WIKA P-30, P-31 Specifications

Page 1
Electronic pressure measurement
Pressure transmitter for precision measurements Model P-30, standard version
Model P-31, ush version
for further approvals see page 5
Applications
Measurement and test benches
Calibration technology
Laboratories
Plant construction and machine building
Special features
WIKA data sheet PE 81.54
Accuracy 0.1 %, without additional temperature error in a range of 50 ... 140 °F (10 ... 60 °C)
Optional accuracy of 0.05 % (full scale) available
Fast measuring rates up to 1 kHz
Analogue, USB and CANopen output signals available
On-site calibration possible using product software
Description
Precise
The model P-30 and P-31 pressure transmitters have been developed for precision measurements. Through the use of a special WIKA pressure sensor, precision measurements with a maximum measuring deviation of as low as 0.05 % of span are guaranteed. Due to their active temperature compensa­tion these pressure transmitters have no additional tempera­ture error in the range of 50 ... 140 °F (10 ... 60 ° C).
Fast
The high measuring and output rates of up to 1 kHz make the measured value available as quickly as possible.
Compact
The compact design makes the pressure transmitter ideal for mounting into test benches, such as 19" racks.
Fig. left: Pressure transmitter model P-30
Versatile
The models P-30 and P-31 oer a wide selection of electrical
connections, process connections, and measuring ranges, and output signals. In addition to the standard analog signals, USB and CANopen
Via a USB service interface and the WIKA conguration
software "EasyCom", the models P-30 and P-31 can quickly and easily be adjusted on site.
Thanks to the simple-to-use software "Wika data logger“, the USB version can also be used to save measured values and create customised reports.
versions are also available.
Page 1 of 8WIKA data sheet PE 81.54 ∙ 02/2016
Page 2
Measuring ranges
Relative pressure
bar Measuring range 0 ... 0.25 0 ... 0.4 0 ... 0.6 0 ... 1 0 ... 1.6 0 ... 2.5 0 ... 4
Overpressure limit 1.5 2.4 3.6 4 6.4 7.5 12
Measuring range 0 ... 6 0 ... 10 0 ... 16 0 ... 25 0 ... 40 0 ... 60 0 ... 100
Overpressure limit 18 30 48 75 80 120 200
Measuring range 0 ... 160 0 ... 250 0 ... 400 0 ... 600 0 ... 1,000
Overpressure limit 320 500 800 1,200 1,500
psi Measuring range 0 ... 5 0 ... 10 0 ... 15 0 ... 25 0 ... 30 0 ... 50 0 ... 100
Overpressure limit 20 40 45 75 90 150 300
Measuring range 0 ... 160 0 ... 200 0 ... 300 0 ... 500 0 ... 1,000 0 ... 1,500 0 ... 2,000
Overpressure limit 480 600 900 1,000 1,500 2,250 3,000
Measuring range 0 ... 3,000 0 ... 5,000 0 ... 10,000
Overpressure limit 4,500 7,500 15,000
1) not available for model P-31
Absolute pressure
bar Measuring range 0 ... 0.25
Overpressure limit 1.5 2.4 3.6 4 3.6 4.8 7.5
Measuring range 0 ... 4 0 ... 6 0 ... 10 0 ... 16 0 ... 25
Overpressure limit 12 18 30 48 48
psi Measuring range 0 ... 5 0 ... 10 0 ... 15 0 ... 25 0 ... 30 0 ... 50 0 ... 100
Overpressure limit 20 40 45 75 90 150 300
Measuring range 0 ... 160 0 ... 200 0 ... 300
Overpressure limit 480 600 600
2) only available with an accuracy of 0.1 % of spann
2)
0 ... 0.4 0 ... 0.6 0 ... 1 0.8 ... 1.2 2)0 ... 1.6 0 ... 2.5
1)
Vacuum and +/- measuring range
bar Measuring range -1 … 0 -0.6 … 0 -0.4 … 0 -0.25 … 0 -1 … +0.6
Overpressure limit 1.5 1.5 1.5 1.5 3.2
Measuring range -1 … +1 -1 … +1.5 -1 … +3 -1 … +5 -1 … +9
Overpressure limit 4 5 8 12 20
Measuring range -1 … +15
Overpressure limit 32
psi Measuring range -30 inHg ... 0 -30 inHg ... +15 -30 inHg ... +30 -30 inHg ... +50 -30 inHg ... +100
Overpressure limit 22.5 60 90 135 240
Measuring range -30 inHg ... +160 -30 inHg ... +200
Overpressure limit 360 450
The given measuring ranges are also available in mbar, kg/cm2 and MPa. Other measuring ranges on request
Vacuum resistance
Ye s
WIKA data sheet PE 81.54 ∙ 02/2016Page 2 of 8
Page 3
Output signal
Accuracy data
Signal type Signal
Current (2-wire) 4 ... 20 mA Current (3-wire) 4 ... 20 mA
0 ... 20 mA
Voltage (3-wire) DC 0 ... 10 V
DC 0 ... 5 V USB per P-30/P-31 interface protocol CANopen
per CiA DS404
Voltage supply
Power supply
The permissible power supply depends on the corresponding output signal.
4 ... 20 mA (2-wire): DC 9 ...30 V
4 ... 20 mA (3-wire): DC 9 ...30 V
0 ... 20 mA (3-wire): DC 9 ...30 V
DC 0 ... 5 V: DC 9 ...30 V
DC 0 ... 10 V: DC 14 ... 30 V
USB: DC 4,5 ... 5,5 V
CANopen
Total current consumption
The total current consumption is dependent on the respective signal type.
Current (2-wire): max. 25 mA
Current (3-wire): max. 45 mA
Voltage (3-wire): max. 10 mA
USB: 40 mA
CANopen
Load
Current (2-wire): ≤ (power supply - 9 V) / 0,02 A
Current (3-wire): ≤ (power supply - 9 V) / 0,02 A
Voltage (3-wire):
: DC 9 ...30 V
: 60 mA
> max. output signal / 1 mA
Accuracy at reference conditions
Accuracy
Standard ≤ ±0,1 % of span Option ≤ ±0,05 % of span
1) For +/- measuring ranges and measuring range ≤ 0.4 bar on request
1)
Including non-linearity, hysteresis, non-repeatability, zero
oset and end value deviation (corresponds to measured
error per IEC 61298-2). Calibrated in vertical mounting position with process connection facing downwards.
Non-linearity (per IEC 61298-2)
≤ ±0.04 % of span BFSL
Temperature error
In the range of -4 ... 176 °F (-20 ... +80 °C) the instrument is actively compensated.
-4 ... 50 °F (-20 ... +10 °C): ≤ ±0,2 % of span/10 K
50 ... 140 °F (10 ... 60 °C): no additional error
140 ... 176 °F (60 ... 80 °C): ≤ ±0,2 % of span/10 K
1) For the optional accuracy at reference conditions of ≤ ±0.05 % of span there is an ad­ditional temperature error of ≤ ±0.05 % of span.
1)
Total error band 50 ... 140 °F (10 ... 60 °C)
≤ ±0.1 % of span
Long-term stability
≤ ±0.1 % of span/year
Adjustability
Adjustment via the "EasyCom 2011" or "EasyCom CANopen®" software
Zero point: -5 ... +10 % of span Span: -50 ... +5 % of span
Measuring rate
The measuring rate is dependent on the respective signal type.
2-wire: 2 ms
3-wire 1 ms
USB 3 ms
CANopen
: 1 ms
WIKA data sheet PE 81.54 ∙ 02/2016
Page 3 of 8
Page 4
1
2
3
4
3
1
2
Reference conditions
Electrical connections
Temperature
59 ... 77 °F (15 ... 25 °C)
Atmospheric pressure
12. 5 ... 15.4 psi (860 ... 1,060 mbar)
Humidity
45 ... 75 % relative
Power supply
DC 24 V
DC 5 V with USB version
Warm-up time
< 10 min
Mounting position
Process connection lower mount (LM)
Operating conditions
Ingress protection (per IEC 60529)
The ingress protection depends on the type of electrical connection.
Angular connector DIN 175301-803 A: IP 65
Circular connector M12 x 1 (4-pin): IP 67
Circular connector M16 x 0.75 (5-pin): IP 67
Bayonet connector: IP 67
CANopen M12 x 1 (5-pin): IP 67
USB: IP 67
Cable outlet: IP 67
Short-circuit resistance
S+ vs. U-
CAN-High/CAN-Low vs. U+/U
-
Reverse polarity protection
U+ vs. U
-
Overvoltage protection
DC 36 V (not with USB version)
Insulation voltage
DC 500 V
Connection diagrams
Circular connector M12 x 1 (4-pin)
2-wire 3-wire
U
1 1
+
U
3 3
-
- 4
S
+
Angular connector DIN 175301-803 A
2-wire 3-wire
U
1 1
+
U
2 2
-
- 3
S
+
Circular connector M16 x 0.75 (5-pin)
2-wire 3-wire
U
3 3
+
U
1 4
-
- 1
S
+
The stated ingress protection only applies when plugged in using mating connectors that have the appropriate ingress protection.
Vibration resistance
10 g (IEC 60068-2-6, under resonance)
Shock resistance
200 g (IEC 60068-2-27, mechanical)
Service life
10 million load cycles
Free fall test
The instrument is resistant to an impact onto concrete from a height of 3.3 ft (1 m).
Temperatures
Ambient: -4 ... 176 °F (-20 ... +80 °C)
Medium: -4 ... 221 °F (-20 ... +105 °C)
Storage: -40 ... 185 °F (-40 ... +85 °C)
Bayonet connector
2-wire 3-wire
U
A A
+
U
B B
-
- C
S
+
Circular connector M12 x 1 (5-pin), CANopen
2-wire
U
+
U
-
Shield 1
CAN-High
CAN-Low
2
3
4
5
Cable outlet unshielded
2-wire 3-wire
U
brown brown
+
U
blue blue
-
- black
S
+
Cable lengths on request.
WIKA data sheet PE 81.54 ∙ 02/2016Page 4 of 8
Page 5
Process connections
CE conformity
Model P-30
Standard Thread size
EN 837 G ¼ B
G ¼ female
G ½ B DIN 3852-E G ¼ A ANSI/ASME B1.20.1 ¼ NPT
½ NPT
- M18 x 1.5 male with G ¼ female
- G ½ male with G ¼ female
Other connections on request
Model P-31
Standard Thread size
EN 837 G ½ B with ush diaphragm
G 1 B with ush diaphragm
Sealings
Thread size Standard Option
G ¼ B Without Cu
Stainless steel
G ½ B Without Cu
Stainless steel
G ¼ A Without NBR
FPM/FKM
Pressure equipment directive
97/23/EC, PS > 200 bar; module A, pressure accessory
EMC directive
2004/108/EC, EN 61326 emission (group 1, class B) and immunity (industrial application)
Approvals
GOST-R, import certicate, Russia
CRN, safety (e.g. electr. safety, overpressure, ...), Canada
Certicates
Accuracy test report (included in the delivery)
2.2 test report per EN 10204
3.1 inspection certicate per EN 10204
1) option
Approvals and certicates, see website
1)
1)
For all other process connections no sealings are available.
Materials
Wetted parts
Stainless steel
Additionally Elgiloy® for measuring ranges > 25 bar
For sealing materials see "Process connections"
Non-wetted parts
Stainless steel
Manufacturer’s declaration
RoHS conformity
Yes, instruments with bayonet connector are not RoHS-compliant
WIKA data sheet PE 81.54 ∙ 02/2016
Page 5 of 8
Page 6
~
~
Dimensions in mm
Pressure transmitters
with M12 x 1 circular connector with M16 x 0.75 circular connector
with angular connector DIN 175301-803 form A
with cable outlet
with bayonet connector
with USB connector type A
~
~
WIKA data sheet PE 81.54 ∙ 02/2016Page 6 of 8
Page 7
Process connections for model P-30
G L1
G ¼ A DIN 3852-E 12
G L1 L2 L3 D1
G ¼
20 13 10 Ø 25
Process connections for model P-31
G L1
G ¼ B EN 837 13 G ½ B EN 837 20
G1 G2 L1 L2 L3
G ½ B G ¼ 20 15,5 13
G L1
¼ NPT 13 ½ NPT 19
G1 G2 L1 L2 L3
M18 x 1,5 G ¼ 20 12 9
G L1 L2 L3 D1
G ½ B
23 20,5 10 Ø 18
G1 L1 L2 L3 D1
G 1 B 23 20,5 10 30
Accessories
CANopen design
Designation Order no.
Y-connector (M12 x 1 female connector, male/female connector)
Terminating resistor (120 Ω, M12 x 1 connector)
Bus cable 0.5 m (M12 x 1 male/female connector) Bus cable 2 m (M12 x 1 male/female connector) Software EasyCom CANopen, incl. PCAN-USB adapter, cable set and power supply P-30/P-31 software CD
Analogue design
Designation Order no.
P-30/P-31 USB service interface, incl. WIKA software CD
Software
The full software is available to download as freeware from the following path. www.wika.com / Download / Software / Electronic Pressure Measurement
2344526
2308274
2308240
2308258
7483167
11478901
13193075
WIKA data sheet PE 81.54 ∙ 02/2016
Page 7 of 8
Page 8
Ordering information
Model / Measuring range / Output signal / Accuracy at reference conditions / Process connection / Sealing / Electrical connection
© 2011 WIKA Alexander Wiegand SE & Co. KG, all rights reserved. The specications given in this document represent the state of engineering at the time of publishing. We reserve the right to make modications to the specications and materials.
Page 8 of 8
WIKA data sheet PE 81.54 ∙ 02/2016
WIKA Instrument Corporation
1000 Wiegand Boulevard Lawrenceville, GA 30043-5868 Tel: 888-WIKA-USA • 770-513-8200 Fax: 770-338-5118 E-Mail: info@wika.com www.wika.com
MM/YYYY country code based on 10/2014 EN