WIKA CF2S User Manual

Operating Instructions
WIKA Operating Instructions CF2S V1.2 • 08/2004
WIKA Alexander Wiegand GmbH & Co. KG
Alexander-Wiegand-Straße 30 63911 Klingenberg/Germany Phone (+49) 93 72/132-0 Fax (+49) 93 72/132-406 E-Mail www.wika.de
info@wika.de
Operating Instructions Temperature Indicating Controller CF2S
Thank you for your purchase of our Microcomputer based Temperature Indicating Controller CF2S. This manual contains instructions for the mounting, functions, operations and notes when operating the CF2S. For model confirmation and unit specifications, please read this instruction manual carefully before starting operation.
To prevent accidents arising from the use of this controller, please ensure the operator using it receives this manual.
***** Note to users ***** Before operating this controller, make sure you have understood the following.
Warning
Turn the power supply to the instrument OFF before wiring or checking. Working or touching the terminal with the power switched ON may result in an Electric Shock which could cause severe injury or death. Moreover, the instrument must be grounded before the power supply to the instrument is turned on.
Caution
• Do not apply a commercial power source to the sensor connected to the input terminal nor allow the power source to come into contact with the sensor, as the input circuit may be burnt out.
• This instrument should be used according to the specifications described in the manual. If it is used outside the specifications, it may malfunction or cause fire.
• Be sure to follow the warnings, cautions and notices. If not, it could cause serious injury or malfunction.
• Specifications of the CF2S and the contents of this instruction manual are subject to change without notice.
• Care has been taken to assure that the contents of this instruction manual are correct, but if there are any doubts, mistakes or questions, please inform our sales department.
• Be sure to turn the power supplied to the instrunment OFF when cleaning.
• Wipe the instrument using dry soft cloth.
(If the paint thinner is used for wiping, the instrument may be deformed or discolored.)
• The display parts are more easily damaged. Do not strike them with hard objects
or press hard on them.
• Any unauthorized transfer or copying of this document, in part or in whole, is prohibited.
• WIKA is not responsible for any damages or secondary damages incurred as a result of using this product, including any indirect damages.
• Setup by the DIP switch and the rotary switch inside of the controller is required before the power is turned on.
• It is recommended that the PID auto-tuning be performed on the trial run.
• It is advised to provide the protective device against such environmental conditions which may cause damage to the device or contribute to the deterioration of its parts.
V1.2 • 08/2004 - 2 -
Operating Instructions Temperature Indicating Controller CF2S
--- CONTENTS ---
1. Model names -------------------------------------------------------------------------------- 5
2. Name and functions of the sections -------------------------------------------- 6
3. Setup
3.1 Taking the internal assembly out ----------------------------------------------------- 7
3.2 Switch setting (multi-function) --------------------------------------------------------- 7
3.3 Insertion of the internal assembly ---------------------------------------------------- 10
4. Operations
4.1 Operating flow chart -------------------------------------------------------------------- 11
4.2 Operations (1) PV/SV display mode ------------------------------------------------------------------ 13 (2) Main setting mode --------------------------------------------------------------------- 14 (3) Sub setting mode
Setting value memory number selection ----------------------------------------- 14 Auto-tuning Perform/Cancel --------------------------------------------------------- 14 Proportional band setting ------------------------------------------------------------- 15 Integral time setting -------------------------------------------------------------------- 15 Derivative time setting ----------------------------------------------------------------- 15 Proportional cycle setting ------------------------------------------------------------- 15
Manual reset setting ------------------------------------------------------------------- 15 Alarm 1, 2 (A1, A2) settings ---------------------------------------------------------- 16 Loop break alarm time setting ------------------------------------------------------- 16 Loop break alarm span setting ------------------------------------------------------ 16
(4) Auxiliary function setting mode 1
Setting value lock designation ------------------------------------------------------ 17 Main setting value high limit setting ------------------------------------------------ 18 Main setting value low limit setting ------------------------------------------------- 18 Sensor correction setting ------------------------------------------------------------- 18 Instrument number setting ----------------------------------------------------------- 18 Transfer rate selection ---------------------------------------------------------------- 18 Communication protocol selection ------------------------------------------------- 18
(5) Auxiliary function setting mode 2
Scaling high limit value setting ------------------------------------------------------ 19 Scaling low limit value setting ------------------------------------------------------- 19 PV filter time constant setting ------------------------------------------------------- 19 Main output high limit setting -------------------------------------------------------- 19 Main output low limit setting --------------------------------------------------------- 19 Main output ON/OFF action hysteresis setting --------------------------------- 20 Alarm 1, 2 (A1, A2) action Energized/Deenergized selections ------------- 20 Alarm 1, 2 (A1, A2) hysteresis settings ------------------------------------------- 20 Alarm 1, 2 (A1, A2) action delayed timer settings ------------------------------ 21 Display selection when control output is off ------------------------------------- 21 Main setting value rising rate setting ---------------------------------------------- 21 Main setting value falling rate setting --------------------------------------------- 21
V1.2 • 08/2004 - 3 -
Operating Instructions Temperature Indicating Controller CF2S
(6) Program mode
Program control change -------------------------------------------------------------- 22
Step 1 to 7 time setting --------------------------------------------------------------- 23 (7) Control output OFF function -------------------------------------------------------- 24 (8) Output manipulating value and Step rest time indication mode ----------- 25
5. Setting value memory number external selection ----------------------- 25
6. Running
6.1 When using the CF2S as a Temperature controller ------------------------------ 26
6.2 When using the CF2S as a Simplified program controller ---------------------- 26
7. Action explanations
7.1 Standard action drawings --------------------------------------------------------------- 27
7.2 ON/OFF action drawings ---------------------------------------------------------------- 28
7.3 Pattern end action drawing ------------------------------------------------------------- 28
7.4 Alarm 1, 2 (A1, A2) action drawings -------------------------------------------------- 29
8. Control actions
8.1 Fuzzy self-tuning -------------------------------------------------------------------------- 31
8.2 Explanations of PID ---------------------------------------------------------------------- 31
8.3 PID auto-tuning of this controller ----------------------------------------------------- 32
9. Other functions ---------------------------------------------------------------------------- 33
10. Mounting to control panel
10.1 Site selection ------------------------------------------------------------------------------- 34
10.2 External dimension drawing ------------------------------------------------------------ 34
10.3 Panel cutout drawing --------------------------------------------------------------------- 34
10.4 Mounting ------------------------------------------------------------------------------------- 35
11. Wiring connection
11.1 Terminal arrangement -------------------------------------------------------------------- 36
11.2 Wiring connection examples ------------------------------------------------------------ 38
12. Specifications
12.1 Standard specifications ------------------------------------------------------------------ 39
12.2 Optional specifications ------------------------------------------------------------------- 42
13. Troubleshooting --------------------------------------------------------------------------- 44
14. Character table ----------------------------------------------------------------------------- 46
V1.2 • 08/2004 - 4 -
Operating Instructions Temperature Indicating Controller CF2S
1. Model names
CF2S -3 A- / M - - - - ..... Series name CF2S
3 Control action PID
(1)
Alarm A Alarm action, output relais Output
R Relais contact output S Non contact voltage output (0/12 VDC)
A Current output 4 ... 20 mA Input M Multi-range input Power supply
Mounting provision
H L
AC 100 ... 240 V, 50 ... 60 Hz AC/DC 24 V
FT One-touch mounting bracket (for mounting
panel thickness 1 to 3 mm)
FS Screw bracket (for mounting panel thickness
1 to 15 mm)
Instrument configuration
B Factory adjustment, device will be configured
by the operator
?/# To customers specification
Options
2AS 2AR 2AL
CR2 CR5 IP4
(3)
Alarm 2 output
(3)
Loop break alarm output
(3)
Alarm 2 output and loop break alarm output with common terminals
(4)
Serial communication RS 232C
(4)
Serial communication RS 485
(5)
IP 54 front protection
KAB Terminal cover
(2)
(1) Fuzzy self-tuning PID, PID, PD and ON/OFF action are selectable by internal DIP switch. (2) 12 types of alarm action and no alarm action are selectable by internal rotary switch and DIP switch. (3)(4) Please note that only one option from every group can be selected. (5) Condition for this option is the screw type mounting bracket (FS).
V1.2 • 08/2004 - 5 -
Operating Instructions Temperature Indicating Controller CF2S
2. Name and functions of the sections
(1)
(3)
(2)
[Fig. 2-1]
(1) PV display
Indicates the Process variable (PV) with the red LED.
(2) SV display
Indicates the Setting value (SV), Manipulating value (MV) or Time (TIME) with the green LED.
(3) Control output indicator (OUT)
The green LED is lit when the Control output is on. (For the current output type, it blinks corresponding to the manipulating value at 0.125 seconds of period.)
(4) Alarm 1 (A1) output action or Pattern end 1 output indicator (A1)
The red LED is lit when the Alarm 1 (A1) output or Pattern end 1 output is on.
(5) Alarm 2 (A2) output action or Pattern end 2 output indicator (A2) [Option]
The red LED is lit when the Alarm 2 (A2) output or Pattern end 2 output is on.
(6) Auto-tuning action indicator (AT)
The yellow LED blinks during auto-tuning.
(4)
(5)
(6)
Increase key : Increases the numeric value on the SV display during setting mode.
Decrease key : Decreases the numeric value on the SV display during setting mode.
Mode key : Selects the setting mode.
OUT/OFF key : Performs the control output ON or OFF, or program control Start or Stop.
In any mode, if the When the function is working, the function cannot be released. Even if the insrument power is turned off and on again, and the function is not released. To release the function, press the
The setting value is registered by pressing the In any setting item, if the key is pressed for approx. 3 seconds, the mode returns to the PV/SV display.
key is pressed for approx. 1 second, the Control output OFF function will work.
key for 1 second.
key.
V1.2 • 08/2004 - 6 -
Operating Instructions Temperature Indicating Controller CF2S
A
A
3. Setup
3.1 Taking the internal assembly out
Warning
Do not take the inner assembly out nor touch the terminal with the power supply on. Touching the terminal with the power switched ON may result in an Electric Shock which could cause severe injury or death.
Before the power supply to this instrument is on, take the internal assembly out from the case by pushing the hook (bottom of the instrument) in the direction indicated by the arrow and holding the notches.
[Fig. 3.1-1]
3.2 Switch setting (multi-function)
Using a small slotted screwdriver and tweezers, set the Sensor input, Alarm 1 action, Alarm 2 action, Control action, Heating (reverse)/Cooling (direct) action, Alarm 1 and 2 standby functions, Unit °C/°F and Program starting form Auto/Manual change by rotary switch and DIP switch by the following procedure. The rotary switch A2 (SW301) is equipped only when the option A2 is applied.
Control action designation Alarm 2 action type designation
Heating/Cooling action designation Alarm 1 action type designation
larm 1 standby action designation Sensor input designation
larm 2 standby action designation
°C/°F designation
Sensor input designation
Program start Auto/Manual
designation
Rotary switch
DIP switch
[Fig. 3.2-1]
V1.2 • 08/2004 - 7 -
Operating Instructions Temperature Indicating Controller CF2S
The following items can be designated by the DIP switch (SW304). Factory adjusted as all switches OFF [
].
[Table 3.2-1]
Item
Control action
action
Sw. No.
1
and
2
3
Fuzzy self-tuning PID action No.1: OFF No.2: OFF PID action No.1: ON No.2: OFF PD action No.1: OFF No.2: ON ON/OFF action No.1: ON No.2: ON Heating (reverse) action No.3: OFFHeating/Cooling Cooling (direct) action No.3: ON
Designation Switch status
No standby action No.4: OFFAlarm 1 (A1)
standby action
4
Standby action No.4: ON
No standby action No.5: OFFAlarm 2 (A2)
standby action
5
Standby action No.5: ON
(*1)
°C/°F 6
Sensor input
(*2)
°C No.6: OFF °F No.6: ON K, J, R, B, N, PL- , Pt100, JPt100
7
S, E, T, C,
(With decimal point)
No.7: OFF
No.7: ON Pt100 (Without decimal point) Manual start No.8: OFFProgram start
Auto/Manual
8
Automatic start No.8: ON
*1: The standby function will not work if the option A2 is not applied. *2: The sensor input can be designated by the combination of this item and the rotary switch sensor input (SW303). (See page 11.)
Program start: This item is available for the program control.
Manual start : Program starts by pressing the
key.
Automatic start : Set program is automatically started from step 1 by turning the power
supply to the instrument on (after 2 seconds of warm-up status).
V1.2 • 08/2004 - 8 -
Operating Instructions Temperature Indicating Controller CF2S
Select the sensor type by rotary switch sensor input (SW303). Factory adjusted as K, [
[Table 3.2-2]
].
Rotary Sw.
No.
0OFFK -200 to 1370 °C -320 to 2500 °F
1 OFF J -200 to 1000 °C -320 to 1800 °F
2 OFF R 0 to 1760 °C 0 to 3200 °F
3 OFF B 0 to 1820 °C 0 to 3300 °F
4OFF PL- 0 to 1390 °C 0 to 2500 °F
5 OFF N 0 to 1300 °C 0 to 2300 °F
6 OFF Pt100 -199.9 to 850.0 °C -199.9 to 999.9 °F
7 OFF JPt100 -199.9 to 500.0 °C -199.9 to 900.0 °F
0 ON S 0 to 1760 °C 0 to 3200 °F
DIP Sw. No. 7
Type of
sensor
Scale range
1 ON E 0 to 1000 °C 0 to 1800 °F
2 ON T -199.9 to 400.0 °C -199.9 to 750.0 °F
3 ON C (W/Re5-26) 0 to 2315 °C 0 to 4200 °F
7 ON Pt100 -200 to 850 °C -320 to 1560 °F
V1.2 • 08/2004 - 9 -
Operating Instructions Temperature Indicating Controller CF2S
The alarm action type and the pattern end output when program control can be designated by the rotary switch A1 (SW302) and A2 (SW301). The rotary switch A2 (SW301) is equipped only when the option A2 is applied.
Rotary switch A1 (SW302): Alarm 1 (A1) action or Pattern end 1 output. Rotary switch A2 (SW301): Alarm 2 (A2) action or Pattern end 2 output.
Factory adjusted as No alarm action, [
[Table 3.2-3]
Alarm 1 action
No alarm action 0 No alarm action 0
High limit alarm 1 High limit alarm 1
Low limit alarm 2 Low limit alarm 2
High/Low limits alarm 3 High/Low limits alarm 3
Hi/Lo limit range alarm 4 Hi/Lo limit range alarm 4
Process high alarm 5 Process high alarm 5
Process low alarm 6 Process low alarm 6
A1 Rotary Sw. (Sw302) No.
].
Alarm 2 action
A2 Rotary Sw. (Sw301) No.
Pattern end 1 output 7 Pattern end 2 output 7
3.3 Insertion of the internal assembly When the setup is completed, insert the internal assembly into the case. Firmly insert the assembly until it is locked by the hook at the bottom of the instrument. (There will be a clicking sound.)
Caution
Do not confuse the top and bottom of the internal assembly. If inserting the assembly into the case by force in the wrong direction, the PCB may be damaged.
V1.2 • 08/2004 - 10 -
Operating Instructions Temperature Indicating Controller CF2S
4. Operations
4.1 Operating flow chart
Instrument power ON
Warm-up status (For approx. 2s, displays as shown on page 15.)
1s
PV/SV display mode
+ + 3s
[Main setting mode] [Sub setting mode] [Auxiliary function setting mode 1]
Control output OFF function
3s
Output manipulating value
Step rest time *1
Main setting Setting value memory No. selection Setting value lock
PV
PV
Auto-tuning Perform/Cancel
PV
*2
designation
PV
Main setting value
Proportional band setting high limit setting
PV
Integral time setting
PV
Derivative time setting
PV
Proportional cycle setting
PV
Manual reset setting
PV
Alarm 1 (A1) setting
PV
Transfer rate selection
Alarm 2 (A2) setting
PV
setting
PV
PV
Main setting value low limit setting
PV
Sensor correcting
value setting
PV
Instrument number
PV
Communication protocol
Loop break alarm time setting selection
PV
Loop break alarm span setting
PV
PV
V1.2 • 08/2004 - 11 -
Operating Instructions Temperature Indicating Controller CF2S
: Options + : Press the key while the key is being pressed. + 3s: Press the key for approx. 3 seconds while the key is being pressed. + + 3s: Press the key for approx. 3 seconds
while the
and keys are being pressed.
+ 3s : Pressing the key, press the key for approx. 3 seconds.
*1: Step rest time is indicated only during program control. *2: If the Auto-tuning is performed, the mode reverts to the PV/SV display.
If the Auto-tuning is released, the mode will move to “Proportional band setting”.
*3: In the mode Program control change, if is selected and the key is
pressed, the display reverts to the PV/SV display mode.
Set the temperature setting of program in the Sub setting mode, Auxiliary fuction setting mode 1 and 2.
+ + 3s + 3s
[Auxiliary function setting mode 2] [Program mode]
Scaling high limit Alarm 1 (A1) hysteresis Program control change
setting setting
PV
PV
PV
Scaling low limit Alarm 2 (A2) hysteresis Step 1 time setting
setting setting
PV
PV
PV
PV filter time constant Alarm 1 (A1) action Step 2 time setting
setting delayed timer setting
PV PV
PV
Main output high limit Alarm 2 (A2) action Step 3 time setting
setting delayed timer setting
PV PV
PV
Main output low limit Display selection when Step 4 time setting
setting control output off
PV PV
PV
Main output ON/OFF Main setting value Step 5 time setting
action hysteresis setting rising rate setting
PV PV
PV
*3
Alarm 1 (A1) action Main setting value Step 6 time setting
Energized/Deenergized falling rate setting
PV
PV
PV
Alarm 2 (A2) action Step 7 time setting
Energized/Deenergized
PV PV
V1.2 • 08/2004 - 12 -
Operating Instructions Temperature Indicating Controller CF2S
4.2 Operations
The PV display indicates the sensor type and the SV display indicates the rated value selected by Sensor input designation (page 10, 11) for approx. 2 seconds after the power is turned on. See [table 4.2-1]. During this time, all outputs and LED indicators are in their off status. After that, the actual temperature is displayed on the PV display, main setting value on the SV display and starts control.
(1) PV/SV display mode
Instrument power ON
Warm-up status (for 2s)
[Table 4.2-1] Rated value
[Table 4.2-1]
Input
PV display SV display PV display SV display
°C °F
K
J
R
B
PL-ll
N
S
E
T
C
Pt100
JPt100
Pt100
PV/SV display mode Mode during control.
Actual Main The contents of setting items nor setting values
Temperature setting value cannot be changed.
V1.2 • 08/2004 - 13 -
Operating Instructions Temperature Indicating Controller CF2S
(2) Main setting mode
In the PV/SV display mode, if the
key is pressed, the Main setting mode will be selected.
The setting value can be increased or decreased by pressing the and keys.
If the
key is pressed, the setting value will be registered and the controller will
revert to the PV/SV display.
PV/SV display mode
Actual Main setting Temperature value
Main setting
Setting Mode to set the setting value of the main control.
value
Setting range: From Main setting value low limit
to Main setting value high limit
[Factory adjusted as 0°C]
(3) Sub setting mode
In the PV/SV display mode, if the
key is pressed while the key is being pressed,
Sub setting mode will be selected.
The setting value can be increased or decreased by pressing the
If the
key is pressed, the setting value is registered and the next setting item will be selected.
PV/SV display mode
and keys.
Actual Main setting temperature value
+
Setting value memory Mode to select the memory number (file) number selection to be set or to be called.
Selected Selecting items: 1 to 7
value [Factory adjusted as Memory number 1.]
Auto-tuning Mode to designate the auto-tuning Perform or Cancel.
Perform/Cancel If Auto-tuning Perform is designated, and the Designation key is pressed, the controller reverts to PV/SV display mode.
(When Perform) To set the other values, cancel the Auto-tuning.
When the ON/OFF action or PD action is selected in the
Control action designation (page 10), this display is not available.
V1.2 • 08/2004 - 14 -
Operating Instructions Temperature Indicating Controller CF2S
SV
(Auto-tuning Cancel)
SV
If the Auto-tuning is released during the process,
Proportional band setting Mode to set the proportional band.
Setting value Setting range: 0.1 to 999.9%
Integral time setting Mode to set the integral time.
Setting value Setting range: 0 to 3600s
[Factory adjusted as 200s]
Derivative time setting Mode to set the derivative time. Setting value Setting range: 0 to 3600s
[Factory adjusted as 50s]
(Auto-tuning Perform) [Factory adjusted as Auto-tuning Cancel]
the PID value reverts to the former value.
[Factory adjusted as 2.5%] This display is not available when the ON/OFF action is
selected in Control action designation (page 10).
Setting the value to 0 disables the function. This display is not available when the ON/OFF action or PD action is selected in Control action designation (page 10).
Setting the value to 0 disables the function. This display is not available when the ON/OFF action is selected in Control action designation (page 10).
Proportional cycle setting Mode to set the proportional cycle
Setting value Setting range: 1 to 120s
[Factory adjusted as 30s to Relay contact output type
and 3s to Non-contact voltage output type] This display is not available when the ON/OFF action is selected in Control action designation (page 10). This setting item is not available to Current output type.
With the Relay contact output type, if the proportional cycle time is decreased, the frequency of the relay action increases, and the life of the relay contact is shortened.
Manual reset setting Mode to set the reset value to correct the offset
Setting value (Deviation between SV and PV when equilibrium status).
Setting range: proportional band converted value [However, in the range, –199.9 to 999.9°C (°F)] [Factory adjusted as 0.0°C] This display is indicated only when PD action is designated in the Control action designation (page 10). Manual reset setting method: When SV>PV, set the positive + value (SV–PV). When SV<PV, set the negative – value (SV–PV).
V1.2 • 08/2004 - 15 -
Loading...
+ 32 hidden pages