Model AV550G
Zirconia Oxygen Analyzer
Averaging Converter
IM 11M12D01-01E
IM 11M12D01-01E
3rd Edition
Introduction
The EXAxt Series AV550G Zirconia Oxygen Analyzer Averaging Converter is designed
for oxygen measurement at multiple points in flue ducts of industrial furnaces and can
be used to optimize the combustion process. For the AV550G system, various types of
EXAxt ZR Series detectors as well as optional accessories are available. The best
measurement can be achieved by selecting instruments appropriate to your application.
This manual provides information, such as installation, operation, inspection and
maintenance procedures, about the instruments used in the AV550G averaging converter
measurement system. Any sections concerning instruments not included in your system
may be skipped.
Before using the instruments, read any descriptions related to your instruments and
system to ensure the best performance.
Regarding the HART communication, refer to IM 11M12D01-51E. IM 11M12D01-51E
has been published as "Model EXAxt AV550G HART protocol."
Regarding the FOUNDATION Fieldbus communication, refer to IM 11M12D01-61E.
IM 11M12D01-61E has been published as "Model EXAxt AV550G Fieldbus
communication."
The models and description items in this manual are as follows:
Models and descriptions in this manual
Description in this manual
Model Product Name Specification Installation Operation Maintenance CMPL
ZR22G General-use detectorss s ss
ZR22G High-temperature detector (0.15m)ss ss
AV550G Averaging Converterss sss
ZO21R-L Probe protectorss
ZO21P-H High-temperature probe adaptersss
ZA8F Flow setting unit (for manual calibration use)ss s
- Auxiliary ejector for high-temperature use
(Part No. E7046EC, EN)sss
- Calibration gas unit case (Part No. E7044KF)ss
- Check valve (Part No. K9292DN, K9292DS)ss
- Dust filter for the detector (Part No. K9471UA)ss
- Dust guard protector (Part No. K9471UC) ss
CMPL : Customer Maintenance Parts List
T.Int.1E
User's manuals of instruments related to the AV550G Averaging Converter are as
follows:
Model
ZR22G
ZO21DW
Title
Separate Type Zirconia Oxygen Analyzer
In-Site Type Zirconia Oxygen Analyzer
1. OverviewEquipment models and system configuration examples
2. SpecificationsStandard specification, model code (or part number),
dimension drawing for each equipment
3. InstallationInstallation method for each equipment
4. PipingExamples of piping in three standard system
configurations
5. WiringWiring procedures such as “Power supply wiring”, “output
signal wiring” or others
6. ComponentsMajor parts and function are described in this manual
7. Startup Basic procedure to start operation of AV550G. Chapter 7
enables you to operate the equipment immediately.
8. Detailed Data SettingDetails of key operations and displays
9. CalibrationDescribes the calibration procedure required in the course
of operation.
10. Other FunctionsOther functions described
11. Inspection andHow to conduct maintenance of AV550G and procedures
Maintenancefor replacement of deteriorated parts
12. TroubleshootingThis chapter describes measures to be taken when an
abnormal condition occurs.
CMPL (parts list)User replaceable parts list
: Read and completely understand before operating the equipment.
s
: Read before operating the equipment, and refer to it whenever necessary.
s
: Recommended to read it at least once.
n
s
s
s
s
s
n
n
s
s
s
s
s
s
s
n
n
s
s
n
n
n
s
n
n
n
n
s
s
s
T.Int.2E
ii
IM 11M12D01-01E
r For the safe use of this equipment
CAUTION
The cell (sensor) at the tip of the detector is made of ceramic (zirconia element). Do not
drop the detector or subject it to pressure stress.
• Do NOT allow the sensor (probe tip) to make contact with anything when installing
the detector.
• Avoid any water dropping directly on the probe (sensor) of the detector when installing it.
• Check the calibration gas piping before introducing the calibration gas to ensure that
there is no leakage of the gas. If there is any leakage of the gas, the moisture drawn
from the measured gas may damage the sensor.
• The detector (especially at the tip) becomes very hot. Be sure to handle it with gloves.
DANGER
AV550G is very heavy. Handle it with care. Be sure not to accidentally drop it. Handle
safely to avoid injury.
Connect the power supply cord only after confirming that the supply voltage matches
the rating of this equipment. In addition, confirm that the power is switched off when
connecting power supply.
Some process gas is dangerous to people. When removing this equipment from the
process line for maintenance or other reasons, protect yourself from potential poisoning
by using a protective mask or ventilating the area well.
When carrying the AV550G Averaging Converter, mark sure this is done by two or
more people.
(1) About This Manual
• This manual should be passed on to the end user.
• The contents of this manual are subject to change without prior notice.
• The contents of this manual shall not be reproduced or copied, in part or in whole,
without permission.
• This manual explains the functions contained in this product, but does not warrant that
those will suit the particular purpose of the user.
• Every effort has been made to ensure accuracy in the preparation of this manual.
However, should any errors or omissions come to the attention of the user, please
contact the nearest Yokogawa Electric representative or sales office.
• This manual does not cover the special specifications. This manual may not be
changed on any change of specification, construction and parts when the change does
not affect the functions or performance of the product.
• If the product is used in a manner not specified in this manual, the safety of this
product may be impaired.
IM 11M12D01-01E
iii
(2) Safety and Modification Precautions
• Follow the safety precautions in this manual when using the product to ensure protection and safety of personnel, product and system containing the product.
(3) The following safety symbols are used on the product as well as in this manual.
DANGER
This symbol indicates that the operator must follow the instructions laid out in this
manual in order to avoid the risk of personnel injury electric shock, or fatalities. The
manual describes what special care the operator must exercise to avoid such risks.
WARNING
This symbol indicates that the operator must refer to the instructions in this manual in
order to prevent the instrument (hardware) or software from being damaged, or a system
failure from occurring.
CAUTION
This symbol draws attention to information essential for understanding the operation and
functions.
Tip
This symbol gives information that complements the present topic.
SEE ALSO
This symbol identifies a source to which to refer.
Protective Conductor Terminal
Function Earth Terminal (Do not use this terminal as the protective conductor
terminal.)
Alternating current
iv
IM 11M12D01-01E
• Special descriptions in this manual
This manual indicates operation keys, displays and drawings on the product as follows:
• Displays on the panel 0 0.
(Ex. message display 0 BASE 0)
(Ex. data display 0 1020 )
IM 11M12D01-01E
v
r NOTICE
• Specification check
When the instrument arrives, unpack the package with care and check that the
instrument has not been damaged during transportation. In addition, please check that
the specification matches the order, and required accessories are not missing. Specifications can be checked by the model codes on the nameplate. Refer to Chapter 2
Specifications for the list of model codes.
• Details on operation parameters
When the AV550G Averaging Converter arrives at the user site, it will operate based
on the operation parameters (initial data) set before shipping from the factory.
Ensure that the initial data is suitable for the operation conditions before conducting
analysis. Where necessary, set the instrument parameters for appropriate operation.
For details of setting data, refer to chapters 7 to 10.
When user changes the operation parameter, it is recommended to note down the
changed setting data.
r After -Sales Warranty
d Do not modify the product.
d During the warranty period, for repair under warranty carry or send the product to the
local sales representative or service office. Yokogawa will replace or repair any
damaged parts and return the product to you.
d Before returning a product for repair under warranty, provide us with the model
name and serial number and a description of the problem. Any diagrams or data
explaining the problem would also be appreciated.
d If we replace the product with a new one, we won’t provide you with a repair report.
d Yokogawa warrants the product for the period stated in the pre-purchase quotation.
Yokogawa shall conduct defined warranty service based on its standard. When the
customer site is located outside of the service area, a fee for dispatching the maintenance engineer will be charged to the customer.
d In the following cases, customer will be charged repair fee regardless of warranty
period.
• Failure of components which are out of scope of warranty stated in instruction
manual.
• Failure caused by usage of software, hardware or auxiliary equipment, which
Yokogawa Electric did not supply.
• Failure due to improper or insufficient maintenance by user.
• Failure due to modification, misuse or outside-of-specifications operation which
Yokogawa does not authorize.
• Failure due to power supply (voltage, frequency) being outside specifications or
abnormal.
• Failure caused by any usage out of scope of recommended usage.
• Any damage from fire, earthquake, storms and floods, lightning, disturbances, riots,
warfare, radiation and other natural changes.
vi
IM 11M12D01-01E
d Yokogawa does not warrant conformance with the specific application at the user
site. Yokogawa will not bear direct/indirect responsibility for damage due to a specific
application.
d Yokogawa Electric will not bear responsibility when the user configures the product
into systems or resells the product.
d Maintenance service and supplying repair parts will be covered for five years after
the production ends. For repair for this product, please contact the nearest sales office
described in this instruction manual.
IM 11M12D01-01E
vii
viii
IM 11M12D01-01E
Contents
Introduction........................................................................................................................... i
r For the safe use of this equipment .......................................................................... iii
r NOTICE ....................................................................................................................vi
r After-Sales Warranty................................................................................................vi
12.2.3 Remedies when Alarms are Generated .............................................. 12-9
12.3 Countermeasures When Measured Value Shows Error ........................... 12-16
12.3.1 Measured Value Higher Than True Value....................................... 12-16
12.3.2 Measured Value Lower Than True Value ....................................... 12-17
12.3.3 Measurements Sometimes Show Abnormal Values ........................ 12-17
Customer Maintenance Parts List ................................................CMPL 11M12D01-01E
Customer Maintenance Parts List ................................................CMPL 11M12A01-02E
Customer Maintenance Parts List ................................................CMPL 11M06B02-01E
Customer Maintenance Parts List ................................................CMPL 11M03B01-10E
Customer Maintenance Parts List ................................................CMPL 11M03B01-05E
Revision Record .................................................................................................................... i
IM 11M12D01-01E
xiii
1.Overview
Zirconia oxygen analyzers are used in combustion facilities to measure the flue gas
oxygen concentration. Boiler operators use the oxygen measurement to optimize fuel
usage, minimize atmospheric emissions and reduce energy consumption.
A multiple point oxygen measurement system may be required for situations when gas
stratification in the flue duct affects combustion control. The AV550G Averaging
Converter can accept inputs from up to eight zirconia oxygen detectors. It sends output
signals for the individual as well as averages of multiple oxygen concentrations. A
robust multipoint converter reduces installation and maintenance costs.
A large 5.7-inch color LCD display shows various measurement, setup, calibration, and
trend screens. Its intuitive touch screen is easy to read and makes set up and maintenance simple. Other standard features include new self-diagnostics and a hot swap
function that allows a desired probe to be disconnected/reconnected for inspection or
maintenance just by turning off the power of the relevant channel.
The ZR22G separate-type detector uses a highly reliable zirconia sensor and its heater
assembly can be replaced in the field. The in situ probe is mounted on the duct wall and
directly measures the oxygen concentration of a sample gas at a temperature of up to
7008C. For higher temperature applications up to 14008C, the ZO21P-H High
Temperature Probe Protector is available for use in conjunction with a 0.15-m ZR22G
general-purpose detector.
1. Overview
The averaging converter system is ideal for combustion control in large utility boilers or
various industrial furnaces.
This chapter explains system configurations with some typical examples.
1.1 System Configuration
The AV550G Zirconia Oxygen Analyzer Averaging Converter system can be configured
by selecting detectors and an averaging converter that meet the individual requirements
and flow setting units for calibration.
Subsection 1.1.1 presents a typical system configuration using flow setting units for
manual calibration. A typical system configuration to perform automatic calibration is
provided in Subsection 1.1.2.
1.1.1System Configuration Using Flow Setting Units for Manual Calibration
This system consists of detectors, an averaging converter, and flow setting units (Model
ZA8F), as shown in Figure 1.1. Note that the ZA8F Flow Setting Units are required as
many as the detectors connected to the averaging converter.
A reference gas needs to be supplied at a constant flow rate to the detectors. This
reference gas must be clean, dry air having a constant percentage of oxygen. Typically,
instrument air that has been dehumidified down to a dew point of approximately -208C
and is free from oil mist or dust, is used as the air source. This air is also used as a span
gas for the detectors during calibration. A zero gas for calibration is supplied from a
cylinder to the detectors.
Process gas temperature 0 to 7008CProcess gas temperature 0 to 14008C
Mounting
Horizontal
to
vertical
Vertical
Horizontal
to
vertical
Horizontal
to
vertical
Vertical
Insertion
length
0.4
to
2 m
2.5 m
or more
3 m
or less
0.4
to
2 m
2.5 m
or more
General-use ProbeHigh temperature detector
Probe Protector
(ZO21R)
Gas Flow
Sample inlet
Filter
(K9471UA)
or
Dust Guard
Protector
(K9471UC)
Model / Part Number/
SpecificationsManual Calibration
(A or B)
h
XP, E7057G
Detector
(ZO21DW-L)
Detector
(ZR22G)
+
Detector
(ZR22G)
Detector
(ZR22G)
h
Application
Boiler
Heating
furnace
For pulverized
coal boiler
with gas flow
velocity
10 m/s or more
Black liquid
recovery boiler
Cement
Kiln
Sample
outlet
Probe protector
for high
temperature use
ZO21P-H
Temperature:
Probe material SUS310S 8008C
Probe material SiC 14008C
Mounting: V ertical downwards
Insertion length: 1.0m, 1.5m
When duct pressure is atmospheric
or negative, attach air ejector.
High temperature auxiliary
ejector (E7046EC, E7046EN)
Needle
valve
Inlet
System 1
s
ssd
s
s
s
s
s
Absorption
structure
Sample inlet
Automatic Calibration
High
temperature
detector
Pressure gauge
Ejector
assy.
Blow
System 2
s
ssd
s
s
s
s
s
s
s
s
Tbl 1-2.eps
Application
Heating
furnace
F1.4E.EPS
IM 11M12D01-01E
1-3
1-4
IM 11M12D01-01E
2.Specifications
This chapter describes the specifications for the following:
ZR22GGeneral-use separate-type detector (See Section 2.2.1)
ZO21R-LProbe protector (See Section 2.2.2)
ZR22G (0.15 m)High-temperature separate-type detector (See Section 2.3.1)
ZO21P-HAdapter for High temperature probe (See Section 2.3.2)
AV550GAveraging converter (See Section 2.4)
ZA8FFlow setting unit (See Section 2.5.1)
Other equipment (See Section 2.6)
2.1General Specifications
2.1.1Standard Specifications
2. Specifications
Measured Object: Oxygen concentration in combustion exhaust gas and mixed gas
(excluding inflammable gases)
To check if sampled gases containing a corrosive gas such as ammonia or
chlorine may be applicable to our oxygen gas analyzer, contact with
YOKOGAWA or its agent.
Measurement System: Zirconia system
Oxygen concentration: 0.01 to 100 vol% O
Output Signal: 4 to 20 mA DC (maximum load resistance 550V)
Measurement Range: Any setting in the range of 0 to 5 through 0 to 100 vol% O2 (in
1 vol% O2), or partial range
Digital Communication (HART): 250 to 550V, depending on number of field devices
connected to the loop (multi-drop mode).
Note: HART is a registered trademark of the HART Communication Foundation.
Display Range:0 to 100 vol% O
Warm-up Time:Approx. 20 min.
Repeatability: (Excluding the case where the reference air is by natural convection)
60.5% Maximum value of range setting. (range up to 0 to 25 vol% O2)
2
2
IM 11M12D01-01E
61% Maximum value of range setting.
(range from 0 to 25 vol% O2 up to 0 to 100 vol% O2)
Linearity:(Excluding standard gas tolerance)
2-1
(Excluding the case where the reference air flow is natural convection)
(Use oxygen of known concentration (in the measuring range) as the zero and
span calibration gas.)
61% Maximum value of range setting ; from 0 to 5 vol% O2 to 0 to 25
vol% O2 range
(Sample gas pressure: within 64.9 kPa)
63% Maximum value of range setting ; from 0 to 25 vol%O2 to 0 to 50 vol%
O2 range
(Sample gas pressure: within 60.49 kPa)
65% Maximum value of range setting ; from 0 to 50 vol% O2 to 0 to 100 vol%
O2 range (Sample gas pressure: within 60.49 kPa)
Drift: (Excluding the first two weeks in use)
(Excluding the case where the reference air flow is natural convection.)
Both zero and span 62% Maximum value of range setting/month
Response Time : Response of 90% within 5 seconds. (Measured from when gas is
introduced from calibration gas inlet and analog output start changing.)
2-2
IM 11M12D01-01E
2.2General-use Separate-type Detector and Related
Equipment
General-use separate-type detector ZR22G can be used in combination with the probe
protector ZO21R-L (see Section 2.2.2).
2.2.1ZR22G General-use Separate-type Detector
Sample Gas Temperature: 0 to 7008C (Probe only)
It is necessary to mount the cell using Inconel cell-bolts when the temperature
is 6008C or greater.
700 to 14008C (with High Temperature Probe Adapter)
For high-temperature sample gas, apply 0.15 m length probe and High tempera
ture Probe Adapter ZO21P-H.
Sample Gas Pressure: -5 to +250 kPa (When the pressure in the furnace exceeds 3 kPa,
it is recommended that you compensate for this pressure. When the pressure in
the furnace exceeds 5 kPa, you must perform pressure compensation.)
For 0.15m probe, 0.5 to 5 kPa.
2. Specifications
No pressure fluctuation in the furnace should be allowed.
Note: When the detector is used in conjunction with a check valve and a ZA8F
Flow Setting Unit, the maximum pressure of sample gas is 150 kPa. When with
a check valve and a ZR40H Auto Calibration Unit, it is 200 kPa. If the pressure
of your sample gas exceeds these limits, consult with Yokogawa.
Probe Length: 0.15, 0.4, 0.7, 1.0, 1.5, 2.0, 2.5, 3.0, 3.6, 4.2, 4.8, 5.4 m
Probe Material: SUS 316 (JIS)
Ambient Temperature: -20 to +1508C
Reference Air System: Natural Convection, Instrument Air, Pressure Compensation
(other than for probe length 0.15 m)
Instrument Air System (excluding Natural Convection) :
Pressure; 200 kPa + the pressure inside the furnace (It is recommended to use
air which has been dehumidified by cooling to dew point -208C or less, and had
dust or oil mist removed.)
Consumption; Approx. 1 Nl/min
Material in Contact with Gas: SUS 316 (JIS), Zirconia, SUS 304 (JIS) (flange),
Cover; Moss green (Munsell 2.5Y8.4/1.2)
Finish: Polyurethane corrosion-resistant coating
Gas Connection:Rc1/4 or 1/4FNPT
Wiring Connection: G1/2, Pg13.5, M20 by 1.5 mm, 1/2 NPT
Installation:Flange mounting
Probe Mounting Angle:Horizontal to vertically downward.
When the probe insertion length is 2 m or less, can install at angles from
horizontal to vertically downward.
When the probe insertion length is 2.5 or more, mount vertically downward
(within 658), and if using a probe protector install at angles between horizontal
and vertically downward (within 658).
Weight:
Insertion length of 0.4 m:approx. 6 kg (JIS 5K 65) / approx. 11 kg (ANSI 150 4)
Insertion length of 1.0 m:approx. 8 kg (JIS 5K 65)) / approx. 13 kg (ANSI 150 4)
Insertion length of 1.5 m:approx. 10 kg (JIS 5K 65)) / approx. 15 kg (ANSI 150 4)
Insertion length of 2.0 m:approx. 12 kg (JIS 5K 65)) / approx. 17 kg (ANSI 150 4)
Insertion length of 3.0 m:approx. 15 kg (JIS 5K 65)) / approx. 20 kg (ANSI 150 4)
Insertion length of 3.6 m:approx. 17 kg (JIS 5K 65)) / approx. 22 kg (ANSI 150 4)
Insertion length of 4.2 m:approx. 19 kg (JIS 5K 65)) / approx. 24 kg (ANSI 150 4)
Insertion length of 4.8 m:approx. 21 kg (JIS 5K 65)) / approx. 26 kg (ANSI 150 4)
Insertion length of 5.4 m:approx. 23 kg (JIS 5K 65)) / approx. 28 kg (ANSI 150 4)
2-4
IM 11M12D01-01E
d Model and Codes
n
Model
ZR22G
Length
-015
-040
-070
-100
-150
-200
-250
-300
-360
-420
-480
-540
Wetted material
-S
-C
Flange
(*3)
Reference air-C
Gas Thread-R
Connection box thread
Instruction manual
Options
Suffix code
-A
-B
-C
-E
-F
-G
-K
-L
-M
-P
-Q
-R
-S
-W
-E
-P
Valves
Filter
Tag plates
2. Specifications
Style: S2
Option
code
Separate type Zirconia Oxygen/ High Temperature Humidity Analyzer, Detector
0.15 m (for high temperature use)(*1)
0.4 m
0.7 m
1.0 m
1.5 m
2.0 m
2.5 m(*2)
3.0 m (*2)
3.6 m(*2)
4.2 m(*2)
4.8 m(*2)
5.4 m(*2)
SUS316
Stainless steel with Inconel calibration gas tube(*11)
ANSI Class 150 2 RF SUS304
ANSI Class 150 3 RF SUS304
ANSI Class 150 4 RF SUS304
DIN PN10 DN50 A SUS304
DIN PN10 DN80 A SUS304
DIN PN10 DN100 A SUS304
JIS 5K 65 FF SUS304
JIS 10K 65 FF SUS304
JIS 10K 80 FF SUS304
JIS 10K 100 FF SUS304
JIS 5K 32 FF SUS304 (for high temperature use) (*4)
JPI Class 150 4 RF SUS304
JPI Class 150 3 RF SUS304
Westinghouse
Natural convection
External connection (Instrument air) (*12)
Pressure compensated(*12)
Rc 1/4
-T
-P
-G
-M
-T
-Q
-J
-E
-A
1/4 FNPT
G1/2
Pg13.5
M20 x1.5 mm
1/2NPT
Quick connect(*9)
Japanese
English
Always -A
DERAKANE coating (*10)/D
Inconel bolt(*5)/C
/CV
/SV
/F1
/F2
/SCT
/PT
Check valve(*6)
Stop valve(*6)
Dust Filter(*7)
Dust Guard Protector (*7)
Stainless steel tag plate(*8)
Printed tag plate(*8)
Description
*1 Used with the ZO21P High Temperature Probe Adapter. Select flange (-Q).
*2 When installing horizontally the probe whose insertion length is 2.5 meters or more, use the Probe Protector. Be sure to specify
ZO21R-L-200-h. Specify the flange suffix code either -C or -K.
*3 The thickness of the flange depends on its dimensions.
*4 Not used in conjunction with —P (pressure compensation) for reference air. The flange thickness does not conform to JIS specificatio
*5 Inconel probe bolts and U shape pipe are used. Use this option for high temperature use (ranging from 600 to 700 8C).
*6 Specify either /CV or /SV option code.
*7 Not used with the high temperature humidity analyzer.
*8 Specify either /SCT or /PT option code.
*9 Not waterproof, avoid rain. Operating maximum temperature is 808C. Available only in the U.S.
*10 Available only in the U.S. DERAKANE is a registered trademark of the Dow Chemical Company.
*11 Recommended if measured gas contains corrosive gas like chlorine.
*12 Piping for reference air must be installed to supply reference air constantly at a specified flow rate.
IM 11M12D01-01E
T2.1.EPS
2-5
EXTERNAL DIMENSIONS
1. Model ZR22G Separate type Zirconia Oxygen Analyzer, Detectors
[50.8
L=0.15, 0.4, 0.7, 1.0,
1.5, 2.0, 2.5, 3.0
3.6, 4.2, 4.8, 5.4 (m)
L
t
Flange
283 to 292
155 to 16369
25
Rc1/4 or 1/4NPT
Calibration gas inlet
[124
85
Rc1/4 or 1/4NPT
Reference air inlet
C
2-G1/2,2-1/2NPT etc.
Cable connection port
[A
48
[B
Flange
Flange
ANSI Class 150 2 RF SUS304
ANSI Class 150 3 RF SUS304
ANSI Class 150 4 RF SUS304
DIN PN10 DN50 A SUS304
DIN PN10 DN80 A SUS304
DIN PN10 DN100 A SUS304
JIS 5K 65 FF SUS304
JIS 10K 65 FF SUS304
JIS 10K 80 FF SUS304
JIS 10K 100 FF SUS304
JIS 5K 32 FF SUS304
JPI Class 150 4 RF SUS304
JPI Class 150 3 RF SUS304
Westinghouse
Model ZR22G...-P (with pressure compensation) Separate type Zirconia Oxygen Analyzer, Detectors
L
[50.8
L=0.4, 0.7, 1.0, 1.5,
2.0, 2.5, 3.0, 3.6,
4.2, 4.8, 5.4 (m)
L
Reference gas outlet
Flange
ANSI Class 150 2 RF SUS304
ANSI Class 150 3 RF SUS304
ANSI Class 150 4 RF SUS304
DIN PN10 DN50 A SUS304
DIN PN10 DN80 A SUS304
DIN PN10 DN100 A SUS304
JIS 5K 65 FF SUS304
JIS 10K 65 FF SUS304
JIS 10K 80 FF SUS304
JIS 10K 100 FF SUS304
JPI Class 150 4 RF SUS304
JPI Class 150 3 RF SUS304
Westinghouse
This probe protector is required for the general-use detector when it is used for oxygen
concentration measurements in powdered coal boilers or in fluidized furnaces to prevent
abrasion due to dust particles when gas flow exceeds 10 m/s.
When using a ZR22G general-use separate-type detector in the horizontal position, be
sure to select a probe protector ZO21R-L-hhh-h*B to support the probe.
Insertion Length: 1.05 m, 1.55 m, 2.05 m.
Flange: JIS 5K 65A FF SUS304. ANSI Class 150 4 FF (without serration). However,
flange thickness is different.
Material: SUS316 (JIS), SUS304 (JIS) (Flange)
Weight: 1.05 m; Approx. 6/10 kg (JIS/ANSI), 1.55 m; Approx. 9/13 kg (JIS/ANSI),
2.05 m; Approx. 12/16 kg (JIS/ANSI)
Installation: Bolts, nuts, and washers are provided for detector, probe adapter and
process-side flange.
ModelSuffix code
ZO21R
Insertion
length
Flange ( *1)
*1 Thickness of flange depends on dimensions of flange.
Gas flow
-L
-100
-150
-200
-J
-A
Style code
[60.5
l (Insertion length)
l=1050,1550,2050
Option
code
*B
Washer (12)
Mounting nut (M12)
SUS316
Probe Protector (0 to 700 8C)
1.05 m (3.5 ft)
1.55 m (5.1 ft)
2.05 m (6.8 ft)
JIS 5K 65 FF SUS304
ANSI Class 150 4 FF SUS304
Style B
Description
D
t
T2.2E.EPS
Flange <1>
(with bolts, nuts,
gasket and washer)
gasket (t1.5)
[A
[B
[B
Unit ; mm
2-8
Flange<1>
JIS 5K 65 FF
SUS304
ANSI Class
150 4 FF SUS304
A
155
228.6
Dimensions of
holes on opposing surface
Ct
B
130
190.5
4 -[15
8 -[19
5
12
C
D
40
50
F2.3E.EPS
IM 11M12D01-01E
2. Specifications
2.3High-Temperature Separate-type Detector and Related
Equipment
Standard Specifications
Construction: Water-resistant, non-explosionproof
Probe length: 0.15 m
Terminal box: Aluminium alloy
Probe material: Probe material in contact with gas: SUS 316 (JIS) (Probe), SUS 304
(JIS) (Flange), Zirconia (Sensor), Hastelloy B, (Inconel 600, 601)
Weight: Approx. 3 kg
Installation: Flange mounting (The use of high-temperature detector probe adapter
ZO21P-H is necessary.)
Flange standard: JIS 5 K 32 FF equivalent (thickness varies)
Mounting angle: Any angle between horizontal and vertical (high-temperature probe is
fitted with an adapter) Reference gas and calibration gas piping
connection:Rc 1/4 or 1/4 NPT female
Cable inlet: G 1/2, Pg 13.5, M20 3 15, 1/2 NPT
Ambient temperature: -20 to 1508C
Sample gas temperature: 0 to 7008C (temperature at the measuring point of the sam-
pling gas. 0 to 7508C or 0 to 14008C when the probe adapter for high
temperature is used.
Temperature of the probe adapter shall not exceed 3008C to protect the
gasket and avoid the bolts seizing together.
Sample gas pressure: -0.5 to +5 kPa: when used at the range of more than 0 to
25 vol% O2, -0.5 to +0.5 kPa. (An auxiliary ejector is required for
negative pressure application.)
Model and Code
Refer to “Model and Code” in page 2-5.
External Dimensions
Refer to the Figure in page 2-6.
IM 11M12D01-01E
2-9
2.3.2ZO21P-H Adapter for High-Temperature Probe
The probe adapter is used to lower the sample gas to a temperature below 7008C (below
3008C at probe adapter surface) before it is fed to the detector.
Insertion length: 1 m, 1.5 m
Material in Contact with Gas:
SUS 316 (JIS), Zirconia, SiC or SUS 310S, SUS 304(JIS) (flange)
Probe Material: SiC, SUS 310S (JIS)
Installation:Flange mounting (FF type or RF type)
Probe Mounting Angle: Vertically downward within 658
Where the probe material is SUS 310S, horizontal mounting is available.
Construction: Non explosion-proof. Rainproof construction
Weight:Insertion length of 1.0 m: approx. 6.5 kg (JIS) / approx. 8.5 kg (ANSI)
Insertion length of 1.5 m: approx. 7.5 kg (JIS) / approx. 9.5 kg (ANSI)
Sample gas temperature:0 to 14008C (when using SiC probe) 0 to 8008C (when
using SUS 310S probe adapter)
Sample gas pressure: -0.5 to + 5 kPa. When using in the range of 0 to 25 vol% O2 or
more, the sample gas pressure should be in the range of -0.5 to +0.5 kPa.
(Where the sample gas pressure for the high-temperature probe is
negative, an auxiliary ejector is necessary.)
d Model and Codes
ModelSuffix code
ZO21P
Material
Insertion
length
Flange
Style code
.......................................
-H
.....................................
-A
.....................................
-B
...............................
-100
...............................
-150
-J
-N
-M
-L
-A
-R
-Q
-T
-S
-E
Option
code
.............................
............................
............................
............................
............................
............................
............................
............................
............................
............................
......................
*A
Description
High Temperature Probe Adapter
SiC
SUS 310S
1.0 m
1.5 m
JIS 5K 50 FF SUS304
JIS 10K 65 FF SUS304
JIS 10K 80 FF SUS304
JIS 10K 100 FF SUS304
ANSI Class 150 4 RF SUS304
ANSI Class 150 2 1/2 RF SUS304
ANSI Class 150 3 RF SUS304
JPI Class 150 3 RF SUS304
JPI Class 150 4 RF SUS304
DIN PN10 DN50 A SUS304
Style A
T2.3E.EPS
2-10
IM 11M12D01-01E
External Dimensions
Measured gas outlet
R1/2 (Note2)
Approx. 351
180
f60.5
2. Specifications
Unit: mm
Flange (Thickness)
JIS 5K 32 FF
Gasket (Thickness 3)
f60.5
170
Approx, 215
Approx. 48
Approx, 100
(Insertion length) (Note1)
f30
[A
[B
fA
f52 over
f115
Flange <1>
Flange provided
by customer
High temperature
Probe SiC pipe
C
110
f124
85
Detector (ZR22G)
69
Reference air inlet <2>
25
Pipe hole (2-G1/2,2-1/2NPT.etc)
Calibration gas inlet <3>
(Note 1) 1=1000 or 1500 (mm)
(Note 2) Sample gas outlet
(if the sample gas pressure is negative,
connect the auxiliary ejector.)
42
IM 11M12D01-01E
<1> Flange
JIS 5K 50 FF SUS304130
ANSI Class 150 4 RF SUS304
<2>,<3> joint
Rc 1/4
1/4 FNPT
228.6
AB
105
190.5
C
4-[15
8-[19
F2.4E.EPS
2-11
Loading...
+ 204 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.