The sensor electronics with the stored sensor data are in the connecting
head of the electrode. The data include, among other things, the sensor type
and series number. The data are recalled by the meter when the sensor is
connected and are used for measurement and for measured value documentation.
The digital transmission technique guarantees the failure-free communication with the meter even with long connection cables. The sensor firmware
can be updated via the meter.
2Commissioning, measuring, checking
2.1Commissioning
Scope of delivery Electrode 4210W or 4211
Operating manual
CommissioningPrepare the electrode for measuring as follows:
Open the refilling opening for the reference electrolyte solution. Depend-
ing on the model, the stopper of the refilling opening is an elastomer stopper or a slider.
The refilling opening must always be open during measurement!
ba76138d0506/2018
Remove the watering cap from the electrode tip. Possible salt deposits in
the area of the watering cap do not affect the measuring characteristics
and can easily be removed with deionized water.
Please keep the watering cap. It is required for the electrode to
be stored. Always keep the watering cap clean.
Connect the sensor to the meter.
The sensor is immediately ready to measure.
4211– via the sensor cable to a free IDS connec-
tor on the meter
3
Commissioning, measuring, checking4210W / 4211
4210W– via a connecting cable (accessory) to a
free IDS connector on the meter
or
– wireless via an IDS WA-S adapter (acces-
sory) to a WA-capable meter
Accessories for the connection of the 4211W
sensor to the meter: See chapter 7 W
PARTSANDACCESSORIES.
EAR
Opening and closing the IDS plug connection,
see section 2.2O
PLUGCONNECTION.
PENINGANDCLOSINGTHE IDS
Measure with the electrode according to the operating manual of the meter
and observe the following rules while doing so:
4
ba76138d0506/2018
4210W / 4211Commissioning, measuring, checking
2
4
3
1
Plug connection
closed and locked
Guidance
Groove
2.2Opening and closing the IDS plug connection
This section only applies to variants with IDS plug (4210W).
Opening the plug
connection
Closing the plug
connection
If necessary, clean the plug connection
Open the locking device (step 1)
Use your thumb and index finger to press the clips of the connector togeth-
er, and pull the connector out of the plug (step 2).
Make sure that the plug connection is perfectly dry and clean.
Align the guidance of the coupling to the groove in the plug and insert the
coupling into the unlocked plug up to the stop (step 3)
Close the locking device (step 4)
ba76138d0506/2018
5
Commissioning, measuring, checking4210W / 4211
2.3Measuring: General rules
Make sure the refilling opening for the reference electrolyte solution is
open.
Avoid the carryover of sample solution from one measurement to the next
as follows:
– Shortly rinse the sample beakers with the solution the beakers are to
be filled with next.
– Between measurements, rinse the electrode with the solution that fol-
lows. Alternatively, you can also rinse the electrode with deionized
water and then carefully dab it dry.
Immerse the electrode in the solution in a vertical or slightly tilted position.
Make sure the immersion depth is correct. The junction must be com-
pletely submersed in the solution. The junction is in the area of the bottom
end of the shaft (see arrow).
Conversion to normal hydrogen elec-
trode
4210W / 4211
At the same time, the level of the reference electrolyte must be at least 2
cm above the level of the solution.
U
= U
H
with:U
Meas
H
U
Meas
U
Ref
+ U
Ref
= ORP, referring to the normal hydrogen electrode
= Measured ORP
= Voltage of the reference system compared to the normal
hydrogen electrode
U
is temperature dependent and can be taken from the following table (see
Ref
also DIN 38404-6):
T (°C)U
[mV]T (°C)U
Ref
Ref
[mV]
4210W / 42114210W / 4211
0+22435+200
5+22140+196
10+21745+192
6
ba76138d0506/2018
4210W / 4211Storage
Checking with
Zobell ORP buffer
solution
(YSI 3682)
T (°C)U
[mV]T (°C)U
Ref
Ref
[mV]
4210W / 42114210W / 4211
15+21450+188
20+21155+184
25+20760+180
30+203
URef is temperature dependent and can be taken from the following table:
T (°C)U [mV]T (°C)U [mV]
4210W / 42114210W / 4211
0+26230+214
5+25435+206
10+24640+199
15+23845+191
20+23050+183
During short mea-
suring breaks
Overnight or lon-
ger
25+222
3Storage
Immerse the electrode in the reference electrolyte with the refilling opening
open.
electrodeReference electrolyte Model
(see page 10)
4210W / 42113 mol/l KCl, Ag
Prior to the next measurement, shortly rinse the electrode with the test
sample or deionized water.
Insert the clean electrode into the watering cap filled with reference electrolyte and shut the refilling opening.
During longer storing periods, salt sediments may develop on the
watering cap. They do not affect the measuring characteristics
and can easily be removed with deionized water when the electrode is put into operation again.
+
freeKCl-250 (250 ml)
ba76138d0506/2018
7
Aging4210W / 4211
4Aging
ORP electrodes are consumables. Every ORP electrode undergoes a natural
aging process. Extreme operating conditions can considerably shorten the
lifetime of the electrode. These are:
The warranty does not cover failure caused by measuring conditions and
mechanical damage.
5Maintenance and cleaning
During operation, a small amount of reference electrolyte leaks through the
junction from the electrode into the test sample. If the level of reference electrolyte becomes too low with time, refill it through the refilling opening.
Refilling the refer-
ence electrolyte
Refilling is very easy using a dropping bottle. Proceed as follows:
Cut off the tip of the dropping bottle at a right angle until the opening in the
tip can be seen
Open the refilling opening of the electrode
Press the tip of the dropping bottle into the refilling opening while turning
it slightly
Pump several small quantities of the reference electrolyte into the stem
using the dropper bottle
Pull the dropping bottle out of the refilling opening while turning it slightly
as necessary.
CleaningRemove water-soluble contamination by rinsing with deionized water.
8
ba76138d0506/2018
4210W / 4211Technical data
Remove other contamination as follows:
ContaminationCleaning procedure
Fat and oilRinse with water containing household washing-up
liquid
Lime and hydroxide
Rinse with citric acid (10 % by weight)
deposits
After cleaningRinse the electrode with deionized water.
6Technical data
Measurement and
application char-
acteristics
Accuracy of the
IDS measuring
technique
General dataReference electrolyte3 mol/l KCl, Ag
mV measuring range- 1250.0 ... + 1250.0
Allowed temperature range0 ... 100 °C (32 ... 212 °F)
Typical applicationLaboratory
Measured parameterAccuracy (± 1 digit)
U [mV]± 0.2
T [°C]± 0.1
JunctionCeramic
Double junctionYes
Electrode material and shapePlatinum / circle
+
free
Connection cableLengths4211: 1.5 m
4210W: 1.5 / 3 m
Diameter4.3 mm
Smallest allowed
bend radius
Fixed installation:20 mm
Flexible use:60 mm
Plug typeSocket, 4 -Ppoles
Shaft dimensions,
material
Shaft length 120 mm
Shaft diameter12 mm
Shaft materialGlass
IDS plug Plastic parts: Glass fiber reinforced
Noryl, TPU, TPC-ET, POM, PVC,
PEEK, PBT
O-ring: FPM
Contacts gold-plated
ba76138d0506/2018
9
Wear parts and accessories4210W / 4211
IDS plugConnection type4-P-pin, watertight plug connection with
locking device, reverse polarity protected
StorageWith watering cap; filled with KCl 3 mol/L, Ag+ free
7Wear parts and accessories
Maintenance
equipment
Connection cable
4211 - meter
Radio connection
4210W - meter
DescriptionModelOrder no.
Reference electrolyte solution 250 ml
to fill the watering cap
(KCl 3 mol/l, Ag
Zobell ORP buffer solution (125 ml)3682061320Y
DescriptionModelOrder no.
IDS connection cable, 1.5 mIDS-CABLE-
IDS connection cable, 3 mIDS-CABLE-3903 851Y
DescriptionModelOrder no.
WA capable IDS meter
+ radio module for IDS meter
Radio module for plug head sensorIDS WA-S108 141Y
+
-free)
KCl-250109 705Y
903 850Y
1.5
see Internet
10
8Disposal
At the end of its operational lifetime, the electrode must be returned to the
disposal or return system statutory in your country (electronic waste). If you
have any questions,
please contact your supplier.
ba76138d0506/2018
4210W / 4211Contact Information
Contact Information
Ordering & Technical Support
Telephone
Fax
:(937) 767-1058
Email
Mail:YSI Incorporated
Internet
When placing an order please have the following information available:
YSI account number (if available)Name and Phone Number
Model number or brief descriptionBilling and shipping address
QuantityPurchase Order or Credit Card
:(800) 897-4151
(937) 767-7241
Monday through Friday, 8:00 AM to 5:00 PM ET
:info@ysi.com
1725 Brannum Lane
Yellow Springs, OH 45387
USA
:www.ysi.com
Service Information
YSI has authorized service centers throughout the United States and
Internationally. For the nearest service center information, please visit
www.ysi.com
directly at 800-897-4151.
When returning a product for service, include the Product Return form
with cleaning certification. The form must be completely filled out for an
YSI Service Center to accept the instrument for service. The Product
Return form may be downloaded at www.ysi.com
‘Support‘ tab.
and click ‘Support’ or contact YSI Technical Support
and clicking on the
ba76138d0506/2018
11
1) The tissue in plants that brings water upward from the roots;
2) a leading global water technology company.
We're 12,500 people unified in a common purpose: creating innovative solutions
to meet our world's water needs. Developing new technologies that will improve
the way water is used, conserved, and re-used in the future is central to our work.
We move, treat, analyze, and return water to the environment, and we help people
use water efficiently, in their homes, buildings, factories and farms. In more than
150 countries, we have strong, long-standing relationships with customers who
know us for our powerful combination of leading product brands and applications
expertise, backed by a legacy of innovation.
For more information on how Xylem can help you, go to www.xyleminc.com
Xylem |'m|zīlə
YSI
1725 Brannum Lane
Yellow Springs, OH 45387
Tel: +1 937-767-7241; 800-765-4974
Fax: +1 937-767-1058
Email: info@ysi.com
Web: www.ysi.com