General
Specifications
YPC4000 is a combination Calibrator / HART® Communicator
that reduces the time spent on field calibrations and the number
of instruments that have to be carried into the field. The
YPC4000 Series delivers the versatility of a modular calibrator
and the convenience of HART® communications in one handheld
unit. Users can view up to four measurements simultaneously on
the high resolution LCD or compare HART® PV and AO values to
corresponding values measured. Available sensor modules
include a variety of differential, gauge, vacuum and absolute
pressure (for EJA/EJX) as well as temperature (for YTA. A
voltage and current meter is integral to the base unit. All HART®
field devices can be configured, polled, and trimmed using
HART® communications. Device Specific commands for
supported devices are included.
FEATURES & ADVANTAGES
USER INTERFACE – 2.6” x 2.6” display, four soft keys (definitions
change with mode requirements) and eight hard keys provide user
interface. Display is configurable for 1 to 4 simultaneous
measurements.
MODULES – select from a variety of positive lock measurement
modules for excellent versatility.
FIELD RECERTIFICATION – sensor modules and integral V / I can
be recalibrated for zero, span, & linearity through the base unit.
MEASUREMENT UNITS – customary English and Metric System
units are selectable
USER CONVENIENCE – Auto shut off, Damping, Min/Max value
capture, display Hold, sensor Zero, direct Contrast adjustment &
tare.
SECURITY LOCKOUT – YPC4000 settings can be locked out
entirely or by specific feature.
DIAGNOSTICS – YPC4000 self-tests upon power up and upon
installation of sensor modules. Advisories are posted on the
display.
FIELD UPDATES – uses Internet downloads to update and add
capability / HART® support
DEDICATED DOF MEMORY – Memory for 500 HART® profiles.
ALL-IN-ONE TOOL – takes the place of multiple handheld
instruments, simplifies workload and improves productivity
DOCUMENTING – Documenting memory and firmware for use
with DMS software. Stores up to 200 HART® configuration files
and over 100 calibration results (as-found & as-left data [up to 21
points each], time/date info, tag). Communicate with PC via DB-9
connector.
INTRINSIC SAFETY – Class I, Div I areas per CSA 22.2 & UL
913 by MET Laboratories .
OPTIONAL DMS SOFTWARE – The Device Management
System software is a full feature software tool for calibration and
configuration management of smart and conventional field
devices. DMS provides user-friendly navigation and organization
of field devices, scheduling of calibration procedures, and
storage of smart field device configuration data.
The calibration management features are extended to the
Calibration / Configuration equipment as well. Calibration
and configuration change reporting, certificates, and
device history are standard features in DMS.
SPECIFICATIONS
BASE UNIT: 9.00"L x 4.70"W x 3.58"D ABS plastic case with
shock absorbing boot, 28 ozs, three bays for measurement
modules
DISPLAY: 128 x 128 pixel graphic display, 2.6” x 2.6”
viewable area, 5-digit resolution
POWER: 6 AA alkaline or NiMH batteries; optional 100
– 240 Vac power adapter
TEMPERATURE RANGE: Storage:-40°F to 140°F (-40°C to
60°C) Operating: 23°F to 122 F (-5°C to 50°C)
CONNECTIONS: Pressure: 1/8” FNPT, 316ss Electrical:
Std. banana jacks (¾” center) for ±50 mA/Volts, HART® &
Loop leads Data transfer: DB-9 for RS-232C
NIST TRACEABLE ACCURACY:
Pressure: ± 0.025% to ± 0.10% of F.S including all effects of
linearity, repeatability, hysteresis and temperature over the
range 23° F to 122° F.
Current / Voltage: ± 0.025% of F.S. ± 0.001%
of Span per ° F (ref. temp. 70° F).
PRESSURE MODULE LIMITS / MEDIAS: GI / AI modules:
2X range / 316ss compatible DN modules: 2X range when
pressurized on high side only, 150 PSI when applied
simultaneously to high and low sides / dry gases only
MODULES: 2.3”L x 1.3”W x 2.6”D, 8.3 ozs or less. Positivelock with spring latch. Depress latch to release. Electrical
connections are PC board contacts to spring loaded receivers
on base unit. Cap plugs protect any empty bay(s).