WIKA MAB User Manual

NI-221E
INSTRUCTION MANUAL
Rev. 1 11/02
DIAPHRAGM PRESSURE SWITCHES
WEATHERPROOF:SERIES MW EXPLOSIONPROOF: SERIES MA
Models: MW-MWH Models: MA-MAH-MAG
A = pressure attachment B = cable entry
WEIGHT 1,8 kg dimensions in mm WEIGHT 3,2 kg dimensions in mm
NOTE: dimensions and weights are not binding unless released on certified drawings.
CAUTION
Before installing, using or carrying out maintenance on the instrument it is necessary to read and understand the indications given in the attached Instruction Manual.
The instrument must only be installed and maintained by qualified personnel.
INSTALLATION IS TO BE CARRIED OUT ONLY AFTER CHECKING THAT INSTRUMENT CHARACTERISTICS
ARE CONSISTENT WITH PROCESS AND PLANT REQUIREMENTS.
The functional features of the instrument and its degree of protection are shown on the identification plate fixed to the case.
CONTENTS:
1
GENERAL NOTES
2 OPERATING PRINCIPLE 3 IDENTIFICATION PLATE AND MARKINGS 4 SET POINT REGULATION 5 SET POINT CALIBRATION 6 MOUNTING AND CONNECTIONS 7 INSTRUMENT PLUMBING 8 PUTTING INTO OPERATION 9 FUNCTIONAL VERIFICATION 10 TROUBLESHOOTING 11 STOPPING AND DISMOUNTING 12 DEMOLITION
SAFETY INSTRUCTIONS FOR USE IN HAZARDOUS ATMOSPHERES. RECOMMENDATIONS FOR PRESSURE SWITCH SAFE USE.
A = pressure attachment B = cable entry
RELATED DOCUMENT
To authentified document with certificate
N° CESI 02 ATEX 144
All data, statements and recommendations supplied with this manual are based on information believed by us to be
The present document is the property of ETTORE CELLA SPA and may not be reproduced in any form, nor used for any purpose other than that for which it is supplied.
INSTRUCTION MANUAL
1 GENERAL NOTES
1.1 FOREWORD
The wrong choice of a series or a model, as well as the incorrect installation, lead to malfunction and reduce in­strument life. Failure to abide by the indications given in this manual can cause damage to the instrument, the environment and persons.
1.2 ALLOWED OVERRANGE
Pressure exceeding the working range can be occasion­ally tolerated provided they remain within the limits stated
in the instrument features (vacuum or proof pressure). Continuous pressures exceeding the working range can be applied to the instrument provided they are clearly stated in the instrument features. The current and voltage values stated in the technical specifications and ratings must not be exceeded. Transitory overages can have a destructive effect on the switch.
1.3 MECHANICAL VIBRATION
Can generally lead to the wearing of some parts of the instrument or cause false actuation. It is therefore recom­mended that the instrument be installed in a place where there are no vibrations. In cases where this is impossible it is a good idea to take measures to lessen the effects (elastic supports, installation with the peg of the mi­croswitch positioned at right angles to the vibration plane).
1.4 TEMPERATURE
Due to the temperature of both the environment and the process fluid, the temperature of the instrument could exceed the allowed limits (normally from -20° to +70°C). Therefore, in case it does, suitable measures (protection against heat radiation, fluid separators, cooling coils, heated lockers) must be taken.
2 OPERATING PRINCIPLE
• The pressure, acting on the sensitive element (dia-
phragm), determines its elastic deformation which is used to activate one or two electric microswitches regulated at set point values. The microswitches are of the rapid re­lease type with automatic rearm. When the pressure moves away from the set values, returning towards the normal values, the switch is rearmed. The differential gap (difference between the set point value and the reset value) can be set or adjustable (letter R in the contact codes).
• Piston activated diaphragm instruments (MWH - MAH)
differ from the former as the pressure, acting on the piston
Fig. 2 - Electrical connections and adjustment screws
NI-221E
Rev. 1 11/02
element, generates a force which, applied to the dia­phragm, causes an elastic deformation. This deformation is then used as in the previous case, to activate one or two microswitches.
3 IDENTIFICATION PLATE AND MARKINGS
The instrument is fitted with a metal plate bearing all its functional characteristics and – in case of explosionproof execution (Series MA) – also the markings prescribed by standard CEI EN 50014. Fig.1 shows the plate mounted on explosionproof instruments.
Fig. 1 - Explosionproof instruments plate
1 CE marking and identification number of the notified body
responsible for production surveillance. 2 Apparatus classification according to ATEX 949 CE directive. 3 Type of protection and ambient temperature limits of opera-
tion. 4 Notified body that issued the type certificate and number of
said certificate.
4 SET POINT REGULATION
Each microswitch is independent and can be regulated by means of a screw (for adjustment) in such a way that it is released when the pressure reaches (increasing or de­creasing) the desired value (set point). The instrument is usually supplied with the switches set at the setting range value nearest to zero (factory calibration). The instru­ment is supplied with an adhesive rating plate showing the set point calibration value. With factory calibration the values are not indicated on the ratings, as these are tem­porary and will be modified with the definitive values. Prior to installation the instrument must be calibrated and the definitive calibration values written on the adhesive rating plate using a suitable indelible ink pen.
1- Microswitch 1 set point
calibration screw
2- Microswitch 2 set point
calibration screw
3- Electrical connection identifi-
cation plate
4- Terminal block
Designation of the contacts: C - Common NA - Normally open NC - Normally closed
INSTRUCTION MANUAL
If the instrument has been ordered with a specific cali- bration, it is a good rule to check the calibration values marked on the relevant adhesive label, prior to installation.
The position of the adjustment screw is given in figure 2. The effect that the direction of rotation of the adjustment screw has is described on the adhesive plate.
5 SET POIT CALIBRATION
In order to proceed with the calibration and the periodical functional verification of the instrument a suitable calibra- tion circuit (fig. 4) and an adequate pressure source is required.
5.1 PRELIMINARY OPERATIONS
5.1.1 Weatherproof pressure switches (Series MW)
Remove the blocking device fixed to the side of the in­strument case (Fig. 3). Remove the cover by rotating it in an anticlockwise direc­tion.
Fig. 3 - Weatherproof pressure switch blocking device
NI-221E
Rev. 1 11/02
If the circuit is open at the working pressure, the switch closes the circuit as the pressure increases when the desired value is reached.
• If the circuit is closed at the working pressure, the switch opens the circuit as the pressure decreases when the desired value is reached.
Connection of C and NC terminals
If the circuit is closed at the working pressure, the switch opens the circuit as the pressure increases when the
desired value is reached.
• If the circuit is open at the working pressure, the switch closes the circuit as the pressure decreases when the desired value is reached. The test instrument should have a measurement range approximately equal to or slightly wider than the pressure switch range and should have an accuracy consistent with the precision required to calibrate the set point. The pressure switch must be mounted in the normal in­stallation position, i.e. with the pressure connection point­ing downwards.
Fig. 5 - Calibration circuit
a - Plumbing wire b - Plumbing
5.1.2 Explosionproof pressure switches (Series MA)
CAUTION: do not open the cover of explosionproof pres-
sure switches (Series MA) when energized, in explosive atmospheres.
Loosen the locking headless screw situated on the cover using a 1,5 hexagonal key then unscrew the cover. (Fig.
4).
Fig. 4 - Explosionproof pressure switch blocking device
c - Nut d - Blocking bracket
5.2 CALIBRATION CIRCUIT AND OPERATIONS
Prepare the control circuit as indicated in Fig.5. The warning lamps should be connected to contact 1 or 2 in the NO or NC position according to the required contact action.
Connection of C and NO terminals
PS - Pressure switch CA - Test pressure gauge V1 - Inlet valve V2 - Discharge valve P - Pressure source
Avoid forcing the elastic support of the microswitch by hand or with tools. This could affect the instrument func­tioning.
CAUTION: if the switch is of the kind with adjustable dead band (letter R in the contact codes) before proceeding with the following operations it is necessary to proceed with the adjustment of the dead band (see attachment NI-
706).
Increase the pressure in the circuit up to the desired set point value for the first microswitch. Use a wide bladed screwdriver, as indicated on the adhesive plate, until the relative lamp turns on (or turns off).
- If the instrument is equipped with only one contact the calibration is complete.
- If it is equipped with two contacts continue in the follow­ing manner. Vary the pressure until the desired set point value for the second microswitch is reached. Act on the adjustment screw of the second contact. Repeat calibrating operations on the first contact, then on the second contact, until the required set point precision is obtained. This is necessary due to the reciprocal influence which the microswitches have on the sensitive element of the instrument. Check the calibration values (varying the pressure in the circuit accordingly) and record them on the adhesive plate using a pen with indelible ink.
5.3 FINAL OPERATIONS
Disconnect the instrument from the calibration circuit.
Test fluid: air for P 10 bar water for P > 10 bar
INSTRUCTION MANUAL
5.3.1 Weatherproof pressure switches (Series MW)
Take the cover, ensure that the sealing gasket is correctly fitted into its seat, and insert the cover onto the case, with the blocking gap positioned in correspondence to the blocking bracket. Turn the cover clockwise closing it tightly. Mount the blocking device as in Fig. 3.
5.3.2 Explosionproof pressure switches (Series MA).
Screw on the cover and block it using the headless screw with which it is equipped (Fig. 4)
Mount on pressure connection and cable entry the protec­tion caps supplied with the instrument. The protection caps should only be definitively removed during the con­nection steps (see § 6).
6 MOUNTING AND CONNECTIONS
6.1 MOUNTING Surface mount the instrument by means of the holes
provided, or pipe mount using the appropriate bracket (see Fig. 9). The chosen position must be such that vibrations, the possibility of shocks or temperature changes are within tolerable limits. The above also applies to direct mounting. With gas or vapour process fluid, the instrument must be positioned higher than the pipe inlet (see Fig. 8). With a liquid process fluid, the instrument can be positioned higher or lower, indifferently (see Fig. 7 e 8). ). In this case, during set point calibration the negative or positive
head must be taken into account (distance h in Fig.7 e 8).
6.2 PRESSURE CONNECTIONS
For a correct installation it is necessary to: Mount a shut-off valve with drain (root valve) on the proc­ess tube to allow the instrument to be excluded and the connection tubing to be drained. It is recommended that said valve has a capstan-blocking device aimed at pre­venting it being activated casually and without authorisa­tion. Mount a service valve near the instrument to permit possible functional verification on site. It is recommended that the service valve is closed with a plug to prevent the outlet of the process fluid caused by the incorrect use of said valve. Mount a three-piece joint onto the threaded attachment of the instrument to permit the easy mounting or removal of the instrument itself. Carry out the connec­tion using a flexible tube in such a way that variations in the temperature of the tube itself do not force the instru­ment attachment. Ensure that all the pressure connections are airtight. It is important that there are no leakages in the circuit. Close the root valve and the relative drain device. Close the service valve using a safety plug.
6.3 ELECTRICAL CONNECTIONS
It is recommended to carry out the electrical connections according to the applicable standards. In case of explo­sionproof instruments (Series MA) see also the Standard EN-60079-14. If the electrical connection is carried out in a protected tube, it shall be made so that condensate is prevented from entering instrument enclosure.
The arrangement shown in Fig. 7 or 8 is therefore recom­mended..
CAUTION: fittings used for the electrical connection of
the pressure switch Series MA (explosionproof) shall be certified to Standards EN 50014 and 50018, and shall guarantee instrument degree of protection (IP65).
Check that there is no power in the lines. Remove the cover and carry out the cabling and connec­tions to the terminal block (see Fig. 2). Flexible cables with a maximum section of 1.2 mm (16AWG) are recommended using the pre-insulated fork
2
NI-221E
Rev. 1 11/02
thimbles. Do not touch the adjustment screws and do not bend the elastic microswitch supports in order to prevent the instrument calibration being altered. Ensure that no deposits or wire ends remain inside the case.
As soon as connection steps are completed, mount the cover on and make sure it is tight and blocked See Fig. 3 and 4.
6.4 SPECIAL NOTE FOR INSTALLATION OF CATEGORY 1 / 2 G PRESSURE SWITCHES
Explosionproof pressure switches (Series MA) can be installed on processes requiring apparatus of group II category 1 in an ambient requiring apparatus of group II category 2 (see Fig. 6).
Fig. 6 - Installation of Group II Cat. 1 / 2 G instruments
7 INSTRUMENT PLUMBING
7.1 Weatherproof p ressure switches (Series MW)
The plumbing, aimed as a guarantee against possible tampering of the calibration and electrical connections, can be carried out using a flexible steel wire (c) inserted into the holes in the screw (a) and the bracket (e) provided for this purpose (see Fig. 3).
7.2 Explosionproof pressure switches (Series MA )
Plumbing is not necessary as the cover is blocked with a headless screw and the instrument does not have to be opened when installed (see Fig. 4).
8 PUTTING INTO OPERATION
As the signal transmitted by the instrument is used in a complex system, it is necessary that the means of putting it into operation are established by those in charge of the plant.
The instrument comes into operation as soon as the root valve is opened. Any possible drainage of the connection tubing can be carried out by removing the safety plug and opening the service valve with the necessary caution.
In case of explosionproof instruments (Series MA), initial inspections are to be carried out according to customer procedures and at least in accordance with Standard EN­60079-17.
NI-221E
INSTRUCTION MANUAL
9 FUNCTIONAL VERIFICATION
This will be carried out according to the Client’s control procedures. Series MW and MWH instruments can be verified on the plant if installed as illustrated in Fig. 7 e 8.
The instruments Series MA may be checked on site only if apparatus suitable for explosive atmospheres are used and provided that the electric line is not energized.
If this is not the case it is necessary to stop operation, dismount by means of the three-piece joints and carry out the verification in a test room.
10 TROUBLESHOOTING
IMPORTANT NOTE: operations involving replacement of essential components must be carried out at our work­shop, especially for instruments with explosionproof certificate; this is to guarantee the user the total and cor­rect restoration of the product original characteristics.
MALFUNCTION PROBABLE CAUSE REMEDY
Set point shift
Poor repeatabil­ity
Slow response
No actuation or undue actuation
Undue actua­tion
11 STOPPING AND DISMOUNTING
Before proceeding with these operations ensure that the plant or machines have been put into the conditions foreseen to allow these operations.
With reference to Figures 7 e 8
Remove the power supply (signal) from the electrical line. Close the root valve (6) and open the drain. Remove the plug (2), open the valve (3) and wait until the process fluid has drained from the tubing through the drain.
Do not dispose of the process fluid into the environment, if this can cause pollution or damage to people.
Unscrew the three-piece joint (8).
CAUTION: do not open the cover of explosionproof pres­sure switches (Series MA) when energized, in explosive atmospheres.
Unscrew the three-piece joint (10) (electrical cable tubing).
Permanent deformation of the sensitive ele-
ment due to fatigue or non-tolerated overages.
Variation of the elastic features of the sensitive
element due to its chemical corrosion.
O-ring wear (only MWH and MAH) O-ring wear (only MWH and MAH)
Air bubbles or condensation (only for types
with pressure <1 bar).
Clogged or obstructed connection line. Root valve partially closed. Too viscous fluid. Root valve closed.
Microswitch contacts damaged. Loosened electrical joints. Interrupted or short-circuited electrical line. Accidental shocks or excessive mechanical
vibrations.
CAUTION: do not open the cover of explosionproof pres-
sure switches (Series MA) when energized, in explosive atmospheres.
Verification consists in checking the calibration values and possibly regulating the adjustment screw (see §4 ) Verification consists in checking the calibration value and possibly regulating the adjustment bush (see §5).
In case of explosionproof instruments (Series MA), in­spections of the electrical installation are to be carried out also according to customer procedures and at least in accordance with Standard EN-60079-17.
Recalibrate or replace the sensitive element. Recalibrate or replace the sensitive element
with another made of a suitable material. If necessary apply a fluid separator.
Replace the piston subgroup and recalibrate. Replace the piston subgroup and recalibrate.
Drain the process connection line and if
necessary modify it.
Check and clean line. Open valve. Provide instrument with suitable fluid separa-
tor.
Open the valve. Replace the Microswitch. Check all electrical joints. Check the conditions of the electrical line. Modify the mounting.
Remove the instrument cover and disconnect the electrical cables from the terminal block and earth screws. Remove the screws fixing the case to the panel (or pipe) and remove the instrument, taking care to slide the electri­cal conductors out from the case. Mount instrument cover. Insulate and protect cables around, if any. Temporarily plug pipes not connected to the instrument.
In case of explosionproof instruments (Series MA) it is recommended to follow - at least – the standard EN­60079-17 for the withdrawal from service of electrical apparatus.
12 DEMOLITION
The instruments are mainly made of stainless steel and aluminium and therefore, once the electrical parts have been dismounted and the parts coming into contact with fluids which could be harmful to people or the environment have been properly dealt with, they can be scrapped.
Rev. 1 11/02
NI-221E
INSTRUCTION MANUAL
WEATHERPROOF EXPLOSIONPROOF
Fig. 7 - Example of connections Fig. 8 - Example of cpnnections
Rev. 1 11/02
Fig. 9 - Example of mounting
Surface mounting
Montaggio a parete
2” pipe mounting
Montaggio su staffa da 2”
NOTE With gas or vapour process fluid, the instrument must be positioned higher than the pipe inlet (see Fig.
8). With a liquid process fluid, the instrument can be positioned higher or lower, indifferently (see Fig. 7 e 8). In this case, during set point calibration the negative or positive head must be taken into account. (distance h in Fig.7 and 8).
LEGEND 1 - Fitting 2 - Drain plug 3 - Service valve 4 - Piping 5 - Three piece fitting 6 - Root valve with drain 7 - Process piping 8 - Three piece fitting 9 - “T” fitting 10 - Three piece fitting 11 - Curve 12 - Blocking joint 13 - M6 screws (No4) 14 - Bracket for 2” pipe 15 - Horizontal pipe 16 - Vertical pipe
ETTORE CELLA SPA Viale de Gasperi, 48 - Casella Postale (P.O. Box) 96 - I 20010 Bareggio (MILANO) ITALY
Telefoni +39 029036.1146/1147/1237/1241 - FAX +39 029036.1331 E-mail: cella@ecellaspa.com
Loading...