Yokogawa ZR22, ZR402 User Manual

User's Manual
Model ZR22G, ZR402G Separate type Zirconia High Temperature Humidity Analyzer
IM 11M12A01-03E
IM 11M12A01-03E
7th Edition

Introduction

The EXAxt ZR Separate-type Zirconia High-temperature Humidity Analyzer has been developed for humidity control in various industrial processes. This analyzer basically consists of a detector and a converter. You can select between several versions based upon your application.
Optional accessories are also available to improve measurements and automate calibra­tion. An optimal control system can be realized by adding appropriate options.
This instruction manual refers to almost all of the equipment related to the EXAxt ZR. You may skip any section(s) on the equipment which is/are not included in your system.
Regarding the HART communication protocol,refer to the IM 11M12A01-51E. IM 11M12A01-51E has been published as Model EXAxt ZR Series HART Protocol The all-in-one version (with sensor and analyzer integrated in one body) is described in
IM 11M12A01-05E < Before using the equipment, please read any related descriptions in this manual for the
equipment and the system you have, on appropriate use and operation of the EXAxt ZR. > Models and descriptions in this manual are as follows:
Models and descriptions in this manual
Description in this manual
Model Product Name Specification Installation Operation Maintenance CMPL
ZR22G General-use detector s s s s ZR402G Converter s s s s s ZH21B Dust protector s s ZA8F Flow setting unit (for manual calibration use) s s s ZR40H Automatic Calibration unit s s s s
- Calibration gas unit case (Part No. E7044KF) s s
- Check valve (Part No. K9292DN, K9292DS) s s ZO21S Standard gas unit s s s s
T.Int.1E
CMPL : Customer Maintenance Parts List
IM 11M12A01-03E 7th Edition: Sep. 2006 (YK) All Rights Reserved, Copyright © 2000, Yokogawa Electric Corporation
IM 11M12A01-03E
This manual consists of twelve chapters. Please refer to the reference chapters for installation, operation and maintenance.
Table of Contents
References
Chapter Outline Installation Operation Maintenance
1. Overview Equipment models and system configuration examples
2. Specifications Standard specification, model code (or part number), dimension drawing for each equipment
3. Installation Installation method for each equipment
4. Piping Examples of piping in three standard system configurations
5. Wiring Wiring procedures such as Power supply wiring , output signal wiring or others
6. Components These are described in this manual
7. Startup Basic procedure to start operation of EXAxt ZR. Chapter 7 enables you to operate the equipment immediately.
8. Detailed Data Setting Details of key operations and displays
9. Calibration Describes the calibration procedure required in the course of operation.
10. Other Functions Other functions described
11. Inspection and How to conduct maintenance of EXAxt ZR and procedures
Maintenance for replacement of deteriorated parts
12. Troubleshooting This chapter describes measures to be taken when an abnormal condition occurs.
CMPL (parts list) User replaceable parts list
: Read and completely understand before operating the equipment.
s
: Read before operating the equipment, and refer to whenever necessary.
s
: Recommended to read it at least once.
n
s
s
s
s
s
n
n
s
s
s
s
s
s
s
n
n
s
s
n
n
n
s
n
n
n
n
s
s
s
T.Int.2E
ii
IM 1M12A01-03E

r For the safe use of this equipment

CAUTION
The cell (sensor) at the tip of the detector is made of ceramic (zirconia element). Do not drop the detector or subject it to pressure stress.
• Do NOT allow the sensor (probe tip) to make contact with anything when installing the detector.
• Avoid any water dropping directly on the probe (sensor) of the detector when install­ing it.
• Check the calibration gas piping before introducing the calibration gas to ensure that there is no leakage of the gas. If there is any leakage of the gas, the moisture drawn from the measuring gas can damage the sensor.
• The detector (especially at the tip) becomes very hot. Be sure to handle it with gloves.
EXAxt ZR is very heavy. Handle it with care. Be sure not to accidentally drop it. Handle safely to avoid injury.
DANGER
(1) About This Manual
Connect the power supply cord only after confirming that the supply voltage matches the rating of this equipment. In addition, confirm that the power is switched off when connecting power supply.
Some process gas is dangerous to people. When removing this equipment from the process line for maintenance or other reasons, protect yourself from potential poisoningby using a protective mask or ventilating the area well.
• This manual should be passed on to the end user.
• The contents of this manual are subject to change without prior notice.
• The contents of this manual shall not be reproduced or copied, in part or in whole, without permission.
• This manual explains the functions contained in this product, but does not warrant that those will suit the particular purpose of the user.
• Every effort has been made to ensure accuracy in the preparation of this manual. However, should any errors or omissions come to the attention of the user, please contact the nearest Yokogawa Electric representative or sales office.
• This manual does not cover the special specifications. This manual may not be changed on any change of specification, construction and parts when the change does not affect the functions or performance of the product.
• If the product is used in a manner not specified in this manual, safety of this product may be impaired.
IM 11M12A01-03E
iii
(2) Safety and Modification Precautions
• Follow the safety precautions in this manual when using the product to ensure protec­tion and safety of personnel, product and system containing the product.
(3) The following safety symbols are used on the product as well as in this manual.
DANGER
This symbol indicates that the operator must follow the instructions laid out in this manual in order to avoid the risk of personnel injuries or fatalities such as electric shock. The manual describes what special care the operator must exercise to prevent such risks.
WARNING
This symbol indicates that the operator must refer to the instructions in this manual in order to prevent the instrument (hardware) or software from being damaged, or a system failure from occurring.
CAUTION
This symbol draws attention to information essential for understanding the operation and functions.
Tip
This symbol gives information that complements the present topic.
SEE ALSO
This symbol identifies a source to which to refer.
Protective Ground Terminal
Function Ground Terminal (Do not use this terminal as the protective ground
terminal.)
Alternating current
iv
IM 1M12A01-03E
• Special descriptions in this manual
This manual indicates operation keys, displays and drawings on the product as follows:
• Operation keys, Enclosed in [ ], displays on the panel 0 0.
• Drawing for flashing
Indicated by gray characters. (Flashing) (lit)
(Ex. [MODE] key) (Ex. message display 0 BASE 0)
(Ex. data display 0 102 0 lit, 0 102 0 flashing)
IM 11M12A01-03E
v
r NOTICE
• Specification check When the instrument arrives, unpack the package with care and check that the instrument has not been damaged during transportation. In addition, please check that the specification matches the order, and required accessories are not missing. Specifi­cations can be checked by the model codes on the nameplate. Refer to Chapter 2 Specifications for the list of model codes.
• Details on operation parameters When the EXAxt ZR Separate-type High Temperature Humidity Analyzer arrives at the user site, it will operate based set before shipping parameters (initial data) on the departure from the factory. Ensure that the initial data is suitable for the operation conditions before conducting analysis. Where necessary, set the instrument parameters for appropriate operation. For details of setting data, refer to chapters 7 to 10. When user changes the operation parameter, it is recommended to note down the changed setting data.
r After -Sales Warranty
d Do not modify the product.
d During the warranty period, for repair under warranty carry or send the product to the
local sales representative or service office. Yokogawa will replace or repair any damaged parts and return the product to you.
d Before returning a product for repair under warranty, provide us with the model
name and serial number and a description of the problem. Any diagrams or data explaining the problem would also be appreciated.
d If we replace the product with a new one, we won’t provide you with a repair report. d Yokogawa warrants the product for the period stated in the pre-purchase quotation.
Yokogawa shall conduct defined warranty service based on its standard. When the customer site is located outside of the service area, a fee for dispatching the mainte­nance engineer will be charged to the customer.
d In the following cases, customer will be charged repair fee regardless of warranty
period.
• Failure of components which are out of scope of warranty stated in instruction
manual.
• Failure caused by usage of software, hardware or auxiliary equipment, which
Yokogawa Electric did not supply.
• Failure due to improper or insufficient maintenance by user.
• Failure due to modification, misuse or outside-of-specifications operation which
Yokogawa does not authorize.
• Failure due to power supply (voltage, frequency) being outside specifications or
abnormal.
• Failure caused by any usage out of scope of recommended usage.
• Any damage from fire, earthquake, storms and floods, lightning, disturbances, riots,
warfare, radiation and other natural changes.
vi
IM 1M12A01-03E
d Yokogawa does not warrant conformance with the specific application at the user
site. Yokogawa will not bear direct/indirect responsibility for damage due to a specific application.
d Yokogawa Electric will not bear responsibility when the user configures the product
into systems or resells the product.
d Maintenance service and supplying repair parts will be covered for five years after
the production ends. For repair for this product, please contact the nearest sales office described in this instruction manual.
IM 11M12A01-03E
vii
Contents
Introduction........................................................................................................................... i
r For the safe use of this equipment ............................................................................... iii
r NOTICE.........................................................................................................................vi
r After-Sales Warranty ................................................................................................... vi
1. Overview ...................................................................................................................... 1-1
1.1 < EXAxt ZR > System Configuration............................................................ 1-2
1.1.1 System 1 .................................................................................................. 1-2
1.1.2 System 2 .................................................................................................. 1-3
1.1.3 System 3 .................................................................................................. 1-4
1.2 < EXAxtZR > System Components ............................................................... 1-5
1.2.1 System Components ................................................................................ 1-5
1.2.2 Detectors and Accessories ....................................................................... 1-5
2. Specifications ............................................................................................................... 2-1
2.1 General Specifications .................................................................................... 2-1
2.1.1 Standard Specifications ........................................................................... 2-1
2.2 General-use Separate-type Detector and Related Equipment ........................ 2-3
2.2.1 ZR22G General-use Separate-type Detector ........................................... 2-3
2.2.2 ZH21B Dust Protector ............................................................................. 2-8
2.3 ZR402G Separate-type Converter ................................................................... 2-9
2.3.1 Standard Specifications ........................................................................... 2-9
2.3.2 Function ................................................................................................. 2-10
2.4 ZA8F Flow Setting Unit and ZR40H Automatic Calibration Unit ............. 2-16
2.4.1 ZA8F Flow Setting Unit........................................................................ 2-16
2.4.2 ZR40H Automatic Calibration Unit ...................................................... 2-18
2.5 ZO21S Standard Gas Unit ............................................................................ 2-21
2.6 Other Equipment ........................................................................................... 2-22
2.6.1 Stop Valve (Part Number: L9852CB or G7016XH) ............................ 2-22
2.6.2 Check Valve (Part Number: K9292DN or K9292DS) ......................... 2-23
2.6.3 Air Set................................................................................................... 2-24
2.6.4 Zero-gas Cylinder (Part Number: G7001ZC) ....................................... 2-26
2.6.5 Pressure Regulator for Gas Cylinder
(Part Number: G7013XF or G7014XF) .............................................. 2-26
2.6.6 Case Assembly for Calibration-gas Cylinder
(Part Number: E7044KF)..................................................................... 2-27
2.6.7 Model ZR22A Heater Assembly ........................................................... 2-28
3. Installation ................................................................................................................... 3-1
3.1 Installation of the Detector ............................................................................. 3-1
3.1.1 Location ................................................................................................... 3-1
3.1.2 Probe Insertion Hole ................................................................................ 3-2
3.1.3 Installation of the Detector ...................................................................... 3-3
3.1.4 Installation of ZH21B Dust Protector ..................................................... 3-3
viii
IM 1M12A01-03E
3.2 Installation of the Converter ........................................................................... 3-4
3.2.1 Location .................................................................................................. 3-4
3.2.2 Mounting of the Converter ...................................................................... 3-4
3.3 Installation of ZA8F Flow Setting Unit ......................................................... 3-7
3.3.1 Location ................................................................................................... 3-7
3.3.2 Mounting of ZA8F Flow Setting Unit .................................................... 3-7
3.4 Installation of ZR40H Automatic Calibration Unit........................................ 3-9
3.4.1 Location ................................................................................................... 3-9
3.4.2 Mounting of ZR40H Automatic Calibration Unit .................................. 3-9
3.5 Installation of E7044KF Case Assembly for
the Calibration-gas Cylinder ......................................................................... 3-11
3.5.1 Location ................................................................................................. 3-11
3.5.2 Mounting................................................................................................ 3-11
3.6 Insulation Resistance Test ............................................................................. 3-12
3.7 External Dimensions of Detectors with Pressure Compensation................. 3-13
4. Piping............................................................................................................................4-1
4.1 Piping for System 1 ........................................................................................ 4-1
4.1.1 Parts Required for Piping in System 1 ................................................... 4-2
4.1.2 Connection to the Calibration Gas Inlet ................................................. 4-2
4.1.3 Connection to the Reference Gas Inlet ................................................... 4-2
4.2 Piping for System 2 ........................................................................................ 4-3
4.2.1 Piping Parts for System 2........................................................................ 4-3
4.2.2 Piping for the Calibration Gas ................................................................ 4-4
4.2.3 Piping for the Reference Gas .................................................................. 4-4
4.3 Piping for System 3 ........................................................................................ 4-5
5. Wiring ........................................................................................................................... 5-1
5.1 General ............................................................................................................ 5-1
5.1.1 Terminals for the External Wiring in the Converter .............................. 5-3
5.1.2 Wiring ...................................................................................................... 5-3
5.1.3 Mounting of Cable Gland........................................................................ 5-4
5.2 Wiring for Detector Output ............................................................................. 5-5
5.2.1 Cable Specifications ................................................................................ 5-5
5.2.2 Connection to the Detector...................................................................... 5-6
5.2.3 Connection to the Converter ................................................................... 5-7
5.3 Wiring for Power to Detector Heater ............................................................. 5-8
5.3.1 Cable Specifications ................................................................................ 5-8
5.3.2 Connection to Detector............................................................................ 5-8
5.3.3 Connection to Converter ......................................................................... 5-9
5.4 Wiring for Analog Output ............................................................................. 5-10
5.4.1 Cable Specifications .............................................................................. 5-10
5.4.2 Wiring Procedure................................................................................... 5-10
5.5 Power and Grounding Wiring....................................................................... 5-11
5.5.1 Power Wiring......................................................................................... 5-11
5.5.2 Grounding Wiring ................................................................................. 5-11
5.6 Wiring for Contact Output............................................................................ 5-12
5.6.1 Cable Specifications .............................................................................. 5-12
5.6.2 Wiring Procedure................................................................................... 5-12
5.7 Wiring for ZR40H Automatic Calibration Unit ........................................... 5-13
5.7.1 Cable Specifications .................................................................................. 5-14
5.7.2 Wiring Procedure ...................................................................................... 5-14
IM 11M12A01-03E
ix
5.8 Wiring for Contact Input ............................................................................... 5-15
5.8.1 Cable Specifications .............................................................................. 5-15
5.8.2 Wiring Procedure................................................................................... 5-15
5.9 Temperature Input Wiring ............................................................................ 5-16
5.9.1 Applicable Temperature Transmitter .................................................... 5-16
5.9.2 Cable Specifications .............................................................................. 5-16
5.9.3 Wiring Procedure................................................................................... 5-16
6. Components ................................................................................................................. 6-1
6.1 ZR22G Detector .............................................................................................. 6-1
6.1.1 General-purpose Detector ........................................................................ 6-1
6.2 ZR402G Converter.......................................................................................... 6-2
6.3 ZA8F Flow Setting Unit, ZR40H Automatic Calibration Unit ..................... 6-3
7. Startup..........................................................................................................................7-1
7.1 Checking Piping and Wiring Connections ..................................................... 7-1
7.2 Checking Valve Setup ..................................................................................... 7-1
7.3 Supplying Power to the Converter .................................................................. 7-2
7.4 Touchpanel Switch Operations ....................................................................... 7-3
7.4.1 Basic Panel and Switch ........................................................................... 7-3
7.4.2 Display Configuration (for High-temperature Humidity Analyzer) ....... 7-4
7.4.3 Display Functions .................................................................................... 7-5
7.4.4 Entering Numeric and Text Data ............................................................ 7-5
7.5 Confirmation of Converter Type Setting........................................................ 7-7
7.6 Confirmation of Detector Type Setting .......................................................... 7-8
7.7 Current Output Setting.................................................................................... 7-8
7.7.1 Analog Output Setting ............................................................................. 7-8
7.7.2 Minimum Current (4 mA) and Maximum Current (20 mA) Settings ... 7-9
7.8 Setting Display Item ..................................................................................... 7-10
7.9 Checking Current Loop ................................................................................. 7-12
7.10 Checking Contact I/O ................................................................................... 7-13
7.10.1 Checking Contact Output ...................................................................... 7-13
7.10.2 Checking Calibration Contact Output................................................... 7-14
7.10.3 Checking Input Contacts ....................................................................... 7-14
7.11 Calibration ..................................................................................................... 7-15
7.11.1 Calibration Setup................................................................................... 7-15
7.11.2 Manual Calibration................................................................................ 7-16
8. Detailed Data Setting.................................................................................................. 8-1
8.1 Current Output Setting.................................................................................... 8-1
8.1.1 About Input Ranges ................................................................................. 8-1
8.1.2 Setting Minimum Current (4 mA) and Maximum Current (20 mA) ..... 8-4
8.1.3 Entering Output Damping Constants ...................................................... 8-4
8.1.4 Selection of Output Mode ....................................................................... 8-4
8.1.5 Default Values ......................................................................................... 8-5
8.2 Output Hold Setting ........................................................................................ 8-6
8.2.1 Definition of Equipment Status ............................................................... 8-6
8.2.2 Preference Order of Output Hold Value ................................................. 8-7
8.2.3 Output Hold Setting................................................................................. 8-8
8.2.4 Default Values ......................................................................................... 8-8
x
IM 1M12A01-03E
8.3 Alarm Setting .................................................................................................. 8-9
8.3.1 Alarm Values ........................................................................................... 8-9
8.3.2 Alarm Output Actions ............................................................................. 8-9
8.3.3 Alarm Setting Procedure ....................................................................... 8-10
8.3.4 Default Values ....................................................................................... 8-11
8.4 Output Contact Setup .................................................................................... 8-12
8.4.1 Output Contact....................................................................................... 8-12
8.4.2 Setting Procedure................................................................................... 8-12
8.4.3 Default Values ....................................................................................... 8-15
8.5 Input Contact Settings................................................................................... 8-16
8.5.1 Input Contact Functions ........................................................................ 8-16
8.5.2 Setting Procedure................................................................................... 8-17
8.5.3 Default Values ....................................................................................... 8-17
8.6 Other Settings................................................................................................ 8-18
8.6.1 Setting the Date-and-Time .................................................................... 8-18
8.6.2 Setting Periods over which Average Values Are Calculated and Periods over
which Maximum and Minimum Values Are Monitored ...................... 8-19
8.6.3 Setting Measurement Gas Temperature and Pressure .......................... 8-20
8.6.4 Setting Purging ...................................................................................... 8-22
8.6.5 Setting Passwords .................................................................................. 8-23
9. Calibration ...................................................................................................................9-1
9.1 Calibration Briefs ............................................................................................ 9-1
9.1.1 Measurement Principle of Zirconia Humidity Analyzer ........................ 9-1
9.1.2 Calibration Gas ........................................................................................ 9-3
9.1.3 Compensation .......................................................................................... 9-4
9.1.4 Characteristic Data from a Sensor Measured During Calibration .......... 9-5
9.2 Calibration Procedures .................................................................................... 9-6
9.2.1 Calibration Setting ................................................................................... 9-6
9.2.2 Default Values ......................................................................................... 9-9
9.2.3 Calibration ............................................................................................. 9-10
10.Other Functions ........................................................................................................ 10-1
10.1 Display........................................................................................................... 10-1
10.1.1 Detailed Display .................................................................................... 10-1
10.1.2 Trend Graph ........................................................................................... 10-5
10.1.3 Auto(matic) Display-Revert Time......................................................... 10-7
10.1.4 Entering Tag Name ............................................................................... 10-8
10.1.5 Language Selection ............................................................................... 10-8
10.2 Blowback....................................................................................................... 10-9
10.2.1 Blowback Setup..................................................................................... 10-9
10.3 Operational Data Initialization.................................................................... 10-13
10.4 Reset ............................................................................................................ 10-17
10.5 Handling of the ZO21S Standard Gas Unit................................................ 10-18
10.5.1 Standard Gas Unit Component Identification ..................................... 10-18
10.5.2 Installing Gas Cylinders ...................................................................... 10-19
10.5.3 Calibration Gas Flow........................................................................... 10-20
10.6 Methods of Operating Valves in the ZA8F Flow Setting Unit .................. 10-23
10.6.1 Preparation Before Calibration ............................................................ 10-23
10.6.2 Operating the Span Gas Flow Setting Valve...................................... 10-23
10.6.3 Operating the Zero Gas Flow Setting Valve ...................................... 10-24
10.6.4 Operation After Calibration................................................................. 10-24
IM 11M12A01-03E
xi
11.Inspection and Maintenance .................................................................................... 11-1
11.1 Inspection and Maintenance of the Detector ................................................ 11-2
11.1.1 Cleaning the Calibration Gas Tube....................................................... 11-2
11.1.2 Replacing the Sensor Assembly ............................................................ 11-3
11.1.3 Replacement of the Heater Unit .......................................................... 11-5
11.1.4 Replacement of O-ring .......................................................................... 11-8
11.1.5 Stopping and Re-starting Operation ........................................................ 11-8
11.2 Inspection and Maintenance of the Converter.............................................. 11-9
11.2.1 Replacing Fuses..................................................................................... 11-9
11.2.2 Cleaning............................................................................................... 11-10
11.2.3 Adjust LCD screen contrast ................................................................ 11-10
11.3 Replacing Flowmeter in ZR40H Autocalibration Unit .............................. 11-11
12.Troubleshooting.........................................................................................................12-1
12.1 Displays and Measures to Take When Errors Occur ................................... 12-1
12.1.1 What is an Error? .................................................................................. 12-1
12.1.2 Measures to Take When an Error Occurs ............................................. 12-2
12.2 Displays and Measures to Take When Alarms are Generated ..................... 12-5
12.2.1 What is an Alarm? ................................................................................. 12-5
12.2.2 Measures Taken When Alarms are Generated.................................. 12-7
12.3 Countermeasures When Measured Value Shows Error ............................. 12-13
12.3.1 Measured Value Higher Than True Value.......................................... 12-13
12.3.2 Measured Value Lower Than True Value .......................................... 12-14
12.3.3 Measurements Sometimes Show Abnormal Values........................... 12-14
Customer Maintenance Parts List ................................................ CMPL 11M12A01-03E
Customer Maintenance Parts List ................................................ CMPL 11M12C01-01E
Customer Maintenance Parts List ................................................ CMPL 11M12A01-11E
Customer Maintenance Parts List .................................................... CMPL 11M3D1-01E
Revision Record .................................................................................................................... i
xii
IM 1M12A01-03E
1. Overview
The EXAxt ZR Separate-type Zirconia High-temperature Humidity Analyzer is used to measure the humidity of hot gases continuously in driers which use hot gas or electricity as the heat source. It can also be used in a variety of manufacturing applications in humidifiers, as well as in driers, for humidity measurement and control. It can help improve productivity in these application fields.
The ZR402G Separate-type converter is equipped with an LCD touchpanel which has various setting displays, a calibration display, humidity trend display, with easier operation and improvement of display functions. The converter is equipped with various standard functions such as measurement and calculation as well as maintenance func­tions including a self-test. Analyzer calibration can also be fully automated and ZR40H, the automatic calibration unit, is provided. Choose the detector which best suits your needs so that an optimal humidity control system can be obtained.
Some examples of typical system configuration are illustrated following pages :

1. Overview

IM 11M12A01-03E
1-1

1.1 < EXAxt ZR > System Configuration

The system configuration should be determined by the conditions; e.g. whether the calibration gas flow should be automated. The system configuration can be classified into three basic patterns as follows:

1.1.1 System 1

This is the simplest system consisting of a detector and a converter. This system can be implemented for monitoring humidity in a production process such as food production. No piping is required for the reference gas (air) which is fed in at the installation site. The handy ZO21S standard gas unit is used for calibration.
Zero gas from this unit and span gas (air) is sent to the detector through a tube which is connected during calibration.
CAUTION
• A needle (stop) valve should be connected to the calibration gas inlet of the detector. The valve should be fully closed unless calibration is in progress.
• As this system uses ambient air for the reference gas, measuring accuracy will be affected by the installation location.
Separate type Zirconia High Temperature Humidity Analyzer, Model ZR22G Detector
Stop valve
Signal (6-core shield cable)
Heater (2-core)
Calibration gas
Figure 1.1
Model ZR402G Converter
EXA
ZR402G
100 to 240 V AC
,
Contact input Analog outout, contact output
Digital output (HART)
Model ZO21S Standard gas unit
,
F1.1E.EPS
1-2
IM 11M12A01-03E

1.1.2 System 2

1. Overview
This system is for accurate monitoring and controlling humidity when the installation environment is polluted with gases other than the air. Instrument air (clean and dry air of oxygen concentration 21%) is used for the reference gas and the span gas for calibra­tion. Zero gas is fed in from a cylinder during calibration. The gas flow is controlled by the ZA8F flow setting unit (for manual valve operation).
Separate type Zirconia High Temperature Humidity Analyzer, Model ZR22G Detector
Check valve or Stop valve
Calibration gas
Signal (6-core shield cable)
Heater (2-core)
Reference gas
Model ZA8F flow setting unit
Figure1.2
Model ZR402G Converter
Flowmeter
Needle
EXA
ZR402G
valve
100 to 240 V AC
,
Contact input Analog outout, contact output
Digital output (HART)
Air Set
Instrument air Span gas (Same as Zero gas calibration unit)
Calibration gas pressure regulator
Calibration gas unit case
Zero gas cylinder
F1.2E.EPS
IM 11M12A01-03E
1-3

1.1.3 System 3

This system is for accurate monitoring and controlling of humidity. Instrument air (clean and dry air) is used for the reference gas and the span gas for calibration. Zero gas is fed in from a cylinder during calibration. The calibration gas flow is controlled automatically by the ZR40H automatic calibration unit.
This system is similar to system 2, except that the calibration gas flow is automated using the ZR40H automatic calibration unit.
Separate type Zirconia High Temperature Humidity Analyzer, Model ZR22G Detector
Check valve
*2
Reference gas
Calibration gas
Signal (6-core shield cable)
Heater (2-core shield cable)
*1
Model ZR40H Auto-Calibration unit
Flowmeter
Model ZR402G Converter
EXA
ZR402G
,
Needle
vaive
Air Set
*3
100 to 240 V AC
Contact input Analog output, contact output Digital output (HART)
lnstrument air
Calibration gas pressure regulator
Calibration gas unit case
Zero gas cylinder
F1.3E.EPS
Figure1.3
*1 Shield cable : Use shielded signal cables, and connect the shield to the FG terminal of the converter. *2 Select the desired probe from the Probe Configuration table on page 1-5
1-4
*3 100% N2 gas cannot be used as the zero gas. Use approximately 1% of O2 gas (N2­ based).
IM 11M12A01-03E

1.2 < EXAxtZR > System Components

1.2.1 System Components

Model or Part No. Product Name System 1 System 2 System 3
ZR22G Detector, Separate-type High-temperature
Humidity Analyzer, Detector
ZR402G Converter, Separate-type High-temperature
Humidity Analyzer, Converter ZH21B Dust protector ZO21S Standard gas unit ZA8F Flow setting unit (for manual calibration) ZR40H Automatic calibration unit (for separate type) L9852CB/G7016XH Stop valve K9292DN/K9292DS Check valve K9473XH/K9473XJ, Air set G7004XF/K9473XG G7001ZC Zero gas cylinder G7013XF/G7014XF Pressure regulator for gas cylinder E7044KF Case assembly for calibration-gas cylinder
ZR22A Heater Assembly
(Spare parts for Model ZR22G)
d : The essential products for the system s : Selected depending on your applications
(d) : Select either
1. Overview
Separate type
d d d
d d d
s s s d
d
d
d (d)
(d) d
d d
d d d d d d
s s s
T1.1E.EPS

1.2.2 Detectors and Accessories

dGeneral-use Detector (sample gas temperature: 0 to 7008C)
Standard detector Model
ZR22G
Mounting angle
The probe insertion length: 0.4 to 2m
Vertical to horizontal
The probe insertion length: 2.5 to 3m
Vertical
Detector with dust protector
Model
ZR22G-040
and
ZH21B
Mounting angle
Vertical to horizontal
(The probe insertion length: 0.4m)
F1.4E.EPS
IM 11M12A01-03E
1-5
2. Specifications
This chapter describes the specifications for the following: ZR22G General-use separate-type detector (See Section 2.2.1) ZH21B Dust protector (See Section 2.2.2) ZR402G Separate-type converter (See Section 2.3) ZA8F Flow setting unit (See Section 2.4.1) ZR40H Automatic calibration unit(See Section 2.4.2) ZO21S Standard gas unit(See Section 2.5)

2.1 General Specifications

2.1.1 Standard Specifications

High Temperature Humidity Analyzer
Oxygen concentration in mixed gas which consists of water vapor and air is proportional to the volume rate of air, so the volume rate of water vapor can be calculated by the oxygen concentration.

2. Specifications

Measured Object:Water vapor (in vol%) in mixed gases (air and water vapor) Measured System: Zirconia system Measured Range: 0.01 to 100 vol% O2, 0 to 100 vol% H2O or 0 to 1.000 kg/kg Output Signal: 4 to 20 mA DC (maximum load resistance 550V) Oxygen concentration;
Any setting in the range of 0 to 5 through 0 to 100 vol%O2 (in 1 vol% O2), or partial range.
Moisture quantity: 0 to 25 through 0 to 100 vol% H2O (in 1 vol% H2O), or partial
range. Mixture ratio 0 to 0.2 through 0 to 1.000 kg/kg (in 0.001 kg/kg), or
partial range.
Digital Communication (HART): 250 to 550V, depending on quantity of field devices
connected to the loop (multi-drop mode). (Note) HART is a registered trademark of the HART Communication
Foundation.
Display Range: Oxygen concentration; 0 to100 vol% O2 , Moisture quantity; 0 to 100
vol% H2O Mixture ratio; 0 to 1 kg/kg Relative humidity; 0 to 100 %RH Dew point; -40 to 3708 C (Note) Those values are calculated by temperature and absolute
pressure. Then accurate temperature and pressure value must be input to the converter.
Warm-up Time: Approx. 20 min.
(Note) These characteristics are calculated by oxygen concentration measured in air which include water vapor.
IM 11M12A01-03E
2-1
Repeatability: (Note1)
61 vol% H2O (sample gas pressure 2 kPa or less)
Linearity: (Excluding standard gas tolerance)
(Note1) (Use oxygen of known concentration (in the measuring range) as the
zero and span calibration gas.)
62 vol% H2O; (Sample gas pressure: within 60.49 kPa) 63 vol% H2O; (Sample gas pressure: 2 kPa or less)
Drift: (Excluding the first two weeks in use)
(Note1) both zero and span 63 vol% H2O/month
Response Time: Response of 90 % within 5 seconds. (Measured after gas is introduced
from calibration-gas inlet and analog output starts changing.) (Note1) These tolerances do not apply to the pressure compensated
version, or where natural convection is used for the reference air.
2-2
IM 11M12A01-03E

2.2 General-use Separate-type Detector and Related Equipment

The “Detector with dust protector” consists of ZR22G general-use separate-type detector and ZH21B dust protector (refer to Section 2.2.2).

2.2.1 ZR22G General-use Separate-type Detector

Sample Gas Temperature: 0 to 7008 C (Probe only)
It is necessary to mount the cell using Inconel cell-bolts when the temperature is 6008 C or greater.
Sample Gas Pressure: -5 to +20 kPa (When the pressure in the process exceeds 3kPa,
it is recommended that you compensate the pressure. When the pressure in the process exceeds 5kPa, you must perform pressure compensation.)
No pressure fluctuation in the process should be allowed. Probe Length: 0.4, 0.7, 1.0, 1.5, 2.0, 2.5, 3.0 m Probe Material: SUS 316 (JIS) Ambient Temperature: -20 to +1508 C Reference Air System: Natural Convection, Instrument Air Instrument Air System (excluding Natural Convection):
Pressure; 200 kPa + the pressure inside the dryer, (It is recommended to
use air which has been dehumidified by cooling to dewpoint -208 C or
less, and with dust or oil mist removed.)
Consumption; Approx. 1Nl/min
Note: When the detector is used in conjunction with a check valve and a
ZA8F Flow Setting Unit, the maximum pressure of sample gas is 150
kPa. When with a check valve and a ZR40H Auto Calibration Unit, it is
200 kPa. If the pressure of your sample gas exceeds these limits, consult
with Yokogawa. Material in Contact with Gas: SUS 316 (JIS), Zirconia, SUS 304 (JIS) (flange),
Hastelloy B, (Inconel 600, 601) Construction: Heater and thermocouple replaceable construction. Non explosion-proof
JIS C0920 / equivalent to IP44D. Equivalent to NEMA 4X / IP66 (Achieved when the cable entry is completely sealed with a cable gland in the recirculation pressure compensated version.)
2. Specifications
IM 11M12A01-03E
Terminal Box Case: Material; Aluminium alloy Terminal Box Paint Color:
Case; Mint green (Munsell 5.6BG3.3/2.9) Cover; Mint green (Munsell 5.6BG3.3/2.9)
Finish: Polyurethane corrosion-resistant coating Gas Connection: Rc 1/4 or 1/4 FNPT
Wiring Connection: G1/2, Pg13.5, M20 by 1.5mm, 1/2 NPT Installation: Flange mounting
2-3
Probe Mounting Angle: Horizontal to vertically downward.
When the probe insertion length is 2 m or less, installing at angles from
horizontal to vertically downward is available.
When the probe insertion length exceeds 2.5 m, mount vertically down-
ward (within 658 ) and use a probe protector. Weight:
Insertion length of 0.4 m: approx. 6 kg (JIS 5K 65) / approx. 11 kg (ANSI 150 4) Insertion length of 1.0 m: approx. 8 kg (JIS 5K 65) / approx. 13 kg (ANSI 150 4) Insertion length of 1.5 m: approx. 10 kg (JIS 5K 65) / approx. 15 kg (ANSI 150 4) Insertion length of 2.0 m: approx. 12 kg (JIS 5K 65) / approx. 17 kg (ANSI 150 4) Insertion length of 3.0 m: approx. 15 kg (JIS 5K 65) / approx. 20 kg (ANSI 150 4)
2-4
IM 11M12A01-03E
2. Specifications
Model and Codes
Model ZR22G
Suffix code
Option
code
Description
Separate type Detector of Zirconia High Temperature Humidity
[ Style : S2 ]
Analyzer
Length
-040
-070
-100
-150
-200
-250
-300
Wetted material
-S
-C
Flange (*2)
-A
-B
-C
-E
-F
-G
-K
-L
-M
-P
-R
-S
-W
Reference air
-C
-E
-P
Gas Thread
-T
Connection box thred
-P
-G
-M
-T
-Q
Instruction manual
-J
-E
-A
Options
/D /C /CV /SV
/PT
-R
/SCT
0.4 m
0.7 m
1.0 m
1.5 m
2.0 m
2.5 m (*1)
3.0 m (*1) SUS316 Stainless steel with lnconel calibration gas tube ANSI Class 150 2 RF SUS304 ANSI Class 150 3 RF SUS304 ANSI Class 150 4 RF SUS304 DIN PN10 DN50 SUS304 DIN PN10 DN80 SUS304 DIN PN10 DN100 SUS304 JIS 5K 65 FF SUS304 JIS 10K 65 FF SUS304 JIS 10K 80 FF SUS304 JIS 10K 100 FF SUS304 JPI Class 150 4 RF SUS304 JPI Class 150 3 RF SUS304 Westinghouse
Natural convection External connection (Instrument air) (*8) Pressure Compensation (*8) Rc 1/4 1/4 NPT(F) G1/2 Pg13.5 M20 x1.5 mm 1/2NPT Quick connect (*6) Japanese English Always-A DERAKANE coating (*7) Inconel bolt (*3) Check valve (*4) Stop valve (*4) Stainless steel tag plate (*5) Printed tag plate (*5)
*1 When installing horizontally the probe whose insertion length is 2.5 meters or more, use the Probe Protector. Be sure to specify ZO21R-L-hhh-h. Specify the flange suffix code either -C or -K. *2 The thickness of the flange depends on its dimensions. *3 Inconel probe bolts and U shape pipe for calibration are used. Use this option for high temperature use (ranging from 600 to 700 8C). *4 Specify either /CV or /SV option code. *5 Specify either /SCT or /PT option code. *6 Not waterproof, protect from rain.Operating maximum temperature is 80 8C. Available only in the U.S. *7 Available only in the U.S. DERAKANE is a registered trademark of the Dow Chemical Company. *8 Piping for reference air must be installed to supply reference air constantly at a specified flow rate.
T2.1E.EPS
IM 11M12A01-03E
2-5
EXTERNAL DIMENSIONS Model ZR22G Detectors Separate type Zirconia High Temperature Humidity Analyzer
[50.8
L=0.15, 0.4, 0.7, 1.0,
1.5, 2.0, 2.5, 3.0 (m)
L
t
Flange
283 to 292
155 to 163 69
25
85
Rc1/4 or 1/4NPT Reference air inlet
2-G1/2,2-1/2NPT etc. Cable connection port
48
Rc1/4 or 1/4NPT Calibration gas inlet
[124
C
[A
[B
Flange
Flange
ANSI Class 150 2 RF SUS304 ANSI Class 150 3 RF SUS304 ANSI Class 150 4 RF SUS304 DIN PN10 DN50 SUS304 DIN PN10 DN80 SUS304 DIN PN10 DN100 SUS304 JIS 5K 65 FF SUS304 JIS 10K 65 FF SUS304 JIS 10K 80 FF SUS304 JIS 10K 100 FF SUS304 JIS 5K 32 FF SUS304 JPI Class 150 4 RF SUS304 JPI Class 150 3 RF SUS304 Westinghouse
AB C
152.4
190.5
228.6 165 200 220 155 175 185 210 115 229 190 155
120.6
152.4
190.5 125 160 180 130 140 150 175
90
190.5
152.4 127
4 - [19 4 - [19 8 - [19 4 - [18 8 - [18 8 - [18 4 - [15 4 - [19 8 - [19 8 - [19 4 - [15 8 - [19 4 - [19
4 - [11.5
19 24 24 18 20 20 14 18 18 18
24 24 14
t
[A
[B
Flange
5
C
F07_01.EPS
2-6
IM 11M12A01-03E
2. Specifications
Model ZR22G...-P Detectors(with pressure compensation), Separate type Zirconia High Temperature Humidity Analyzer, Detectors
L
f50.8
L=0.15, 0.4, 0.7, 1.0,
1.5, 2.0, 2.5, 3.0 (m)
L
Reference gas outlet
Flange
ANSI Class 150 2 RF SUS304 ANSI Class 150 3 RF SUS304 ANSI Class 150 4 RF SUS304 DIN PN10 DN50 SUS304 DIN PN10 DN80 SUS304 DIN PN10 DN100 SUS304 JIS 5K 65 FF SUS304 JIS 10K 65 FF SUS304 JIS 10K 80 FF SUS304 JIS 10K 100 FF SUS304 JPI Class 150 4 RF SUS304 JPI Class 150 3 RF SUS304 Westinghouse
Flange
t
303
156
t
25
PIPING
:B
PIPING : A
AB C
152.4
190.5
228.6 165 200 220 155 175 185 210 229 190 155
120.6
152.4
190.5
190.5
152.4
125 160 180 130 140 150 175
127
4 - [19 4 - [19 8 - [19 4 - [18 8 - [18 8 - [18 4 - [15 4 - [19 8 - [19 8 - [19 8 - [19 4 - [19
4 - [11.5
85
Rc1/4 or 1/4NPT Reference air inlet
87
2-G1/2, 2-1/2NPT etc. Cable connection port
48
Rc1/4 or 1/4NPT Calibration gas inlet
Stop Valve
t
PIPING
19 24 24 18 20 20 14 18 18 18 24 24 14
A B B A B B A A B B B B A
f124
C
[A
[B
Flange
C
[A
[B
Flange
F07_02.EPS
IM 11M12A01-03E
2-7

2.2.2 ZH21B Dust Protector

This protector is designed to protect the probe output from dust agitation (i.e., to prevent combustible materials from entering the probe cell where humidity measurements are made) in a dusty environment.
d Standard Specification
Insertion length: 0.428m Flange: JIS 5K 80 FF SUS304 or ANSI Class 150 4 FF SUS304 (However,
Material: SUS 316(JIS), SUS 304(JIS) (flange) Weight: Approximately 6kg (JIS), approximately 8.5kg (ANSI) Mounting: Mounted on the probe or process flange with bolts and associated
flange thickness is different)
nuts and washers.
Model Suffix code
ZH21B
Insertion length
Flange ( *1)
Style code
* The flange thickness varies.
Specify the probe ZR22G-040-h-K in code of (1)
-040 0.428 m
-A
Option
code
Dust Protector (0 to 6008C)
-J
*B
ZR22G-040-h-C in code of (2)
JIS 5K 80 FF SUS304 (1) ANSI Class 150 4B FF SUS304 (2)
Style B
Dust protector
72
[
Install facing upwards
C
[B
Description
428
T2.2E.EPS
76.3
[
D
[B
Unit ; mm
t
[
A
C
2-8
Insertion hole;minimum[80
ANSI flange
Size of each part of flange
Flange
JIS 5K 80 FF SUS304
ANSI Class 150 4B FF SUS304
A
180
228.6
B
145
190.5
four holes of
diameter 19
eight holes of diameter 19
Ct
12
12
2.3E-1.EPS
JIS flange
D
40
50
Insertion hole;minimum[80
F2.3E.EPS
IM 11M12A01-03E

2.3 ZR402G Separate-type Converter

2.3.1 Standard Specifications

Operated using an LCD touchscreen on the converter. Display: LCD display of size 320 by 240 dot with touchscreen. Output Signal: 4 to 20 mA DC, two points (maximum load resistance 550 V) Contact Output Signal: four points (one is fail-safe, normally open) Contact Input: two points Analog Input: one point (thermal input 4-20 mA) Auto-calibration Output: Two points (for dedicated auto-calibration unit) Ambient Temperature: -20 to +558C Storage Temperature: -30 to +708C Humidity Ambient: 0 to 95%RH (Non-condensing) Installation Altitude: 2000 m or less Category based on IEC 1010: II (Note)
2. Specifications
Pollution degree based on IEC 1010: 2 (Note)
Note: Installation category, called over-voltage category, specifies impulse withstand­ing voltage. Category II is for electrical equipment. Pollution degree indicates the degree of existence of solid, liquid, gas or other inclusions which may reduce breakdown voltage. Degree 2 is the normal indoor environment.
Power Supply Voltage: Ratings; 100 to 240 V AC Acceptable range; 85 to 264 V AC Power Supply Frequency: Ratings; 50/60 Hz Acceptable range; 45 to 66 Hz Power Consumption: Max. 300 W, approx. 100 W for ordinary use. Safety and EMC conforming standards Safety: EN61010-1
CSA C22.2 No.61010-1 UL61010-1
EMC: EN 61326 Class A
EN 55011 Class A Group 1 EN 61000-3-2 AS/NZS CISPR 11
Maximum Distance between Probe and Converter: Conductor two-way resistance must be 10 V or less (when a 1.25 mm2 cable or equivalent is used, 300 m or less.)
Construction: Outdoor installation, equivalent to NEMA 4 (with conduit holes completely sealed with a plastic cable gland optional)
Wiring Connection: G1/2, Pg13.5, M20 by 1.5 mm, 1/2 NPT, eight holes Installation: Panel, wall or 2B pipe mounting Case: Aluminum alloy Paint Color: Door: Sliver gray (Munsell 3.2PB7.4/1.2)
Case: Sliver gray (Munsell 3.2PB7.4/1.2)
Finish: Polyurethane corrosion-resistant coating Weight: Approx. 6 kg
IM 11M12A01-03E
2-9

2.3.2 Function

Display Functions:
Value Display; Displays values of the measured Oxygen concentration,
moisture quantity, mixture ratio etc.
Graph Display; Displays trends of measured oxygen concentration moisture
quantity, mixture ratio etc.
Data Display; Displays various useful data for maintenance, such as cell
temperature, reference junction temperature, maximum/
minimum moisture quantity and the like.
Status Messages; Indicates an alarm or error occurrence with flashing of the corre-
sponding icon. Indicates status such as warming-up, calibrating, or the like by symbols.
Alarm, Error Display: Displays alarms such as “Abnormal moisture quantity” or
errors such as “Abnormal cell e.m.f.” when any such status occurs.
Calibration Functions:
Auto-Calibration; Requires the Auto-calibration Unit. It calibrates automatically at
specified intervals.
Semi-auto Calibration; Requires the Auto-calibration Unit. Input calibration direction
on the touchpanel or contact, then it calibrates automatically afterwards.
Manual Calibration; Calibration with opening/closing the valve of calibration gas in
operation interactively with an LCD panel.
Blowback Function:
Output through the contact at a set period and time. Auto/semi-auto selectable.
Maintenance Functions:
Can update data settings during daily operation and checking. Display data settings, calibration data settings, blowback data settings,
current output loop check, input/output contact check.
Setup Functions:
Initial settings suit for the plant conditions when installing the con verter. Equipment settings, current output data settings, alarm data settings, contact data settings, other settings.
Self-diagnosis:
This function diagnoses conditions of the converter or the probe and indicates when any abnormal condition occurs.
Password Functions:
Enter your password to operate the analyzer excepting data display. Individual passwords can be set for maintenance and setup.
2-10
IM 11M12A01-03E
Display and setting content:
Measuring-related items: Oxygen concentration (vol% O2), moisture quantity(vol% H2O),
Display Related Items: Oxygen concentration (vol% O2), Moisture quantity (vol% H2O),
Calibration Setting Items: Span gas concentration (vol% O2), zero gas concentration
Output Related Items: Analog output/output mode selection, value conditions when
Alarm Related Items: Oxygen concentration high-alarm/high-high alarm limit values
2. Specifications
mixture ratio (kg/kg), relative humidity (%RH) and dew point (8 C)
mixture ratio (kg/kg), relative humidity (%RH), dew point (8 C), cell tempera­ture (8 C), thermocouple reference junction temperature (8 C), maximum/ minimum/average oxygen concentration (vol% O2), maxi-mum/minimum/ average moisture quantity (vol% H2O), maximum/minimum/average mixture ratio (kg/kg), cell e.m.f. (mV), output 1, 2 current (mA), cell response time (seconds), cell internal resistance (V), cell condition (in four grades), heater on­time ratio (%), calibration history (ten times), time (year/month/day/hour/ minute)
(vol% O2), calibration mode (auto, semi-auto, manual), calibration type and method (zero-span calibration, zero calibration only, span calibration only), stabilization time (min.sec), calibration time (min.sec), calibration period (day/ hour), starting time (year/month/day/hour/minute)
warming-up/maintenance/calibrating(or blowback)/abnormal, oxygen concentra­tion at 4mA/20mA (vol% O2), moisture quantity at 4mA/20mA (vol% H2O), mixture ratio at 4mA/20mA (kg/kg), time constant.
(vol% O2), Oxygen concentration low-alarm/low-low alarm limit values (vol% O2), moisture quantity high-alarm/high-high alarm limit values (vol% H2O), moisture quantity low-alarm/low-low alarm limit values (vol% H2O), mixture ratio high-alarm/high-high alarm limit value (kg/kg), mixture ratio low-alarm/ low-low alarm limit values (kg/kg), oxygen concentration alarm hysteresis (vol% O2), moisture quantity alarm hysteresis (vol% H2O), mixture ratio alarm hysteresis (kg/kg), oxygen concentration/moisture quantity/mixture ratio alarm detection, alarm delay (seconds).
Contact Related Items: Selection of contact input 1 and 2, selection of contact output 1
to 3 (abnormal, high-high alarm, high-alarm, low-alarm, low-low alarm, maintenance, calibrating, range switching, warming-up, calibration-gas pres­sure-decrease, temperature high-alarm, blow back, unburnt gas detected)
Converter Output: Two points mA analog output (4 to 20 mA DC (maximum load
resistance of 550V) and one of two mA digital output points (HART) (mini­mum load resistance of 250V).
Range: any setting between 0 to 5 through 0 to 100 vol% O2, 0 to 25 through 0 to 100
vol% H2O, 0 to 0.200 through 0 to 1.000 kg/kg or partial range is available. For the log output, the minimum range values are fixed at 0.1 vol% O2 for the
oxygen concentration, 0.1 vol% H2O for the moisture quantity, and 0.01 kg/kg for the mixture ratio.
4 to 20 mA DC linear or log can be selected. Input/output isolation provided. Output damping 0-255 (sec.) Can select hold or non-hold, and set preset value for hold.
IM 11M12A01-03E
2-11
Loading...
+ 174 hidden pages