WIKA BWX User Manual

A
A
INSTRUCTION MANUAL
Rev. 1 12/02
BOURDON TUBE PRESSURE SWITCHES
WEATHERPROOF: SERIES BW EXPLOSIONPROOF: SERIES BAX
B
NI-222E
B
A pressure attachment B cable entry WEIGHT 2 kg dimensions in mm WEIGHT 3,7 kg dimensions in mm
NOTE: dimensions and weights are not binding unless released on certified drawings.
CAUTION
Before installing, using or carrying out maintenance on the instrument it is necessary to read and understand the indications given in the attached Instruction Manual.
The instrument must only be installed and maintained by qualified personnel.
INSTALLATION IS TO BE CARRIED OUT ONLY AFTER CHECKING THAT INSTRUMENT CHARACTERISTICS
ARE CONSISTENT WITH PROCESS AND PLANT REQUIREMENTS.
The functional features of the instrument and its degree of protection are shown on the identification plate fixed to the case.
CONTENTS:
1 - GENERAL NOTES 2 - OPERATING PRINCIPLE 3 - IDENTIFICATION PLATE AND MARKINGS 4 - SET POINT REGULATION 5 - SET POINT CALIBRATION 6 - MOUNTING AND CONNECTIONS 7 - INSTRUMENT PLUMBING 8 - PUTTING INTO OPERATION 9 - FUNCTIONAL VERIFICATION 10 - TROUBLESHOOTING 11 - STOPPING AND DISMOUNTING 12 - DEMOLITION
SAFETY INSTRUCTIONS FOR USE IN HAZARDOUS ATMOSPHERES. RECOMMENDATIONS FOR PRESSURE SWITCH SAFE USE.
A pressure attachment B cable entry
RELATED DOCUMENT
To authentified document with certificate
N° CESI 03 ATEX 005
All data, statements and recommendations supplied with this manual are based on information believed by us to be
reliable. As the conditions of effective use are beyond our control, our products are sold under the condition that the user himself evaluates such conditions before following our recommendations for the purpose or use foreseen by him.
NI-222E
INSTRUCTION MANUAL
Rev. 1 12/02
1 - GENERAL NOTES
1.1 FOREWORD
The wrong choice of a series or a model, as well as the incorrect installation, lead to malfunction and reduce instrument life. Failure to abide by the indications given in this manual can cause damage to the instrument, the environment and persons.
1.2 ALLOWED OVERRANGE
Pressure exceeding the working range can be occa­sionally tolerated provided they remain within the limits
stated in the instrument features (vacuum or proof pres­sure). Continuous pressures exceeding the working range can be applied to the instrument provided they are clearly stated in the instrument features. The current and voltage values stated in the technical specifications and ratings must not be exceeded. Transitory overranges can have a destructive effect on the switch.
1.3 MECHANICAL VIBRATIONS
Can generally lead to the wearing of some parts of the instrument or cause false actuation. It is therefore rec­ommended that the instrument be installed in a place where there are no vibrations. In cases where this is impossible it is a good idea to take measures to lessen the effects (elastic supports, installation with the peg of the microswitch positioned at right angles to the vibration plane, etc.).
1.4 TEMPERATURE
Due to the temperature of both the environment and the process fluid, the temperature of the instrument could exceed the allowed limits (normally from -20° to +70°C). Therefore, in case it does, suitable measures (protection against heat radiation, fluid separators, cooling coils, heated lockers) must be taken.
2 - OPERATING PRINCIPLE
The pressure, acting on the sensitive element (Bourdon tube), determines its elastic deformation which is used to activate one or two electric microswitches regulated at set point values. The microswitches are of the rapid release type with automatic rearm. When the pressure moves away from the set values, returning towards the normal values, the switch is rearmed. The differential gap (difference between the set point value and the reset value) can be set or adjustable (letter R in the contact codes).
3 - IDENTIFICATION PLATE AND MARKINGS
The instrument is fitted with a metal plate bearing all its functional characteristics and – in case of explosionproof execution (Series BAX) – also the markings prescribed by standard CEI EN 50014. Fig. 2 shows the plate mounted on explosionproof instruments.
Fig. 1 - Explosionproof instruments plate
1 CE marking and identification number of the notified body
responsible for production surveillance.
2 Apparatus classification according to ATEX 94/9 CE direc-
tive.
3 Type of protection and ambient temperature limits of opera-
tion.
4 Notified body that issued the type certificate and number of
said certificate.
4 - SET POINT REGULATION
Each microswitch is independent and can be regulated by means of a screw (for adjustment) in such a way that it is released when the pressure reaches (increasing or decreasing) the desired value (set point). The instrument is usually supplied with the switches set at the setting range value nearest to zero (factory cali- bration). The instrument is supplied with an adhesive rating plate showing the set point calibration value. With factory calibration the values are not indicated on the ratings as these are temporary and will be modified with the defini­tive values. Prior to installation the instrument must be calibrated and the definitive calibration values written on the adhe­sive rating plate using a suitable indelible ink pen.
If the instrument has been ordered with a specific cali- bration, it is a good rule to check the calibration values marked on the relevant adhesive label, prior to installa­tion.
The position of the adjustment screws is given in figure The effect that the direction of rotation of the adjustment screws has is described on the adhesive plate.
NI-222E
INSTRUCTION MANUAL
Rev. 1 12/02
Fig. 2 -
One contact instruments
1 - Microswitch set point calibration screws 3 - Terminal block 4 - Electrical connection identification plate
5 - SET POINT CALIBRATION
In order to proceed with the calibration and the periodical functional verification of the instrument a suitable cali- bration circuit (fig. 5) and an adequate pressure source is required.
5.1 PRELIMINARY OPERATIONS
Weatherproof pressure switches (BWX)
Remove the blocking device fixed to the side of the in­strument case and the adjustment screw access plate (Fig. 3). Remove the cover by rotating it in an anticlock­wise direction.
Fig. 3 - Weatherproof pressure switch blocking device -
Electrical connections and adjustment screws
Two contact instruments
1 - Microswitch 2 set point calibration screws 2 - Microswitch 1 set point calibration screws 3 - Terminal block 4 - Electrical connection identification plate
Explosionproof pressure switches (BAX) (Fig. 4)
CAUTION: do not open the cover of explosionproof
pressure switches (Series BAX) when energized, in explosive atmospheres.
Loosen the locking headless screw situated on the cover using a 1,5 hexagonal key then unscrew the cover. Remove the internal blocking device inserted on the closure plugs and slide out the plugs.
Fig. 4 -
Explosionproof pressure switch blocking device
Microswitch electrical circuit
State of the contacts at at­mospheric pressure
Designation of the contacts: C - common NA - Normally open NC - Normally closed
a - Plumbing wire b - Plumbing c - Blocking nut
d - Blocking bracket e - Adjustment screw access plate
5.2 CALIBRATION CIRCUIT AND OPERATIONS
Prepare the control circuit as indicated in Fig.5. The warning lamps should be connected to contact 1 or 2 in the NO or NC position according to the required contact action.
Connection of C and NO terminals
If the circuit is open at the working pressure, the switch closes the circuit as the pressure increases when the desired value is reached.
If the circuit is closed at the working pressure, the switch opens the circuit as the pressure decreases when the desired value is reached.
Connection of C and NC terminals
If the circuit is closed at the working pressure, the switch opens the circuit as the pressure increases when the desired value is reached.
Loading...
+ 4 hidden pages