WIKA T24.10 Operating Instructions Manual

D
GB
Operating instructions Betriebsanleitung
Temperature transmitter, model T24.10
Temperature transmitter, model T24.10
Temperatur-Transmitter, Typ T24.10
2
2370242.09 08/2013 GB/D
WIKA operating instructions temperature transmitter, model T24
© 2011 WIKA Alexander Wiegand SE & Co. KG All rights reserved. / Alle Rechte vorbehalten. WIKA
®
is a registered trademark in various countries.
WIKA
®
ist eine geschützte Marke in verschiedenen Ländern.
Prior to starting any work, read the operating instructions! Keep for later use!
Vor Beginn aller Arbeiten Betriebsanleitung lesen! Zum späteren Gebrauch aufbewahren!
D
GB
Operating instructions model T24.10
Page 3 - 32
Betriebsanleitung Typ T24.10
Seite 33 - 62
Further languages can be found at www.wika.com.
GB
1. General information 4
2. Safety 6
3. Specications 11
4. Design and function 12
5. Transport, packaging and storage 12
6. Commissioning, operation 14
7. T24 conguration software WIKA_TT 21
8. Information on mounting and operation in hazardous areas (Europe) 23
9. Further country-specic approvals 27
10. Maintenance 28
11. Faults 28
12. Return and disposal 30
Appendix 1: FM installation drawing 31
Appendix 2: EC declaration of conformity 32
WIKA operating instructions temperature transmitter, model T24 3
2370242.09 08/2013 GB/D
Contents
Contents
Declarations of conformity can be found online at www.wika.com.
GB
4 WIKA operating instructions temperature transmitter, model T24
2370242.09 08/2013 GB/D
1. General information
The temperature transmitter described in the operating instructions has been manufactured using state-of-the-art technology. All components are subject to stringent quality and environmental
criteria during production. Our management systems are certied to
ISO 9001 and ISO 14001.
These operating instructions contain important information on handling the instrument. Working safely requires that all safety instructions and work instructions are observed.
Observe the local accident prevention regulations and general
safety regulations, in eect for the instrument's range of use.
The operating instructions are part of the product and must be kept in the immediate vicinity of the instrument and readily accessible to skilled personnel at any time.
Skilled personnel must have carefully read and understood the operating instructions prior to beginning any work.
The manufacturer‘s liability is void in the case of any damage caused by using the product contrary to its designated use, non-compliance with these operating instructions, assignment
of insuciently qualied skilled personnel or unauthorised modications to the instrument.
The general terms and conditions, contained in the sales documentation, shall apply.
Subject to technical modications.
Further information:
- Internet address: www.wika.de / www.wika.com
- Relevant data sheet: TE 24.01
- Application consultant:
Tel.: +49 9372 132-0 Fax: +49 9372 132-406
info@wika.de
1. General information
GB
WIKA operating instructions temperature transmitter, model T24 5
2370242.09 08/2013 GB/D
Explanation of symbols
WARNING!
... indicates a potentially dangerous situation that can result in serious injury or death, if not avoided.
CAUTION!
... indicates a potentially dangerous situation that can result in light injuries or damage to equipment or the environment, if not avoided.
Information
… points out useful tips, recommendations and informati-
on for ecient and trouble-free operation.
DANGER!
...identies hazards caused by electrical power. Should
the safety instructions not be observed, there is a risk of serious or fatal injury.
WARNING!
... indicates a potentially dangerous situation in a poten­tially explosive atmosphere, resulting in serious injury or death, if not avoided.
1. General information
GB
6 WIKA operating instructions temperature transmitter, model T24
2370242.09 08/2013 GB/D
2. Safety
WARNING!
Before installation, commissioning and operation, ensure that the appropriate temperature transmitter has been
selected in terms of measuring range, design and specic
measuring conditions. Non-observance can result in serious injury and/or damage to the equipment.
WARNING!
This is a Class A product. In a domestic environment this product may cause radio interference in which case the user may be required to take adequate measures. EMC directive 2004/108/EC, EN 61326 Emission (Group 1, Class A) and Immunity (industrial locations)
Further important safety instructions can be found in the individual chapters of these operating instructions.
2.1 Intended use
The temperature transmitter T24.10 is a congurable transmitter for
use with resistance thermometers (RTD) and resistance sensors.
The temperature transmitter has been designed and built solely for the intended use described here, and may only be used accordingly.
The technical specications contained in these operating instructions
must be observed. Improper handling or operation of the instrument
outside of its technical specications requires the instrument to be
taken out of service immediately and inspected by an authorised WIKA service engineer.
2. Safety
GB
WIKA operating instructions temperature transmitter, model T24 7
2370242.09 08/2013 GB/D
If the instrument is transported from a cold into a warm environment, the formation of condensation may result in instrument malfunction. Before putting it back into operation, wait for the instrument temperature and the room temperature to equalise.
The manufacturer shall not be liable for claims of any type based on operation contrary to the intended use.
2.2 Personnel qualication
WARNING! Risk of injury if qualication is insucient!
Improper handling can result in considerable injury and damage to equipment.
The activities described in these operating instructions may only be carried out by skilled personnel who have
the qualications described below.
Keep unqualied personnel away from hazardous
areas.
Skilled personnel
Skilled personnel are understood to be personnel who, based on their technical training, knowledge of measurement and control
technology and on their experience and knowledge of country-specic
regulations, current standards and directives, are capable of carrying out the work described and independently recognising potential hazards.
Special operating conditions require further appropriate knowledge, e.g. of aggressive media.
2. Safety
GB
8 WIKA operating instructions temperature transmitter, model T24
2370242.09 08/2013 GB/D
2.3 Additional safety instructions for instruments per ATEX
WARNING!
Non-observance of these instructions and their contents may result in the loss of explosion protection.
WARNING!
Observe the applicable regulations for the use of Ex-class instruments (e. g.: EN 60079-0: 2006, EN 60079-11: 2007, EN 60079-26: 2007, EN 60079-27: 2006 + 2008, EN 61241-0: 2006, EN 61241-11: 2006, EN 60079-15: 2005).
Do not use transmitters with any damage to the exterior!
Do not congure the transmitter, programming unit
(PU-448) and PC in hazardous areas!
2.4 Special hazards
WARNING!
Observe the information given in the applicable type
examination certicate and the relevant country­specic regulations for installation and use in potentially
explosive atmospheres (e.g. IEC 60079-14, NEC, CEC). Non-observance can result in serious injury and/or damage to the equipment.
For additional important safety instructions for instruments with ATEX approval, see chapter 8 “Information on mounting and operation in hazardous areas (Europe)”.
WARNING!
For hazardous media such as oxygen, acetylene,
ammable or toxic gases or liquids, and refrigeration
plants, compressors, etc., in addition to all standard regulations, the appropriate existing codes or regulations must also be followed.
2. Safety
GB
WIKA operating instructions temperature transmitter, model T24 9
2370242.09 08/2013 GB/D
WARNING!
Protection from electrostatic discharge (ESD) required. The proper use of grounded work surfaces and personal wrist straps is required when working with exposed circuitry (printed circuit boards), in order to prevent static discharge from damaging sensitive electronic components.
To ensure safe working on the instrument, the operating company must ensure
that suitable rst-aid equipment is available and aid is
provided whenever required,
that the operating personnel are regularly instructed
in all topics regarding work safety, rst aid and
environmental protection and knows the operating instructions and, in particular, the safety instructions contained therein.
DANGER!
Danger of death caused by electric current
Upon contact with live parts, there is a direct danger of death.
Electrical instruments may only be installed and connected by skilled electrical personnel.
Operation using a defective power supply unit (e.g. short circuit from the mains voltage to the output voltage) can result in life-threatening voltages at the instrument!
WARNING!
Residual media in dismounted instruments can result in a risk to persons, the environment and equipment. Take
sucient precautionary measures.
Do not use this instrument in safety or Emergency Stop devices. Incorrect use of the instrument result in injury.
2. Safety
GB
10 WIKA operating instructions temperature transmitter, model T24
2370242.09 08/2013 GB/D
Should a failure occur, aggressive media with extremely high temperature and under high pressure or vacuum may be present at the instrument.
2.5 Labelling / safety marks
Product label
Output signal
Power supply
Permissible ambient temperature
Pin assignment
Year of manufacture
Sensor, Pt100 or RTD
Model
Explanation of symbols
Before mounting and commissioning the instrument, ensure you read the operating instructions!
CE, Communauté Européenne
Instruments bearing this mark comply with the relevant European directives.
ATEX European Explosion Protection Directive
(Atmosphère = AT, explosible = EX) Instruments bearing this mark comply with the requirements of the European Directive 94/9/EC (ATEX) on explosion protection.
2. Safety
GB
WIKA operating instructions temperature transmitter, model T24 11
2370242.09 08/2013 GB/D
3. Specications
Specications Model T24.10
Ambient/storage temperature
Standard
Option
-40 ... +85 °C
-40 ... +105 °C
1)
Climate class per DIN EN 60654-1 Cx (-40 ... +85 °C, 5 ... 95 % r. h.) Max. permissible humidity
per DIN IEC 60068-2-30 var. 2
relative humidity 100 %,
condensation allowed Vibration per DIN EN 60068-2-6 10 ... 2000 Hz 10 g Shock DIN EN 60068-2-27 Salt fog DIN EN 60068-2-11
Housing material Plastic, PBT, glass bre
reinforced Ingress protection per IEC 60529/ EN 60529
Case
Terminals
IP 66/IP 67
IP 00
1) -40 ... +105 °C only without explosion protection
For further specications see WIKA data sheet TE 24.01 and the order
documentation.
For further important safety instructions for the operation in hazardous areas, see chapter 2.3 “Additional safety instructions for instruments per ATEX”.
3. Specications
GB
12 WIKA operating instructions temperature transmitter, model T24
2370242.09 08/2013 GB/D
4. Design and function
4.1 Description
The analogue temperature transmitter is used for converting a resistance value into a proportional current signal (4 ... 20 mA). The analogue signal is transmitted to a series-connected logic unit. For cable break monitoring the logic unit must be capable of recognising HI alarms (> 21 mA) and LO alarms (< 3.6 mA).
The electrical components of the transmitter are assembled in a plastic case and completely encapsulated. Due to the internal analogue signal processing, this transmitter can also be used in multi­plexed systems.
The temperature transmitter described in these operating instructions features a galvanic connection internally between the sensor input and the analogue output. There must not be any external conductive connection (e.g. via earth) between the connected sensor and the analogue output!
4.2 Scope of delivery
Cross-check scope of delivery with the delivery note.
5. Transport, packaging and storage
5.1 Transport
Check the instrument for any damage that may have been caused during transportation. Obvious damage must be reported immediately.
5.2 Packaging
Do not remove packaging until just before mounting. Keep the packaging as it will provide optimum protection during transport (e.g. change in installation site, sending for repair).
4. Design and function / 5. Transport, ...
GB
WIKA operating instructions temperature transmitter, model T24 13
2370242.09 08/2013 GB/D
5.3 Storage
Permissible conditions at the place of storage:
Storage temperature: -40 ... +85 °C
Humidity: 95 % relative humidity (condensation allowed)
Avoidance of exposure to the following factors:
Direct sunlight or proximity to hot objects
Mechanical vibration, mechanical shock (putting it down hard)
Soot, vapour, dust and corrosive gases
Potentially explosive environments, ammable atmospheres
Store the instrument in its original packaging in a location that fulls
the conditions listed above. If the original packaging is not available, pack and store the instrument as described below:
1. Wrap the instrument in an antistatic plastic lm.
2. Place the instrument, along with shock-absorbent material, in the packaging.
3. If stored for a prolonged period of time (more than 30 days), place a bag containing a desiccant inside the packaging.
WARNING!
Before storing the instrument (following operation), remove any residual media. This is of particular impor­tance if the medium is hazardous to health, e.g. caustic, toxic, carcinogenic, radioactive, etc.
5. Transport, packaging and storage
GB
14 WIKA operating instructions temperature transmitter, model T24
2370242.09 08/2013 GB/D
6. Commissioning, operation
In hazardous areas, only use temperature transmitters that are approved for those hazardous areas. The approval is marked on the product label.
6.1 Mounting
The model T24.10 transmitters are designed to be mounted on a measuring insert in a Form B DIN connection head. The connection wires of the measuring insert must be approx. 40 mm long and insulated.
Mounting example:
2476890.A1
Mounting on the measuring insert
Mount the transmitter on the circular plate of the measuring insert using two countersunk M3 screws per ISO 2009.
Check the screw length before xing the transmitter to the measuring
insert: Insert the screw into the circular plate and verify length of 4 mm!
6. Commissioning, operation
GB
2363122.01
WIKA operating instructions temperature transmitter, model T24 15
2370242.09 08/2013 GB/D
Mounting in connection head
Insert the measuring insert with the mounted transmitter into the protective sheath and secure into the connecting head using screws in pressure springs.
2476890.B1
6.2 Electrical connections
WARNING!
Observe the safety-related maximum values for the connection of the power supply and the sensors, see chapter 8.3 “Safety-related maximum values”.
When working on the transmitter (e.g. installation/removal, maintenance work) take measures to prevent electrostatic discharge from the terminals.
Input 4 ... 20 mA - loop
Recommended tools for terminal screws: Pozidriv screw-driver size 2 (ISO 8764). max. tightening torque 0.4 Nm
6. Commissioning, operation
GB
16 WIKA operating instructions temperature transmitter, model T24
2370242.09 08/2013 GB/D
6.2.1 Power supply / 4 ... 20 mA-loop
The T24 is a 2-wire temperature transmitter and depending on the version, it can be supplied with various types of power supply. Connect
the positive pole of the power supply to the terminal marked with ⊕
and the negative pole of the power supply to the terminal marked with
⊖. With exible leads we recommend the use of crimped connector
sleeves.
The T24 temperature transmitter requires a minimum terminal voltage of DC 10 V. The load must not be too high as, otherwise, in the case of relatively high currents, the terminal voltage at the transmitter will be too low.
Maximum permissible load depending on the supply voltage:
Load diagram
Load R
A
en Ω
Voltage UB en V
Ex ia Ex nL/nA/ic
6.2.2 Connecting the sensor
The connection of a resistance thermometer (e.g. per ICE 60751 or similar) into a 2-, 3- or 4-wire sensor connection. The sensor input of the
transmitter has to be congured in accordance with the actually used
sensor connection type, otherwise a complete use of the possibilities of connection compensation is not possible; furthermore this may cause
additional measuring errors (see chapter 6.3 “Conguration”).
6. Commissioning, operation
GB
WIKA operating instructions temperature transmitter, model T24 17
2370242.09 08/2013 GB/D
If a 2-wire conguration is used:
Set a jumper between the input terminals and . Otherwise the transmitter will signal a sensor burnout and set the output signal
< 3.6 mA or > 21.0 mA, dependent on conguration.
6.3 Conguration
Sensor type, sensor connection, measuring range, signalling as well
as further parameters can be congured (see data sheet TE 24.01).
The temperature transmitters are delivered with a basic conguration (Pt100, 3-wire, 0 ... +150 °C, downscale) or congured according to the customer‘s specications within the given conguration possibilities.
For customer-dened congurations, the input and measuring range
is given on the product label. The initial value of the measuring range is between -200 ... +200 °C (T24.10.2Px) or between -150 ... +150 °C (T24.10.1Px). The possible end value of the measuring range is dependent on the respective initial value of the range. For possible combinations of initial value of measuring range/end of measuring range see the diagram on page 19. For the purpose of clarity this dependency is represented in 50 °C steps in this diagram, though
intermediate values are also congurable. Possible combinations of
the lower and upper limits of the measuring range can be check easily
by the software WIKA_TT. Amendments to conguration should be noted on the product label with a waterproof bre-tipped pen.
To congure the T24 temperature transmitter, a simulation
of the input value is not required. A sensor simulation is only required for the functional test.
6. Commissioning, operation
GB
18 WIKA operating instructions temperature transmitter, model T24
2370242.09 08/2013 GB/D
Conguration via the PC
To congure the transmitter, the WIKA_TT conguration software and the PU-448 programming unit are also required. Therefore WIKA oers an optional conguration set (order no. 11606304), consisting of the
following components:
PU-448 programming unit to connect between a Windows PC and the transmitter
Quick connector model magWIK
For further important instructions for the programming unit see chapter 6.4 “Connection of the model PU-448
programming unit” and for the WIKA_TT conguration software see chapter 7 “T24 conguration software
WIKA_TT”.
6. Commissioning, operation
GB
WIKA operating instructions temperature transmitter, model T24 19
2370242.09 08/2013 GB/D
minimum maximum
-150 ... +380 -150 ... +850
-100 ... +180 -100 ... +850
-50 ... -20 -50 ... +850
-40 ... -20 -40 ... +850
0 ... +20 0 ... +850
+10 ... +30 +10 ... +850
+20 ... +50 +20 ... +850
+50 ... +230 +50 ... +850
+100 ... +530 +100 ... +850
+150 ... +830 +150 ... +850
-150 °C
-100 °C
-50 °C
-40 °C
0 °C
+10 °C
+20 °C
+50 °C
+100 °C
+150 °C
-200 -100 0 +100 +200 +300 +400 +500 +600 +700 +800
minimum maximum
-200 ... +350 -200 ... +850
-150 ... +150 -150 ... +850
-100 ... -50 -100 ... +850
-50 ... 0 -50 ... +850
0 ... +50 0 ... +850
+50 ... +100 +50 ... +850
+100 ... +150 +100 ... +850
+150 ... +450 +150 ... +850
+200 ... +750 +200 ... +850
-200 °C
-150 °C
-100 °C
-50 °C
0 °C
+50 °C
+100 °C
+150 °C
+200 °C
-200 -100 0 +100 +200 +300 +400 +500 +600 +700 +800
6. Commissioning, operation
Diagram for measuring ranges model T24.10.1Px
Measuring range in °C
Range of possible end of measuring range in °C
Start of
measuring range
Diagram for measuring ranges model T24.10.2Px
Measuring range in °C
Range of possible end of measuring range in °C
Start of
measuring range
GB
20 WIKA operating instructions temperature transmitter, model T24
2370242.09 08/2013 GB/D
6.4 Connection of the model PU-448 programming unit
WARNING!
Transmitter, programming unit and PC must be used
outside the hazardous area during conguration.
Tighten the screws of the terminals to avoid communication errors. The electrical connection of the terminals for the head-mounted versions is achieved using the patented magWIK quick connector.
During the conguration, no additional power supply is
needed - the power for the PU-448 and the transmitter is supplied from the PC, directly through the USB connection.
The wiring is made as shown in the diagram. The connection of the plus and minus terminals of the transmitter must be made to the red and black magWIK terminals respectively.
T24.10
Connection PU-448 ↔ temperature transmitter
(option: magWIK quick connector)
3214338.04
6. Commissioning, operation
Loading...
+ 44 hidden pages