WIKA MWB, MW, MWH, MAB, MAH Instruction Manual

...
INSTRUCTION MANUAL
NI-221WE
REV. 5 12/16
WEATHERPROOF AND INTRINSICALLY SAFE: MODEL MW FLAMEPROOF: MODEL MA Versions: MWB-MW-MWH-MWG Versions: MAB-MA-MAH-MAG
A
= pressure attachment
B
= cable entry
For surface mounting use two screws M6
WEIGHT 1,8 kg Dimensions in mm
NOTE: dimensions and weights are not binding unless released on certified drawings.
CAUTION
Before installing, using or carrying out maintenance on the instrument it is necessary to read and understand the indications given
in the attached Instruction Manual.
The instrument must only be installed and maintained by qualified personnel. INSTALLATION IS TO BE CARRIED OUT ONLY AFTER CHECKING THAT INSTRUMENT CHARACTERISTICS ARE
CONSISTENT WITH PROCESS AND PLANT REQUIREMENTS.
The functional features of the instrument and its degree of protection are shown on the identification plate fixed to the case.
CONTENTS:
GENERAL NOTES
1 2OPERATING PRINCIPLE 3MODEL CODE 4IDENTIFICATION PLATE AND MARKINGS 5SET POINT REGULATION 6SET POIT CALIBRATION 7MOUNTING AND CONNECTIONS 8INSTRUMENT PLUMBING 9SAFETY INTEGRITY LEVEL (SIL) INSTALLATION REQUIREMENTS 10PUTTING INTO OPERATION 11VISUAL INSPECTION 12FUNCTIONAL VERIFICATION 13STOPPING AND DISMOUNTING 14DISPOSAL 15TROUBLESHOOTING
SAFETY INSTRUCTIONS FOR USE IN HAZARDOUS ATMOSPHERES.
RECOMMENDATIONS FOR PRESSURE SWITCH SAFE USE.
DIAPHRAGM PRESSURE SWITCHES
A
= pressure attachment
B
= cable entry
For surface mounting use four screws M6
WEIGHT 3,2 kg Dimensions in mm
RELATED DOCUMENT
To authentified document with certificate
N° IECEx PRE 16.0067X N° IECEx PRE 16.0074X
All data, statements and recommendations supplied with this manual are based on information believed by us to be reliable. As the
The present document is the property of ALEXANDER WIEGAND SE &Co and may not be reproduced in any form, nor used for any purpose other than that for which it is supplied.
1
GENERAL NOTES
1.1
FOREWORD
The wrong choice of a model/ version, as well as the incorrect in­stallation, lead to malfunction and reduce instrument life. Failure to abide by the indications given in this manual can cause damage to the instrument, the environment and persons.
1.2
ALLOWED OVERR ANGE
Pressure exceeding the working range can be occasionally toler­ated provided they remain within the limits stated in the instrument features (vacuum or proof pressure). Continuous pressures ex­ceeding the working range can be applied to the instrument pro­vided they are clearly stated in the instrument features. The current and voltage values stated in the technical specifications and rat­ings must not be exceeded. Transitory overages can have a de­structive effect on the switch.
1.3
MECHANICAL VIBRATION
Can generally lead to the wearing of some parts of the instrument or cause spurious action. It is therefore recommended that the in­strument be installed in a place where there are no vibrations. In cases where this is impossible it is a good idea to take measures to lessen the effects (elastic supports, installation with the switch plunger of the microswitch positioned at right angles to the vibra­tion plane).
1.4
TEMPERATURE
Due to the temperature of both the environment and the process fluid, the temperature of the instrument could exceed the allowed limits (normally from -40° to +60°C). Therefore, in case it does, suitable measures (protection against heat radiation, fluid separa­tors, cooling coils, heated lockers) must be taken. The process fluid or its impurities must not however solidify in the instrument.
2
OPERATING PRINCIPLE
The pressure, acting on the sensitive element (diaphragm), deter­mines its elastic deformation which is used to activate one or two electric microswitches regulated at set point values. The micro­switches are of the rapid release type with automatic rearm. When the pressure moves away from the set values, returning towards the normal values, the switch is rearmed. The dead band (gap be­tween the set point value and the reset value) can be fixed or ad­justable (letter R in the contact codes). The versions with sensing element H or G are piston activated di­aphragm instruments. The pressure, acting on a piston element, generates a force which, applied to the diaphragm, causes an elas­tic deformation.
3
MODEL CODE
See Annex 1
4
IDENTIFICATION PLATE AND MARKINGS
The instrument is fitted with a metal plate bearing all its functional characteristics and in case of flameproof or intrinsic safety execu­tion also the markings prescribed by standard IEC/EN 60079-0. Fig.1 shows the plate mounted on flameproof instruments.
Fig. 1 - Flameproof instruments plate
1 Notified body that issued the type certificate and number of said
certificate.
2 CE marking and identification number of the notified body re-
sponsible for production surveillance.
3 Apparatus classification according to ATEX 2014/34/EU di-
rective.
4 Type of protection and ambient temperature limits of operation.
INSTRUCTION MANUAL
NI-221WE
Rev. 5 12/16
The following table gives the relationship between hazardous ar­eas, Atex Categories and Equipment Protection Level (EPL) listed on the flameproof instrument nameplate
Hazardous area
Gas, vapours, fog Zone 0 1G Ga
Gas, vapours, fog Zone 1 2G or 1G Gb or Ga
Gas, vapours, fog Zone 2 3G, 2G or 1G Gc, Gb or Ga
Dust Zone 20 1D Da
Dust Zone 21 2D or 1D Db or Da
Dust Zone 22 3D, 2D or 1D Dc, Db or Da
5
SET POINT REGULATION
Each microswitch is independent and can be adjusted by means of a screw (for adjustment) to snap when the pressure reaches (increasing or decreasing) the desired value (set point). The instru­ment is usually supplied with the switches adjusted at the setting range value nearest to zero (factory calibration). The instrument is supplied with a label showing the set point calibration value. With factory calibration the values are not indicated, as these are tem­porary and will be modified with the definitive values. Prior to in­stallation the instrument must be calibrated and the definitive cal­ibration values written on the label.
If the instrument has been ordered with a specific calibration, check the calibration values marked on the relevant label, prior to installation.
The position of the adjustment screw is given in figure 2.
Fig. 2 - Electrical connections and adjustment screws
1- Microswitch 1 set point calibration screw 2- Microswitch 2 set point calibration screw 3- Electrical connection identification plate 4- Terminal block
Condition of the contacts at atmosphere pressure Designation of the contacts:
C - Common NA - Normally open
NC - Normally closed The effect that the direction of rotation of the adjustment screw has is described on the label.
6
SET POIT CALIBRATION
In order to proceed with the calibration and the periodical functional verification of the instrument a suitable calibration circuit (Fig. 3) and an adequate pressure source is required. The test instrument should have a measurement range approximately equal to or slightly wider than the pressure switch range and should have an accuracy consistent with the accuracy required to calibrate the set point.
Categories ac-
cording to 2014/34/EU Di­rective (ATEX)
EPL
2 of 9
Fig. 3 - Calibration circuit
PS - Pressure switch CA - Test gauge V1 - Inlet valve V2 - Discharge valve P - Pressure source
6.1
PRELIMINARY OPERATIONS
6.1.1
Weatherproof pressure switches (Model MW) Remove the blocking device fixed to the side of the instrument case (Fig. 4).Remove the cover by rotating it in an anticlockwise direction.
Fig. 4 - Weatherproof pressure switch blocking device
a - Plumbing wire b - Plumbing
6.1.2
Flameproof pressure switches (Model MA) CAUTION: do not open the cover of pressure switches when en-
ergized, in explosive atmospheres. Loosen the locking headless screw situated on the cover using a
1,5 hexagonal key then unscrew the cover (fig.5).
Fig. 5 - Flameproof pressure switch blocking device
Test fluid: air for P < 10 bar water for P > 10 bar
c - Nut d - Blocking bracket
INSTRUCTION MANUAL
NI-221WE
Rev. 5 12/16
The pressure switch must be mounted in the normal installation position, i.e. with the pressure connection pointing downwards. Avoid forcing the elastic support of the microswitch by hand or with tools. This could affect the instrument functioning.
CAUTION: if the switch is of the kind with adjustable dead band (letter R in the contact codes) before proceeding with the following operations it is necessary to proceed with the adjustment of the dead band.
Increase the pressure in the circuit up to the desired set point value for the first microswitch. Use a wide bladed screwdriver, as indi­cated on the label, turn the screw until the relative lamp turns on (or turns off).
- If the instrument is equipped with only one contact the calibration is complete.
- If it is equipped with two contacts continue in the following man­ner. Vary the pressure until the desired set point value for the sec­ond microswitch is reached. Act on the adjustment screw of the second contact. Repeat calibrating operations on the first contact, then on the sec­ond contact, until the required set point accuracy is obtained. This is necessary due to the reciprocal influence which the micro­switches have on the sensitive element of the instrument.
CAUTION: if the two set point are different they must be different for much of 5% of the adjustable span.
6.3
CALIBRATION OF INSTRUMENT WITH SENSING
ELEMENT H AND G
The pressure switches equipped with sensor H and G, are instruments with a piston sensing element. To perform an high accuracy calibration, due to the sensing element, it is necessary to adjust the set point by varying the pressure from the normal working pressure of operation to the set point.
6.4
CHECK OF SET POINT
Generate the normal working pressure and wait the pressure sta­bilisation. Vary the pressure into the circuit and record the set point value. Write the set point values on the adhesive label. Note: the repeatability should be checked verifying for three times the set point (Pi) starting always from the same pressure value (Pw). The pressure cycle should be slowly to give the possibility to record the set point with accuracy.
6.5
ADJUSTMENT OF DEAD BAND (LETTER R ON THE
MODEL CODE)
The dead band can be adjusted only on the instrument is equipped with a microswitch, which allows adjusting (Letter R on the model code). Adjustment may be obtained by rotating the wheel placed on the microswitch (Fig. 6). In order to carry out this operation it is advisable to insert thumb and forefinger of the left hand in the instrument. The instrument is normally delivered adjusted on the minimum value of its range (factory calibration).
Fig. 6 – Adjustment of dead band
6.2
CALIBRATION CIRCUIT AND OPERATIONS
Prepare the calibration circuit as indicated in Fig.3. The warning lamps should be connected to contact 1 or 2 in the NO or NC position according to the required contact action.
Connection of C and NO terminals
• If the circuit is open at the working pressure, the switch closes the circuit as the pressure increases when the desired value is reached.
• If the circuit is closed at the working pressure, the switch opens the circuit as the pressure decreases when the desired value is reached.
Connection of C and NC terminals
If the circuit is closed at the working pressure, the switch opens
the circuit as the pressure increases when the desired value is reached.
• If the circuit is open at the working pressure, the switch closes the circuit as the pressure decreases when the desired value is reached.
Mark Dead band adjustment wheel Set point adjustment screw
3 of 9
Loading...
+ 6 hidden pages