Model ZR22G, ZR402G
Separate type
Zirconia High Temperature Humidity Analyzer
IM 11M12A01-03E
IM 11M12A01-03E
7th Edition
Introduction
The EXAxt ZR Separate-type Zirconia High-temperature Humidity Analyzer has been
developed for humidity control in various industrial processes. This analyzer basically
consists of a detector and a converter. You can select between several versions based
upon your application.
Optional accessories are also available to improve measurements and automate calibration. An optimal control system can be realized by adding appropriate options.
This instruction manual refers to almost all of the equipment related to the EXAxt ZR.
You may skip any section(s) on the equipment which is/are not included in your system.
Regarding the HART communication protocol,refer to the IM 11M12A01-51E.
IM 11M12A01-51E has been published as Model EXAxt ZR Series HART Protocol
The all-in-one version (with sensor and analyzer integrated in one body) is described in
IM 11M12A01-05E
< Before using the equipment, please read any related descriptions in this manual for the
equipment and the system you have, on appropriate use and operation of the EXAxt ZR. >
Models and descriptions in this manual are as follows:
Models and descriptions in this manual
Description in this manual
Model Product Name Specification Installation Operation Maintenance CMPL
ZR22G General-use detectorssss
ZR402G Converterss sss
ZH21B Dust protectorss
ZA8F Flow setting unit (for manual calibration use)ss s
ZR40H Automatic Calibration unitss ss
- Calibration gas unit case (Part No. E7044KF)ss
- Check valve (Part No. K9292DN, K9292DS)ss
ZO21S Standard gas unitssss
1. OverviewEquipment models and system configuration examples
2. SpecificationsStandard specification, model code (or part number),
dimension drawing for each equipment
3. InstallationInstallation method for each equipment
4. PipingExamples of piping in three standard system
configurations
5. WiringWiring procedures such as Power supply wiring , output
signal wiring or others
6. ComponentsThese are described in this manual
7. Startup Basic procedure to start operation of EXAxt ZR. Chapter 7
enables you to operate the equipment immediately.
8. Detailed Data SettingDetails of key operations and displays
9. CalibrationDescribes the calibration procedure required in the course
of operation.
10. Other FunctionsOther functions described
11. Inspection andHow to conduct maintenance of EXAxt ZR and procedures
Maintenancefor replacement of deteriorated parts
12. TroubleshootingThis chapter describes measures to be taken when an
abnormal condition occurs.
CMPL (parts list)User replaceable parts list
: Read and completely understand before operating the equipment.
s
: Read before operating the equipment, and refer to whenever necessary.
s
: Recommended to read it at least once.
n
s
s
s
s
s
n
n
s
s
s
s
s
s
s
n
n
s
s
n
n
n
s
n
n
n
n
s
s
s
T.Int.2E
ii
IM 1M12A01-03E
r For the safe use of this equipment
CAUTION
The cell (sensor) at the tip of the detector is made of ceramic (zirconia element). Do not
drop the detector or subject it to pressure stress.
• Do NOT allow the sensor (probe tip) to make contact with anything when installing
the detector.
• Avoid any water dropping directly on the probe (sensor) of the detector when installing it.
• Check the calibration gas piping before introducing the calibration gas to ensure that
there is no leakage of the gas. If there is any leakage of the gas, the moisture drawn
from the measuring gas can damage the sensor.
• The detector (especially at the tip) becomes very hot. Be sure to handle it with gloves.
EXAxt ZR is very heavy. Handle it with care. Be sure not to accidentally drop it.
Handle safely to avoid injury.
DANGER
(1) About This Manual
Connect the power supply cord only after confirming that the supply voltage matches
the rating of this equipment. In addition, confirm that the power is switched off when
connecting power supply.
Some process gas is dangerous to people. When removing this equipment from the
process line for maintenance or other reasons, protect yourself from potential
poisoningby using a protective mask or ventilating the area well.
• This manual should be passed on to the end user.
• The contents of this manual are subject to change without prior notice.
• The contents of this manual shall not be reproduced or copied, in part or in whole,
without permission.
• This manual explains the functions contained in this product, but does not warrant that
those will suit the particular purpose of the user.
• Every effort has been made to ensure accuracy in the preparation of this manual.
However, should any errors or omissions come to the attention of the user, please
contact the nearest Yokogawa Electric representative or sales office.
• This manual does not cover the special specifications. This manual may not be
changed on any change of specification, construction and parts when the change does
not affect the functions or performance of the product.
• If the product is used in a manner not specified in this manual, safety of this product
may be impaired.
IM 11M12A01-03E
iii
(2) Safety and Modification Precautions
• Follow the safety precautions in this manual when using the product to ensure protection and safety of personnel, product and system containing the product.
(3) The following safety symbols are used on the product as well as in this manual.
DANGER
This symbol indicates that the operator must follow the instructions laid out in this
manual in order to avoid the risk of personnel injuries or fatalities such as electric shock.
The manual describes what special care the operator must exercise to prevent such risks.
WARNING
This symbol indicates that the operator must refer to the instructions in this manual in
order to prevent the instrument (hardware) or software from being damaged, or a system
failure from occurring.
CAUTION
This symbol draws attention to information essential for understanding the operation and
functions.
Tip
This symbol gives information that complements the present topic.
SEE ALSO
This symbol identifies a source to which to refer.
Protective Ground Terminal
Function Ground Terminal (Do not use this terminal as the protective ground
terminal.)
Alternating current
iv
IM 1M12A01-03E
• Special descriptions in this manual
This manual indicates operation keys, displays and drawings on the product as follows:
• Operation keys, Enclosed in [ ], displays on the panel 0 0.
• Drawing for flashing
Indicated by gray characters. (Flashing) (lit)
(Ex. [MODE] key)
(Ex. message display 0 BASE 0)
(Ex. data display 0 102 0 lit, 0 102 0 flashing)
IM 11M12A01-03E
v
r NOTICE
• Specification check
When the instrument arrives, unpack the package with care and check that the
instrument has not been damaged during transportation. In addition, please check that
the specification matches the order, and required accessories are not missing. Specifications can be checked by the model codes on the nameplate. Refer to Chapter 2
Specifications for the list of model codes.
• Details on operation parameters
When the EXAxt ZR Separate-type High Temperature Humidity Analyzer arrives at
the user site, it will operate based set before shipping parameters (initial data) on the
departure from the factory.
Ensure that the initial data is suitable for the operation conditions before conducting
analysis. Where necessary, set the instrument parameters for appropriate operation.
For details of setting data, refer to chapters 7 to 10.
When user changes the operation parameter, it is recommended to note down the
changed setting data.
r After -Sales Warranty
d Do not modify the product.
d During the warranty period, for repair under warranty carry or send the product to the
local sales representative or service office. Yokogawa will replace or repair any
damaged parts and return the product to you.
d Before returning a product for repair under warranty, provide us with the model
name and serial number and a description of the problem. Any diagrams or data
explaining the problem would also be appreciated.
d If we replace the product with a new one, we won’t provide you with a repair report.
d Yokogawa warrants the product for the period stated in the pre-purchase quotation.
Yokogawa shall conduct defined warranty service based on its standard. When the
customer site is located outside of the service area, a fee for dispatching the maintenance engineer will be charged to the customer.
d In the following cases, customer will be charged repair fee regardless of warranty
period.
• Failure of components which are out of scope of warranty stated in instruction
manual.
• Failure caused by usage of software, hardware or auxiliary equipment, which
Yokogawa Electric did not supply.
• Failure due to improper or insufficient maintenance by user.
• Failure due to modification, misuse or outside-of-specifications operation which
Yokogawa does not authorize.
• Failure due to power supply (voltage, frequency) being outside specifications or
abnormal.
• Failure caused by any usage out of scope of recommended usage.
• Any damage from fire, earthquake, storms and floods, lightning, disturbances, riots,
warfare, radiation and other natural changes.
vi
IM 1M12A01-03E
d Yokogawa does not warrant conformance with the specific application at the user
site. Yokogawa will not bear direct/indirect responsibility for damage due to a specific
application.
d Yokogawa Electric will not bear responsibility when the user configures the product
into systems or resells the product.
d Maintenance service and supplying repair parts will be covered for five years after
the production ends. For repair for this product, please contact the nearest sales office
described in this instruction manual.
IM 11M12A01-03E
vii
Contents
Introduction........................................................................................................................... i
r For the safe use of this equipment ............................................................................... iii
r NOTICE.........................................................................................................................vi
r After-Sales Warranty ................................................................................................... vi
12.1Displays and Measures to Take When Errors Occur ................................... 12-1
12.1.1 What is an Error? .................................................................................. 12-1
12.1.2 Measures to Take When an Error Occurs ............................................. 12-2
12.2Displays and Measures to Take When Alarms are Generated ..................... 12-5
12.2.1 What is an Alarm? ................................................................................. 12-5
12.2.2 Measures Taken When Alarms are Generated.................................. 12-7
12.3Countermeasures When Measured Value Shows Error ............................. 12-13
12.3.1 Measured Value Higher Than True Value.......................................... 12-13
12.3.2 Measured Value Lower Than True Value .......................................... 12-14
12.3.3 Measurements Sometimes Show Abnormal Values........................... 12-14
Customer Maintenance Parts List ................................................ CMPL 11M12A01-03E
Customer Maintenance Parts List ................................................ CMPL 11M12C01-01E
Customer Maintenance Parts List ................................................ CMPL 11M12A01-11E
Customer Maintenance Parts List .................................................... CMPL 11M3D1-01E
Revision Record .................................................................................................................... i
xii
IM 1M12A01-03E
1.Overview
The EXAxt ZR Separate-type Zirconia High-temperature Humidity Analyzer is used to
measure the humidity of hot gases continuously in driers which use hot gas or electricity
as the heat source. It can also be used in a variety of manufacturing applications in
humidifiers, as well as in driers, for humidity measurement and control. It can help
improve productivity in these application fields.
The ZR402G Separate-type converter is equipped with an LCD touchpanel which has
various setting displays, a calibration display, humidity trend display, with easier
operation and improvement of display functions. The converter is equipped with various
standard functions such as measurement and calculation as well as maintenance functions including a self-test. Analyzer calibration can also be fully automated and ZR40H,
the automatic calibration unit, is provided. Choose the detector which best suits your
needs so that an optimal humidity control system can be obtained.
Some examples of typical system configuration are illustrated following pages :
1. Overview
IM 11M12A01-03E
1-1
1.1< EXAxt ZR > System Configuration
The system configuration should be determined by the conditions; e.g. whether the
calibration gas flow should be automated. The system configuration can be classified
into three basic patterns as follows:
1.1.1System 1
This is the simplest system consisting of a detector and a converter. This system can be
implemented for monitoring humidity in a production process such as food production.
No piping is required for the reference gas (air) which is fed in at the installation site.
The handy ZO21S standard gas unit is used for calibration.
Zero gas from this unit and span gas (air) is sent to the detector through a tube which is
connected during calibration.
CAUTION
• A needle (stop) valve should be connected to the calibration gas inlet of the detector.
The valve should be fully closed unless calibration is in progress.
• As this system uses ambient air for the reference gas, measuring accuracy will be
affected by the installation location.
Separate type Zirconia
High Temperature Humidity Analyzer,
Model ZR22G Detector
Stop valve
Signal
(6-core shield cable)
Heater (2-core)
Calibration gas
Figure 1.1
Model ZR402G Converter
EXA
ZR402G
100 to 240 V AC
,
Contact input
Analog outout, contact output
Digital output (HART)
Model ZO21S Standard gas unit
,
F1.1E.EPS
1-2
IM 11M12A01-03E
1.1.2System 2
1. Overview
This system is for accurate monitoring and controlling humidity when the installation
environment is polluted with gases other than the air. Instrument air (clean and dry air
of oxygen concentration 21%) is used for the reference gas and the span gas for calibration. Zero gas is fed in from a cylinder during calibration. The gas flow is controlled by
the ZA8F flow setting unit (for manual valve operation).
Separate type Zirconia
High Temperature Humidity Analyzer,
Model ZR22G Detector
Check valve
or Stop valve
Calibration gas
Signal
(6-core shield cable)
Heater (2-core)
Reference
gas
Model ZA8F flow setting unit
Figure1.2
Model ZR402G Converter
Flowmeter
Needle
EXA
ZR402G
valve
100 to 240 V AC
,
Contact input
Analog outout, contact output
Digital output (HART)
Air Set
Instrument air
Span gas (Same as Zero gas
calibration unit)
Calibration gas pressure regulator
Calibration gas
unit case
Zero gas cylinder
F1.2E.EPS
IM 11M12A01-03E
1-3
1.1.3System 3
This system is for accurate monitoring and controlling of humidity. Instrument air
(clean and dry air) is used for the reference gas and the span gas for calibration. Zero
gas is fed in from a cylinder during calibration. The calibration gas flow is controlled
automatically by the ZR40H automatic calibration unit.
This system is similar to system 2, except that the calibration gas flow is automated
using the ZR40H automatic calibration unit.
Separate type Zirconia
High Temperature Humidity Analyzer,
Model ZR22G Detector
Check valve
*2
Reference gas
Calibration gas
Signal
(6-core shield cable)
Heater (2-core shield cable)
*1
Model ZR40H Auto-Calibration unit
Flowmeter
Model ZR402G Converter
EXA
ZR402G
,
Needle
vaive
Air Set
*3
100 to 240 V AC
Contact input
Analog output, contact output
Digital output (HART)
lnstrument air
Calibration gas pressure regulator
Calibration gas
unit case
Zero gas cylinder
F1.3E.EPS
Figure1.3
*1 Shield cable : Use shielded signal cables, and connect the shield to the FG terminal
of the converter.
*2 Select the desired probe from the Probe Configuration table on page 1-5
1-4
*3 100% N2 gas cannot be used as the zero gas. Use approximately 1% of O2 gas (N2 based).
IM 11M12A01-03E
1.2< EXAxtZR > System Components
1.2.1System Components
Model or Part No. Product Name System 1 System 2 System 3
ZR22GDetector, Separate-type High-temperature
Humidity Analyzer, Detector
ZR402GConverter, Separate-type High-temperature
Humidity Analyzer, Converter
ZH21BDust protector
ZO21SStandard gas unit
ZA8FFlow setting unit (for manual calibration)
ZR40HAutomatic calibration unit (for separate type)
L9852CB/G7016XHStop valve
K9292DN/K9292DSCheck valve
K9473XH/K9473XJ,Air set
G7004XF/K9473XG
G7001ZCZero gas cylinder
G7013XF/G7014XFPressure regulator for gas cylinder
E7044KFCase assembly for calibration-gas cylinder
ZR22AHeater Assembly
(Spare parts for Model ZR22G)
d : The essential products for the system
s : Selected depending on your applications
(d) : Select either
1. Overview
Separate type
d d d
d d d
s s s
d
d
d
d (d)
(d) d
d d
d d
d d
d d
s s s
T1.1E.EPS
1.2.2Detectors and Accessories
dGeneral-use Detector (sample gas temperature: 0 to 7008C)
Standard detectorModel
ZR22G
Mounting angle
The probe insertion length: 0.4 to 2m
Vertical to horizontal
The probe insertion length: 2.5 to 3m
Vertical
Detector with dust protector
Model
ZR22G-040
and
ZH21B
Mounting angle
Vertical to horizontal
(The probe insertion length: 0.4m)
F1.4E.EPS
IM 11M12A01-03E
1-5
2.Specifications
This chapter describes the specifications for the following:
ZR22GGeneral-use separate-type detector (See Section 2.2.1)
ZH21BDust protector(See Section 2.2.2)
ZR402GSeparate-type converter(See Section 2.3)
ZA8FFlow setting unit (See Section 2.4.1)
ZR40HAutomatic calibration unit(See Section 2.4.2)
ZO21SStandard gas unit(See Section 2.5)
2.1General Specifications
2.1.1Standard Specifications
High Temperature Humidity Analyzer
Oxygen concentration in mixed gas which consists of water vapor and air is proportional
to the volume rate of air, so the volume rate of water vapor can be calculated by the
oxygen concentration.
2. Specifications
Measured Object:Water vapor (in vol%) in mixed gases (air and water vapor)
Measured System: Zirconia system
Measured Range: 0.01 to 100 vol% O2, 0 to 100 vol% H2O or 0 to 1.000 kg/kg
Output Signal: 4 to 20 mA DC (maximum load resistance 550V) Oxygen concentration;
Any setting in the range of 0 to 5 through 0 to 100 vol%O2 (in 1
vol% O2), or partial range.
Moisture quantity: 0 to 25 through 0 to 100 vol% H2O (in 1 vol% H2O), or partial
range.
Mixture ratio 0 to 0.2 through 0 to 1.000 kg/kg (in 0.001 kg/kg), or
partial range.
Digital Communication (HART): 250 to 550V, depending on quantity of field devices
connected to the loop (multi-drop mode).
(Note) HART is a registered trademark of the HART Communication
vol% H2O
Mixture ratio; 0 to 1 kg/kg
Relative humidity; 0 to 100 %RH
Dew point; -40 to 3708 C
(Note) Those values are calculated by temperature and absolute
pressure. Then accurate temperature and pressure value must be
input to the converter.
Warm-up Time: Approx. 20 min.
(Note) These characteristics are calculated by oxygen concentration
measured in air which include water vapor.
IM 11M12A01-03E
2-1
Repeatability: (Note1)
61 vol% H2O (sample gas pressure 2 kPa or less)
Linearity:(Excluding standard gas tolerance)
(Note1)
(Use oxygen of known concentration (in the measuring range) as the
zero and span calibration gas.)
62 vol% H2O; (Sample gas pressure: within 60.49 kPa)
63 vol% H2O; (Sample gas pressure: 2 kPa or less)
Drift:(Excluding the first two weeks in use)
(Note1)
both zero and span 63 vol% H2O/month
Response Time: Response of 90 % within 5 seconds. (Measured after gas is introduced
from calibration-gas inlet and analog output starts changing.)
(Note1) These tolerances do not apply to the pressure compensated
version, or where natural convection is used for the reference air.
2-2
IM 11M12A01-03E
2.2General-use Separate-type Detector and Related
Equipment
The “Detector with dust protector” consists of ZR22G general-use separate-type detector
and ZH21B dust protector (refer to Section 2.2.2).
2.2.1ZR22G General-use Separate-type Detector
Sample Gas Temperature: 0 to 7008 C (Probe only)
It is necessary to mount the cell using Inconel cell-bolts when the
temperature is 6008 C or greater.
Sample Gas Pressure: -5 to +20 kPa (When the pressure in the process exceeds 3kPa,
it is recommended that you compensate the pressure. When the pressure
in the process exceeds 5kPa, you must perform pressure compensation.)
No pressure fluctuation in the process should be allowed.
Probe Length: 0.4, 0.7, 1.0, 1.5, 2.0, 2.5, 3.0 m
Probe Material: SUS 316 (JIS)
Ambient Temperature: -20 to +1508 C
Reference Air System: Natural Convection, Instrument Air
Instrument Air System (excluding Natural Convection):
Pressure; 200 kPa + the pressure inside the dryer, (It is recommended to
use air which has been dehumidified by cooling to dewpoint -208 C or
less, and with dust or oil mist removed.)
Consumption; Approx. 1Nl/min
Note: When the detector is used in conjunction with a check valve and a
ZA8F Flow Setting Unit, the maximum pressure of sample gas is 150
kPa. When with a check valve and a ZR40H Auto Calibration Unit, it is
200 kPa. If the pressure of your sample gas exceeds these limits, consult
with Yokogawa.
Material in Contact with Gas: SUS 316 (JIS), Zirconia, SUS 304 (JIS) (flange),
Hastelloy B, (Inconel 600, 601)
Construction: Heater and thermocouple replaceable construction. Non explosion-proof
JIS C0920 / equivalent to IP44D. Equivalent to NEMA 4X / IP66
(Achieved when the cable entry is completely sealed with a cable
gland in the recirculation pressure compensated version.)
Probe Mounting Angle: Horizontal to vertically downward.
When the probe insertion length is 2 m or less, installing at angles from
horizontal to vertically downward is available.
When the probe insertion length exceeds 2.5 m, mount vertically down-
ward (within 658 ) and use a probe protector.
Weight:
Insertion length of 0.4 m: approx. 6 kg (JIS 5K 65) / approx. 11 kg (ANSI 150 4)
Insertion length of 1.0 m: approx. 8 kg (JIS 5K 65) / approx. 13 kg (ANSI 150 4)
Insertion length of 1.5 m: approx. 10 kg (JIS 5K 65) / approx. 15 kg (ANSI 150 4)
Insertion length of 2.0 m: approx. 12 kg (JIS 5K 65) / approx. 17 kg (ANSI 150 4)
Insertion length of 3.0 m:approx. 15 kg (JIS 5K 65) / approx. 20 kg (ANSI 150 4)
2-4
IM 11M12A01-03E
2. Specifications
Model and Codes
Model
ZR22G
Suffix code
Option
code
Description
Separate type Detector of Zirconia High Temperature Humidity
[ Style : S2 ]
Analyzer
Length
-040
-070
-100
-150
-200
-250
-300
Wetted material
-S
-C
Flange
(*2)
-A
-B
-C
-E
-F
-G
-K
-L
-M
-P
-R
-S
-W
Reference air
-C
-E
-P
Gas Thread
-T
Connection box thred
-P
-G
-M
-T
-Q
Instruction manual
-J
-E
-A
Options
/D
/C
/CV
/SV
/PT
-R
/SCT
0.4 m
0.7 m
1.0 m
1.5 m
2.0 m
2.5 m (*1)
3.0 m (*1)
SUS316
Stainless steel with lnconel calibration gas tube
ANSI Class 150 2 RF SUS304
ANSI Class 150 3 RF SUS304
ANSI Class 150 4 RF SUS304
DIN PN10 DN50 SUS304
DIN PN10 DN80 SUS304
DIN PN10 DN100 SUS304
JIS 5K 65 FF SUS304
JIS 10K 65 FF SUS304
JIS 10K 80 FF SUS304
JIS 10K 100 FF SUS304
JPI Class 150 4 RF SUS304
JPI Class 150 3 RF SUS304
Westinghouse
Natural convection
External connection (Instrument air) (*8)
Pressure Compensation (*8)
Rc 1/4
1/4 NPT(F)
G1/2
Pg13.5
M20 x1.5 mm
1/2NPT
Quick connect (*6)
Japanese
English
Always-A
DERAKANE coating (*7)
Inconel bolt (*3)
Check valve (*4)
Stop valve (*4)
Stainless steel tag plate (*5)
Printed tag plate (*5)
*1 When installing horizontally the probe whose insertion length is 2.5 meters or more, use the Probe Protector. Be sure to specify
ZO21R-L-hhh-h. Specify the flange suffix code either -C or -K.
*2 The thickness of the flange depends on its dimensions.
*3 Inconel probe bolts and U shape pipe for calibration are used. Use this option for high temperature use (ranging from 600 to 700 8C).
*4 Specify either /CV or /SV option code.
*5 Specify either /SCT or /PT option code.
*6 Not waterproof, protect from rain.Operating maximum temperature is 80 8C. Available only in the U.S.
*7 Available only in the U.S. DERAKANE is a registered trademark of the Dow Chemical Company.
*8 Piping for reference air must be installed to supply reference air constantly at a specified flow rate.
T2.1E.EPS
IM 11M12A01-03E
2-5
EXTERNAL DIMENSIONS
Model ZR22G Detectors Separate type Zirconia High Temperature Humidity Analyzer
[50.8
L=0.15, 0.4, 0.7, 1.0,
1.5, 2.0, 2.5, 3.0 (m)
L
t
Flange
283 to 292
155 to 16369
25
85
Rc1/4 or 1/4NPT
Reference air inlet
2-G1/2,2-1/2NPT etc.
Cable connection port
48
Rc1/4 or 1/4NPT
Calibration gas inlet
[124
C
[A
[B
Flange
Flange
ANSI Class 150 2 RF SUS304
ANSI Class 150 3 RF SUS304
ANSI Class 150 4 RF SUS304
DIN PN10 DN50 SUS304
DIN PN10 DN80 SUS304
DIN PN10 DN100 SUS304
JIS 5K 65 FF SUS304
JIS 10K 65 FF SUS304
JIS 10K 80 FF SUS304
JIS 10K 100 FF SUS304
JIS 5K 32 FF SUS304
JPI Class 150 4 RF SUS304
JPI Class 150 3 RF SUS304
Westinghouse
This protector is designed to protect the probe output from dust agitation (i.e., to prevent
combustible materials from entering the probe cell where humidity measurements are
made) in a dusty environment.
dStandard Specification
Insertion length: 0.428m
Flange:JIS 5K 80 FF SUS304 or ANSI Class 150 4 FF SUS304 (However,
Material:SUS 316(JIS), SUS 304(JIS) (flange)
Weight:Approximately 6kg (JIS), approximately 8.5kg (ANSI)
Mounting:Mounted on the probe or process flange with bolts and associated
flange thickness is different)
nuts and washers.
ModelSuffixcode
ZH21B
Insertionlength
Flange(*1)
Stylecode
*Theflangethicknessvaries.
SpecifytheprobeZR22G-040-h-Kincodeof(1)
-0400.428m
-A
Option
code
DustProtector(0to6008C)
-J
*B
ZR22G-040-h-Cincodeof(2)
JIS5K80FFSUS304(1)ANSIClass1504BFFSUS304(2)
StyleB
Dust protector
72
[
Install facing upwards
C
[B
Description
428
T2.2E.EPS
76.3
[
D
[B
Unit ; mm
t
[
A
C
2-8
Insertion hole;minimum[80
ANSI flange
Size of each part of flange
Flange
JIS5K80FFSUS304
ANSIClass1504BFFSUS304
A
180
228.6
B
145
190.5
fourholesof
diameter19
eightholesofdiameter19
Ct
12
12
2.3E-1.EPS
JIS flange
D
40
50
Insertion hole;minimum[80
F2.3E.EPS
IM 11M12A01-03E
2.3ZR402G Separate-type Converter
2.3.1Standard Specifications
Operated using an LCD touchscreen on the converter.
Display: LCD display of size 320 by 240 dot with touchscreen.
Output Signal: 4 to 20 mA DC, two points (maximum load resistance 550 V)
Contact Output Signal: four points (one is fail-safe, normally open)
Contact Input: two points
Analog Input: one point (thermal input 4-20 mA)
Auto-calibration Output: Two points (for dedicated auto-calibration unit)
Ambient Temperature: -20 to +558C
Storage Temperature: -30 to +708C
Humidity Ambient: 0 to 95%RH (Non-condensing)
Installation Altitude: 2000 m or less
Category based on IEC 1010: II (Note)
2. Specifications
Pollution degree based on IEC 1010: 2 (Note)
Note: Installation category, called over-voltage category, specifies impulse withstanding voltage. Category II is for electrical equipment.
Pollution degree indicates the degree of existence of solid, liquid, gas or other
inclusions which may reduce breakdown voltage. Degree 2 is the normal indoor
environment.
Power Supply Voltage: Ratings; 100 to 240 V AC
Acceptable range; 85 to 264 V AC
Power Supply Frequency:Ratings; 50/60 Hz Acceptable range; 45 to 66 Hz
Power Consumption: Max. 300 W, approx. 100 W for ordinary use.
Safety and EMC conforming standards
Safety: EN61010-1
CSA C22.2 No.61010-1
UL61010-1
EMC:EN 61326 Class A
EN 55011 Class A Group 1
EN 61000-3-2
AS/NZS CISPR 11
Maximum Distance between Probe and Converter: Conductor two-way resistance must
be 10 V or less (when a 1.25 mm2 cable or equivalent is used, 300 m or less.)
Construction: Outdoor installation, equivalent to NEMA 4 (with conduit holes
completely sealed with a plastic cable gland optional)
Wiring Connection: G1/2, Pg13.5, M20 by 1.5 mm, 1/2 NPT, eight holes
Installation: Panel, wall or 2B pipe mounting
Case: Aluminum alloy
Paint Color: Door: Sliver gray (Munsell 3.2PB7.4/1.2)
Case: Sliver gray (Munsell 3.2PB7.4/1.2)
Finish: Polyurethane corrosion-resistant coating
Weight: Approx. 6 kg
IM 11M12A01-03E
2-9
2.3.2Function
Display Functions:
Value Display;Displays values of the measured Oxygen concentration,
moisture quantity, mixture ratio etc.
Graph Display;Displays trends of measured oxygen concentration moisture
quantity, mixture ratio etc.
Data Display; Displays various useful data for maintenance, such as cell
Status Messages;Indicates an alarm or error occurrence with flashing of the corre-
sponding icon. Indicates status such as warming-up, calibrating,
or the like by symbols.
Alarm, Error Display: Displays alarms such as “Abnormal moisture quantity” or
errors such as “Abnormal cell e.m.f.” when any such status
occurs.
Calibration Functions:
Auto-Calibration;Requires the Auto-calibration Unit. It calibrates automatically at
specified intervals.
Semi-auto Calibration; Requires the Auto-calibration Unit. Input calibration direction
on the touchpanel or contact, then it calibrates automatically
afterwards.
Manual Calibration; Calibration with opening/closing the valve of calibration gas in
operation interactively with an LCD panel.
Blowback Function:
Output through the contact at a set period and time. Auto/semi-auto
selectable.
Maintenance Functions:
Can update data settings during daily operation and checking.
Display data settings, calibration data settings, blowback data settings,
current output loop check, input/output contact check.
Setup Functions:
Initial settings suit for the plant conditions when installing the con
verter. Equipment settings, current output data settings, alarm data
settings, contact data settings, other settings.
Self-diagnosis:
This function diagnoses conditions of the converter or the probe and
indicates when any abnormal condition occurs.
Password Functions:
Enter your password to operate the analyzer excepting data display.
Individual passwords can be set for maintenance and setup.
Calibration Setting Items: Span gas concentration (vol% O2), zero gas concentration
Output Related Items: Analog output/output mode selection, value conditions when
Alarm Related Items: Oxygen concentration high-alarm/high-high alarm limit values
2. Specifications
mixture ratio (kg/kg), relative humidity (%RH) and dew point (8 C)
mixture ratio (kg/kg), relative humidity (%RH), dew point (8 C), cell temperature (8 C), thermocouple reference junction temperature (8 C), maximum/
minimum/average oxygen concentration (vol% O2), maxi-mum/minimum/
average moisture quantity (vol% H2O), maximum/minimum/average mixture
ratio (kg/kg), cell e.m.f. (mV), output 1, 2 current (mA), cell response time
(seconds), cell internal resistance (V), cell condition (in four grades), heater ontime ratio (%), calibration history (ten times), time (year/month/day/hour/
minute)
(vol% O2), calibration mode (auto, semi-auto, manual), calibration type and
method (zero-span calibration, zero calibration only, span calibration only),
stabilization time (min.sec), calibration time (min.sec), calibration period (day/
hour), starting time (year/month/day/hour/minute)
warming-up/maintenance/calibrating(or blowback)/abnormal, oxygen concentration at 4mA/20mA (vol% O2), moisture quantity at 4mA/20mA (vol% H2O),
mixture ratio at 4mA/20mA (kg/kg), time constant.
Contact Related Items: Selection of contact input 1 and 2, selection of contact output 1
to 3 (abnormal, high-high alarm, high-alarm, low-alarm, low-low alarm,
maintenance, calibrating, range switching, warming-up, calibration-gas pressure-decrease, temperature high-alarm, blow back, unburnt gas detected)
Converter Output: Two points mA analog output (4 to 20 mA DC (maximum load
resistance of 550V) and one of two mA digital output points (HART) (minimum load resistance of 250V).
Range: any setting between 0 to 5 through 0 to 100 vol% O2, 0 to 25 through 0 to 100
vol% H2O, 0 to 0.200 through 0 to 1.000 kg/kg or partial range is available.
For the log output, the minimum range values are fixed at 0.1 vol% O2 for the
oxygen concentration, 0.1 vol% H2O for the moisture quantity, and 0.01 kg/kg
for the mixture ratio.
4 to 20 mA DC linear or log can be selected.
Input/output isolation provided.
Output damping 0-255 (sec.)
Can select hold or non-hold, and set preset value for hold.
IM 11M12A01-03E
2-11
Loading...
+ 174 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.