Yokogawa ZR202 User Manual

User’s
R
Manual
Model ZR202G Integrated type Zirconia Oxygen Analyzer
IM 11M12A01-04E
7th Edition

Introduction

The EXAxt ZR Integrated type Zirconia Oxygen Analyzer has been developed for combustion control in various industrial processes. There are several version of this analyzer so you can select one that matches your application.
Optional accessories are also available to improve measurement accuracy and automate calibration. An optimal control system can be realized by adding appropriate options.
This instruction manual describes almost all of the equipment related to the EXAxt ZR. You may skip any section(s) regarding equipment which is not included in your system.
Regarding the HART Communication Protocol, refer to IM 11M12A01-51E.
IM11M12A01-51E has been published as ''Model EXAxt ZR series HART protocol''.
Regarding Separate type Zirconia Oxygen Analyzer, refer to IM 11M12A01-02E.
< Before using the equipment, please read any descriptions in this manual related to the equipment and system that you have, on appropriate use and operation of the EXAxt ZR. >
Models and descriptions in this manual are listed below.
<Introduction>
i
Model
ZR202G Integrated type Oxygen Analyzer
ZO21R Probe Protector
ZA8F Flow Setting Unit (for manual
ZR20H Automatic Calibration Unit
Case Assembly for calibration
Check valve (Part No. K9292DN,
Dust Filter for the detector (Part
Dust Guard Protector (Part No.
ZO21S Standard Gas Unit
CMPL: Customer Maintenance Parts List
Product Name
calibration use)
gas cylinder (Part No. E7044KF)
K9292DS)
No. K9471UA)
K9471UC)
Description in this manual
Specication Installation Operation Maintenance CMPL
Media No.IM 11M12A01-04E (CD) 7th Edition :March 2012 (YK) All Rights Reserved Copyright © 2000, Yokogawa Electric Corporation
IM 11M12A01-04E
ii
IM 11M12A01-04E
<Introduction>
This manual consists of twelve chapters. Please refer to the reference chapters for installation, opera­tion and maintenance.
Chapter
1. Overview Equipment models and system conguration examples
2. Specications Standard specication, model code (or part number),
3. Installation Installation method for each equipment
4. Piping Examples of piping in three standard system
5. Wiring Wiring procedures such as “Power supply wiring”,
6. Components Major parts and functions are described
7. Startup Basic procedure to start operation of EXAxt ZR.
8. Detailed Data
Setting
9. Calibration Describes the calibration procedure required in the
10. Other
Functions
11. Inspection
and Maintenance
12. Troubleshoot-
ing
CMPL (parts list) User replaceable parts list
: Read and completely understand before operating the equipment. : Read before operating the equipment, and refer to it whenever necessary. : Recommended to read at least once.
dimension drawing for each equipment
congurations
“output signal wiring” or others
Chapter 7enables you to operate the equipment immediately.
Details of key operations and displays
course of operation.
Other functions described
How to conduct maintenance of EXAxt ZR and procedures for replacement of deteriorated parts
This chapter describes measures to be taken when an abnormal condition occurs.
Outline
Relates to
Installation Operation Maintenance

For the safe use of this equipment

WARNING
CAUTION
EXAxt ZR is very heavy. Be sure not to accidentally drop it. Handle safely to avoid injury.
Connect the power supply cord only after conrming that the supply voltage matches the rating of this equipment. In addition, conrm that the power is switched off when connecting power supply.
Some process gas is dangerous to people. When removing this equipment from the process line for maintenance or other reasons, protect yourself from potential poisoning by using a protective mask or ventilating the area well
The cell (sensor) at the tip of the probe is made of ceramic (zirconia element). Do not drop the equip­ment or subject it to pressure stress.
<Introduction>
iii
• Do NOT allow the sensor (probe tip) to make contact with anything when installing the analyzer.
• Avoid water dropping directly on the probe (sensor) of the analyzer when installing it.
• Check the calibration gas piping before introducing the calibration gas to ensure that there is no leakage of the gas. If there is any leakage of the gas, the moisture drawn from the sample gas may damage the sensor.
• The probe (especially at the tip) becomes very hot. Be sure to handle it with gloves.
(1) About This Manual
The contents of this manual shall not be reproduced or copied, in part or in whole, without per-
mission.
This manual explains the functions contained in this product, but does not warrant that those will
suit the particular purpose of the user.
Every effort has been made to ensure accuracy in the preparation of this manual. However,
should any errors or omissions come to the attention of the user, please contact the nearest
Yokogawa Electric representative or sales ofce.
This manual does not cover the special specications. This manual may not be changed on any
change of specication, construction and parts when the change does not affect the functions or
performance of the product.
If the product is used in a manner not specied in this manual, the safety of this product may be
impaired.
IM 11M12A01-04E
iv
IM 11M12A01-04E
<Introduction>
WARNING
CAUTION
IMPORTANT
NOTE
(2) Safety and Modication Precautions
Follow the safety precautions in this manual when using the product to ensure protection and
safety of personnel, product and system containing the product.
The following safety symbols and wordings are used on the product as well as in this manual.
(3) The following safety symbols are used in this manual.
Throughout this user’s manual, you will find several different types of symbols are used to identify different sections of text. This section describes these icons.
Indicates a potentially hazardous situation which, if not avoided, could result in death or serious injury.
Indicates a potentially hazardous situation which, if not avoided, may result in minor or moderate injury. It may also be used to alert against unsafe practices.
Indicates that operating the hardware or software in this manner may damage it or lead to system failure.
Draws attention to information essential for understanding the operation and features.
TIP
Identies additional information.
SEE ALSO
Identies a source to be referred to.
Clicking a reference displayed in green can call up its source, while clicking a reference displayed
in black cannot.
<Introduction>
9
8
7
6
5
4
3
2
1
0
W
V
U
T
S
R
Q
P
O
N
M
L
K
J
I
H
G
F
E
D
C
B
A
Alphanumerics Alphanumerics AlphanumericsLED Display LED Display LED Display
Z
Y
Special descriptions in this manual
This manual indicates operation keys, displays and drawings on the product as follows:
v
• Operation keys, displays on the panel Enclosed in [ ]. (Ex. "MODE" key)
(Ex. message display "BASE") (Ex. data display "102" lit, "102" ashing)
• Drawing for ashing
Indicated by gray characters (Flashing) (lit)
• Displays on the LCD display panel
IM 11M12A01-04E
vi
IM 11M12A01-04E
<Introduction>

NOTICE

• Specication check
When the instrument arrives, unpack the package with care and check that the instrument has not
been damaged during transportation. In addition, please check that the specication matches the order, and required accessories are not missing. Specications can be checked by the model codes on the nameplate. Refer to Chapter 2 specications for the list of model codes.
• Details on operation parameters
When the EXAxt ZR Integrated type Zirconia Oxygen Analyzer arrives at the user site, it will operate based on the operation parameters (initial data) set before shipping from the factory.
Ensure that the initial data is suitable for the operation conditions before conducting analysis.
Where necessary, set the instrument parameters for appropriate operation. For details on setting data, refer to Chapters 7 to 10.
When user changes the operation parameter, it is recommended to note down the changed setting data.
• How to dispose the batteries:
This is an explanation about the new EU Battery Directive (DIRECTIVE 2006/66/EC). This directive is only valid in the EU.
Batteries are included in this product. Batteries incorporated into this product cannot be removed by yourself. Dispose them together with this product.
When you dispose this product in the EU, contact your local Yokogawa Europe B.V.ofce. Do not
dispose them as domestic household waste.
Battery type: silver oxide battery
Notice:
The symbol (see above) means they shall be sorted out and collected as ordained in ANNEX II in DIRECTIVE 2006/66/EC

After-Sales Warranty

Do not modify the product.
During the warranty period, for repair under warranty carry or send the product to the local sales
representative or service ofce. Yokogawa will replace or repair any damaged parts and return
the product to you.
Before returning a product for repair under warranty, provide us with the model name and serial
number and a description of the problem. Any diagrams or data explaining the problem would also be appreciated.
If we replace the product with a new one, we won’t provide you with a repair report.
Yokogawa warrants the product for the period stated in the pre-purchase quotation. Yokogawa
shall conduct dened warranty service based on its standard. When the customer site is located
outside of the service area, a fee for dispatching the maintenance engineer will be charged to the customer.
In the following cases, customer will be charged repair fee regardless of warranty period.
• Failure of components which are out of scope of warranty stated in instruction manual.
• Failure caused by usage of software, hardware or auxiliary equipment, which Yokogawa Elec-
tric did not supply.
<Introduction>
vii
• Failure due to improper or insufcient maintenance by user.
• Failure due to modication, misuse or outside-of-specications operation which Yokogawa
does not authorize.
• Failure due to power supply (voltage, frequency) being outside specications or abnormal.
• Failure caused by any usage out of scope of recommended usage.
• Any damage from re, earthquake, storms and oods, lightning, disturbances, riots, warfare,
radiation and other natural changes.
Yokogawa does not warrant conformance with the specic application at the user site. Yokogawa
will not bear direct/indirect responsibility for damage due to a specic application.
Yokogawa Electric will not bear responsibility when the user congures the product into systems
or resells the product.
Maintenance service and supplying repair parts will be covered for ve years after the production
ends. For repair for this product, please contact the nearest sales ofce described in this instruc-
tion manual.
IM 11M12A01-04E
viii
<Introduction>
IM 11M12A01-04E
<CONTENTS>
Model ZR202G Integrated type Zirconia Oxygen Analyzer
IM 11M12A01-04E 7th Edition

CONTENTS

Introduction ..............................................................................................................i
For the safe use of this equipment .............................................................................. iii
NOTICE ........................................................................................................................... vi
After-Sales Warranty .................................................................................................... vii
1. Overview .................................................................................................... 1-1
1.1 <EXAxtZR>SystemConguration ............................................................... 1-1
1.1.1 System 1 ............................................................................................
1.1.2 System 2 ............................................................................................
1.1.3 System 3 ............................................................................................
1.2 < EXAxt ZR > System Components ................................................................
1.2.1 System Components .........................................................................
1.2.2 Oxygen Analyzer and Accessories ....................................................
Toc-1
1-1
1-2
1-3
1-4
1-4
1-4
2. Specications ........................................................................................... 2-1
2.1 GeneralSpecications ..................................................................................... 2-1
2.1.1 StandardSpecications .....................................................................
2.1.2 ZR202G Integrated type Zirconia Oxygen Analyzer ..........................
2.1.3 ZO21R Probe Protector .....................................................................
2.2 ZA8F Flow Setting Unit and ZR20H Automatic Calibration Unit ................
2.2.1 ZA8F Flow Setting Unit ....................................................................
2.2.2 ZR20H Automatic Calibration Unit ...................................................
2.3 ZO21S Standard Gas Unit ..............................................................................
2.4 Other Equipment .............................................................................................
2.4.1 Dust Filter for Oxygen Analyzer (part no. K9471UA) .......................
2.4.2 Dust Guard Protector (part no. K9471UC) ......................................
2.4.3 Stop Valve (part no. L9852CB or G7016XH) ...................................
2.4.4 Check Valve (part no. K9292DN or K9292DS) ................................
2.4.5 Air Set ...............................................................................................
2.4.6 Zero Gas Cylinder (part no. G7001ZC) ...........................................
2.4.7 Pressure Regulator (G7013XF or G7014XF) for Gas Cylinder .......
2.4.8 Case Assembly (E7044KF) for Calibration gas Cylinder .................
2.4.9 ZR202A Heater Assembly ...............................................................
2-1
2-2
2-9
2-10
2-10
2-12
2-14
2-15
2-15
2-15
2-16
2-16
2-17
2-18
2-18
2-19
2-19
IM 11M12A01-04E
Toc-2
<CONTENTS>
3. Installation ................................................................................................. 3-1
3.1 Installation of ZR202G Zirconia Oxygen Analyzer ........................................3-1
3.1.1 Installation Location ...........................................................................
3.1.2 Probe Insertion Hole ..........................................................................
3.1.3 Installation of the Probe .....................................................................
3.1.4 Installation of the Dust Filter (K9471UA), Dust Guard Protector
(K9471UC) Probe Protector (ZO21R) ...............................................3-3
3.2 Installation of ZA8F Flow Setting Unit ............................................................
3.2.1 Installation Location ...........................................................................
3.2.2 Mounting of ZA8F Flow Setting Unit ..................................................
3.3 Installation of ZR20H Automatic Calibration Unit .........................................
3.3.1 Installation Location ...........................................................................
3.3.2 Mounting of ZR20H Automatic Calibration Unit .................................
3.4 Installation of the Case Assembly (E7044KF) for
Calibration Gas Cylinder ..................................................................................3-9
3.4.1 Installation Location ...........................................................................
3.4.2 Mounting ............................................................................................
3.5 Insulation Resistance Test .............................................................................
3-1
3-2
3-2
3-5
3-5
3-6
3-7
3-7
3-7
3-9
3-9
3-10
4. Piping ......................................................................................................... 4-1
4.1 Piping for System 1 ...........................................................................................4-1
4.1.1 Piping Parts for System 1 ..................................................................
4.1.2 Piping for the Calibration Gas ............................................................
4.1.3 Piping for the Reference Gas
4.2 Piping for System 2 ...........................................................................................
4.2.1 Piping Parts for System 2 ..................................................................
4.2.2 Piping for the Calibration Gas ............................................................
4.2.3 Piping for the Reference Gas
4.3 Piping for System 3 ...........................................................................................
4.4 Piping for the Oxygen Analyzer with Pressure Compensation ...................
4.4.1 Piping Parts for Oxygen Analyzer with Pressure Compensation ......
4.4.2 Piping for the Calibration Gas ............................................................
4.4.3 Piping for the Reference Gas
.............................................................4-2
.............................................................4-4
.............................................................4-8
4-2
4-2
4-3
4-3
4-3
4-4
4-6
4-8
4-8
5. Wiring ......................................................................................................... 5-1
5.1 General ...............................................................................................................5-1
5.1.1 Terminals for the External Wiring .......................................................
5.1.2 Wiring .................................................................................................
5.1.3 Mounting of Cable Gland ...................................................................
5.2 Wiring for Analog Output .................................................................................
5.2.1 Cable Specications ..........................................................................
5.2.2 Wiring Procedure ...............................................................................
5-2
5-2
5-3
5-3
5-3
5-3
IM 11M12A01-04E
<CONTENTS>
5.3 Wiring Power and Ground Terminals .............................................................5-4
5.3.1 Wiring for Power Line .........................................................................
5.3.2 Wiring for Ground Terminals ..............................................................
5.4 Wiring for Contact Output ................................................................................
5.4.1 Cable Specications ..........................................................................
5.4.2 Wiring Procedure ...............................................................................
5.5 Wiring for Contact Input ...................................................................................
5.5.1 Cable Specications ..........................................................................
5.5.2 Wiring Procedure ...............................................................................
Toc-3
5-4
5-4
5-5
5-5
5-5
5-5
5-5
5-6
6. Components ............................................................................................. 6-1
6.1 ZR202G Zirconia Oxygen Analyzer .................................................................6-1
6.1.1 Integrated type Zirconia Oxygen Analyzer ........................................
6.2 ZA8F Flow Setting Unit, ZR20H Automatic Calibration Unit ........................
6-1
6-2
7. Startup ....................................................................................................... 7-1
7.1 Checking Piping and Wiring Connections .....................................................7-2
7.2 Valve Setup ........................................................................................................
7.3 Supplying Power to Converter ........................................................................
7.4 Operation of Infrared Switch ...........................................................................
7.4.1 Display and Switches .........................................................................
7.4.2 Display Conguration .........................................................................
7.4.3 Entering Parameter Code Selection Display .....................................
7.4.4 Selecting Parameter Codes ...............................................................
7.4.5 Changing Set Values .........................................................................
7.5 Conrmation of Equipment Type Setting .......................................................
7.6 Selection of Measurement Gas .......................................................................
7.7 Output Range Setting .......................................................................................
7.7.1 Minimum Current (4 mA) and Maximum Current (20 mA) Settings ..
7.8 Checking Current Loop ..................................................................................
7.9 Checking Contact I/O ......................................................................................
7.9.1 Contact Output Check .....................................................................
7.9.2 Checking Calibration Contact Output ..............................................
7.9.3 Checking Input Contacts ..................................................................
7.10 Calibration .......................................................................................................
7.10.1 Calibration Setup .............................................................................
7.10.2 Manual Calibration ...........................................................................
7-2
7-2
7-3
7-3
7-4
7-5
7-6
7-6
7-8
7-9
7-9
7-9
7-11
7-11
7-12
7-13
7-14
7-15
7-15
7-17
IM 11M12A01-04E
Toc-4
<CONTENTS>
8. Detailed Data Setting ............................................................................... 8-1
8.1 Setting Display Item ..........................................................................................8-1
8.2 Current Output Setting .....................................................................................
8.2.1 Setting Minimum Oxygen Concentration ( at 4 mA) and Maximum
Oxygen Concentration ( at 20 mA) ....................................................8-2
8.2.2 Entering Output Damping Constants .................................................
8.2.3 Selection of Output Mode ..................................................................
8.2.4 Default Values ....................................................................................
8.3 Output Hold Setting ..........................................................................................
8.3.1 Denition of Equipment Status ..........................................................
8.3.2 Preference Order of Output Hold Value .............................................
8.3.3 Output Hold Setting ............................................................................
8.3.4 Default Values ....................................................................................
8.4 Setting Oxygen Concentration Alarms
8.4.1 Alarm Values ......................................................................................
8.4.2 Alarm Output Actions .........................................................................
8.4.3 Alarm Setting
8.4.4 Default Values ....................................................................................
8.5 Output Contact Setup .......................................................................................
8.5.1 Output Contact ...................................................................................
8.5.2 Setting Output Contact ......................................................................
8.5.3 Default Values ..................................................................................
8.6 Input Contact Settings ....................................................................................
8.6.1 Setting Input Contact .......................................................................
8.6.2 Default Values ..................................................................................
8.7 Other Settings .................................................................................................
8.7.1 Setting the Date-and-Time ...............................................................
8.7.2 Setting Periods over which Average Values are Calculated and Periods
over which Maximum and Minimum Values Are Monitored.............8-13
8.7.3 Setting Fuels ....................................................................................
8.7.4 Setting Purging ................................................................................
......................................................................................8-7
...........................................................8-6
8-1
8-2
8-2
8-2
8-3
8-3
8-5
8-5
8-5
8-6
8-6
8-8
8-8
8-8
8-9
8-10
8-11
8-11
8-11
8-12
8-12
8-14
8-18
9. Calibration ................................................................................................. 9-1
IM 11M12A01-04E
9.1 Calibration Briefs ..............................................................................................9-1
9.1.1 Principle of Measurement ..................................................................
9.1.2 Calibration Gas ..................................................................................
9.1.3 Compensation ....................................................................................
9.1.4 Characteristic Data from a Sensor Measured During Calibration .....
9.2 Calibration Procedures ....................................................................................
9.2.1 Calibration Setting ..............................................................................
9.2.2 Calibration ..........................................................................................
9-1
9-2
9-2
9-4
9-5
9-5
9-8
<CONTENTS>
Toc-5
10. Other Functions ...................................................................................... 10-1
10.1 Detailed Display ..............................................................................................10-1
10.1.1 Air Ratio
10.1.2 Cell Temperature ..............................................................................
10.1.3 C. J. Temperature
10.1.4 Amount of Water Vapor in Exhaust Gas ..........................................
10.1.5 Cell Voltage ......................................................................................
10.1.6 Thermocouple Voltage .....................................................................
10.1.7 Cold Junction Voltage ......................................................................
10.1.8 Current Output .................................................................................
10.1.9 Response Time ................................................................................
10.1.10 Cell’s Internal Resistance ...............................................................
10.1.11 Robustness of a Cell .......................................................................
10.1.12 Heater On-Time Ratio .....................................................................
10.1.13 Oxygen Concentration (with time constant) ....................................
10.1.14 Maximum Oxygen Concentration ...................................................
10.1.15 Minimum Oxygen Concentration ....................................................
10.1.16 Average Oxygen Concentration ......................................................
10.1.17 Span and Zero Correction Ratios ...................................................
10.1.18 History of Calibration Time ..............................................................
10.1.19 Time .................................................................................................
10.1.20 Software Revision ...........................................................................
10.2 Operational Data Initialization ......................................................................
10.3 Initialization Procedure ..................................................................................
10.4 Reset .................................................................................................................
10.5 Handling of the ZO21S Standard Gas Unit .................................................
10.5.1 Standard Gas Unit Component Identication ................................
10.5.2 Installing Gas Cylinders .................................................................
10.5.3 Calibration Gas Flow ......................................................................
10.6 Methods of Operating Valves in the ZA8F Flow Setting Unit ...................
10.6.1 Preparation Before Calibration ......................................................
10.6.2 Operating the Span Gas Flow Setting Valve .................................
10.6.3 Operating the Zero Gas Flow Setting Valve ..................................
10.6.4 Treatment After Calibration ............................................................
............................................................................................10-3
10-3
.............................................................................10-3
10-3
10-4
10-4
10-4
10-4
10-5
10-5
10-5
10-6
10-6
10-6
10-6
10-6
10-6
10-7
10-7
10-7
10-7
10-8
10-9
10-16
10-16
10-16
10-17
10-19
10-19
10-19
10-19
10-20
11. Inspection and Maintenance ................................................................. 11-1
11.1 Inspection and Maintenance of the Detector ............................................... 11-1
11.1.1 Cleaning the Calibration Gas Tube ..................................................
11.1.2 Replacing the Sensor Assembly ......................................................
11.1.3 Replacement of the Heater Assembly .............................................
11.1.4 Replacement of Dust Filter ..............................................................
11.1.5 Replacement of O-ring .....................................................................
11.1.6 Stopping and Re-starting Operation ................................................
11-1
11-2
11-4
11-6
11-6
11-6
IM 11M12A01-04E
Toc-6
<CONTENTS>
11.2 Inspection and Maintenance of the Converter ............................................ 11-8
11.2.1 Replacing Fuses ..............................................................................
11.3 Replacement of Flowmeter for ZR20H Automatic Calibration Unit ........
11-8
11-10
12. Troubleshooting ..................................................................................... 12-1
12.1 Displays and Measures to Take When Errors Occur .................................. 12-1
12.1.1 What is an Error? .............................................................................
12.1.2 Measures to Take When an Error Occurs .......................................
12.2 Displays and Measures to Take When Alarms are Generated ...................
12.2.1 What is an Alarm? ............................................................................
12.3 Measures When Measured Value Shows an Error ......................................
12.3.1 Measured Value Higher Than True Value ........................................
12.3.2 Measured Value Lower Than True Value ......................................
12.3.3 Measurements Sometimes Show Abnormal Values .....................
12-1
12-2
12-4
12-4
12-9
12-9
12-10
12-10
Customer Maintenance Parts List ........................................CMPL 11M12A01- 04E
Customer Maintenance Parts List ........................................CMPL 11M12A01-12E
Customer Maintenance Parts List ...........................................CMPL 11M3D1- 01E
Revision Information ...............................................................................................i
IM 11M12A01-04E
<1.Overview>
~
ZR202G Integrated type Zirconia Oxygen Analyzer
ZO21S Standard gas unit
Stop valve
Calibration gas
F1.1E.ai
~
100 to 240 V AC
100/110/115 200/220/240 V AC
Contact input Analog output, contact output
Digital output (HART)

1. Overview

The EXAxt ZR Integrated type Zirconia Oxygen Analyzer is used to monitor and control the oxygen concentration in combustion gases, in boilers and industrial furnaces, for wide application in industries which consume considerable energy-such as steel, electric power, oil and petrochemical, ceramics, pulp and paper, food, or textiles, as well as incinerators and medium/small boilers. It can help conserve energy in these industries. The EXAxt ZR also contributes to preservation of the earth’s environment in preventing global warming and air pollution by controlling complete combustion to reduce CO2, SOx and NOx.
The EXAxt ZR Integrated type Zirconia Oxygen Analyzer integrates both probe and converter. The
analyzers need not use a sampling device, and allow direct installation of the probe in the wall of a ue
or furnace to measure the concentration of oxygen in the stack gas of the temperature up to 700°C.
The probe uses a high-reliability Zirconia sensor and a heater assembly that can be replaced on site.
The analyzer is equipped with three infrared switches, which enable the user to operate the equipment without opening the cover on site. Analyzer calibration can also be fully automated and the automatic calibration unit is provided. Choose the equipment which best suits your needs so that an optimal combustion control system can be obtained.
Some examples of typical system conguration are illustrated below:
1-1
1.1 <EXAxtZR>SystemConguration
The system configuration should be determined by the conditions; e.g. whether the calibration is to be automated, and whether flammable gas is present and requires safety precautions. The system
conguration can be classied into three basic patterns as follows:

1.1.1 System 1

This is the simplest system consisting of an integrated type analyzer. This system can be implemented for monitoring oxygen concentration in the combustion gases boiler. No piping is required for the reference gas (air) which is fed in at the installation site. The ZO21S standard gas unit is used for calibration.
Zero gas from this unit and span gas (air) is sent to the probe through a tube which is connected during calibration.
Figure1.1 ExampleofSystem1
IM 11M12A01-04E
<1. Overview>
1-2
IM 11M12A01-04E
CAUTION
~
ZR202G Integrated type Zirconia Oxygen Analyzer
F1.2E.ai
ZA8F flow setting unit
Reference gas
Calibration gas
Needle
valve
Flowmeter
Instrument air
Air Set
Calibration
gas pressure
regulator
Zero gas cylinder
Calibration gas unit case
Stop valve
or
Check valve
Span gas(Same as Zero gas calibration)
100 to 240 V AC Contact input
Analog output, contact output Digital output (HART)
• As this system uses ambient air for the reference gas, measuring accuracy will be affected by the installation location.
• A stop valve should be connected to the calibration gas inlet of the equipment. The valve should be fully closed unless calibration is in progress.

1.1.2 System 2

This system is for monitoring and controlling oxygen concentration in the combustion gases of a large-size boiler or heating furnace. Instrument air (clean and dry air of oxygen concentration 21%) is used as the reference gas and the span gas for calibration. Zero gas is fed from a cylinder during
calibration. The gas ow is controlled by the ZA8F ow setting unit (for manual valve operation).
Figure 1.2 Example of System 2
<1.Overview>
~
100 to 240 V AC
Automatic calibration unit
Reference gas
Calibration gas (Zero)
Contact input Analog output, contact output
Digital output (HART)
Air Set
Instrument air
Calibration gas unit case
Calibration
gas pressure
regulator
Zero gas cylinder
ZR202G Integrated type Zirconia Oxygen Analyzer with automatic calibration unit (ZR202G----A-----A)
F1.3E.ai
Span gas
ZR20H
*3
Note: The installation temperature limits range for integrated type analyzer is -20 to 55°C.
*2
*1

1.1.3 System 3

This example, System 3, represents typical applications in large boilers and heating furnaces, where is a need to monitor and control oxygen concentration. The reference gas and calibration-time span gas are (clean, dry) instrument air. Zero gas is supplied from a gas cylinder.
System 3 uses the ZR20H automatic calibration unit, with auto-switching of the calibration gas.
A “combustible gas detected” contact input turns off power to the heater. There’s also contact output from the converter that can be used to operate a purge gas valve to supply air to the sensor.
1-3
*1 Shield cable:
Use shielded signal cables, and connect the shields to the FG terminal of the converter.
*2 Select the desired probe from the Probe Conguration table on page 1-4.
*3 When a zirconia oxygen analyzer is used, 100% N
Use approx. 1 vol% O2 gas (N2-based).
gas cannot be used as the zero gas.
2
Figure 1.3 Example of System 3
IM 11M12A01-04E
1-4
System Components
ZR202G Integrated type Zirconia Oxygen Analyzers ZO21R Probe Protector for Zirconia Oxygen Analyzers
K9471UA Dust Filter for Oxygen Analyzer
ZO21S Standard Gas Unit ZA8F Flow Setting Unit for manual calibration
ZR20H Automatic Calibration Unit for Integrated type Analyzer
L9852CB, G7016XH Stop Valve for Calibration gas line K9292DN,K9292DS Check Valve for Calibration gas line
G7003XF/K9473XK, G7004XF/K9473XG Air Set
G7001ZC Zero gas Cylinder
G7013XF, G7014XF Pressure Regulator for Gas Cylinder
E7044KF Case Assembly for Calibration gas Cylinder
Integrated type
System configuration
Example 3Example 1
Example 2
( )
( )
T1.1.ai
: Items required for the above system example
: To be selected depending on each application. For details, refer to corresponding chapter.
: Select either
( )
ZR202A Heater Assembly (Spare Parts for ZR202G)
K9471UC Dust Guard Protector
F1.4E.ai
Detector (ZR202G)
Detector (ZR202G)
Detector(ZR202G)
Probe Protector
(ZO21R)
Dust Filter
(K9471UA)
or
Dust Guard
Protector
(K9471UC)
+
Sample gas temperature 0 to 700°C
Mounting
Insertion length
General-use Probe
Application
Horizontal
to
vertical
Horizontal
to
vertical
Horizontal
to
vertical
Vertical
Vertical
0.4 to
2 m
0.4
to
2 m
2.5 m
or more
3 m
or less
2.5 m
or more
• Boiler
• Heating furnace
• For pulverized coal boiler with gas flow velocity 10 m/sec or more
• Cement Kiln
• Black liquid recovery boiler
• Cement Kiln
Gas Flow
Sample inlet
<1. Overview>

1.2 < EXAxt ZR > System Components

1.2.1 System Components

1.2.2 Oxygen Analyzer and Accessories

IM 11M12A01-04E
<2. Specications>
2. Specications
This chapter describes the specications for the following:
ZR202G General-use Integrated type Zirconia Oxygen Analyzer (See Section 2.1.2)
ZO21R-L Probe Protector (See Section 2.1.3)
ZA8F Flow Setting Unit (See Section 2.2.1)
ZR20H Automatic Calibration Unit (See Section 2.2.2)
ZO21S Standard Gas Unit (See Section 2.3)
K9471UA Dust Filter for Oxygen Analyzer (See Section 2.4)
2.1 General Specications
2.1.1 Standard Specications
Measured Object: Oxygen concentration in combustion exhaust gas and mixed gas (excluding
inammable gases). May not be applicable corrosive gas such as ammonia,
chlorine is present-check with YOKOGAWA.)
2-1
Measured System: Zirconia system
Oxygen Concentration: 0.01 to 100 vol% O
Output Signal: 4 to 20 mA DC (maximum load resistance 550Ω)
Measurement Range: Any setting in the range of 0 to 5 through 0 to 100 vol% O
partial range
Digital Communication (HART): 250 to 550Ω, depending on number of eld devices connected to the
loop (multi-drop mode).
Note: HART is a registered trademark of the HART Communication Foundation.
Display Range: 0 to 100 vol% O
Warm-up Time: Approx. 20 min.
Repeatability: (Excluding the case where the reference gas is by natural convection)
± 0.5% Maximum value of set range; less than 0 to 25 vol% O2 range
± 1 % Maximum value of set range; 0 to 25 vol%
Linearity: (Excluding standard gas tolerance)
(Excluding the case where the reference gas is by natural convection) (Use oxygen of known concentration (with in the measuring range) as the zero and span calibration gases.) ± 1% Maximum value of set range; less than 0 to 25 vol% O2 range
± 3% Maximum value of set range; 0 to 25 vol% O2 or more and less than 0
± 5% Maximum value of set range; 0 to 50 vol %O2 or more and up to 0 to
2
(in 1 vol%O2), or
2
2
O2 or more and up to 0 to
100 vol% O2 range
(Sample gas pressure: within ± 4.9 kPa)
to 50 vol% O2 range (Sample gas pressure: within ± 0.49 kPa)
100 vol% O2 range (Sample gas pressure: within ± 0.49 kPa)
IM 11M12A01-04E
<2. Specications>
2-2
IM 11M12A01-04E
Drift: (Excluding the rst two weeks in use)
(Excluding the case where the reference gas is by natural convection.) Both zero and span ± 2% Maximum value of set range/month
Response Time: Response of 90% within 5 seconds. (Measured after gas is introduced from
calibration gas inlet and analog output starts changing.)

2.1.2 ZR202G Integrated type Zirconia Oxygen Analyzer

Can be operated in the eld without opening the cover using optical switches.
Display: 6-digit LCD
Switch: Three optical switches
Output Signal: 4 to 20 mA DC, one point (maximum load resistance 550 Ω)
Digital Communication (HART): 250 to 550 Ω, depending on quantity of eld devices connected to the
loop (multi-drop mode).
Note: HART is a registered trademark of the HART Communication Foundation.
Contact Output Signal: Two points (one is fail-safe, normally open)
Contact Input Signal: Two points
Sample Gas Temperature: 0 to 700°C
It is necessary to mount the cell using Inconel cell-bolts when the temperature measures more than 600°C.
High-temperature service ― greater than 700°C ― is not available.
Sample Gas Pressure: -5 to +250 kPa (When the pressure in the furnace exceeds 3 kPa, it is
recommended to compensate the pressure. When the pressure in the furnace exceeds 5 kPa, you must perform pressure compensation.)
No pressure uctuation in the furnace be allowed.
Note: When the detector is used in conjunction with a check valve and the ZA8F Flow Setting Unit, the maximum pressure of sample
gas is 150 kPa. When with a check valve and the ZR20H Automatic Calibration Unit, it is 200 kPa. If the pressure of your sample gas exceeds these limits, consult with Yokogawa.
Probe Length: 0.4, 0.7, 1.0, 1.5, 2.0, 2.5, 3.0 m
Probe Material: SUS 316 (JIS)
Ambient Temperature: -20 to +55°C (- 5 to +70°C on the case surface)
Storage Temperature: -30 to +70°C
Humidity Range: 0 to 95%RH (non-condensing)
Installation Altitude: 2000 m or less
Category based on IEC 1010: II (Note)
Pollution degree based on IEC 1010: 2 (Note)
Note: Installation category, called over-voltage category, species impulse withstand voltage. Category II is for electrical equipment.
Pollution degree indicates the degree of existence of solid, liquid, gas or other inclusions which may reduce dielectric strength. Degree 2 is the normal indoor environment.
Power Supply Voltage: Ratings; 100 to 240 V AC
Acceptable range; 85 to 264 V AC
Power Supply Frequency: Ratings; 50/60 Hz
Acceptable range; 45 to 66 Hz
Power Consumption: Max. 300 W, approx. 100 W for ordinary use.
<2. Specications>
CAUTION
Safety and EMC conforming standards
Safety: EN61010-1, CSA C22.2 No.61010-1,
EMC: EN 61326 Class A,
CISPR 11
KC Marking: Korea Electromagnetic Conformity Standard
This instrument is a Class A product, and it is designed for use in the industrial environment. Please use this instrument in the industrial environment only.
Reference Gas System: Natural Convection, Instrument Air, Pressure Compensation
Instrument Air System (excluding Natural Convection):
Pressure; 200 kPa + the pressure inside the furnace (It is recommended
to use air which is dehumidied by cooling to dew point -20°C or
less, and dust or oil mist are removed.)
Consumption; Approx. 1Nl/min
EN 55011 Class A Group 1, EN 61000-3-2, AS/NZS
UL61010-1
2-3
Material in Contact with Gas: SUS 316 (JIS), Zirconia, SUS 304 (JIS) (ange), Hastelloy B, (Inconel
600, 601)
Construction: Heater and thermocouple replaceable construction. Non explosion-proof JIS
C0920 / equivalent to IP44D. Equivalent to NEMA 4X/IP66 (Achieved when the cable entry is completely sealed with a cable gland in the recirculation pressure compensated version.)
Gas Connection: Rc 1/4 or 1/4 FNPT
Wiring Connection: G1/2, Pg13.5, M20 by 1.5mm, 1/2 NPT select one type (4 pieces)
Installation: Flange mounting
Probe Mounting Angle: Horizontal to vertically downward.
When the probe insertion length is 2 m or less, installing at angles from horizontal to vertically downward is available. When the probe insertion length is 2.5m or more, mount vertically downward (within ± 5°), and if installing at angles from horizontal to vertically downward (within ± 5°), use a probe protector.
Case: Aluminum alloy
Paint Color: Cover; Mint green (Munsell 5.6BG3.3/2.9)
Case; Mint green (Munsell 5.6BG3.3/2.9)
Finish: Polyurethane corrosion-resistance coating
Insertion length of 0.4m: approx. 8 kg (JIS 5K 65) / approx. 13 kg (ANSI 150 4)
Insertion length of 1.0m: approx. 10 kg (JIS 5K 65) / approx. 15 kg (ANSI 150 4)
Insertion length of 1.5m: approx. 12 kg (JIS 5K 65) / approx. 17 kg (ANSI 150 4)
Insertion length of 2.0m: approx. 14 kg (JIS 5K 65) / approx. 19 kg (ANSI 150 4)
Insertion length of 3.0m: approx. 17 kg (JIS 5K 65) / approx. 22 kg (ANSI 150 4)
Weight:
IM 11M12A01-04E
<2. Specications>
2-4
IM 11M12A01-04E
Functions
Display Function: Displays values of the measured oxygen concentration, etc.
Alarm, Error Display: Displays alarms such as “AL-06” or errors such as “Err-01” when any such
status occurs.
Calibration Functions: Auto-calibration; Requires the Automatic Calibration Unit. It calibrates
Maintenance Functions: Can operate updated data settings in daily operation and checking. Display
Setup Functions: Initial settings suit for the plant conditions when installing the converter.
Display and setting content:
Display Related Items: Oxygen concentration (vol% O2), Output current value (mA), air ratio,
automatically at specied intervals.
Semi-auto Calibration; Requires the Automatic Calibration Unit. Input
calibration start signal by optical switch or contact, then it calibrates automatically afterwards.
Manual Calibration; Calibration with opening/closing the valve of calibration
gas in operation interactively with the optical switch.
data settings, calibration data settings, test settings (current output loop check, input/output contact check).
Current output data settings, alarm data settings, contact data settings, other settings.
moisture quantity (in hot gases) (vol% H2O), Cell temperature (°C ), thermocouple reference junction temperature (°C ), maximum/minimum/ average oxygen concentration (vol% O2), cell e.m.f. (mV), cell internal
resistance (Ω), cell condition (in four grades), heater on-time rate (%),
calibration record (ten times), time (year/month/day/hour/minute)
Calibration Setting Items: Span gas concentration (vol% O2), zero gas concentration (vol%O2),
calibration mode (auto, semi-auto, manual), calibration type and method (zero-span calibration, zero calibration only, span calibration only), stabilization time (min.sec), calibration time (min.sec), calibration period (day/hour), starting time (year/month/day/hour/minute)
Output Related Items: Analog output/output mode selection, output conditions when warming-up/
maintenance/calibrating/abnormal, 4 mA/20 mA point oxygen concentration (vol% O2), time constant, preset values when warming-up/maintenance/ calibrating/abnormal, output preset values on abnormal
Alarm Related Items: Oxygen concentration high alarm/high-high alarm limit values (vol%
Oxygen concentration low alarm/low-low alarm limit values (vol% O2), Oxygen concentration alarm hysteresis (vol% O2), Oxygen concentration alarm detection, alarm delay (seconds)
Contact Related Items: Selection of contact input 1 and 2, selection of contact output 1 and
2 (abnormal, high-high alarm, high alarm, low alarm, low-low alarm, maintenance, calibrating, range switching, warming-up, calibration gas
pressure decrease, ameout gas detection (answer-back of contact input)
O2),
<2. Specications>
Converter Output: One mA analog output point (4 to 20 mA DC (maximum load resistance of
550Ω)) with mA digital output point (HART) (minimum load resistance of 250Ω).
2-5
Range; Any setting between 0 to 5 through 0 to 100 vol%
O2 in 1 vol% O2, and partial range is available (Maximum range value/minimum range value 1.3 or more) For the log output, the minimum range value is xed at 0.1 vol% O2. 4 to 20 mA DC linear or log can be selected. Input/output isolation.
Output damping; 0 to 255 seconds.
Hold/non-hold selection, preset value setting possible with hold.
Contact Output: Two points, contact capacity 30V DC 3A, 250V AC 3A (resistive load)
One of the output points can be selected to ether normally energized or nor­mally de-energized status. Delayed functions (0 to 255 seconds) and hysteresis function (0 to 9.9 vol% O2) can be added to high/low alarms. The following functions are programmable for contact outputs. (1) Abnormal, (2) High-high alarm, (3) High alarm, (4) Low-low alarm, (5) Low alarm, (6) Maintenance, (7) Calibration, (8) Range switching answer­back, (9) Warm-up, (10) Calibration gas pressure decrease (answer-back of contact input), (11) Flameout gas detection (answer-back of contact input).
Contact Input: Two points, voltage-free contacts
The following functions are programmable for contact inputs: (1) Calibration gas pressure decrease alarm, (2) Range switching (switched
range is xed), (3) External calibration start, (4) Process alarm (if this signal
is received, the heater power turns off)
Self-diagnosis: Abnormal cell, abnormal cell temperature (low/high), abnormal calibration,
A/D converter abnormal, digital circuit abnormal
Calibration: Method; zero/span calibration
Calibration mode; Auto, semi-auto and manual (All are operated using
optical switches). Either zero or span can be skipped.
Zero calibration gas concentration setting range; 0.3 to 100 vol% O2 (in 0.01
vol% O2).
Span calibration gas concentration setting range; 4.5 to 100 vol% O2 (in 0.01
vol% O2).
Use nitrogen-balanced mixed gas containing 0 to 10 vol% O2 scale of
oxygen for standard zero gas and 80 to 100 vol% O2 scale of oxygen for standard span gas. Calibration period; date/time setting: maximum 255 days
IM 11M12A01-04E
<2. Specications>
2-6
IM 11M12A01-04E
Model and Codes
Model Sufx code Option code Description
ZR202G - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - Integrated type Zirconia Oxygen Analyzer
Length -040
Wetted material -S
Flange (*2)
Auto Calibration -N
Reference gas -C
Gas Thread -R
Connection box thread -P
Instruction manual -J
-- -A - - - - - - - - - - Always -A
Options
Valves
Tag plates
NAMUR NE43 compliant
-070
-100
-150
-200
-250
-300
-C
-A
-B
-C
-E
-F
-G
-K
-L
-M
-P
-R
-S
-W
-A
-B
-E
-P
-T
-G
-M
-T
-E
-C
*1 For the horizontally installed probe whose insertion length is 2.5 m or more, use the Probe Protector. Be sure to specify
ZO21R-L-200-. Specify the ange sufx code either -C or -K.
*2 The thickness of the ange depends on its dimensions.
*3 Inconel probe bolts and U shape pipe are used. Use this option for high temperature use (ranging from 600 to 700°C). *4 Specify either /CV or /SV option code. *5 Specify either /SCT or /PT option code. *6 No need to specify the option codes, /CV and /SV, since the check valves are provided with the Automatic Calibration Unit.
Automatic calibration cannot be used when natural convection is selected as reference gas.
*7 Sun shield hood is still effective even if scratched. Hood is necessary for outdoor installation out of sun shield roof. *8 Recommended if sample gas contains corrosive gas like chlorine.
*9 Piping for reference gas must be installed to supply reference gas constantly at a specied ow rate.
*10 Output signal limits: 3.8 to 20.5 mA. Specify either /C2 or /C3 option code.
- - - - - - - - - -
- - - - - - - - - -
- - - - - - - - - -
- - - - - - - - - -
- - - - - - - - - -
- - - - - - - - - -
- - - - - - - - - -
- - - - - - - - - -
- - - - - - - - - -
- - - - - - - - - -
- - - - - - - - - -
- - - - - - - - - -
- - - - - - - - - -
- - - - - - - - - -
- - - - - - - - - -
- - - - - - - - - -
- - - - - - - - - -
- - - - - - - - - -
- - - - - - - - - -
- - - - - - - - - -
- - - - - - - - - -
- - - - - - - - - -
- - - - - - - - - -
- - - - - - - - - -
- - - - - - - - - -
- - - - - - - - - -
- - - - - - - - - -
- - - - - - - - - -
- - - - - - - - - -
- - - - - - - - - -
- - - - - - - - - -
- - - - - - - - - -
- - - - - - - - - -
- - - - - - - - - -
- - - - - - - - - -
- - - - - - - - - -
- - - - - - - - - -
/C Inconel bolt (*3)
/CV /SV
/H Hood (*7)
/F1 /F2
/SCT /PT
/C2
/C3
0.4 m
0.7 m
1.0 m
1.5 m
2.0 m
2.5 m (*1)
3.0 m (*1)
SUS316 Stainless steel with Inconel calibration gas tube (*8)
ANSI Class 150 2 RF SUS304 ANSI Class 150 3 RF SUS304 ANSI Class 150 4 RF SUS304 DIN PN10 DN50 A SUS304 DIN PN10 DN80 A SUS304 DIN PN10 DN100 A SUS304 JIS 5K 65 FF SUS304 JIS 10K 65 FF SUS304 JIS 10K 80 FF SUS304 JIS 10K 100 FF SUS304 JPI Class 150 4 RF SUS304 JPI Class 150 3 RF SUS304 Westinghouse
Not required Horizontal mounting (*6) Vertical mounting (*6)
Natural convection External connection (Instrument air) (*9) Pressure compensated (*9)
Rc 1/4 1/4 NPT (Female)
G1/2 Pg13.5 M20x1.5 mm 1/2NPT
Japanese English Chinese
Check valve (*4) Stop valve (*4)
Dust Filter Dust Guard Protector
Stainless steel tag plate (*5) Printed tag plate (*5)
Failure alarm down-scale: Output status at CPU failure and hardware error is 3.6 mA or less (*10) Failure alarm up-scale: Output status at CPU failure and hardware error is 21.0 mA or more (*10)
<2. Specications>
F11_01.ai
25
t
153 to 164
125
L
170
252 to 265
49
48.5
338 to 351
122
L= 0.4, 0.7,
1.0, 1.5, 2.0,
2.5, 3.0 (m)
4-G1/2,2-1/2NPT etc. Cable connection port
Display side
ANSI Class 150 2 RF SUS304 ANSI Class 150 3 RF SUS304 ANSI Class 150 4 RF SUS304 DIN PN10 DN50 A SUS304 DIN PN10 DN80 A SUS304 DIN PN10 DN100 A SUS304 JIS 5K 65 FF SUS304 JIS 10K 65 FF SUS304 JIS 10K 80 FF SUS304 JIS 10K 100 FF SUS304 JPI Class 150 4 RF SUS304 JPI Class 150 3 RF SUS304 Westinghouse
152.4
190.5
228.6 165 200 220 155 175 185 210 229 190 155
120.6
152.4
190.5 125 160 180 130 140 150 175
190.5
152.4 127
4 - Ø19 4 - Ø19 8 - Ø19 4 - Ø18 8 - Ø18 8 - Ø18 4 - Ø15 4 - Ø19 8 - Ø19 8 - Ø19 8 - Ø19 4 - Ø19
4 - Ø11.5
A B C
19 24 24 18 20 20 14 18 18 18 24 24 14
t
Ø50.8
Rc1/4 or 1/4NPT Reference gas inlet
Rc1/4 or 1/4NPT Calibration gas inlet
Ø123
Terminal side
C
Flange
Flange
Ø
B
C
Flange
Ø
A
Flange
Ø
A
Ø
B
• External Dimensions
Model ZR202G Integrated type Zirconia Oxygen Analyzers Unit: mm
2-7
• Standard Accessories
Fuse A1113EF 1 3.15 A
Allen wrench L9827AB 1 For lock screw
Item Parts. No. Q'ty Description
IM 11M12A01-04E
<2. Specications>
2-8
IM 11M12A01-04E
F11_02.EPS
Flange
156 ± 3
125
170
256 ± 4
49
48.5
Ø123
342 ± 4
25
t
122
Terminal side
PIPING:A
PIPING
:B
PIPING
A B B A B B A A B B B B A
ANSI Class 150 2 RF SUS304 ANSI Class 150 3 RF SUS304 ANSI Class 150 4 RF SUS304 DIN PN10 DN50 A SUS304 DIN PN10 DN80 A SUS304 DIN PN10 DN100 A SUS304 JIS 5K 65 FF SUS304 JIS 10K 65 FF SUS304 JIS 10K 80 FF SUS304 JIS 10K 100 FF SUS304 JPI Class 150 4 RF SUS304 JPI Class 150 3 RF SUS304 Westinghouse
152.4
190.5
228.6 165 200 220 155 175 185 210 229 190 155
120.6
152.4
190.5 125 160 180 130 140 150 175
190.5
152.4 127
4 - Ø19 4 - Ø19 8 - Ø19 4 - Ø18 8 - Ø18 8 - Ø18 4 - Ø15 4 - Ø19 8 - Ø19 8 - Ø19 8 - Ø19 4 - Ø19
4 - Ø11.5
A B C
19 24 24 18 20 20 14 18 18 18 24 24 14
t
L
L= 0.4, 0.7,
1.0, 1.5, 2.0,
2.5, 3.0 (m)
4-G1/2,2-1/2NPT etc. Cable connection port
Ø50.8
Rc1/4 or 1/4NPT Reference gas inlet
Rc1/4 or 1/4NPT Calibration gas inlet
Reference gas outlet
Display side
C
Flange
ØB
C
Flange
ØA
Flange
Stop
valve
ØA
ØB
274
150
150
± 4
± 3
± 3
F13.ai
Material of HOOD : Aluminum Food Weight : Approx. 800g
Model ZR202G...-P Integrated type Zirconia Oxygen Analyzer with pressure compensation
Unit: mm
• Standard Accessories
• Hood (Option code /H) Unit: mm
Item Parts. No. Q'ty Description
Fuse A1113EF 1 3.15 A Allen wrench L9827AB 1 For lock screw
<2. Specications>
F2-3E.ai
D
Ø
B
C
Ø
B
Flange <1> (with bolts, nuts and washer)
gasket (t1.5)
SUS316
Dimensions of holes on opposing surface
ØA
Gas flow
Washer (12)
Mounting nut (M12)
l (Insert length)
l=1050,1550,2050
Ø60.5

2.1.3 ZO21R Probe Protector

This probe protector is required for the general-use detector when it is used for oxygen concentration
measurements in powdered coal boilers or in uidized furnaces to prevent abrasion due to dust particles when gas ow exceeds 10 m/sec.
When probe length of the ZR202G is 2.5 m or more and mounted horizontally, be sure to select a probe protector ZO21R-L-200-*B to support the probe.
Insertion Length: 1.05 m, 1.55 m, 2.05 m.
Flange: JIS 5K 65A FF equivalent, ANSI Class 150 4 FF (without serration)
equivalent or DIN PN10 DN50A equivalent. However, ange thickness is
different.
Material: SUS316 (JIS), SUS304 (JIS) (Flange)
2-9
Weight: 1.05m; Approx. 6/10 kg (JIS/ANSI), 1.55 m; Approx. 9/13 kg
Installation: Bolts, nuts, and washers are provided for detector, probe adapter and
• Model and Codes
Model Sufx code Option code Description
ZO21R -L - - - - - - - - - - - - Probe Protector(0 to 700°C)
Insertion length
Flange ( *1) -J
Style code *B - - - - - - - - - - - - Style B
*1 Thickness of ange depends on dimensions of ange.
• External Dimension
(JIS/ANSI), 2.05 m; Approx. 12/16 kg (JIS/ANSI)
process-side ange.
-100
-150
-200
-A
-E
- - - - - - - - - - - -
- - - - - - - - - - - -
- - - - - - - - - - - -
- - - - - - - - - - - -
- - - - - - - - - - - -
- - - - - - - - - - - -
1.05 m (3.5 ft)
1.55 m (5.1 ft)
2.05 m (6.8 ft)
JIS 5K 65 FF SUS304 ANSI Class 150 4 FF SUS304 DIN PN10 DN50A SUS304
Unit: mm
Flange<1> A B C t D
JIS 5K 65 FF SUS304 155 130 4 - Ø15 5 40
ANSI Class 150 4 FF SUS304 228.6 190.5 8 - Ø19 12 50
DIN PN10 DN50A SUS304 165 125 4 - Ø18 12 50
IM 11M12A01-04E
<2. Specications>
2-10
IM 11M12A01-04E
NOTE

2.2 ZA8F Flow Setting Unit and ZR20H Automatic Calibration Unit

2.2.1 ZA8F Flow Setting Unit

This ow setting unit is applied to the reference gas and the calibration gas in a system conguration
(System 2).
This unit consists of a owmeter and ow control valves to control the ow of calibration gas and
reference gas.
• Standard Specications
Flowmeter: Calibration gas; 0.1 to 1.0 l/min. Reference gas; 0.1 to 1.0 l/min.
Construction: Dust-proof and rainproof construction
Case Material: SPCC (Cold rolled steel sheet)
Painting: Baked epoxy resin, Dark-green (Munsell 2.0 GY 3.1/0.5 or equivalent)
Tube Connections: Rc1/4 or 1/4FNPT
Reference Gas Pressure: Clean air supply of sample gas pressure plus approx. 50 kPa G (or
sample gas pressure plus approx.150 kPa G when a check valve is used, maximum pressure rating is 300 kPa) (pressure at inlet of the automatic calibration unit)
Air Consumption: Approx. 1.5 l/min
Weight: Approx. 2 kg
Used instrument air for span calibration gas, if without instrument air is used, contact YOKOGAWA.
• Model and Codes
Model Sufx code Option code Description
ZA8F - - - - - - - - - - - - - - - - - - - - - - - Flow setting unit
Joint
Style code *C - - - - - - - - - - - - Style C
-J
-A
- - - - - - - - - - - -
- - - - - - - - - - - -
Rc 1/4 With 1/4 NPT adapter
<2. Specications>
CHECK
OUT
REF OUT
REFERENCE
SPAN
ZERO
REFERENCE CHECK
AIR
IN
ZERO
IN
SPAN
IN
CHECK OUT
PIPNG INSIDE THE FLOW SETTING UNIT
Unit : mm (inch)
Weight : Approx. 2.3 kg
Flowmeter Flowmeter
ZERO
GAS IN
AIR IN
Instrument air Approx. 1.5 l/min.
Air Set
REF OUT
SPAN
GAS IN
20
2035
35
35
35
35
70
32
7
222.8
235.8
140
180
7
Reference gas outlet
Calibration gas outlet
ø6 Hole
Span gas inlet
Zero gas inlet
2B mounting pipe
Instrument air inlet
F2.6E.ai
Air pressure ; without check valve ; sample gas pressure + approx.50 kPaG with check valve ; sample gas pressure + approx.150 kPaG
Piping connection port A
ZA8F-J*C
ZA8F-A*C
5 - Rc1/4
5 - 1/4NPT
Model
Piping connection port A
• External Dimensions
2-11
IM 11M12A01-04E
Loading...
+ 127 hidden pages