7.Maintenance, cleaning and service (recalibration)21
7.1Maintenance21
7.2Cleaning21
7.3Recalibration21
8.Faults22
9.Return and disposal22
9.1Return22
9.2Disposal23
10.Accessories23
GB
11570351.02 10/2010 GB/D/F/E/I
3WIKA operating instructions, model CEP1000
1. General information
1. General information
■
GB
The loop calibrator described in the operating instructions has been manufactured
using state-of-the-art technology.
All components are subject to stringent quality and environmental criteria during
production. Our management systems are certied to ISO 9001 and ISO 14001.
■
These operating instructions contain important information on handling the instrument. Working safely requires that all safety instructions and work instructions are
observed.
■
Observe the relevant local accident prevention regulations and general safety regulations for the instrument's range of use.
■
The operating instructions are part of the instrument and must be kept in the immediate vicinity of the instrument and readily accessible to skilled personnel at any time.
■
Skilled personnel must have carefully read and understood the operating instructions, prior to beginning any work.
■
The manufacturer's liability is void in the case of any damage caused by using the
product contrary to its intended use, non-compliance with these operating instructions, assignment of insuciently qualied skilled personnel or unauthorised modications to the instrument.
■
The general terms and conditions, contained in the sales documentation, shall apply.
■
Subject to technical modications.
■
Factory calibrations/DKD/DAkkS calibrations are carried out in accordance with
international standards.
... indicates a potentially dangerous situation that can result in serious
injury or death, if not avoided.
CAUTION!
... indicates a potentially dangerous situation that can result in light
injuries or damage to equipment or the environment, if not avoided.
Information
... points out useful tips, recommendations and information for ecient
and trouble-free operation.
DANGER!
...identies hazards caused by electric power. Should the safety instructions not be observed, there is a risk of serious or fatal injury.
GB
WARNING!
Before installation, commissioning and operation, ensure that the
appropriate loop calibrator been selected in terms of measuring range,
design and specic measuring conditions.
Non-observance can result in serious injury and/or damage to equipment.
Further important safety instructions can be found in the individual
chapters of these operating instructions.
11570351.02 10/2010 GB/D/F/E/I
5WIKA operating instructions, model CEP1000
2. Safety
2.1 Intended use
The precision loop calibrator CEP1000 is designed for high-accuracy measuring and
simulation of current and voltage.
GB
The instrument has been designed and built solely for the intended use described here,
and may only be used accordingly.
The technical specications contained in these operating instructions must be
observed. Should the instrument be improperly handled or operated outside of its
technical specications, it has to be taken out of service immediately and inspected by
an authorised WIKA service engineer.
Handle electronic precision measuring instruments with the required care (protect from
humidity, impacts, strong magnetic elds, static electricity and extreme temperatures,
do not insert any objects into the instrument or its openings). Plugs and sockets must be
protected from contamination.
If the instrument is transported from a cold into a warm environment, the formation of
condensation may result in the instrument malfunctioning. Before putting it back into
operation, wait for the instrument temperature and the room temperature to equalise.
The manufacturer shall not be liable for claims of any type based on operation contrary
to the intended use.
2.2 Personnel qualication
WARNING!
Risk of injury should qualication be insucient!
Improper handling can result in considerable injury and damage to
equipment.
■
The activities described in these operating instructions may only
be carried out by skilled personnel who have the qualications
described below.
■
Keep unqualied personnel away from hazardous areas.
Skilled personnel
Skilled personnel are understood to be personnel who, based on their technical training,
knowledge of measurement and control technology and on their experience and knowl-
edge of country-specic regulations, current standards and directives, are capable of
carrying out the work described and independently recognising potential hazards.
Special operating conditions require further appropriate knowledge, e.g. of aggressive
media.
6WIKA operating instructions, model CEP1000
11570351.02 10/2010 GB/D/F/E/I
2. Safety
2.3 Special hazards
WARNING!
For hazardous media such as oxygen, acetylene, ammable or toxic
gases or liquids, and refrigeration plants, compressors, etc., in addition
to all standard regulations, the appropriate existing codes or regulations must also be followed.
WARNING!
Protection from electrostatic discharge (ESD) required.
The proper use of grounded work surfaces and personal wrist straps is
required when working with exposed circuitry (printed circuit boards),
in order to prevent static discharge from damaging sensitive electronic
components.
To ensure safe working on the instrument, the operating company must
ensure that the operating personnel are regularly instructed in all topics
regarding work safety, rst aid and environmental protection and knows
the operating instructions and, in particular, the safety instructions
contained therein.
DANGER!
Danger of death caused by electric current!
Operation using a defective power supply unit (e.g. short circuit from
the mains voltage to the output voltage) can result in life-threatening
voltages at the instrument!
GB
11570351.02 10/2010 GB/D/F/E/I
7WIKA operating instructions, model CEP1000
3. Specications
3. Specications
GB
Specications
Input
Current range0.000 … 24.000 mA
Voltage rangeDC 0.000 … 28.000 V
Input protectionFuseless, up to AC 250 V
Output
With HART™-resistor950 Ω
Loop supply voltageDC 24 V
Range selectDecade; incremental to 0.001 mA steps
Special featuresAuto step/ramp, Auto “%-Error” function, built-in resistor
Permissible
■
Operating temperature-10 … +55 °C
■
Storage temperature-20 … +70 °C
Display5 1/2-digit with character size 9 mm and %-announce-
RangemA, V, %
Recording optionsMIN-/MAX-storage
Power supplyDC 9 V battery
■
Battery life12 hours (12 mA into 250 Ω continuous)
■
Low battery indicatordisplayed icon near the end of battery life
EMC / CE conformity2004/108/EC, EN 61326 Emission (Group 1, Class B) and
Ingress protectionIP 52
Calibration3.1 calibration certicate per DIN EN 10204 (option: DKD/
Dimensions144.7 x 80.0 x 36.3 mm
Weightapprox. 340 g
-25.00 … +125.00 %
-25.00 … +125.00 %
for HART™ communications
ment
Immunity (portable equipment)
DAkkS-calibration certicate)
For further specications see WIKA data sheet CT 81.01 and the order documentation.
8WIKA operating instructions, model CEP1000
11570351.02 10/2010 GB/D/F/E/I
3. Specications / 4. Design and function
Dimensions in mm
Side viewFront view
GB
4. Design and function
4.1 Short description / Description
The CEP1000 is a loop calibrator operating in the range of 4 ... 20 mA designed for
both troubleshooting and calibration of circuits without separate power supply. Very
high accuracy and diverse special functions make the instrument to a user-friendly and
highly exible calibration instrument.
The CEP1000 can be used as current reference standard for up to 24 mA, it can also be
used to simulate a 2-wire connection or to perform measurements in circuits with up to
24 mA without separate power supply.
By means of the calibrator it is possible to measure process signal voltages of up to 28 V
direct current with an accuracy of 0.015 % of the measurement value.
11570351.02 10/2010 GB/D/F/E/I
9WIKA operating instructions, model CEP1000
4. Design and function
Using the digital knob user interface as well as the decade output control the user
can set both large and incremental changes of the output signal conveniently. Multiple
unique functions such as the automatic step/ramp function allow easy application of the
GB
CEP1000.
Besides the CEP1000 can supply a 2-wire transmitters with voltage and at the same
time measure the emerging current. All measurements are displayed in mA as well as in
percent of the measuring range.
4.2 Scope of delivery
■
Precision loop calibrator model CEP1000
■
Operating instructions
■
Neoprene case
■
Calibration certicate 3.1 per to DIN EN 10204
■
9 V battery
Cross-check scope of delivery with delivery note.
4.3 Front foil of the CEP1000
1) Mode indication
2) Step / ramp indication
3) On/O
4) Current loop key: Activation of
the 24 V power supply in the mA
measuring mode
5) %Error key: indicates the actual
vs. ideal error in %
6) Digital „knob“: output control
7) Simultaneous % display: 4 mA =
0.00 %, 20 mA = 100.00 %
8) Mode key
9) Auto key: Switching between
automatic step and ramp function
(slow or fast)
10) 25 %-key: manual activation of the
step function 4, 8, 12, 16, 20 mA
11) Cursor keys: arrow keys to select
the required decades
1
2
3
4
5
6
7
8
9
10
11
Figure 1
10WIKA operating instructions, model CEP1000
11570351.02 10/2010 GB/D/F/E/I
4. Design and function / 5. Transport, packaging and storage
4.4 Battery selection
The CEP1000 is operated with a normal 9 V alkaline battery or optionally with a
rechargeable 9 V Ni-Cd battery. For the most application cases the alkaline battery with
9 V is sucient. However, for the intensive use the 9 V Ni-Cd battery is the better choice.
The Ni-Cd battery for 9 V available optionally with an output current of 12 mA and full
battery enables continuous operation for approx. 3 hours (the alkaline battery lasts
approximately 12 hours). The optionally available charging device charges the battery
over night (10 to 12 hours long) and powers the instrument in the stationary mode. At
the same time the conservation charging of the Ni-Cd battery is carried out.
WARNING!
Never connect the power supply unit/charging device if a 9 V alkaline
cell is inserted.
4.5 Input protection
The device has an input protection without a fuse for up to 250 V of direct or alternating
current for the period of time of up to 30 seconds.
5. Transport, packaging and storage
5.1 Transport
Check loop calibrator for any damage that may have been caused by transport.
Obvious damage must be reported immediately.
GB
5.2 Packaging
Do not remove packaging until just before mounting.
Keep the packaging as it will provide optimum protection during transport (e.g. change
in installation site, sending for repair).
5.3 Storage
Permissible conditions at the place of storage:
■
Storage temperature: -20 ... +70 °C
■
Humidity: 0 ... 90 % relative humidity (no condensation)
Avoid exposure to the following factors:
■
Direct sunlight or proximity to hot objects
■
Mechanical vibration, mechanical shock (putting it down hard)
5. Transport, packaging and storage / 6. Commissioning ...
Store the instrument in its original packaging in a location that fulls the conditions
listed above. If the original packaging is not available, pack and store the instrument as
described below:
GB
1. Wrap the instrument in an antistatic plastic lm.
2. Place the instrument, along with shock-absorbent material, in the packaging.
3. If stored for a prolonged period of time (more than 30 days), place a bag, containing
a desiccant, inside the packaging.
WARNING!
Before storing the instrument (following operation), remove any residual
media. This is of particular importance if the medium is hazardous to
health, e.g. caustic, toxic, carcinogenic, radioactive, etc.
6. Commissioning, operation
DANGER!
To avoid electric shock and injuries:
■
Do not connect voltage higher than the nominal voltage. For the
supported voltage ranges refer to "3. Specications."
■
Observe all safety measures for the instruments.
■
Do not use the calibrator if it is damaged. Prior to use check the
housing of the calibrator. Check carefully if plastic parts are missing
and if there are any cracks. Pay special attention to insulation
around connections.
■
Select the proper function and correct measuring range.
■
The battery case must be closed and locked before the calibrator is
put into operation.
■
Disconnect the test leads from the calibrator prior to opening the
battery case.
■
Check the test leads for signs of damaged insulation or metal parts
with natural nish.
■
When using the probe keep the ngers away from the probe
contacts. Hold the probes behind the nger protection at the test
leads. Do not use the calibrator in case of malfunction. The instrument protection can be impaired. In case of doubt the instrument
must be repaired.
■
Do not operate the instrument in the environment with explosive
gases, vapours or dusts.
■
Disconnect the capillary prior to switching to another measurement
or before using as a simulator.
■
Use only the specied spare parts for repairs of the calibrator.
12WIKA operating instructions, model CEP1000
11570351.02 10/2010 GB/D/F/E/I
6. Commissioning, operation
■
To avoid faulty measurements that can lead to electric shock or
injuries, replace the battery immediately as soon as a dead battery
is indicated.
WARNING!
To avoid damages of the calibrator or devices during testing, use the
correct terminals, functions and measuring ranges for measurement or
simulation.
6.1. First steps
6.1.1 Battery saving function
To avoid discharge of the battery in case of unintended switching on of the CEP1000,
the battery saving function is activated ex works and set to 30 minutes. The battery
saving function turns o the CEP1000 automatically if no key is pressed within 30
minutes.
To change the turn-o time or deactivate it, the following steps must be carried out when
turning on:
1) Turn o the instrument.
2) Press the "AUTO" key.
3) Turn on the instrument again.
4) After turning on wait for 1 second and then release the "AUTO" key.
5) The set period of time appears on the display for the battery saving function
(between "O" and 30 minutes).
6) Set the required time using the knob (between "O" and 30 minutes).
7) Press the "AUTO" key again to save the value and switch the instrument back to
normal operation. The saved value is stored in the memory until it is changed by the
operator.
GB
11570351.02 10/2010 GB/D/F/E/I
13WIKA operating instructions, model CEP1000
6. Commissioning, operation
6.1.2 HART™ resistance conguration
Due to internal 250
Communicator during calibration or installation of a HART™ transmitter. The internal
GB
resistance replaces the resistance of 250
model 275. To activate this function if necessary, the internal jumper must be removed.
The instrument is supplied with an inserted jumper (see gure 2). To activate the resistance, remove the jumper. The jumper can be either stored outside of the instrument or
just plugged onto a contact pin in the instrument, thus, the circuit is interrupted.
Ω resistance of the CEP1000 it is possible to use Rosemount 275
Ω required normally for the communicator
-J8 (HART™ JUMPER)
Remove the jumper to activate the HART™
resistance (250 Ω).
LCD
CEP1000
Figure 2
If the 250
maximum load of 950
Ω resistance is activated, it is possible to energise a
Ω instead of 1200 Ω. In most cases this would not
cause any problems for the calibration of products incompatible with
HART™.
14WIKA operating instructions, model CEP1000
11570351.02 10/2010 GB/D/F/E/I
6. Commissioning, operation
6.2 Basic operating modes
The CEP1000 has 4 operating modes for the mA range and one operating mode for
voltage measurement. The following text and gures explain dierent operating modes.
6.2.1 Milliampere source
Figure 3 shows a typical application, for which the CEP1000 is used as a source of
current for the range from 4 ... 20 mA for the instrument to be tested.
( + )
GB
Instrument with an input
( - )
of 4 ... 20 mA
SOURCE
4 ... 20 mA
CEP 1000
mA
Max. load: 1200 Ω without HART™
950 Ω with HART™
MAX
MODE
AUTO
MIN
%ERROR
LOOPPWR
Calibration Line
25%
Figure 3 - Current source for 4 ... 20 mA
1) Turn on the instrument. As a standard the operating mode source of current in the
mA range is set.
2) Connect the instrument to be calibrated.
For the instrument to function as a source of current, the circuit must be
closed. In case of an interrupted circuit the LCD display "OL" is blinking
and indicating an overload or an interruption of the circuit.
11570351.02 10/2010 GB/D/F/E/I
15WIKA operating instructions, model CEP1000
6. Commissioning, operation
3) The output signal can be adjusted to the set point by means of the knob. For this
purpose select the decade to be set using the arrow keys.
4) The output signal can also be increased incrementally in steps of 4 mA (25 %) using
GB
the "25%" key.
5) The automatic step function and automatic ramp function allow automatic change of
the output signal.
For further information refer to the section "6.3.1 Automatic operation modes" of the
instruction.
6.2.2 Milliampere simulation
Figure 4 shows the way the CEP1000 is connected if it is supposed to function as a
2-wire transmitter with an external power supply. In this case the circuit is powered via
an external voltage source and the instrument just regulates the current in the circuit.
The regulating procedure of the output signal is the same as for the application as the
source of current in the milliampere range.
( + )
Power supply DC 30 V
( - )
( + )
Instrument with an input
( - )
of 4 ... 20 mA
max.
SIM
LOOPPWR
Calibration Line
MODE
%ERROR
CEP 1000
mA
MAX
AUTO
MIN
25%
4 ... 20 mA
The instrument is operated in the simulation mode as a
two-wire sensor.
Figure 4: Simulation of a sensor with a current output for 4 ... 20 mA
16WIKA operating instructions, model CEP1000
11570351.02 10/2010 GB/D/F/E/I
6. Commissioning, operation
6.2.3 Milliampere measurement
Figure 5 shows the way the CEP1000 must be connected to carry out a measurement in
a circuit for 4 ... 20 mA. During this application the instrument just measures the current
in the circuit and displays it on the main display in
indicates the percentage of current for the measuring range 4 ... 20 mA.
6.2.4 Milliampere measurement with 24 V power supply
For this case gure 5 also shows the correct connection. However, in this case a 2-wire
transmitter is tested, it requires an external power supply for the output current. For this
application the instrument supplies the circuit with an operating voltage of 24 V and at
the same time measures the emerging current. To enable the function with separate
power supply, press the "LOOP PWR" key if the option milliampere measurement is
selected.
mA
. The lower (smaller) scale range
( + )
GB
Instrument with an
output for 4 ... 20 mA
MEASURE
CEP 1000
mA
4 ... 20 mA
( - )
Maximum input signal: 24 mA
LOOPPWR
Calibration Line
MODE
%ERROR
AUTO
MAX
MIN
25%
To measure the output signal of a two-wire sensor use
the option measurement with separate power supply.
Figure 5: Measurement in the range of 4 ... 20 mA
The measuring mode with separate power supply can be activated for
the option measuring in the milliampere range.
11570351.02 10/2010 GB/D/F/E/I
17WIKA operating instructions, model CEP1000
6. Commissioning, operation
6.2.5 Voltage measurement
Figure 6 shows the connection of CEP1000 required to measure direct voltage of up
to 28 V. To activate the voltage measurement, just press the "MODE" key to search the
GB
individual functions until the voltage measurement is displayed.
The input impedance of the CEP1000 is higher than 1 MΩ in this mode.
( + )
instrument with an output
for 0 ... 28 V
MEASURE
LOOPPWR
MODE
%ERROR
CEP 1000
V
MAX
AUTO
MIN
25%
( - )
Maximum input signal: DC 28 V
Calibration Line
Figure 6: Measurement in the range of DC 0 ... 28 V
18WIKA operating instructions, model CEP1000
11570351.02 10/2010 GB/D/F/E/I
6. Commissioning, operation
6.3 Advanced operating modes
6.3.1 Automatic step function/automatic ramp mode
The CEP1000 is capable of increasing the output current in the mA range in steps of
25 % automatically with intervals of 5 seconds. The output current in the mA range
can be increased linearly in form of the ramp function from 4 mA up to 20 mA and then
reduced again automatically to 4 mA.
To activate the automatic step/ramp function proceed as follows:
1) The automatic step/ramp function can be used for the application as milliampere
source as well as for the milliampere simulation. Switch the instrument into the corresponding operating mode.
2) Connect the CEP1000 to the instrument to be tested.
3) For the automatic step function press the "AUTO" key once, for a slow ramp function
twice and for the quick ramp function three times. After every key stroke a small
conrmation window appears in the lower left corner of the LCD display, it indicates
which step function or ramp function has been selected.
4) As soon as the step function/ramp function has been activated, it is functioning
constantly until another key is pressed.
6.3.2 Fault percentage function
This unique function calculates the fault percentage of the measuring range (a measur-
ing range of 4 ... 20 mA gives a span of 16 mA). Using the function for calculation of fault
percentage the user can "Zero Out" the milliampere indication, thus, enabling indication
of deviation from the ideal value on the display as a fault percentage.
Procedure:
1) Connect the CEP1000 to the instrument to be tested.
2) Set a mode with milliampere output for the instrument (source or simulation).
3) Set the required output value.
4) If the instrument to be tested does not measure the exact value, press and hold
the "%ERROR" key and adjust the knob so that the instrument displays the correct
value. Press and keep holding the “%ERROR” key further on.
5) The indication in the percentage eld corresponds to the fault or deviation as a
percentage of the span.
6) After the "%ERROR" key has been released, the indication switches back to normal
operation.
GB
11570351.02 10/2010 GB/D/F/E/I
19WIKA operating instructions, model CEP1000
6. Commissioning, operation
6.3.3 Data acquisition for minimum and maximum
If the CEP1000 is operated in a measuring mode (milliampere, milliampere with power
supply or voltage), the minimum and maximum values for the corresponding input mode
GB
are monitored and stored permanently. To activate the data acquisition proceed as
follows:
1) Switch the CEP1000 to a measuring mode.
2) Connect the instrument to the instrument to be tested.
3) Press both keys (“MIN” and “MAX”) to clear the memory for the minimum and
maximum values. The “CLR” message is displayed on the LCD display, if the
memory has been deleted.
4) Acquire the minimum value and the maximum value using the instrument as long as
required.
If necessary, deactivate the battery saving function prior to calling the
data acquisition function to avoid premature switching o of the calibrator.
5) The saved value can be accessed any time by pressing the "MIN" or "MAX" key.
6) When switching o the instrument or switching to another operating mode the
memory will be cleared. Back up the data before switching o.
20WIKA operating instructions, model CEP1000
11570351.02 10/2010 GB/D/F/E/I
7. Maintenance, cleaning and service (recalibration)
7. Maintenance, cleaning and service (recalibration)
7.1 Maintenance
The precision loop calibrator CEP1000 is maintenance-free.
Repairs must only be carried out by the manufacturer. This does not apply to the battery
replacement.
7.2 Cleaning
CAUTION!
■
Clean the instrument with a moist cloth.
■
Electrical connections must not come into contact with moisture.
For information on returning the instrument see chapter "9.1 Return".
7.3 Recalibration
DKD/DAkkS certicate - Certicates:
We recommend that the instrument is regularly recalibrated by the manufacturer, with
time intervals of approx. 12 months. Every factory recalibration includes, additionally, an
extensive free-of-charge check of all system parameters with respect to their compli-
ance with the specication. The basic settings will be corrected if necessary.
GB
11570351.02 10/2010 GB/D/F/E/I
21WIKA operating instructions, model CEP1000
8. Faults / 9. Return and disposal
8. Faults
GB
Indication Type of faultMeasures
Battery is low, functioning is only
guaranteed for a short period of time
OL
-OL
No display or
undenable
characters, the
instrument does
not respond to
key press
Reading is signicantly above or
below the measuring range
Battery is dead.Insert new batteries
System errorTurn o the CEP1000, wait for a
The CEP1000 is defectiveSend in for repair
CAUTION!
If faults cannot be eliminated by means of the measures listed above,
the instrument must be shut down immediately, and it must be ensured
that pressure and/or signal are no longer present, and it must be
prevented from being inadvertently put back into service.
In this case, contact the manufacturer.
If a return is needed, please follow the instructions given in chapter
"9.1 Return".
9. Return and disposal
Insert new batteries
Test: if the measurement parameter
is within the approved measuring
range of the instrument?
short period of time, switch on again
WARNING!
Residual media in dismounted loop calibrator can result in a risk to
persons, the environment and equipment. Take sucient precautionary
measures.
9.1 Return
WARNING!
Strictly observe when shipping the instrument:
All instruments delivered to WIKA must be free from any kind of hazard-
ous substances (acids, bases, solutions, etc.).
22WIKA operating instructions, model CEP1000
11570351.02 10/2010 GB/D/F/E/I
9. Return and disposal / 10. Accessories
When returning the loop calibrator, use the original packaging or a suitable transport
package.
To avoid damage:
1. Wrap the instrument in an antistatic plastic lm.
2. Place the instrument, along with the shock-absorbent material, in the packaging.
Place shock-absorbent material evenly on all sides of the transport packaging.
3. If possible, place a bag, containing a desiccant, inside the packaging.
4. Label the shipment as transport of a highly sensitive measuring instrument.
Enclose the completed return form with the instrument.
The return form is available on the internet:
www.wika.de / Service / Return
9.2 Disposal
Incorrect disposal can put the environment at risk.
Dispose of instrument components and packaging materials in an environmentally
compatible way and in accordance with the country-specic waste disposal regulations.
This marking on the instruments indicates that they must not be
disposed of in domestic waste. The disposal is carried out by return to
the manufacturer or by the corresponding municipal authorities. See EU
regulation 2002/96/EC.
GB
10. Accessories
For accessories such as 9 V Ni-Cd battery (rechargeable), AC power supply unit/charging device for Europe, Japan, the UK or USA etc. refer to the current WIKA price list:
Testing and calibration technology.
11570351.02 10/2010 GB/D/F/E/I
23WIKA operating instructions, model CEP1000
GB
24WIKA operating instructions, model CEP1000
11570351.02 10/2010 GB/D/F/E/I
Inhalt
Inhalt
1.Allgemeines26
2.Sicherheit27
2.1Bestimmungsgemäße Verwendung28
2.2Personalqualikation28
2.3Besondere Gefahren29
3.Technische Daten30
4.Aufbau und Funktion31
4.1Kurzbeschreibung / Beschreibung31
4.2Lieferumfang32
4.3Frontfolie des CEP100032
4.4Batterieauswahl 33
4.5Eingangsschutz33
5.Transport, Verpackung und Lagerung33
5.1Transport33
5.2Verpackung33
5.3Lagerung33
6.Inbetriebnahme, Betrieb34
6.1Erste Schritte35
6.1.1Batteriesparfunktion35
6.1.2HART™ Widerstandskonguration36
6.2Grundlegende Betriebsarten37
6.2.1Milliampere-Quelle 37
6.2.2Milliampere-Simulation38
6.2.3Milliampere-Messung39
6.2.4Milliampere-Messung mit 24 V Spannungsversorgung39
7.Wartung, Reinigung und Service (Rekalibrierung)43
7.1Wartung43
7.2Reinigung43
7.3Rekalibrierung43
8.Störungen44
9.Rücksendung und Entsorgung44
9.2Rücksendung44
9.3Entsorgung45
10.Zubehör45
D
11570351.02 10/2010 GB/D/F/E/I
25WIKA Betriebsanleitung, Typ CEP1000
1. Allgemeines
1. Allgemeines
■
Der in der Betriebsanleitung beschriebene Stromschleifenkalibrator wird nach den
neuesten Erkenntnissen gefertigt.
Alle Komponenten unterliegen während der Fertigung strengen Qualitäts- und
D
Umweltkriterien. Unsere Managementsysteme sind nach ISO 9001 und ISO 14001
zertiziert.
■
Diese Betriebsanleitung gibt wichtige Hinweise zum Umgang mit dem Gerät. Voraussetzung für sicheres Arbeiten ist die Einhaltung aller angegebenen Sicherheitshinweise und Handlungsanweisungen.
■
Die für den Einsatzbereich des Gerätes geltenden örtlichen Unfallverhütungsvorschriften und allgemeinen Sicherheitsbestimmungen einhalten.
■
Die Betriebsanleitung ist Produktbestandteil und muss in unmittelbarer Nähe des
Gerätes für das Fachpersonal jederzeit zugänglich aufbewahrt werden.
■
Das Fachpersonal muss die Betriebsanleitung vor Beginn aller Arbeiten sorgfältig
durchgelesen und verstanden haben.
■
Die Haftung des Herstellers erlischt bei Schäden durch bestimmungswidrige Verwen-
dung, Nichtbeachten dieser Betriebsanleitung, Einsatz ungenügend qualizierten
Fachpersonals sowie eigenmächtiger Veränderung am Gerät.
■
Es gelten die allgemeinen Geschäftsbedingungen in den Verkaufsunterlagen.
■
Technische Änderungen vorbehalten.
■
Werkskalibrierungen/DKD/DAkkS-Kalibrierungen erfolgen nach internationalen
Normen.
■
Weitere Informationen:
- Internet-Adresse:www.wika.de / www.wika.com
- zugehöriges Datenblatt:
- Anwendungsberater:
CT 81.01
Tel.: (+49) 9372/132-9986
Fax: (+49) 9372/132-217
E-Mail:
testequip@wika.de
26WIKA Betriebsanleitung, Typ CEP1000
11570351.02 10/2010 GB/D/F/E/I
1. Allgemeines / 2. Sicherheit
Symbolerklärung
WARNUNG!
… weist auf eine möglicherweise gefährliche Situation hin, die zum Tod
oder zu schweren Verletzungen führen kann, wenn sie nicht gemieden
wird.
VORSICHT!
… weist auf eine möglicherweise gefährliche Situation hin, die zu
geringfügigen oder leichten Verletzungen bzw. Sach- und Umweltschäden führen kann, wenn sie nicht gemieden wird.
Information
… hebt nützliche Tipps und Empfehlungen sowie Informationen für
einen ezienten und störungsfreien Betrieb hervor.
GEFAHR!
…kennzeichnet Gefährdungen durch elektrischen Strom. Bei Nichtbeachtung der Sicherheitshinweise besteht die Gefahr schwerer oder
tödlicher Verletzungen.
D
2. Sicherheit
WARNUNG!
Vor Montage, Inbetriebnahme und Betrieb sicherstellen, dass der richtige Stromschleifenkalibrator hinsichtlich Messbereich, Ausführung und
spezischen Messbedingungen ausgewählt wurde.
Bei Nichtbeachten können schwere Körperverletzungen und/oder
Sachschäden auftreten.
Weitere wichtige Sicherheitshinweise benden sich in den einzelnen
Kapiteln dieser Betriebsanleitung.
11570351.02 10/2010 GB/D/F/E/I
27WIKA Betriebsanleitung, Typ CEP1000
2. Sicherheit
2.1 Bestimmungsgemäße Verwendung
Dieser Präzise Stromschleifenkalibrator CEP1000 dient zum genauen messen und
simulieren von Strom und Spannung.
Das Gerät ist ausschließlich für den hier beschriebenen bestimmungsgemäßen
Verwendungszweck konzipiert und konstruiert und darf nur dementsprechend verwen-
D
det werden.
Die technischen Spezikationen in dieser Betriebsanleitung sind einzuhalten. Eine
unsachgemäße Handhabung oder ein Betreiben des Gerätes außerhalb der techni-
schen Spezikationen macht die sofortige Stilllegung und Überprüfung durch einen
Elektronische Präzisionsmessgeräte mit erforderlicher Sorgfalt behandeln (vor Nässe,
Stößen, starken Magnetfeldern, statische Elektrizität und extremen Temperaturen schüt-
zen, keine Gegenstände in das Gerät bzw. Önungen einführen). Stecker und Buchsen
vor Verschmutzung schützen.
Wird das Gerät von einer kalten in eine warme Umgebung transportiert, so kann durch
Kondensatbildung eine Störung der Gerätefunktion eintreten. Vor einer erneuten
Inbetriebnahme die Angleichung der Gerätetemperatur an die Raumtemperatur abwarten.
Ansprüche jeglicher Art aufgrund von nicht bestimmungsgemäßer Verwendung sind
ausgeschlossen.
2.2 Personalqualikation
WARNUNG!
Verletzungsgefahr bei unzureichender Qualikation!
Unsachgemäßer Umgang kann zu erheblichen Personen- und
Sachschäden führen.
■
Die in dieser Betriebsanleitung beschriebenen Tätigkeiten nur
durch Fachpersonal nachfolgend beschriebener Qualikation
durchführen lassen.
■
Unqualiziertes Personal von den Gefahrenbereichen fernhalten.
Fachpersonal
Das Fachpersonal ist aufgrund seiner fachlichen Ausbildung, seiner Kenntnisse der
Mess- und Regelungstechnik und seiner Erfahrungen sowie Kenntnis der landesspezi-
schen Vorschriften, geltenden Normen und Richtlinien in der Lage, die beschriebenen
Arbeiten auszuführen und mögliche Gefahren selbstständig zu erkennen.
28WIKA Betriebsanleitung, Typ CEP1000
11570351.02 10/2010 GB/D/F/E/I
2. Sicherheit
Spezielle Einsatzbedingungen verlangen weiteres entsprechendes Wissen, z. B. über
aggressive Medien.
2.3 Besondere Gefahren
WARNUNG!
Bei gefährlichen Messstoen wie z. B. Sauersto, Acetylen, brennbaren oder giftigen Stoen, sowie bei Kälteanlagen, Kompressoren etc.
müssen über die gesamten allgemeinen Regeln hinaus die einschlägigen Vorschriften beachtet werden.
WARNUNG!
Schutz vor elektrostatischer Entladung (ESD) erforderlich!
Die ordnungsgemäße Verwendung geerdeter Arbeitsächen und
persönlicher Armbänder ist bei Arbeiten mit oenen Schaltkreisen
(Leiterplatten) erforderlich, um die Beschädigung empndlicher elektro-
nischer Bauteile durch elektrostatische Entladung zu vermeiden.
Für ein sicheres Arbeiten am Gerät muss der Betreiber sicherstellen,
dass das Bedienpersonal regelmäßig in allen zutreenden Fragen
von Arbeitssicherheit, Erste-Hilfe und Umweltschutz unterwiesen wird,
sowie die Betriebsanleitung und insbesondere die darin enthaltenen
Sicherheitshinweise kennt.
D
GEFAHR!
Lebensgefahr durch elektrischen Strom!
Bei Betrieb mit einem defekten Netzgerät (z. B. Kurzschluss von
Netzspannung zur Ausgangsspannung) können am Gerät lebens-
gefährliche Spannungen auftreten!
11570351.02 10/2010 GB/D/F/E/I
29WIKA Betriebsanleitung, Typ CEP1000
3. Technische Daten
3. Technische Daten
Technische Daten
Eingang
D
Strombereich0,000 … 24,000 mA
SpannungsbereichDC 0,000 … 28,000 V
EingangsschutzSicherungslos bis AC 250 V
Ausgang
Strombereich0,000 … 24,000 mA
Messunsicherheit± 0,015 % des Messwerts ± 2 µA
Auösung1 µA, 1 mV
Maximale Last
■
Ohne HART™-Widerstand
■
Mit HART™-Widerstand
Versorgungs-Spannungsausgang
BereichsauswahlZehnerstellen; Schrittweise in 0,001 mA Schritten
BesonderheitenStufen- und Rampenfunktion, Automatische „%-Fehler“-
Zulässige
■
Betriebstemperatur
■
Lagertemperatur
Display5 1/2-stellig mit 9 mm Ziernhöhe und %-Anzeige
AnzeigemA, V, %
AufzeichnungsoptionenMIN-/MAX-Speicher
HilfsenergieDC 9 V Batterie
■
Batterielebensdauer
■
Batteriestandsanzeige
EMV/CE-Zulassung2004/108/EC, EN 61326 Emission (Gruppe 1, Klasse B)
SchutzartIP 52
KalibrierungWerkskalibrierschein 3.1 nach DIN EN 10204 (optional:
Abmessungen144,7 x 80,0 x 36,3 mm
Gewichtca. 340 g
-25,00 … +125,00 %
-25,00 … +125,00 %
1200 Ω
950 Ω
DC 24 V
Funktion, eingebauter Widerstand für HART™ Kommunikation
-10 … +55 °C
-20 … +70 °C
12 Stunden (12 mA / 250 Ω Dauerbelastung)
Symbolanzeige bei schwacher Batterie
Störfestigkeit (tragbares Gerät)
DKD/DAkkS Kalibrierschein)
Weitere technische Daten siehe WIKA Datenblatt CT 81.01 und Bestellunterlagen.
30WIKA Betriebsanleitung, Typ CEP1000
11570351.02 10/2010 GB/D/F/E/I
Loading...
+ 82 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.