Yokogawa XL100 Data sheet

Portable Data Station XL100
Network-Based 8-/16-Channel Data Logger Robust and Easy to Operate
Patent Pending
T M
World First Instrument
To Obtain IPv6 Ready Logo Phase-1 !
ID : 01-000273
Special Prize
INTEROP T oky o 2005
Best of Show Award
Won Special Facility Product Prize!
LF XL100-E
Supports inter-channel calculation, scaling, statistical operation (maximum value, minimum value, average value, root-mean-square value, peak value).
Alarm can be output for up to 4 channels, and the alarm conditions can be set for each channel. Alarm status can be checked from a distance via E-mail delivery.
Allows setting of temperature (thermocouple, resistance temperature detector) and voltage for each channel. Each input channel is insulated plus the unit is equipped with pulse input (1 channel) and logic input (2 channels).
Allows data saving for 8 channels at 100ms (maximum speed). Data can be saved in either binary or ASCII (can be imported to Excel*) format. Binary data can be converted to CSV data using the software supplied with the unit.
*Excel is a registered trademark of Microsoft Corporation.
In addition to the internal memory (16MB), the unit can use an external storage medium such as a compact flash memory card and SD card. (Max. 512MB) The data can also be copied to a USB memory, making data transfer easy after measurement.
Compact
Compact
Compact
Size
Size
Size
External dimensions: 155(W)3155(H)3355 (D) mm Weight: Approx. 800g
Space
Space
Space
Saving
Saving
Saving
*Only compact flash memory cards, SD cards and USB memories that have been verified by YOKOGAWA can be used.
The unit can be used in a limited space since wiring is only required on the sides of the unit.
Introduction of TFT LCD with Wide Viewing Angle
Use of a TFT LCD with wide viewing angle reduces eye strain during use. Various display contents (waveform, digital, bar graph, waveform + digital) can be switched from one to another easily.
Displays data for up to 8 channels in one screen. Up to four groups be displayed by switching from one group to another.
(32 channels)
Detachable T erminal Block
The terminal block can be detached in one single operation making wiring work easy.
can
Protection on LCD
Resistance to impact can be improved by use of rubber boot (standard accessory). The rubber boot is detachable.
Shown with the rubber boot attached
Entry of dust can be prevented by using the side cover.
The unit can be operated in hot field sites. The allowed operating humidity range is 5 to 85% RH.
Supports pre-trigger and trigger delay. Pre-trigger starts data collection before a trigger is caused, and trigger delay starts data collection after a trigger is caused. (Settable number of triggers is 0 to 600 for both.) Single mode (only once) or continuous mode (each time a trigger is caused) can be selected.
Auto-Set & auto--run Function can be performed by simply installing an external storage medium (compact flash memory card, SD card, USB memory). If an automatic measurement setting file is stored in the external storage medium, the data of the file can be read automatically to start measurement.
Displays the collected data in waveform. It also lets you create a setting file and convert binary data to CSV data so that the data can be processed using Excel *.
* Excel is a registered trademark of Microsoft Corporation.
OfficeField
Ethernet
Use of Ethernet (10BASE-T, 100BASE-TX) supports standard protocols, such as HTTP, FTP, SMTP and SNTP, eliminating the need for special software. In addition, to prevent illegal access to the unit via the network, user authentication to allow only the registered users to access the unit can be set. Up to 7 persons (1 administrator, 6 users) can be set.
Web monitoring
This is a screen example and may differ from the actual one.
Use of the HTTP protocol allows monitoring the main unit's screen in a Web browser and operating the unit. The screen can be refreshed automatically (2, 5, 10, 30 seconds) or manually. The screen has two pages: the monitor page to monitor the unit's screen, and the operator page to operate the unit. To prevent illegal access, user authentication can be set for each of these pages.
This function allows you to access the unit from PC via FTP protocol, output a list of files stored in the internal memory or external storage medium and transfer/delete the files.
This function allows transfer of the measurement data file, alarm data file and log file created in the internal memory or external storage medium to the specified FTP server. Two FTP servers (primary and secondary) can be set as the file transfer destination, and the files are transferred to the secondary FTP server in the case of failure with the primary FTP server.
This function delivers E-mail when an alarm occurs, when alarm is cleared, when power is restored from power failure, when a medium related error or FTP client related error occurs, or when the specified time is reached. This function can be set for up to two groups, and the delivery destination can be specified for each group.
Te r minal block (
for 16 channels
)
Simple Operation
Measurement (free running, logging) review (to check measured data), file management and setting can be performed using direct keys. Operations can be performed with one hand. Furthermore, guidance for setting items is displayed on the screen to reduce inconvenience of referring to the instruction manual.
Built-in terminal screwdriver
Built-in rechargeable battery
External inputs
/outputs
External storage
medium RS-232 RS-485
LAN USB
Use of a lithium ion battery allows continuous measurement for up to 8 hours.* The battery can be charged during measurement. The battery can also be used as backup power in case of power failure.
*When operated under YOKOGAWA's recommended conditions
• Built-in ModBus protocol
For serial communication (RS-232, RS-485), ModBus RTU and ModBus ASCII protocols are provided as standard features in addition to the dedicated protocol. They allow the unit to be connected to another device to collect data during temperature/voltage measurement. Dedicated 32 communication channels are provided.
Connection example
Tempera-
ture
Analog
Connecting to Yokogawa's CW120 series clamp power meter enables collection of temperature data and current/voltage data, providing ultimate energy saving solution. Two or more power meters can be connected. (Max: 31 units)
RS-485
Clamp-on power meter
(CW120 series)
Allows one-to-one connection with PC. This function allows you to collect memory data and change settings made to the unit. RS-232-compatible printer (97010) is available to print data to paper.
Optional printer (97010)
Item
Shape of input terminal Numbers of input Measurement interval Resolution of A/D converter Average function Each mode Measurement screen display Input item
Alarm output
Trigger function
Computing function
Communication function
Data saving
Additional function Indicator Power supply Operation Temp. & RH range Outside dimension & Weight Supplied Standard Accessories
Specification
Details
Push-in type terminal 8ch or 16ch (Channel-to-Channel isolation, terminal "b" is shared by resistance temperature detector inputs) 100 ms (for 8-channel type only), 200 ms, 500 ms, 1 s, 2 s, 5 s, 10 s, 20 s, 30 s, 1 min, 2 min, 5 min, 10 min, 20 min, 30 min, 1 h 16 bits Moving average processing and can be set for each channel (1, 2, 5, 10, and 20 times) Free running, Logging, Review, Logging + Review, File processing, setting Waveform, Digital, Bar graph, Waveform + Digital, Alarm summary, Log
DCV TC (without reference junction guarantee accuracy) RTD (3 wire) Reference junction guarantee accuracy Maximum input voltage Logic (DI) (2ch)
Pulse (1ch) Number of output: 4ch, Type: Level (Hi/Lo), window (IN/OUT) / Output form: Open collector / Output hold form: Hold/Not Hold Alarm sound: ON/OFF / Output update:every measuring period / Delay time of delay alarm: For each channel, select from 1 s to 1 h Setting:Pre trigger, Trigger delay Condition:Non, External, Level (Leading edge/Falling edge), Window (IN/OUT), Logic (Hi), Alarm, Time, Timer ((timer can only be used to stop logging)) Mode: One time, Repeat Timing Start, Stop, Start/Stop (Can be set condition separately) Differential computation Linear scaling
Statistical calculation LAN
USB1.1 (1 port), RS-232, RS-485 (Protocol: Dedicated protocol, ModBus RTU, ModBus ASCII) Data saving media: Internal memory (16 MB), CF card, SD card, USB memory (USB memory copies function only) / Data saving method: Manual or Logging data Stored data: Instantaneous value (Measured value/Calculated value) / Data format: Binary or ASCII Other possible data saving: Image data (BMP form), Setting data (CSV form), Alarm data (CSV form), Log data (CSV form) Auto setup & auto run by external memory, Auto battery change when instantaneous power failure occurred, Key lock, key login, Clock function
3.5-inch TFT color LCD (320240 pixels) AC100 to 240 V±10% (50/60 Hz), Lithium ion battery (Approx. 8 hours, when operated under the YOKOGAWA-recommended conditions) 0 to 50˚C, 5 to 85% RH (no condensation) Approx. 155 (W)155 (H)55 (D), Approx. 800 g (Without battery and rubber boot) Safety: EN61010-1, Emission: EN61326 class A, EN55011 class A Group1, EN61000-3-2, EN61000-3-3, Immunity: EN61326 Annex A Terminal block unit (8ch or 16ch), AC adapter, Quick manual, CD-ROM (Standard software, Instruction manual, Communication function manual, Quick manual), rubber boot, Screw driver (for terminal block)
100 mV, 500 mV, 1 V, 5 V, 10 V, 30 V, 1-5 V f.s. K, E, J, T, L, U, N W, R, S, B Pt 100/JPt 100 ±1˚C DC30 V Voltage input Contact input Instantaneous value, Integration, revolution
Between any two channels Possible calculation range: DCV, TC, RTD, Pluse / Scaling possible range: -30000 to 30000 Decimal point position: Selectable / Unit: Possible for optional setting (Maximum 6 characters) /each item (e.g.length, area) Displays value from logging start to until now. Calculation item: Max. value, Min. value, AVE value, Peak value, RMS value Media: Ethernet (10BASE-T/100BASE-TX), Protocol: TCP/IP, FTP, HTTP, SMTP, SNTP Function: Web server function, FTP client function, FTP server function, Real time monitoring function, E-mail delivery function, Time adjustment function
±0.1% of f.s. ±0.05% of f.s. ±1˚C ±0.05% of f.s. ±2˚C ±0.05% of f.s. ±0.5˚C
OFF: Less than 0.9 V, ON: Over 2.1 V Contact ON/OFF
Max. ResolutionAccuracyRangeItem
10V (for 100 mV range)
0.1˚C 1˚C
0.1˚C
Optional accessories and Spares
Optional accessories
Spares
Name
Type-K TC Carrying case Lithium ion battery Digital I/O cable Communication cable Printer cable Printer Printer thermal paper AC adapter for printer AC adapter for printer Terminal block unit (16ch) Terminal block unit (8ch) Rubber boot AC adapter (Suffix code)
Model No.
90060 93037 94009 91029 91011 91010 97010 97080 94006 94007 95050 95051 93036 94010
-D
-F
-H
-R
-S
5 meter 4 sets To store the main unit and accessories 2,400 mAh, 7.4 V For pulse/logic inputs and alarm outputs, 3 m RS-232 communication cable for PC (9 pin) RS-232 cable for printer Includes 1 roll thermal paper and 1 battery pack 10 rolls/set Power supply 200-240 V Power supply 100-120 V 8ch type 16ch type For protecting impact Power cord For UL/CSA Standard For VDE Standard For GB Standard For SAA Standard For BS Standard
Description
Model number and suffix code
Suffix codeModel
-1
-2
XL100
Related products (Connectable to Datum-Y)
Clamp on power meter
CW120 series
RS - 485 ModBus
World Wide Web site at
http://www.yokogawa.com/MCC
YOKOGAWA METERS & INSTRUMENTS CORPORATION International Sales Dept.
YOKOGAWA CORPORATION OF AMERICA (U.S.A.) Phone: +1-770-253-7000 Facsimile: +1-770-251-2088 YOKOGAWA EUROPE B. V. (THE NETHERLANDS) Phone: +31-334-64-1611 Facsimile: +31-334-64-1610 YOKOGAWA ENGINEERING ASIA PTE. LTD. (SINGAPORE) YOKOGAWA AMERICA DO SUL LTDA (BRAZIL) Phone: +55-11-5681-2400 Facsimile: YOKOGAWA MEASURING INSTRUMENTS KOREA CORPORATION (KOREA) YOKOGAWA AUSTRALIA PTY. LTD. (AUSTRALIA) Phone: +61-2-8870-1100 Facsimile: +61-2-8870-1111 YOKOGAWA INDIA LTD. (INDIA) Phone: +91-80-2227-1513 Facsimile: +91-80-2227-4270 YOKOGAWA SHANGHAI TRADING CO., LTD. (CHINA) Phone: +86-21-6880-8107 Facsimile: +86-21-6880-4987 YOKOGAWA MIDDLE EAST E.C. (BAHRAIN) Phone: +973-358100 Facsimile: +973-336100 YOKOGAWA ELECTRIC CIS LTD. (RUSSIAN FEDERATION)
2-9-32 Nakacho, Musashino-shi, Tokyo, 180-8750 Japan Phone: +81-422-52-3530 Facsimile: +81-422-55-8954
Phone: +65-6241-9933 Facsimile: +65-6241-2606
Phone: +82-2-551-0660 Facsimile: +82-2-551-0665
+55-11-5681-1274/4434
Phone: +7-095-737-7868 Facsimile: +7-095-737-7869
With 8ch terminal block unit With 16ch terminal block unit
-D
Power cord (UL/CSA Standard)
-F
Power cord (VDE Standard)
-H
Power cord (GB Standard)
-R
Power cord (SAA Standard)
-S
Power cord (BS Standard)
Digital illumination
meter 510 series
analog 0 to 1V
Clamp tester
CL series
analog
0 to 1V
NOTICE
Before using the product, read the instruction manual
carefully to ensure proper and safe operation.
Represented by:
Specification
MIK-ES14
Subject to change without notice.
All Rights Reserved. Copyright © 2005, Yokogawa Meters & Instruments Corporation.