Digital Thermometer [ Operation Manual ]
This manual describes the specications and handling precautions of
the digital thermometer. Before using this product, thoroughly read this manual to understand how to use it properly.
IM TX10-02E Safety Precautions, Specications etc. IM TX10-02-02EN: Operation Manual (This manual)
Store this manual in an easily accessible place for quick reference.
Printed in Japan
All Rights Reserved. Copyright © 2015 Yokogawa Test & Measurement Corporation
IM TX10-02-02EN
This user's manual describes the TX1003, which has more functions than the TX1002.
However, with the exception of the
key, operation of the TX1002 is performed in
the same way.
7. Components and Functional Description
7.1 Components
(4) (5) (6) (7) (8)
(1) (2)
(3)
value is shown.
is shown.
correction mode.
(11)(10)
Channel A
(chA)
Channel B
(chB)
Resolution select key
Maximum and minimum value record key Data record key
auto power off release
YOKOGAWA
(For connecting the probe)
Display
Data hold key
Power ON/OFF key Thermocouple type
select key (Type K, J, E, T)
Input channel select key
Data call-up key
[ Rear side ] Battery box
(2) Displayed when stored data
(3) Displayed in simplified
(4) Data hold status (5) Maximum-minimum record mode (6) Relative display mode (7) Battery alarm (8) Auto power-off status (9) Unit (10) Selected channel
7.2 Functional Description
To turn the power on and off
Key Operation and description Display
Power on
Power off
■ Press the POWER key to turn on the power.
• The auto power-off function turns off the power automatically about 10 minutes after the last key operation of any key other than the key.
• To disable the auto power-off function, turn off the power first
if the power is already turned on.
Then simultaneously press the and key. This will keep the instrument turned on continuously. (If the instrument is used with the auto power-off function disabled, pay attention to the battery consumption.)
■ To turn off the power, press this key for 1 second or longer.
After the power has been turned on, all the elements of the display light up.
AUTO OFF
To select the thermocouple type
To select a thermocouple type from K, J, E, and T:
+
If the power is already turned on, turn it off.
■ Press and hold down the key while pressing the key to turn on the power.
The thermocouple type and unit on the display flash, indicating that the instrument is ready for changing thermocouple type.
■ Every time the TC TYPE key is pressed the thermocouple type.
K, J, E, T
■ Press the POWER/SET key to accept the setting.
TIP
The factory-set default is type K.
To select the input channel
3rd Edition:
Contact information of Yokogawa ofces worldwide is provided on the following sheet.
PIM 113-01Z2 Inquiries List of worldwide contacts
To select the resolution
To select a display resolution:
■ Press the RANGE key. Every time this key is pressed, the display resolution alternates between 0.1°C and 1°C (within the range of -200.0°C to 199.9°C). If 0.1°C is selected, a resolution of 0.1°C is achieved when the display value is 199.9°C or below; when it goes over 200°C, the resolution changes to 1°C, and the number of display digits is switched automatically. When “chA-B” is selected, the display is switched to the input value of channel A or B, whichever is greater.
TIP
• The resolution setting is maintained even after the power is turned off. The next time the power is turned on, the instrument starts up with the setting that was set before the power was turned off.
• Factory-set default is 0.1°C.
To turn relative display on and off
Relative display
Cancel relative display
To select relative display:
■ Press the REL key to select a relative display. The display shows the difference (relative value) of a subsequent measured value (Dx) from the reading given as a reference value (D1) when this key was pressed.
Relative display value = Dx–D1
The reference value affects measurements on all channels (channel A, channel B and channel A-B). Any value among MAX, MIN and MEM can also be used as the reference value.
■ To cancel relative display, press the REL key again.
To turn simplied correction on and off
To set the correction value for simplified correction:
The current temperature reading can be corrected to a value based on the reference thermometer. By thus eliminating probe-specific errors, you can take precision measurements over a frequently used temperature range.
When you change the measurement probe, the correction value must be set again regardless of the thermocouple type.
■ Press and hold the REL key about 3 seconds.
The ADJ on the display flashes, indicating that the instrument is ready for setting.
■ Carry out the correction.
While measuring the reference temperature, correct the display value to an optional value using the [ ] and [ ] keys.
(The correction range is ±20°C of the measured value. Resolution of the correction value is determined according to
the resolution setting.)
TIP
Holding down the [ ] or [ ] key increases the speed with which the display value changes. (a value continues to change)
■ Press the POWER/SET key to accept the setting.
Simplified correction mode
Cancel simplified correction mode
Use this key to enter and exit the simplified correction mode:
■ To enter, press the REL key for 1 to 2 seconds.
■ To exit, press the REL key for 1 to 2 seconds again.
ADJ
ADJ flashes.
(TX1003 only)
To hold data
Hold
Release hold
To select the input channel from channel A, B, or A-B:
■ Press the CH key. With each press, the display changes between A, B, and A-B, allowing you to select the input channel for measurement.
• Channel switching is disabled during the RCD mode.
• By switching the channels when data is being read, you can display the data of each channel: maximum and minimum values and measured data saved using the Memory-in function.
Use this key to hold the measured value on LCD.
■ To hold, press the DATA HOLD key.
■ To cancel holding, press the DATA HOLD key again.
To record maximum and minimum values
NOTE
CAUTION
Attention
Burnout indication
Recording of the maximum and minimum values:
Detection and recording of the max./min. values starts
Recording
Cancel recording
when the RECORD key is pressed and continues until this key is pressed again or the power is turned off.
■ To start recording, press the RECORD key.
Enable the RCD mode after selecting the channel.
■ To stop the recording, press the RECORD key again.
Measurement on channel A-B is done as described below.
Press the key to read the minimum and maximum values.
With the key, select channel A or B. The display shows the MIN and MAX values for each channel.
Temperature
Channel-A MIN value
Channel A-B MAX value
(°C)
Channel-B MIN value
(Start of MIN/MAX recording) (End of MIN/MAX recording)
• The maximum and minimum values are stored in internal memory even after recording is cancelled. The data will be updated when the RCD mode is enabled again.
• If you enable the RCD mode when the REL mode is active, the reading changes to a relative value. Nevertheless, the minimum and maximum of measured values (Dx) are stored in internal memory.
• Recorded maximum and minimum values can be referenced even after the power is turned off and on again because they are stored in memory until the RCD mode is enabled next time.
• Auto power-off is cancelled during the RCD mode.
Channel-A MAX value
chA
Channel A-B MIN value
chB
Channel-B MAX value
Time (t)
To store data in memory
To store data in internal memory (maximum of 10 data items):
■ Press the DATA HOLD key to hold the measured value.
■ Press the RECORD key.
The display automatically shows the smallest memory number under which no data has been recorded. At this point, the MEM symbol turns on to indicate that the thermometer is in the “Memory-in” mode.
■ Select a memory number. Use the [ ▲ ] and [ ▼ ] keys to select a desired memory number under which you want to record data. In this step, you can also check the existing data value.
TIP
Pressing the key while a memory number is being displayed cancels the memory-in function (returns to measurement mode).
■ Press the RECORD key to store the data.
TIP
The “HOLD” and “MEM” symbols indicate that the value recorded with the memory-in function is being held and shown.
Press the key to return to normal measurement mode.
To read maximum and minimum values and stored data values
To read maximum and minimum values and stored data values from memory:
■ When the READ key is pressed,
the display changes in the sequence shown in the figure on the right.
■ Select the memory number. For stored data, select the memory number using the [ ▲ ] or [ ▼ ] key. After the memory number has been displayed for a few seconds, the value stored in that memory is displayed.
Measured
value
Maximum
value
Minimum
value
Stored
value
Memory number selection
(or chB or chA-chB)
Source input channel of the recorded data
MAX
chA
MIN
MEM
• When the probe is not connected to the input connector (or the sensor opens), a burnout mark is displayed.
• For the 2-channel mode, the burnout mark is displayed when the probe is not connected to the selected channel (channel A and B, or channel A-B)
To clear the memory
To clear the recording memory:
■ Press the READ key to display stored data.
■ Select the memory number you wish to clear.
Use the [ ▲ ] and [ ▼ ] keys to select a desired memory number.
■ Press the RECORD key to clear the data. To clear all the stored data in the memory at one time, press and hold the RECORD key for 1 to 2 seconds.
TIP
This operation does not affect the maximum and minimum values.
8. Measurement
This instrument can connect any of the 4 types of thermocouples:
Type K, J, E, or T (it is set to Type K at shipment from the factory). Use an omega-type connector corresponding to the TC type for the probe.
[ Procedure ]
(1) Load the batteries (see section "9. Battery Replacement"). (2) Press the
key to turn on the power.
When the power has been turned on, all indications on the display light up
for about 2 seconds, then the measured value appears.
When the probe is not connected to the input connector, a burnout mark is displayed.
Burnout indication:   - - - -
(3) Check the thermocouple type.
If you wish to change the current setting, see "To select a thermocouple type" of section "6.2 Functional Description."
(4) Select the input channel. (TX1003 only)
Press the
key and select one channel from channel A, B, and A-B.
(see "To select the input channel.")
(5) Connect the probe to the input connector.
If you are using the 2-channel model (TX1003), connect the probe(s) to the channel(s) you are using. (The measured value appears.)
(6) When you have nished measurement, turn off the power by
pressing the
key.
• The power may be turned off by the auto power-off function during measurement. In this case, press the
key to restart the instrument and
continue measurement.
• If the power is turned off and then on again, the instrument restarts with the same settings of the resolution (0.1°C or 1°C), input channel (channel A, B or A-B) and thermocouple type (K, J, E or T) as before
it was turned off.
For other operations such as display hold, relative display, and maximum and minimum record/readout, see section "7.2 Functional Description."
Key operation of this instrument is comparatively easy.
However, it is necessary to make sure that the instrument is not mistakenly set to a state other than that intended. Note that, under normal measurement,
“HOLD,” “RCD,” “REL,” “MAX,” “MIN,” and “MEM” are not displayed.
9. Battery Replacement
• When inserting batteries, be careful not to mistake the polarity as this can damage the instrument.
• When the instrument is to be stored for a long period of time, remove the batteries.
• Do not leave dead batteries in the instrument as doing so may result in a failure or
malfunction of the instrument due to a leakage in battery liquid.
• Both batteries should be replaced with new ones at the same time as replacing
only one of the batteries may result in the charge leaking from the new battery to the old.
• This data reading is disabled during thermocouple type selecting or correction value setting.
• If you press the READ key when the REL mode is active, each of the MIN, MAX and MEM values is shown as a relative value based on the reference value (D1).
• The number of display digits of maximum and minimum/stored data depends on the current resolution setting, not the number of display digits at the time when the data was measured.
Channel switching and display when data is being read (TX1003 only):
■ Single channel (either channel A or B)
Using the key, display the channel used when the data was measured. All of the maximum, minimum, and stored data values can be displayed.
■ Two channels (channel A and B measurement)
• Pressing the key to select channel A or B shows the minimum and maximum values, as well as the memory value, measured on that channel.
• Pressing the key to select ‘chA-chB’ displays the stored values of the difference between channels A and B as well as the maximum or minimum difference values.
• Lors de l'insertion des piles, faire en sorte de ne pas confondre de polarité, car cela risquerait d'endommager l'instrument.
• Si l'instrument doit être stocké pendant une longue période, retirer les piles.
• Ne pas laisser de vieilles piles usagées dans l'instrument, car cela risquerait sinon d'entraîner une défaillance ou un dysfonctionnement de l'instrument dû(e) à une fuite du
liquide des piles.
• Les deux piles doivent être remplacées par des nouvelles en même moment, car le fait de n'en remplacer qu'une seule risque d'entraîner un transfert de charge de la nouvelle pile vers l'ancienne.
When the batteries approach the end of their lives, the
mark appears
on the display. If this mark is displayed, replace the batteries.
To replace the batteries:
(1) Remove the battery box cover from the back (rear) of the instrument.
Then replace the two AA alkaline dry batteries (LR6).
(2) Ret the cover to the battery box.