WIKA BWX User Manual

Page 1
A
A
INSTRUCTION MANUAL
Rev. 1 12/02
BOURDON TUBE PRESSURE SWITCHES
WEATHERPROOF: SERIES BW EXPLOSIONPROOF: SERIES BAX
B
NI-222E
B
A pressure attachment B cable entry WEIGHT 2 kg dimensions in mm WEIGHT 3,7 kg dimensions in mm
NOTE: dimensions and weights are not binding unless released on certified drawings.
CAUTION
Before installing, using or carrying out maintenance on the instrument it is necessary to read and understand the indications given in the attached Instruction Manual.
The instrument must only be installed and maintained by qualified personnel.
INSTALLATION IS TO BE CARRIED OUT ONLY AFTER CHECKING THAT INSTRUMENT CHARACTERISTICS
ARE CONSISTENT WITH PROCESS AND PLANT REQUIREMENTS.
The functional features of the instrument and its degree of protection are shown on the identification plate fixed to the case.
CONTENTS:
1 - GENERAL NOTES 2 - OPERATING PRINCIPLE 3 - IDENTIFICATION PLATE AND MARKINGS 4 - SET POINT REGULATION 5 - SET POINT CALIBRATION 6 - MOUNTING AND CONNECTIONS 7 - INSTRUMENT PLUMBING 8 - PUTTING INTO OPERATION 9 - FUNCTIONAL VERIFICATION 10 - TROUBLESHOOTING 11 - STOPPING AND DISMOUNTING 12 - DEMOLITION
SAFETY INSTRUCTIONS FOR USE IN HAZARDOUS ATMOSPHERES. RECOMMENDATIONS FOR PRESSURE SWITCH SAFE USE.
A pressure attachment B cable entry
RELATED DOCUMENT
To authentified document with certificate
N° CESI 03 ATEX 005
All data, statements and recommendations supplied with this manual are based on information believed by us to be
reliable. As the conditions of effective use are beyond our control, our products are sold under the condition that the user himself evaluates such conditions before following our recommendations for the purpose or use foreseen by him.
Page 2
NI-222E
INSTRUCTION MANUAL
Rev. 1 12/02
1 - GENERAL NOTES
1.1 FOREWORD
The wrong choice of a series or a model, as well as the incorrect installation, lead to malfunction and reduce instrument life. Failure to abide by the indications given in this manual can cause damage to the instrument, the environment and persons.
1.2 ALLOWED OVERRANGE
Pressure exceeding the working range can be occa­sionally tolerated provided they remain within the limits
stated in the instrument features (vacuum or proof pres­sure). Continuous pressures exceeding the working range can be applied to the instrument provided they are clearly stated in the instrument features. The current and voltage values stated in the technical specifications and ratings must not be exceeded. Transitory overranges can have a destructive effect on the switch.
1.3 MECHANICAL VIBRATIONS
Can generally lead to the wearing of some parts of the instrument or cause false actuation. It is therefore rec­ommended that the instrument be installed in a place where there are no vibrations. In cases where this is impossible it is a good idea to take measures to lessen the effects (elastic supports, installation with the peg of the microswitch positioned at right angles to the vibration plane, etc.).
1.4 TEMPERATURE
Due to the temperature of both the environment and the process fluid, the temperature of the instrument could exceed the allowed limits (normally from -20° to +70°C). Therefore, in case it does, suitable measures (protection against heat radiation, fluid separators, cooling coils, heated lockers) must be taken.
2 - OPERATING PRINCIPLE
The pressure, acting on the sensitive element (Bourdon tube), determines its elastic deformation which is used to activate one or two electric microswitches regulated at set point values. The microswitches are of the rapid release type with automatic rearm. When the pressure moves away from the set values, returning towards the normal values, the switch is rearmed. The differential gap (difference between the set point value and the reset value) can be set or adjustable (letter R in the contact codes).
3 - IDENTIFICATION PLATE AND MARKINGS
The instrument is fitted with a metal plate bearing all its functional characteristics and – in case of explosionproof execution (Series BAX) – also the markings prescribed by standard CEI EN 50014. Fig. 2 shows the plate mounted on explosionproof instruments.
Fig. 1 - Explosionproof instruments plate
1 CE marking and identification number of the notified body
responsible for production surveillance.
2 Apparatus classification according to ATEX 94/9 CE direc-
tive.
3 Type of protection and ambient temperature limits of opera-
tion.
4 Notified body that issued the type certificate and number of
said certificate.
4 - SET POINT REGULATION
Each microswitch is independent and can be regulated by means of a screw (for adjustment) in such a way that it is released when the pressure reaches (increasing or decreasing) the desired value (set point). The instrument is usually supplied with the switches set at the setting range value nearest to zero (factory cali- bration). The instrument is supplied with an adhesive rating plate showing the set point calibration value. With factory calibration the values are not indicated on the ratings as these are temporary and will be modified with the defini­tive values. Prior to installation the instrument must be calibrated and the definitive calibration values written on the adhe­sive rating plate using a suitable indelible ink pen.
If the instrument has been ordered with a specific cali- bration, it is a good rule to check the calibration values marked on the relevant adhesive label, prior to installa­tion.
The position of the adjustment screws is given in figure The effect that the direction of rotation of the adjustment screws has is described on the adhesive plate.
Page 3
NI-222E
INSTRUCTION MANUAL
Rev. 1 12/02
Fig. 2 -
One contact instruments
1 - Microswitch set point calibration screws 3 - Terminal block 4 - Electrical connection identification plate
5 - SET POINT CALIBRATION
In order to proceed with the calibration and the periodical functional verification of the instrument a suitable cali- bration circuit (fig. 5) and an adequate pressure source is required.
5.1 PRELIMINARY OPERATIONS
Weatherproof pressure switches (BWX)
Remove the blocking device fixed to the side of the in­strument case and the adjustment screw access plate (Fig. 3). Remove the cover by rotating it in an anticlock­wise direction.
Fig. 3 - Weatherproof pressure switch blocking device -
Electrical connections and adjustment screws
Two contact instruments
1 - Microswitch 2 set point calibration screws 2 - Microswitch 1 set point calibration screws 3 - Terminal block 4 - Electrical connection identification plate
Explosionproof pressure switches (BAX) (Fig. 4)
CAUTION: do not open the cover of explosionproof
pressure switches (Series BAX) when energized, in explosive atmospheres.
Loosen the locking headless screw situated on the cover using a 1,5 hexagonal key then unscrew the cover. Remove the internal blocking device inserted on the closure plugs and slide out the plugs.
Fig. 4 -
Explosionproof pressure switch blocking device
Microswitch electrical circuit
State of the contacts at at­mospheric pressure
Designation of the contacts: C - common NA - Normally open NC - Normally closed
a - Plumbing wire b - Plumbing c - Blocking nut
d - Blocking bracket e - Adjustment screw access plate
5.2 CALIBRATION CIRCUIT AND OPERATIONS
Prepare the control circuit as indicated in Fig.5. The warning lamps should be connected to contact 1 or 2 in the NO or NC position according to the required contact action.
Connection of C and NO terminals
If the circuit is open at the working pressure, the switch closes the circuit as the pressure increases when the desired value is reached.
If the circuit is closed at the working pressure, the switch opens the circuit as the pressure decreases when the desired value is reached.
Connection of C and NC terminals
If the circuit is closed at the working pressure, the switch opens the circuit as the pressure increases when the desired value is reached.
Page 4
If the circuit is open at the working pressure, the
switch closes the circuit as the pressure decreases when the desired value is reached. The test instrument should have a measurement range approximately equal to or slightly wider than the pressure switch range and should have an accuracy consistent with the precision required to calibrate the set point. The pressure switch must be mounted in the normal installation position, i.e. with the pressure connection pointing downwards.
Fig. 5 -
PS - Pressure switch CA - Test pressure gauge V1 - Inlet valve V2 - Inlet valve P - Pressure source
Avoid forcing the elastic support of the microswitch by hand or with tools. This could affect the instrument func­tioning.
CAUTION: if the switch is of the kind with adjustable dead band (letter R in the contact codes) before pro­ceeding with the following operations it is necessary to proceed with the adjustment of the dead band (see at­tachment NI-705 for model BWXUR and NI-704 for model BAXUR)
Increase the pressure in the circuit up to the desired set point value for the first microswitch. Use a wide bladed screwdriver, as indicated on the ad­hesive plate, until the relative lamp turns on (or turns off).
- If the instrument is equipped with only one contact the
calibration is complete.
- If it is equipped with two contacts continue in the fol-
lowing manner. Vary the pressure until the desired set point value for the second microswitch is reached. Act on the adjustment screw of the second contact. Repeat former operations on the first contact, then op­erations on the second contact, until the required set point precision is obtained. This is necessary due to the reciprocal influence which the microswitches have on the sensitive element of the instrument. Check the calibration values (varying the pressure in the circuit accordingly) and record them on the adhesive plate using a pen with indelible ink.
5.3 FINAL OPERATIONS
Disconnect the instrument from the calibration circuit
Weatherproof pressure switches (Series BWX)
Take the cover, ensure that the sealing gasket is cor­rectly fitted into its seat, and insert the cover onto the case, with the blocking gap positioned in correspon­dence to the blocking bracket. Turn the cover clockwise closing it tightly. Mount the adjustment screw access plate, then the blocking device as in Fig. 3.
Calibration circuit
INSTRUCTION MANUAL
Test fluid: air for P 10 bar water for P > 10 bar
NI-222E
Rev. 1 12/02
Explosionproof pressure switches (Series BAX).
Insert the closure plugs of the adjustment screw access holes, block them using the internal device and if nec­essary seal them with plumbing. Screw on the cover and block it using the headless screw with which it is equipped (Fig. 4).
Replace the supplied protection caps on the pressure attachment and cable conduit.
IMPORTANT: the protection caps should only be defini­tively removed during the connection steps (see §6).
6 - MOUNTING AND CONNECTIONS
6.1 MOUNTING Surface mount the instrument by means of the holes
provided, or pipe mount using the appropriate bracket (see Fig. 9). The chosen position must be such that vibrations, the possibility of shocks or temperature changes are within tolerable limits. The above also ap­plies to direct mounting. With gas or vapour process fluid, the instrument must be positioned higher than the pipe inlet (see Fig. 8). With a liquid process fluid, the instrument can be positioned higher or lower, indiffer­ently (see Fig. 7 e 8). ). In this case, during set point calibration the negative or positive head must be taken into account (distance h in fig.7 and 8).
6.2 PRESSURE CONNECTIONS
For a correct installation it is necessary to: Mount a shut-off valve with drain (root valve) on the process tube to allow the instrument to be excluded and the connection tubing to be drained. It is recommended that said valve has a capstan-blocking device aimed at preventing it being activated casually and without au­thorisation. Mount a service valve near the instrument to permit possible functional verification on site. It is recom­mended that the service valve is closed with a plug to prevent the outlet of the process fluid caused by the incorrect use of said valve. Mount a three-piece joint onto the threaded attachment of the instrument to permit the easy mounting or removal of the instrument itself. Carry out the connection using a flexible tube in such a way that variations in the temperature of the tube itself do not force the instrument attachment. Ensure that all the pressure connections are airtight. It is important that there are no leakage in the circuit. Close the root valve and the relative drain device. Close the service valve using a safety plug.
6.3 ELECTRICAL CONNECTIONS
It is recommended to carry out the electrical connections according to the applicable standards. In case of explo­sionproof instruments (Series BAX) see also the Stan­dard EN-60079-14. If the electrical connection is carried out in a protected tube, it shall be made so that conden­sate is prevented from entering instrument enclosure.
The arrangement shown in Fig. 7 or 8 is therefore rec­ommended.
Page 5
INSTRUCTION MANUAL
CAUTION: fittings used for the electrical connection of
the pressure switch Series BAX (explosionproof) shall be certified to Standards EN 50014 and 50018, and shall guarantee instrument degree of protection (IP65).
Check that there is no power in the lines. Remove the cover and carry out the cabling and connec­tions to the terminal block (see Fig. 2). Flexible cables with a maximum section of 1.2 mm (16AWG) are recommended using the pre-insulated fork thimbles. Do not touch the adjustment screws and do not bend the elastic microswitch supports in order to prevent the instrument calibration being altered. Ensure that no deposits or wire ends remain inside the case.
As soon as connection steps are completed, mount the cover on and make sure it is tight and blocked See Fig. 3 and 4.
6.4 SPECIAL NOTE FOR INSTALLATION OF
CATEGORY 1 / 2 G PRESSURE SWITCHES
Explosionproof pressure switches (Series BAX) can be installed on processes requiring apparatus of group II category 1 in an ambient requiring apparatus of group II category 2 (see Fig. 6).
Fig. 6 - Installation of Group II Cat. 1 / 2 G instruments
2
NI-222E
Rev. 1 12/02
8 - PUTTING INTO OPERATION
As the signal transmitted by the instrument is used in a complex system, it is necessary that the means of put­ting it into operation are established by those in charge of the plant.
The instrument comes into operation as soon as the root valve is opened. Any possible drainage of the connection tubing can be carried out by removing the safety plug and opening the service valve with the necessary
caution.
In case of explosionproof instruments (Series BAX), initial inspections are to be carried out according to cus­tomer procedures and at least in accordance with Stan­dard EN-60079-17.
9 - FUNCTIONAL VERIFICATION
This will be carried out according to the Client’s control procedures. Series BWX instruments can be verified on the plant if installed as illustrated in Fig. 7 and 8.
The instruments Series BAX may be checked on site only if apparatus suitable for explosives atmosphere are used and provided that the electric line is not energized.
If this is not the case it is necessary to stop operation, dismount by means of the three-piece joints and carry out the verification in a test room.
CAUTION: do not open the cover of explosionproof pressure switches (Series BAX) when energized, in explosive atmospheres.
Verification consists in checking the calibration value and possibly regulating the adjustment screw (see §4 ).
In case of explosionproof instruments (Series BAX), inspections of the electrical installation are to be carried out also according to customer procedures and at least in accordance with Standard EN-60079-17.
7 - INSTRUMENT PLUMBING Weatherproof pressure switches (Series BWX)
The plumbing, aimed as a guarantee against possible tampering of the calibration and electrical connections, can be carried out using a flexible steel wire (c) inserted into the holes in the screw (a) and the bracket (e) pro­vided for this purpose (see Fig. 3).
Explosionproof pressure switches (Series BAX)
Plumbing is not necessary as the cover is blocked with a headless screw, the closure plugs of the adjustment screw access holes are blocked by means of the internal blocking device and the instrument does not have to be opened when installed.
Page 6
NI-222E
10 - TROUBLESHOOTING
IMPORTANT NOTE: operations involving replacement of essential components must be carried out at our work­shop, especially for instruments with explosionproof certificate; this is to guarantee the user the total and cor­rect restoration of the product original characteristics.
MALFUNCTION PROBABLE CAUSE REMEDY
Set point shift
Slow response
No actuation or undue actuation
Undue actuation
11 - STOPPING AND DISMOUNTING
Before proceeding with these operations ensure that the plant or machines have been put into the conditions foreseen to allow these operations.
With reference to figures 7 and 8
Remove the power supply (signal) from the electrical line. Close the root valve (6) and open the drain Remove the plug (2), open the valve (3) and wait until the process fluid has drained from the tubing through the drain.
Do not dispose of the process fluid into the environment, if this can cause pollution or damage to people.
Unscrew the three-piece joint (8).
CAUTION: do not open the cover of explosionproof pressure switches (Series BAX) when energized, in explosive atmospheres.
Unscrew the three-piece joint (10) (electrical cable tub­ing).
Permanent deformation of the sensitive
element due to fatigue or non-tolerated over­ranges.
Variation of the elastic features of the sensi-
tive element due to its chemical corrosion.
Clogged or obstructed connection line. Root valve partially closed. Too viscous fluid. Root valve closed. Microswitch contacts damaged. Loosened electrical joints. Interrupted or short-circuited electrical line. Accidental shocks or excessive mechanical
vibrations.
INSTRUCTION MANUAL
Recalibrate or replace the sensitive element.
Recalibrate or replace the sensitive element
with another made of a suitable material. If necessary apply a fluid separator.
Check and clean line. Open valve. Provide instrument with suitable fluid separator. Open the valve Replace the Microswitch. Check all electrical joints. Check the conditions of the electrical line. Modify the mounting.
Remove the instrument cover and disconnect the electri­cal cables from the terminal block and earth screws. Remove the screws fixing the case to the panel (or pipe) and remove the instrument, taking care to slide the elec­trical conductors out from the case. Replace the instrument cover. Insulate and protect the conductors remaining on the plant. Temporarily plug the tubing (4).
In case of explosionproof instruments (Series BAX) it is recommended to follow - at least – the standard EN­60079-17 for the withdrawal from service of electrical apparatus.
12 - DEMOLITION
The instruments are mainly made of stainless steel and aluminium and therefore, once the electrical parts have been dismounted and the parts coming into contact with fluids which could be harmful to people or the environ­ment have been properly dealt with, they can be scrapped.
Rev. 1 12/02
Page 7
NI-222E
INSTRUCTION MANUAL
WEATHERPROOF EXPLOSIONPROOF
Fig. 7 - Example of connections - Fig. 8 - Example of connections -
Rev. 1 12/02
Fig. 9 - Example of mounting -
Surface mounting
Montaggio a parete
2” pipe mounting (vertical pipe)
Montaggio su staffa per tubo 2" (palina verticale)
NOTE With gas or vapour process fluid, the instrument must be positioned higher than the pipe inlet (see Fig. 8). With a liquid process fluid, the in­strument can be positioned higher or lower, indifferently (see Fig. 7 and 8). In this case, during set point calibration the negative or positive head must be taken into account. (Distance h in Fig.7 and 8).
LEGEND
1 - Fitting 2 - Drain plug 3 - Service valve 4 - Piping 5 - Three piece fitting 6 - Root valve with drain 7 - Process piping 8 - Three piece fitting 9 - “T” fitting 10 - Three piece fitting 11- Curve 12 - Blocking joint 13 - M6 screws (No. 4) 14 - Bracket for 2” pipe 15 - Vertical pipe 16 - Horizontal pipe
2” pipe mounting (horizontal pipe)
Montaggio su staffa per tubo 2" (palina orizzontale)
ETTORE CELLA SPA Viale de Gasperi, 48 - Casella Postale (P.O. Box) 96 - I 20010 Bareggio (MILANO) ITALY
Telefoni +39 029036.1146/7 - 029036.1237/41 - FAX +39 029036.1331 Email: CELLA@ECELLASPA.COM
Loading...