional Plus Water Quality Instrument
Model Numbers
Instrument/Accessory:Professional Plus (6050000) / ProComm (605604)
measurement, control, and laboratory use – EMC
requirements – Part 1: General Requirements
•EN61326-2-3:2006,Electricalequipmentfor
measurement, control and laboratory use – EMC
requirements – Part 2-3: Particular Requirements
– Test configuration, operational conditions,
and performance criteria for transducers with
integrated or remote signal conditioning.
•EN61000-3-2:2006,Electromagnetic
compatibility (EMC) – Part 3-2: Limits
– Limits for harmonic current emissions
(equipment input current < 16A per phase).
•EN61000-3-3:1995+A1:2001+A2:2005,
Electromagnetic compatibility (EMC) – Part
3: Limits – Section 3: Limitation of voltage
fluctuations and flicker in low-voltage supply
systems for equipment with rated current < 16A.
Supplementary
Information:
Authorized EU
Representative
All performance met the continuous
unmonitored operation criteria as follows:
1. ESD, EN61000-4-2, Performance Criterion B
2. Radiated Immunity, EN61000-43, Performance Criterion A
3. EFT, EN61000-4-4, (EFT) Performance Criterion B
4. Surge, EN61000-4-5, Performance Criterion B
5. Conducted Immunity, EN610004-6, Performance Criterion A
6. Voltage Interrupts, EN61000-411, Performance Criterion B
7. RF Emissions, EN55011:1998,
A1:1999 Class B equipment
YSI Hydrodata Ltd
Unit 8, Business Centre West, Avenue 1
Letchworth, Hertfordshire, SG6 2HB UK
The undersigned hereby declares on behalf of the named manufacturer under our
sole responsibility that the listed product conforms to the requirements for electrical
equipment under US FCC Part 15 and ICES-003 for unintentional radiators.
Tested using ANSI C63.4-2003 (excluding
sections 4.1, 5.2, 5.7, 9, and 14)
Signed: Lisa M. Abel Date: 22 February 2008
Title: Director of Quality
Signed: Lisa M. Abel Date: 22 February 2008
Title: Director of Quality
78
79
The undersigned hereby declares on behalf of the named manufacturer under
our sole responsibility that the listed product conforms with the Australian and
New Zealand Electromagnetic Compatibility (EMC) requirements for generic
products to be used in residential, commercial, and light industrial environments.
Compatibility (EMC) – Generic
emission standard – Part 1: Residential,
commercial, and light industry.
605790
RECYCLING
Batteries are removed and sent to battery recyclers for dedicated metals. •
When the time comes for you to recycle, follow the easy steps outlined at
www.ysi.com.
CONTACT INFORMATION
orderInG and technIcal support
Telephone: 800 897 4151 (US)
+1 937 767 7241 (Globally) Monday through Friday, 8:00 AM to 5:00 ET
Mail: YSI Incorporated 1725 Brannum Lane Yellow Springs, OH 45387 USA
Internet: www.ysi.com
When placing an order please have the following available:
1.) YSI account number (if available)
2.) Name and phone number
3.) Purchase Order or Credit Card
4.) Model Number or brief description
5.) Billing and shipping addresses
6.) Quantity Telephone: 800 897 4151 (US)
YSI is committed to reducing the environmental footprint in the course of
doing business. Even though materials reduction is the ultimate goal, we know
there must be a concerted effort to responsibly deal with materials after they’ve
served a long, productive life-cycle. YSI’s recycling program ensures that old
equipment is processed in an environmentally friendly way, reducing the amount
of materials going to landfills.
Printed Circuit Boards are sent to facilities that process and reclaim as much •
material for recycling as possible.
Plastics enter a material recycling process and are not incinerated or sent •
to landfills.
serVIce InforMatIon
YSI has authorized service centers throughout the United States and
Internationally. For the nearest service center information, please visit www.
ysi.com and click ‘Support’ or contact YSI Technical Support directly at 800897-4151.
When returning a product for service, include the Product Return form with
cleaning certification. The form must be completely filled out for a YSI Service
Center to accept the instrument for service. The form may be downloaded from
www.ysi.com by clicking on the ‘Support” tab, then the Product Return Form
button.