Thank you for purchasing the YOKOGAWA DL1540C/DL1540CL Digital
Oscilloscope.
This User’s Manual contains useful infor mation about the functions and operating
procedures of the instrument, mainly DL1540C. It also contains precautions that
should be observed during use. Main differences in the specifications between
DL1540C and DL1540CL is that DL1540CL allows record length of up to 2M words.
Due to these differences, functions and operations differ slightly among these
models. For a detailed description, refer to the relevant sections of this manual. To
ensure correct use of the instrument, please read this manual thoroughly before
operating it.
Keep the manual in a safe place for quick reference whenever a question arises.
Two manuals are provided with the instrument in addition to this manual.
DL1540C/1540CLIM701530-11EDescribes the communications functions
Communication Interfacefor the GP-IB/RS-232-C interface.
User’s Manual
• The contents of this manual are subject to change without prior notice as a result
of improvements in the instrument’s performance and functions. Display contents
illustrated in this manual may differ slightly from what actually appears on your
screen.
• Every effort has been made in the preparation of this manual to ensure the
accuracy of its contents. However, should you have any questions or find any
errors, please contact your nearest YOKOGAWA representative as listed on the
back cover of this manual.
• Copying or reproduction of all or any part of the contents of this manual without
YOKOGAWA’s permission is strictly prohibited.
Trademarks
MS-DOS is a registered trademark of Microsoft Corporation.
HP-GL is a registered trademark of Hewlett-Packard Company.
PostScript is a registered trademark of Adobe Systems Incorporated.
Unpack the box and check the contents before operating the instrument. If the
wrong instrument or accessories have been delivered, if some accessories are
missing or if they appear abnormal, contact the dealer from which you purchased
them.
Main Body
Check that the model name and suffix code given on the name plate of the rear
panel match those on your order.
MODEL
SUFFIX
NO.
MODEL (Type Name)
701530 : DL1540C, 701540 : DL1540CL
Made in Japan
SUFFIX (Suffix Code)
Suffix CodeSpecifications
Power voltage100-120 VAC/200-240 VAC
Power cord -DUL/CSA Standards Power Cord (Part No.: A1006WD)
*1 /F5 includes the enhanced trigger function for the DL1540CL.
*2 /V1 and /C8 cannot be specified simultaneously.
Example : DL1540C, UL/CSA standard power cord and full options → 701530-D/B5/E1/
F1/V1
[Maximum Rated Voltage: 125 V, Maximum Rated Current: 7 A]
[Maximum Rated Voltage: 250 V, Maximum Rated Current: 10 A]
[Maximum Rated Voltage: 250 V, Maximum Rated Current: 10 A]
[Maximum Rated Voltage: 240 V, Maximum Rated Current: 10 A]
/E1 With two additional 150 MHz probes (700998)
/F1 Additional trigger (OR, Pattern, Width)
/F5*1I2C-bus analysis function
/C8*22.1-GB internal HDD
/V1*2VGA output
NO. (Instrument No.)
When contacting the dealer from which you purchased your instrument, please
quote the instrument No.
2
IM 701530-01E
Standard Accessories
Checking the Contents of the Package
The following standard accessories are supplied with the instrument. Make sure
that all items are present and undamaged.
Power cord (one of the following power cords
is supplied according to the instrument's suffix codes)
UL/CSA standard
A1006WD
D
150MHz probe
(10:1, 1:1 selectable)(x2)
700998
Power supply fuse
(in fuse holder)
A1351EF
VDE standard
A1009WD
Printer roll
∗
1
chart
B9850NX
Front cover
B9957DG
BS standard
A1054WD
F
∗
2
User's manual (this manual)
Communication interface manual
Operation guide
Q
Rubber feet(x4)
A9088ZMx2
SAA standard
A1024WD
R
Soft case
B9918EZ
(x1)
(x1)
(x1)
IM 701530-01E
*1 For DL1540C/DL1540CL, a total of four passive probes are supplied (if the two
optional probes are included).
*2 A roll chart will be supplied only when the instrument is equipped with a built-in printer.
3
Checking the Contents of the Package
Optional Accessories
The following optional accessories are available. On receiving these optional
accessories, make sure that all the items that you ordered have been supplied and
that they are undamaged.
If you have any questions regarding optional accessories, or if you wish to place an
order, contact the dealer from whom you purchased the instrument.
Optional Spare Parts
Mini-clip convertor
B9852CR
Carrying case
700915
BNC adapter
B9852CS
Front cover (clear type)
700917
Ground lead
B9852CT
50 Ω Through
terminator
700976
The following optional spare parts are available. On receiving these optional spare
parts, make sure that all the items that you ordered have been supplied and that
they are undamaged.
If you have any questions regarding optional spare parts, or if you wish to place an
order, contact the dealer from whom you purchased the instrument.
Roll chartB9850NX 5Thermo-sensible paper, Total length: 30 m
Power supply fuseA1351EF 2Time lag, 3.15 A, 250 V
Note
It is recommended that the packing box be kept in a safe place. The box can be used when you
need to transport the instrument somewhere.
4
IM 701530-01E
Safety Precautions
This instrument is an IEC safety class I instrument (provided with terminal for
protective grounding). The following general safety precautions must be observed
during all phases of operation, service, and repair of this instrument. If this
instrument is used in a manner not specified in this manual, the protection provided
by this instrument may be impaired. Also, YOKOGAWA Electric Corporation
assumes no liability for the customer’s failure to comply with these requirements.
The following symbols are used on this instrument.
To avoid injury, death of personnel or damage to the instrument, the operator must
refer to an explanation in the User’s Manual or Service manual.
Function Grounding Terminal (This terminals shold not be used as a “Protective
grounding terminal”.)
Alternating current
ON(power)
OFF(power)
In-position of a bistable push control
Out-position of a bistable push control
IM 701530-01E
5
Safety Precaustions
WARNING
Power Supply
Ensure the source voltage matches the voltage of the power supply before turning
on the power.
Power Cord and Plug
To prevent an electric shock or fire, be sure to use the power supply cord supplied
by YOKOGAWA. The main power plug must be plugged in an outlet with protective
grounding terminal. Do not invalidate protection by using an extension cord without
protective grounding.
Protective Grounding
Make sure to connect the protective grounding to prevent an electric shock before
turning on the power.
Necessity of Protective Grounding
Never cut off the internal or external protective grounding wire or disconnect the
wiring of protective grounding terminal. Doing so poses a potential shock hazard.
Defect of Protective Grounding and Fuse
Do not operate the instrument when protective grounding or fuse might be
defective.
Fuse
To prevent a fire, make sure to use fuses with specified standard (current, voltage,
type). Before replacing the fuses, turn off the power and disconnect the power
source. Do not use a different fuse or short-circuit the fuse holder.
Do not Operate in an Explosive Atmosphere
Do not operate the instrument in the presence of flammable liquids or vapors.
Operation of any electrical instrument in such an environment constitutes a safety
hazard.
Do not Remove any Covers
There are some areas with high voltage. Do not remove any cover if the power
supply is connected. The cover should be removed by qualified personnel only.
External Connection
To ground securely, connect the protective grounding before connecting to
measurement or control unit.
6
IM 701530-01E
How to Use this Manual
Structure of the Manual
This User’s Manual consists of 16 chapters, an Appendix and an Index as described
below.
Chapter TitleDescription
1FunctionsDescribes measurement principles and functions. Although
2Name and Use ofGives the name of each part and describes how to use it.
Each PartThe corresponding reference pages for panel keys and the
3Before StartingDescribes points to watch during use and describes how to
Observation andinstall the instrument, connect the power cord, turn the
Measurement ofpower switch ON/OFF and connect probes as well as how to
Waveformsset the date and time.
4CommonDescribes the operations used to make settings, such as
Operationsstart/stop acquisition, auto set-up and initialization of each
5Setting the VerticalDescribes how to turn channels ON/OFF, how to make
and Horizontalsettings, such as input coupling, probe attenuation and
Axesvertical sensitivity, that are related to the ver tical axis
6Activating a Trigger Describes trigger setting operations, such as those for
7ChangingDescribes how to make acquisition condition settings, such
Acquisitionas acquisition mode, sampling mode, input filter and
and Displayhysteresis, as well as how to make display condition
Conditionssettings, such as waveform expansion, interpolation,
8AnalyzingDescribes how to measure waveforms using cursors,
Waveformsperformautomatic measurement of waveform parameters
9Performing aDescribes how to perform a GO/NO-GO action by setting the
GO/NO-GO Actionwaveform zone or setting limits for waveform parameters,
10Outputting Displayed Describes how to output displayed waveforms (screen
Waveforms andhardcopy) and set-up information to the built-in printer or
Setting Parametersplotter connected to the GP-IB interface.
to the Printer/Plotter
11Storing and Recalling Describes how to store displayed waveforms or settings in the
Data from theinternal memory, ho w to recall stored wa v eform data and display
Internal Memorythe waveforms, and how to recall settings and change them.
12Saving andDescribes how to save waveform data or set-up data to a
Loading Datafloppy disk, how to load saved data, how to initialize a floppy
from a Floppy Diskdisk and how to delete data.
13Saving andDescribes how to save the waveform data or set-up data to
Loading Data fromthe SCSI device or the internal hard disk. Describes how to
a SCSI Device orload the saved data, initialize the disk, delete the data, and
the Internal Hard Disk other operations.
14Other OperationsDescribes operations such as those required for trigger output.
15Troubleshooting,Describes how to diagnose the cause of trouble and how to
Maintenance andcorrect it, as well as describing each message that may
Inspectionappear on the screen and how to perform a self test.
16SpecificationsDescribes the specifications.
Appendix
IndexImportant terms used in this manual with corresponding
this section does not give a description of the operating
methods, it will help you to understand the basic operations.
rotary knobs are also given.
setting as well as operations, such as snapshot, clear trace
and calibration, that are used frequently during operation.
(voltage axis), as well as how to make settings related to the
horizontal axis (time axis).
setting trigger mode, trigger type, trigger source and trigger
level, that enable you to determine acquisition timing.
accumulation, X-Y display and graticule.
and waveform math, and how to display the power spectrum
obtained by FFT computation.
then detecting whether the input signal waveform is within
the specified range (zone or limits).
Shows a menu map, describes the relationship between the time
axis, sample rate and record length, and describes the data
format which must be specified when waveform data is saved.
reference page numbers are given.
IM 701530-01E
7
Conventions Used in this Manual
Unit
k ........... Denotes “1000”. Example: 100kS/s
K ........... Denotes “1024” or “1002”.
• Example of when K indicates “1024” : 640 KB (storage capacity of a
floppy disk)
• Example of when K indicates “1002” : 100 KW (record length for
acquisition memory)
Used Characters
Alphanumerics enclosed in double quotation marks usually refer to characters and
set values that appear on the screen and panel.
The SHIFT + xxx key refers to first pressing the SHIFT key (the indicator above the
SHIFT key lights), and then pressing the xxx key to obtain another, specified,
function.
Symbols Used
The following symbol marks are used in this User's Manual.
To avoid injury or death of personnel, or damage to the
instrument, the operator must refer to the User's Manual. In
the User's Manual, these symbols appear on the pages to
which the operator must refer.
WARNING
CAUTION
NoteProvides information that is important for proper operation
Symbol Marks Used for Descriptions of Operations
The following symbol marks are used in Chapters 3 to 14 to distinguish certain
feg|ures in descriptions.
Describes precautions that should be observed to prevent
the danger of serious injury or death to the user.
Describes precautions that should be observed to prevent
the danger of minor or moderate injury to the user, or the
damage to the property.
of the instrument.
Relevant KeysIndicates the relevant panel key which are
necessary to carry out the operation.
Operating ProcedureCarr y out steps in the order shown. The
operating procedures are given with the
assumption that you are not familiar with the
operation. Thus, it may not be necessary to carry
out all the steps when changing settings.
Keys and ProcedureShows the above mentioned relevant keys and
the operating procedure in one block.
ExplanationDescribes settings and restrictions relating to the
operation. A detailed description of the function
is not provided. For a detailed description of the
function, refer to Chapter 1.
Appendix 4 How to Calculate the Area of a Waveform ............................................................App-11
Appendix 5 Output Data Format used when Saving Waveform Data (Example DL1540CL) ..App-12
12
IM 701530-01E
Chapter 1Functions
1.1System Configuration and Block Diagram
System Configuration
1
Functions
Block Diagram
CH1
CH2
CH3
CH4
External trigger input/
External clock input
HP-GL plotter
Hard copy
Set-up information list
YOKOGAWA AG series
arbitrary waveform generator
Arbitrary waveform output
Personal computer
Remote control
Data analysis
Vertical control circuit
Input coupling
circuit
Data transfer
Pre-
ATT
amplifier
Front end
control circuit
External printer
Hard copy
Communication
SCSI device
Saving / loading data
Multiplex circuit
A/D
Timebase
Trigger
circuit
Object to be measured
Data
processing circuit
Acquisition
control circuit
Built-in printer
Hard copy
Set-up information list
GO/NO-GO judgement output
Trigger output
RGB video signal output
External clock input
/ trigger input
Signal input
Acquisition memory
Display
buffer
Display
control
circuit
CPU
Floppy disk
Saving / loading data
Color LCD display
Keyboard
Printer
GP-IB
FDD
Half pitch interface
connector
(included the output
for GO/NO-GO
judgement)
Trigger output
(Optional)
IM 701530-01E
Signal Flow
A signal input via the input terminal is sent to the vertical control circuit. The vertical
control circuit consists of an input coupling circuit, attenuator (ATT) and preamplifier, and it adjusts the voltage and amplitude of the signal according to the
input coupling, probe attenuation, vertical sensitivity and
offset voltage settings, before sending the signal to the multiplex circuit. The signal
will then be sent to the A/D converter matching settings such as time axis and so
on. The A/D converter samples the signal at a sample rate of 100 MS/s to convert it
to digital data. The digital data is then stored in the acquisition memory by the data
processing circuit at a sample rate which matches the time axis setting.
The data sent out from the acquisition memory by the data processing circuit is then
processed according to the invert waveform and averaging settings, then subjected
to interpolation and P-P compression (to obtain the maximum/minimum values
within each specified intervals), then converted into display data by the display
control circuit and finally memorized in the wavefor m display buffer.
The data read from the waveform display buffer and non-waveform display data
read from the graphic display buffer are combined, and the combined image is
displayed on the color LCD display.
1-1
1.2Setting the Vertical and Horizontal Axes
Vertical
control
circuit
Input terminal
Input coupling ≡ page 5-2. ≡
When you only want to observe the amplitude of an alternating current signal, or
when you just want to observe the signal relative to a given reference voltage,
eliminating the direct current components from the input signal makes observation
easier. You may also want to check the ground level or observe the input signal
waveform with the offset voltage removed. In this case, you can change the input
coupling setting. This will switch the coupling method, which determines how the
input signal is input to the vertical control circuit (voltage axis).
The input coupling method can be chosen from the following.
AC
The input signal is sent through a capacitor to the
attenuator in the vertical control circuit. This method
can be used when you just want to observe the
amplitude of the alternating current signal,
eliminating the DC components from the input
signal, or when you want to observe the signal
relative to a given reference voltage.
DC
The input signal is sent directly to the attenuator in
the vertical control circuit.
This method can be used when you want to observe
both the DC and AC components of the vertical
input signal. This method also enables observation
of the input signal with the offset voltage (DC
voltage) eliminated. This function is useful when you
want to observe ripples in a DC output signal.
Input terminal
Vertical
control
circuit
GND
The ground signal, not the input signal, is connected
to the attenuator in the vertical control circuit. This
method enables observation of the ground level on
the screen.
Probe Attenuation ≡ page 5-4. ≡
A probe is usually used to connect the circuit to be measured to an input terminal.
Use of a probe provides the following advantages.
• the voltage and current of the circuit to be measured are not disturbed;
• a signal can be input without distortion;
• the measurement voltage range of the oscilloscope can be widened.
A 150 MHz passive probe is supplied with the instrument. The probe attenuates the
input signal by 1/10. When a probe is used, the probe attenuation must match the
instrument’s attenuation setting so that the input voltage can be measured directly.
Set the instrument’s attenuation to 10:1 if the supplied probe is to be used.
In addition to 10:1, attenuations of 1:1, 100:1 and 1000:1 are provided. When you
use a probe other than the one supplied with the instrument, set the instrument’s
attenuation so that it matches the probe’s attenuation.
Input terminal
Vertical
control
circuit
1-2
IM 701530-01E
Inverted Wa veform Display ≡ page 5-5. ≡
1div=1V1div=500mV
When V/div is switched from 1 V/div to 500 mV/div
The voltage axis is inverted about the ground as shown below, and the wavefor m is
displayed. In other words, + voltage is converted to – voltage and – voltage to +
voltage.
Original waveform (not inverted)
Ground level
Vertical sensitivity ≡ page 5-6. ≡
The V/div (vertical sensitivity) setting is used to adjust the amplitude of the
displayed waveform so that the waveform can be observed easily.
The V/div setting is made by setting the voltage value per division on the screen
grid. The set-up value for the vertical sensitivity is displayed according to the
probe’s attenuation setting.
The V/div setting can be either “VAR” or “CAL”. In case of the “CAL” setting,
attenuators with different attenuation rates are used to change the ver tical
sensitivity, which then changes in steps of 1 V/div → 2 V/div → 5 V/div.
In case of the “VAR” setting, fine adjustment of the sensitivity set under the “CAL”
setting can be performed. The setting range is approximately 0.4 to 2.5 times the
value of the “CAL” setting. Digital data are taken from the “CAL” setting and after
calculating they will be displayed at the “VAR” setting. The measurement accuracy
of the voltage axis using the “VAR” setting however, is the same as the
measurement accuracy set at the “CAL” setting.
1.2 Setting the Vertical and Horizontal Axes
1
Functions
Inverted waveform
Note
Vertical sensitivity setting and measurement resolution
For accurate voltage measurement, set the V/div so that the maximum amplitude of the displayed
waveform is displayed using the full height of the screen (8 divisions).
The instrument uses an 8-bit A/D converter to sample the input signal with a resolution of 255
levels. When displaying a waveform, 25 levels are used for each division of the grid. Thus, in
normal display mode (not expansion display mode), the full range (255 levels) of the A/D converter
is equivalent to 10.24 div.
From this, it can be understood that increasing the V/div setting decreases the voltage per div
(25 levels), thereby increasing the display (measurement) resolution.
A/D converter
full range:
255 levels
IM 701530-01E
Display range: 200 levels
25 levels
1-3
1.2 Setting the Vertical and Horizontal Axes
Vertical Position of the Waveform ≡ page 5-7. ≡
Since a total of four input waveforms can be displayed, they may overlap each
other, making observation difficult. In this case, the wavefor ms can be moved in the
vertical direction so that they can be observed more easily.
Furthermore, full scale in the vertical direction is eight divisions of the grid as
mentioned earlier, and the voltage display range is eight times the vertical
sensitivity setting (voltage value per division). Thus, increasing the vertical
sensitivity results in a narrower voltage display range. However, this function allows
the ground level (0 V level) to be shifted up and down up to four divisions from the
center of the waveform frame, so that observation of waveforms is possible over the
entire voltage display range (V/div value x 16 divisions). This is because the input
voltage range of the A/D converter shifts due to the change in the frame position.
However, when acquisition is stopped and displayed waveforms are not updated,
the input voltage range does not shift even if the frame position is moved. As a
result, acquisition within a voltage display range exceeding 10.24 divisions (that is,
equivalent to the full range of the A/D converter) becomes impossible, thereby
resulting in a discontinuous waveform
as shown on the right.
Position range
= 16 divisions
Time Axis ≡ pages 5-8 and 5-10. ≡
Selection of the timebase
With the default settings, sampling timing is controlled by the clock signal output
from the timebase circuit of the instrument (refer to the Block Diagram, page 1-1).
The sampling timing can be controlled by an external clock signal instead of the
clock signal from the timebase circuit.
For DL1540C/DL1540CL, an external clock signal can be input to the CLOCK IN
terminal on the rear panel or to the CH4 input terminal. When an external clock is
input to the CLOCK IN terminal, the clock must be of TTL level, however up to four
waveforms can be observed at the same time. On the other hand, if an external
clock is input to the input terminal, the clock can be of the same signal level normal
input signals. This also allows you to check the clock signal waveform before
observing the desired waveform.
This external clock function is useful when you are observing a signal whose period varies or
when you are observing a waveform by synchronizing it with the clock signal to be
measured.
CH1
Display area
= 8 divisions
A/D
1-4
CH4
EXT CLOCK IN
Internal timebase
IM 701530-01E
1div=500µs
10div
1div=1ms
501points
0500
Sampling data
Time axis
Display record length
Display record length
Record length of
acquisition memory
P-P
compression
Voltage axis
1.2 Setting the Vertical and Horizontal Axes
Setting the time axis
When using the internal clock, set the time axis scale as a time duration per division
of the grid (T/div). The setting range is 2 ns/div to 50 s/div. The time range in which
waveform is displayed is “time axis setting x 10”, as the display range along the
horizontal axis is 10 divisions.
Note
Display of time axis direction
The sampled data is read into the acquisition memory, and a waveform is displayed based on
this data. The number of data stored into the acquisition memory differs depending on settings
such as time axis settings, trigger mode, and acquisition mode.
The number of display points in the time axis direction on a 10-div screen is 501 points (i.e. 50
points per division). Thus, the number of displayed points and record length might slightly differ
and this will be treated as follows (for more details on the relation between time axis, trigger
mode, acquisition mode, record length of acquisition memory and displayed record length,
refer to Appendix 2).
• When the record length is too long
The data is displayed in the time axis direction after first being compressed per one division.
Depending on the time axis settings, there will be times when all data of full record length of the
acquisition memory can be displayed, and times when the data can only be displayed partially.
The part which does not appear on your screen, can be viewed by moving the horizontal position
of the waveform (refer to page 1-7).
• When the record length is too short
The missing data will be interpolated (refer to page 1-15).
* Record length is expressed in units of words. One word refers to one point of sampled data.
For example, 10 KW means 10020 points. (K represents 1002.)
1
Functions
IM 701530-01E
Relationship between the time axis setting, sample rate and record length
If you change the time axis setting, the sample rate and the record length of the acquisition
memory changes too. This is because a linear relation exists between “record length of the
acquisition memory” and “time required for acquisition of all data x sample rate”. For more
detailed information, refer to Appendix 2/Appendix 3.
1-5
1.2 Setting the Vertical and Horizontal Axes
Relationship between time axis setting and sampling mode
The sampling method (sampling mode) for an input signal changes according to the time axis
setting as described hereafter.
For T/div settings in the range between 5 µs/div and 100 ns/div, it is possible to switch real-time
sampling area to repetitive sampling area. But, for DL1540CL, note that the time axis range
over which this feature is actually available will vary according to the maximum displayable
record length, as shown in Appendix 3.
• “50s/div to 100 ns/div” → Real-time sampling mode
When changing the time axis settings, the sample rate will change and sampling can be performed
at a sample rate of maximum 200 MS/s (or 100 MS/s when ch. 3 and 4 are also used). The input
signal is sampled sequentially, and data is stored in the acquisition memory.
In this mode, the waveform can only be displayed correctly at frequencies up to half the sample
rate, due to Nyquist’s theorem*. Sample rate is expressed in S/s (number of samples per second).
Thus, this mode is suitable for observation of a waveform which fluctuates more slowly than the
sample rate.
* If the sample rate is later than the frequency of the input signal, high frequency components
will be lost. In this case, a phenomenon in which high frequency components change to lower
frequency components occurs, due to Nyquist’s theorem. This phenomenon is called aliasing.
Aliasing signalInput signalSampling point
• 50 ns/div to 5 ns/div →Repetitive sampling mode
If T/div is set so that the sample rate exceeds 200 MS/s (or 100 MS/s), the sampling mode is
switched to the repetitive sampling mode automatically. In this mode, sampling is performed
only once for each cycle of a repetitive signal and several cycles are needed to form a waveform.
It appears as if the signal is sampled at a sample rate higher than the actual rate. An apparent
sample rate of up to 20 GS/s can be used. However, the maximum observable frequency is 150
MHz due to the instrument’s frequency characteristics. With DL1540CL, it is not possible to set
T/div if the maximum display record length is 400 KW or more.
There are two repetitive sampling methods: sequential sampling, in which a signal is sampled
sequentially at a fixed interval, and random sampling, in which a signal is sampled at random to
produce a waveform. This instrument uses a random sampling method which also enables
observation of the waveform up to the trigger point.
Time axis setting and roll mode display
If trigger mode (described on page 1-10) is set to auto-mode or auto-level mode, and T/div is
set between 50 ms/div and 50 s/div, the display will not be updated by trigger anymore (update
mode), but the mode will switch to roll mode when new data is acquired. In roll mode, the oldest
data is deleted, and the waveform shifts from right to left on the screen. A waveform can be
observed in the same way as it is recorded on a pen recorder. This mode is useful when you are
observing a signal which repeats or which fluctuates slowly. This mode is also useful when you
want to detect glitches (fast spikes on a waveform) which occur intermittently. With DL1540CL,
the time axis range for roll mode varies depending on the maximum display record length.
Refer to Appendix 3.
* Roll mode is also turned on when trigger mode is switched to single (short/long) mode, but
the displayed waveform stops when a trigger is activated since trigger setting is effective in
single (short/ long) mode.
1-6
IM 701530-01E
Displayed range before
moving (+5 div)
Displayed range
after moving to –5 div
Displayed range
after moving to 0 div
Center of the
waveform display
moves to 0 div
Center of the
waveform display
moves to –5 div
Display record length
Record length of acquisition memory
1.2 Setting the Vertical and Horizontal Axes
Horizontal Position of the Waveform ≡ page 5-12 ≡
In case you acquired more data than can be displayed on one (10 div) screen (i.e.
the record length* of the acquisition memory is longer than the record length of one
screen), then you can view the data by moving the display position in the time axis
direction. This might be the case when using the roll mode display or when the
trigger mode is set to single long mode (for DL1540C). The displayed waveform can
be moved over a span of +/- 5 div from the center of the screen.
* For more details on the relation between record length of acquisition memory and
displayed record length, refer to page 1-5, Appendix 2 and Appendix 3.
Moving the horizontal position
In roll mode, the display position can be moved after data acquisition is stopped. If
the instrument is in a mode other than roll mode, the display position can be moved
even during data acquisition. The following example shows how the display
position is moved in roll mode.
Note
Relationship between the display range on the screen and the display record length,
acquisition memory’s record length and trigger position (see page 1-11) is given below.
• Display range on the screen = Acquisition memory’s record length
The entire waveform data saved in the acquisition memory can be displayed without the
need for moving the horizontal position of the waveform.
Trigger position
1
Functions
Displayed range:10 div
(Displayed record length)
Record length of
acquisition memory
• Display range on the screen x 2 > Acquisition memory’s record length > Display range on the
screen
Waveform datas which are out of the display range on the screen are acquired in the
acquisition memory (up to 10 div) located before (on the left of) the trigger position. The
part which is out of the display range on the screen can be displayed by adjusting the
horizontal position of the waveform as described earlier.
Trigger position
Waveform data exceeding
the screen display range
Record length of
acquisition memory
Displayed range:10 div
(Displayed record length)
• Display range on the screen x 2 = Acquisition memory’s record length > Display range on the
screen
When the length of the waveform data acquired in the acquisition memory located before (on the left of) the
trigger position exceeds 10 div, waveform data will be acquired in the acquisition memory located after (on
the right of) the trigger position. Waveform data of up to 20 div is acquired. The part which is out of the
display range on the screen can be displayed by adjusting the horizontal position of the waveform as
described earlier.
Trigger position
Displayed range:10 div
(Displayed record length)
IM 701530-01E
10 div
10 div
Record length of acquisition memory
1-7
1.3Setting a Trigger
Trigger level
A trigger is activated at this point if
"Rise" ( ) is selected.
Trigger source
WIDTH
trigger is activated
[OUT]
trigger is activated
[IN]
Window
Window
trigger is activated
CENTER
WIDTH
CENTER
Trigger Type / Trigger Source / Trigg er Level ≡ pages 6-1 to 6-11. ≡
Trigger type: Selects the type of trigger. The following types are available as
described below: Edge trigger, TV trigger and window trigger.
Furthermore, the OR trigger, pattern trigger and width trigger are
available as an option.
Trigger source : Selects the signal for the selected trigger type.
Trigger level: Sets the voltage level used to judge trigger conditions such as
trigger slope (rise/fall of a signal) and trigger state (high/low
level).
Edge trigger → page 6-2.
The edge trigger is the simplest type of trigger and uses a single trigger source to
activate a trigger. A trigger is activated when the trigger source exceeds (rises
above) or drops (falls) below the preset trigger level*.
In addition to input signals (CH1 to CH4), the external trigger input signal and the
commercial power supply signal can be used as a trigger source.
* “A trigger is activated” refers to the condition in which trigger conditions are satisfied and a
waveform is displayed.
Window T rigger → page 6-4.
A certain voltage range (window) is set and a trigger is activated when the trigger
source level enters this voltage range (IN) or exits from this voltage range (OUT).
TV Trig ger → page 6-5.
The TV trigger is used when you are observing a video signal, and is compatible
with NTSC, PAL and HDTV broadcasting systems.
1-8
IM 701530-01E
1.3 Setting a Trigger
OR T rigger (option) → page 6-8.
Multiple trigger sources are selected, and a trigger is activated when one of the
trigger conditions set for each trigger source becomes true. Trigger conditions are
established by setting the trigger slope.
When CH1 = , CH2 =
CH1
CH2
1
Functions
Trigger
TriggerTrigger
Pattern trigger (option) → page 6-9.
Multiple trigger sources are selected, and a trigger is activated when all of the
trigger conditions set for each trigger source become true or false. Trigger
conditions are established by setting combinations of the state (High or Low) of
each trigger source.
Furthermore, one of the trigger sources can be used as the clock signal, and
triggering is synchronized with this clock signal.
Example: a trigger is activated
when CH1:L, CH2:L, CH3:L and CH4:L
CH1
CH2
CH3
CH4
CH1
LH L LL
CH2
CH3
CH4
L: Low level, H: High level
HHHHLLLL
H
HLL
Trigger is activated.
H
L
HLLH
IM 701530-01E
Width trigger (option) → page 6-11.
If the pulse width of a trigger source is narrower or wider than some specified time,
then a trigger is activated accordingly.
A trigger can also be activated by a single trigger source.
TWTWTW
Trigger is activated.
TW: Specified
pulse width
1-9
Half the amplitude
Half the amplitude
Trigger level Amplitude
1.3 Setting a Trigger
Trigger Mode ≡ page 6-13. ≡
Conditions for updating displayed waveforms are set. The following six types of
trigger mode are available.
Auto-mode
Displayed waveforms are updated each time a trigger is activated within a specified
time (approximately 100 ms, referred to as the time-out period) and are updated
automatically after each time-out period.
Auto-level mode
Waveforms are displayed in the same way as in Auto-mode if a trigger is activated
within the time-out period. If no trigger is activated, the center value of the
amplitude of the trigger source (page 1-7) is detected and the trigger level is
changed automatically to this center value, then a (edge) trigger is activated to
update the displayed waveforms.
Normal mode
Displayed waveforms are updated only when a trigger is activated. Displayed
waveforms will not be updated if no trigger is activated.
Single (short) mode
When a trigger is activated, displayed waveforms are updated only once, then
acquisition stops. This mode is useful when you are observing a single-shot signal.
(Called single-shot mode for DL1540C, and single mode for DL1540CL)
Single long mode
As with the single (short) mode, when a trigger is activated, displayed waveforms are
updated only once, then acquisition stops. However, the length of acquired data is
longer in this mode (up to 120 K words). This mode is available only with DL1540C.
Single (N) mode
This mode is useful when using the sequential store function (refer to page 1-14).
Waveforms are acquired and stored in different memory areas each time a trigger is
activated, then acquisition stops, and the waveforms are displayed. Acquisition is
performed the specified number of times. Acquired waveforms can be displayed
together, or they can be displayed individually. This mode is useful when you want
to detect a sudden abnormality in a waveform.
1st acquisition2nd acquisitionNth acquisition
Action-On T rigger ≡ page 6-15. ≡
The displayed waveform can be output to the optional built-in printer or saved to a
floppy disk each time a trigger is activated.
Trigger Coupling ≡ page 6-17. ≡
1-10
Input coupling can also be switched for trigger sources as it is for input signals.
Select the type of input coupling which is most suitable for the trigger source signal.
The following two types of input coupling are available for trigger source signals.
• DC : The trigger source signal is used as the trigger source without any process.
• AC : The trigger source signal is used as the trigger source after DC the content
has been removed from it. A trigger can always be activated if the trigger
level is set to 0 V as long as the signal’s amplitude is one division or more.
IM 701530-01E
HF Rejection ≡ page 6-17. ≡
Input signal
Repetitive period: T
Trigger level
Trigger source
signal
t
Trigger signal restricted by hold-off time "t" (when "Rise" is selected as the trigger slope)
Set HF rejection to ON when you want to remove high frequencies exceeding 15
kHz from the trigger source. This prevents a trigger from being activated
unexpectedly due to high frequency noise.
Trigger Position ≡ page 6-18. ≡
The trigger position indicates which position of the waveform in the acquisition memory
will actually be displayed on the screen. The trigger position can be moved on the
display from the center (0 div) to +/-4 div, which enables observation of the waveform up
to the trigger point (the pre-trigger section). You can also set the trigger position in the
range of +/-5div in steps of 1/50 div. The trigger point refers to the point at which a
trigger is activated. In case the trigger delay (to be explained here after) is set to 0s, the
trigger point and the trigger position refer to the same location.
Observation of the waveform up to the trigger position is made possible by
constantly storing sampled data in the acquisition memory (i.e. removing the oldest
data each time new data is acquired), retaining data in the memory and displaying it
on the screen each time a trigger is activated.
Trigger position
setting range
1.3 Setting a Trigger
1
Functions
Pre-trigger rangePost-trigger range
0 %100 %
Trigger Delay ≡ page 6-19. ≡
Normally, the waveform around the trigger point is displayed. However the trigger
delay function enables display of a waveform which has been acquired after a
specified time (called the delay time) has elapsed following activation of a trigger.
Trigger point
Trigger Hold-off ≡ page 6-20. ≡
The trigger hold-off function temporarily stops detection of the next trigger once a
trigger has been activated. For example, when observing a pulse train signal, such
as a PCM code, display of the waveform can be synchronized with repetitive cycles;
or when using the history memory function, you may want to change the repetitive
period, as shown below.
Trigger position
Acquisition memory record length
T (trigger position)
Delay time
IM 701530-01E
1-11
1.4Setting the Acquisition and Display Conditions
Acquisition Modes ≡ page 7-1. ≡
When storing sampled data in the acquisition memory, it is possible to perform
processing on specified data and display the resultant waveform. The following data
processing methods are available.
Normal mode
In this mode, sampled data is stored in the acquisition memory without processing.
Envelope mode
In normal mode and averaging mode, the sample rate (the number of times data is
acquired per second in the acquisition memory) drops if T/div is increased (refer to
Appendix 2/Appendix 3). In envelope mode, the maximum and minimum values are
obtained at each acquisition of the data (sampled at 100 MS/s), then the maximum
and minimum values are stored as a pair in the acquisition memory (one pair takes
up 2 words of memory).
Envelope mode is useful when you want to avoid aliasing (page 1-6), since the
sample rate is kept high irrespective of the time axis setting (T/div). Furthermore,
envelope mode is also useful when you want to detect glitches (pulsing signals
which rise very fast) or display an envelope of a modulating signal.
Averaging mode
Averaging is a process in which waveforms are acquired repeatedly to obtain the
average of waveform data of the same timing (the same time in relation to the
trigger point).
When the trigger mode is not single mode, exponential averaging is performed on
the waveform data using the following equation, then the resultant waveform is
stored in the acquisition memory and displayed on screen. The attenuation constant
can be set between 2 and 256 (in steps of 2
Exponential averaging
An = {(N–1)An–1+Xn}
An
Xn
N
This averaging process is useful when you want to eliminate random noise.
Sampling Mode ≡ page 7-1. ≡
As explained earlier in “Relationship between the time axis setting and sampling
mode” (page 1-6), data sampling can be performed either in real-time or in
repetitive sampling mode depending on the T/div setting (for DL1540CL, the T/div
and maximum displayable record length). If T/div is set between 50 ns/div and 5 ns/
div, data sampling will be performed in the repetitive sampling mode. (But, for
DL1540CL, note that these T/div settings are not available if maximum displayable
record length is set to a high value.) And, if T/div is set between 5 µs/div and 100
ns/div, data sampling can be performed both in real-time or repetitive sampling
mode and schould be selected beforehand. But, for DL1540CL, the range over
which this feature is actually available will again vary according to the maximum
displayable record length. For details, refer to Appendix 2/Appendix 3.
1
N
: Value obtained after nth averaging
: nth measured value
: Attenuation constant
Envelope
n
).
1-12
IM 701530-01E
VoltageVoltage
Time
Time
A
B
C
D
E
F
a
b
c
G
Input signal
Smoothing
data
1.4 Setting the Acquisition and Display Conditions
Record Length (Maximum Displayable Record Length) ≡ page 7-1. ≡ , (for DL1540CL)
The term “record length” refers to the number of data points (per channel) read into
acquisition memory. “Displayed record length” refers to the number of these data
points that actually appear on the screen. (Note that sampling rate and record
length will vary according to the T/div setting; see page 1-5). The term “maximum
displayable record length” refers to the maximum number of points that can be
displayed on the screen. This value can be selected by the user: the available
selections are 1 KW, 10 KW, 100 KW, 400 KW, 1 MW, and 2 MW.
In general, the acquisition-memory record length and the displayed record length
are equivalent. But if the maximum displayable record length is set to 1 KW, 10 KW,
or 100 KW and operation is in roll mode, then the acquisition acquisition -memory
record length will be double the maximum displayable record length. As an
example: if maximum displayable record length is set to 100 KW and the T/div is 50
ms/div, then the record length within acquisition memory will be 200 KW (200, 400
points). Only 100 KW of this length can appear on the screen at any given time. But
the waveform’s horizontal position can be shifted as necessary to enable viewing
over the entire 200 KW record (see page 1-7). High record-length settings are
useful for monitoring repetitive long-period signals and rapid sustained single-pulse
signals. But note that the three highest settings (400 KW, 1 MW, and 2 MW) involve
various operational restrictions. For details about these restrictions, refer to Section
7.1.
For details of the record length, refer to Appendix 2/Appendix 3.
1
Functions
Input Filter ≡ page 7-3. ≡
Smoothing
The smoothing function enables the instrument to ignore small fluctuations in
rapidly changing waveform data so that the overall characteristic of the data can be
understood. The function acts as a low-pass filter. A weighted moving average value
is obtained for each set of five points of sampled data, then the result is displayed
as a wavefor m. This process is useful when you want to eliminate low levels of
noise on input signals. Furthermore, smoothing is performed on the data stored in
the acquisition memory, thus it can still be performed after acquisition has stopped,
allowing the elimination of noise on single-shot signals.
Bandwidth limit
High frequency noise of 20 MHz or higher can be eliminated from the input signal.
Expanded Wavef orm ≡ page 7-4. ≡
IM 701530-01E
Waveforms can be expanded in the time axis direction. This function is useful when
you want to change the T/div setting after the waveform has been displayed in
single mode or when you want to extend the acquisition time to observe a particular
part of the waveform thoroughly.
As explained in “Display of the time axis direction” on page 1-5, the number of data
points stored in the acquisition memory in normal display mode is greater than the
number of display points (501 points) stored in the time axis direction. Hence, the
displayed waveform is P-P compressed in the time axis direction. In other words,
more than one waveform data value is displayed at same individual points.
1-13
expansion
(zoom) box
Pre-zoom waveform
expanded waveform
Display example in case N=100 times
Displaying all waveforms
(ALL)
Displaying newest waveform
(Display Number=0)
Displaying oldest waveform
(Display Number=–99)
1.4 Setting the Acquisition and Display Conditions
The zoom function described here reduces the compression ratio so as to enable
observation of individual data points from acquisition memory. The maximum
selectable magnification rate depends on the displayed record length: 10x zoom at
1 KW length; 100x at 10 KW; 1000x at 100 KW, 5000x at 400 KW; 10,000x at 1
MW; and 25,000x at 2 MW. Specifically, the waveform can be zoomed until the
number of displayed samples per division falls to 10 (until the displayed record
length for the full screen falls to 100 points). If you zoom in such that the number of
points per division is less than 50, an interpolation function will automatically supply
interpolation over the time axis direction, as described on the next page. Zoom
position is specified in units of grid division.
* Display record length of 400 KW, 1 MW and 2 MW is available on DL1540CL.
Display in zooming mode
As shown below, it is possible to select to display a pre-zoom waveform (MAIN) and
its expanded waveform (ZOOM) simultaneously, or to display the expanded
waveform only.
In the MainZoom display, a box is displayed in the display frame for the pre-zoom
waveform to indicate the position of the expanded waveform.
MainZoom displayZOOM display
Sequential Store ≡ page 7-7. ≡
In the real-time sampling mode, waveform data will be stored to the acquisition
memory only a set number of times, and all waveforms can be displayed. This stops
automatically after acquisition. This function is operable when the trigger mode is
set to single(N) mode. The maximum number of waveforms that can be stored into
acqsuisition memory under this mode varies according to model and the maximum
displayable record length in the range from 5 to 100 waveforms. The acquisitionmemory record length (the number of data points per waveform for a given channel)
is equivalent to the maximum displayable record length. Sequential-store operation
is not available at maximum displayable record lengths of 400KW or above.
Once the specified number of waveforms have been stored, you can display any of
the waveforms individually or all of them together, so that it is possible to derive a
time series of the wavefor m variation. The drawings below illustrate how stored data
can be displayed (assuming sequential storage of 100 waveforms).
1-14
IM 701530-01E
History Memory ≡ page 7-9. ≡
The oscilloscope automatically retains the last N waveforms recorded by real-time
sampling, where N is equivalent to the maximum number of waveforms that can be
stored by sequential-store. (The record-length for history storage is the same as
that for sequential-store.) The oscilloscope retains all waveforms for the first N
triggers; then, for each subsequent trigger, the oscilloscope deletes the oldest
stored wavefor m. You are free to switch the display from the current (newest)
waveform to any of other N-1 waveforms in the history. The illustration below shows
how data can be displayed, assuming N=100.
Saved waveform data of previous 100 triggers
1.4 Setting the Acquisition and Display Conditions
1
Functions
Current waveform display
(Display Number = 0)
Display Number 0
Any former waveform display
(display number is selectable in the range –1 to –99)
Display Number –25
Wide Screen ≡ page 7-11. ≡
The area for displaying the waveform can be switched from normal mode at 501
dots (horizontal) X 401 dots (vertical) to wide mode at 601 dots (horizontal) X 401
dots (vertical).
Display Interpolation ≡ page 7-12. ≡
This instrument displays waveforms at 50 dots* per division in both the vertical and
horizontal axis directions, as shown below. However, if the dots corresponding to
the sampled data are not contiguous, the instrument will interpolate between them.
The interpolation method can be selected from the following.
* Dots are also called pixels, and are the smallest units used to construct waveforms and
characters on the screen.
Line interpolation
Interpolates between two dots using a straight line.
Sine interpolation
Generates interpolation data using the function sin(X)/X then interpolates between
two dots using resulting sine curve. Sine interpolation is suitable for observation of
sine waves.
IM 701530-01E
1-15
1.4 Setting the Acquisition and Display Conditions
Pulse interpolation
Draws a horizontal line to the time axis position of the next data point, then interpolates
between the two dots by drawing a vertical line from the next data point.
No interpolation
Draws only a horizontal line to the time axis position of the next dat point.
Accumulated W aveform Display ≡ page 7-13. ≡
The time that old waveforms remain on the screen is made longer than the
waveform update interval so that new waveforms can be accumulated on top of old
waveforms. In addition, waveforms can be displayed in different colors depending
on the frequency of occurrence of the waveforms.
This function is useful when you want to observe jitters and temporary turbulence in
waveforms.
X-Y Wavef orm Display ≡ page 7-15. ≡
The horizontal axis (X-axis) is used as the voltage axis for the input signal to CH1
and the vertical axis (Y-axis) is used as the voltage axis for the input signal to CH2
for observing the relationship between the voltages between the two signals. The XY waveforms for CH1 and CH2 can be displayed simultaneously. In addition, they
can also be displayed simultaneously with a normal V-T waveform (a waveform
displayed using voltage and time axes).
Use of this X-Y waveform display function enables measurement of the phase angle
between two sine wave signals. For example, two X-Y sine waveforms are
displayed to obtain an X-Y waveform (called a Lissajous waveform), from which the
phase angle can be obtained.
Phase
angle 0°
Phase
angle 45°
Phase
angle 90°
Frequency
ratio
(X:Y)
Lissajous waveform
1:11:21:3
1-16
IM 701530-01E
1.4 Setting the Acquisition and Display Conditions
Other Display Modes Settings ≡ pages 7-16 to 7-18. ≡
Graticule/scale/% marker
Under the default settings, the graticule (including the grid and frame), scale or %
marker is displayed on the display, as shown below, enabling the easy observation
of wavefor ms. The type of graticule can be changed and scale display and %
marker can be turned ON/OFF.
Graticule: Grid, Scale: ON, % Marker: ON
Graticule: Frame, Scale: OFF, % Marker: OFF
1
Functions
Intensity
The intensity for items displayed on the LCD display such as characters, graticule
and so on can be adjusted. Adjust the intensity according to your needs.
IM 701530-01E
1-17
Loading...
+ 205 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.