Yokogawa Corporation of America
2 Dart Road, Newnan, Georgia U.S.A. 30265
Tel: 1-800-258-2552 Fax: 1-770-254-0928
IM 11Y01B01-01E-A
6th Edition
Introduction
Thank you for purchase the TDLS200 Tunable Diode Laser Analyzer. Please read the following respective
documents before installing and using the TDLS200.
Notes on Handling User’s Manuals
• This manual should be passed on to the end user.
• The contents of this manual are subject to change without prior notice.
• The contents of this manual shall not be reproduced or copied, in part or in whole, without permission.
• This manual explains the functions contained in this product, but does not warrant that they are suitable
for the particular purpose of the user.
• Every effort has been made to ensure accuracy in the preparation of this manual. However, when you
realize mistaken expressions or omissions, please contact the nearest Yokogawa Electric representative
or sales ofce.
• This manual does not cover the special specications. This manual may be left unchanged on any
change of specication, construction or parts when the change does not affect the functions or
performance of the product.
• If the product is not used in a manner specied in this manual, the safety of this product may be
impaired.
i
Yokogawa is not responsible for damage to the instrument, poor performance of the instrument or losses
|resulting from such, if the problems are caused by:
• Improper operation by the user.
• Use of the instrument in improper applications
• Use of the instrument in an improper environment or improper utility program
• Repair or modication of the related instrument by an engineer not authorized by Yokogawa.
Drawing Conventions
Some drawings may be partially emphasized, simplied, or omitted, for the convenience of description.
Some screen images depicted in the user’s manual may have different display positions or character types
(e.g., the upper / lower case). Also note that some of the images contained in this user’s manual are display
examples.
Safety, Protection, and Modification of the Product
•In order to protect the system controlled by the product and the product itself and ensure safe operation, observe the safety precautions described in this user’s manual. We assume no liability for safety if users fail to observe these instructions when operating the product.
•If this instrument is used in a manner not specified in this user’s manual, the protection provided by this instrument may be impaired.
•If any protection or safety circuit is required for the system controlled by the product or for the product itselfprepare it separately.
•Be sure to use the spare parts approved by Yokogawa Electric Corporation (hereafter simply referred to as YOKOGAWA) when replacing parts or consumables.
•Modification of the product is strictly prohibited.
•The following safety symbols are used on the product as well as in this manual.
Safety, Protection, and Modication of the Product
• In order to protect the system controlled by the product and the product itself and ensure safe operation,
observe the safety precautions described in this user’s manual. We assume no liability for safety if users
fail to observe these instructions when operating the product.
• If this instrument is used in a manner not specied in this user’s manual, the protection provided by this
instrument may be impaired.
• If any protection or safety circuit is required for the system controlled by the product or for the product
itself, prepare it separately.
• Be sure to use the spare parts approved by Yokogawa Electric Corporation (hereafter simply referred to
as YOKOGAWA) when replacing parts or consumables.
• Modication of the product is strictly prohibited.
• The following safety symbols are used on the product as well as in this manual.
This symbol indicates that an operator must follow the instructions laid out in this manual in
order to avoid the risks, for the human body, of injury, electric shock, or fatalities. The
manual describes what special care the operator must take to avoid such risks.
DANGER
WARNING
This symbol indicates that the operator must refer to the instructions in this manual in order
to prevent the instrument (hardware) or software from being damaged, or a system failure
from occurring.
ii
CAUTION
This symbol gives information essential for understanding the operations and functions.
Note!
This symbol indicates information that complements the present topic.
The product is provided on an “as is” basis. YOKOGAWA shall have neither liability nor responsibility to any
person or entity with respect to any direct or indirect loss or damage arising from using the product or any
defect of the product that YOKOGAWA cannot predict in advance.
IM 11Y01B01-01E-A 6th Edition :Feb 13, 2013-00
TDLS200
CAUTION
SAFETY should be considered rst and foremost importance when working on the equipment described in this
manual. All persons using this manual in conjunction with the equipment must evaluate all aspects of the task
for potential risks, hazards and dangerous situations that may exist or potentially exist. Please take appropriate
action to prevent ALL POTENTIAL ACCIDENTS.
AVOID SHOCK AND IMPACT TO THE ANALYZER THE LASERS CAN BE PERMANENTLY DAMAGED
Laser Safety & Classication according to FDA Regulations. The TDLS200 is Registered with the United States
FDA as a Laser Product.
WARNING
THIS ANALYZER CONTAINS A LASER PRODUCT THAT IS GENERALLY IN ACCORDANCE WITH THE REGULATIONS FOR THE ADMINISTRATION AND ENFORCEMENT OF THE RADIATION CONTROL FOR HEALTH AND
SAFETY ACT OF 1968 (TITLE 21, CODE OF FEDERAL REGULATIONS, SUBCHAPTER J). REFER SECTION
1002.10 OF THE REGULATIONS REFERENCED ABOVE.
CAUTION INVISIBLE LASER RADIATION AVOID DIRECT EXPOSURE
MAXIMUM OUTPUT POWER < 1 MW (Oxygen)
MAXIMUM OUTPUT POWER < 20 mW (other Gases) DURING NORMAL OPERATION THIS ANALYZER IS:
CLASS I LASER PRODUCT (according to IEC 60825-1)
CAUTION
iii
The Instrument is packed carefully with shock absorbing materials, nevertheless, the instrument may be
damaged or broken if subjected to strong shock, such as if the instrument is dropped. Handle with care.
Warranty and service
Yokogawa products and parts are guaranteed free from defects in workmanship and material under normal use
and service for a period of (typically) 12 months from the date of shipment from the manufacturer. Individual
sales organizations can deviate from the typical warranty period, and the conditions of sale relating to the original purchase order should be consulted. Damage caused by wear and tear, inadequate maintenance, corrosion,
or by the effects of chemical processes are excluded from this warranty coverage.
In the event of warranty claim, the defective goods should be sent (freight paid) to the service department of the
relevant sales organization for repair or replacement (at Yokogawa discretion). The following information must be
included in the letter accompanying the returned goods:
• Part number, model code and serial
• Number
• Original purchase order and date
• Length of time in service and a description of the process
• Description of the fault, and the circumstances of failure
• Process/environmental conditions that may be related to the failure of the device.
• A statement whether warranty or nonwarranty service is requested
• Complete shipping and billing instructions for return of material, plus the name and phone number of a
contact person who can be reached for further information.
Returned goods that have been in contact with process uids must be decontaminated/ disinfected before
shipment. Goods should carry a certicate to this effect, for the health and safety of our employees. Material
safety data sheets should also be included for all components of the processes to which the equipment has
been exposed.
IM 11Y01B01-01E-A 6th Edition :Feb 13, 2013-00
iv
DANGER
WARNING
DANGER
Dont install “general purpose type” instruments in the hazardous area.
CAUTION
The intrument is packed carefully with shock absorbing materials, nevertheless, the instrument may be damaged
or broken if subjected to strong shock, such as if the instrument is dropped. Handle with care.
IM 11Y01B01-01E-A 6th Edition :Feb 13, 2013-00
TOC-1
TABLE OF CONTENTS
Introduction .......................................................................................................................................................... i
Safety Precutions ............................................................................................................................................... ii
3 General Specications .............................................................................................................................3-1
3.1 Model & Sufx Code ............................................................................................................................3-4
4.1 Launch Unit .........................................................................................................................................4-2
4.2 Main Electronics Housing ....................................................................................................................4-3
4.4 Check Gas Flow Cell (for On-Line) ......................................................................................................4-7
4.5 Detect Unit ..........................................................................................................................................4-8
4.6 Process Interface .................................................................................................................................4-9
5 Installation and Wiring .............................................................................................................................5-1
5.1 Process Measurement Point Considerations .....................................................................................5-1
5.2 Position of Process Flanges for Launch and Detect Units .................................................................5-2
5.3 Process Flange Welding Alignment and Line-Up ............................................................................... 5-4
5.4 Process Flange Clear Aperture ...........................................................................................................5-5
5.5 Mounting the Launch and Detect Units to the Process Flange .........................................................5-5
5.5.1 Process Window Purge Gas Connection .....................................................................................5-6
5.6 Mounting the Process Interface .........................................................................................................5-6
5.7 Typical Purge Gas Conguration, In-Situ ...........................................................................................5-7
5.8 Typical Purge Gas Conguration, Extractive trace ppm H2O system.................................................5-7
6.4 Reference Peak Lock with 2nd Absorption gas ...............................................................................6-22
6.5 Large Aperture Optics ......................................................................................................................6-26
6.5.1 LAO Installation, Alignment & Dector Gain .................................................................................6-27
6.5.2 Adjustment of Dector Gain for LAO ...........................................................................................6-28
6.5.3 Dector Gain Adjustment Service Tips ........................................................................................6-30
IM 11Y01B01-01E-A 6th Edition :Feb 13, 2013-00
6.6 Valve Control Logic ...........................................................................................................................6-30
6.7 Introduction for H2Oppm measurements in Methane Gas ...............................................................6-32
6.8 Introduction to Gas Temperature Predictions with High Temperature Oxygen Measurements .......6-38
6.9 Controlling the Analyzer Remotely or Locally via external PC/Laptop2 ...........................................6-34
6.9.1 Instructions for Connecting an External Computer to the Analyzer ...........................................6-35
6.9.2 Using Ultra-VNC Software ..........................................................................................................6-36
6.9.3 Remote Interface Unit (RIU) ........................................................................................................6-37
10. Data Files And Format ............................................................................................................................10-1
10.1 Conguring Data Capture .................................................................................................................. 10-5
10.2 Downloading (Transfering/Exporting) the Data ..................................................................................10-8
IM 11Y01B01-01E-A 6th Edition :Feb 13, 2013-00
<1 QUICK START>
1-1
1 QUICK START
StepTitleDescription
1.0
1.1Ensure the process connections match the supplied process interface.
1.2Ensure the appropriate utilities are available and ready for connection. These may include electrical
1.3Ensure you comply with any local and/or site specic safety requirements.
1.4Read the appropriate sections of the Instruction Manual BEFORE starting any installation work –
2.0
2.1Attach the process interface (alignment anges) to the site installed anges (or isolation valves as
2.2Carefully mount the Launch and Detect Units to their alignment anges using the quick connect
2.3Mount optional equipment such as Universal Power Supply (UPS), Universal Remote Display (URD),
2.4
Preparation
Installation
Ambient
Temperature
3.0
3.1Connect the appropriate electrical power supply.
3.2Connect the Launch to Detect interconnect cable (supplied with analyzer) according to the supplied
3.3Connect any analog I/O signals to the analog I/O Board. Outputs land on TB8 and Inputs land on TB9.
3.4Connect any other equipment such as URD, Ethernet, solenoid valves, digital I/O, etc.
3.5
Wiring
Carefully un-pack and check equipment for any obvious damage. This includes anges, Cables, Power
Supplies, manuals and any other supplied options.
NOTES: There are 14 ferrules in the accessory bag for tubing-piping. The number of ferrule that are
required for actual tubing-piping are different by application. Please see tubing-piping gure
specic to project for exact detail.
power, nitrogen purge gas, instrument air, validation gas, etc.
Contact Yokogawa Laser Analysis Division or Local Agent if any doubts!
If separate process isolation anges have been provided for corrosive service, then install to the
process/stack ange/isolation valves.
appropriate).
If installing Large Aperture Optics, ensure the detect system is correctly mounted and purged to
prevent damage to the large optical element.
coupling.
Remote Interface Unit (RIU), etc.
The analyzer and some accessories (such as LAO, RIU, UPS, URD, alignment anges, etc.) are
suitable for -20 to +50oC ambient operating temperature. Accessories and Options are available to
increase these the operating conditions – please consult Yokogawa for further details.
Ensure that all wiring will enable the analyzer launch and detect units to be freely moved from their
process location to an adjacent off-line calibration cell. This will entail the use of tray rated cables and/
or exible conduit and/or other suitable armored cable. Rigid conduit systems are not recommended.
• 24 VDC to TB1 on the analyzer (launch Unit) backplane. Check that the actual voltage
is >23.5VDC otherwise the SBC and other devices will not function!
• 110/240 50/60 Hz to UPS or URD, then take 24 VDC to analyzer
wiring detail (TB7 on the Launch and TB 13 on the Detect Unit).
Check terminations and ensure all cable shields are landed per supplied wiring details.
4.0
Utilities
4.1Connect the appropriate process window purge gas (nitrogen for oxygen analyzers) and make site
4.2Connect the appropriate analyzer on-line check gas ow cell gas (nitrogen for oxygen analyzers) and
4.3Connect and check any other required utility connections (such as steam trace for heated isolation
NOTE! – All purge, Validation Gas and other gas utility lines should be thoroughly cleaned, dried
and purged prior to connecting to the analyzer
– Failure to do so can result in serious damage to the TDLS200 or contamination to the
internal optical elements.
Connect the appropriate analyzer purge gas (nitrogen for oxygen analyzers) and make site connections
per the supplied purge gas sequence details (including any Hazardous area purge system). Start the
purge gas ow accordingly.
ATEX purge requires dual regulators at the inlet purge gas supply to prevent overpressure damage in
the event of a single regulator failure!
connections per the supplied purge gas sequence details. Start the window purge gas ow
accordingly – ensuring that any isolation valves are open.
make site connections per the supplied purge gas sequence details. Start the purge gas ow
accordingly.
anges or ow cells) or secondary window purges for lethal service gases. Start other utilities
accordingly.
IM 11Y01B01-01E-A 6th Edition :Feb 13, 2013-00
<1 QUICK START>
4.4
5.0
5.1Use the internal On-Off switch to power-up the analyzer.
5.2Observe the various LED clusters on the backplane and FPGA boards. All blue LEDs located on the
5.3Observe the Green power indicator on the SBC.
5.4
6.0
6.1If there is an installed optional Mini Display (4x20 VFD) – Observe the status line message.
6.2If there is no installed User Interface, then connect a laptop PC via Ethernet to the SBC mounted
6.3
Power-Up
Checking
Alignment
Leak-check all connections and ensure pressure ratings are not exceeded!
Apply power to the analyzer and using a multi-meter, check for 24VDC power at TB1 on the launch
unit back plane.
lower right side of the back-plane should be on.
Observe the LEDs on the analog I/O board.
If there is an installed optional 6.5” Display and Keypad – Observe the Main Menu messages and
status information.
on the backplane. Initiate the supplied VNC software from the laptop to initiate a VNC session with
the ‘blind’ analyzer and observe the analyzer Main Menu via the laptop.
At this time there may be one or more alarm message due to low transmission, out of range
parameters or other – nal system conguration is still required!
Please also note that the analyzer laser temperature control is disabled for the initializing
period (5 minutes) – this means that even manual control of the laser temperature is disabled
during this period.
Initially, observe the Transmission value through the appropriate user interface. The objective is to
adjust alignment until the maximum transmission value is obtained. Perfect alignment in a clear
process gas will yield close to 100% transmission.
If the analyzer displays a Warning “Validation Required”, this indicates there is no target gas absorption peak found at start-up.
7.0
7.1Start by adjusting the Launch unit alignment ange nuts up-down and left right. Look for increases
7.2When it has been maximized at the launch side, adjust the detect unit accordingly.
7.3Further adjustment can be made by maximizing the raw detector voltage signal (available at test
7.4
8.0
Alignment – check
Detector Gain
Congure
BASIC
8.1Enter the Basic Menu and go to Congure.
8.2
8.3
8.4
8.5If any other parameters are required to be set (such as analog I/O ranges, alarms levels, Auto
Optical Path
Gas Pressure
Gas Temperature
Introduce some measured gas into the optical path and re-start or perform a validation with target
gas. This will ensure that the analyzer is correctly tuned to the measurement gas absorption peak.
If this Warning cannot be cleared by either method, please contact Yokogawa Laser Analysis
Division or your local agent for further assistance.
If you have 100% certainty that the analyzer is already measuring the process gas and validation is
not currently possible then, this alarm can be cleared via the Advanced Calibrate & Validate menu.
Initially, observe the Transmission value through the appropriate user interface. The objective is
to adjust alignment until the maximum transmission value is obtained. Perfect alignment a clear
process gas will yield close to 100% transmission.
and decreases in transmission strength to aid in the alignment.
points on both launch and detect). The signal should be maximized and will not exceed 5.3V DC for
low temperature (<600C process) or 9.9V DC for high temperature (>600C process).
For Large Aperture Optics (LAO) systems, please refer to the Detector Gain Adjustment section of
this User Guide to ensure correct functionality and adjustment.
By way of the appropriate user interface, the correct process parameters and other parameters can
now be entered.
Enter in the correct optical path length.
Enter in the correct process gas pressure (if Active, see Advanced Congure).
Enter in the correct process gas temperature (if Active, see Advanced Congure).
Validation sequences) then the Advanced Menu needs to be accessed.
Advanced Menu access is Password protected and should only be used by skilled and
trained persons - Contact Yokogawa Laser Analysis Division or Local Agent if any doubts!
IM 11Y01B01-01E-A 6th Edition :Feb 13, 2013-00
<1 QUICK START> 1-3
2-3
9.0
Congure
Using the correct password (Default 1234), enter in to the Advanced Menu, then the Congure.
ADVANCED
9.1Select the desired measurement units (English or Metric selected on an individual parameter basis).
9.2
9.3
9.4
9.5Congure the system I/O by entering in to the System I/O sub menu in Congure.
9.6If the Analog I/O board is installed, then select Analog Output and set the appropriate 4mA and
9.7Select what mode (Block, Track or Hold) the 4-20mA outputs are to be when the analyzer is in
9.8Congure Digital outputs – Warnings and Faults. Many of these will be factory preset so if unsure
9.9Go to the Data screen and set the appropriate parameters for and ‘Spectrum Capture’. These will
9.10Go to the Trends screen and review/plot several of the listed parameters to check analyzer
9.11
Optical Path
Gas Pressure
Gas Temperature
Non- Process
Parameters
10.0
10.1To Export Data, simply insert an empty USB memory stick in to a USB port on the launch unit back
10.2Close out the VNC software and disconnect the service PC – if connected.
10.3Ensure the doors/lids are closed and tightly sealed.
10.4The system is now in normal operation mode.
10.5
Normal Operation
Enter in the correct optical path length.
Select Fixed or Active.
If Fixed, enter in the correct process gas pressure.
If Active, enter in the 4-20mA input signal range proportional to the pressure range.
“Control” mode is not applicable to TDLS200
Select Fixed or Active.
If Fixed, enter in the correct process gas temperature.
If Active, enter in the 4-20mA input signal range proportional to the temperature range.
Active ambient and Active Peaks may also be used, refer to project specic and application
specic details.
“Control” mode is not applicable to TDLS200
20mA values for Ch1 Concentration and Ch2 Transmission.
Warning, Fault and Calibration Modes.
about any settings then leave as Factory Default. Select and set level for Alarm Limit to either the
Measured Gas orTransmission.
ensure the analyzer stores important information during operation that may be used to verify
operation/status/diagnostics and other trouble shooting.
performance over a period of time.
If the application use gas containing the target gas (e.g. Oxygen measurement with
Instrument Air Purge) then the Non-Process parameters should be congured as detailed later in
this manual under the Software Section. Non-Process Parameters should also be congured if
using a linelocking gas in the validation cell (e.g. CO for combustion).
When the site/eld conguration is complete and the analyzer has operated for at least two hours
without any functional alarms, then perform an export data routine.
plane. The data transfer may take several minutes.
DO NOT REMOVE THE MEMORY STICK DURING THIS TIME!
We RECOMMEND sending all the Exported Data les to Yokogawa Laser Analysis Division
along with any notes and comments. We will then be able to store these les on a master
record for future reference.
Please carefully read the appropriate Sections of this Instruction Manual. The TDLS200 Tunable Diode
Laser (TDL) Analyzer is a technologically advanced instrument that requires the appropriate care when
handling, installing and operating.
Failure to do so may result in damage and can void any warranties!
If there is any doubt about any aspect of the Instrument or its use, please contact Yokogawa Laser
Analysis Division and/or your authorized Representative/Distributor.
IM 11Y01B01-01E-A 6th Edition :Feb 13, 2013-00
<2. INTRODUCTION AND GENERAL DESCRIPTION>2-32-1
2 INTRODUCTION AND GENERAL DESCRIPTION
The TDLS200 TDLS analyzer is designed to measure
selected target gases in gas phase samples directly
at the process point (across stack, across pipe,
etc.), close coupled/by-pass leg or in full extractive
systems (ow cell).
The analyzer measures free molecules on a path
averaged basis. Unless there is an extractive
sampling system up-stream that removes water (or
other condensables) then the measurements are
considered to be on a ‘Wet Basis’.
Measurements are possible (with correct analyzer
conguration) at the following conditions:
• Gas temperatures up to 1500˚C (2730˚F)
• Gas pressures up to 10 BarG (145 psig)
• High Particulate loading (as a function of mea
surement path length)
Each application may differ in maximum
limitations depending upon the combination of
gas temperature, gas pressure, optical path length
and concentration of the gas being measured. The
standard analyzer is designed for operation in a Safe
Area (General Purpose). The addition of a Purge
System facilitates operation in Hazardous Areas in
accordance with the relevant UL, CSA and ATEX
standards for gaseous releases.
The basic TDLS200 analyzer comprises two units,
the Launch Control Unit and Detect Unit.
Various Process Interface congurations are
available for connecting the analyzer to the
measurement point. Several options may be added
to the standard analyzer such as:
2.1 Functional Description
Tunable Diode Laser Spectroscopy (or TDLS)
measurements are based on absorption
spectroscopy. The TDLS200 Analyzer is a TDLS
system and operates by measuring the amount of
laser light that is absorbed (lost) as it travels through
the gas being measured. In the simplest form a
TDLS analyzer consists of a laser that produces
infrared light, optical lenses to focus the laser light
through the gas to be measured and then on to a
detector, the detector, and electronics that control
the laser and translate the detector signal into a
signal representing the gas concentration. Gas
molecules absorb light at specic colors, called
absorption lines. This absorption follows Beer’s Law.
Using a Tunable Diode Laser as a light source for
spectroscopy has the following benets:
• Sensitivity. As low as 10-6 by volume, lower
with path length enhancement.
• Selectivity. The narrow line width of the laser is
able to resolve single absorption lines. This
provides more choices of a particular peak
to use for measurement, usually allowing one
isolated peak to be used.
• Power. Diode lasers have power ranging from
0.5 mW to 20 mW. Also, being highly coherent
this allows measurement in optically thick
environments (high particulate loading).
• Monochromatic, no dispersive element (lter,
etc.) required. Light source itself is selective.
• Mini Display
• 6.5” screen and keypad
• Display sun shield
• Optional Universal Power Supply (with or
without a Mini Display)
• Remote Interface Unit (not required for normal
operation)
• Hazardous Area purge systems
• Other options may also be added.
• Tunable Wavelength can be swept across the
entire absorption feature, this allows resonant
(peak) and non resonant (baseline)
measurement during every scan. By
measuring the baseline and peak power at the
detector, transmission can uctuate rapidly by
large amounts without affecting the
measurement. This is useful for high particulate
applications.
IM 11Y01B01-01E-A 6th Edition :Feb 13, 2013-00
<2. INTRODUCTION AND GENERAL DESCRIPTION> 2-2
Currentramptolaser
SignalatDetector
Currentramptolaser
SignalatDetector
ProcessedDetectorSignal
2.1.1 Measurement
• During measurement the laser is held at a xed temperature. This is the coarse wavelength adjustment.
• A current ramp is fed to the laser. This is the ne
wavelength adjustment. Figure 1.
• The current is ramped to scan across the wavelength region desired.
Figure 1.
• The collimated light passes through the gas to be
measured. The amount of light absorbed by the peak is
proportional to the analyte concentration.
• The light is then focused on a detector. Figure 2.
• This signal is used to quantify the light absorbed by the analyte.
Figure 3.
Figure 2.
Currentramptolaser
Figure 3.
2.2 Instrument Check
Upon delivery, unpack the instrument carefully and inspect it to ensure that it was not damaged during
shipment. If damage is found, retain the original packing materials (including the outer box) and then
immediately notify the carrier and the relevant Yokogawa sales ofce.
TDLS Analyzer
MODEL
TDLS200
SUFFIX
STYLE
SUPPLY--- 24.0 VDC
MAX 120W
AMB TEMP-20 TO 50
NO.
THIS PRODUCT COMPLIES WITH
21 CFR PART 1040.10
Made in USA
KCC-REMYCA-EEN999
Make sure the model number on the
nameplate of the instrument agrees with your
order.
The nameplate will also contain the serial
number and any relevant certication marks.
Be sure to apply correct power to the unit, as
detailed on the nameplate.
For products used within the European
Community or other countries requiring the CE
mark and/or ATEX classication, the following
labels are attached (as appropriate):
IM 11Y01B01-01E-A 6th Edition :Feb 13, 2013-00
<2. INTRODUCTION AND GENERAL DESCRIPTION>2-22-3
TDLS200 Instruction Manual V2.1
For Zone 2 (CAT 3) ATEX use the following labels will be attached as appropriate:
For YR-200 (Remote Interface Unit, RIU) Zone 2 (CAT 3) ATEX use the following labels will be attached
as appropriate:
IM 11Y01B01-01E-A 6th Edition :Feb 13, 2013-00
<2. INTRODUCTION AND GENERAL DESCRIPTION> 2-4
CAUTION - For Cleaning of the labels and LCD window, please use wet cloth to avoid electrostatic
condition.
NOTE - ATEX Hazardous Area Operation:
Product MUST NOT be used in Zone 0 (CAT 1) locations
Product MUST NOT be used in Group I (Dust/Grain) locations
Product MUST NOT be used in Group III (Fibers) locations
Conditions of Certication
On loss off purge an alarm shell be made to inform the user, action shall then be taken by the user to
ensure continued use is safe.
A functional test shall be carried out in accordance with clause 17.1 of EN 60079-2:2007 to verify the
parameters of the Purge Control Unit when tted.
A leakage test shall be carried out in accordance with clause 17.2 of EN 60079-2:2007. The
manufacturer shall record and retain these results.
Only Lithium batteries specied in manual are to be used in this enclosure.
Special Conditions of Certication:
A suitability certied Purge Control Unit must be sued with the TDLS Analyzer that is capable providing
the requirements listed on label/certicate and that either provides a suitable exhaust through a particle
barrier of to a safe area.
When installed there shall be a minimum of two pressure regulators in the air/nitrogen supply line.
Materials of Construction
The analyzer incorporates a variety of materials in its construction and they should therefore be
used in an appropriate manner. Any chemicals (liquid or gas) that may have a detrimental effect on
the product’s structural integrity should not be allowed come in contact.
The electronic enclosures are constructed from Aluminum Alloy AL Si 12 (ASTM A413) and have a
protective epoxy powder coated surface nish.
The welded bodies are constructed of stainless steel grade 316
The fasteners are constructed of stainless steel grade 18-8
The windows (when tted) are constructed of laminated safety glass
IM 11Y01B01-01E-A 6th Edition :Feb 13, 2013-00
<2. INTRODUCTION AND GENERAL DESCRIPTION>2-42-5
Maintenance Work by Qualied Personnel
Unqualied work on the product may result in severe personal injury and/or extensive damage to
property. If the Warnings contained herein are not adhered to the result may also be severe
personal injury and/or extensive damage to property.
This product is designed such that maintenance work must be carried out by trained personnel.
Trained personnel are considered as below:
- Engineers familiar with the safety approaches of process analytical instrumentation (and/or
general automation technology) and who have read and understood the content of this
User Guide.
- Trained start-up/commissioning analyzer technicians who have read and understood the
content of this Instruction Manual.
WARNING – Battery replacement
Replacement Battery Installation (Type CR2032 located on CPU).
The battery MUST be factory installed and cannot be installed by others at site (soldered connections,
required) – Contact factory for further assistance
IM 11Y01B01-01E-A 6th Edition :Feb 13, 2013-00
3 GENERAL SPECIFICATIONS
0.5A@125 VAC
<3. GENERAL SPECIFICATIONS>
3G with purge system EEx pz II T5
Class 1 Div.2 Group BCD with integral purge
kit
KC mark: KCC-REM-YCA-EEN999
3-1
USB1 and USB2 connection for data
transfer using memory stick, data storage
in CF card (result les, spectra capture,
conguration data, etc.) Capture rate is
congurable typical capacity for results and
spectra is 14 days.
2”, 3” or 4” 150# ANSI RF or adaptors for
DN50 PN16, and DN80 PN16
Response time: 2-20 seconds, plus transport time for
extractive systems when applicable
Drift: Application Dependant
Installation Specifications
Hazardous Area: Zone 1: Contact Yokogawa Zone 2: ATEX group II Cat. 3G with purge
system EEx pz II T5 (-20< Ta <50C)
By Design:Non-Hazardous Area; Purge required forATEX zone 1&2 and NEC Class 1
Division 1&2
Maximum Distance between Launch and Detect:
30 m (±90ft)
Maximum interconnecting cable 50m
Wetted Parts: Analyzer & standard Alignment Flange - 316
SS, BK-7 Glass, Teflon encapsulated Viton
and Silicone RTV sealant.
Basic System Configuration
The TruePeak can be installed in a number of ways depending on
process requirements. The most typical installation types are shown
below, however other installation methods are possible, please contact
Yokogawa with your application details.
Cross Stack/Pipe Configuration
• Measures directly across process pipe or vessel
• Typically has nitrogen or other purge gas protecting
process windows
• Span Validation via serial flow cell (see Operation Specifications).
• Full calibration requires removal from process
• May require pressure and temperature inputs
(Application Dependant)
• Multiple methods to increase Optical Path Length (OPL) if needed
• 5 meter interconnection cable standard
Close Coupled Extractive / Bypass Configuration*
Optional: Isolation Flanges and Flow Cells - 316 SS,
Sapphire, Kalrez Also available in Monel A400, Hastelloy
C-276, Carpenter 20, Titanium Grade 2 and
others on request.
Utilities:
Instrument Air may be used as a purge gas in principle for all of the
below applications, but this will depend on the application type and the
required precision of the measurement.
Oxygen Analyzer N
CO Analyzer N2 or Instrument Air
CO
Analyzer N2 or other non-CO2 containing inert gas
2
H
O ppm Analyzer N2 with <20ppm levels H2O for feed to
2
H
O % Analyzer N
2
2
optional Dryer Package
2
Flow Rate: • 5-30 L/min for window purge
• 2 L/min for validation, calibration and optical
purge
SIL Assessment:
The TDLS200 has a FMEDA assessment by exida and is
classified as a Type B1 device in compliance with the following
standards; IEC 61508 or EN 954-1. Functional Safety of Electrical/
electronic/programmable electronic related systems; SIL 1
capability for single device.
* The TDLS200 is not SIL certified as standard; to be certified the unit
must be specified and designed from the beginning to meet all SIL
specifications.
Cal
inlet
PROCESS
• Measures across a section of pipe where process flow
is directed
• The measurement section can be isolated from process flow for full
calibration/validation, zero and span
• Process pressure and temperature can be controlled
or the analyzer may require pressure and temperature inputs
(Application Dependant)
• Length of measurement section dependant on accuracy requirements
and process conditions
Extractive Configuration*
process
sample
inlet
Zero check gas,
at grade
Span check gas.
at grade
Flow-Cell-Outlet
• Sample is fully extracted from process
(and may be conditioned before measurement).
• Flow cells are available with ability to purge in front of windows
(balanced flow cell) if required.
• Process pressure and temperature can be controlled
or the analyzer may require pressure and temperature inputs
(Application Dependant)
• Length of flow cell dependant on accuracy requirements
and process conditions
* Contact Yokogawa for further details
IM 11Y01B01-01E-A 6th Edition :Feb 13, 2013-00
<3. GENERAL SPECIFICATIONS>
3-3
Standard Accessories
Calibration Cell: - Used for off-line calibrations and
validations
- Stainless steel 316 with free standing
frame
- Connects Launch and Detect with
72.6cm (28.6") OPL
Flow Cells:- Used for extracted sample streams at any
location
- 316SS low volume fixed alignment; 50ºC,
5.5 bar (80psig) max
- Enhanced for 200ºC, 20 Bar (290psig),
Sapphire window, Kalrez o-rings and can
be constructed from 316SS, Monel A400,
Hastelloy C-276, Carpenter 20 and other
materials on request to suit the process
Isolation Flanges:- Used for additional protection for in-situ
or by-pass installations
- 2” or 3” 150# or 300# ANSI RF, 4”150#,
DN80 PN16 welded 5/8” or M16” bolt
studs included sapphire 20 Bar (290 psig)
or BK-7 5.5bar (80 psig) isolation window
- Kalrez window seal o-ring rated max
200ºC
- 316SS, Monel A400, Hastelloy C-276,
Carpenter 20, other on request
Note: Must use in conjunction with alignment flanges
Utility Panel: - Used for convenient field installation of
utilities, configurations for
- Single, dual or four analyzers
- Manual or automatic on-line validation
(controlled by analyzer)
- Safe area (GP), Div 2 purged or nonpurged, ATEX CAT 2G components
- Purge flowmeters with integral needle
valve, glass tube variable area
- Swagelok double ferrule stainless steel tube
fittings and tubing standard
- Panel mounted or fiberglass (NEMA 4X/
IP65), with viewing window
- 5A 24VDC power supply, output to
analyzer – requires VAC input power
Display and Software Functions
TruePeak Software has multiple levels, the default (or start page) is
the Main Menu:
Main Menu Displays: - Concentration & Units (% or ppm)
Options/U ----------------------- Ext.USB Port IP66 (NOT ATEX)
/P ----------------------- Pressure Comp Curve
/D ----------------------- Diverging Beam No Large Aperture Optics
IM 11Y01B01-01E-A 6th Edition :Feb 13, 2013-00
TDLS200 TDL Analyzer Instruction Manual V2.1
Launch to Detect Interconnect (cable)
Launch Unit:
•Main Electronics Housing
•User Interface (optional)
•Laser Assembly
•Check GasFlowCell (for
On-Line Validation)
Detect Unit:
•Detect Electronics
Housing
•Detector
Assembly
Process Interface:
•Analyzer detachable from process interface for Off-Line
calibration/ service.
•Flanged O-Ring Alignment
•Flanged Metal Bellows Seal Alignment
•Flow Cell
•Isolation Flanges
•By-Pass Piping
•Custom designs for specific applications.
Hazardous Area Purge (optional)
•
NEC/CSAClass 1,Div.2,Gr. A-D
•ATEXZone2 Cat3
4 ANALYZER COMPONENTS
<4. ANALYZER COMPONENTS>
4-1
Launch Unit:
• Main Electronics Housing
• User Interface (optional)
• Laser Assembly
• Check Gas Flow Cell (for
On-Line Validation)
Hazardous Area Purge (optional)
• NEC/CSA Class 1, Div. 2, Gr. A-D
• ATEX Zone 2 Cat 3
Launch Unit:
• Analyzer detachable from process
interface for Off-Line calibration / service.
• Flanged O-Ring Alignment
• Flanged Metal Bellows Seal Alignment
• Flow Cell
• Isolation Flanges
• By-Pass Piping
• Custom designs for specic applications.
Detect Unit:
• Detect Electronics
Housing
• Detector Assembly
Launch to Detect Interconnect (cable)
Figure 4 - System Overview
• The Launch Unit and Detect Unit are connected to each other via a Tray Rated 4-pair shielded twisted
pair cable.
• The Launch Unit requires a single 24VDC power supply (by customer or via optional Power Supply
Unit).
• Nitrogen purge gas is required to prevent ambient oxygen ingress however, for other target gases it
may be possible to use Instrument Air for purging.
• The Process Interfaces are available in various formats, sizes and materials to suit the desired
measurement/installation.
• The available Remote Interface Unit (RIU) can be located typically up to 100m (330ft) away from the
Launch Unit. The RIU also requires a 24VDC power supply. The RIU connects to the Launch Unit on
Ethernet (10-base-T 10/100) via CAT5e eld rated cable.
• The available Universal Remote Display (URD) can be located typically up to 40m (120ft) away from the
Launch Unit. The URD requires an AC power input that is connected to a universal power supply with
24VDC output power supply (for the analyzer). The URD connects to the Launch Unit via multipair
sheilded twisted pair cable.
IM 11Y01B01-01E-A 6th Edition :Feb 13, 2013-00
4.1 Launch Unit
Main Electronics Housing
• Back Plane circuit board
• Single Board Computer (SBC)
• FPGA signal Processing board
• Analog I/O circuit board
• Field electrical terminals are located on Back Plane
(and optional Analog I/O board).
• Optional Mini Display (4x20 VFD) shown
<4. ANALYZER COMPONENTS>
Check Gas Flow Cell
Short cell (gas tight chamber)
allows Zero Gas or Span gas to
flow through the measuring path for
on-line validation)
4-2
Figure 5 - Launch Unit - Optional Keypad and Display
Laser Housing and Laser Module
• Laser diode and collimating lens assembly
• Laser module designed to be field
replaceable and purged to prevent
ambient air ingress.
• Housed in a stainless steel body with
O-rings seals, attached to the main
Die cast copper free aluminum grade AL Si 12 alloy (A413.0) with a powder coat exterior nish. The copper free
aluminum alloy is particularly resistant to salt atmospheres, sulfur gases and galvanic corrosion.
An externally hinged door opening to the left incorporates a weather tight gasket seal and four captive fastening
screws (stainless steel). The external dimensions are approx 16” W x 12” H x 7”
D (400mm x 300mm x 180mm).
The environmental protection rating is considered IP65 (EN 60529) or NEMA 4X.
Cable entries are located on the bottom face of the enclosure. They are typically ¾” Myers hubs that have ¾”
NPT female threads. Each has a ground lug to facilitate the grounding of cable shields to the analyzer chassis.
When an analyzer has been supplied with the optional Mini Display (4x20 VFD), the normally blank
(blind) door has a different conguration. The center of the door has a cut-out measuring approx 3”
W x 1” H (75mm x 25mm). A clear laminated safety glass window is mounted to the inside of the door with stainless steel fasteners and a weather tight gasket. This allows for external viewing of the actual VFD display without
opening the door.
When an analyzer has been supplied with the optional integral 6.5” display and keypad, then the normally blank
(blind) door has a different conguration. The left hand side of the door has a cutout measuring
approx 5” W x 4” H (130mm x 100mm). A clear laminated safety glass window is mounted to the inside of the
door with stainless steel fasteners and a weather tight gasket. This allows for external viewing of the actual
display without opening the door. The right hand side of the door accommodates a keypad (30 keys, stainless
steel) which is also operated externally without opening the door.
Backplane Circuit Board
Large (approx. 10” H x 15” W) printed circuit board that mounts inside the enclosure. The board has
several integral circuits and several connectors to accommodate various plug-in boards. The board is
designed such that any eld terminations are located along the lower edge of the board via pluggable
terminal blocks for customer or eld cable interface.
All components and devices on the board are designed for extended temperature (-20 to +80ºC) and low drift
operation.
The Backplane Circuit Board contains the following integrated circuits:
• DC Power Input
• DC Power Distribution
• Watchdog Circuit
• Display Backlight Power Interrupt
• Alarm Relays
• Remote Calibration Initiation
• Calibration Valve Driver Relays
• Laser Temperature and Current Control
• Board temperature
DC Power Input
There are four pluggable screw terminals located on the lower right hand side of the Back Plane. These are used
for connecting the 24VDC power input supply.
There is an adjacent On/Off miniature toggle switch and re-settable thermal fuse.
The single 24DVC power supply is distributed to various output power channels. Each output power channel
has the appropriate DC-DC converter, regulator(s), ltering capacitors and status LEDs, etc.
IM 11Y01B01-01E-A 6th Edition :Feb 13, 2013-00
<4. ANALYZER COMPONENTS>
Relay Coil
24VDC
TB3
DGND
wired outputs to prevent
4-4
Watchdog Power Interrupts
The power output channels for microprocessors have control logic lines (TTL activated). These allow for
watchdog interrupt/reset functionality.
Alarm Relays
There are three alarm relay circuits on the board. These are capable of actuating Form C Single Pole Double
Throw (SPDT) relays. The three connections of each relay (Common, Normally Open and Normally Closed) are
routed through the board to eld terminals.
The contacts are rated for a maximum of 1A @ 24VDC.
The pluggable eld terminals are mounted on the lower edge of the board, just to the left side of
the DC power input terminals. The appropriate relay(s) is actuated when there is an analyzer
Warning, Fault and/or Level Alarm.
Remote Validation/ Remote Calibration Initiation
A validation/calibration routine can be initiated from a remote location (up to 300m away) using contact
closures. The Back Plane has circuitry such that it can monitor for a return voltage. The return voltage comes
from remote Volt Free Contacts (VFCs) at the customer DCS (or other control system).
The circuits include suitable protection against inadvertent shorting/grounding of the supply 24VDC or the
application of excess power to the monitoring circuit. There are three sets of remote contact monitoring circuits
on the Back Plane.
Valve Relays
There are three calibration valve relay circuits on the board. These are capable of actuating Form C SPDT
relays. The common pole is connected to 24VDC power and the normally open pole is routed to the eld
terminal block. Digital ground is also routed to the terminal block TB3 as shown below.
FPGA
TTL out
Relay Coil
Drive Circuit
Relay Coil
Relay
Relay
Contacts
Contacts
24VDC to
24VDC to C
C and to NO
C to NO
NOTE; Use ferrite coil or direction diode on TB6
24VDC to external
switching spikes
Figure 8 - Calibration Valve Relay Diagram
24VDC 12W max to
solenoid valve when
external solenoid valve
relay is engergized
when relay is energized
IM 11Y01B01-01E-A 6th Edition :Feb 13, 2013-00
<4. ANALYZER COMPONENTS>
Connections of each relay (Common and Normally Open) are routed through the board to eld terminals.
The contacts are rated for a maximum of 1A @ 24VDC (or 0.5A @ 125VAC).
The pluggable eld terminals are mounted on the lower edge of the board, just to the left side of the DC power
input terminals.
The appropriate relay(s) is actuated when a calibration gas check valve is to be initiated.
Laser Temperature & Current Control
The board has two main laser control function circuits, temperature control and laser current control.
Board Temperature
The board has a temperature sensing chip/circuit that monitors temperature of the board inside the main
electronics enclosure. The sensor is located on the top edge of the Back Plane.
Backplane Circuit Board Power & Signal Routing
The Back Plane carries out several routing functions for both power and signals: I/O for Detect Unit is routed
through the Back Plane from one set of pluggable eld terminals (located lower left hand of Back Plane) to the
appropriate destination. Terminals are provided for:
• Analog DC power (x3)
• Raw Detector Signal (differential voltage) (x2)
• Detect Unit Temperature (differential voltage) (x2)
Analog I/O Board outputs the analyzer results and reads input process gas compensation values (pressure and
temperature). The board has power status LEDs as well as voltage test points for the input and
output channels.
• Output channels (three) are ranged 0-20mA. They can be assigned to measured values Oxygen,
Transmission or compensation signal re-transmission.
• Input Channels (two) are used by the analyzer to read active values for process gas temperature
and/or process gas pressure. These are application dependant and may or may not be required
inputs. There are two channels, one for temperature and one for pressure. Each may be used to
read 4-20mA signals that are isolated or to read and loop power (with integral 24VDC) signals.
IM 11Y01B01-01E-A 6th Edition :Feb 13, 2013-00
<4. ANALYZER COMPONENTS>
4-6
Optional Mini Display (4x20 VFD) mounts on the analyzer enclosure door. The display itself is an indus-
trial grade 4 line 20 character vacuum uorescence display (VFD) that is self illuminating (i.e. no back light
required).
Optional 6.5” Display is an industrial grade 6.5” VGA color TFT LCD Module that has a built-in CCFL
backlight. Both the display and interface board are mounted to a cover plate that attaches to the inside of
the enclosure door.
Optional Keypad is an industrial rated 30 key unit that has a PS/2 (6-pos miniDIN) interface direct to the
SBC. It has an Ingress Protection Rating of IP65 equivalent to NEMA 4X and is of low prole design.
Backplane Field Terminal Blocks:
• TB1 - 24VDC Power input 80 w (and optional purge power)
Laser Assembly Body is a stainless steel mechanical pipe
housing that accommodates the module and protects it from the
environment. The body has two Swagelok style tube fittings welded
on that serve as inlet and outlet ports for the nitrogen purge gas. The
body attaches to the Main Electronics Housing with an O-Ring seal
and several stainless steel fasteners. At the other end of the body
there is a standard adaptor piece welded in place. This adaptor
can fit several different Process interface systems as well as an off-
line calibration cell. The adaptor also accommodates the standard
Process isolation window holder.
Laser Module
Laser Module is a mechanical component that holds both the
laser diode and the lens holder. The assembly is factory setup, permanently configured and can be replaced in the field if
necessary.
Body
Laser
Module
Figure 9 - Laser Assembly
Window
Figure 10 - Laser Module
IM 11Y01B01-01E-A 6th Edition :Feb 13, 2013-00
<4. ANALYZER COMPONENTS>
Laser Diode is either a Vertical Cavity Surface-Emitting Laser (VCSEL) or Distributed Feedback (DFB) that
outputs at wavelengths in the 750nm to 2400nm range (invisible)
depending on the target gas being measured. The primary
output wavelength of the laser is controlled by a thermoelectric
cooling module (Peltier Element). The laser diode is permanently
attached to the module. Collimating Lens is an optical
component that collimates the diverging light source.
4.4 Check Gas Flow Cell (for On-Line Validation and/or
Laser
Module
Check Gas
Body
Line Locking
Flow Cell
The Check or Check Gas Flow Cell is a short chamber that exists
between the laser collimating lens and the standard Process
isolation window. The cell is sealed with double O-rings and is
in series with the measurement optical path. The body has two
Figure 11 - Check Gas Flow Cell
Swagelok style tube fittings welded on that serve as inlet and
outlet ports for the nitrogen purge gas or calibration check gas as
appropriate.
The Check gas flow cell is used for performing on-line validations (or Dynamic Spiking) while the analyzer is mounted
on the Process. This feature allows for the analyzer to be validated without removing it from the Process location.
By introducing a gas of known target gas concentration, at a given temperature and pressure, the analyzer can
determine if the Validation routine has been PASSED or FAILED.
Window
This cell can also be used for Line Locking applications, such as %CO for combustion applications. Refer to
Non-Process Parameters for details of how to configure the software when implementing a line-locking application.
Please also refer to project specific drawings for detail of how to configure the tubing/valving when implementing line
locking.
The various parameters that enable the validation are all configurable within the TDLS200 software. Refer to the
Validation and Calibration section of this User Guide for further details.
IM 11Y01B01-01E-A 6th Edition :Feb 13, 2013-00
4.5 Detect Unit
<4. ANALYZER COMPONENTS>
4-8
Detect Electronics Housing
• Detector Circuit Board
Process Interface
Detect unit
Detect
Electronic
Housing
Detect
Assembly
Detector Housing and Detector Module
• Detector and focusing lens assembly
• Detector module designed to be field replaceable
and purged to prevent ambient air ingress.
• Housed in a stainless steel body with O-rings
seals, attached to the detect electronics housing.
Figure 12 - Detect Unit
Detect or Electronics Housing
The Enclosure is die cast copper free aluminum grade AL Si 12 alloy (A413.0) with a powder coat exterior finish.
The copper free aluminum alloy is particularly resistant to salt atmospheres, sulfur gases and galvanic corrosion. A
removable cover (lid) incorporates a weather tight gasket seal and four captive fastening screws (stainless steel). The
external dimensions are approx 7” W x 7” H x 4” D (180mm x 180mm x 100mm).
The environmental protection rating is considered IP65 (EN 60529) or NEMA 4X.
The cable entry located on the bottom face of the enclosure. It is typically a ¾” Myers hub that has a ¾” NPT
female thread. It has a ground lug to facilitate the grounding of cable shields to the analyzer chassis.
Detector Circuit Board
Detector Circuit Board main function is to convert detector photocurrent into voltage and send it to be digitized.
LEDs are incorporated to provide simple diagnostic of available power. The board has a temperature sensing chip/
circuit that monitors the ambient temperature inside the detect electronics enclosure. The sensor is located on the
top edge of the detect board to ensure the maximum temperature reading is monitored.
The board is medium size (approx. 4” H x 6” W) printed circuit board that mounts inside the enclosure. The field
terminations are located along the lower edge of the board via pluggable terminal block. All components and
devices on the board are designed for extended temperature and low drift operation.
IM 11Y01B01-01E-A 6th Edition :Feb 13, 2013-00
<4. ANALYZER COMPONENTS>
TDLS200 TDL Analyzer Instruction Manual V2.1
The cable entry located on the bottom face of the enclosure. It is typically a ¾” Myers hub that has a ¾” NPT female thread. It has a ground lug to facilitate the grounding of cable shields to the analyzer chassis.
Detector Circuit Board main function is to convert detector photocurrent into voltage and send it
to be sampled.LEDs are incorporated to provide simple diagnostic of available power. The board has a temperature sensing chip/circuit that monitors the ambient temperature inside the
detect electronics enclosure. The sensor is located on the top edge of the detect board to ensurethe maximum temperature reading is monitored.
The board is medium size (approx. 4” H x 6” W)printed circuit board that mounts inside the enclosure. The field terminations are located along the lower edge of the board via pluggable terminal block. All components and devices on the board are designed for extended temperature and low drift operation.
2.3.7 Process Interface
An appropriate Process Interface is selected to suit the process/stack installation. The analyzer is detachable from the process interface to facilitate Off-Line calibration and service.
Process Interface Options
There are several systems available as well as custom designs for specific applications.
Flanged O-Ring Alignment comprises typical 2” or 3”
process flange with a large diameter O-Ring seal, typically used for stack or inert applications that are non-corrosive.
Flanged Metal Bellows Seal comprises typical 2” or
3” process flange with a metal bellows seal and external mechanical alignment system, typically used when gas containment is important.
Flow Cell may be used when the process gas has
been extracted or is used in a by-pass flow loop.This allows for heat tracing (if necessary) and easy introduction of both Zero and Span gases.
Isolation Flanges are supplied with process windows
mounted in the flanges themselves typically for very corrosive and/or high pressure applications.
By-Pass Piping may be used when the process gas
line is of small diameter (typically <10”) and when no suitable section of pipe work exists on this which the analyzer can be mounted. By pass piping systems may be provided or supplied at site. Heat tracingmay also be supplied or provided at site.
Off Line Calibration Cell is used for off line
verification of the system. This cell is used to flowZero and Span gases when the analyzer is not connected to the process.
4.6 Process Interface
An appropriate Process Interface is selected to suit the process/stack installation. The analyzer is
detachable from the process interface to facilitate Off-Line calibration and service.
Process Interface Options
There are several systems available as well as custom designs for specic applications.
Flanged with Metal Alignment
• Flanged O-Ring Alignment comprises typical 2” or 3”process
ange with a large diameter O-Ring seal, typically used for
stack or inert applications that are non-corrosive.
• Flanged Metal Bellows Seal comprises typical 2” or 3”
process ange with a metal bellows seal and external
mechanical alignment system, typically used when gas containment is important.
• LAO-Large Aperature Optics, for long path combustion
application.
extracted or is used in a by-pass ow loop. This allows for heat tracing (if necessary) and easy introduction of both Zero and Span gases.
mounted in the anges themselves typically for very corrosive and/or high pressure applications.
small diameter (typically <10”) and when no suitable section of pipe work exists on this which the analyzer can be mounted. By pass piping systems may be provided or supplied at site. Heat tracing may also be supplied or
provided at site.
Typical Cross Pipe/Stack
Typical Flow Cell / By-Pass
• Flow Cell may be used when the process gas has been
• Isolation Flanges are supplied with process windows
• By-Pass Piping may be used when the process gas line is of
• Off Line Calibration Cell is used for off line verication of the
system. This cell is used to ow Zero and Span gases when the analyzer is not connected to the process.
Figure 13 - Process Interface Examples
IM 11Y01B01-01E-A 6th Edition :Feb 13, 2013-00
<4. ANALYZER COMPONENTS>
4.7 Analyzer Connections
Launch – Detect Interconnect
The two units are connected to each other via a four, twisted pair cable suitable for tray installation outdoors.
Pluggable terminals strips are provided at both units to enable eld termination of the cable. The cable pairs are
individually shielded as well as an overall shield.
The cable specications are as below.
ItemSpecication
Number fo Pairs4, individually shielded
Total Number of Conductors9 (includes 1 comm.)
2
AWG18 (0.75 mm
Conductors7 x 26 stranding, Bare Copper
Inner ShieldAluminum Foil-Polyester tape, 100% coverage with 20
SuitabilityIndoor, Outdoor, Burial and Sunlight Resistant
Power Limited Tray Rated Cable
Nom. Conductor DC resistance @ 20˚C5.86 Ohms/1000 ft (305 m)
Nom. Outer Shield DC resistance @ 20˚C4.75 Ohms/1000 ft (305 m)
Max. Operating Voltage - UL300 V RMS
Conductor IdenticationNumbered pairs, black & white conductors
Typical Manufacturer & Part No.Belden Type 1475 A
The maximum cable length should not exceed 150 ft (46 m).
Please ensure that the Launch to Detect cable is properly terminated and that all grounding and shielding
details are correct per installation drawings-Espically important for CE/A TEX
installations.
IM 11Y01B01-01E-A 6th Edition :Feb 13, 2013-00
<4. ANALYZER COMPONENTS>
4.8 Communications
Stand Alone Options
The analyzer is capable of fully independent operation with no external computer or interface required. A number
of options are available for a built in user interface (mounted on Launch Unit):
• Blind with no display or keypad. Access to the analyzer through; Ethernet connection (local or remote
computer), Remote Interface Unit (RIU), Universal Remote Display (remote display only - no keypad) with
menu access via external computer.
• Mini display which is an Integral display 4X20 smart VFD (cycles information). No keypad,
menu access via local or remote external computer (Ethernet connected).
• Keypad with 6.5” display.
• Regardless of the user interface selected the analyzer will continuously record results,
diagnostics and spectra. Data can be transferred from the analyzer via USB or Compact
Flash.
Remote Interface Options
A number of options are available for remote access to the analyzer
Remote Interface Unit (RIU) model YR200 shown below, allows remote analyzer control and data
transfer from analyzer to RIU (data can be transferred from RIU via USB memory stick or
Compact Flash card.
• Allows multi-unit eld communication via
central user interface
• Not required for individual analyzer
operation, interface and data transfer only
• Connects with 1-8 analyzers via Ethernet
switch
• Integral Keypad and 6.5” display
External Computer via Ethernet. A
separate computer can be connected to the
analyzers locally or through an Ethernet
network to allow analyzer control and data
transfer
• All eld electrical terminals are located on the Back Plane.
A single RIU can be used in conjunction with up to 8 analyzers via Ethernet (more with additional/custom
Ethernet switches).
IM 11Y01B01-01E-A 6th Edition :Feb 13, 2013-00
<4. ANALYZER COMPONENTS>
TDLS200 TDL Analyzer Instruction Manual V2.1
The unit acts as a remote interface for the analyzer. Should the physical location of the actual analyzer(s) be
inconvenient for easy access, then the RIU can be used.
It can be mounted up to 100m (330ft) away from the analyzer(s) using the standard 10-BaseT twisted pair
wiring method. It communicates to the analyzer(s) through a Virtual Network Connection (VNC). If there is more
than one analyzer connected to the RIU, then they are routed via an industrial Ethernet switch. Up to four
analyzers can be routed through one RIU switch.
The RIU Enclosure is die cast copper free aluminum grade AL Si 12 alloy (A413.0) with a powder coat exterior
nish. The copper free aluminum alloy is particularly resistant to salt atmospheres, sulfur gases and galvanic
corrosion. An externally hinged door opening to the left incorporates a weather tight gasket seal and four
captive fastening screws (stainless steel). The external dimensions are approx 16” W x 12” H x 7” D (400mm x
300mm x 180mm). Wall mounting brackets are included with the RIU.
4-12
The environmental protection rating is considered IP65 (EN 60529) or NEMA 4X. Cable entries are located on the
bottom face of the enclosure. They are typically ¾” Myers hubs that have ¾” NPT female threads. Each has a
ground lug to facilitate the grounding of cable shields to the chassis.
The RIU is supplied with standard integral display and keypad.
RIU Interconnect to Launch Control Unit(s)
When connecting just one analyzer to the RIU there are two twisted pair wires to consider , there
are only four wires to be terminated to make the 10/100 Ethernet connection.
Analyzer
SBC
Analyzer TB6
Tx Tx Rc Rc
RIU
SBC
RIU
Tx Tx Rc Rc
Figure 15 – Connecting RUI to Analyzer(s)
IM 11Y01B01-01E-A 6th Edition :Feb 13, 2013-00
<4. ANALYZER COMPONENTS>
TDLS200 TDL Analyzer Instruction Manual V2.1
RIUOptionalEthernetSwitch
If there is more than one analyzer connected to the RIU, then they are routed via an industrial Ethernet switch. Up to four analyzers can be routed through one RIU switch. The switch is powered by 24VDC from the back-plane and includes several status LEDs.
RIU Optional Ethernet Switch
If there is more than one analyzer connected to the RIU, then they are routed via an industrial Ethernet switch.
Up to four analyzers can be routed through one RIU switch. The switch is powered by 24VDC from the
back-plane and includes several status LEDs.
RIU
SBC
Analyzer 1
SBC
Analyzer TB6
Figure 16 – RIU Ethernet Switch
Analyzer 2
SBC
Analyzer TB6
FeedThrough
Board
Ethernet
Switch
Feed-through
Board
RIU Optional Feed-through Board
To facilitate the connection of more than one analyzer to the RIU, an optional Feed-through board can be used.
The board has pluggable screw terminals that allow for the landing of eld cables from the analyzers at the RIU.
RIU Hazardous Area Purging
The standard RIU is designed for operation in Safe Areas (General Purpose). An optional Z-Type purge control
system can be tted to the RIU and it includes a local indicator and pressure switch alarm contacts. When
applied, the purge system allows for operation in:-
• NEC/CSA Class 1, Division 2, Groups A-D
• ATEX Zone 2 CAT 3 (dual regulators at the inlet MUST be used)
The purge gas may be either Instrument Air or Nitrogen.
4.9 Purge Systems
The TDLS200 Analyzer requires a continuous nitrogen gas purge to prevent ambient oxygen ingress to the optical path, when oxygen is the measured gas. The ow rate can be minimized as long as it prevents any ambient
oxygen ingress to the measurement optical path. Other purge gases may be used as long as they do not contain
any of the measured gas and they are clean, dry, etc.
For hazardous area operation, the same nitrogen purge gas is used to purge the entire analyzer (including
non-optical path sections such as the electronics). The process interface may also require purging to maintain
clear windows, refer to Process Window Purge details separately.
IM 11Y01B01-01E-A 6th Edition :Feb 13, 2013-00
<4. ANALYZER COMPONENTS>
TDLS200 TDL Analyzer Instruction Manual V2.1
Purging Analyzer for Safe Area
The block diagram below shows the sections of the analyzer that require nitrogen purging. The purging should be carried in sequence typically as shown below.
TDLS200 TDL Analyzer Instruction Manual V2.1
Purging Analyzer for Safe Area
The block diagram below shows the sections of the analyzer that require nitrogen purging. The purging should be carried in sequence typically as shown below.
Figure 12 - Purging for Safe Areas
Purging Analyzer for Hazardous Areas
Z-Purged designed in accordance with NEC/CSA Class 1, Division 2, Groups A-D ATEX Zone 2 CAT 3 (Certified)
The block diagram below shows the sections of the analyzer that require nitrogen purging. A Z-Type purge control system is fitted the Main Electronics Housing and it includes a local indicatorand pressure switch alarm contacts. The purging should be carried in sequence typically as shown below.
MainElectronic Housing
LaserModule
Check Gas Flow Cell
Process
Interface
Process
Interface
DetectModule
DetectElectronic Housing
Nitrogen or I/A Purge Gas
Nitrogen or I/A Purge Gas
4-14
Purging Analyzer for Safe Area.
The block diagram below shows the sections of the analyzer that require nitrogen purging. The purging should
be carried in sequence typically as shown below.
Nitrogen or I/A Purge
Gas
Main
Electronic
Housing
Laser
Laser
Module
Module
Nitrogen or I/A Purge
Nitrogen or I/A Purge Gas
Gas
Check Gas
Flow Cell
Process
Interface
Process
Interface
Detect
Detect
Module
Module
Detect
Electronic
Housing
Figure 17 – Purging for Safe Areas
Purging Analyzer for Hazardous Areas
• NEC/CSA Class 1, Division 2, Groups A-D
• ATEX Zone 2 CAT 3
The block diagram below shows the sections of the analyzer that require nitrogen purging. A Z-Type
purge control system is tted the Main Electronics Housing and it includes a local indicator and pressure switch
alarm contacts. The purging should be carried in sequence typically as shown below.
Nitrogen or I/A Purge
Nitrogen or I/A Purge
Gas
Gas
Main
Electronic
Housing &
Purge
System
Laser
Laser
Module
Module
Nitrogen or I/A Purge
Gas
Check Gas
Flow Cell
Process
Interface
Detect
Detect
Module
Module
Detect
Electronic
Housing
Process
Interface
Figure 18 – Purging for Hazardous Areas
IM 11Y01B01-01E-A 6th Edition :Feb 13, 2013-00
<5. INSTALLATION AND WIRING> 5-1
5 INSTALLATION AND WIRING
Detailed Installation, Wiring, Utility Drawings are included on a Project Basis. Please contact
Yokogawa for any project specic documentation to ensure correct installation. Drawings provided herein
are considered for standard installation use only
5.1 Process Measurement Point Considerations
The following criteria should be considered when selecting the installation point in respect to the process
conditions:
• Process Gas Flow Conditions – Laminar, homogenous gas concentration distribution conditions
across the measurement point are recommended.
For circular ducts/stacks this condition is generally at least three unimpaired diameters (D) before and
after a process bend. For rectangular cross sections, the hydraulic ductdiameter (D) is derived from:
D = (4 x duct cross sectional area) / duct circumference
If neither situation exists or is possible, then distribution of the unimpaired section of duct should be
66% on the inlet side and 34% on the outlet side. Proling of the proposed measurement point may be
required to ensure that a correct installation point is selected.
• Process Gas Temperature – It is recommended that the analyzer be installed at a location where
temperature uctuations are minimized. Generally as a guide, if the temperature of the gas at the point
where the analyzer is to be installed is to vary by more than +/-10˚C (+/-18˚F) then an “Active” input
signal should be used for compensation.
Ensure the analyzer has been selected and congured to suit the maximum operating gas temperature.
Lower gas temperatures generally lead to better measurements.
• Process Gas Pressure – It is recommended that the analyzer be installed at a location where pressure
uctuations are minimized. Generally as a guide, if the pressure of the gas at the point where the
analyzer is to be installed is to vary by more than +/- 0.05 Bar (+/- 0.725 psi) then an “Active” input
signal should be used for compensation.
Ensure the analyzer has been selected and congured to suit the maximum operating gas pressure.
Ensure the process isolation windows have been selected and congured to suite the maximum design
gas pressure.
Lower gas pressures generally lead to better measurements.
• Process Dust/Particulate Matter – It is recommended that the analyzer be installed at
a location where dust loadings are minimized. Dust and other particulate matter will
reduce the optical transmission of the measuring laser beam. Within limits, the loss of
optical transmission does not effect the measurement however a Warning alarm will
be initiated when the transmission falls below allowable limits. The amount of dust
loading is also dependant upon the optical path length – Consult Factory for further
details.
Lower dust loads generally lead to better measurements.
IM 11Y01B01-01E-A 6th Edition :Feb 13, 2013-00
<5. INSTALLATION AND WIRING>5-15-2
TDLS200 TDL Analyzer Instruction Manual V2.1
3.2 Position of Process Flanges for Launch and Detect Units:
Processflangesshouldbelocated ontheprocesssuchthat theLaunch andDetectUnitscanbeinstalled, accessedremoved in a safe and convenientmanner.
•Clearance for installationand removal of purgeinsertion tubes (if applicable)
•Accessto processisolationvalves
•Saferoutingforinterconnectingcables
•
Ambient conditions in accordance with analyzer limits
•Accessto appropriate utilities
•Adjacent space for mounting to C
alibrationCellwhen off-line
5.2 Position of Process Flanges for Launch and Detect Units:
Process anges should be located on the process such that the Launch and Detect Units can be installed,
accessed and removed in a safe and convenient manner.
The following criteria/Check List should be met at a minimum:
• Good, Safe Engineering practices
• Local codes and regulations for such equipment installation
• Appropriate hazardous area (if applicable) precautions
• Owner Company best practice and engineering standards
• Access for personnel to stand in front of launch and Detect Units
• Clearance for installation and removal of Launch and Detect (see below)
• Clearance for installation and removal of purge insertion tubes (if applicable)
• Access to process isolation valves
• Safe routing for interconnecting cables
• Ambient conditions in accordance with analyzer limits
• Access to appropriate utilities
• Adjacent space for mounting to Calibration Cell when off-line
Figure 23 - Launch and Detect Unit-In -Situ Installation Overview
Figure 22 - and Detect Unit – In-Situ Installation Overview
Figure 24 - Launch and Detect Unit-In-Situ with Insertion Purge Tube Installation Overview
5.2 Fast Loop Line Size and Response Times.....................................9
The following safety symbols are used onthe product as well as in this manual.
DANGER
This symbol indicates that an operatormust follow the instructions laid out inthis manual in order to avoid the risks, forthe human body, of injury, electric shockor fatalities. The manual describes whatspecial care the operator must take toavoid such risks.
WARNING
This symbol indicates that an operator mustrefer to the instructions in this manual inorder to prevent the instrument (hardware)or software from being damaged, or asystem failure from occurring.
CAUTION
This symbol gives information essentialfor understanding the operations and
functions.
Note!
Figure 25 - Launch or Detect Unit Installation Dimensions
The standard ange sizes are either 2” or 3” 150# R.F. ANSI as well as DN50 and DN80. Please check the exact
ange size specied and provided for the particular installation. Other ange sizes and a variety of materials (to
suit the process) are available so please check these details prior to installing the anges on the process.
Note: The process isolation valves should have at least a 1½” (38mm) diameter clear bore size
(aperture) to ensure there is sufcient tolerance to align the laser beam after installation. Ducts and
Stacks that have thin and exible walls should be reinforced to ensure that the laser beam alignment is
maintained at all times. Rigid mounting for the process anges is highly recommended to ensure a
lignment is maintained.
In situations where the process anges are mounted to these thin and exible duct/stack walls, a larger
reinforcing plate should be welded around the mounting ange area to increase the attaching region. The gure
below depicts a typical suggestion however; it is the installer’s responsibility to ensure appropriately rigid
installation is provided for the analyzer.
Figure 26 - Suggested Reinforcing Plate for Launch or Detect Unit Flanges- reduce exing with mechnical
reinforcement to duck/stack walls
IM 11Y01B01-01E-A 6th Edition :Feb 13, 2013-00
<5. INSTALLATION AND WIRING>5-35-4
TDLS200 TDL Analyzer Instruction Manual V2.1
Figure 21 - Reinforcing Plate for Launch or Detect Unit Flanges
3.3 Process Flange Welding Alignment and Line-Up
The Launch and Detect units are provided with alignment mechanisms that allow for some manual adjustment of the laser beam direction in both planes. It is however recommended that the following angular tolerances be adhered to as closely as possible.
5.3 Process Flange Welding Alignment and Line-Up
The Launch and Detect units are provided with alignment mechanisms that allow for some manual adjustment
of the laser beam direction in both planes. It is however recommended that the following angular tolerances be
adhered to as closely as possible.
PROCESS FLANGE
ANGULAR TOLERANCE
Angular tolerance
PROCESS FLANGE
BOLT ALIGNMENT
NOZZLE FLANGE BOLT PATTERN MUST BE AS
INDICATED TO ASSURE PROPER MOUNTING OF
ANALYZER LAUNCH AND DETECT UNITS
COMBINED ANGULAR OFFSET OF BOTH NOZZLES
MUST NOT EXCEED 2” IN ANY DIRECTION. BEST
INSTALLATION WILL HAVE NO ANGULAR OFFSET
Figure 27 - Angular Alignment Tolerances for Launch or Detect Unit Flanges
5.2 Fast Loop Line Size and Response Times.....................................9
The following safety symbols are used onthe product as well as in this manual.
DANGER
This symbol indicates that an operatormust follow the instructions laid out inthis manual in order to avoid the risks, forthe human body, of injury, electric shockor fatalities. The manual describes whatspecial care the operator must take toavoid such risks.
WARNING
This symbol indicates that an operator mustrefer to the instructions in this manual inorder to prevent the instrument (hardware)or software from being damaged, or asystem failure from occurring.
CAUTION
This symbol gives information essentialfor understanding the operations and
5.2 Fast Loop Line Size and Response Times.....................................9
The following safety symbols are used onthe product as well as in this manual.
DANGER
This symbol indicates that an operatormust follow the instructions laid out inthis manual in order to avoid the risks, forthe human body, of injury, electric shockor fatalities. The manual describes whatspecial care the operator must take toavoid such risks.
WARNING
This symbol indicates that an operator mustrefer to the instructions in this manual inorder to prevent the instrument (hardware)or software from being damaged, or asystem failure from occurring.
CAUTION
This symbol gives information essentialfor understanding the operations and
functions.
Note!
5.4 Process Flange Clear Aperture
The Launch and Detect unit anges (and insertion tubes if used) should be installed in such a way that a
minimum clear through aperture (opening) diameter of 1 ½” (38 mm) when the distance between the 2” 150#
ange faces does not exceed 6ft (~2 m).
Note: Larger clear through apertures may be required when operating over longer distances –
CONSULT FACTORY for further details.
PROCESS FLANGE
CLEAR APERTURE TOLERANCE
Figure 28 - Clear Aperture for Launch or Detect Unit Flanges
5.5 Mounting the Launch and Detect Units to the Process Flange
Securely bolt the Launch and Detect Units to the process anges using the standard bolt holes provided. Ensure
the correct size bolts, nuts, and gasket are used in accordance with the ange specications and in accordance
with the process specications when applicable.
NOTE: If the process isolation valve ange is excessively hot due to the process temperature or radiant
heat, then a thermal insolating ange gasket should be used in order to minimize the heat transfer to the
analyzer ange face.
It is generally benecial to make the ange of the Launch and Detect Units and the ange of the process
concentric with each other. Due to the large clearance provided by standard anges and bolts, it is possible to
mount the two anges in an un-concentric manner – this should be avoided to aid laser beam alignment.
LATERAL OFFSET OF NOZZELS MUST BE MINIZED AS TO
ALLOW FOR MINMUM 1.75” [45MM] CLEAR APERTURE.
BEST INSTALLATION WILL HAVE NO LATEAL OFFSET.
IM 11Y01B01-01E-A 6th Edition :Feb 13, 2013-00
<5. INSTALLATION AND WIRING>5-55-6
5.5.1 Process Window Purge Gas Connection
In order to keep the process windows clean (prevent fouling by process gas) it is necessary to purge the
windows with a clean dry gas of sufciently low dew point.
When measuring Oxygen, Nitrogen should be used for purging the windows. The purge gas or nitrogen should
be clean (<0.5 μ particulate), dry (-40˚C dew point), oil free.
The process anges are provided with two diagonally opposed inlet ports (typically ¼” OD tube). Use Swagelok
(or equal) double ferrule tube ttings and connect both ports with ¼” OD stainless steel tubing to the purge gas
supply. The exact purge gas ow rate will be dependant upon the process conditions that exist at the ange
connection and therefore, the ow rate could be anywhere from 5lts/min to 50lts/min
(~10 SCFH to ~100 SCFH).
Ensure the purge gas line is clean and dry prior to connecting to the ange to ensure any condensate or
debris is not blown on to the windows at initial start-up. The same applies to both Launch Unit and Detect
Unit process anges, with and without any insertion purge tubes installed.
5.6 Mounting the Process Interface – standard 2”, 3”. 4” 150# ANSI RF or DN - 50/80
Alignment Nuts (x8)
Window Purge Port (x2)
Analyzer Quick Connect
Alignment Studs (x4)
Analyzer Mounting Flange
Process Flange
Alignment Bellows
Figure 29 3” 150# Welded Metal Bellows Alignment Flange Shown above, with analyzer quick connect
IM 11Y01B01-01E-A 6th Edition :Feb 13, 2013-00
<5. INSTALLATION AND WIRING>5-75-7
5.7 Typical Purge Gas Conguration, In-Situ
Please refer to project specic details, the following is a typical standard in-situ conguration:
Figure 30
5.8 Typical Purge Gas Conguration, Extractive trace ppm H2O system:
Please refer to project specic details, the following is a typical standard enhanced ow cell conguration:
Figure 31
IM 11Y01B01-01E-A 6th Edition :Feb 13, 2013-00
<5. INSTALLATION AND WIRING> 5-8
Typical Purge Large Aperture Optics (LAO) combustion O2, CO/CH4
Figure 32
5.9 Dimensional Drawings
Figure 33 - General Installation Clearance Requirements
IM 11Y01B01-01E-A 6th Edition :Feb 13, 2013-00
<5. INSTALLATION AND WIRING>5-85-9
Figure 34 -Off-Line Calibration Overview
Figure 35 - Launch and Detect unit Dimensions
IM 11Y01B01-01E-A 6th Edition :Feb 13, 2013-00
Figure 36 - Remote Interface Unit
<5. INSTALLATION AND WIRING> 5-10
Figure 37 - Universal remote Display (URD)
IM 11Y01B01-01E-A 6th Edition :Feb 13, 2013-00
Figure 38 - Universal Power Supply (UPS)
<5. INSTALLATION AND WIRING>5-105-11
Figure 39 - Alignment Bellows * Other ange sizes available to meet application needs
IM 11Y01B01-01E-A 6th Edition :Feb 13, 2013-00
<5. INSTALLATION AND WIRING> 5-12
Figure 40 - Large Apenture * Other ange sizes available to meet application needs
Figure 41 - Isolation Flanges, model IF200
IM 11Y01B01-01E-A 6th Edition :Feb 13, 2013-00
Figure 42 - Insertion Tube Detail
<5. INSTALLATION AND WIRING>5-125-13
Figure 43 - Off-Line Calibration Gas and Purge Detail
IM 11Y01B01-01E-A 6th Edition :Feb 13, 2013-00
<5. INSTALLATION AND WIRING> 5-14
Figure 44 - Utility Panel Tubing Detail Standard Single, Auto - validation
Figure 45 - Utility Panel Tubing Detail Standard Dual Auto - validation
IM 11Y01B01-01E-A 6th Edition :Feb 13, 2013-00
5.10 Wiring
5.10.1 Wiring of Launch for the US version
<5. INSTALLATION AND WIRING>5-145-15
Figure 46
IM 11Y01B01-01E-A 6th Edition :Feb 13, 2013-00
<5. INSTALLATION AND WIRING> 5-16
5.10.2 Launch and Detect Unit Wiring – Standard for CE/ATEX
SPECIAL NOTE: Noise sources magnetically coupled to the 4-pair launch to detect cable can cause errors to
the measurement if the frequency and applied total voltage meet or exceed the following; Frequency range of
0.1 to 7.5 MHz and Total induced voltage of 3Vrms.
A combination of these parameters will cause measurement error outside of the analyzer performance
specication. In practice, these potential errors can be easily avoided by routing the 4-pair launch to detect
cable away from electromagnetic interference sources that could exhibit these parameters. Such sources might
include power transformers, electric motors, electric welding machinery, high voltage power lines, etc. Analyzer
grounding wires and any other I/O lines such as 8-pair launch to utility panel cable or A I/O cables to/from the
DCS should also be installed with similar basic practices to also ensure there is no additional adverse inuence
on any of the I/O signals.
Launch and Detect Unit Wiring – Standard GP/Div2 (non CE/ATEX)
Launch Unit (to Detect) Terminations at TB-7
Detect Unit (to Launch) Terminations at TB-13
IM 11Y01B01-01E-A 6th Edition :Feb 13, 2013-00
<5. INSTALLATION AND WIRING>5-175-18
5.10.3 Launch Unit Terminations CE/ATEX/GP/Div2 – all units:
SEE FOLLOWING NOTES:
Please NOTE that ALL analog output signals from TB-8 are POWERED BY THE ANALYER!
Outputs from TB-9 can be congured for POWERED inputs or LOOP POWERED external transmitter!
PLEASE CHECK CAREFULLY BEFORE APPLYING POWER TO THE ANALYZER!
5.2 Fast Loop Line Size and Response Times.....................................9
The following safety symbols are used onthe product as well as in this manual.
DANGER
This symbol indicates that an operatormust follow the instructions laid out inthis manual in order to avoid the risks, forthe human body, of injury, electric shockor fatalities. The manual describes whatspecial care the operator must take toavoid such risks.
WARNING
This symbol indicates that an operator mustrefer to the instructions in this manual inorder to prevent the instrument (hardware)or software from being damaged, or asystem failure from occurring.
CAUTION
This symbol gives information essentialfor understanding the operations and
functions.
Note!
5.11 Hazardous Area Systems
The TDLS200 Analyzer requires a continuous nitrogen or I/A gas purge to prevent ambient oxygen ingress to the
optical path, when oxygen is the measured gas. The ow rate can be minimized as long as it prevents any ambient
oxygen ingress to the measurement optical path. Other purge gases may be used as long as they do not contain
any of the measured gas and are clean, dry, etc. If using a purge gas that also contains the measured gas (e.g.
purge with instrument air and measuring process/combustion oxygen) then the Non-Process Parameters software
feature/parameters will have to be implemented.
For hazardous area operation, the same nitrogen purge gas is used to purge the entire analyzer (including
non-optical path sections such as the electronics).
The process interface may also require purging to maintain clear windows, refer to Process Window Purge details
separately.
NOTE: Please also refer to any separate Purge System Original Manufacturers Operating Instructions and
Manuals in conjunction with this User Guide.
The Purge Systems are not manufactured by Yokogawa Laser Analysis Division.
Please also refer to separate detailed manufacturer’s instructions and start-up information for any Zone 1
or Division 1 automatic purge controller unit operational details. Failure to follow the manufacturer’s
guideline can result in damages and/or non-functionality of the purging system!
IM 11Y01B01-01E-A 6th Edition :Feb 13, 2013-00
<5. INSTALLATION AND WIRING>5-195-20
TDLS200 TDL Analyzer Instruction Manual V2.1
3.9.2 Purging Analyzer for Hazardous Areas (without On-Line Validation)
•Z-Purged designed in accordance with NEC/CSA Class 1, Division 2, Groups A-D
•Certified ATEX Zone 2 CAT 3 – Dual regulators must be used on the inlet!
The block diagram below shows the sections of the analyzer that require nitrogen purging. A Z-Type purge control system is fitted the Launch Unit and it includes a local indicator (Bright Green, rugged light) and pressure switch alarm contacts (open on loss of purge pressure).
TDLS200 TDL Analyzer Instruction Manual V2.1
3.9 Hazardous Area Systems
The TDLS200 Analyzer requires a continuous nitrogen gas purge to prevent ambient oxygen ingress to the optical path, when oxygen is the measured gas. The flow rate can be minimized as long as it prevents any ambient oxygen ingress to the measurement optical path. Other purge gases may be used as long as they do not contain any of the measured gas and are clean, dry, etc.
For hazardous area operation, the same nitrogen purge gas is used to purge the entire analyzer (including non-optical path sections such as the electronics).
The process interface may also require purging to maintain clear windows, refer to Process Window Purge details separately.
NOTE: Please also refer to any separate Purge System Original Manufacturers Operating
Instructions and Manuals in conjunction with this Instruction Manual.
The Purge Systems are not manufactured byYokogawa Laser Analysis Division.
3.9.1 Purging Analyzer for Hazardous Areas (with On-Line Validation)
•Z-Purged designed in accordance with NEC/CSA Class 1, Division 2, Groups A-D Certified ATEX Zone 2 CAT 3 – Dual regulators must be used on the inlet!
The block diagram below shows the sections of the analyzer that require nitrogen purging. A Z-Type purge control system is fitted the Launch Unit and it includes a local indicator (Bright Green, rugged light) and pressure switch alarm contacts (open on loss of purge pressure). The purging should be carried in sequence typically as shown below. All purge gas connections are ¼” od Tube fittings.
5.11.1 Purging Analyzer for Hazardous Areas (with On-Line Validation)
• NEC/CSA Class 1, Division 2, Groups A-D
• ATEX Zone 2 CAT 3 Dual regulators must be used on the inlet!
The block diagram below shows the sections of the analyzer that require nitrogen purging. A Z-Type purge control
system is tted the Launch Unit and it includes a local indicator (Bright Green, rugged light) and pressure switch
alarm contacts (open on loss of purge pressure). The purging should be carried in sequence typically as shown
below. All purge gas connections are ¼” od Tube ttings.
Nitrogen or I/A
Purge Gas
Nitrogen orI/A
Nitrogen or I/A Purge
Purge Gas
Gas
Main
Main
Electric
Electronic
Housing &
Housing
Purge
& Purge
System
System
Laser
Module
Check Gas
Flow Cell
Process
Interface
Interface
Process
Detect
Module
Figure 47 - Purge Flow Diagram when using on line validation
5.11.2 Purging Analyzer for Hazardous Areas (without On-Line Validation)
• NEC/CSA Class 1, Division 2, Groups A-D
• ATEX Zone 2 CAT 3 – Dual regulators must be used on the inlet!
The block diagram below shows the sections of the analyzer that require nitrogen purging. A Z-Type purge
control system is tted the Launch Unit and it includes a local indicator (Bright Green, rugged light) and pressure switch alarm contacts (open on loss of purge pressure).
Nitrogen or I/A Purge Gas
Main
Main
Electronic
Electric
Housing &
Housing
Purge
& Purge
System
System
Laser
Module
Check Gas
Check Gas
Flow Cell
Flow Cell
Process
Interface
Process
Detect
Interface
Module
Detect
Electronic
Detect
Electronic
Housing
Housing
Detect
Detect
Electronic
Electronic
Housing
Housing
Figure 48 - Purge Flow Diagram when not using on line validation
IM 11Y01B01-01E-A 6th Edition :Feb 13, 2013-00
<5. INSTALLATION AND WIRING> 5-21
TDLS200 TDL Analyzer Instruction Manual V2.1
3.9.2 Purging Analyzer for Hazardous Areas (without On-Line Validation)
•Z-Purged designed in accordance with NEC/CSA Class 1, Division 2, Groups A-D
•Certified ATEX Zone 2 CAT 3 – Dual regulators must be used on the inlet!
The block diagram below shows the sections of the analyzer that require nitrogen purging. A Z-Type purge control system is fitted the Launch Unit and it includes a local indicator (Bright Green, rugged light) and pressure switch alarm contacts (open on loss of purge pressure).
Figure 28 - Purge Flow Diagram when not using on line validation
3.9.3 Purging Analyzer and Universal Power Supplyand/orURD for HazardousAreas(withOn-LineValidation)
•Z-Purged designed in accordance with NEC/CSA Class 1, Division 2, Groups A-D
•Certified ATEX Zone 2 CAT 3 – Dual regulators must be used on the inlet!
The block diagram below shows the sections of the analyzer that require nitrogen purging when using in conjunction with either or the Universal Power Supply and Universal Remote Display. A Z-Type purge control system is fitted the Launch Unit and it includes a local indicator (Bright Green, rugged light) and pressure switch alarm contacts (open on loss of purge pressure).
Main
Electronic
Housing &
Purge
System
Laser
Module
Check Gas
Flow Cell
Process
Interface
Process
Interface
Detect
Module
Detect
Electronic
Housing
Nitrogen orI/A Purge Gas
TDLS200 TDL Analyzer Instruction Manual V2.1
3.9.4 Purging Analyzer and Universal PowerSupplyand/orURD(notusingOn-Line
Validation)
•Z-Purged designed in accordance with NEC/CSA Class 1, Division 2, Groups A-D
•Certified ATEX Zone 2 CAT 3 – Dual regulators must be used on the inlet!
The block diagram below shows the sections of the analyzer that require nitrogen purging. A Z-Type purge control system is fitted the Launch Unit and it includes a local indicator (Bright Green, rugged light) and pressure switch alarm contacts (open on loss of purge pressure). The purging should be carried in sequence typically as shown below. All purge gas connections are ¼” od Tube fittings.
5.11.3 Purging Analyzer and Universal Power Supply and/or URD for Hazardous
Areas (with On-Line Validation)
• NEC/CSA Class 1, Division 2, Groups A-D
• ATEX Zone 2 CAT 3 – Dual regulators must be used on the inlet!
The block diagram below shows the sections of the analyzer that require nitrogen purging when using in
conjunction with either or the Universal Power Supply and Universal Remote Display. AZ-Type purge control
system is tted the Launch Unit and it includes a local indicator (Bright Green, rugged light) and pressure switch
alarm contacts (open on loss of purge pressure).
Nitrogen or I/A Purge
Gas
Figure 49 - Purge Flow Diagram when using on line validation
5.11.4 Purging Analyzer and Universal Power Supply and/or URD (not using On-Line Validation)
Main
Electronic
Housing &
Purge
System
Laser
Module
Nitrogen or I/A Purge
Nitrogen orI/A Purge
Gas
Gas
Check Gas
Check Gas
Flow Cell
Flow Cell
Power
Supply
or URD
Interface
Process
Detect
Process
Interface
Module
Detect
Detect
Electronic
Electronic
Housing
Housing
• NEC/CSA Class 1, Division 2, Groups A-D
• ATEX Zone 2 CAT 3 – Dual regulators must be used on the inlet!
The block diagram below shows the sections of the analyzer that require nitrogen purging. A Z-Type purge
control system is tted the Launch Unit and it includes a local indicator (Bright Green, rugged light) and
pressure switch alarm contacts (open on loss of purge pressure). The purging should be carried in sequence
typically as shown below. All purge gas connections are ¼” od Tube ttings.
Nitrogen or I/A Purge Gas
Main
Main
Electronic
Electronic
Housing &
Housing
Purge
& Pureg
System
System
Laser
Module
Check Gas
Check Gas
Flow Cell
Flow Cell
Process
Interface
Interface
Process
Detect
Module
Detect
Detect
Electronic
Electronic
Housing
Housing
Power
Power
Supply
Supply
or URD
or URD
Figure 50 - Purging Analyzer for Hazardous Areas -NO- On-Line Validation
IM 11Y01B01-01E-A 6th Edition :Feb 13, 2013-00
<5. INSTALLATION AND WIRING>5-215-22
5.12 Cyclops Division 2/Zone 2 Purge Indictor, with switch
Type Z purging reduces the classication within a protected electronics enclosures from Division 2 or Zone 2
to nonhazardous. Failure to maintain pressure within the protected enclosure shall be detected by an alarm
or indicator at the electronics enclosure. The dilution purge time shall be a manual operation and once the
electronics enclosure has been purged of ignitable or ammable concentrations, only positive pressure of at least
0.20 inches H2O (0.50 mbar) is required to be maintained within the electronics enclosure and it is not necessary
to remove power from the protected equipment upon the loss of purge pressure. The CYCLOPS
Z – Purge Indicator is used to provide safe monitoring of electrical equipment in Division 2 and Zone 2 hazardous
areas, which can be used to prevent the possibility of re or explosion inside the enclosure of energized electrical
equipment, a protective gas supply is used to dilute potentially ammable materials to an acceptable level,
creating a safe area for the electrical equipment within the enclosure. Positive pressure prevents the ingress
of ammable materials in the surrounding atmosphere from entering into the enclosure, as long as positive
pressure is maintained. After the enclosure is purged, power may be manually applied to the protected electrical
equipment.
The CYCLOPS Z – Purge Indicator provides an objective evidence of the presence of adequate positive purge
pressure within the electrical equipment enclosure. A normally open differential pressure switch continuously
compares the pressure inside the monitored electronics enclosure with respect to the atmospheric pressure
surrounding the electronics enclosure. When the monitored electronics enclosure register’s a pressure of at
least 0.20 inches H2O (0.50 mbar) above the reference atmospheric pressure a manually controlled dilution
time cycle may then begin. Typically, a minimum dilution time cycle is specied to ensure that at least ve
times the volume of free space in the electronics enclosure is exchanged before power is manually applied to
the electrical equipment. (The number of exchanged volumes may be higher in some situations). After the
manual dilution time cycle has elapsed and the monitored electronics enclosure pressure is being maintained
above 0.20 inches H2O (0.50 mbar), power may be manually applied to the electrical equipment within the
purged electronics enclosure. The CYCLOPS Z – Purge Indicator is designed to indicate the presence of purge
pressure from one pressure reference point. Several electronics enclosures can be installed in series with purge
gas being introduced into the rst electronics enclosure and the CYCLOPS Z – Purge Indicator monitoring the
last electronics enclosure in the series; multiple electronics enclosures can now be monitored using only one
CYCLOPS Z – Purge Indicator. The pressure inside the monitored electronics enclosures must maintain at least
0.20 inches H2O (0.50 mbar) higher than the atmospheric pressure surrounding the electronics enclosure. This
ensures that hazardous materials are not going to ingress into the pressurized and now protected electronics
enclosures. If any of the electronics enclosures installed in the series door is opened, pressure will show to be
below the required 0.20 inches H2O (0.50 mbar) in all electronics enclosures. The exhaust vent which comes as
part of the CYCLOPS Z – Purge Indicator casing, can exhaust purge gas from enclosures with volumes up to 15
cubic feet (425 liters).
Cyclops Features
Certied for installation and use in ATEX and IECEx for
Type Z – Purge, II 3 G Ex nA nL [pz] IIC T6
For Zone 2 gas hazardous areas
Certied for installation and use in ATEX and IECEx for
Type Z – Purge, II 3 G Ex nA nL [pz] IIC T6
For Zone 2 gas hazardous areas
IM 11Y01B01-01E-A 6th Edition :Feb 13, 2013-00
<5. INSTALLATION AND WIRING> 5-23
Normal Operating Conditions
Power
Certied for installation and use in ATEX and IECEx for
Type Z – Purge, II 3 G Ex nA nL [pz] IIC T6
For Zone 2 gas hazardous areas
Manual Dilution Cycle Time To
Energizing Electrical Equipment
CYCLOPS Z – Purge Indicator,
Minimum Pressure
WARNING: The number of exchanged volumes may be higher in some situations. Refer to
TDLS-200 ATEX Purge Warning Labels for Details.
NOTE: Instrument Air and Nitrogen purge gases have different purge time requirement. It is
important to use clean, dry purge gases to ensure the pressure switch contacts do not foul and
cause subsequent operating issues (i.e. non-functionality of the Cyclops).
Typically, dilution cycle time is to ensure that at least ve (5)
times the volume of free space in the enclosure of protective
gas supply is exchanged before power is applied to the
electrical equipment. Ten (10) times volumes for motors,
generators and other rotating electrical machinery.
Green indicator light remains on to show purge pressure being
maintained above 0.20 inches H2O (0.50 mbar) in electronics
enclosure being monitored.
Utility Requirements
Purge Protective Gas Supply
Pressure to Pressure Regulator
Purge Protective Gas Supply
Quality
Power Input / Consumption
Voltage
20 psig (1.4 Bar) minimum (Suggested to compensate for
enclosure leak rate)
Water and oil-free, - 40°F (- 40°C) dew point, particles ≤ 5μ, ISA
grade hydrocarbon free
0.5 Watts maximum
4VDC model (19VDC to 28VDC) 47 to 63 Hz
CYCLOPS Z – Purge Indica-
tor
Mains Supply Fluctuation
Not to Exceed 10%
Environmental Conditions
Operating Temperature Range- 40°F to 150°F (- 40°C to 65°C)
Remote Initiate: control channel
Time Initate: frequency
Settings: valve, purge, tiime, AO
mode gastype, conc, opl, temp,
pres
Factory Calibration
Remote Initiate: control channel
Time Initiate: frequency
Settings: conc, opl, temp, pres
Factory Calibration
Pres, temp, opi, gas, type, conc
Remote Initate: control channel
Time Initiate: frequency
Settings: conc, opl, temp, pres,
gastype, valve,purgetimes, AO
mode
6-3
Data
Online ValidationCheck Gas 1
Clear Validation Alarms
Alarm History
Cal History
Spectrum CaptureManual
Check Gas 2*
*(similar to Gas 1)
AutomaticUpdated
ManualLocal Initate
Automatic
Relative
Absolute
Warning
Fault
Remote Initate: control channel
Time Initiate: frequency
Settings: conc, opl, temp, pres,
gastype, valve,purgetimes, AO
mode
IM 11Y01B01-01E-A 6th Edition :Feb 13, 2013-00
<6. BASIC OPERATION>
Online MenuLevel 1 Menu Level 2 MenuLevel 3 MenuLevel 4 MenuLevel 5 Menu
TrendsRefreshRefresh Current Trend screen
Gas 1 Concentration
STDEV of Gas 1 Concentra-
tion*
Gas 2 Concentration*
STEV of Gas 2 Concentration*
Transmission*
Laser Temp Setpoint*
Laser Temp in DegC*
Peak Center Position*
Gas Temperature*
Gas Pressure*
*(Similar to Gas 1 Concentration)
Display TextDescription
Line 1 - MeasurementO2 xx.x %Measured gas and unit of measurement For Gas 1
System OKNormal Operation condition with no active alarms
WARNING Det Sig Low
WARNING Trans Low
WARNING Spectr Noise
WARNING Gas Pres
WARNING Gas Temp
WARNING Gas Level
WARNING Board Temp
Line 3 - Status
FAULT Laser Temp
FAULT Det Sig High
FAULT Det Sig Lost
FAULT Peak Response
FAULT Peak Center
Zero Calibrating…
Span Calibrating…
Ofine Validating…
Online Validating…
Data Transferring…
Transfer Success
Transfer Failure
Yokogawa TDLS
SN 76-1xxx-05-xx
AO1: CONC xx-xx%/ppm/ppb/mg/m3
or mg/Nm3
AO2: TRANS xx-xx%
AO3: TEMP xx-xxF/C/K
Line 4 - Information
10.0.0.35
TEMP Act/Con/Fox xx F/C/K
PRES Act/Con/Fix xx.x PsiA/BarA/KPa/
torr/atm
OPL xx.x in/ft/cm/m
Launch xx deg C
Detect xx deg C
Min
Max
Minutes
and/or Gas 2 as congured
(0-100% range)
initialization of the analyzer
WARNING Conditions
FAULT Conditions
Validation Status
Data Transfer Status
Analyzer Name
Analyzer Serial No.
Congured 4-20mA output for AO1, AO2, & AO3
Static IP Address
Process Gas Temperature used for gas
concentration calculation
Process Gas Pressure used for gas
concentration calculation
Optical Path Length over which the analyzer is
measuring the target gas
Launch unit internal temperature
Detect unit internal temperature
IM 11Y01B01-01E-A 6th Edition :Feb 13, 2013-00
6.2 Software Guide
NOTE: At any time in the Main Menu, press the F5 key (only on Screen & Keypad versions)
to toggle the LCD backlight on/off. If there is no keypad (or VNC) activity for 30 minutes
then the LCD backlight will automatically switch off, press any key to restore backlight.
<6. BASIC OPERATION>
MAIN MENU (Home Page)
Display of Concentration & Transmission (or 2nd Concentration)
Status Window – notication of initiating, working properly,
warnings or faults
Gas Temperature
Gas Pressure
Selection of Basic or Advanced Menu
Tag No and Serial No. (SN.) congured to analyzer
Active Alarm Display Button
Analyzer Shut Down Button
After selection of either Basic or Advanced Menu you will see
the Output Selection screen. PLEASE note the RIGHT ARROW
key access to the MENU (older version of software require no
other key or Press 9 to access MENU)
This allows control of the analog output while the user is
working in the analyzer software. Enter mA value form 0-20.
• Block will hold outputs at user selectable mA (example
shows 3.8mA) until return to Main Screen but
NOTE CHANNEL 2 ALWAYS TRACKS!
• Track will allow outputs to continue to report
concentration and transmission until return to Main
Screen
• Hold will hold outputs at their current value until return to
Main Screen NOTE CHANNEL 2 ALWAYS TRACKS!
6-5
BASIC MENU
Congure – allows setting of Path Length, Gas
Temperature, Gas Pressure
View Spectra – user will select display of raw detector
signal or absorption spectra
Data – Alarm History, Calibration History
Trends – Allows for the displaying of data in a trend format
IM 11Y01B01-01E-A 6th Edition :Feb 13, 2013-00
<6. BASIC OPERATION>
BASIC CONFIGURE
PATH LENGTH – allows adjustment of the optical path (distance the laser is exposed to the process gas).
PRESSURE – allows adjustment of the gas pressure value if
using xed pressure. If the analyzer is using active pressure
compensation (live signal fed from pressure transducer) no
changes are allowed. Active pressure compensation settings
are found in Advanced Menu.
TEMPERATURE – allows adjustment of the gas pressure value
if using xed pressure. If the analyzer is using active pressure
compensation (live signal fed from pressure transducer) or
active ambient, no changes are allowed. Active pressure
compensation settings are found in Advanced Menu.
IP ADDRESS – displays the analyzer IP address
SERIAL NO. – displays analyzer serial number
VERSION – software version number
The spectra screen (raw detect, left or absorption, below)
allows capture and view of current spectra.
The screen auto scales the vertical (Y) axis; this will result in a
visually noisy spectra when at low gas levels.
In fact the spectra may not be noisy, but simply that the display
range is extremely low.
IM 11Y01B01-01E-A 6th Edition :Feb 13, 2013-00
<6. BASIC OPERATION>
The BASIC DATA MENU allows the user to select:
ALARM HISTORY – displays the last 17 alarms and faults
with brief description, date and time
CALIBRATION HISTORY - displays the last 17 calibration
events with adjustment amount, date and time
The TREND SCREEN is identical for BASIC or ADVANCED
MENUS. It allows the user to trend up to the last 750
minutes (of current day) of data for:
REFRESH - The trend will NOT update automatically, use
the refresh button to update the trend
CONC. – analyzer reading of gas concentration (shown here
as O2, or whatever Gas 1 is congured)
STDEV1 – the standard deviation of 25 consecutive
concentration readings (for gas concentration 1)
Gas2. – analyzer reading of gas 2 concentration
STDEV2 – the standard deviation of 25 consecutive
concentration readings (for gas concentration 2)
TRANS. – transmission % of laser light through the process
gas
LTS – analyzer laser temperature set point
LT – analyzer laser temperature
PCP – peak center position for the absorption peak
TEMP – process gas temperature
PRES – process gas pressure
Alongside the selection buttons the current value is
displayed.
When selecting the information to trend user will be
prompted to enter minimum value, maximum value and time
to trend.
6-7
IM 11Y01B01-01E-A 6th Edition :Feb 13, 2013-00
<6. BASIC OPERATION>
ADVANCED CONFIGURE MENU
PROCESS PATH LENGTH – allows user to enter in a new optical
path length (distance laser is exposed to process gas)
PRESSURE – allows selection of ACTIVE (analyzer fed pressure
value from external transducer) or FIXED (value entered into
software) process gas pressure. In Active mode, a BackUp value can be entered, in case of active input failure.
CONTROLLED is not applicable for TDLS-200.
TEMPERATURE – allows selection of ACTIVE (analyzer fed
temperature value from external transducer), ACTIVE AMBIENT
(ambient gas temperature derived from internal sensor) or FIXED
(value entered into software) process gas temperature. In Active
mode, a Back-Up value can be entered, in case of active input
failure. ACTIVE PEAKS is used for special Oxygen applications
only, please consult with Yokogawa directly. CONTROLLED is
not applicable for TDLS-200.
NON-PROCESS PARAMETERS – allows mathematical
subtraction of purge gases that contain the target gas. Example
would be Instrument air purge when measuring Oxygen or CO
line locking gas for combustion CO applications.
UNITS – selection as dened below, independently
SYSTEM I/O – allows set up and assigning of analyzer Analog
and Digital I/O
SYSTEM – displays analyzer information (serial number, Fat date,
password, software version, launch/detect unit temperatures,
etc.), allows setting of date/time, TCP/IP.
VALVE CONTROL ¬– allows for manual and/or automatic control
of the valve driver output signals
SIGNAL PROCESSING – Factory set parameters only
LASER SPECTRA & CONTROL – displays spectra and allows
manual control of laser
PROCESS PATH LENGTH
Enter the distance over which the laser will be exposed to
the process gas, this excludes any purge paths. Consider
just the distance of process gas exposure to the laser beam
path. Consult Yokogawa is any assistance required. For bypass applications with window purges, typically the center
line distance from inlet-outlet pipes is used. For combustion
applications, the distance inside the refractory for example.
NOTE: for extractive applications, this will match the ow cell
condition, typically 40”.
NOTE: when using an off-line calibration cell the standard optical
path will be 28.6” (72.6cm)
IM 11Y01B01-01E-A 6th Edition :Feb 13, 2013-00
<6. BASIC OPERATION>
PROCESS PRESSURE
FIXED is used when the process gas pressure will not vary
under normal operating conditions when the measurement is
required. If there is any process pressure variation, then the
results are typically affected proportionally according to gas
law. Fixed is most suited to ~atmospheric conditions, such as
vent lines and combustion.
ACTIVE is used when the highest degree of accuracy is
required under variable process pressure conditions or when
specied for the given application. The range must match the
4-20mA input range and a back-up value may be entered in
case the input signal fails
CONTROL is not used in TDLS-200
PROCESS TEMPERATURE
FIXED is used when the process gas temperature will not vary
under normal operating conditions when the measurement is
required. If there is any process temperature variation, then
the results are typically affected according to the specic
application.
ACTIVE INPUT is used when the highest degree of accuracy
is required under variable process temperature conditions or
when specied for the given application. The range must match
the 4-20mA input range and a back-up value may be entered in
case the input signal fails
ACTIVE AMBIENT is used when the process gas generally
follows ambient temperature. It is not as accurate as an active
input but it is more accurate than a Fixed value
ACTIVE PEAKS is used for special high temperature oxygen
combustion application when the gas is above 800˚C – used
only when approved by Yokogawa and not functional <800˚C
CONTROL is not used in TDLS-200
6-9
NON-PROCESS PARAMETERS
This software feature allows the user to enter all necessary
parameters associated with the Non-Process Parameters
conguration.
Refer to detail section of User Guide for more information
and follow the on-screen directions for programming
details
IM 11Y01B01-01E-A 6th Edition :Feb 13, 2013-00
<6. BASIC OPERATION>
UNITS
Path Length, select the appropriate units of measure for Path
Length: in/ft/cm/m
Pressure, select the appropriate units of measure for Pressure:
psiA, barA, kPa, torr, atm
Temperature, select the appropriate units of measure for
Temperature: F, C, K
SYSTEM I/O - ANALOG OUTPUT
CHANNEL 1 to 3 – conguring each 4 to 20mA channel to
output Concentration, Transmission, Gas Temperature, Gas
Pressure or None. NOTE, Channel 2 DOES NOT follow
the Block/Track/Hold modes and is typically assigned to
Transmission
WARNING MODE – setting of mA output response during
analyzer warnings (Block, Track, Hold)
FAULT MODE – setting of mA output response during analyzer
warnings (Block, Track, Hold)
FIELD LOOP CHECK – allows specied 4-20mA output levels
to check and distinguish between the three analog output
connections; select analog output channel 1, 2, or 3 to check
and input new value to output
AO CH CALIBRATION – Pre-Calibrated at factory and not
normally required. Allows calibration of 4 to 20mA output
channels; follow onscreen instructions.
SYSTEM I/O - ANALOG INPUT, Field Checking
Review the displayed mA values to eld check the incoming
analog signals
IM 11Y01B01-01E-A 6th Edition :Feb 13, 2013-00
<6. BASIC OPERATION>
SYSTEM I/O - DIGITAL OUTPUT
Setting of Digital Output assignments (DO 1-3)
CHANNEL 1 WARNINGS – Setting of levels that will trigger
analyzer warning and subsequent DO
CHANNEL 2 FAULTS – Setting of levels that will trigger analyzer
fault and subsequent DO
CHANNEL 3 USER ALARM – Setting of either Concentration
or Transmission level (high or low), calibration/validation status
that will trigger analyzer user alarm DO/status
FIELD LOOP CHECK – allows for convenient eld loop
checking of the digital alarm output contacts, follow the on-
screen directions
CHANNEL 1 - WARNINGS
Menu allows setting of various analyzer WARNINGS conditions.
WARNING is an event that will indicate that maintenance is
require and the analyzer is still operational.
– PLEASE ADJUST WITH ONLY FACTORY ASSISTANCE –
WARNINGS CONDITIONS ARE IMPORTANT SETTINGS
6-11
DETECTOR SIGNAL LOW – lower raw detector signal limit
TRANSMISSION LOW – lower limit of transmission. Note, when
transmission falls below this value, the automatic peak tuning
function is disabled. This is to prevent excessive noise (caused
by low transmission) interfering with correct peak tuning.
SPECTRUM NOISE HIGH – factory set, do not adjust
PROCESS PRESSURE OUT OF RANGE – typically set for the
min-max expected process pressure. Note, if operating outside
the known conditions, measurement accuracy may be affected
PROCESS TEMPERATURE OUT OF RANGE – typically set for
the min-max expected process temperature. Note, if operating
outside the known conditions, measurement accuracy may be
affected
CONCENTRATION OUT OF RANGE – set for any desired alarm
points
BOARD TEMPERATURE OUT OF RANGE – used to indicate
excessively hot-cold ambient conditions for the analyzer
DO-WARNING DELAY – to avoid single event alarms that have
proven to be short terms events, enter a number of readings to
delay before the alarm becomes active
IM 11Y01B01-01E-A 6th Edition :Feb 13, 2013-00
<6. BASIC OPERATION>
CHANNEL 2 FAULTS
Menu allows setting of various analyzer FAULT conditions.
FAULT is an event that will eliminate the measurement
integrity, it is an indication that maintenance is require and the
analyzer is not operational.
– ONLY ADJUST WITH FACTORY ASSISTANCE – FAULT
CONDITIONS ARE CRITICAL SETTINGS THAT CAN
RESULT IN DAMAGE TO THE ANALYZER IF IMPROPERLY
PROGRAMMED -
LASER TEMPERATURE OUT OF RANGE – upper and lower
fault conditions for laser temperature
DETECTOR SIGNAL HIGH – upper raw detector signal limit
DETECTOR SIGNAL LOST – lower raw detector signal limit
MEASUREMENT PEAK NO RESPONSE – Used when Line
Locking gas and/or Non-Process Parameters are enabled
to detect the loss of absorption peak, i.e. the measured
absorption peak falls below the set threshold value – consult
factory for further details if attempting to implement LineLocking gas and/or non-process parameters. To disable
this Fault (only under guidance from factory authorized
personnel) please enter a value of -1
PEAK CENTER OUT OF RANGE – loss of peak center control
VALIDATION FAILURE – Allows the user to rene the most
appropriate +/-% for validation PASS/FAIL criteria. Some
process and conditions of application will require a large
PASS/FAIL criteria (as much as +/-50%) due to the variables
and dynamic nature of the application. On-Line Validation
can be considered as Response Checking. Please consult
Yokogawa for further information and help in determining what
an appropriate Pass/Fail criteria is for the specic application.
DO FAULT DELAY - this feature allows the end user to
minimize potentially non-impacting nuisance alarms by adding
some delay before the analyzer reports the Fault Alarm
IM 11Y01B01-01E-A 6th Edition :Feb 13, 2013-00
<6. BASIC OPERATION>
CHANNEL 3 - USER ALARMS
Enables digital output alarm for concentration value (High or
Low), Transmission (High or Low) or Validation/Calibration (the
contact changes state during a validation or calibration)
DIGITAL INPUT – STATUS CHECK
The user can check the status of remote Digital Inputs
6-13
IM 11Y01B01-01E-A 6th Edition :Feb 13, 2013-00
<6. BASIC OPERATION>
SYSTEM
Some settings are not adjustable by user, user adjustment is
possible for:
PASSWORD – changes password for ADVANCED menu
access
DATE & TIME – changes analyzer date and time
TCP/IP – the analyzers real IP address used for Ethernet
communications can be changed via this menu option.
External keyboard with Windows key no longer required as in
older software versions.
ADJUSTABLE RESISTORS: After detector gain adjustment in
the eld, these values should be entered – these values are
then stored to the syste.cfg le.
SETTINGS USED ONLY FOR STORING FACTORY DATA:
Analyzer S/N, Laser S/N, Software Version,
VALVE CONTROL
Typically used when the analyzer is congured with a ow
cell in an ofine application. These values can also be
used manually to introduce on-line validation gases. Valve
conguration conicts will be noted on-screen if applicable
VALVE 1 – Manual (On-Off Toggle), Time Sequence (next valve
and time seconds to next valve) or Remote Control (select D/I
channel for control) capability, follow the on-screen directions
to program Valve 1
VALVE 2 – Manual (On-Off Toggle), Time Sequence (next valve
and time seconds to next valve) or Remote Control (select D/I
channel for control) capability, follow the on-screen directions
to program Valve 2
VALVE 3 – Manual (On-Off Toggle), Time Sequence (next valve
and time seconds to next valve) or Remote Control (select D/I
channel for control) capability, follow the on-screen directions
to program Valve 3
6-14
IM 11Y01B01-01E-A 6th Edition :Feb 13, 2013-00
<6. BASIC OPERATION>
LASER SPECTRA & CONTROL
Displays Raw Detector Signal and Absorption Spectrum as
well as Gas Concentration, Gas Temperature, Gas Pressure,
Transmission, Laser Temperature, Peak Center Position and
other parameters
CAPTURE – allows a manual spectra capture (user will be
prompted to enter a unique le name for captured spectra)
CONTROL MODE – allows selection of Automatic (laser
temperature is controlled to keep peak centered using peak
center position as set point) or Manual (laser temperature is
controlled using integral laser temperature sensor). Important
auto-tune peak-height threshold values should not be adjusted
without factory guidance – ONLY ADJUST WITH FACTORY
ASSISTANCE
Manual adjustment of the laser temperature set-point (LTS)
may be done with care and factory guidance – used only
to re-adjust the absorption peak position if the peak has
moved outside of the normal operating range – advanced
troubleshooting.
LASER TEMP SETPOINT (LTS) – In manual mode allows
adjustment of laser temperature – ONLY ADJUST WITH
FACTORY ASSISTANCE. NOTE, the analyzer will return to
automatic mode if un-touched for 30 minute!
LTSP LIMITS – setting of guard limits for laser temperature set
point – ONLY ADJUST WITH FACTORY ASSISTANCE
MAX CURRENT – setting of center point for laser current ramp
–ONLY ADJUST WITH FACTORY ASSISTANCE
FAST UP-DATE – this function can be Enabled (faster
update times) or Disabled (normal operation) as necessary.
When enabled, the analyzer up-date is approximate twice
as fast as normal to allow for faster/easier alignment during
commissioning or troubleshooting. Normal up-date time is
restored automatically after 30 minutes when the screen saver/
backlight feature also enables
ADVANCED CALIBRATE & VALIDATE MENU
OFFLINE CALIBRATIONS – allows zero calibration, zero offset,
span calibration, transmission adjustment,
OFFLINE VALIDATIONS – allows manual or automatic
conguration of check gases 1 or 2 or a separate two gas
check
ONLINE VALIDATIONS – allows manual or automatic
conguration
CLEAR VALIDATION ALARMS – allows the user to clear
the validation alarm if the user knows that the analyzer is
functioning and tuned to correct absorption peak.
IM 11Y01B01-01E-A 6th Edition :Feb 13, 2013-00
<6. BASIC OPERATION>
OFFLINE CALIBRATIONS
ZERO CALIBRATION – manual or automatic calibration of
Zero - ensure there is no absorption peak feature before
performing a zero calibration, failure to do so can result is false
low readings later when the un-desired target gas has been
removed.
ZERO OFFSET – allows manual adjustment of Zero by applying
a concentration offset – use only with factory guidance
SPAN CALIBRATION – manual or automatic calibration of
Span – use only when sure the purge paths are correctly
congured
TRANSMISSION – adjustment of transmission value
DARK CURRENT – DISABLED feature, Factory use only
PEAK SEARCH – DISABLED feature that initiates a system
scan of absorption peaks to validate current peak selection is
correct, Factory use only
OFFLINE VALIDATIONS
CHECK GAS 1 to 2 – allows manual or automatic conguration
up to 2 check gasses.
6-16
Two Gas Validation – allows the user to congure two
validation gas checks sequentially.
An example would be Oxygen analyzers used for Marine
Vapor Recovery applications whereby a single “validation”
command shall execute both a zero and span gas check.
Results for the two gas off-line validation are shown for both
gases along with their corresponding results
IM 11Y01B01-01E-A 6th Edition :Feb 13, 2013-00
<6. BASIC OPERATION>
ONLINE VALIDATIONS
CHECK GAS 1 – allows user to select and congure the on-line
validations for check gas #1. The conguration options include
Manual or Automatic.
Automatic has selections for Local Initiate, Remote Initiate, Time
Initiate as well as Settings for these options
CHECK GAS 2 – allows user to select and congure the on-line
validations for check gas #2. The conguration options include
Manual or Automatic.
Automatic has selections for Local Initiate, Remote Initiate, Time
Initiate as well as Settings for these options
ADVANCED DATA MENU
ALARM HISTORY – shows chronological list of analyzer’s most
recent alarms
CAL HISTORY – shows chronological list of analyzer’s recent
calibrations
SPECTRUM CAPTURE – selection of; AUTOMATIC (user will
be prompted to select capture interval, number of UPDATES
to trigger capture, RELATIVE concentration level trigger which
is a % of reading change, or ABSOLUTE concentration level to
trigger capture); in addition the software will prompt for number
of spectra to capture when a Warning or Fault occurs. MANAUL
selection will result in spectra capture only when requested by
user.
CAUTION! By setting a high spectrum capture rate (i.e. a low
number between up-dates), the memory can very quickly ll
up and because the system uses a First-In First-Out basis, you
may lose many data les for results (.res) and spectrum (.spe).
As a guide, spectrum is captured every 200-300 up-dates and
1-5 every Warning and 1-5 every Fault. The spectrum capture
rate can be increased by lowering the up-dates value on a short
term basis for a particular monitoring period but please ensure it
is returned to normal rates.
An example could be the monitoring of a dirty process whereby
transmission becomes very low during certain known process
operating conditions. The spectrum capture rate could be
changed to 2 for a 1 hour per period when these conditions
exist and then set back to 200-300 after. This then gives a lot of
spectrum data to review for the specic operating conditions.
IM 11Y01B01-01E-A 6th Edition :Feb 13, 2013-00
<6. BASIC OPERATION>
6.3 Non-Process Parameters
Non-Process Parameters is the Yokogawa Laser Analysis Division term used to dene regions of the optical path
that may be purged with a gas containing the actual target (measured) gas. The most common application of this
is to use Instrument Air (~20.9% O2) as the purge gas for analyzers measuring Oxygen in the process.
Another common use of this feature is “Line Locking” whereby some target gas (typically %CO for CO/CH4
combustion analyzers) is locked into the validation cell at all times and therefore its absorption contribution must
be accounted for.
The valve functionality of Line Locking has four stages as shown below.
6-18
Figure 51
IM 11Y01B01-01E-A 6th Edition :Feb 13, 2013-00
<6. BASIC OPERATION>
If measuring CO in the process gas and the purge gas is Instrument Air, then these parameters are not applicable
because the CO concentrations typically found in Instrument Air are below practical detection limits.
In order for the analyzer to measure correctly under these purge conditions, the analyzer must know the correct
parameters such that the measured output value has been compensated i.e. the oxygen in the purge gas has
been taken into account when determining the process oxygen concentration.
Oxygen present in the Instrument Air purge may exist in several section of the optical path including:
• Inside the Launch unit body, laser module section
• Inside the Launch unit body, online validation cell
• Inside the Launch unit alignment ange
• Inside the launch side process nozzle (and valve)
• Inside the Detect unit body
• Inside the Detect unit alignment ange
• Inside the Detect side process nozzle (and valve)
It is important that these dimensions are known as they will need to be programmed in – if in doubt, please
contact Yokogawa Laser Analysis Division.Calibration – The analyzer MUST be calibrated (Zero and Span) as
per the normal methods outlined in the standard User’s Guide. When performing a Zero Calibration,
ensure that the entire optical path is purged with Nitrogen. When performing a Span Calibration, ensure the
correct procedures are followed!
IM 11Y01B01-01E-A 6th Edition :Feb 13, 2013-00
<6. BASIC OPERATION>
ADVANCED CONFIGURE MENU (UPDATED)
The Advanced Congure Menu has been updated with a sub-
section titled Non-Process Parameters.
NON-PROCESS PARAMETERS
These non-process parameters are for the measured gas in the
optical path but outside of the process path length.
These parameters MUST be entered for an accurate
measurement if the purge is not nitrogen (when measuring
Oxygen) or when a line locking gas is being used (such as %
CO for combustion CO/CH4 applications)
6-20
NON-PROCESS PATH LENGTH
This is the optical path length between the laser and detector
excluding the process path length.
The factory values entered include the analyzer internals and
the alignment anges. The distance of process isolation valves
and process nozzles (on both Launch and Detect sides) MUST
be added to the existing factory values.
NON-PROCESS PRESSURE
This is the pressure of the non-process gas. Typically, this
will be close to atmospheric pressure of 1.01 BarA or 14.7
PsiA. Check the actual operating conditions and enter the
appropriate value. Contact Yokogawa if unsure.
IM 11Y01B01-01E-A 6th Edition :Feb 13, 2013-00
<6. BASIC OPERATION>
NON-PROCESS TEMPERATURE
This is the temperature of the non-process gas with two modes
of input:
FIXED – manual input of xed temperature value
ACTIVE AMBIENT - ambient gas temperature derived from
integral sensor on detector circuit with offset adjustment
(typically -5 deg C) and an adjustable coefcient value (1.0
shown left).
To derive the optimum coefcient value, please contact
Yokogawa with installation and application details. This will
ensure the optimal coefcient value taking into account the
temperature gradient from ambient to process gas temperature
for the non-process purge gases.
NON-PROCESS ALARM SELECTION
This allows the selection or disabling of alarm when then non-
process gas peak height falls below the entered threshold
value.
In this example, the analyzer is congured to FAULT Alarm
when the peak height is lower than 0.001 au (see absorption
spectrum). This value can be determined with assistance from
the Yokogawa factory to suit the specic application.
This alarm function can be used to help detect the leakage or
loss of line locking gas, for example %CO on combustion CO/
CH4 applications
To disable this alarm, simply enter -1 as the threshold value
with the toggle in either Fault or Warning position.
NON-PROCESS ALARM SELECTION
This allows the selection or disabling of alarm when then non-
process gas peak height falls below the entered threshold
value.
In this example, the analyzer is congured to FAULT Alarm
when the peak height is lower than 0.001 au (see absorption
spectrum). This value can be determined with assistance from
the Yokogawa factory to suit the specic application.
This alarm function can be used to help detect the leakage or
loss of line locking gas, for example %CO on combustion CO/
CH4 applications
To disable this alarm, simply enter -1 as the threshold value
with the toggle in either Fault or Warning position.
IM 11Y01B01-01E-A 6th Edition :Feb 13, 2013-00
6.4 Reference Peak Lock with 2nd Absorption Gas
TDLS200 TDL Analyzer Instruction Manual V2.1
4.5 Reference Peak Lock with 2nd Absorption gas
In some measurement circumstances, instead of using the measurement gas (i.e. NH3) absorption peak to lock the laser wavelength tuning range, the analyzer uses an adjacent (i.e. H2O) absorption peak near the measured absorption peak as the reference. This is used when the measured gas concentration is normally zero or very low such that its absorption peak is not large enough for the wavelength locking function. The adjacent absorption peak level, however, is typically always high and therefore provides a large enough absorption peak for the wavelength locking function.
For the following explanation of “Reference Peak Lock” we will use NH3 as the “Measured Gas” and H2O as the “Reference Gas”. The TDLS is measuring NH3 but is using an adjacent H2O absorption peak to maintain laser wavelength control.
There are two wavelength regions that need to be mathematically manipulated for an absorption spectrum: reference peak region and concentration prediction region. In the reference peak region, the adjacent (i.e. H2O) peak position is calculated for the laser wavelength lock. In the concentration prediction region, area integration is performed for measured gas (i.e. NH3) concentration prediction. For example, the following spectrum is captured from a test with an NH3analyzer. The reference peak region is at 150~180, and the measurement gas concentration prediction region is at 100~140.
For Off-Line Zero Calibration, make sure to purge Nitrogen (or dry Instrument Air) everywhere through the optical path so that there is no measured gas or reference peak gas (i.e. NH3 or H2O) absorption in the spectra. For span calibration it is best to leave the laser
temperature control as it is (there will be peak center out of range fault). Factory software access can further allow a change of the reference peak position set point to 124.5 (NH3 peak locking), oruse manual laser temperature control mode.
Below: Typical Process Gas Spectra with NH3 and H2O present
<6. BASIC OPERATION>
6-22
In some measurement circumstances, instead of using the measurement gas (i.e. NH
the laser wavelength tuning range, the analyzer uses an a H2O absorption peak near the NH3 absorption peak
as the reference. This is because in the typical application process, NH3 concentration is normally zero or very
low such that its absorption peak is not large enough for wavelength lock function. Moisture levels, however,
are typically always high (about 10%) and therefore provide a large enough absorption peak for the wavelength
locking function. Other applications using this same approach may include CO/CO2 or in some combustion
applications, measure CO and lock onto an adjacent H2O peak.
There are two wavelength regions that need to be mathematically manipulated for an absorption spectrum:
reference peak region and concentration prediction region. In the reference peak region, the H
calculated for the laser wavelength lock. In the concentration prediction region, area integration is performed for
NH3 concentration prediction. For example, the following spectrum is captured from a furnance test with a NH3
analyzer. The reference peak region is at 150~180, and the measurement gas concentration prediction region is
at 100~140.
) absorption peak to lock
3
O peak position is
2
For Off-Line Zero Calibration, make sure to purge Nitrogen (or dry Instrument Air) everywhere through the
optical path so that there is no NH3 or H2O absorption in the spectra. For span calibration it is best to leave
the laser temperature control as it is (there will be peak center out of range fault). Factory software access can
further allow a change of the reference peak position set point to 124.5 (NH3 peak locking in the example shown
below), or use manual laser temperature control mode.
NH
3
H2O
Figure 52 - Typical Process Gas Spectra with NH3 and H2O present
IM 11Y01B01-01E-A 6th Edition :Feb 13, 2013-00
<6. BASIC OPERATION>
TDLS200 TDL Analyzer Instruction Manual V2.1
Following pages are the various “Laser Spectra& Control” screen captures that depict the differentscenariosdepending on what process gas(es) are present in theOPL and what gas is purgedthrough theanalyzer:
Note that the actual magnitude of the absorptionunits (au) asindicatedon the “AbsorptionSpectrum”Y-Scale (vertical) will vary dependingupon the actual gas concentrations,optical pathlengths, gas temperatures and gaspressures.
Following pages are the various “Laser Spectra & Control” screen captures that depict the different scenarios
depending on what process gas(es) are present in the OPL and what gas is purged through the analyzer:
Note that the actual magnitude of the absorption units (au) as indicated on the “Absorption Spectrum” Y-Scale
(vertical) will vary depending upon the actual gas concentrations, optical path lengths, gas temperatures and gas
pressures.
Off-Line Calibration Conditions:
Process/Calibration OPL: N2
Process/Calibration OPL: N2
Analyzer Purge: N2
Analyzer Purge: N
This is how the absorption spectra will appear
This is how the absorption spectra will appear when
when there is neither NH
there is neither measured or reference gas (i.e.NH
optical path (Process OPL and Analyzer Purge).
and Hand Analyzer Purge).
O) in the entire optical path (Process OPL
2
2
nor H2O in the entire
3
3
This is the spectra appearance
required for an Off-Line Zero
This is the spectra appearance
required for an Off-Line Zero
Calibration of the analyzer.
Calibration of the analyzer.
Process/Calibration OPL: ~20ppm NH
Process/Calibration OPL: ~20 ppm NH3
Analyzer Purge: N
Analyzer Purge: N
In this Absorption Spectra view, there is practi-
In this Absorption Spectra view, there is practically
cally zero H2O absorption peak in the approxi-
zero reference gas (H
mate peak center position (PCP) 167 region.
approximate peak center position (PCP) 167 region.
There is approximate 20 ppm NH3 (@ approx
There is approximate 20 ppm of me a s ured gas
PCP ~120) absorption in this spectra based on
(NH
) (@ approx PCP ~120) absorption in this
3
spectra based on 72.6cm OPL, ambient
72.6cm OPL, ambient temperature and ambient
temperature and ambient pressure.
pressure.
This is the spectra appearance
This is the spectra appearance
required for an Off-Line Span
required for an Off-Line Span
Calibration of the analyzer.
Calibration of the analyzer.
2
2
O) absorption peak in the
2
3
IM 11Y01B01-01E-A 6th Edition :Feb 13, 2013-00
On-Line Process Conditions – Analyzer Purged with Nitrogen:
TDLS200 TDL Analyzer Instruction Manual V2.1
On-Line Process Conditions – Analyzer Purged with Nitrogen:
Process OPL: Measured and Reference gas (NH3 and H2O)
Analyzer Purge: N2
In this Absorption Spectra view, the reference gas (H2O) absorption peak can be seen at approximate peak center position (PCP) 167. This absorption peak is from the reference gas contained in the process stream ( H2O) as there is no reference gas (H2O) in the analyzer purge gas.
TDLS200 TDL Analyzer Instruction Manual V2.1
On-Line Process Conditions – Analyzer Purged with Nitrogen:
Process OPL: Measured and Reference gas (NH3 and H2O)
Analyzer Purge: N2
In this Absorption Spectra view, the reference gas (H2O) absorption peak can be seen at approximate peak center position (PCP) 167. This absorption peak is from the reference gas contained in the process stream ( H2O) as there is no reference gas (H2O) in the analyzer purge gas.
There is also approximate 20ppm NH3 (measured gas) (@ approx PCP 120) absorption in this spectra based on 72.6cm OPL, ambient temperature and ambient pressure.
Process OPL: Zero Measured Gas (NH3) + ~2%H2O (Reference Gas)
Analyzer Purge: N2
In this Absorption Spectra view, the large reference gas (H2O) absorption peak can be seen at approximate peak center position (PCP) 167.
TDLS200 TDL Analyzer Instruction Manual V2.1
On-Line Process Conditions – Analyzer Purged with Nitrogen:
Process OPL: Measured and Reference gas (NH3 and H2O)
Analyzer Purge: N2
In this Absorption Spectra view, the reference gas (H2O) absorption peak can be seen at approximate peak center position (PCP) 167. This absorption peak is from the reference gas contained in the process stream ( H2O) as there is no reference gas (H2O) in the analyzer purge gas.
There is also approximate 20ppm NH3 (measured gas) (@ approx PCP 120) absorption in this spectra based on 72.6cm OPL, ambient temperature and ambient pressure.
Process OPL: Zero Measured Gas (NH3) + ~2%H2O (Reference Gas)
Analyzer Purge: N2
In this Absorption Spectra view, the large reference gas (H2O) absorption peak can be seen at approximate peak center position (PCP) 167.
There is no measured gas (NH3) absorption in the process .
Process OPL: ~20ppm Measured Gas (NH3) + ZeroReferenceGas(H2O)
Analyzer Purge: N2
In this Absorption Spectra view, there is practically zero reference gas contained in the process stream. No absorption from H2O at the approximate peakcenter position (PCP) 167 region.
There is approximate 20ppm measured gas (NH3)
Process OPL: NH3 and H2O
Analyzer Purge: N
In this Absorption Spectra view, the Process H2O
absorption peak can be seen at approximate
peak center position (PCP) 167. This absorption
peak is Process H2O as there is no reference gas
H2O in the N2 analyzer purge gas.
There is also approximate 20 ppm NH3 (@ approx PCP
120) absorption in this spectra based on 72.6 cm OPL,
ambient temperature and ambient pressure.
Process OPL: Zero NH3 + ~2% H2O Air
Analyzer Purge: N
In this Absorption Spectra view, the large process
H2O absorption peak can be seen at
approximate peak center position (PCP) 167.
There is no NH3 absorption in the spectra .
<6. BASIC OPERATION>
2
2
6-24
Process OPL: ~20 ppm NH3 + Zero H2O
Analyzer Purge: N
In this Absorption Spectra view, there is practically
zero process H2O absorbtion peak in the approximate
peak center position (PCP) 167 region.
There is approximate 20 ppm NH3 (@ approx PCP
~120) absorption in this spectra based on 72.6 cm
OPL, ambient temperature and ambient pressure.
2
IM 11Y01B01-01E-A 6th Edition :Feb 13, 2013-00
<6. BASIC OPERATION>
On-Line Process Conditions – Analyzer Purged with Nitrogen:
When purging with Instrument Air (as opposed to ambient air), the H2O levels are much smaller (due to it
having been dried and having a dew point typically in the order of -40˚C) and therefore the H2O
concentration is not large enough to produce an absorption peak that can be seen.
Process OPL: Zero NH3 + Zero H2O
Analyzer Purge: Air (typical ~2-4% H2O)
In this Absorption Spectra view, the analyzer air purge H2O
absorption peak can be seen at approximate peak center position
(PCP) 167.
There is no NH3 absorption in the spectra.
Process OPL: ~20 ppm NH3 + Zero H2O
Analyzer Purge: Air (typical ~2-4% H2O)
In this Absorption Spectra view, the H2O absorption (analyzer air
purge) peak can be seen at approximate peak center position (PCP)
167.
There is approximate 20 ppm NH3 absorption (approx 120PCP)
in this spectra based on 72.6 cm OPL, ambient temperature and
ambient pressure.
Process: Zero NH3 + ~2% H2O
Analyzer Purge: Air (typical ~2-4% H2O)
In this Absorption Spectra view, the process gas H2O (~2% 0.7 m
@ STP) and analyzer air purge H2O absorption peak can be seen at
approximate peak center position (PCP) 167.
There is no NH3 absorption in the spectra.
IM 11Y01B01-01E-A 6th Edition :Feb 13, 2013-00
<6. BASIC OPERATION>
TDLS200 TDL Analyzer Instruction Manual V2.1
4.6 Large Aperture Optics
For in-situ application (typically large scale combustion systems) the optical path lengths are generally very long (7-30m for large scale combustion and ethylene furnaces). Standard TDL optics and laser beam configurations are unsuitable because of the mechanical stability of these large scalecombustion systems. Yokogawa Laser Analysis Division therefore designed and developed the concept of a diverging beam (i.e. a laser beam that expands over distance) and alarge aperture optics detector scheme (i.e. a large target for the laser to hit).
The general concept of diverging beam and large aperture optics is shown below:
6.5 Large Aperture Optics
For in-situ application (typically large scale combustion systems) the optical path lengths are generally very long
(7-30 m for large scale combustion and ethylene furnaces). Standard TDLS optics and laser beam congurations
are unsuitable because of the mechanical stability of these large scale combustion systems. Yokogawa Laser
Analysis Division therefore designed and developed the concept of a diverging beam (i.e. a laser beam that
expands over distance) and a large aperture optics detector scheme (i.e. a large target for the laser to hit). The
general concept of diverging beam and large aperture optics is shown below:
Figure 52
In a standard TruePeak TDLS200 analyzer, the laser beam exiting the launch unit is normally collimated parallel
before hitting the opposing detect unit. The collimated beam size is typically less than 1” diameter. However,
this optical layout is not appropriate for long-path applications (the dimension of process is longer than 30 feet).
During initial installation, it is also difcult to align the laser beam so that it can hit the targeted detect unit over
a long distance (small changes in the launch unit angle are magnied over long distances). Also it is almost
impossible to keep good alignment with varying ambient and process conditions – especially during cold starts
and shut-downs (the most extreme thermal changes on the mechanical structures).
6-26
To resolve the above issues, Yokogawa Laser Analysis Division has developed a diverging beam and large
aperture optics strategy. At the launch side, the output laser beam has a small diverging angle. For example, the
beam size is about 20” (~50 cm) diameter at 60’ (~18 m) optical distance. The optical aperture at the receive unit
is enlarged from the original 1¾” diameter. With these two changes, it is much easier to do initial alignment and
keep good transmission during a wide range of operating conditions.
Please note that when an analyzer is mounted on the standard 0.726m (28.6”) OPL off-line calibration cell, there
is a large amount of laser light on the detector. Once moved onto a long path installation, the diverging beam
power is spread out over a larger area and hence weaker, this means that once installed on the long path, the
detector gain may require some manual adjustment (increase typically) – please use the following procedure.
IM 11Y01B01-01E-A 6th Edition :Feb 13, 2013-00
<6. BASIC OPERATION>
6.5.1 Large Aperture Optics Installation, Alignment & Detector Gain
The alignment of the Large Aperture Optics (or LAO) is quite similar to the alignment method of a standard
TDLS200. To prepare for the alignment you will need some form of screen at the analyzer this can be a launch
unit with a screen and keypad, a mini display, or a laptop PC connected through VNC/Ethernet. Go into “Laser
Spectra & Control” observe the “Trans. (%)”. You will need to set up the alignment by rst getting both the
launch and detect sides as perpendicular to the optical plane as possible. This will allow for the initial signal to
be found more easily. Note, the oxygen analyzers tend to be easier to align than CO/CH4 analyzers due to the
optical/detection internal congurations.
Once this has been done, adjust the launch unit till a change in transmission can be seen. Do this by loosening
the nuts from the launch unit and manually point the launch unit till some transmission is observed.
Note: the change in transmission might be very small depending on the incident angle on the detect optic. It
is easy to pass over the correct launch alignment without knowing it. If a change cannot be found,
repeat with smaller adjustments in both the X-axis and Y-axis. If no transmission can be established
by movement of the launch unit, check the general alignment of the detect unit to ensure its axis is
generally aligned with the launch unit axis.
Once some transmission is observed align the detect side so that the transmission is at a maximum. To do
this, back off all four of the setscrews. With two people, one to do the adjustments and one to watch the
transmission, slowly adjust the detect alignment by loosening one direction and tightening the opposite till the
maximum transmission can be found. Place setscrews so that the alignment is secure.
With the detector aligned begin vertical adjustments of the launch unit using the studs for small, ne
adjustments to attain maximum transmission. Once maximum transmission has been found in the vertical
position, adjust the horizontal plane for maximum transmission. With maximum transmission found in the
horizontal position repeat the alignment in the vertical position for maximum transmission. Once the maximum
transmission has been found once again in the vertical direction, align the horizontal direction again to nalize
maximum transmission alignment.
With the analyzer at maximum transmission, the beam needs to be centered on the detector. Adjust the laser
vertical plane till transmission is almost lost i.e. the edge of the beam is almost missing the detector lens.
Adjust the laser in the opposite vertical plane till the transmission is almost lost once again. Take note on how
many quarter turns that it takes to get from one side of the beam to the other. Repeat back to the other side
of the beam so that the transmission is almost lost once again. Make sure that the number of quarter turns is
similar. Adjust the beam back by half the number of quarter turns needed. The beam will be centered vertically.
Repeat for the horizontal direction using quarter turns once again. This will fully center the beam on the detector.
Now the detector will need to be aligned once again. Make sure that all backing nuts and setscrews are
securely fastened.
Carefully check the detector Large Aperture Optics bevel mounting seal is not leaking at the perimeter,
this can happen if too much bias is asserted. For oxygen analyzer and combustion processes with
negative pressure, this will cause false high readings as ambient air (20.9% O2) is drawn into the optical
path via a gap at the bevel seal edge – please clamp down all bolts and adjust the large bevel ange to
ensure a gas tight seal is maintained at all times.
IM 11Y01B01-01E-A 6th Edition :Feb 13, 2013-00
<6. BASIC OPERATION>
6.5.2 Adjustment of Detector Gain for Large Aperture Optics
This section of the User’s Guide gives the guidance and procedures to adjust the detector board gain after the
analyzer is installed and powered up. This operation should be done by qualied personnel. Read instructions
fully before starting this operation. Any doubts or questions, please contact Yokogawa.
PHOTOS OF DETECTOR BOARD
6-28
RESISTOR BAG
For a TDLS200 analyzer with diverging laser beam conguration (either with or without large aperture detector
lens), a resistor bag is attached inside the detector box. The resistor bag contains the resistors with the following
values. All of them are ¼ W, 5% tolerance metal lm through-hole resistors.
Resistor Bag for Analyzers With Diverging Beam Optics
1. This procedure requires wearing a grounding strap connected to one of the grounding lugs of the
analyzer to prevent any electrostatic damage.
2. Open the detector enclosure and put on a grounding strap.
3. Identify the gain resistors R21, R22 and R23 on the detector board as shown in the photo above. They are
all plugged in sockets instead of being soldered, easy to be modied by a pair of small pliers. The existing
gain resistors have factory resistance values optimized on a calibration cell.
IM 11Y01B01-01E-A 6th Edition :Feb 13, 2013-00
<6. BASIC OPERATION>
4. [Optional] For a high-process-temperature application and if the current process temperature is higher than 600°C,
apply a multi-meter to measure the voltage across R21. If the measured voltage absolute value is greater than 5V,
change R21 to the 100Ω resistor from the resistor bag. Keep the original R21 in the resistor bag as it can be used
for future ofine calibration.
5. Remove R22 and R23 from the detector board and replace them with the 11kΩ resistors from the resistor bag.
Please keep the original R22 and R23 in the resistor bag as they can be used for future ofine calibration.
6. Optimize the analyzer alignment for both the launch and detector sides. Stop the alignment if Detector Signal High
fault activates.
7. In the TruePeak user interface software, navigate to Advanced Menu (password) -> Congure -> Laser Spectra &
Control screen to check the raw detector signal. Write down the raw detector signal MAX and MIN values for later
use.
8. Change R22 and R23 accordingly based on the current raw detector signal. Perform one item of the following
selections (a, b, or c).
a. If Detector Signal High fault is active, change both R22 and R23 to the next smaller value available in the resistor
bag, and then go back to STEP 6. For example, if the current R22 and R23 values are 11kΩ, the new R22 and R23
values should be 7.5kΩ. NOTE: please access Active Alarm in the main user interface panel to check if Detector
Signal High fault is active.
b. If the raw detector signal is within requirement, remove R22 and R23 (and R21 if it was changed to 100Ω in STEP
3) and cut their leads properly to t the sockets tight and low, and then go to STEP 9. NOTE: (1) for a low-process-
temperature analyzer (with no capacitor on R3), the raw detector signal is within requirement if MAX is between
0.0 and 4.0; (2) for a high-process-temperature analyzer (with a capacitor on R3), the raw detector signal is within
requirement if MIN is between -4.0 and -8.5.
c. If the raw detector signal is too low (other than a and b), change either R22 or R23 to the next greater value
available in the resistor bag, and then go back to STEP 7. How to decide which resistor to change:
i. If the current R22 value is greater than the current R23 value, change R23 to the next greater value available
in the resistor bag.
ii. If the current R22 value is same as the current R23 value, change R22 to the next greater value available in
the resistor bag.
iii. If the current R22 and R23 are already 24kΩ (the greatest value available in the resistor bag), please
contact Yokogawa for assistance.
IM 11Y01B01-01E-A 6th Edition :Feb 13, 2013-00
<6. BASIC OPERATION>
9. Close the detector enclosure. Write down the nal values of R21, R22 and R23 below for record.
10. Keep the resistor bag and this procedure document by customer. DO NOT leave them in the detector box or
throw them away.
11. Enter these new resistor values into the analyzer conguration under Advanced Menu, Congure, System,
Adjustable Resistors Detect – R22, Detect R23 and Detect R-23. NOTE, all resistors values are entered in
kOhm units.
6.5.3 Detector Gain Adjustment Service Tips
• How to remove a gain resistor? On the detector board, nd the location of the gain resistor. Please note that
the resistor is not soldered but resting in sockets. Carefully remove the resistor from the socket on the board.
Since the analyzer is still powered on, please take extreme caution to prevent a short circuit on the board, i.e.
loose wires or touching adjacent components with pliers or tool used to remove resistor.
• How to install a new gain resistor? First try the new resistor by molding and clipping the resistor leads to t
into the socket. Then carefully insert the resistor into the socket on the detector board. Since the analyzer is
still powered on, please take extreme caution to prevent a short circuit on the board, i.e. loose wires or
touching adjacent components with pliers or tool used to install resistor.
• Please keep the factory/original gain resistors in the resistor bag as they will be used in the future for ofine
calibration or test.
6-30
• The raw detector signal (MAX-MIN) is proportional to R21·R22·R23. Usually R21 is maintained as factory value.
R22 and R23 are optimized in the eld. We want to avoid the situation where one of R22 and R23 resistors has
extreme low resistance and the other one has extreme high resistance. The analyzer gives best performance
when R22 and R23 are balanced. Based on this relation between raw detector signal and gain resistors,
service technician or customer can select R22 and R23 faster with the help of a calculator.
• Multiple alignment actions might be needed if Detector Signal High fault happens. Reduce the gain resistor
rst as described in STEP 8-a, and then optimize the alignment again.
6.6 Valve Control Logic
The TDLS200 has three valve driver outputs (24VDC, 11W max each) available at Launch unit
connection terminal TB-3. those can be used for calibration/validation functions and/or stream switching
functions if being used in an extractive installation. The valve control logic is outlined below:
Manual Switch
• Turn the valve on.
o If time sequence is on, record valve_start[0] (current minute).
o No need to turn off other valves. Multiple valves can be on at the same time.
• Turn the valve off.
IM 11Y01B01-01E-A 6th Edition :Feb 13, 2013-00
<6. BASIC OPERATION>
TIME SEQUENCE
• It can happen only when there is no active auto calibration or validation.
• When time counting is up,
o Turn off the current valve.
o Turn on the next valve.
o Start the new time counting.
o Multiple valve time sequences are allowed. When congure the parameters, please pay attention.
Time sequence is not recommended to use with remote control at the same time to avoid valve
chaos.
REMOTE CONTROL
• The remote initiate contacts are found at TB-2 and the external contact MUST BE VOLTAGE FREE.
The TDLS200 circuits are low level current monitoring so any externally applied voltage/current
may cause damage to the analyzer, hence use voltage free contact only.
• It can happen only when no auto calibration or validation is on-going.
• Direct control from DCS and or local/remote switch(s)
• Contact signal -> valve on (please keep remote contacts closed for at least 10seconds to ensure the
analyzer reads/registers the contacts at measurement up-date)
• Open signal -> valve off (please keep remote contacts closed for at least 10seconds to ensure the analyzer
reads/registers the contacts at measurement up-date).
AUTO CALIBRATION & VALIDATION – see separate section of User Guide for details
1. Remember the current valve status.
2. Turn on the assigned valve, turn off other valves.
3. Resume the original valve status before auto calibration or validation.
IM 11Y01B01-01E-A 6th Edition :Feb 13, 2013-00
<6. BASIC OPERATION>
AUTO ONLINE VALIDATION – see separate section of User Guide for details
1. Remember the current valve status.
2. Turn on the assigned valve(s), turn off other valves. Two valves could be
open for auto online validation with blocked line lock gas in validation cell.
3. If only one valve is used, turn off the assigned valve. If two valves are used, turn off the first valve.
4. If two valves are used, turn off the second valve.
6.7 Introduction for H2Oppm Measurements in Methane Gas
The measurement of H2Oppm in methane gas (LNG/NG) is specifically outlined in this section. Items contained
within this section will supersede any prior information within the User’s Guide, particularly items pertaining to
calibration methods. Please also note the special requirements needed for the flow cell measuring conditions as
these can affect the measured H2Oppm concentration values.
Spectroscopic detection of moisture in natural gas presents a specific challenge for traditional TDLS methods,
because a weak (at low ppm moisture content) absorption line of water must be detected on a high absorption
background of methane and other components (C2/C3) of the natural gas. This challenge has been successfully
overcome with TDLS200 by using a special (previously proven) approach to spectroscopic data processing.
However, reliability of results relies on stable and well defined conditions in the flow cell of the TDLS200 analyzer.
Note that typically temperature of the process/calibration gases within the flow cells is to be maintained
at 50.0±0.5 ºC and the gas pressure controlled at 30±0.25 psig. Maintaining these conditions ensures welldefined absorption of methane and reliable ppm moisture concentration readings.
6-32
At the same time, this method eliminates the need for routine analyzer calibrations (zero or span). This is due
to the fact that the strength of molecular absorption is an intrinsic property of the chemical species. When the
gas temperature, pressure, and its bulk composition are defined, absorption spectrum depends only on the
absorption path length, which is also fixed in the supplied analyzers. These analyzers are factory calibrated and
Therefore, the H2Oppm in Natural Gas analyzers DO NOT require actual field calibration by end users
– they can however be Validated. Calibration procedure described in the standard TDLS200 User Guide is
intended for the different measurements/techniques of data processing and must not be used in these units. If
applied to these H2O-NGas analyzers, it will disrupt the file system with unpredictable/erroneous results requiring
factory assistance to restore proper functionality.
At all times, the analyzer internal optics must be purged with dry nitrogen (<0.25 ppm H2O content). Any moisture
present within the analyzer internal optics purge gas (or on-validation cell purge gas) will be added to the actual
process gas measurement of H2Oppm.
DO NOT ZERO or SPAN CALIBRATE THE ANALYZER without prior written approval from the Yokogawa
Laser Analysis Division Factory!
Validation (response check) of the analyzer can be performed in two possible ways:
(1) Off-Line Method (interrupting process flow through the cell): A pure dry (<5 ppm H2O) methane (100%
CH4) validation gas can be supplied to the flow cell (at same flow conditions 50ºC and 30 psig); analyzer
readings for CH4 % must be 100±2%. Note, at this time the analyzer will also indicate any value of
residual moisture in the pure dry methane validation gas standard.
(2) On-Line Method (no interruption of process flow through the cell) to observe a step change in the
H2O ppm output value: Nitrogen gas with typical 100-120 ppm H2O (+/-10 ppm) content can be switched
through the on-line validation cell at 10psig (for example, equipped with a 10 psig pressure relief valve
on validation cell outlet). The analyzer readings for H2O will be increased by approximately 10-12 ppm
(based on 110 ppm*pathval.cell/pathflow cell) or roughly equivalent to 10% of the cylinder certified
H2O ppm value. This is an observation only of step change to the introduction of H2O into the optical
path, sometimes considered a “response check” or “analyte spiking”.
IM 11Y01B01-01E-A 6th Edition :Feb 13, 2013-00
Normal Operation – TDLS200 TruePeak Software Main
Menu
The analyzer is configured to measure not only H2O ppm
but it will also indicate the CH4 % concentration value, as
shown on the Main Menu display below: Note that the flow
cell gas temperature is also shown as Active (based on flow
cell sensor active input) and pressure is input as a Fixed
value (shown in psiA units, based on the external sample
system regulator fixed setting).
Note, during normal process operation with 90100% methane gas flowing in the cell, the laser
temperature set point (LTSP) is configured to control
the methane peak position at data point 220 (as
shown red in absorption spectrum, right):
<6. BASIC OPERATION>
6-33
The actual measurement of H
peak is made in the data point region of 160 – note
that H2Oppm is not visible to the eye within this
absorption spectrum when <100 ppm (as shown
blue in absorption spectrum, right):
6.8 Introduction to Gas Temperature Predictions with High Temperature Oxygen Measurements
For some high temperature oxygen measurements operating at the correct conditions, the analyzer can predict
the gas temperature as well as the oxygen concentration. The TruePeak TDL analyzer scans a tunable diode
laser over a small wavelength range across measurement gas absorption line profile. The measurement gas
concentration can be predicted from the absorption peak with combustion parameters of optical path length, gas
temperature and gas pressure. The gas temperature has significant effect on the absorption line intensity, typically
weaker absorption at higher temperature for oxygen. However, there are several oxygen lines whose absorption
intensity has different temperature response. By simultaneously comparing two oxygen lines with different
temperature effect on absorption intensity, we can calculate the gas temperature.
O ppm absorption
2
IM 11Y01B01-01E-A 6th Edition :Feb 13, 2013-00
<6. BASIC OPERATION>
The following graph illustrates an example. At 800 degC, there are two oxygen absorption peaks, and their ratio
is 8.6. At 1500 degC, the left oxygen absorption peak increases while the right peak drops significantly. The peak
ration is 2.2. So by calculating the peak ratio of these two oxygen peaks, the gas temperature can be predicted.
Figure 54
The requirements for accurately predicting process gas temperature with high temperature oxygen
analyzers:
• Gas temperature should be higher than 750˚C so it’s ideal for combustion zone gas applications (typical
measured range 750-1,500˚C or 1380-2,730˚F).
• It needs sufficient oxygen absorption so longer path length and/or higher oxygen levels are preferred – contact
Yokogawa Laser Analysis Division with application details
6.9 Controlling the Analyzer Remotely or Locally via external PC/Laptop
A number of methods are available to connect to the TDLS200 analyzer.
• Direct Access with optional Keypad and Display
• Direct Access using VNC and an external computer (Ethernet)
• Remote Access using VNC via network (external computer)
• Remote Access using optional Remote Interface Unit (RIU with VAC software)
Whether directly connected, or connected via network, operating the analyzer with an external connection allows
two basic functions:
• Remote control of the analyzer via TDLS200 software allows full control of the analyzer. The user will see the
same screen with the same access functions as if controlling using a built in keypad and display.
• Data transfer via VIRTUAL ANALYZER CONTROLLER (VAC) software allows download
of data files to/from the analyzer.
IM 11Y01B01-01E-A 6th Edition :Feb 13, 2013-00
<6. BASIC OPERATION>
6.9.1 Instructions for Connecting an External Computer to the Analyzer
Requires Windows 98SE or later (on computer to be connected to the analyzer) and crossover Ethernet cable.
The analyzer is provided with a CD that includes the VNCviewer program. Windows 7 will support the UltraVNC software and associated functionality. Contact your local Yokogawa agent for a free copy of the necessary
“VNCviewer.exe” le that will enable the VNC connection with the analyzer. This VNCviewer.exe le should be
loaded on to the connecting PCs desktop ready for use when connecting with the analyzer.
• From Control Panel – Network Connections, make sure Ethernet Local Area Connection is set to
Enabled status. Disable wireless and any other networking connections.
• Connect crossover Ethernet cable from client system to Analyzer.
• On the computer, go to Control Panel – Network Connections - Local Area Connection, Internet
Protocol (TCP/IP) Properties.
o Set IP to 10.255.255.254 & Subnet Mask to 255.0.0.0. Select OK to accept changes on Internet
Protocol (TCP/IP) and Local Area Connection Properties.
o Start Ultra-VNC software, running from Desktop using the Guide below:
6-35
SBC Ethernet Port
Figure 55 - Connecting External Computer to the analyzer
IM 11Y01B01-01E-A 6th Edition :Feb 13, 2013-00
<6. BASIC OPERATION>
TDLS200 TDL Analyzer Instruction Manual V2.1
TDLS200 TDL Analyzer Instruction Manual V2.1
4.8.2 Using Ultra-VNC Software
Start the Ultra-VNC software by double-clicking on the “vncviewer.exe” ICON (as shown below):
Within the VNC Server field, enter the correct IP address for the analyzer to which you are connecting then click on “Connect” button. If a successful connection is established then use the default password for entering the VNC connection screen is 1234 – see screen below that shows an example IP address of 10.0.2.14 (for an analyzer Serial number XX-1214-XX etc.):
DO NOT attempt to change of the “Quick Options” or anyother settings on this menu!
TDLS200 TDL Analyzer Instruction Manual V2.1
4.8.2 Using Ultra-VNC Software
Start the Ultra-VNC software by double-clicking on the “vncviewer.exe” ICON (as shown below):
Within the VNC Server field, enter the correct IP address for the analyzer to which you are connecting then click on “Connect” button. If a successful connection is established then use the default password for entering the VNC connection screen is 1234 – see screen below that shows an example IP address of 10.0.2.14 (for an analyzer Serial number XX-1214-XX etc.):
DO NOT attempt to change of the “Quick Options” or anyother settings on this menu!
If the analyzer connection cannot be established (see error message below) then check the PC IP settings, connection wires/Cat5 cable and IP address.
3.3 Inlet Temperature Section................................................................6
The following safety symbols are used onthe product as well as in this manual.
DANGER
This symbol indicates that an operatormust follow the instructions laid out inthis manual in order to avoid the risks, forthe human body, of injury, electric shockor fatalities. The manual describes whatspecial care the operator must take toavoid such risks.
WARNING
6.9.2 Using Ultra-VNC Software
Start the Ultra-VNC software by double-clicking on the “vncviewer.exe” ICON (as shown below):
Within the VNC Server eld, enter the correct IP address for the analyzer to which you are connecting then click
on “Connect” button. If a successful connection is established then use the default password for entering the
VNC connection screen is 1234 – see screen below that shows an example IP address of 10.0.2.14 (for an analyzer Serial number XX-1214-XX etc.):
For analyzers with Serial/Tracking number XX.2xxx-xx, use IP address 10.0.20xx, for example 76H-2018-12-Ex
would have an IP address 10.0.20.18
For analyzers with Serial/Tracking number XX.3xxx-xx, use IP address 10.0.30xx, for example 23-3007-13-Ex
would have an IP address 10.0.30.7
DO NOT attempt to change of the “Quick Options” or any other settings on this menu!
If the analyzer connection cannot be established (see error message below) then check the PC IP settings,
connection wires/Cat5 cable and IP address.
IM 11Y01B01-01E-A 6th Edition :Feb 13, 2013-00
<6. BASIC OPERATION>
6.9.3 OPTIONAL Remote Interface Unit (RIU)
The OPTIONAL RIU runs the Virtual Analyzer Controller (VAC) software described below.
The RIU VAC software is designed to allow communication
between an analyzer and an appropriate interface.
The primary functions of the software are:
Create a virtual network computing connection to an analyzer thus
allowing for control of the analyzer, typically for start-up, service,
calibration, etc.
Allow for le transfer from an analyzer to a local USB port (for
memory device)
Create connections by name and/or IP address
After selecting “Connect F2” you will be allowed to select the
analyzer (with description and IP address) you wish to control
remotely.
This will allow full analyzer control as if you were at the analyzer
itself.
Please remember that when you have nished working on the
analyzer via this RIU VAC software, you must END SESSION
to the analyzer by pressing the Control+Backspace keys
simultaneously – DO NOT SHUT DOWN analyzer!!!
After selecting “Copy Data – F3” you will be allowed to select the
analyzer (with description and IP address) you wish receive data
from.
This will initiate a data transfer for all results and conguration les
stored on the analyzer.
NOTE:
Please ensure there is a ‘clean’ (or empty) SanDisk USB memory
stick inserted into either of the two USB ports of the RIU.
When copying data from an analyzer via the RIU, a dedicated
folder with the serial number is NOT created, the data les are
dumped to the root directory of the USB stick.
IM 11Y01B01-01E-A 6th Edition :Feb 13, 2013-00
<6. BASIC OPERATION>
After selecting “Conguration – F4” you will be asked to Enter
Password – the default password is 1234, then press F2 to proceed
and then you will be allowed to select the analyzer (with description
and IP address) you wish to congure.
You will be given the following menu choices:
Create Connection - F2: This will allow programming of Tag Name
and IP Address for future connections
Delete Connection - F3: This is to delete an existing connection
Options - F4: (see next section)
6-39
Password - Ctl-Ins: Allows changes to the access password
VAC Options (Conguration Menu):
Data Dump Options. This sets the directories to receive data from
(analyzer) and send data to (on system running VAC software)
Source Directory: This is the analyzer data le folder; it should not be
changed without factory consultation.
Target Directory: This is the remote computer or RIU directory to
receive data les
File Masks: These are the extensions of the les to be transferred; it
should not be changed without factory consultation.
VNC Viewer Location species the location of the software to
remotely control an analyzer; it should not be changed without factory
consultation.
APPLY-F2 must be selected to save changes
From the VAC Main Menu Ctrl-F1 will bring up the Help Screen.
This will give at system text help describing the shortcut keys and their
function.
IM 11Y01B01-01E-A 6th Edition :Feb 13, 2013-00
<6. BASIC OPERATION>
To shut down the RIU, press the Esc key and the screen will appear as
shown. Press 9 to continue with the shut-down process and when the
RIU screen turns white, the power can be switched off.
Note, there is no watch-dog in the RIU so un-like the TDLS-200, it will
not automatically re-start after a period of time if power is not switched
off.
ALWAYS “END VNC SESSION” TO ANALYZER WHEN DONE – THIS
WILL PREVENT THE ANALYZER AND/OR RIU FROM ‘HANGING’!
DO NOT LEAVE THE RIU PERMANENTLY CONNECTED TO
ANALYZER – CONNECT ONLY WHEN IN USE
IM 11Y01B01-01E-A 6th Edition :Feb 13, 2013-00
<7. ROUTINE MAINTENANCE>
7 ROUTINE MAINTENANCE
The TDLS200 TDL analyzer requires little routine maintenance if it has been installed, set-up and calibrated
correctly. This section will outline the routine maintenance procedures.
7.1 Maintaining Good Transmission
The % Transmission of the laser light through the process is the most important variable to consider for routine
maintenance and troubleshooting. Under normal operating conditions (nonfailure) transmission will be affected
by:
• Alignment of the Launch and Detect units (covered below)
• Window fouling. For most applications where solids or liquids are expected in the process a window
purge is recommended (see installation section)
• Particulate in the process. Particulate in the process will block the laser light to some extent. To
maximize performance in particulate laden processes it is important to maximize alignment and ensure
windows are clean.
7.1.1 Maintaining Clean Windows
If the window purge supplies have been set-up correctly on a reliable supply line, then under normal operating
conditions there should be no window fouling. Window fouling may be caused by one or more conditions:
• Loss of window purge gas ow/pressure which then allows the dirty process gas to contact the windows
and leave deposits.
• Loss of window purge gas ow/pressure which then allows the hot process gas to contact the window,
condense any liquids and leave deposits.
• Over pressure of the process gas which allows the process gas to overcome the window purge gas and
contaminate the window.
• Contact on the window by an incompatible gas (such as HF on the standard BK-7 Borosilicate crown
glass).
• Contamination of the purge gas supply, either by particulate matter, oil or other.
7-1
If the windows are contaminated they may be cleaned using the following procedure:
• Remove analyzer from process (isolate, etc. if necessary).
• Use a clean, dry instrument air or nitrogen pressure supply to rst blow off any particulate matter.
• Using warm water and mild soap detergent, gently clean the window surface with a soft, non-abrasive
cloth.
• If the deposits do not come off then use a small amount of IPA (Isopropyl Alcohol) and a soft,
non-abrasive cloth.
• Use the same clean, dry instrument air or nitrogen pressure supply to blow dry the surface.
• Carefully check the entire surface of the window from different angles to ensure it is thoroughly cleaned
and ready for service.
• If the window does not appear to clean up well, then replace the window assembly with a new one.
• If the window appears to have an etched surface then it has probably been contaminated with HF or
other similar corrosive gas.
IM 11Y01B01-01E-A 6th Edition :Feb 13, 2013-00
<7. ROUTINE MAINTENANCE>
TDLS200 TDL Analyzer Instruction Manual V2.1
If the windows are contaminated they may be cleaned using the following procedure:
•Remove analyzer from process (isolate, etc. if necessary)
•Use a clean, dry instrument air or nitrogen pressure supply to first blow off any particulate
matter.
•Using warm water and mild soap detergent, gently clean the window surface with a soft, non-abrasive cloth.
•If the deposits do not come off then use a small amount of IPA (Isopropyl Alcohol) and a soft, non-abrasive cloth.
•Use the same clean, dry instrument air or nitrogen pressure supply to blow dry the surface.
•Carefully check the entire surface of the window from different angles to ensure it is
thoroughly cleaned and ready for service.
•If the window does not appear to clean up well, then replace the window assembly with a new one.
•If the window appears to have an etched surface then it has probably been contaminated with HF or other similar corrosive gas.
TDLS200 TDL Analyzer Instruction Manual V2.1
o
seated
positio
Window surface to
be cleaned
O- ring correctly
O-ring correctly
seated in position
in
Offset screw position
f
r window
Figure 56 - Window Replacement
n
7-2
Figure 57 - Window Cleaning
IM 11Y01B01-01E-A 6th Edition :Feb 13, 2013-00
<7. ROUTINE MAINTENANCE>
7.1.2 Replacing Windows
If the windows are contaminated they may be cleaned using the following procedure:
• If the analyzer has not yet been shut down, then please shut-down the analyzer properly and remove
power to ensure the laser if OFF
• Remove Launch or Detect (as appropriate) from the process interface by removing
the one xed hole fastener (5/32” Allen Wrench), loosening the remaining three
fasteners and then twisting and pull off the unit.
• Remove the four retaining screws (3/32” Allen Wrench, T-Handle) and lock washers
from the window holder recessed inside the quick connect body.
• Remove the window that requires replacement.
• Ensure the new replacement window is of the same specication as the window
being removed. Standard Windows are 0.25” Thick wedged, BK-7 borosilicate crown
glass mounted in stainless steel retaining ring.
• The window holder assemblies can only be installed in one way by virtue of the
mounting hole placement.
• Ensure a new O-Ring (of same specication) is installed at the same time. Viton is the
standard material however, certain applications may require other materials – check
original specications.
• With the new O-Ring in position, carefully locate the new window holder in place with
the smooth glass surface against the O-Ring.
• Carefully tighten all four screws (ensuring the lock washers are in place) using a
T-handle wrench. Keep tightening all the fasteners in turn by ¼ turns thus ensuring an
even torque loading.
• Hand tighten the screws until they are very rmly tightened (a T handle wrench as
shown below will torque its own shaft by about ¼ turn when the screws are fastened
securely).
IM 11Y01B01-01E-A 6th Edition :Feb 13, 2013-00
TDLS200 TDL Analyzer Instruction Manual V2.1
5.2 Alignment
7.2 Alignment
<7. ROUTINE MAINTENANCE>
8-4
7-4
Alignment Studs (x4)
Alignment Nuts (x8)
Window Purge Port (x2)
Analyzer Quick Connect
Figure 58 - Alignment Mechanism
Alignment Nuts
(x8)
Analyzer Quick
Connect
WindowPurge
Port (x2)
Alignment Bellows
Alignment Studs
Analyzer Mounting Flange
(x4)
Alignment
Bellows
Analyzer
Mounting Flange
Process Flange
ProcessFlange
7.3 Data Reporting, Storage and Retrieval
The TDLS200 analyzer has been designed with extensive data reporting capabilities. All data is available in
the analyzer as a text le for import into a spreadsheet for analysis
Data stored in the analyzer:
• Results. Every measurement the gas concentration, transmission, diagnostic data are stored.
• Spectra. The analyzer records spectra at a timed interval, in the event of an analyzer warning or fault
(including concentration values) and manually via the user interface.
• Calibration History is stored during every calibration or validation event.
• Alarm Fault History
• Events History which includes any changes made to the system settings
All data can be retrieved using a USB flash drive (at the analyzer), via the RIU, or over an Ethernet connection.
Please refer to Section 10 for complete details of the available data
IM 11Y01B01-01E-A 6th Edition :Feb 13, 2013-00
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.