YSI 4211, 4210W Operating Manual

OPERATIONS MANUAL
ba76138e05 06/2018
4210W 4211
ORP ELECTRODE
4210W / 4211
For the most recent version of the manual, please visit
www.ysi.com
.
Contact YSI
800-765-4974 Email: info@ysi.com Internet: www.ysi.com
Copyright © 2018 Xylem Inc.
2 ba76138e05 06/2018
4210W / 4211 General information
Refilling opening

1 General information

Automatic sensor
recognition
The sensor electronics with the stored sensor data are in the connecting head of the electrode. The data include, among other things, the sensor type and series number. The data are recalled by the meter when the sensor is connected and are used for measurement and for measured value documen­tation.
The digital transmission technique guarantees the failure-free communica­tion with the meter even with long connection cables. The sensor firmware can be updated via the meter.

2 Commissioning, measuring, checking

2.1 Commissioning

Scope of delivery Electrode 4210W or 4211
Operating manual
Commissioning Prepare the electrode for measuring as follows:
Open the refilling opening for the reference electrolyte solution. Depend-
ing on the model, the stopper of the refilling opening is an elastomer stop­per or a slider.
The refilling opening must always be open during measurement!
ba76138d05 06/2018
Remove the watering cap from the electrode tip. Possible salt deposits in
the area of the watering cap do not affect the measuring characteristics and can easily be removed with deionized water.
Please keep the watering cap. It is required for the electrode to be stored. Always keep the watering cap clean.
Connect the sensor to the meter.
The sensor is immediately ready to measure.
4211 – via the sensor cable to a free IDS connec-
tor on the meter
3
Commissioning, measuring, checking 4210W / 4211
4210W – via a connecting cable (accessory) to a
free IDS connector on the meter
or – wireless via an IDS WA-S adapter (acces-
sory) to a WA-capable meter
Accessories for the connection of the 4211W sensor to the meter: See chapter 7 W
PARTS AND ACCESSORIES.
EAR
Opening and closing the IDS plug connection, see section 2.2O
PLUG CONNECTION.
PENING AND CLOSING THE IDS
Measure with the electrode according to the operating manual of the meter
and observe the following rules while doing so:
4
ba76138d05 06/2018
4210W / 4211 Commissioning, measuring, checking
2
4
3
1
Plug connection closed and locked
Guidance
Groove

2.2 Opening and closing the IDS plug connection

This section only applies to variants with IDS plug (4210W).
Opening the plug
connection
Closing the plug
connection
If necessary, clean the plug connection
Open the locking device (step 1)
Use your thumb and index finger to press the clips of the connector togeth-
er, and pull the connector out of the plug (step 2).
Make sure that the plug connection is perfectly dry and clean.
Align the guidance of the coupling to the groove in the plug and insert the
coupling into the unlocked plug up to the stop (step 3)
Close the locking device (step 4)
ba76138d05 06/2018
5
Commissioning, measuring, checking 4210W / 4211

2.3 Measuring: General rules

Make sure the refilling opening for the reference electrolyte solution is
open.
Avoid the carryover of sample solution from one measurement to the next
as follows:
– Shortly rinse the sample beakers with the solution the beakers are to
be filled with next.
– Between measurements, rinse the electrode with the solution that fol-
lows. Alternatively, you can also rinse the electrode with deionized water and then carefully dab it dry.
Immerse the electrode in the solution in a vertical or slightly tilted position.
Make sure the immersion depth is correct. The junction must be com-
pletely submersed in the solution. The junction is in the area of the bottom end of the shaft (see arrow).
Conversion to nor­mal hydrogen elec-
trode
4210W / 4211
At the same time, the level of the reference electrolyte must be at least 2 cm above the level of the solution.
U
= U
H
with: U
Meas
H
U
Meas
U
Ref
+ U
Ref
= ORP, referring to the normal hydrogen electrode
= Measured ORP
= Voltage of the reference system compared to the normal
hydrogen electrode
U
is temperature dependent and can be taken from the following table (see
Ref
also DIN 38404-6):
T (°C) U
[mV] T (°C) U
Ref
Ref
[mV]
4210W / 4211 4210W / 4211
0 +224 35 +200
5 +221 40 +196
10 +217 45 +192
6
ba76138d05 06/2018
4210W / 4211 Storage
Checking with
Zobell ORP buffer
solution
(YSI 3682)
T (°C) U
[mV] T (°C) U
Ref
Ref
[mV]
4210W / 4211 4210W / 4211
15 +214 50 +188
20 +211 55 +184
25 +207 60 +180
30 +203
URef is temperature dependent and can be taken from the following table:
T (°C) U [mV] T (°C) U [mV]
4210W / 4211 4210W / 4211
0 +262 30 +214
5 +254 35 +206
10 +246 40 +199
15 +238 45 +191
20 +230 50 +183
During short mea-
suring breaks
Overnight or lon-
ger
25 +222

3Storage

Immerse the electrode in the reference electrolyte with the refilling opening open.
electrode Reference electrolyte Model
(see page 10)
4210W / 4211 3 mol/l KCl, Ag
Prior to the next measurement, shortly rinse the electrode with the test sample or deionized water.
Insert the clean electrode into the watering cap filled with reference electro­lyte and shut the refilling opening.
During longer storing periods, salt sediments may develop on the watering cap. They do not affect the measuring characteristics and can easily be removed with deionized water when the elec­trode is put into operation again.
+
free KCl-250 (250 ml)
ba76138d05 06/2018
7
Aging 4210W / 4211

4Aging

ORP electrodes are consumables. Every ORP electrode undergoes a natural aging process. Extreme operating conditions can considerably shorten the lifetime of the electrode. These are:
Strong acids or lyes, hydrofluoric acid, organic solvents, oils, fats, bro-
mides, sulfides, iodides, proteins
High temperatures
High changes in pH and temperature.
The warranty does not cover failure caused by measuring conditions and mechanical damage.

5 Maintenance and cleaning

During operation, a small amount of reference electrolyte leaks through the junction from the electrode into the test sample. If the level of reference elec­trolyte becomes too low with time, refill it through the refilling opening.
Refilling the refer-
ence electrolyte
Refilling is very easy using a dropping bottle. Proceed as follows:
Cut off the tip of the dropping bottle at a right angle until the opening in the
tip can be seen
Open the refilling opening of the electrode
Press the tip of the dropping bottle into the refilling opening while turning
it slightly
Pump several small quantities of the reference electrolyte into the stem
using the dropper bottle
Pull the dropping bottle out of the refilling opening while turning it slightly
as necessary.
Cleaning Remove water-soluble contamination by rinsing with deionized water.
8
ba76138d05 06/2018
4210W / 4211 Technical data
Remove other contamination as follows:
Contamination Cleaning procedure
Fat and oil Rinse with water containing household washing-up
liquid
Lime and hydroxide
Rinse with citric acid (10 % by weight)
deposits
After cleaning Rinse the electrode with deionized water.

6 Technical data

Measurement and
application char-
acteristics
Accuracy of the
IDS measuring
technique
General data Reference electrolyte 3 mol/l KCl, Ag
mV measuring range - 1250.0 ... + 1250.0
Allowed temperature range 0 ... 100 °C (32 ... 212 °F)
Typical application Laboratory
Measured parameter Accuracy (± 1 digit)
U [mV] ± 0.2
T [°C] ± 0.1
Junction Ceramic
Double junction Yes
Electrode material and shape Platinum / circle
+
free
Connection cable Lengths 4211: 1.5 m
4210W: 1.5 / 3 m
Diameter 4.3 mm
Smallest allowed bend radius
Fixed installation:20 mm Flexible use:60 mm
Plug type Socket, 4 -Ppoles
Shaft dimensions,
material
Shaft length 120 mm
Shaft diameter 12 mm
Shaft material Glass
IDS plug Plastic parts: Glass fiber reinforced
Noryl, TPU, TPC-ET, POM, PVC, PEEK, PBT
O-ring: FPM
Contacts gold-plated
ba76138d05 06/2018
9
Wear parts and accessories 4210W / 4211
IDS plug Connection type 4-P-pin, watertight plug connection with
locking device, reverse polarity protected
Storage With watering cap; filled with KCl 3 mol/L, Ag+ free

7 Wear parts and accessories

Maintenance
equipment
Connection cable
4211 - meter
Radio connection
4210W - meter
Description Model Order no.
Reference electrolyte solution 250 ml to fill the watering cap (KCl 3 mol/l, Ag
Zobell ORP buffer solution (125 ml) 3682 061320Y
Description Model Order no.
IDS connection cable, 1.5 m IDS-CABLE-
IDS connection cable, 3 m IDS-CABLE-3 903 851Y
Description Model Order no.
WA capable IDS meter + radio module for IDS meter
Radio module for plug head sensor IDS WA-S 108 141Y
+
-free)
KCl-250 109 705Y
903 850Y
1.5
see Internet
10

8 Disposal

At the end of its operational lifetime, the electrode must be returned to the disposal or return system statutory in your country (electronic waste). If you have any questions, please contact your supplier.
ba76138d05 06/2018
4210W / 4211 Contact Information

Contact Information

Ordering & Technical Support

Telephone
Fax
: (937) 767-1058
Email
Mail: YSI Incorporated
Internet
When placing an order please have the following information available:
YSI account number (if available) Name and Phone Number Model number or brief description Billing and shipping address Quantity Purchase Order or Credit Card
: (800) 897-4151
(937) 767-7241 Monday through Friday, 8:00 AM to 5:00 PM ET
: info@ysi.com
1725 Brannum Lane Yellow Springs, OH 45387 USA
: www.ysi.com

Service Information

YSI has authorized service centers throughout the United States and Internationally. For the nearest service center information, please visit
www.ysi.com
directly at 800-897-4151.
When returning a product for service, include the Product Return form with cleaning certification. The form must be completely filled out for an YSI Service Center to accept the instrument for service. The Product Return form may be downloaded at www.ysi.com ‘Support‘ tab.
and click ‘Support’ or contact YSI Technical Support
and clicking on the
ba76138d05 06/2018
11
1) The tissue in plants that brings water upward from the roots;
2) a leading global water technology company.
We're 12,500 people unified in a common purpose: creating innovative solutions to meet our world's water needs. Developing new technologies that will improve the way water is used, conserved, and re-used in the future is central to our work. We move, treat, analyze, and return water to the environment, and we help people use water efficiently, in their homes, buildings, factories and farms. In more than 150 countries, we have strong, long-standing relationships with customers who know us for our powerful combination of leading product brands and applications expertise, backed by a legacy of innovation.
For more information on how Xylem can help you, go to www.xyleminc.com
Xylem |' m|zīlə
YSI 1725 Brannum Lane Yellow Springs, OH 45387 Tel: +1 937-767-7241; 800-765-4974 Fax: +1 937-767-1058 Email: info@ysi.com Web: www.ysi.com
©Xylem Inc
Loading...