AC/DC Medical Desktop
130 Watts
PMP130 Series
Specification
All specifications typical at nominal line, full load and 25°C
Input
Input Voltage • 85-130 VAC or 180 to 264 VAC
auto ranging
Input Frequency • 47 to 63 Hz
Input Current • 3.0 A (rms) for 115 VAC
1.7 A (rms) for 230 VAC
Safety Ground • 50 µA max at 110 VAC 60 Hz
Leakage Current • 100 µA max at 240 VAC 50 Hz
Output
Total Output Power • See Tables
Ripple and Noise • 1% pk-pk max
Overvoltage • Set at 112-132% of its nominal
Protection output voltage
Overcurrent • Output protected to short
Protection: circuit conditions
Temperature • ±0.04%/°C
Coefficient:
Transient Response • Maximum excursion of 4% or
better on all models. Recovers
to 1% of final value within
500 µs after a 25% step load
change
Remote Sense • Optional, please consult factory
XPiQ inc.
Intelligent Design Quality Product
•
General
Efficiency • 70% minimum
Hold-up time • 10 msec minimum at 110 VAC
Line Regulation • ±0.5% maximum at full load
Inrush Current • 15 A at 115 VAC or
Withstand Voltage • 4000 VAC from input to output
Insulation Resistance • 10 Mohm minimum from
Switching Frequency • 60 kHz to 75 kHz
Environmental
Operating • 0 °C to +70 °C, derate linearly from
Temperature 100% load at +50 °C to 50% load
Storage • -40 °C to +85 °C
Temperature
Relative Humidity • 5% to 95% non-condensing
MTBF • 100,000 hours minimum at full load
EMC & Safety
EMI Requirements • In compliance with radiated and
Safety Requirements • Approved to: (not for patient contact)
CE Marked - LVD & EMC
•
Universal Input
•
100% Burn-In
•
Low Leakage Current
•
4000 VAC Isolation
•
Input Surge Protection
•
Line & Neutral Fused
30 A at 230 VAC at +25°C cold start
1500 VAC from input to ground
500 VAC from output to ground
output to ground
at +70 °C
at +25 °C ambient
conducted limits of CISPR 11
(EN 55011) Class B
a) UL 2601
b) CSA C22.2 No. 601.1
c) IEC 601-1/EN60601-1
•
www.XPiQ.com
PH: 508 429.9883
FAX: 800 226.2100
Email: sales@xpiq.com
Holliston, MA 01746 USA
•
OUTPUT VOLTAGE & CURRENT RATINGS PMP130
Maximum Output Model
Output Power Vnom Imin Imax Tol. Number
130 W 12 V 0 A 10.8 A 5% PMP130-12
130 W 15 V 0 A 8.6 A 4% PMP130-13
130 W 18 V 0 A 7.2 A 4% PMP130-13-1
130 W 24 V 0 A 5.4 A 3% PMP130-14
130 W 30 V 0 A 4.3 A 3% PMP130-16
130 W 36 V 0 A 3.6 A 2% PMP130-17
130 W 48 V 0 A 2.7 A 2% PMP130-18
Pin Chart
Model 1 2 3 4 5 6 7 8
All Models Return Return Output Return Output Return Output Output
+V +V +V +V
Mechanical Details