The sensor electronics with the stored sensor data is in the connecting head
of the electrode. The data include, among other things, the sensor type and
series number. The data is recalled by the meter when the sensor is
connected and is used for measurement and for measured value documentation.
The digital transmission technique guarantees the failure-free communication with the meter even with long connection cables. If the sensor firmware
is enhanced by YSI, it can be updated via the meter.
Technical data
General dataReference electrolyte3 mol/l KCl, Ag+free
JunctionCeramic
Measurement and
application
characteristics
Electrode material and
shape
mV measuring range- 1250.0 ... + 1250.0
Allowed temperature range0 ... 100 °C (32 ... 212 °F)
Typical applicationLaboratory
Platinum / circle
Shaft dimensions,
material, electrical
connection
Connection cableLength1.5 m
Accuracy of the
IDS measuring
technique
Shaft length 120 mm
Shaft diameter12 mm
Shaft materialGlass
Combination electrode connection
Meter connectionDigital plug
Diameter4.3 mm
Smallest allowed bend
radius
Plug typeSocket, 4 pins
Measured parameterAccuracy (± 1 digit)
U [mV]± 0.2
Fixed cable
Fixed installation:20 mm
Flexible use:60 mm
ba76138e0102/2013
3
Page 4
Commissioning, measuring, checking4210
Refilling
opening
Commissioning, measuring, checking
Note
For ORP electrodes with platinum electrode, please follow the instructions in
section A
CommissioningPrepare the electrode for measuring as follows:
 Open the refilling opening for the reference electrolyte solution. Depend-
ing on the model, the stopper of the refilling opening is an elastomer stopper or a slider.
The refilling opening must always be open during measurement!
CTIVATINGPLATINUMELECTRODES.
General rules for
measuring
 Remove the watering cap from the electrode tip. Possible salt deposits in
the area of the watering cap do not affect the measuring characteristics
and can easily be removed with deionized water.
Note
Please keep the watering cap. It is required for the electrode to be stored.
Always keep the watering cap clean.
 Connect the combination electrode to the meter.
 Measure with the electrode according to the operating manual of the meter
and observe the following rules while doing so:
 Make sure the refilling opening for the reference electrolyte solution is
open.
 Avoid the carryover of sample solution from one measurement to the next
as follows:
– Shortly rinse the sample beakers with the solution the beakers are to
be filled with next.
– Between measurements, rinse the electrode with the solution that fol-
lows. Alternatively, you can also rinse the electrode with deionized water and then carefully dab it dry.
 Immerse the electrode in the solution in a vertical or slightly tilted position.
4
 Make sure the immersion depth is correct. The junction must be complete-
ly submersed in the solution. The junction is in the area of the bottom end
of the shaft (see arrow).
ba76138e0102/2013
Page 5
4210Commissioning, measuring, checking
4210
At the same time, the level of the reference electrolyte must be at least 2
cm above the level of the solution.
Conversion to
normal hydrogen
electrode
T (°C)T (°F)U
032+2243595+200
541+22140104+196
1050+21745113+192
1559+21450122+188
U
= U
H
with:U
Meas
H
U
Meas
U
Ref
+ U
Ref
= ORP, referring to the normal hydrogen electrode
= Measured ORP
= Voltage of the reference system compared to the normal
hydrogen electrode
U
is temperature dependent and can be taken from the following table (see
Ref
also DIN 38404-6):
[mV]T (°C)T (°F)U
Ref
Ref
[mV]
42104210
2068+21155131+184
2577+20760140+180
3086+203
ba76138e0102/2013
5
Page 6
Commissioning, measuring, checking4210
Checking with
Zobell ORP buffer
solution (YSI 3682)
U is temperature dependent and can be taken from the following table:
T (°C)U [mV]T (°C)U [mV]
42104210
0+26230+214
5+25435+206
10+24640+199
15+23845+191
20+23050+183
25+222
6
ba76138e0102/2013
Page 7
4210Aging
Storage
During short
measuring breaks
Overnight or
longer
Immerse the electrode in the reference electrolyte with the refilling opening
open.
ElectrodeReference electrolyteModel
(see page 10)
+
42103 mol/l KCl, Ag
Prior to the next measurement, shortly rinse the electrode with the test
sample or deionized water.
Insert the clean electrode into the watering cap filled with reference
electrolyte and shut the refilling opening.
Note
During longer storing periods, salt sediments may develop on the watering
cap. They do not affect the measuring characteristics and can easily be removed with deionized water when the electrode is put into operation again.
freeKCl-250 (250 ml)
Aging
Every ORP electrode undergoes a natural aging process. Extreme operating
conditions can considerably shorten the lifetime of the electrode. These are:
The warranty does not cover failure caused by measuring conditions and
mechanical damage.
ba76138e0102/2013
7
Page 8
Aging4210
Maintenance and cleaning
During operation, a small amount of reference electrolyte leaks through the
junction from the electrode into the test sample. If the level of reference
electrolyte becomes too low with time, refill it through the refilling opening.
Refilling the
reference
electrolyte
Refilling is very easy using a dropping bottle. Proceed as follows:
 Cut off the tip of the dropping bottle at a right angle until the opening in the
tip can be seen
 Open the refilling opening of the electrode
 Press the tip of the dropping bottle into the refilling opening while turning
it slightly
 Pump several small quantities of the reference electrolyte into the stem
using the dropper bottle
 Pull the dropping bottle out of the refilling opening while turning it slightly
as necessary.
8
ba76138e0102/2013
Page 9
4210Activating platinum electrodes
1000
800
600
400
200
20481012146
0
pH value
U
H
in mV
oxidizing pretreatment
reducing pretreatment
CleaningRemove water-soluble contamination by rinsing with deionized water. Re-
move other contamination as follows:
ContaminationCleaning procedure
Fat and oilRinse with water containing household wash-
ing-up liquid
Lime and hydroxide deposits Rinse with citric acid (10 % by weight)
After cleaningRinse the electrode with deionized water.
Activating platinum electrodes
First-time
activation during
installation and as
required
Activation during
very long set-up
times
For first-time activation use the activation powder from the SORT/RH reagent
set (component of the SORT/RH reagent set). Immerse the moist (but not
dripping) platinum electrode into the activation powder and rotate the
electrode several times in the powder. Then remove the activation powder
under flowing water with a soft brush (e.g. toothbrush).
When changing from oxidizing to reducing test solutions and vice versa this
can result in set-up times that can take significantly more than an hour. In this
case pretreatment (activation) of the platinum surface can shorten the set-up
time. The type of pretreatment (reducing or oxidizing) is based on the pH value and the ORP voltage (U
) of the test solution where the latter must be es-
H
timated for the first measurement.
The type of pretreatment can then be determined using the following diagram
where U
is based on the normal hydrogen electrode:
H
ba76138e0102/2013
9
Page 10
Wear parts and accessories4210
Oxidizing pretreatment
Immerse the platinum electrode for two to three
days in a sulfuric acid clorina solution (0,5 g clorina
powder, 100 ml H
O dist., 2-3 ml 20% sulfuric acid).
2
Clorina powder for producing the solution is included in the SORT/RH reagent set.
The diaphragm must not be immersed in the
Note:
clorina solution!
Reducing pretreatment
When the electrode is ready for the test immerse it
in the Zobell ORP buffer solution and wait for a stable measured value.
Wear parts and accessories
DescriptionModelOrder no.
Reference electrolyte solution 3 mol/l KCl,
+
free (250 ml)
Ag
Reagent set for regenerating ORP platinum
electrodes, consisting of 10 g activation powder and 30 g clorina powder
KCl-250109 705Y
SORT/RH109 730Y
Zobell ORP buffer solution (125 ml)3682061320
10
ba76138e0102/2013
Page 11
4210Contact Information
Contact Information
Ordering & Technical Support
Telephone
Fax
:(937) 767-1058
Email
Mail:YSI Incorporated
Internet
When placing an order please have the following information available:
YSI account number (if available)Name and Phone Number
Model number or brief descriptionBilling and shipping address
QuantityPurchase Order or Credit Card
:(800) 897-4151
(937) 767-7241
Monday through Friday, 8:00 AM to 5:00 PM ET
:environmental@ysi.com
1725 Brannum Lane
Yellow Springs, OH 45387
USA
:www.ysi.com
Service Information
YSI has authorized service centers throughout the United States and
Internationally. For the nearest service center information, please visit
www.ysi.com
directly at 800-897-4151.
When returning a product for service, include the Product Return form
with cleaning certification. The form must be completely filled out for an
YSI Service Center to accept the instrument for service. The Product
Return form may be downloaded at www.ysi.com
‘Support‘ tab.
and click ‘Support’ or contact YSI Technical Support
and clicking on the
ba76138d0102/2013
11
Page 12
Contact Information4210
12
ba76138d0102/2013
Page 13
Page 14
1) The tissue in plants that brings water upward from the roots;
2) a leading global water technology company.
We're 12,500 people unified in a common purpose: creating innovative solutions
to meet our world's water needs. Developing new technologies that will improve
the way water is used, conserved, and re-used in the future is central to our work.
We move, treat, analyze, and return water to the environment, and we help people
use water efficiently, in their homes, buildings, factories and farms. In more than
150 countries, we have strong, long-standing relationships with customers who
know us for our powerful combination of leading product brands and applications
expertise, backed by a legacy of innovation.
For more information on how Xylem can help you, go to www.xyleminc.com
Xylem |'m|zīlə
YSI
1725 Brannum Lane
Yellow Springs, OH 45387
Tel: +1 937-767-7241; 800-765-4974
Fax: +1 937-767-1058
Email: environmental@ysi.com
Web: www.ysi.com