INTRODUCTION
The Model 41342LC/LF Platinum Temperature Probe is an accurate 1000 ohm Platinum RTD temperature sensor and 4-20 mA line driver interface mounted in a weatherproof junction box. The probe is available in Celsius or Fahrenheit calibration. The probe is designed for easy installation in YOUNG Multi-plate and Aspirated Radiation Shields.
INSTALLATION
For accurate measurements, the temperature probe should be installed in a protective radiation shield. Use of the probe without a radiation shield may result in large errors. YOUNG naturally ventilated or motor aspirated shields are recommended. For best performance, the probe and shield should be placed in a location with good air circulation clear of large masses (buildings, pavement, solar panels...), exhaust vents, electrical machinery, motors, water fountains and sprinklers
MAINTENANCE
The temperature probe is designed to offer years of service with minimal maintenance. If necessary, the probe may be periodically checked or recalibrated using normal bath calibration methods. NIST traceable calibration is available from YOUNG at nominal cost.
WARRANTY
This product is warranted to be free of defects in materials and construction for a period of 12 months from date of initial purchase. Liability is limited to repair or replacement of defective item. A copy of the warranty policy may be obtained from R. M. Young Company.
SPECIFICATIONS
Power Requirement: 12 - 30 VDC, 20mA Calibrated measuring range: -50 to +50°C (suffix C)
-50 to +150°F (suffix F) Accuracy at 0°C: ±0.3°C Time Constant: 42 seconds in 43408 shield. Sensor type: 1000 Platinum RTD Output signal: 4-20 mA Recommended Cable: 2 conductor shielded,
22 AWG (#18641)
Recommended Radiation Shields: Model 43408P Gill Aspirated Radiation Shield Model 41002P Gill Multi-Plate Radiation Shield
CE COMPLIANCE
This product has been tested and shown to comply with European CE requirements for the EMC Directive. Please note that shielded cable must be used.
Declaration of Conformity
Application of Council Directives: 89/336/EEC Standards to which Conformity is Declared: EN 50082-1 (IEC 801-2, 3, 4)
Manufacturer's Name and Address: R. M. Young Company
Importer's Name and Address: See Shipper or Invoice Type of Equipment: Meteorological Instruments Model Number / Year of Manufacture: 41342 (V, L)/1996
I, the undersigned, hereby declare that the equipment specified conforms to the above Directives and Standards.
Date / Place: Traverse City, Michigan, USA February 19, 1996
David Poinsett R & D Manager, R. M. Young Company