General
Specifications
YPC4000 is a combination Calibrator / HART® Communicator
that reduces the time spent on field calibrations and the number
of instruments that have to be carried into the field. The
YPC4000 Series delivers the versatility of a modular calibrator
and the convenience of HART® communications in one handheld
unit. Users can view up to four measurements simultaneously on
the high resolution LCD or compare HART® PV and AO values to
corresponding values measured. Available sensor modules
include a variety of differential, gauge, vacuum and absolute
pressure (for EJA/EJX) as well as temperature (for YTA. A
voltage and current meter is integral to the base unit. All HART®
field devices can be configured, polled, and trimmed using
HART® communications. Device Specific commands for
supported devices are included.
FEATURES & ADVANTAGES
USER INTERFACE – 2.6” x 2.6” display, four soft keys (definitions
change with mode requirements) and eight hard keys provide user
interface. Display is configurable for 1 to 4 simultaneous
measurements.
MODULES – select from a variety of positive lock measurement
modules for excellent versatility.
FIELD RECERTIFICATION – sensor modules and integral V / I can
be recalibrated for zero, span, & linearity through the base unit.
MEASUREMENT UNITS – customary English and Metric System
units are selectable
USER CONVENIENCE – Auto shut off, Damping, Min/Max value
capture, display Hold, sensor Zero, direct Contrast adjustment &
tare.
SECURITY LOCKOUT – YPC4000 settings can be locked out
entirely or by specific feature.
DIAGNOSTICS – YPC4000 self-tests upon power up and upon
installation of sensor modules. Advisories are posted on the
display.
FIELD UPDATES – uses Internet downloads to update and add
capability / HART® support
DEDICATED DOF MEMORY – Memory for 500 HART® profiles.
ALL-IN-ONE TOOL – takes the place of multiple handheld
instruments, simplifies workload and improves productivity
DOCUMENTING – Documenting memory and firmware for use
with DMS software. Stores up to 200 HART® configuration files
and over 100 calibration results (as-found & as-left data [up to 21
points each], time/date info, tag). Communicate with PC via DB-9
connector.
INTRINSIC SAFETY – Class I, Div I areas per CSA 22.2 & UL
913 by MET Laboratories .
OPTIONAL DMS SOFTWARE – The Device Management
System software is a full feature software tool for calibration and
configuration management of smart and conventional field
devices. DMS provides user-friendly navigation and organization
of field devices, scheduling of calibration procedures, and
storage of smart field device configuration data.
The calibration management features are extended to the
Calibration / Configuration equipment as well. Calibration
and configuration change reporting, certificates, and
device history are standard features in DMS.
SPECIFICATIONS
BASE UNIT: 9.00"L x 4.70"W x 3.58"D ABS plastic case with
shock absorbing boot, 28 ozs, three bays for measurement
modules
DISPLAY: 128 x 128 pixel graphic display, 2.6” x 2.6”
viewable area, 5-digit resolution
POWER: 6 AA alkaline or NiMH batteries; optional 100
– 240 Vac power adapter
TEMPERATURE RANGE: Storage:-40°F to 140°F (-40°C to
60°C) Operating: 23°F to 122 F (-5°C to 50°C)
CONNECTIONS: Pressure: 1/8” FNPT, 316ss Electrical:
Std. banana jacks (¾” center) for ±50 mA/Volts, HART® &
Loop leads Data transfer: DB-9 for RS-232C
NIST TRACEABLE ACCURACY:
Pressure: ± 0.025% to ± 0.10% of F.S including all effects of
linearity, repeatability, hysteresis and temperature over the
range 23° F to 122° F.
Current / Voltage: ± 0.025% of F.S. ± 0.001%
of Span per ° F (ref. temp. 70° F).
PRESSURE MODULE LIMITS / MEDIAS: GI / AI modules:
2X range / 316ss compatible DN modules: 2X range when
pressurized on high side only, 150 PSI when applied
simultaneously to high and low sides / dry gases only
MODULES: 2.3”L x 1.3”W x 2.6”D, 8.3 ozs or less. Positivelock with spring latch. Depress latch to release. Electrical
connections are PC board contacts to spring loaded receivers
on base unit. Cap plugs protect any empty bay(s).
Model YPC4000
Portable Modular Calibrator with HART®
Communications

YPC 4000 Multifunction Calibrator
YPC BASE UNIT MODEL NUMBERS (includes CE mark, standard test lead kit and six AA alkaline batteries)
Calibrator - Base unit wit three (3) modular sensor bays - see available sensor modules on next page;
use VMA0055 for 24Vdc loop supply
HART® Communicator / Calibrator - Base calibrator unit with HART® capability (includes HART® leads, and
250 ohm resistor) - see available sensor modules on next page ; use VMA0055 for 24Vdc loop supply
Yokogawa Corporation of America
Standard documenting (as-found & as-left, requires DMS database software sold separately)
CE Mark standard on all models
Intrinsic Safety certification by MET Laboratories per
C22.2 No. 157, and UL913, Fifth Edition Rev. 2/21/97 for Class I, Div I, Groups A,
Carry-Case and Strap Options
YPC/ Manual Language options
Download Subscription Options
-1 Standard 3-year DOF Download (YPC4010)
YPC4010 base calibrator with HART® - standard documenting - Intrinsic Safety – Rear hand Strap – English – Standard 3 year DOF

Sensor Modules for YPC 4000 Multifunction Tester
SENSOR MODULE MODEL NUMBER, RANGE & ACCURACY
Modules are Intrinsically Safe when used in accordance with Control Doc.
± 0.1% FS from 23° F to 122° F
± 0.1% FS from 23° F to 122° F
± 0.025% FS from 23° F to 122° F
± 0.025% FS from 23° F to 122° F
± 0.025% FS from 23° F to 122° F
± 0.025% FS from 23° F to 122° F
± 0.025% FS from 23° F to 122° F
± 0.025% FS from 23° F to 122° F
± 0.025% FS from 23° F to 122° F
± 0.025% FS from 23° F to 122° F
± 0.025% FS from 23° F to 122° F
± 0.025% FS from 23° F to 122° F
± 0.05% FS from 23° F to 122° F
± 0.025% FS from 23° F to 122° F
± 0.025% FS from 23° F to 122° F
± 0.025% FS from 23° F to 122° F
T/C measure and simulate module supports types B,C,E,J,K,N,R,S,T,M,P
V/I: meas ±55 Vdc, source 0-28Vdc; mA: meas ±100 mA dc, source 0-100 mA dc
RTD measure and simulate module supports 22 different RTDs
KIT OPTIONS
Standard kit includes: hard carrying case (P/N A36750), pressure (B34686) or vacuum (B34700) pump, low pressure connection kit (P/N
A34386).
Standard kit with B34686 pressure pump
Standard kit with B34700 vacuum pump
CARRYING CASES/HANGING STRAP
Soft case w/ shoulder strap for carrying one YPC and installed modules, additional modules, test lead kit, hand
pump & fittings.
ABS plastic case with handle and foam insert for YPC base unit and installed modules,
additional modules, test lead kit, hand pump, RTD temperature probe, fittings & tubing
YPC Hanging/Shoulder Strap, includes factory mounted D-Rings to YPC case, adjustable strap, spring-loaded
swivel snaps
961P pressure pump with integral variator, bleed valve, 1/8" FNPT connection, 0 - 145 PSIG
961V vacuum pump with integral variator, bleed valve, 1/8" FNPT connection, -650 mm Hg
Repair parts kit for pressure pump B34686 (961P)
Repair parts kit for vacuum pump B34700 (961V)

FITTING KITS & ACCESSORIES
Low Pressure Connector Kit (not for use above 250 PSIG) includes:
1/8" OD tubing (18 ft), 1/8" OD "T" fittings (2), 1/8" OD tube by 1/8" MNPT adapters (6),
3/16" barb to 1/8" FNPT (1)fitting and 1/4" MNPT to 1/8" FNPT reducers (2)
1/8" O.D. tubing to 1/8" MNPT connector (not for use above 250 PSIG)
1/8" O.D. tubing union tee (not for use above 250 PSIG)
1/4" MNPT X 1/8" FNPT reducer bushing, brass
10 ft of 1/8" O.D. nylon tubing (not for use above 250 PSIG)
Push to read equalizing manifold for DP modules, brass, 1/8"FNPT connections
High Pressure Connection Kit (not for use above 9,000 PSIG) includes:
3' HP Hose with 1pc 1/8" MNPT & 1pc 1/4" MNPT x HP Quick Connect fitting
High Pressure Connection Kit (not for use above 9,000 PSIG) includes:
1' HP Hose with 1pc 1/8" MNPT and 1pc 1/4" MNPT x HP Quick Connect fitting
5 pcs 1/4" MNPT x HP Quick Connect Fitting (not for use above 9,000 PSIG)
1 pcs 1/8" MNPT HP Quick Connect fitting (not for use above 9,000 PSIG)
ELECTRICAL / COMMUNICATIONS AND POWER ACCESSORIES
HART® Test Lead Kit - std banana plugs, mini-grabbers, 250 ohm resistor (inc'd w/ YPC 401X)
Standard Test Lead Kit - std banana plugs with assorted connectors
Std size 250 ohm shunt resistor req'd for HART® low resistance loops (inc'd w/ YPC 401X)
VMA Test Lead Kit - banana plugs on 9" breakouts both ends, assorted connectors
Adapter, standard banana jack (F) to mini-banana jack (M)
DB-9 cable, Male x Female, 6 ft length
6-pack of AA Alkaline Batteries
6-pack of AA 1.2V NiMH Batteries
NiMH battery charger, holds up to 8 individual AA cells, 8 hours, 120 VAC 60 Hz
Power adapter, 100 Vac to 240 Vac, 50-60 Hz; Universal plug adapter (New Style)
Power adapter, 100 Vac to 240 Vac, 50-60 Hz; Universal plug adapter (Old Style – Prior to 11/06)
24Vdc Power Supply, input: 120Vac 60 Hz, output: 24Vdc 350 mA
Pt 100 RTD, DIN Class "A", 0.00385 alpha, 4 lead, 1/4" O.D. 300 Series SS probe, 10"
length, 3.5" handle and 5 ft coiled cable with connector for RTD1000 temperature module
Mating connector for RTD temperature modules facilitates use of customer's RTD
User cable for RTD temp module, 5 ft cable, 4 lead wires, alligator clip terminations
T/C wire Kit 1 for types E, J, K, N, T; mini t/c connector, 1 m cable
T/C wire Kit 2 for types B, K, R/S; mini t/c connector, 1 m cable
mV measure/simulate cable assembly; mini t/c connector, 3.5 ft. copper cable, alligator clips
OPEN BAY PROTECTION MODULE (included with base YPC as required)
Dummy module protects unused (open) sensor module bays in YPC
Paper instructions for YPC 4000 (PDF instructions sent with each YPC on CD)

DOF Subscriptions for YPC 401X Multi-Function Tester with HART®
Communications
HART® communication support for the YPC4000 products and HART® smart field devices takes two forms. One is a Generic
Device Object File (DOF) shipped with every YPC unit. The Generic DOF enables Universal and Common Practice HART
®
commands to poll, write and execute analog trim and loop test functions on any HART® device. Device Specific Commands are
supported by device specific DOFs. All device specific DOFs available at the time are included with each YPC shipment. New
device specific HART® DOFs are added at a rate of 5 to 8 per month and posted on our Download Site at
http://dpcupdates.1911mptdofs.com/past.aspx
A Subscription to the Download Site provides the user with access to YPC/YHC firmware updates and all posted DOFs for a period
of 3 years after YPC/YHC shipment. Users will gain access to the Download Site by logging in with a User Name (YPC/YHC serial
number, example 950680-S01 including hyphen) and Password (initially assigned by supplier but user can enter his own after first
log in).
Why subscribe: 2 reasons
1. Products remain current. With the release of HART® 6.0 specifications, new devices and new device revisions of older products
will be released. This subscription will enable you to download DOF’s for new devices and new revisions for older devices.
Additionally, the exclusive Field Device Revision Independence (FDRI) allows YHC and YPC to automatically map configurations
from older devices to newer devices by the same manufacturer in the event of field replacement. This feature saves you valuable
downtime, and is exclusive to the YPC/YHC HART® products.
2. Firmware updates. Periodically, enhancements, improvements, and or features are made available. You may also decide to
add new options (Documentation capability for example). The firmware is flashable (i.e. field upgradeable) so that the unit never
leaves the customer's facility for an update – the result is no down time.
Discontinuous Service: Users may elect to allow their subscription to lapse. For these users, there is an increased amount over
the standard annual subscription rate.