4310(W) Overview
1 Overview
1.1 Structure and function
Structure
Automatic sensor
recognition
1 Voltage electrode (inside, 2x)
2 Current electrode (ring, 2x)
3 Temperature sensor in graphite enclosure
4 Shaft
5 Connecting head with active electronics
6 Connection cable (4310)
7 IDS plug (4310W)
The sensor electronics with the stored sensor data is in the connecting head.
The data include, among other things, the sensor type and series number.
With each calibration, the calibration data is written in the sensor and the
calibration history is recorded. The data is recalled by the meter when the
sensor is connected and is used for measurement and for measured value
documentation. Storing the calibration data in the sensor ensures that the
correct cell constant is automatically used if the sensor is operated with
several meters.
The digital transmission technique guarantees the failure-free
communication with the meter even with long connection cables. If the sensor
firmware is enhanced by YSI, it can be updated via the meter.
ba76139d02 10/2017
1.2 Recommended fields of application
On site measurements in rivers, lakes and wastewater
Fish farming
Ground water measurements
Applications in water laboratories
3
Measuring / Operation 4310(W)
Due to the use of high performance PEEK material for the connection
technology 4310 has an improved chemical resistance, especially in acid
media (eg. As electroplating baths).
NOTE
Concentrated or strong oxidizing acids and organic solvents can
damage or destroy the sensor.
2 Measuring / Operation
2.1 Commissioning
Scope of delivery Conductivity measuring cell4310(W)
Operating manual
Commissioning Connect the electrode to the meter.
4310 – via the sensor cable to a free IDS connec-
tor on the meter
4310W – via a connecting cable (accessory) to a
free IDS connector on the meter
or
– wireless via an IDS WA-S adapter (acces-
sory) to a WA-capable meter
Accessories for the connection of the 4310W
sensor to the meter: See chapter 6 W
PARTS AND ACCESSORIES.
EAR
Opening and closing the IDS plug connection,
see section 2.2 O
PLUG CONNECTION (4310W).
IDS
PENING AND CLOSING THE
2.2 Opening and closing the IDS plug connection (4310W)
This section only applies to the IDS plug variant, 4310W.
4
ba76139d02 10/2017
4310(W) Cleaning
2
4
3
1
Plug connection
closed and locked
Guidance
Groove
Opening the plug
connection
Closing the plug
connection
If necessary, clean the plug connection
Open the locking device (step 1)
Use your thumb and index finger to press the clips of the connector
together, and pull the connector out of the plug (step 2).
Make sure that the plug connection is completely dry and clean.
Align the guidance of the connector with the groove in the plug and insert
the connector in the unlocked plug until it catches (step 3).
Close the locking device (step 4).
3 Cleaning
NOTE
Outside cleaning We recommend to clean the sensor thoroughly, especially before measuring
Prior to cleaning, disconnect the sensor from the meter.
low conductivity values.
ba76139d02 10/2017