Contact address in the event of queries:
Adolf Wolf SANOclav
Robert-Bosch-Straße 13
73337 BAD ÜBERKINGEN-HAUSEN
Tel. : (07334) 9623- 0
Fax.: (07334) 9623-12
Type of autoclave :
Serial number :
Year of manufacture :
max. Operating pressure :
SMALL-BATCH STERILIZER
Type KL and TKL-MCS
STAINLESS STEEL 18/10
max. Operating temperature :
Volume of pressure chamber :
Examined pressure :
Date of examination :
__________________________________________________
KL
TKL-MCS
OPERATING INSTRUCTIONS
HIGH-PRESSURE SMALL-BATCH STERILIZER
For sterilization in steam and water under pressure up
to 2 bar = 134 °C and up to 3 bar = 143°C
Edition of 07.01.2003
___________________________________________
6.2 Operating Test of Safety-Valve
Table 4: Safety-Valve
DECLARATION OF CONFORMITY „CE“
Start of the Test:
• Turn the knurled-screw
in counter- clockwise
direction
• After an evident audible crack stop turning
of the knurled-screw
• Now turn the knurledscrew in clockwisedirection back to start
position.
The audible crack while
turning the knurled-screw signals the functioning of the safetyvalve.
If during the test the crack is not audible, the safety-valve is
out of order.
In this case the safety-valve has to be replaced unconditional
before the next use of the sterilizer.
According to the EG-directives
89/336/EWG edition of. 93/31/EWG
73/23/EWG „Low Tension“
Type and models of products
Autoclave Type : TKL – MCS – 53
Autoclave Type : TKL – MCS – 73
Autoclave Type : TKL – MCS - 122
It is declared that the above mentioned products in conception and
type of construction, also in the mode of design traded by us, meet the
above directives.
This declaration is going to loose its validity in case of changes not
agreed by us.
In accordance to following international Regulations:
DIN EN 61010 – 1 ( VDE 0411 – 1 )
DIN EN 61010 – 2 – 041 ( VDE 0411 – 2 – 041 )
pr EN 61000 – 4 – 2 pr EN 61000 – 4 - 4
EN 61000 – 4 – 11 EN 55011 or. EN 55022
In accordance to national directives/technical specifications:
pr DIN 58951 – 2 IEC 801 - 5
„Electromagnectical Compatibility“
WARNING: A Cleaning or Repair of the safety-valve
can only be made by the manufacturer.
Any change or manipulation at the safety-valve, except the operating test, can be followed by a damage of the equipment or
in case to Explosion of the pressure chamber.
⇒ DANGER TO LIFE !
__________________________________________________
page 31
ADOLF WOLF SANOclav
Bad Überkingen , 07.01. 2003
Adolf Wolf .
Managing Director
2. Preheating and after-heating temperature
Preheating: If the media sensor has been activated while
the „F“ key (start key) has not been depressed, the autoclave is preheated to an
„operating readiness“ temperature of 60°C.
After-heating: After sterilization the autoclave is cooled
down to 60°C, when the media sensor
remains activated. This temperature is
maintained until the autoclave is switched off
so that nutrient media can be further
processed.
COPYRIGHT
Firma Wolf
Robert Bosch Straße 13
73337 Bad Überkingen-Hausen/Germany
owns the copyright to these Operating Instructions as of
01.09.96. All rights reserved.
Labor u. pharm. Geräte
5 Optional extras: Port for Documentation
The port for documentation of the sterilization cyclus only works
together with the MCS-processor.
It allows the transfer and documentation of sterilizationprocess
on an external PC.
6 Maintenance and Care
WARNING: It is imperative that maintenance and care
are completed according to the following table to prevent damage and injury and to achieve the required
sterilization result.
6.1 Inspection, Maintenance and
Cleaning
Wolf Labor u. pharm. Geräte does not give any guarantee for
the contents of these Operating Instructions, but particularly no
guarantee or liability for specific purposes.
No part of these Operating Instructions may be transmitted,
reproduced, revised or stored in any form, or translated into
another language. Exemptions from this prohibition can only be
given in writing by Wolf Labor u. pharm. Geräte.
Wolf Labor u. pharm. Geräte reserves the right to update and
correct the contents of these Operating Instructions at any time
without previous notification.
__________________________________________________
page 29
__________________________________________________