YSI Pro20 Quick Start Guide

probe solution according to the
2
Pro20
Pocket Guide
membrane from the sensor.
bubbles from later developing under the membrane.
SETTING SENSOR & MEMBRANE TYPE
4. Fill a new membrane cap with probe solution. Avoid
3. Thoroughly rinse the sensor tip with distilled or
2. Remove, and discard the red protective cap or used
6. Screw the probe sensor guard on moderately tight.
5. Thread the membrane cap onto the sensor, moderately
The instrument’s Sensor Type must be configured for the sensor
installed. Failure to do this may result in damage not covered
under warranty.
The instrument will step you through an initial configuration
when powered on for the first time. This allows you to set the
language, sensor, and membrane options. Use the up or
calibrating and operating the Pro20. It is not intended to
INSTALLING THE DO MEMBRANE
1. Prepare the O
instructions on the bottle. After mixing, allow the
solution to sit for 1 hour. This will help prevent air
tight. A small amount of electrolyte will overflow.
touching the membrane portion of the cap.
deionized water.
SALINITY COMPENSATION CALIBRATION
The Pro20 uses an inputted salinity value in ppt (parts per
thousand) to compensate dissolved oxygen mg/L values.
To adjust the salinity compensation value: This Pocket Guide is meant to serve as a quick reference in
1. Use the up or down arrow keys to highlight the salinity
compensation value to the salinity of the water you are
testing.
box on the run screen, then press enter.
2. Use the up or down arrow keys to adjust the salinity
compensation value.
1. Turn the instrument on and wait 5-15 minutes if using a
3. Press enter to confirm and to save the new salinity
TAKING MEASUREMENTS
Revision A
May 2008
instrument will confirm that the reading was successfully
the probe in the sample at a rate of at least 6 inches per
polarographic sensor.
second until the readings stabilize.
2. To take readings, insert the probe into the sample. Move
saved.
CONTACT INFORMATION
3. Highlight Save and press enter to store the reading. The
YSI
1700/1725 Brannum Lane
Yellow Springs, OH 45387
Tel: 800-765-4974 (937-767-7241) Item # 605645
Website: www.ysi.com
Fax: 937-767-1058 Drawing # A605645
E-Mail: proseries@ysi.com
calibrated with the press of one key when One Touch Cal is
enabled in the System Setup menu.
key for three seconds.
small amount of water and install it on the probe. The
cal/transport sleeve ensures venting to the atmosphere.
Make sure the DO and temperature sensors are not
The following procedure outlines the % saturation calibration
option with and without One Touch Cal enabled.
1. Moisten the sponge in the cal/transport sleeve with a
wait 10 minutes for the DO sensor to stabilize. Galvanic
immersed in the water.
2. Turn the instrument on. If using a polarographic sensor,
values along the bottom of the screen are accurate.
sensors do not require a warm up time.
3. Ensure the barometer reading and salinity correction
4. Press and hold the Cal
Calibrating %DO on the display and automatically
calibrate the sensor to the barometer reading and salinity
correction value.
5. If One Touch Cal is enabled, the instrument will indicate
DO% and temperature readings along with the %
calibration value. Wait at least 3 seconds, then, once the
DO% and temperature readings are stable, press enter to
and press enter. The Pro20 will display the current
6. If One Touch Cal is not enabled, highlight % or %Local
complete the calibration.
7. Calibration Successful will display for a few seconds to
display on the screen. Press the Cal key to exit the error
indicate a successful calibration and then the instrument
will return to the run screen.
message and return to the run screen.
8. If the calibration is unsuccessful, an error message will
down arrow keys to highlight the appropriate language,
sensor, and membrane, then press enter to confirm. To
change the sensor or membrane type after the initial
configuration, press the Menu key.
BAROMETER CALIBRATION
from a mercury barometer, an independent laboratory,
or from a local weather service. If the BP reading has
been corrected to sea level, use the following equation to
determine the true BP in mmHg for your altitude:
The barometer reading must be accurate to ensure accurate
DO % calibrations and DO readings.
If your barometer requires an adjustment:
1. Determine your local barometric pressure (BP) in mmHg
in feet/100)}
True BP = (Corrected BP in mmHG) – {2.5 * (Local Altitude
reading to the local, true barometric pressure.
highlight the barometer box then press enter.
4. Press enter to confirm and save the barometer
3. Use the up or down arrow keys to adjust the barometer
2. From the run screen, use the up or down arrow keys to
adjustment.
DO CALIBRATION
The Pro20 can be calibrated in % saturation, mg/L, or ppm.
Calibration of any option (%, mg/L, or ppm) will
automatically calibrate the others. The Pro20 can be
Loading...