AC/DC Medical Desktop
110 Watts
PMP110 Series
Specification
All specifications typical at nominal line, full load and 25°C
Input
Input Voltage • 85-264 VAC
Input Frequency • 47 to 440 Hz
Input Current • 3.20 A (rms) for 115 VAC
0.90 A (rms) for 230 VAC
Safety Ground • 50 µA max at 110 VAC 60 Hz
Leakage Current • 100 µA ma. at 240 VAC 50 Hz
Output
Total Output Power • 0-110 watts
Ripple and Noise • 1% pk-pk max
Overvoltage • Provided on output #1 only
Protection set at 112-132% of its nominal
output voltage
Overcurrent • All outputs protected to short
Protection: circuit conditions
Temperature • ±0.04%/°C
Coefficient:
Transient Response • Maximum excursion of 4% or
better on all models. Recovers
to 1% of final value within
500 µs after a 25% step load
change
XPiQ inc.
Intelligent Design Quality Product
•
General
Efficiency • 65% minimum
Hold-up time • 12 msec minimum at 115 VAC
Line Regulation • ±0.5% maximum at full load
Inrush Current • 15 A at 115 VAC or
Withstand Voltage • 4000 VAC from input to output
Environmental
Operating • 0 °C to +70 °C, derate linearly from
Temperature 100% load at +50 °C to 50% load
Storage • -40 °C to +85 °C
Temperature
Relative Humidity • 5% to 95% non-condensing
MTBF • 100,000 hours minimum at full load
EMC & Safety
EMI Requirement • Meets conducted and radiated
Safety Requirements • Approved to: (not for patient contact)
CE Marked - LVD & EMC
•
Universal Input
•
100% Burn-In
•
Low Leakage Current
•
4000 VAC Isolation
•
Input Surge Protection
•
Line & Neutral Fused
30 A at 230 VAC at +25°C
cold start
1500 VAC from input to ground
500 VAC from output to ground
at +70 °C
at +25 °C ambient
limits of EN 55011 (CISPR11) Class B
a) UL 2601
b) CSA C22.2 No. 601.1-M90 per CUL
c) IEC 601-1/EN60601-1
•
www.XPiQ.com
PH: 508 429.9883
FAX: 800 226.2100
Email: sales@xpiq.com
Holliston, MA 01746 USA
•
OUTPUT VOLTAGE & CURRENT RATINGS PMP110
Output #1 Output #2 Output #3 Output #4 Model
Vnom Imin Imax Tol. Vnom Imin Imax Ipeak Tol. Vnom Imin Imax Tol. Vnom Imin Imax Tol. Number
+5 V 0 A 10 A 5% +12 V 0 A 5 A 9.0 A 5% PMP110-23
+5 V 0 A 10 A 5% +12 V 0 A 5 A 9.0 A 5% -12 V 0 A 1 A 5% PMP110-31
+5 V 0 A 10 A 5% +15 V 0 A 4 A 7.5 A 5% -15 V 0 A 1 A 5% PMP110-32
+5 V 0 A 10 A 5% +12 V 0 A 5 A 9.0 A 5% -12 V 0 A 1 A 5% -5 V 0.0 A 1 A 5% PMP110-40
+5 V 0 A 10 A 5% +15 V 0 A 4 A 7.5 A 5% -15 V 0 A 1 A 5% +24 V 0.0 A 1 A 5% PMP110-41
+5 V 0 A 10 A 5% +12 V 0 A 5 A 9.0 A 5% -12 V 0 A 1 A 5% +12 V 0.0 A 1 A 5% PMP110-42
+5 V 0 A 10 A 5% +12 V 0 A 5 A 9.0 A 5% -12 V 0 A 1 A 5% +24 V 0.0 A 1 A 5% PMP110-45
+5 V 2 A 10 A 5% +12 V 0 A 5 A 9.0 A 5% –12 V 0 A 1 A 5% +24 V 1.5 A 3 A 10% PMP110-45-1
+5 V 0 A 10 A 5% +24 V 0 A 3 A 5.0 A 5% –12 V 0 A 1 A 5% +12 V 0.0 A 1 A 5% PMP110-45-2
+5 V 0 A 10 A 5% +15 V 0 A 4 A 7.5 A 5% –15 V 0 A 1 A 5% -5 V 0.0 A 1 A 5% PMP110-46
Note
Peak output current with 10% maximum duty cycle for less than 60 seconds. Total peak power must not exceed 130 Watts.
•
Pin Chart
Model 1 2 3 4 5 6 7 8
PMP110-23 Common N.C. Output#1 N.C. Output#2 Common Output#1 P.F.D.
PMP110-31 PMP110-32 Common Output#3 Output#1 N.C Output#2 Common Output#1 P.F.D.
PMP110-40 PMP110-41 Common Output#3 Output#1 Output#4 Output#2 Common Output#1 P.F.D.
PMP110-42 PMP110-45 Return Return
PMP110-45-1 PMP110-45-2
PMP110-46
Return Return
Return Return
Mechanical Details
DC FAN
48.0 (1219.0)
*
DC OUTPUT
8-PIN DIN PLUG
NOTE:
1. Dimensions shown in inches (mm).
2. Tolerance 0.02 (0.5) maximum.
3. Optional output connectors available.
4. Optional ON/OFF switch available.
5. Weight: 1.35 kg.
10.0 (254.0)
6.5 (165.1)
5.0 (127.01)
2.7 (68.58)
RUBBER FOOT
*AMP Receptacle No. 212047-1 or Equivalent
OPTIONAL
ON/OFF SWITCH
3.75 (95.3)
AC INPUT
IEC 320 INLET
www.XPiQ.com
PH: 508 429.9883
FAX: 800 226.2100
Email: sales@xpiq.com
Holliston, MA 01746 USA