OPERATING MANUAL
ScienceLine --- pHT - Pt ScienceLine --- pHT - G ScienceLine --- pHT - Mikro
MODE
INSTRUCCIONES DE
T
814244 04/2014
D’EMPLOI
OPERACIÓN
Science
Line
Science Line
Operating
Technical data .................................................................................................................. 5
Commissioning, measuring, calibration ........................................................................... 6
Storage ............................................................................................................................ 7
Aging ................................................................................................................................ 8
Maintenance and cleaning ............................................................................................... 8
Wear parts and accessories ............................................................................................. 9
Mode d'emploi ..................................................................................................................... 11
Caractéristiques techniques ........................................................................................... 13
Mise en service, mesure, calibration .............................................................................. 14
Stockage ........................................................................................................................ 15
Vieillissement ................................................................................................................. 16
Maintenance et nettoyage .............................................................................................. 16
Pièces d'usure et accessoires ........................................................................................ 17
Instrucciones de
Especificaciones técnicas .............................................................................................. 21
Puesta en funcionamiento, medición, calibración .......................................................... 22
Almacenamiento ............................................................................................................ 23
Envejecimiento ............................................................................................................... 24
Mantenimiento y limpieza ............................................................................................... 24
Piezas de desgaste y accesorios ................................................................................... 25
Contact Information......................................................................................... 27
Manual ......................................................................................................................
operación ....................................................................................................
3
19
ScienceLine --- pHT - Pt ScienceLine --- pHT - G ScienceLine --- pHT - Mikro
OPERATING MANUAL
T
04/2014
Science
pH
COMBINATION
ELECTRODES WITH
Line
SILAMID® DOUBLE JUNCTION
ScienceLine
Contact YSI
1725 Brannum Lane Yellow Springs, OH 45387 USA Tel: +1 937-767-7241
For the most recent version of the manual, please visit www.ysi.com.
800-765-4974 Email: environmental@ysi.com Internet: www.ysi.com
Copyright © 2014 Xylem Inc.
ScienceLine
Technical
data
T 814244
04/2014
5
YSI model
Reference electro-
Junction
NTC
Special features
Platinum
Yes
Silamid double junction
Science pHT-G
KCl 3 mol/L, Ag+ free
Ground
Yes
Silamid double junction
Micro
Platinum
Yes
YSI model
pH measur-
Allowed ture range
Membrane 25 °C
Typical
Science pHT-Pt
0 ... 14
-5 ... 100 °C
< 300 MOhm
Laboratory
Science pHT-G
0 ... 14
-5 ... 100 °C
< 300 MOhm
Laboratory
Micro
0 ... 14
-5 ... 100 °C
< 500 MOhm
Laboratory
YSI model
Shaft
Electrical connection
Length
Ø [mm]
Material
Combination connection
Meter
Cable
pHT-Pt
170
12
Glass
Cable
BNC + 4mm
1 m pHT-G
170
12
Glass
BNC + 4mm
1 m
Science pHT-Micro
200
12/5
Glass
Cable
BNC + 4mm
1 m
Science pHT-Pt
Science pHT-
Technical
General data
lyte KCl 3 mol/L, Ag+ free
KCl 3 mol/L, Ag+ free
Measurement and application characteristics
ing range
data
joint diaphragm
tempera-
Silamid double junction
resistance at
application
Science pHT-
Science
Science
Shaft dimensions, shaft material, electrical connection
[mm]
(70/130)
electrode
Cable
connection
length
6
T 814244
04/2014
Commissioning, measuring,
calibration
Commissioning, measuring,
calibration
Commissioning The electrodes are filled with reference electrolyte solution in the factory.
Prepare the combination electrode for measuring as follows:
The closure of the refilling opening consists of a slide. Open the refilling opening for the reference electrolyte solution.
The refilling opening must always be open during calibration and measurement!
1
2
Refilling opening
Remove the watering cap from the electrode tip. Possible salt deposits in the area of the watering cap do not affect the measuring characteristics and can easily be removed with deionized water.
Note Please keep the watering cap. It is required for the combination electrode to be stored. Always keep the watering cap clean.
Connect the combination electrode to the meter
Calibrate the combination electrode according to the operating manual of the meter. When doing so, observe the general rules on the following page.
Note (only Scinece pHT-G) When the Science pHT-G combination electrode is delivered, the ground sleeve is fixed with a shrinkable tubing. For commissioning, carefully remove the shrinkable tubing with a sharp knife or scalpel. To moisten the ground surfaces hold the combination electrode in a vertical position with the refill opening open and carefully loosen the ground junction by turning it slightly until reference electrolyte solution comes out of the junction. Then put the ground sleeve on again while turning it slightly.
Make sure the refilling opening for the reference electrolyte solution is open.
Avoid the displacement of any solution (sample or buffer solution) from one measurement to the next by taking the following measures:
- Shortly rinse the calibration and sample beakers with the solution the beakers are to be filled with next.
- Between measurements, rinse the combination electrode with the solution that follows. Alternatively, you can also rinse the combination electrode with deionized water and then carefully dab it dry.
ScienceLine
T 814244
04/2014
7
IoLine
ScienceLine
Overnight
longer
or
Immerse the combination electrode in the solution in a vertical or slightly tilted position.
Make sure the immersion depth is correct. The junction must be completely submersed in the solution. At the same time, the level of the reference electrolyte must be at least 2 cm above the level of the solution.
Provide approximately the same stirring conditions for measuring as for calibrating.
The frequency of subsequent calibrations depends on the application. Many meters provide an option where you can enter a calibration interval. After the calibration interval has expired, the meter will automatically remind you of the due calibration.
Storage
With the refilling opening open, immerse the combination electrode in in reference electrolyte (KCl 3 mol/L, Ag+ free). Observe the minimum depth of immersion. Prior to the next measurement, shortly rinse the combination electrode with the test sample or deionized water.
Note Prevent contact of the pH membrane to the beaker bottom to avoid scratches on the pH membrane.
Put the clean combination electrode in the watering cap that is filled with reference electrolyte (KCl 3 mol/L, Ag
Note pH combination electrodes must not be stored dry or in deionized water. The combination electrode could be permanently damaged by this. If the liquid in the irrigation cap has dried up, condition the combination electrode in reference electrolyte (KCl 3 mol/L, Ag+
Note During longer storing periods, salt sediments may develop on the watering cap. They do not affect the measuring characteristics and can easily be re­moved with deionized water when the combination electrode is put into oper­ation again.
+
free) and close the refilling opening.
+
free) for at least 24 hours.
Storage
8
T 814244
04/2014
Aging,
Maintenance
Refilling
the
reference
electrolyte
and cleaning
ScienceLine
Aging
Every pH With aging, the responding behavior becomes slower and the electrode slope and asymmetry change. Moreover, extreme operating conditions can considerably shorten the lifetime of the combination electrode. These are:
Strong acids or lyes, hydrofluoric acid, organic solvents, oils, fats, bro- mides, sulfides, iodides, proteins
High temperatures
High changes in pH and temperatures
The warranty does not cover failure caused by measuring conditions and mechanical damage.
Maintenance
During operation, a small amount of reference electrolyte leaks through the junction from the If the level of reference electrolyte becomes too low with time, refill it through the refilling opening. Refilling is very easy with the enclosed dropping bottle. Proceed as follows:
Cut off the tip of the dropping bottle at a right angle until the opening in the tip can be seen
Open the refilling opening of the combination electrode
Press the tip of the dropping bottle into the refilling opening while turning it slightly
Pump the reference electrolyte in the shaft using the dropping bottle
Pull the dropping bottle out of the refilling opening while turning it slightly if necessary
combination
and cleaning
combination
electrode
electrode into the test sample.
undergoes
a natural aging process.
T 814244
04/2014
9
Contamination
Cleaning procedure
F L
P
Description
Model
Order no.
Reference electrolyte solution 250 ml (
KCl-250
109 705Y
ScienceLine
Wear parts and accessories
Cleaning Remove water-soluble contamination by rinsing with deionized water. Re-
move other contamination as follows:
at and oil Rinse with water containing household
washing-up liquid ime and hydroxide deposits Rinse with citric acid (10 % by weight) roteins Immerse in pepsin cleaning solution PEP/pH
for approx. 1 hour.
Note: Make sure the level of the reference
electrolyte is above that of the cleaning
solution.
Note Hydrofluoric acid, hot phosphoric acid and strong alkaline solutions destroy the glass membrane.
After cleaning Rinse the combination electrode with deionized water and condition it in
reference electrolyte solution (KCl 3 mol/l, Ag
+
free) for at least 1 hour. Then
recalibrate the combination electrode.
Wear parts and accessories
KCl 3 mol/l, Ag+ free)