Model TB750G
Right Angle Scattered Light
Turbidimeter
Quick Start Manual
IM 12E01A06-02E
IM 12E01A06-02E
1st Edition
◆INTRODUCTION
■ Structure of this Manual
This is a supplementary manual to TB750G Right Angle Scattered Light Turbidimeter
for User's Manual IM 12E01A06-01E. This manual describes the specifications,
installation, operation, maintenance, and troubleshooting. To use this instrument
correctly, read this manual and User's Manual IM 12E01A06-01E thoroughly.
■ Before Measurement
The TB750G turbidimeter is preset with defaults prior factory shipment. Before
measurement, verify that these factory default settings meet the operating conditions and
if necessary, reconfigure parameters.
◆Symbol and Notations Used in this Manual
■ Symbol
The following symbol is used in this manual.
CAUTION
■ Notations
Indicates that damage to software or hardware, or system failures could occur if
instructions are not followed.
In this manual, operation keys, display messages, indicators, and LED lamps on the
instrument are denoted in the following manner.
1. Operation keys are denoted by brackets [ ] – for example:
YES
: [YES] key
2. Indicators, messages, and numerical data on the display are denoted as follows:
Message display: {*WASH}
Data display: {2.05} (when lit), {2.05} (when flashing)
●This manual should be read thoroughly before operating the instrument.
●This manual explains the functions contained in this product, but does not warrant
that they will suit the particular purpose of the user.
●The contents of this manual shall not be reproduced or copied, in part or in whole,
without permission.
●The contents of this manual are subject to change without prior notice.
●Every effort has been made to ensure the accuracy in the preparation of this manual.
However, if any errors or omissions are noticed, please contact the nearest Yokogawa
representative or sales office.
■ Protection, Safety, and Modification of the Product
●The safety instructions described in this manual should be strictly observed to ensure
safety both of the product and the system controlled by the product.
●A protection or safety circuit should be installed externally, if needed. Do not attempt
to modify or add such circuit to the inside of the equipment.
■ Limitation of Liability
●Yokogawa grants no warranties other than the express warranty set forth under the
warranty provisions.
●Yokogawa shall not be liable to you or any third party for any damage, including
consequential or incidental damages, arising out of or in connection with the use of
this equipment, defects beyond our knowledge, or any other contingency beyond our
control.
Revision Record .................................................................................................................... 1
IM 12E01A06-02E
3
Contents
4
IM 12E01A06-02E
1.OVERVIEW
1.1Features
The TB750G is a process turbidimeter employing the right angle light scattering method,
and has the following features.
• Highly reliable measurement with excellent linearity and repeatability
- Linearity: ±2% of reading or ±0.01 NTU, whichever is greater
- Repeatability: ±1% of reading or ±0.002 NTU, whichever is greater
- Display resolution: 0.001 NTU
• Easy-to-clean measurement cell
• Compact, lightweight converter and detector
• User configurable analog output range
- Analog output range: 0-0.2 NTU to 0-100 NTU
• Analog output range switching (2 or 3 ranges)
• Enhanced self-diagnostic function as standard
- Light source failure, input element failure, calibration failure, various circuit failures,
etc.
• Detector designed to remove influence of air bubbles
• A wide range of measurement conditions
- Low flow rate: 0.05 to 20 l/min {0.8 to 317 gal/h}
- High pressure: Up to 500 kPa {72 psi}
- Sample temperature: 0 to 50°C {32 to 122°F}
• Detector can be connected for in-line analysis
• 2 analog outputs, 3 relay contact outputs, and 1 serial communication
• Many options available upon request
• Ultrasonic transducer and oscillator for ultrasonic cleaning
• Various head tanks to accommodate application requirements
1. OVERVIEW
1.2Measurement Principle
The measuring system of the TB750G turbidimeter, comprised of a detector and a
converter, complys with USEPA 180.1.
Scatted Light Detection Element
Measuring
Water
IM 12E01A06-02E
LightSource
(Lamp)
Figure 1.1 Measurement Principle
Lens
Scatted Light Detector
Receiver (Darkroom)
MeasurementWindow (Glass)
F0101.EPS
5
2. PIPING AND WIRING
2.PIPING AND WIRING
2.1Piping
Piping methods depend on the application, select the appropriate system configuration,
and install pipes accordingly following examples.
2.1.1System Using Open Head Tank and Zero Turbidity Filter
This is a typical system. A piping diagram is shown in Figure 2.1. Air bubbles in a
water sample are removed by an open head tank and the water sample is introduced into
the detector at a stable flow rate.
Sample
Water
T ap
Water
HeadTank
Sample Water
Valv e
T ap W ater
Valv e
ZeroTurbidity
Filter
Drain
Valv e
Drain
Sample Water
Zero Water
Zero Water
Drain Valv e
H
Drain
Sample Water
Inlet
Sample Water
SupplyValv e
Zero Water SupplyValv e
FlowControl Valve
Sample Water
Outlet
TB750G
Detector
Other TurbidityDetectors
h
Drain Port
Drain Valv e
Drain
Sample Water
Effluent
Drain
F0320.EPS
Figure 2.1 Piping Diagram
• To prevent corrosion of water inside the zero turbidity filter, allow water to flow
through the filter and flow out from the zero water drain valve continuously at a flow
rate of approximately 10 ml/min.
• When using filtered water as zero reference, install a filter with the appropriate pore
size depending on the measuring range.
Measuring range of less than 2 NTU:0.2 µm
Measuring range of 2 NTU or greater:1 µm
6
IM 12E01A06-02E
2.1.2System without Head Tank and Zero Turbidity Filter
This is a simple system where a water sample is taken from the process and directly
introduced into the detector. This system configuration can be used when a water sample
contains a negligible amount of air bubbles or when the turbidity of a water sample is
high and the effect of air bubbles is nonsignificant.
CAUTION
Sample water conditions and ambient temperature must meet the specifications:
maximum pressure: 500 kPa; sample temperature: 0 to 50°C; ambient temperature: -5 to
50°C.
ProcessPiping
FlowControl Valve
Sample Water
Outlet
TB750G
Detector
2. PIPING AND WIRING
Sample Water
Effluent
h
Drain
Sample
Water
T ap W ater
Sample Water
Valv e
T ap W ater
Valv e
ZeroTurbidity
Filter
Sample Water
Zero Water
Zero Water
Drain Valv e
Sample Water
SupplyValv e
Zero Water SupplyValv e
Sample Water
Inlet
Other Turbidity Detectors
Drain Port
Drain Valv e
Drain
F0325.EPS
Figure 2.2 Piping Diagram
• The flow control should not be done on the inlet side of the detector.
• To prevent corrosion inside the zero turbidity filter, allow water to flow through the
filter and flow out from the zero water drain valve continuously at a flow rate of
approximately 10 ml/min.
(Note) Dotted wiring is externalwiring. Use cable with6 to 12 mm OD for wiring.*1Power terminal "G" on detector, detector case, and converter case must be grounded (ground resistance: 100V or less).*2Externalgrounding terminal of ultrasonic oscillator must be grounded (ground resistance: 100V or less). *3Use 2-conductor shielded cable for analog output wiringand serial communication wiring.*4 The wiring configuration is described below in case that RS-232Cserial communication isselected.
Serial com-
munication
(RS-232C)
U1
U2
U3
S
Dedicated cable (1/2/3m)
Grounding*
(100V or less)
5
*
Grounding*
(100V or less)
TXD
RXD
SG
G
1
Ultrasonic transducer
connecting cable
(maximum length:15m)
(customer wiring)
2
RS-232C
Serial commu-
Analog output 1
(4-20mA DC) *
Analog output 2
(4-20mA DC or
0-20mADC)*
3 *4
nication *
(RS-422)
Contact input
SENSOR
GG
RX+
RX-
RS-422
TX+
TX-SG
G
IN1
INPUT
IN2
COM
3
3
+
-
+
-
G
mA1
mA2
RANGE
OUTPUT
S1
S2
FAIL
NO
NC
COM
NO
NC
COM
NO
NC
COM
A
B
C
COM
Contact
output S1
Contact
output S2
Contact
output FAIL
Range
contact output
Grounding*
(100V or less)
1
*5 When option code "/US" isspecified, TUS400G should be purchased separately.
When TUS400G is used in system, the power supply to TB750G should be the same as the supply voltage specified in
the MSCode of TUS400G.
Figure 2.3 Connection Diagram
2.2.2Wiring for Detector
Wire for connections to the following terminals:
(1) Power supply terminals;
(2) Power supply terminals for ultrasonic oscillator (when option code "/US" is
specified); and,
(3) External grounding terminal.
F18.EPS
8
IM 12E01A06-02E
(1) Power Supply Terminals
2. PIPING AND WIRING
Terminals TM3 (Power to Lamp)
Terminal G
(For Converter 2Detector
ConnectingCable)
(2) Power Supply Terminals for
UltrasonicOscillator
Connector for Converter 2Detector Cable
Figure 2.4 Internal View of Detector
(3) External Grounding Terminal (M5 Screw)
(Norminal size of at least 2 mm
Figure 2.5 Grounding Terminal of Detector
F0310.EPS
2
)
F0319.EPS
2.2.3Wiring for Converter
Wire for connections to the following terminals:
(1) Analog output/contact input terminals;
(2) Contact output/range contact output terminals;
(3) Serial communication terminals (when in use); and,
(4) Grounding terminal.
IM 12E01A06-02E
9
Loading...
+ 22 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.