Model ZR202G
Integrated type
Zirconia Oxygen Analyzer
IM 11M12A01-04E
IM 11M12A01-04E
7th Edition
Introduction
The EXAxt ZR Integrated type Zirconia Oxygen Analyzer has been developed for combustion control
in various industrial processes. There are several version of this analyzer so you can select one that
matches your application.
Optional accessories are also available to improve measurement accuracy and automate calibration.
An optimal control system can be realized by adding appropriate options.
This instruction manual describes almost all of the equipment related to the EXAxt ZR. You may skip
any section(s) regarding equipment which is not included in your system.
Regarding the HART Communication Protocol, refer to IM 11M12A01-51E.
IM11M12A01-51E has been published as ''Model EXAxt ZR series HART protocol''.
Regarding Separate type Zirconia Oxygen Analyzer, refer to IM 11M12A01-02E.
< Before using the equipment, please read any descriptions in this manual related to the equipment
and system that you have, on appropriate use and operation of the EXAxt ZR. >
Models and descriptions in this manual are listed below.
This manual consists of twelve chapters. Please refer to the reference chapters for installation, operation and maintenance.
Chapter
1. Overview Equipment models and system conguration
examples
2. Specications Standard specication, model code (or part number),
3. InstallationInstallation method for each equipment
4. PipingExamples of piping in three standard system
5. WiringWiring procedures such as “Power supply wiring”,
6. ComponentsMajor parts and functions are described
7. StartupBasic procedure to start operation of EXAxt ZR.
8. Detailed Data
Setting
9. CalibrationDescribes the calibration procedure required in the
10. Other
Functions
11. Inspection
and
Maintenance
12. Troubleshoot-
ing
CMPL (parts list)User replaceable parts list
: Read and completely understand before operating the equipment.
: Read before operating the equipment, and refer to it whenever necessary.
: Recommended to read at least once.
dimension drawing for each equipment
congurations
“output signal wiring” or others
Chapter 7enables you to operate the equipment
immediately.
Details of key operations and displays
course of operation.
Other functions described
How to conduct maintenance of EXAxt ZR and
procedures for replacement of deteriorated parts
This chapter describes measures to be taken when an
abnormal condition occurs.
Outline
Relates to
Installation Operation Maintenance
For the safe use of this equipment
WARNING
CAUTION
EXAxt ZR is very heavy. Be sure not to accidentally drop it. Handle safely to avoid injury.
Connect the power supply cord only after conrming that the supply voltage matches the rating of this
equipment. In addition, conrm that the power is switched off when connecting power supply.
Some process gas is dangerous to people. When removing this equipment from the process line for
maintenance or other reasons, protect yourself from potential poisoning by using a protective mask or
ventilating the area well
The cell (sensor) at the tip of the probe is made of ceramic (zirconia element). Do not drop the equipment or subject it to pressure stress.
<Introduction>
iii
• Do NOT allow the sensor (probe tip) to make contact with anything when installing the analyzer.
• Avoid water dropping directly on the probe (sensor) of the analyzer when installing it.
• Check the calibration gas piping before introducing the calibration gas to ensure that there is no
leakage of the gas. If there is any leakage of the gas, the moisture drawn from the sample gas
may damage the sensor.
• The probe (especially at the tip) becomes very hot. Be sure to handle it with gloves.
(1) About This Manual
The contents of this manual shall not be reproduced or copied, in part or in whole, without per-
mission.
This manual explains the functions contained in this product, but does not warrant that those will
suit the particular purpose of the user.
Every effort has been made to ensure accuracy in the preparation of this manual. However,
should any errors or omissions come to the attention of the user, please contact the nearest
Yokogawa Electric representative or sales ofce.
This manual does not cover the special specications. This manual may not be changed on any
change of specication, construction and parts when the change does not affect the functions or
performance of the product.
If the product is used in a manner not specied in this manual, the safety of this product may be
impaired.
IM 11M12A01-04E
iv
IM 11M12A01-04E
<Introduction>
WARNING
CAUTION
IMPORTANT
NOTE
(2) Safety and Modication Precautions
Follow the safety precautions in this manual when using the product to ensure protection and
safety of personnel, product and system containing the product.
The following safety symbols and wordings are used on the product as well as in this manual.
(3) The following safety symbols are used in this manual.
Throughout this user’s manual, you will find several different types of symbols are used to identify
different sections of text. This section describes these icons.
Indicates a potentially hazardous situation which, if not avoided, could result in death or serious injury.
Indicates a potentially hazardous situation which, if not avoided, may result in minor or moderate
injury. It may also be used to alert against unsafe practices.
Indicates that operating the hardware or software in this manner may damage it or lead to system
failure.
Draws attention to information essential for understanding the operation and features.
TIP
Identies additional information.
SEE ALSO
Identies a source to be referred to.
Clicking a reference displayed in green can call up its source, while clicking a reference displayed
When the instrument arrives, unpack the package with care and check that the instrument has not
been damaged during transportation. In addition, please check that the specication matches the
order, and required accessories are not missing. Specications can be checked by the model codes
on the nameplate. Refer to Chapter 2 specications for the list of model codes.
• Details on operation parameters
When the EXAxt ZR Integrated type Zirconia Oxygen Analyzer arrives at the user site, it will operate
based on the operation parameters (initial data) set before shipping from the factory.
Ensure that the initial data is suitable for the operation conditions before conducting analysis.
Where necessary, set the instrument parameters for appropriate operation. For details on setting data,
refer to Chapters 7 to 10.
When user changes the operation parameter, it is recommended to note down the changed setting
data.
• How to dispose the batteries:
This is an explanation about the new EU Battery Directive (DIRECTIVE 2006/66/EC). This directive is
only valid in the EU.
Batteries are included in this product. Batteries incorporated into this product cannot be removed by
yourself. Dispose them together with this product.
When you dispose this product in the EU, contact your local Yokogawa Europe B.V.ofce. Do not
dispose them as domestic household waste.
Battery type: silver oxide battery
Notice:
The symbol (see above) means they shall be sorted out and collected as ordained in ANNEX II in
DIRECTIVE 2006/66/EC
After-Sales Warranty
Do not modify the product.
During the warranty period, for repair under warranty carry or send the product to the local sales
representative or service ofce. Yokogawa will replace or repair any damaged parts and return
the product to you.
Before returning a product for repair under warranty, provide us with the model name and serial
number and a description of the problem. Any diagrams or data explaining the problem would
also be appreciated.
If we replace the product with a new one, we won’t provide you with a repair report.
Yokogawa warrants the product for the period stated in the pre-purchase quotation. Yokogawa
shall conduct dened warranty service based on its standard. When the customer site is located
outside of the service area, a fee for dispatching the maintenance engineer will be charged to the
customer.
In the following cases, customer will be charged repair fee regardless of warranty period.
• Failure of components which are out of scope of warranty stated in instruction manual.
• Failure caused by usage of software, hardware or auxiliary equipment, which Yokogawa Elec-
tric did not supply.
<Introduction>
vii
• Failure due to improper or insufcient maintenance by user.
• Failure due to modication, misuse or outside-of-specications operation which Yokogawa
does not authorize.
• Failure due to power supply (voltage, frequency) being outside specications or abnormal.
• Failure caused by any usage out of scope of recommended usage.
• Any damage from re, earthquake, storms and oods, lightning, disturbances, riots, warfare,
radiation and other natural changes.
Yokogawa does not warrant conformance with the specic application at the user site. Yokogawa
will not bear direct/indirect responsibility for damage due to a specic application.
Yokogawa Electric will not bear responsibility when the user congures the product into systems
or resells the product.
Maintenance service and supplying repair parts will be covered for ve years after the production
ends. For repair for this product, please contact the nearest sales ofce described in this instruc-
tion manual.
IM 11M12A01-04E
viii
<Introduction>
IM 11M12A01-04E
<CONTENTS>
Model ZR202G
Integrated type Zirconia Oxygen Analyzer
12.1 Displays and Measures to Take When Errors Occur .................................. 12-1
12.1.1 What is an Error? .............................................................................
12.1.2 Measures to Take When an Error Occurs .......................................
12.2 Displays and Measures to Take When Alarms are Generated ...................
12.2.1 What is an Alarm? ............................................................................
12.3 Measures When Measured Value Shows an Error ......................................
12.3.1 Measured Value Higher Than True Value ........................................
12.3.2 Measured Value Lower Than True Value ......................................
12.3.3 Measurements Sometimes Show Abnormal Values .....................
12-1
12-2
12-4
12-4
12-9
12-9
12-10
12-10
Customer Maintenance Parts List ........................................CMPL 11M12A01- 04E
Customer Maintenance Parts List ........................................CMPL 11M12A01-12E
Customer Maintenance Parts List ...........................................CMPL 11M3D1- 01E
Revision Information ...............................................................................................i
IM 11M12A01-04E
<1.Overview>
~
ZR202G Integrated type Zirconia Oxygen Analyzer
ZO21S Standard gas unit
Stop valve
Calibration gas
F1.1E.ai
~
100 to 240 V AC
100/110/115
200/220/240 V AC
Contact input
Analog output, contact output
Digital output (HART)
1. Overview
The EXAxt ZR Integrated type Zirconia Oxygen Analyzer is used to monitor and control the oxygen
concentration in combustion gases, in boilers and industrial furnaces, for wide application in industries
which consume considerable energy-such as steel, electric power, oil and petrochemical, ceramics,
pulp and paper, food, or textiles, as well as incinerators and medium/small boilers. It can help conserve
energy in these industries. The EXAxt ZR also contributes to preservation of the earth’s environment
in preventing global warming and air pollution by controlling complete combustion to reduce CO2, SOx
and NOx.
The EXAxt ZR Integrated type Zirconia Oxygen Analyzer integrates both probe and converter. The
analyzers need not use a sampling device, and allow direct installation of the probe in the wall of a ue
or furnace to measure the concentration of oxygen in the stack gas of the temperature up to 700°C.
The probe uses a high-reliability Zirconia sensor and a heater assembly that can be replaced on site.
The analyzer is equipped with three infrared switches, which enable the user to operate the equipment
without opening the cover on site. Analyzer calibration can also be fully automated and the automatic
calibration unit is provided. Choose the equipment which best suits your needs so that an optimal
combustion control system can be obtained.
Some examples of typical system conguration are illustrated below:
1-1
1.1<EXAxtZR>SystemConguration
The system configuration should be determined by the conditions; e.g. whether the calibration is to
be automated, and whether flammable gas is present and requires safety precautions. The system
conguration can be classied into three basic patterns as follows:
1.1.1 System 1
This is the simplest system consisting of an integrated type analyzer. This system can be implemented
for monitoring oxygen concentration in the combustion gases boiler. No piping is required for the
reference gas (air) which is fed in at the installation site. The ZO21S standard gas unit is used for
calibration.
Zero gas from this unit and span gas (air) is sent to the probe through a tube which is connected
during calibration.
Figure1.1ExampleofSystem1
IM 11M12A01-04E
<1. Overview>
1-2
IM 11M12A01-04E
CAUTION
~
ZR202G Integrated type Zirconia Oxygen Analyzer
F1.2E.ai
ZA8F flow setting unit
Reference gas
Calibration gas
Needle
valve
Flowmeter
Instrument air
Air Set
Calibration
gas pressure
regulator
Zero gas cylinder
Calibration gas unit case
Stop valve
or
Check valve
Span gas(Same as Zero gas calibration)
100 to 240 V AC
Contact input
Analog output, contact output
Digital output (HART)
• As this system uses ambient air for the reference gas, measuring accuracy will be affected by
the installation location.
• A stop valve should be connected to the calibration gas inlet of the equipment. The valve should
be fully closed unless calibration is in progress.
1.1.2 System 2
This system is for monitoring and controlling oxygen concentration in the combustion gases of a
large-size boiler or heating furnace. Instrument air (clean and dry air of oxygen concentration 21%)
is used as the reference gas and the span gas for calibration. Zero gas is fed from a cylinder during
calibration. The gas ow is controlled by the ZA8F ow setting unit (for manual valve operation).
Figure 1.2 Example of System 2
<1.Overview>
~
100 to 240 V AC
Automatic calibration unit
Reference gas
Calibration gas (Zero)
Contact input
Analog output, contact output
Digital output (HART)
Air Set
Instrument air
Calibration gas unit case
Calibration
gas pressure
regulator
Zero gas cylinder
ZR202G Integrated type Zirconia Oxygen Analyzer
with automatic calibration unit (ZR202G-□-□-□-A-□-□-□-□-A)
F1.3E.ai
Span gas
ZR20H
*3
Note:
The installation temperature limits range for integrated type analyzer is -20 to 55°C.
*2
*1
1.1.3 System 3
This example, System 3, represents typical applications in large boilers and heating furnaces, where
is a need to monitor and control oxygen concentration. The reference gas and calibration-time span
gas are (clean, dry) instrument air. Zero gas is supplied from a gas cylinder.
System 3 uses the ZR20H automatic calibration unit, with auto-switching of the calibration gas.
A “combustible gas detected” contact input turns off power to the heater. There’s also contact output
from the converter that can be used to operate a purge gas valve to supply air to the sensor.
1-3
*1 Shield cable:
Use shielded signal cables, and connect the shields to the FG terminal of the converter.
*2 Select the desired probe from the Probe Conguration table on page 1-4.
*3 When a zirconia oxygen analyzer is used, 100% N
Use approx. 1 vol% O2 gas (N2-based).
gas cannot be used as the zero gas.
2
Figure 1.3 Example of System 3
IM 11M12A01-04E
1-4
System Components
ZR202G Integrated type Zirconia Oxygen Analyzers
ZO21R Probe Protector for Zirconia Oxygen Analyzers
K9471UA Dust Filter for Oxygen Analyzer
ZO21S Standard Gas Unit
ZA8F Flow Setting Unit for manual calibration
ZR20H Automatic Calibration Unit for Integrated type Analyzer
L9852CB, G7016XH Stop Valve for Calibration gas line
K9292DN,K9292DS Check Valve for Calibration gas line
G7003XF/K9473XK, G7004XF/K9473XG Air Set
G7001ZC Zero gas Cylinder
G7013XF, G7014XF Pressure Regulator for Gas Cylinder
E7044KF Case Assembly for Calibration gas Cylinder
Integrated type
System configuration
Example 3Example 1
Example 2
( )
( )
T1.1.ai
: Items required for the above system example
: To be selected depending on each application. For details, refer to corresponding chapter.
: Select either
( )
ZR202A Heater Assembly (Spare Parts for ZR202G)
K9471UC Dust Guard Protector
F1.4E.ai
Detector
(ZR202G)
Detector
(ZR202G)
Detector(ZR202G)
Probe Protector
(ZO21R)
Dust Filter
(K9471UA)
or
Dust Guard
Protector
(K9471UC)
+
Sample gas temperature 0 to 700°C
Mounting
Insertion length
General-use Probe
Application
Horizontal
to
vertical
Horizontal
to
vertical
Horizontal
to
vertical
Vertical
Vertical
0.4
to
2 m
0.4
to
2 m
2.5 m
or more
3 m
or less
2.5 m
or more
• Boiler
• Heating furnace
• For pulverized coal boiler with gas flow velocity
10 m/sec or more
• Cement Kiln
• Black liquid recovery boiler
• Cement Kiln
Gas Flow
Sample inlet
<1. Overview>
1.2 < EXAxt ZR > System Components
1.2.1 System Components
1.2.2 Oxygen Analyzer and Accessories
IM 11M12A01-04E
<2. Specications>
2. Specications
This chapter describes the specications for the following:
ZR202G General-use Integrated type Zirconia Oxygen Analyzer (See Section 2.1.2)
ZO21R-L Probe Protector (See Section 2.1.3)
ZA8F Flow Setting Unit (See Section 2.2.1)
ZR20H Automatic Calibration Unit (See Section 2.2.2)
ZO21S Standard Gas Unit (See Section 2.3)
K9471UA Dust Filter for Oxygen Analyzer (See Section 2.4)
2.1 General Specications
2.1.1 Standard Specications
Measured Object: Oxygen concentration in combustion exhaust gas and mixed gas (excluding
inammable gases). May not be applicable corrosive gas such as ammonia,
chlorine is present-check with YOKOGAWA.)
2-1
Measured System: Zirconia system
Oxygen Concentration: 0.01 to 100 vol% O
Output Signal: 4 to 20 mA DC (maximum load resistance 550Ω)
Measurement Range: Any setting in the range of 0 to 5 through 0 to 100 vol% O
partial range
Digital Communication (HART): 250 to 550Ω, depending on number of eld devices connected to the
loop (multi-drop mode).
Note: HART is a registered trademark of the HART Communication Foundation.
Display Range: 0 to 100 vol% O
Warm-up Time: Approx. 20 min.
Repeatability: (Excluding the case where the reference gas is by natural convection)
± 0.5% Maximum value of set range; less than 0 to 25 vol% O2 range
± 1 % Maximum value of set range; 0 to 25 vol%
Linearity: (Excluding standard gas tolerance)
(Excluding the case where the reference gas is by natural convection)
(Use oxygen of known concentration (with in the measuring range) as the
zero and span calibration gases.)
± 1% Maximum value of set range; less than 0 to 25 vol% O2 range
± 3% Maximum value of set range; 0 to 25 vol% O2 or more and less than 0
± 5% Maximum value of set range; 0 to 50 vol %O2 or more and up to 0 to
2
(in 1 vol%O2), or
2
2
O2 or more and up to 0 to
100 vol% O2 range
(Sample gas pressure: within ± 4.9 kPa)
to 50 vol% O2 range
(Sample gas pressure: within ± 0.49 kPa)
100 vol% O2 range
(Sample gas pressure: within ± 0.49 kPa)
IM 11M12A01-04E
<2. Specications>
2-2
IM 11M12A01-04E
Drift: (Excluding the rst two weeks in use)
(Excluding the case where the reference gas is by natural convection.)
Both zero and span ± 2% Maximum value of set range/month
Response Time: Response of 90% within 5 seconds. (Measured after gas is introduced from
calibration gas inlet and analog output starts changing.)
2.1.2 ZR202G Integrated type Zirconia Oxygen Analyzer
Can be operated in the eld without opening the cover using optical switches.
Display: 6-digit LCD
Switch: Three optical switches
Output Signal: 4 to 20 mA DC, one point (maximum load resistance 550 Ω)
Digital Communication (HART): 250 to 550 Ω, depending on quantity of eld devices connected to the
loop (multi-drop mode).
Note: HART is a registered trademark of the HART Communication Foundation.
Contact Output Signal: Two points (one is fail-safe, normally open)
Contact Input Signal: Two points
Sample Gas Temperature: 0 to 700°C
It is necessary to mount the cell using Inconel cell-bolts when the
temperature measures more than 600°C.
High-temperature service ― greater than 700°C ― is not available.
Sample Gas Pressure: -5 to +250 kPa (When the pressure in the furnace exceeds 3 kPa, it is
recommended to compensate the pressure. When the pressure in the
furnace exceeds 5 kPa, you must perform pressure compensation.)
No pressure uctuation in the furnace be allowed.
Note: When the detector is used in conjunction with a check valve and the ZA8F Flow Setting Unit, the maximum pressure of sample
gas is 150 kPa. When with a check valve and the ZR20H Automatic Calibration Unit, it is 200 kPa. If the pressure of your sample
gas exceeds these limits, consult with Yokogawa.
Probe Length: 0.4, 0.7, 1.0, 1.5, 2.0, 2.5, 3.0 m
Probe Material: SUS 316 (JIS)
Ambient Temperature: -20 to +55°C (- 5 to +70°C on the case surface)
Storage Temperature: -30 to +70°C
Humidity Range: 0 to 95%RH (non-condensing)
Installation Altitude: 2000 m or less
Category based on IEC 1010: II (Note)
Pollution degree based on IEC 1010: 2 (Note)
Note: Installation category, called over-voltage category, species impulse withstand voltage. Category II is for electrical equipment.
Pollution degree indicates the degree of existence of solid, liquid, gas or other inclusions which may reduce dielectric strength.
Degree 2 is the normal indoor environment.
Power Supply Voltage: Ratings; 100 to 240 V AC
Acceptable range; 85 to 264 V AC
Power Supply Frequency: Ratings; 50/60 Hz
Acceptable range; 45 to 66 Hz
Power Consumption: Max. 300 W, approx. 100 W for ordinary use.
<2. Specications>
CAUTION
Safety and EMC conforming standards
Safety: EN61010-1, CSA C22.2 No.61010-1,
EMC: EN 61326 Class A,
CISPR 11
KC Marking: Korea Electromagnetic Conformity Standard
This instrument is a Class A product, and it is designed for use in the industrial environment. Please
use this instrument in the industrial environment only.
Reference Gas System: Natural Convection, Instrument Air, Pressure Compensation
Instrument Air System (excluding Natural Convection):
Pressure; 200 kPa + the pressure inside the furnace (It is recommended
to use air which is dehumidied by cooling to dew point -20°C or
less, and dust or oil mist are removed.)
Consumption; Approx. 1Nl/min
EN 55011 Class A Group 1, EN 61000-3-2, AS/NZS
UL61010-1
2-3
Material in Contact with Gas: SUS 316 (JIS), Zirconia, SUS 304 (JIS) (ange), Hastelloy B, (Inconel
600, 601)
Construction: Heater and thermocouple replaceable construction. Non explosion-proof JIS
C0920 / equivalent to IP44D. Equivalent to NEMA 4X/IP66 (Achieved when
the cable entry is completely sealed with a cable gland in the recirculation
pressure compensated version.)
Gas Connection: Rc 1/4 or 1/4 FNPT
Wiring Connection: G1/2, Pg13.5, M20 by 1.5mm, 1/2 NPT select one type (4 pieces)
Installation: Flange mounting
Probe Mounting Angle: Horizontal to vertically downward.
When the probe insertion length is 2 m or less, installing at angles from
horizontal to vertically downward is available.
When the probe insertion length is 2.5m or more, mount vertically downward
(within ± 5°), and if installing at angles from horizontal to vertically downward
(within ± 5°), use a probe protector.
Case: Aluminum alloy
Paint Color: Cover; Mint green (Munsell 5.6BG3.3/2.9)
Case; Mint green (Munsell 5.6BG3.3/2.9)
Finish: Polyurethane corrosion-resistance coating
Insertion length of 0.4m: approx. 8 kg (JIS 5K 65) / approx. 13 kg (ANSI 150 4)
Insertion length of 1.0m: approx. 10 kg (JIS 5K 65) / approx. 15 kg (ANSI 150 4)
Insertion length of 1.5m: approx. 12 kg (JIS 5K 65) / approx. 17 kg (ANSI 150 4)
Insertion length of 2.0m: approx. 14 kg (JIS 5K 65) / approx. 19 kg (ANSI 150 4)
Insertion length of 3.0m: approx. 17 kg (JIS 5K 65) / approx. 22 kg (ANSI 150 4)
Weight:
IM 11M12A01-04E
<2. Specications>
2-4
IM 11M12A01-04E
Functions
Display Function: Displays values of the measured oxygen concentration, etc.
Alarm, Error Display: Displays alarms such as “AL-06” or errors such as “Err-01” when any such
status occurs.
Calibration Functions: Auto-calibration; Requires the Automatic Calibration Unit. It calibrates
Maintenance Functions: Can operate updated data settings in daily operation and checking. Display
Setup Functions: Initial settings suit for the plant conditions when installing the converter.
Display and setting content:
Display Related Items: Oxygen concentration (vol% O2), Output current value (mA), air ratio,
automatically at specied intervals.
Semi-auto Calibration; Requires the Automatic Calibration Unit. Input
calibration start signal by optical switch or contact, then it calibrates
automatically afterwards.
Manual Calibration; Calibration with opening/closing the valve of calibration
gas in operation interactively with the optical switch.
data settings, calibration data settings, test settings (current output loop
check, input/output contact check).
Current output data settings, alarm data settings, contact data settings, other
settings.
moisture quantity (in hot gases) (vol% H2O), Cell temperature (°C ),
thermocouple reference junction temperature (°C ), maximum/minimum/
average oxygen concentration (vol% O2), cell e.m.f. (mV), cell internal
resistance (Ω), cell condition (in four grades), heater on-time rate (%),
calibration record (ten times), time (year/month/day/hour/minute)
Calibration Setting Items: Span gas concentration (vol% O2), zero gas concentration (vol%O2),
calibration mode (auto, semi-auto, manual), calibration type and method
(zero-span calibration, zero calibration only, span calibration only),
stabilization time (min.sec), calibration time (min.sec), calibration period
(day/hour), starting time (year/month/day/hour/minute)
Output Related Items: Analog output/output mode selection, output conditions when warming-up/
maintenance/calibrating/abnormal, 4 mA/20 mA point oxygen concentration
(vol% O2), time constant, preset values when warming-up/maintenance/
calibrating/abnormal, output preset values on abnormal
Alarm Related Items: Oxygen concentration high alarm/high-high alarm limit values (vol%
Contact Related Items: Selection of contact input 1 and 2, selection of contact output 1 and
2 (abnormal, high-high alarm, high alarm, low alarm, low-low alarm,
maintenance, calibrating, range switching, warming-up, calibration gas
pressure decrease, ameout gas detection (answer-back of contact input)
O2),
<2. Specications>
Converter Output: One mA analog output point (4 to 20 mA DC (maximum load resistance of
550Ω)) with mA digital output point (HART) (minimum load resistance of
250Ω).
2-5
Range; Any setting between 0 to 5 through 0 to 100 vol%
O2 in 1 vol% O2,
and partial range is available (Maximum range value/minimum
range value 1.3 or more)
For the log output, the minimum range value is xed at 0.1 vol% O2.
4 to 20 mA DC linear or log can be selected. Input/output isolation.
Output damping; 0 to 255 seconds.
Hold/non-hold selection, preset value setting possible with hold.
Contact Output: Two points, contact capacity 30V DC 3A, 250V AC 3A (resistive load)
One of the output points can be selected to ether normally energized or normally de-energized status.
Delayed functions (0 to 255 seconds) and hysteresis function (0 to 9.9 vol%
O2) can be added to high/low alarms.
The following functions are programmable for contact outputs.
(1) Abnormal, (2) High-high alarm, (3) High alarm, (4) Low-low alarm, (5)
Low alarm, (6) Maintenance, (7) Calibration, (8) Range switching answerback, (9) Warm-up, (10) Calibration gas pressure decrease (answer-back of
contact input), (11) Flameout gas detection (answer-back of contact input).
Contact Input: Two points, voltage-free contacts
The following functions are programmable for contact inputs:
(1) Calibration gas pressure decrease alarm, (2) Range switching (switched
range is xed), (3) External calibration start, (4) Process alarm (if this signal
is received, the heater power turns off)
Self-diagnosis: Abnormal cell, abnormal cell temperature (low/high), abnormal calibration,
A/D converter abnormal, digital circuit abnormal
Calibration: Method; zero/span calibration
Calibration mode; Auto, semi-auto and manual (All are operated using
optical switches). Either zero or span can be skipped.
Zero calibration gas concentration setting range; 0.3 to 100 vol% O2 (in 0.01
vol% O2).
Span calibration gas concentration setting range; 4.5 to 100 vol% O2 (in 0.01
vol% O2).
Use nitrogen-balanced mixed gas containing 0 to 10 vol% O2 scale of
oxygen for standard zero gas and 80 to 100 vol% O2 scale of oxygen for
standard span gas.
Calibration period; date/time setting: maximum 255 days
*1 For the horizontally installed probe whose insertion length is 2.5 m or more, use the Probe Protector. Be sure to specify
ZO21R-L-200-. Specify the ange sufx code either -C or -K.
*2 The thickness of the ange depends on its dimensions.
*3 Inconel probe bolts and U shape pipe are used. Use this option for high temperature use (ranging from 600 to 700°C).
*4 Specify either /CV or /SV option code.
*5 Specify either /SCT or /PT option code.
*6 No need to specify the option codes, /CV and /SV, since the check valves are provided with the Automatic Calibration Unit.
Automatic calibration cannot be used when natural convection is selected as reference gas.
*7 Sun shield hood is still effective even if scratched. Hood is necessary for outdoor installation out of sun shield roof.
*8 Recommended if sample gas contains corrosive gas like chlorine.
*9 Piping for reference gas must be installed to supply reference gas constantly at a specied ow rate.
*10 Output signal limits: 3.8 to 20.5 mA. Specify either /C2 or /C3 option code.
- - - - - - - - - -
- - - - - - - - - -
- - - - - - - - - -
- - - - - - - - - -
- - - - - - - - - -
- - - - - - - - - -
- - - - - - - - - -
- - - - - - - - - -
- - - - - - - - - -
- - - - - - - - - -
- - - - - - - - - -
- - - - - - - - - -
- - - - - - - - - -
- - - - - - - - - -
- - - - - - - - - -
- - - - - - - - - -
- - - - - - - - - -
- - - - - - - - - -
- - - - - - - - - -
- - - - - - - - - -
- - - - - - - - - -
- - - - - - - - - -
- - - - - - - - - -
- - - - - - - - - -
- - - - - - - - - -
- - - - - - - - - -
- - - - - - - - - -
- - - - - - - - - -
- - - - - - - - - -
- - - - - - - - - -
- - - - - - - - - -
- - - - - - - - - -
- - - - - - - - - -
- - - - - - - - - -
- - - - - - - - - -
- - - - - - - - - -
- - - - - - - - - -
/C Inconel bolt (*3)
/CV
/SV
/HHood (*7)
/F1
/F2
/SCT
/PT
/C2
/C3
0.4 m
0.7 m
1.0 m
1.5 m
2.0 m
2.5 m (*1)
3.0 m (*1)
SUS316
Stainless steel with Inconel calibration gas tube (*8)
ANSI Class 150 2 RF SUS304
ANSI Class 150 3 RF SUS304
ANSI Class 150 4 RF SUS304
DIN PN10 DN50 A SUS304
DIN PN10 DN80 A SUS304
DIN PN10 DN100 A SUS304
JIS 5K 65 FF SUS304
JIS 10K 65 FF SUS304
JIS 10K 80 FF SUS304
JIS 10K 100 FF SUS304
JPI Class 150 4 RF SUS304
JPI Class 150 3 RF SUS304
Westinghouse
Not required
Horizontal mounting (*6)
Vertical mounting (*6)
Stainless steel tag plate (*5)
Printed tag plate (*5)
Failure alarm down-scale:
Output status at CPU failure and hardware error is 3.6 mA or less
(*10)
Failure alarm up-scale:
Output status at CPU failure and hardware error is 21.0 mA or more
(*10)
<2. Specications>
F11_01.ai
25
t
153 to 164
125
L
170
252 to 265
49
48.5
338 to 351
122
L= 0.4, 0.7,
1.0, 1.5, 2.0,
2.5, 3.0 (m)
4-G1/2,2-1/2NPT etc.
Cable connection port
Display side
ANSI Class 150 2 RF SUS304
ANSI Class 150 3 RF SUS304
ANSI Class 150 4 RF SUS304
DIN PN10 DN50 A SUS304
DIN PN10 DN80 A SUS304
DIN PN10 DN100 A SUS304
JIS 5K 65 FF SUS304
JIS 10K 65 FF SUS304
JIS 10K 80 FF SUS304
JIS 10K 100 FF SUS304
JPI Class 150 4 RF SUS304
JPI Class 150 3 RF SUS304
Westinghouse
Model ZR202G Integrated type Zirconia Oxygen Analyzers Unit: mm
2-7
• Standard Accessories
FuseA1113EF13.15 A
Allen wrenchL9827AB1For lock screw
ItemParts. No.Q'tyDescription
IM 11M12A01-04E
<2. Specications>
2-8
IM 11M12A01-04E
F11_02.EPS
Flange
156 ± 3
125
170
256 ± 4
49
48.5
Ø123
342 ± 4
25
t
122
Terminal side
PIPING:A
PIPING
:B
PIPING
A
B
B
A
B
B
A
A
B
B
B
B
A
ANSI Class 150 2 RF SUS304
ANSI Class 150 3 RF SUS304
ANSI Class 150 4 RF SUS304
DIN PN10 DN50 A SUS304
DIN PN10 DN80 A SUS304
DIN PN10 DN100 A SUS304
JIS 5K 65 FF SUS304
JIS 10K 65 FF SUS304
JIS 10K 80 FF SUS304
JIS 10K 100 FF SUS304
JPI Class 150 4 RF SUS304
JPI Class 150 3 RF SUS304
Westinghouse
Reference Gas Pressure: Clean air supply of sample gas pressure plus approx. 50 kPa G (or
sample gas pressure plus approx.150 kPa G when a check valve is used,
maximum pressure rating is 300 kPa) (pressure at inlet of the automatic
calibration unit)
Air Consumption: Approx. 1.5 l/min
Weight: Approx. 2 kg
Used instrument air for span calibration gas, if without instrument air is used, contact YOKOGAWA.