Yokogawa WT210 Data Sheet

ENERGY SAVING TOOLS
Digital Sampling Power Meters
with Superior Cost Performance
Digital Power Meters
WT210
Basic power accuracy: 0.1% ● DC measurement, 0.5 Hz to 100 kHz power frequency range
5 mA range for very low current measurements (model WT210 only)
Line filter function High-speed data update (as fast as 10 readings per second)
Harmonic measurement function available
User calibration capability
/
WT230
www.yokogawa.com/tm/
... and subscribe to “Newswave,” our free e-mail newsletter
Bulletin 7604-00E
The WT230’s advanced specifications and its wide range of functions let you handle all your measurement applications from low-frequency equipment to high frequency inverters using a single power meter. One unit also handles standby low-power measurements and rated-power measurements
(functions available with the WT210 only).
WT230 WT210
Wide range of 5 mA to 20 A
The built-in 5 mA range lets you measure currents as low as 25 µA. This makes it possible to measure very low currents on such things as intermittent control equipment. The wide current range (5 mA to 20 A) means a single power meter can be used for applications such as Energy Star® measuremnts, to measure everything from standby-power to rated-power.
Fr ee
Soft ware Soft ware
WTViewer for the WT210/WT230
Easily Acquire and Manage Power
Measurement Data from Your PC
See 8 pages or
Bulletin 7604- 32E for details.
NEW
Functions and Features of the WT210 and WT230
A Wide Frequency Range Lets You Work on a Variety of Different Applications
Low-frequency Equipment Commercial Power Supplies
NEW
NEW
Low-frequency measurements starting at 0.5 Hz
Low-frequency measurements starting at 0.5 Hz can be used with evaluations of cycloconverter and when a motor are started.
Accuracy Is Assured between 1% and 130%
960 01 751552
NEW
NEW
Power accuracy is even better than in former WT series.
0.1%
1: Maximum display is 140% of the rated input. 2: Conditions apply to accuracy from 110% to 130%.
NEW
NEW
Now you can obtain more precise measurements on high-frequency equipment such as inverters.
With 960 01 Max. 400 Arms With 751552 Max. 1000 Arms
Capture a Variety of Signal Types
Surge current and maximum load state
NEW
MAX hold function for voltage, current, and power
This function lets you keep, on the display, voltage and current peak values, voltage and current rms values, and maximum values for active power, apparent power, and reactive power.
3
NEW
Quick response with display updating as fast as every 0.1 second
With measurement intervals as short as 0.1 second, you can capture transient phenomena with a fine level of detail. You can also reduce the time per measurement for increased through put in production testing.
Half-wave Rectification, Intermittent Control, Distortion Waves
NEW
Measurement of DC components
In addition to using DC inputs, you can obtain precise measurements of signals containing DC components, such as intermittent signals and half­wave rectification signals.
Instead of taking notes, you can use the internal memory to store and recall settings and field measurement data.
Information on the features and functions of Yokogawa's WT210, WT230, accessories, and related products is also available at our web site. http://www.yokogawa.com/tm/
NEW
This function lets you measure fundamental wave rms values for inverter output voltages.
Constantly changing signals
Noisy Signals
Line filter function (fc = 500 Hz)
Inverters
100 kHz power frequency range
130% input1% input
26A
Powerful Tools for Energy Measurement
Extended Energy Measurement Applications
Maximum integration time: 10,000 hours
Time can be set between 1 second and 10,000 hours (416 days) in 1­second increments.
3
Battery equipment applications
Integrating power measurement by polarity
Power and current values can be integrated separately for positive and negative polarities. Integrated values are shown with the decimal point
12
moving according to the integrated value.
Applications for a Variety of Add-on Options
Large-current Measurement Using Current Clamps
External input for current sensor
Select either 50/100/200 mV or 2.5/5/10 V. A current clamp lets you measure currents without needing to disconnect the power supply circuit wiring. 4: Please select /EX1 (2.5/ 5/10 V) option when you use 960 01.
4
Online Power Meter Control and Recording
GP-IB/serial (RS-232-C) interface
This option lets you control the power meter through a PC, or save data to a PC.
GP-IB/serial interface (RS-232-C)
Clamp probe
External input
D/A output
Recording to a Recorder
D/A output
This option lets you output a variety of measurement data, such as voltage, current, and power measurements, with ±5 V rating, for recording on a recorder. The recorder can then be used to check changes in data over time.
Recorder
Intermittent Control Equipment Applications
Average active power display
The power of intermittent control equipment changes significantly over time. The average active power in intermittent operations can be displayed, which is highly effective for consumed-power measurements.
Average active power (W) =
Integrated power (Wh)
Integrated elapsed time (Hours)
3
Instantaneous power
Power
Average power
Time
Power Supply Harmonic Measurements
Calculate voltage, current, reactive power, content ratio, and phase angle relative to fundamental frequency for up to 50 orders. This option is well-suited to power supply environment evaluations. Measurement time is approximately 90% shorter than in former models.
Comparator output
GO/NO-GO Evaluations on Testing Lines
4-channel comparator function
A 4-channel relay contact output (normal-open and normal-close pair) lets you do GO/NO-GO evaluations on production and testing lines.
3: Popular functions on the WT200 were incorporated into WT210 and WT230.
32
Basic Characteristics
Example of Frequency-power Accuracy Characteristics
10
WT230 15 V/1 A range
5
0
Tolerance (% of range)
-5
-10
0.1 1 10 100 1000 10000 100000
Spec and reference values
Frequency (Hz)
Example of WT210 Current Accuracy
10
8
6
4
2
0
-2
-4
-6
-8
Indicated value tolerance (% of reading)
-10
0.1 1 10 100
Input level relative to set range (% of range)
100 mA range: 60 Hz input
Current Input Surge Withstanding Ability Example of Influence of Common Mode Voltage
350
300
250
200
150
100
Surge current value (Arms)
50
0
0.01 0.1 1 10 100
5 mA-20 mA range
0.5 A-20 A range
Input current time width (seconds)
5
4
3
2
1
Tolerance (% of range)
0
-1
0.1 1 10 100 1000 10000 100000
WT230 150 V range WT230 500 mA range Spec and reference values
Frequency (Hz)
Example of D/A Output Response Comparison with Former Models
WT200/WT130 WT210/WT230
0.25% of rng 0.2% of rng
0.35% of rng (WT130)
5 digits (WT200)
Approximately 3 seconds 0.25/0.5/1/2/5 seconds
EXT STOP, EXT RESET, and INTEG BUSY are not added.
No Yes (need /HRM) Yes No
4 digits 5 digits
All six signals listed to the left are added.
display update time
Voltage input terminal Binding post Plug-in terminal (safety terminal)
External input terminal Plug-in terminal (safety terminal) BNC
Voltage and current
basic accuracy
Power basic accuracy 0.3% of rng (WT200)
Frequency range DC, 10 Hz to 20 kHz DC, 0.5 Hz to 100 kHz Assured accuracy range 10% to 130% of range rating 1% to 130% of range rating Display updating interval 0.25 second (fixed) 0.1/0.25/0.5/1/2/5 seconds
V, A, W display digits 4 digits (WT130)
Line filter function No Yes (fc = 500 Hz) Frequency filter function Yes (fc = 300 Hz) Yes (fc = 500 Hz)
Key lock No Yes
Harmonic measurement display
updating interval
Remote signals when EXT HOLD and EXT TRIG are added. EXT START,
comparator is installed
Online data format ASCII ASCII, binary
Waveform data communications output
Addressable mode B for
GP-IB communications
Display digits (factory default) 4 digits 5 digits
Online output data digits (factory default)
Functions Included with the WT200 (but Not Included with the WT130) and Included with the WT210/WT230
MAX hold function
10,000-hour maximum integration time • Integration with few data omissions • Average active power display
Moving decimal point display based on integrated power value
0.2% of rng
5 digits
Pin assign is changed.
4
WT230 WT2 10
Parameter Voltage
15/30/60/150/300/600 V
Input resistance: Approximately 2 M Input capacitance: Approximately 13 pF
Peak voltage of 2.8 kV or rms value of 2.0 kV (whichever is less)
Peak voltage of 2.0 kV or rms value of 1.5 kV (whichever is less)
Peak voltage of 1.5 kV or rms value of 1.0 kV (whichever is less)
600 Vrms (with output connector protective cover),
CAT II / 400 Vrms (without output connector protective cover) CAT II
50/60 Hz, -80 dB or higher (±0.01% of range or less) with voltage input terminals shorted and current input terminals open and external input terminals shorted Reference value (up to 100 kHz): ±((Maximum range rating)/(Range rating) × 0.001 × f% of rng) or less (voltage range and 0.5-20 A current range and external input range
3
) ±((Maximum range rating)/(Range rating) × 0.0002 × f% of rng) or less (WT210; 5-200 mA range) Note: 0.01% or higher. f is in kHz. 3 Decuple the above-formula about the external input range.
Resistance voltage divider
Current
Shunt input system Direct input: 5/10/20/50/100/200 mA (WT210 only)
1
; 0.5/1/2/5/10/20 A (WT210/WT230) External input (optional): 2.5/5/10 V or 50/100/200 mV
Direct input: Approximately 500 m + approximately 0.1 µH (5-200 mA; WT210)
Approximately 6 m + 10 m (max)2 + approximately 0.1 µH (0.5-20 A; WT210) Approximately 6 m approximately 0.1 µH (0.5-20 A; WT230)
External input:
Approximately 100 k (2.5/5/10 V), approximately 20 k (50/100/200 mV)
0.5-20 A (WT210/WT230): Peak current of 450 A or rms value of 300 A (whichever is less) 5-200 mA (WT210): Peak current of 150 A or rms value of 100 A (whichever is less) External input: Peak value of 10 times range or less
0.5-20 A (WT210/WT230): Peak current of 150 A or rms value of 40 A (whichever is less) 5-200 mA (WT210): Peak current of 30 A or rms value of 20 A (whichever is less) External input: Peak value of 10 times range or less
0.5-20 A (WT210/WT230): Peak current of 100 A or rms value of 30 A (whichever is less) 5-200 mA (WT210): Peak current of 30 A or rms value of 20 A (whichever is less) External input: Peak value of 5 times range or less
Input type
Rated values (ranges)
Measuring instrument loss (input resistance)
Maximum instantaneous allowed input (1 cycle, 20 ms duration)
Maximum instantaneous allowed input (1 second duration)
Maximum continuous allowed input
Maximum continuous common mode voltage (with 50/60 Hz input) CMRR 600 Vrms across input terminal and case
Floating input
Input Specifications
Plug-in terminal (safety terminal) Direct input: Large binding post
External input: BNC connector (insulation type)
Input terminal type
A/D converter Simultaneous conversion of voltage and current inputs
Resolution: 16 bits Maximum conversion speed: Approximately 20 µs (approximately 51 kHz)
Range switching Ranges can be set manually, automatically, or through online controls.
Auto-range function Range raising: When a measurement exceeds 130% of the rating, or when the peak value exceeds approximately 300% of the rating Range lowering: When a measurement falls to 30% or less of the rating, and the peak value falls to approximately 300% or less of the rating for the low range
Measurement mode switching Any of the following, selected manually or through online controls: RMS (true rms value measurements for both voltage and current), V MEAN (calibration of
average-value-rectified rms value for voltage; true rms value measurement for current), DC (simple averages for both voltage and current)
Note: Current direct input and external sensor input cannot both be used at the same time. When you operate current input terminals and external input terminals, please be careful. Since these terminals are electrically connected inside the instrument. 1, Connect wires that match the size of the measurement current. 2, Factory setting
Parameter Voltage/current
DC:
0.5 Hz f < 45 Hz: 45 Hz f 66 Hz: 66 Hz < f 1 kHz: 1 kHz < f 10 kHz:
10 kHz < f 100 kHz:
* Add ±10 µA to the current DC accuracy.
1-130% of voltage/current range rating (for accuracy at 110-130%, add the reading tolerance × 0.5 to the above accuracy) Add the accuracy's reading tolerance (three months after calibration) × 0.5 to the accuracy three months after calibration. A low-pass filter can be inserted in the input circuit for measurement. The cutoff frequency (fc) is 500 Hz. Voltage and current: Add 0.2% of rdg at 45-66 Hz. Add 0.5% of rdg below 45 Hz. Power: Add 0.3% of rdg at 45-66 Hz. Add 1% of rdg below 45 Hz. ±0.03% of range/°C at 5-18°C and 28-40°C.
0.1/0.25/0.5/1/2/5 seconds
Data updating rate 0.1 second 0.25 second 0.5 second 1 second 2 seconds 5 seconds Measurement lower limit frequency
25 Hz 10 Hz 5 Hz 2.5 Hz 1.5 Hz 0.5 Hz
±(0.2% or rdg + 0.2% of rng)* ±(0.1% of rdg + 0.2% of rng) ±(0.1% of rdg + 0.1% of rng) ±(0.1% of rdg + 0.2% of rng) ±((0.07 × f)% of rdg + 0.3% of rng)
±((0.5% of rdg + 0.5% of rng)
±((0.04 × (f-10))% of rdg)
Active power
DC:
0.5 Hz f < 45 Hz: 45 Hz f 66 Hz: 66 Hz < f 1 kHz: 1 kHz < f 10 kHz:
10 kHz < f 100 kHz:
* Add ±10 µA
×
voltage reading to the power DC accuracy. For cosϕ = 0 45 Hz f 66 Hz: ±0.2% of VA (VA is a reading value of apparent power) Reference data (up to 100 kHz): ±((0.2 + 0.2 × f)% of VA) Indicated value tolerance for 0 < cosϕ < 1 Add (tanϕ × (effect when cosϕ = 0)% of power reading to the above power accuracy. Note: ϕ is the phase angle between voltage and current.
±(0.3% or rdg + 0.2% of rng)* ±(0.3% of rdg + 0.2% of rng) ±(0.1% of rdg + 0.1% of rng) ±(0.2% of rdg + 0.2% of rng) ±(0.1% of rdg + 0.3% of rng) ±((0.067 × (f-1))% of rdg) ±(0.5% of rdg + 0.5% of rng)
±((0.09 × (f-10))% of rdg)
System Frequency range Crest factor
Accuracy (three months after calibration)
(Conditions)
Temperature: 23±5°C Humidity: 30-75% RH Input waveform: Sinewave
Power factor: cosϕ = 1
In-phase voltage: 0 V DC Frequency filter: ON at 200 Hz or less Scaling: OFF Display digits: 5 digits After CAL is executed
Note: In the accuracy calculation formula, f is in kHz. Power factor effect
Note: In the accuracy calculation formula, f is in kHz.
Effective input range Accuracy (12 months after calibration) Line filter function Accuracy with line filter on
Temperature coefficient Display updating intervals
Measurement lower limit frequency
Digital sampling; sum of averages method
DC, and 0.5 Hz to 100 kHz
3 (with rated input) 300 (with minimum effective input)
Measurement Functions
rng: Range rdg: Reading
Lead/lag is detected correctly when phase difference equal to or greater than ±5° with both voltage and current inputs as sine waves equal to or greater than 50% of rated range-value, and the frequency is between 20 Hz to 2 kHz.
Lead/lag detecting
Specifications
The latest product information is available at our web site http://www.yokogawa.com/tm/. Review the specifications to determine which model is right for you.
Frequency Measurements
Measurement inputs Measurement system Measurement frequency ranges
Accuracy: ±(0.06% of rdg) Conditions: Input equal to at least 30% of voltage/current rated range.
: V1, V2, V3, A1, A2, or A3 (select one) : Reciprocal system
100 ms: 25 Hz f 100 kHz 250 ms: 10 Hz f 100 kHz 500 ms: 5 Hz f ≤ 100 kHz 1 sec: 2.5 Hz f ≤ 100 kHz
2.5 sec: 1.5 Hz f ≤ 50 kHz 5 sec: 0.5 Hz f ≤ 20 kHz
Frequency filter function ON at 200 Hz and below. Frequency filter cutoff frequency: 500 Hz
Communication Functions (Optional for the WT210)
GP-IB or serial interface (RS-232-C) (select one) GP-IB
Electrical and mechanical specifications: Functional specifications: Protocol: Conforms to IEEE Standard 488.2-1992.
Code used: ISO (ASCII) code Addresses: 0-30 talker/listener addresses can be set.
Serial interface (RS-232-C)
Transmission mode Baud rates: 1200, 2400, 4800, 9600 bps
Conform to IEEE Standard 488-1978 (JIS C1901-1987). SH1, AH1, T5, L4, SR1, RL1, PR0, DC1, DT1, C0
: Asynchronous
5
Specifications
Calculation Functions Internal Memory Functions
Single-
Three-phase 3-wire
phase 3-
(2 voltages,
wire Voltage V (V1 + V3)/2 Current A Active power W Reactive
power var, var
Apparent power VA, VA
Power factor PF, PF
Phase angle deg, deg
Notes
1. This equipment's apparent power (VA), reactive power (var), power factor (PF), and phase angle (deg) are calculated from voltage, current, and active power. (Therefore, if the input contains a distorted wave, the v alues may not match those of other measuring instruments based on different measurement principles.)
2.
If either voltage or current falls to 0.5% of the range rating or less , then the apparent power (VA) and reactiv e power (v ar) are display ed as zero , and errors are displayed for power factor (PF) and phase angle (deg).
3. The sign of the var of each phase is displayed with +(positive). In the ∑var calculation, the var value for each phase is calculated with a negative sign if the current input leads the voltage input, and with a positive sign if the current input lags the voltage input. Then the value of ∑ var may be display ed with –(negativ e).
4. Apparent power (V A) and reactive power (v ar) cannot be calculated and displa yed at the harmonics measurement mode.
vari = (VA2 - W2)
VAi = Vi × Ai
Pfi = Wi/VAi W/VA
degi =
-1
cos
(Wi/VAi)
(A1 + A3)/2 W1 + W3
var1 + var3
VA1 + VA3
-1
cos
(W/VA)
2 currents)
2
3
(VA1 + VA3)
Three-phase 3-wire (3 voltages, 3 currents)
(V1 + V2 + V3)/3 (A1 + A2 + A3)/3
3
(VA1 + VA2 + VA3)
3
Three­phase 4­wire
W1 + W2 + W3
var1 + var2 + var3
VA1 + VA2 + V A3
Display Functions
Display unit: 7-segment LED (light-emitting diode) Display areas: 3
Display area
A
V, A, W, VA, var (for each element), integration elapsed time
B
V, A, W, PF, deg (for each element, percentage (content percentage, THD)
C
V, A, W, V/AHz, Vpk, Apk, ±Wh, ±Ah (for each element), MATH
Measurement parameters
V, A, W, VA, var PF deg ±Wh, ±Ah VHz, AHz
Display digits: 4 or 5 digits (selectable by user). Factory default setting is 5 digits.
Units: m, k, M, V, A, W, VA, var, Hz, h±, deg, % Display updating intervals: 0.1/0.25/0.5/1/2/5 seconds Response time: Maximum 2 times the display updating interval (time required
Maximum display: 140% of voltage/current range rating Minimum display: About Vrms, Arms, and Ah, 0.5% of range rating.
Display scaling function
Effective digits: Selected automatically according to the digits in the voltage and Setting range: 0.001 to 9999
Averaging function
There are two averaging methods (selectable by user): Exponential average Moving average
Auto-range monitor
An LED turns on when the input value is outside the range set for the auto-range.
MAX hold function
This function can be used to hold V, A, W, VA, var, Vpk, and Apk at maxim um v alues .
MATH functions System: When a function key on DISPLAY C is pressed to select the
Maximum display
99999
±1.0000
±180.0
999999
99999
for display value to enter accuracy range of final v alue with line filter off, when range rating abruptly changes from 0% to 100%, and from 100% to 0%)
Less than 0.5% is zero suppression.
current ranges.
In cases where response can be set and exponential average is used, the atten uation constant can be selected. In cases where a moving av erage is used, the number of averages N can be selected from 8, 16, 32, and 64.
MATH functions , it is possible to perf orm efficiency (WT230 only) and input crest factor measurements, as well as arithmetic calculations on DISPLAY A and B measurements. In addition, it is possible to display average active power for time-converted integrated power.
Displayed information
Display resolution
0.001%
0.01%
0.1*
0.0001%
Input frequency/20,000
Integration Functions
Display resolution: The minimum display resolution changes together with the Maximum display: -99999 to 999999 MWh/MAh
Modes: Standard integration mode (timer mode), continuous integration Timer: Automatic integration start/stop based on timer setting.
Count over flow: When the integrated value exceeds 999999 MWh/MAh or falls
Accuracy: ±(display accuracy + 0.1% of rdg) Timer accuracy: ±0.02% Remote control: Starting, stopping, and resetting can be controlled through
integrated value.
mode (repeat mode), manual integration mode Setting range: 000 h:00 min:00 sec to 10000 h:00 min:00 sec
(If the time is set to zero, manual mode is automatically set.) to at least -99999 MWh/MAh, the elapsed time is saved and the
operation is stopped.
external contact signals. This function is only available when option /DA4, /DA12 or /CMP is installed.
Measurement data
Stored data WT210 (760401) WT230 (760502) WT230 (760503)
Store interval: Display updating interval and 1 second to 99 hours, 59 minutes, Recall interval: Display updating interval and 1 second to 99 hours, 59 minutes,
Panel setting information
Normal measurement Data for 600 samples Data for 300 samples Data for 200 samples
Harmonic measurement
Data for 30 samples Data for 30 samples Data for 30 samples
and 59 seconds and 59 seconds
(Both can be set in 1-second increments.)
:Four different patterns of panel setting information can be written/
read.
Harmonic Measurement Function (optional)
System: PLL synchronization Measurement frequency range:
Maximum display: 99999 Display digits: 4 or 5 digits (selectable by user).
Measurement parameters: V, A, W, deg (WT210), V1, V2, V3, A1, A2, A3, W1, W2,
Measurement element: These parameters can only be measured simultaneously for
Sampling speed, window width, and analysis orders
The values for these parameters vary according to the input fundamental frequency as shown below.
Fundamental frequency Sampling speed Window width Analysis orders
40 ≤ f < 70 Hz f × 512 Hz 2 periods of f 50 70 ≤ f < 130 Hz f × 256 Hz 4 periods of f 50 130 f < 250 Hz f × 128 Hz 8 periods of f 50
250 f 440 Hz f × 64 Hz 16 periods of f 30 FFT data length: 1024 FFT processed word length: 32 bits Window function: Rectangular Display updating interval:
Accuracy: Add ±0.2% of range to normal measurement accuracy.
Fundamental frequency in range of 40-440 Hz
Factory default setting is 5 digits. W3, deg1, deg2, deg3 (WT230), individual harmonic levels, rms
voltage, rms current, active power, fundamental frequency PF, harmonic distortion rate, individual harmonic content
a single specified input element.
0.25/0.5/1/2/5 seconds Updating is slower during online output according to the communication speed and the number of parameters transferred.
Note: For nth-order component input, add ((nth order reading) × (10/(m+1))%) to the n+mth order and n-mth order.
D/A Output (optional)
Output voltage: ±5 V FS (maximum approximately ±7.5 V) for each rated value Number of outputs: 12 parameters with /DA12 option; 4 parameters with /D A4 option Output data selection Accuracy: ±(equipment accuracy + 0.2% of FS) D/A converter: 12-bit resolution Response time: Maximum 2 times the display updating interval Updating interval: Same as the equipment's display updating interval Temperature coefficient: ±0.05%˚C of FS Output type
Frequency
Integration
Other parameters
: Can be set separately for each channel.
D/A output
7.5V
5.0V
2.5V
0.5V
0
0.5Hz 100Hz10Hz1Hz 1kHz 10kHz
D/A output
7.0V
5.0V
Display
140%
100%
-100%
-140%
Note: For PF and deg, there is no output between ±5 and ±7 V. When an error occurs, approximately 7.5 V are output.
0
t0:
0%
Rated setting time
Output
7.0V
-5.0V
-7.0V
For input equal to 140% of rating
5.0V 0V
-100%-140%
D/A output
7.5V
7.0V
5.0V
For rated input
0V
-5.0V
-7.0V
-7.5V
100kHz
Display value
t0
Integration time
Display value
140%100%
6
External Input (Optional)
Model or part number
751533-E2 751533-J2 751534-E2 751534-J2 751533-E3 751533-J3 751534-E3 751534-J3
Specification For WT210 EIA standalone installation For WT210 JIS standalone installation For WT210 EIA connected installation For WT210 JIS connected installation For WT230 EIA standalone installation For WT230 JIS standalone installation For WT230 EIA connected installation For WT230 JIS connected installation
Order quantity
1 1 1 1 1 1 1 1
Product Rack mounting kit Rack mounting kit Rack mounting kit Rack mounting kit Rack mounting kit Rack mounting kit Rack mounting kit Rack mounting kit
Ask Yokogawa for information on rack mounts in which WT210 and WT230 are combined.
B9317WD
B9284LK
1.5 mm hex wrench External sensor cable
For fastening cable on 758931 For external input; 50 cm
Model number
Description
Single-phase 2-wire Single-phase 3-wire Three-phase 3-wire (2 voltages, 2 currents) Three-phase 3-wire (3 voltages, 3 currents) Three-phase 4-wire
760401
– – – –
760502
✓ ✓ ✓
– –
760503
✓ ✓ ✓ ✓ ✓
Wiring
Model
Select either /EX1 or /EX2 for the voltage output-type current sensor. /EX1: 2.5/5/10 V /EX2: 50/100/200 mV Specifications: See the section on input specifications.
Comparator Output (Optional)
Output method: Normal-open and normal-close relay contact output (pair) Number of output parameters and settings:
Contact capacitance:24 V/0.5 A D/A output (4-channel)
Four parameters; can be set separately on each output channel.
:See section on D/A output (optional)
External Control Signal (with D/A or /CMP Option Only)
External control signals Input: TTL level negative pulse
:EXT-HOLD, EXT-TRIG, EXT-START, EXT-STOP, EXT-RESET,
INTEG-BUSY
General Specifications
Warmup time: Approximately 30 minutes Operating temperature and humidity ranges: 5-40˚C, 20-80% RH (no condensation) Storage temperature Maximum operating elevation: 2000 meters Insulating resistance
Insulating withstand voltage:
Power supply: Free power supply (100-240 V), 50/60 Hz frequency Consumed power: Max 35 VA for WT210, max 55 VA for WT230 External dimensions for WT210:
External dimensions for WT230:
Weight: Approximately 3 kg for WT210, approximately 5 kg for WT230 Safety standard Complying standard EN61010-1
Emission Complying standard EN61326 Class A
Immunity Complying standard EN61326 Annex A
: -25-60˚C (no condensation)
: 50 M or higher at 500 V DC across all of the following areas:
Voltage input terminals (ganged) and case Current input terminals (ganged) and case Voltage input terminals (ganged) and current input terminals (ganged) Voltage input terminals (ganged) of each element Current input terminals (ganged) of each element Voltage input terminals (ganged) and power plug Current input terminals (ganged) and power plug Case and power plug
3700 V for one minute at 50/60 Hz across all of the following areas: Voltage input terminals (ganged) and case Current input terminals (ganged) and case Voltage input terminals (ganged) and current input terminals (ganged) Voltage input terminals (ganged) of each element Current input terminals (ganged) of each element Voltage input terminals (ganged) and power plug Current input terminals (ganged) and power plug 1500 V for one minute at 50/60 Hz across case and po wer plug
Approximately 213 × 88 × 379 mm (WHD) (excluding projections) Approximately 213 × 132 × 379 mm (WHD) (excluding
projections)
Overvoltage category (Installation category) II Pollution degree 2
EN61000-3-2 EN61000-3-3 AS/NZS 2064 Class A
Exterior View
Unit : mm
250 73 23 213
356
359
Model Numbers and Suffix Codes
-D
-F
-R
-Q
-C1
-C2
Suffix code
/C1 /C2
/EX1 /EX2
/HRM
Suffix code
-D
-F
-R
-Q /EX1 /EX2
/HRM
WT210 single-input element model UL/CSA standard VDE
standard
AS
standard
BS
standard GP-IB communication interface Serial (RS-232-C) communication interface External input 2.5/5/10 V External input 50/100/200 mV Harmonic measurement function
/DA4
4-channel DA output
/CMP
Comparator and D/A, 4 channels each
WT230 2-input element model WT230 3-input element model GP-IB communication interface Serial (RS-232-C) communication interface UL/CSA standard VDE
standard
AS
standard
BS
standard External input 2.5/5/10 V External input 50/100/200 mV Harmonic measurement function
/DA12
12-channel DA output
/CMP
Comparator and D/A, 4 channels each
Description
Description
Select one
Select one
Select one
Select one
Select one
Select one
Model number
760401 Power cord
Options
Note: The WT210 communication interface cannot be changed or modified after delivery.
Model number
760502 760503 Interface
Power cord
Options
Standard Accessories
Power cord, Power fuse, Current input protective cover, Rubber feet for the hind feet, 24-pin connector (provided only on options/DA4, /DA12, and /CMP), User’s manual
Wiring Types and Model Numbers
Rack mounts
8819
179
13
213 327
13221
23
28.5
WT210
34
WT230
Accessories (sold separately)
7
Related Products
758917
Measurement leads
Two leads in a set. Use 758917 in combination with 758922 or 758929. Total length: 75 cm Rating: 1000 V, 32 A
758922
Small alligator adapters
For connection to measurement leads (758917). Two in a set. Rating: 300 V
For current measurements with wires connected
960 01 Clamp on Probe
366921
Conversion adapter
758924
Conversion adapter
Measurement frequency range: 20 Hz to 20 kHz
Basic accuracy: 1.0% of reading + 0.2 mA (40 Hz to 1 kHz)
Maximum allowed input: AC 400 Arms
Output: 10 mV/A
A separately sold adapter (366921 or 758924) is required for connection to WT210/WT230. This is a Yokogawa M&C Product. For detailed information, see http://www.yokogawa.com/MCC/clamp.htm#96001 1 Use with low-voltage circuits (42 V or less).
Free Application Software
WTViewer
for the WT210 and WT230
758929
Large alligator adapters
For connection to measurement leads (758917). Two in a set. Rating: 1000 V
758923
Safety terminal adapter set
(spring-hold type) Two adapters in a set.
For high-current measurements up to 1000 Arms
751552 Clamp on Probe
1
Measurement frequency range: 30 Hz to 5 kHz
Basic accuracy: 0.3% of reading
Maximum allowed input: AC 1000 Arms, max 1400 Apk (AC)
Current output type: 1 mA/A
A separately sold fork terminal adapter set (758921), measurement leads (758917), etc. are required for connection to WT210/WT230. For detailed information, see Power Meter Accessory Catalog Bulletin 7515­52E.
Due to the nature of this product, it is possible to touch its metal parts. Therefore, there is a risk of electric shock, so the product must be used with caution.
758917
Measurement leads
758921
Fork Terminal Adapter
DAQLOGGER & GateWT
GateWT is a software package that can collect data measured by digital power meter WT series including WT210 and WT230 through a GP-IB or serial (RS-232) Communication interface. See Bulletin 04L00L00-00E for details.
758931
Safety terminal adapter set
Screw-fastened adapters. Two adapters in a set. 1.5 mm Allen wrench included for tightening.
B9284LK
External sensor cable
For the external input of the WT210 and WT 230. Length: 50 cm
For high precision (0.05% + 40 µA)
751574 Current Transducer
Wide dynamic range: 0-600 A (DC)/600 A peak (AC)
Wide measurement frequency range: DC and up to 100 kHz (-3 dB)
High-precision fundamental accuracy: ±(0.05% of reading + 40 µA)
±15 V DC power supply, connector, and load resistor required.
For detailed information, see Power Meter Accessory Catalog Bulletin 7515-52E.
LabVIEW* Driver Software (Free)
Download this software program from our web site.
Easily Acquire and Manage Power Measurement Data form Your PC
WTViewer for the WT210/WT230 is a software application that allows you to load numeric and waveform data measured with the WT210 or WT 230 Digital Power Meter to a PC via GP-IB or serial (RS-232-C) communications.
Visit our web site to register your product and download this software program.
http://www.yokogawa.com/tm/WT210/
See our web site or the software catalog (Bulletin 7604-32E) for detailed specifications.
Information on the features and functions of Yokogawa’s WT series & PZ, accessories, and related products is also available at our homepage. http://www.yokogawa.com/tm/
Protecting the global environment
Yokogawa's products are developed and produced in facilities that have received ISO14001 approval.
YOKOGAWA ELECTRIC CORPORATION
Test and Measurement Business Div.
/Phone: (81)-55-243-0313, Fax: (81)-55-243-0396
CAUTION
Read the user's manual carefully for correct and safe use of the instrument.
E-mail: tm@csv.yokogawa.co.jp
YOKOGAWA CORPORATION OF AMERICA Phone: (1)-770-253-7000, Fax: (1)-770-251-2088 YOKOGAWA EUROPE B.V. Phone: (31)-33-4641806, Fax: (31)-33-4641807 YOKOGAWA ENGINEERING ASIA PTE. LTD Phone: (65)-62419933, Fax: (65)-62412606
* LabVIEW is a registered trademark of National
Instruments Corporation.
Subject to change without notice.
[Ed : 02/b] Copyright ©2002
Printed in Japan, 307(YG)
MS-12E