Yokogawa TC10 Specifications

01EN
8
TC S
-50 to +1760°C
-58 to +3200°F
Pt100
-200 to +850°C
-328 to +1562°F
Linear 12 to 60 mV
Linear 0 to 5 V
Linear 2 to 10 V
General Specification
GS 05C01E81-01EN
TC10 Temperature Controller
■ General
The TC10 is a compact single loop temperature controller with easy-to-read 3 dynamic colors led display. The short depth of the controller helps save instrument panel space. The TC10 supports easy configuration with codes for quick start-up.
■ Features
3 Dynamic Colors Led Display Compact size: 48 x 48 mm (1/16 DIN), depth 48 mm
+ 14 mm (terminals)
Universal Input: (TC, mV, V, mA, Pt100-Pt1000)  3 configurable alarms as absolute, deviation and
band
4 selectable Set Point Serial Communication (optional): RS-485 Modbus PID control with single or double action with
overshoot control, ON/OFF, ON/OFF Neutral Zone, Auto-tuning, Self-tuning
Standby mode of display (selectable) User calibration for sensor position compensation
Functional Specifications
Control Specifications Control Mode: On/Off heating, On/Off cooling, On/Off with neutral zone (H/C), PID heating, PID cooling, PID double action (H/C) Auto-tuning and Self-tuning algorithms, Overshoot control Alarm Functions Absolute high/low, Absolute inside/outside the band, Sensor break, Deviation high/low, Deviation inside/ outside the band. They are combined with Not active at power up, Latched, Acknowledgeable, and Not active at set point change for Deviation alarm. Digital Input Functions Alarm reset, Alarm acknowledge, Hold of the measured value, Stand-by mode, Manual mode, Heat with SP1 and Cool with SP2, Sequential set point selection, SP1/SP2 selection, Binary selection of the set point, Work in parallel with Up/down key. Communication Function Interface type: Isolated (50 V) RS-485 Protocol: Modbus RTU Baud rate: 1200, 2400, 9600, 19200, 38400 bps Byte format: 8bit with no parity, one stop bit. Instrument address: 1 to 254
Hardware Specifications
Display Specifications Main display: 4 digit height 15.5 mm, 3 color red, green and amber Secondary display: 4 digit height 7 mm, green color Display updating time: 500 ms Universal Input Specifications
TC J -50 to +1000°C -58 to +1832°F
TC K -50 to +1370°C -58 to +2498°F
(*)
TC R -50 to +1760°C -58 to +3200°F
TC T -70 to +400°C -94 to +752°F
Pt1000 -200 to +850°C -328 to +1562°F
Linear 0 to 60 mV
Linear 0 to 20 mA (this selection forces Out 4 = TX)
Linear 4 to 20 mA (this selection forces Out 4 = TX)
Linear 1 to 5 V
Linear 0 to 10 V
Sampling time: 130 ms Resolution: 30000 counts Total Accuracy: ±0.5% of F.S. ±1 digit *: ±1.0% of F.S. ±1 digit
Resistance-temperature detector (RTD) measured current; Pt100: 150 μA, Pt1000: 15.5 μA Response time: 2 second or less, 63% (10 - 90%) (The time required for t r ans mi s sion o utput to reach 63% of the maximum excursion when PV abruptly changes from 10% to 90%)
GS 05C01E81-
All Rights Reserved. Copyright © 2014, Y okogawa Electric Corporation
GS 05C01E81-01EN
6th Edition May. 2018
Output Specifications OUT 1: Relay SPST - NO 4A/250 Vac or voltage to
drive SSR 13V max. @1mA Analog output: 0/4 to 20 mA, galvanically isolated, RL max. 600Ω ±0.2% of F.S. or 0/2 to 10 V, galvanically isolated, RL min.: 500Ω ±
0.3% F.S.
OUT 2: Relay SPST -NO 2A/250 Vac or voltage to drive
SSR 13V max. @1mA, 10.5 min @15mA ±10%
OUT 3: Relay SPST -NO 2A/250 Vac or voltage to drive
SSR 13V max. @1mA, 10.5 min @15mA ±10%
OUT 4: programmable: voltage output to drive SSR 13V
max. @1mA, 10.5 min. @15mA ±10%, 12 VDC (20 mA) transmitter power supply or 2nd digital input
Note: Either control output or retransmission output can
be used for analog output.
Regulatory Compliance
CE marking, UL(USA/CANADA)
EMC Directive:
EN 61326-1 Class A, Table 2 (For use in industrial locations) EN 55011 Class A, Group 1 (During the test, the instrument continues to operate at the measurement accuracy within specification.)
LV Directive:
EN 61010-1, EN 61010-2-030 UL 61010-1 CSA 61010-1 Installation category: II Pollution category: 2
RoHS Directive:
EN 50581
Power Supply Specification and Isolation Voltage
24 VAC/DC (±10% of the nominal value)100 to 240 VAC (-15 to +10% of the nominal value)
Power consumption: 4.5 VA max. (24 VAC/DC)
6.0 VA max. (100 to 240 VAC)
Isolation Voltage
3000 V AC for 1 minute between primary and secondary terminals (Primary terminals = Power
(*)
and relay output terminals, Secondary terminals = Analog I/O signal terminals, contact input terminals, and communication terminals.) *: Power terminals for 24 V AC/DC models are the
secondary terminals.
Environmental Conditions
Normal Operating Conditions
Operating temperature: 0 to 50°C (32 to 122°F) Humidity: 20 to 90% RH, not condensing
Temperature Effects
Analog input: It is part of the global accuracy Reference junction compensation: ±0.1°C/°C or less Analog output: ±0.05% of F.S./°C or less
Storage temperature
Storage temperature: -20 to +70°C (-4 to +158°F) Humidity: 20 to 95% RH, not condensing