ist eine geschützte Marke in verschiedenen Ländern.
Prior to starting any work, read the operating instructions!
Keep for later use!
28 - 54
55 - 80
81 - 106
Vor Beginn aller Arbeiten Betriebsanleitung lesen!
Zum späteren Gebrauch aufbewahren!
Lire le mode d'emploi avant de commencer toute opération !
A conserver pour une utilisation ultérieure !
¡Leer el manual de instrucciones antes de comenzar cualquier trabajo!
¡Guardar el manual para una eventual consulta!
2
WIKA operating instructions calibration system, model BCS10
14051046.01 10/2012 GB/D/F/E
Contents
Contents
1.General information4
2.Safety5
3.Specications9
4.Design and function11
5.Transport, packaging and storage13
6.Commissioning, operation14
7.Maintenance and cleaning25
8.Faults25
9.Dismounting, return and disposal26
GB
10.Accessories27
Appendix 1: EC Declaration of conformity model BCS1028
Declarations of conformity can be found online at www.wika.com.
14051046.01 10/2012 GB/D/F/E
WIKA operating instructions calibration system, model BCS10
3
1. General information
1. General information
■
The pressure transmitter described in the calibration instructions has been designed
and manufactured using state-of-the-art technology. All components are subject
GB
to stringent quality and environmental criteria during production. Our management
systems are certied to ISO 9001 and ISO 14001.
■
These operating instructions contain important information on handling the instrument. Working safely requires that all safety instructions and work instructions are
observed.
■
Observe the relevant local accident prevention regulations and general safety regulations for the instrument's range of use.
■
The operating instructions are part of the product and must be kept in the immediate
vicinity of the instrument and readily accessible to skilled personnel at any time.
■
Skilled personnel must have carefully read and understood the operating instructions
prior to beginning any work.
■
The manufacturer's liability is void in the case of any damage caused by using the
product contrary to its intended use, non-compliance with these operating instructions, assignment of insuciently qualied skilled personnel or unauthorised modications to the instrument.
■
The general terms and conditions contained in the sales documentation shall apply.
■
Subject to technical modications.
■
Factory calibrations / DKD/DAkkS calibrations are carried out in accordance with
international standards.
WIKA operating instructions calibration system, model BCS10
14051046.01 10/2012 GB/D/F/E
1. General information / 2. Safety
Explanation of symbols
WARNING!
... indicates a potentially dangerous situation that can result in serious
injury or death, if not avoided.
CAUTION!
... indicates a potentially dangerous situation that can result in light
injuries or damage to the equipment or the environment, if not avoided.
Information
... points out useful tips, recommendations and information for ecient
and trouble-free operation.
2. Safety
WARNING!
Before installation, commissioning and operation, ensure that the appropriate calibration system has been selected in terms of measuring range,
design and specic measuring conditions.
Non-observance can result in serious injury and/or damage to the equipment.
GB
Further important safety instructions can be found in the individual
chapters of these operating instructions.
14051046.01 10/2012 GB/D/F/E
WIKA operating instructions calibration system, model BCS10
5
2. Safety
2.1 Intended use
The calibration system model BCS10 serves for the inspection of SF6 gas density
measuring instruments.
The instrument has been designed and built solely for the intended use described here,
and may only be used accordingly.
GB
The technical specications contained in these operating instructions must be
observed. Improper handling or operation of the instrument outside of its technical specications requires the instrument to be taken out of service immediately and
inspected by an authorised WIKA service engineer.
The manufacturer shall not be liable for claims of any type based on operation contrary
to the intended use.
Handle electronic precision measuring instruments with the required care (protect from
humidity, impacts, strong magnetic elds, static electricity and extreme temperatures,
do not insert any objects into the instrument or its openings). Plugs and sockets must be
protected from contamination.
If the instrument is transported from a cold into a warm environment, the formation of
condensation may result in instrument malfunction. Before putting it back into operation,
wait for the instrument temperature and the room temperature to equalise.
The manufacturer shall not be liable for claims of any type based on operation contrary
to the intended use.
2.2 Personnel qualication
WARNING!
Risk of injury should qualication be insucient!
Improper handling can result in considerable injury and damage to equipment.
■
The activities described in these operating instructions may only be
carried out by skilled personnel who have the qualications described
below.
■
Keep unqualied personnel away from hazardous areas.
Skilled personnel
Skilled personnel are understood to be personnel who, based on their technical training,
knowledge of measurement and control technology and on their experience and knowledge of country-specic regulations, current standards and directives, are capable of
carrying out the work described and independently recognising potential hazards.
Special operating conditions require further appropriate knowledge, e.g. of aggressive
media.
6
WIKA operating instructions calibration system, model BCS10
14051046.01 10/2012 GB/D/F/E
2. Safety
2.3 Additional safety instructions for SF6 gas in switchgear
The plant operator must ensure that the handling of SF6 gas is only carried out by a
qualied company or by qualied persons which have been specially trained in accordance with IEC 61634, section 4.3.1 or IEC 60480, section 10.3.1.
Valid standards and guidelines for SF
Installation, assembly, commissioning:
■
IEC 61634 (Handling of the SF6 gas)
■
IEC 60376 (new SF6 gas, technical SF6 gas)
■
IEC 60480 (used SF6 gas)
■
CIGRE report 276, 2005 (Practial SF6 gas handling instructions)
Leaks during operation:
■
IEC 60376 (new SF6 gas, technical SF6 gas)
■
IEC 60480 (used SF6 gas)
■
CIGRE 2002 („SF6 gas in the electrical industry“)
Repair work and maintenance:
■
IEC 61634 (Use and handling of SF6 gas in high-voltage switchgear and controlgear)
■
CIGRE 1991 (handling of the SF6 gas)
■
CIGRE report 276, 2005 (Practical SF6 gas handling instructions)
■
CIGRE report 163, 2000 (Guide for SF6 gas mixtures)
gas
6
GB
is a colourless and odourless, chemically neutral, inert and not
SF
6
inammable gas which is approx. ve times heavier than air, not toxic and
not harmful to the ozone layer.
Detailed information is given in IEC 60376 and IEC 61634.
14051046.01 10/2012 GB/D/F/E
WIKA operating instructions calibration system, model BCS10
7
2. Safety
2.4 Personal protective equipment
The personal protective equipment is designed to protect the skilled personnel from
hazards that could impair their safety or health during work. When carrying out the
various tasks on and with the instrument, the skilled personnel must wear personal
protective equipment.
GB
Follow the instructions, displayed in the work area, regarding personal protective equipment!
The required personal protective equipment must be provided by the operating company.
Wear safety goggles!
Protect eyes from ying particles and liquid splashes.
Wear protective gloves!
Protect hands from friction, abrasion, cuts or deep injuries and also from
contact with hot surfaces.
2.5 Special hazards
WARNING!
Residual media in the pressure measuring instrument can result in a risk
to persons, the environment and equipment.
Take sucient precautionary measures.
WARNING!
The wetted parts have been designed exclusively for SF
SF
/N2 mixtures.
6
The pressure inside the test pump can be extremely high. Ensure therefore that all connections of the calibration system are connected correctly.
8
WIKA operating instructions calibration system, model BCS10
gas and
6
14051046.01 10/2012 GB/D/F/E
2. Safety / 3. Specications
2.6 Labelling, safety marking
Product label
Date of manufacture
Model
Article number
Serial number
Measuring range
Adapter
Explanation of symbols
Before mounting and commissioning the instrument, ensure
you read the operating instructions!
GB
CE, Communauté Européenne
Instruments bearing this mark comply with the relevant European directives.
3. Specications
Measuring range
■
Measuring range:0 ... 20 bar relative
■
Overpressure limit:42 bar
■
Burst pressure:140 bar
Accuracy data
■
Accuracy:0.05 % FS
■
Compensated temperature range: 0 ... 50 °C
■
Temperature error:0.005 % of the span/K (outside of the compen-
sated temperature range)
14051046.01 10/2012 GB/D/F/E
WIKA operating instructions calibration system, model BCS10
9
3. Specications
Digital indicator
■
Type of indication:7-segment LCD
■
Digits:5 ½-digit
■
Resolution:0.001
■
GB
Character size:16.53 mm (0.65")
■
Bargraph display:20 segment bargraph, 0 ... 100 %
■
Possible pressure units:bar, psi, kg/cm2, kPa, MPa and 15 other units
Pressure ne-adjustment: By means of the ne-adjustment valve
■
Measuring rate:30/min. to 10/sec. selectable
■
Memory:MIN/MAX values
■
Data logger:Cyclic data logger: automatic recording of up to 8.500
Integrated data logger
values
10
Cycle time: selectable from 1 ... 3.600 sec. in the following
steps
1 s, 2 s, 5 s, 10 s, 30 s, 1 min, 2 min, 5 min, 10 min,
30 min and 1 h
CPGLog data-logger evaluation software is needed to
use the data logger function.
The order number of the CPGlog software is given in the
data sheet SP 60.08.
14051046.01 10/2012 GB/D/F/E
WIKA operating instructions calibration system, model BCS10
3. Specications / 4. Design and function
Voltage supply
■
Power supply:3 x 1.5 V AA alkaline batteries
■
Battery life:approx. 1.500 ... 2.000 h
■
Battery status indication:Icon in display for low battery level
Process connection
■
Test item adapter with quick-acting coupling for G ½
■
Test item adapter with quick-acting coupling for G ¾
■
Test item adapter with quick-acting coupling for M30 x 2
CE conformity
■
EMC directive:2004/108/EC, EN 61 326-1 emission (group 1, class B)
and interference immunity (appendix A)
Plastic case
■
Dimensions in mm:395 x 295 x 106
■
Weight:approx. 4 kg (with contents)
For further specications see WIKA data sheet SP 60.08 and the order documentation.
4. Design and function
4.1 Short description
The modular calibration system model BCS10 serves for pressure generation and
inspection of mechanical and electronic SF
means of comparative measurements at 20 °C.
The test pump can generate a pressure of 0 ... 35 bar. Individual measuring points can
be approached with highest accuracy by means of the ne-adjustment valve.
The results of the measurements can be displayed in one of the 21
customer units using the built-in digital pressure gauge.
gas density measuring instruments by
6
measuring units or in
GB
4.2 Scope of delivery
■
Precision digital pressure gauge model CPG1000 with attached pneumatic test
pump model CPP30
■
Test item adapter with quick-acting coupling for G ½
■
Test item adapter with quick-acting coupling for G ¾
■
Test item adapter with quick-acting coupling for M30 x 2
■
Adjustable overpressure valve
■
Metal flexible hose for connecting the test item, length 2 m
■
Plastic case, incl. foam insert nad operating instructions
Cross-check scope of delivery with delivery note.
14051046.01 10/2012 GB/D/F/E
WIKA operating instructions calibration system, model BCS10
11
4. Design and function
4.3 Design of the calibration system
1) Digital pressure gauge
2) Fine-adjustment valve
GB
3) Pressure relief valve
4) Switching between pressure or
vacuum generation
1
8
5) Pump handles
6) Adjustable knurled nut for
setting the pump performance
2
3
(overpressure protection)
7) Metal protection hose for
7
4
connecting the test item, with
quick-acting coupling on both
sides, length 2 m
8) Adjustable overpressure valve
5
6
4.4 Battery life time
The battery life time amounts to approx. 1.500 hours of operation (without backlighting).
In the upper left corner of the digital indicator you will nd a symbol of the battery capacity. This symbol appears when the battery voltage is too low.
For instructions on batteries see chapter 3 "Specications".
For battery replacement see chapter 6.8 "Battery replacement".
4.5 Serial interface
The digital pressure gauge has as standard a serial interface. The serial data transmission can be used for conguration, calibration and transmission of measured values of
the measuring instrument.
12
WIKA operating instructions calibration system, model BCS10
14051046.01 10/2012 GB/D/F/E
5. Transport, packaging and storage
5. Transport, packaging and storage
5.1 Transport
Check the calibration system for any damage that may have been caused by transport.
Obvious damage must be reported immediately.
5.2 Packaging
Do not remove packaging until just before mounting.
Keep the packaging as it will provide optimum protection during transport (e.g. change
in installation site, sending for repair).
5.3 Storage
Permissible conditions at the place of storage:
see chapter 3 "Specications"
Avoid exposure to the following factors:
■
Direct sunlight or proximity to hot objects
■
Mechanical vibration, mechanical shock (putting it down hard)
Store the calibration system in its original packaging in a location that fulls the conditions listed above. If the original packaging is not available, pack and store the instrument as described below:
1. Wrap the instrument in an antistatic plastic lm.
2. Place the instrument, along with the shock-absorbent material, in the packaging.
3. If stored for a prolonged period of time (more than 30 days), place a bag containing a
desiccant inside the packaging.
GB
14051046.01 10/2012 GB/D/F/E
WIKA operating instructions calibration system, model BCS10
13
6. Commissioning, operation
6. Commissioning, operation
Commissioning may only be carried out by trained and qualied personnel. The calibration system was comprehensively checked ex works for tightness.
GB
WARNING!
Only connect or disconnect test and calibration installations once the
system has been depressurised!
6.1 Switching on and o
To switch on the calibration system,
press the Power key (1). Press the
Power key again to switch o the
calibration system.
After switching on the version of the
rmware is indicated on the display for
approx. 1 second.
6.2 Zero point adjustment
To set the displayed value as zero
point, simply press the ZERO key (3).
The display changes immediately to the
0 value.
6.3 MIN/MAX value
The digital pressure gauge saves the
minimum and maximum pressure.
If the MAX/MIN key (4) is pressed once,
the maximum pressure from the memory is displayed. If the MAX/MIN key is pressed
again, the minimum pressure is displayed from the memory.
1
2
5
4
3
After 2 seconds, the display switches back to the operating mode. To clear the MAX/MIN
memory, press the MAX/MIN key (4) for at least 2 seconds until the message "CLr" is
displayed.
The bargraph display at the bottom of the display indicates the applied pressure relative
to the total measuring range.
Due to a possible Tare setting, the displayed pressure value can dier
from the applied pressure.
14
WIKA operating instructions calibration system, model BCS10
14051046.01 10/2012 GB/D/F/E
6. Commissioning, operation
6.4 Backlighting
The backlighting makes it easier to read the values of the digital indicator. Press the
ENTER key (5) to turn the backlighting on and o.
6.5 Conguration
The calibration system model BCS10 functions in two dierent modes, the operating
mode and the conguration mode.
In the operating mode, the digital pressure gauge measures the pressure and shows it
on the digital indicator.
The conguration mode is used for the setting of the system.
The following setting can be modied:
■
Units
■
Automatic disconnection
■
Damping
■
Sampling rate
■
Tare function
■
Function locking
For the detailed description of the individual points see chapter 6.6 "Functions".
Conguration menu
To call up the conguration menu of the digital pressure gauge, press the CONFIG key
(2).
GB
The individual menu items are set via the ZERO key (3) and the MAX/MIN key (4). Hold
down the key to scroll faster.
To save all changes and exit the conguration menu, press the ENTER key (5).
You can exit the conguration menu at any time by pressing the ENTER
key. That will save all changes and the calibration system will return to the
operating mode.
14051046.01 10/2012 GB/D/F/E
WIKA operating instructions calibration system, model BCS10
15
6. Commissioning, operation
Menu navigation
Operating mode
GB
Unit selection
UNITS SET
Automatic
disconnection
SET AUTO-
Battery status
Temperature display
OFF
BAR
PSI
MPA ...
OFF
1 ... 30 min
°C
°F
Damping
SET DAMP
next page
16
ON
OFF
14051046.01 10/2012 GB/D/F/E
WIKA operating instructions calibration system, model BCS10
6. Commissioning, operation
Sampling rate
SET RATE
Tare function
SET TARE
Function locking
LOCK
FUnC
Operating mode
0.5/SEC
1/SEC
3/SEC
10/SEC
Values are
dependent on
the selected unit
Password
request:
enter 101
Zero point adjustment
(ZERO key)
lock/unlock
GB
The password 101 is factory-set and
cannot be changed.
UnLOC
LOC
Unit selection
lock/unlock
Automatic
disconnection
lock/unlock
next page
14051046.01 10/2012 GB/D/F/E
WIKA operating instructions calibration system, model BCS10
UnLOC
LOC
UnLOC
LOC
17
6. Commissioning, operation
GB
Damping
lock/unlock
Sampling rate
lock/unlock
Tare function
lock/unlock
User-dened unit
(enter scaling)
SET FACTR
UnLOC
LOC
UnLOC
LOC
UnLOC
LOC
Setting see
chapter 6.7
18
Operating mode
14051046.01 10/2012 GB/D/F/E
WIKA operating instructions calibration system, model BCS10
6. Commissioning, operation
6.6 Description of the menu items
Pressure units
The calibration system is factory set (see delivery note).
For the list of the available units see chapter 3 "Specications".
For conguration details of the user-dened units see chapter 6.7 "Conguration of the
function locking".
Automatic disconnection
The automatic disconnection can be adjusted in minute steps in the range from
1 ... 30 minutes. For continuous operation of the instrument, deactivate this menu item
with "OFF".
Battery voltage display
The current battery status is displayed as a voltage value and progress bar indication.
Temperature display
This menu item displays the temperature measured by the internal sensor. The value
can be displayed in °F or °C.
Damping
With enabled damping the measured values are smoothed by the pulsating pressure
sources.
GB
Sampling rate
Species the interval in which the pressure is measured and the display updated.
0.5 measurements per second correspond to 30 measurements per minute.
TARE (tare function)
In this menu item, a constant oset value can be set, which will be subtracted from the
measured pressure.
The TARE value depends on the selected values and can be set on the maximum scale
range.
The progress bar indication always indicates the actual pressure depending on the
entire measuring range, irrespective of the Tare setting. This is done for safety to
indicate that even if "0" is displayed that pressure is being applied to the calibration
system.
Example:
If, for example, TARE is set to 30 bar and the measured pressure is 37 bar, the measured
value displayed will be 7 bar. The pressure of 27 bar would then be displayed as -3 bar.
14051046.01 10/2012 GB/D/F/E
WIKA operating instructions calibration system, model BCS10
19
6. Commissioning, operation
Function locking
The calibration system model BCS10 is delivered from the factory with unlocked access
to all settings and they can be changed.
The access to the adjustable parameters can be limited or completely prohibited to
prevent unauthorized changes of the conguration.
GB
This is achieved by the function locking.
In addition, user dened units can be scaled here.
For conguration of the function locking see chapter 6.7 "Conguration of the function
locking".
6.7 Conguration of the function locking
Scroll to the menu item "FUnC
appears the message "0PWRD". Enter the password "101". The password is factory set
and cannot be changed.
The functions that can be locked or unlocked:
- zero point adjustment (ZERO button)
- setting the units
- auto Power-O Settings
- setting of the damping
- setting of the sampling rate
- tare setting
The command LOC locks the respective function and UnLOC unlocks it again.
Locked functions are not displayed in the conguration menu any more and therefore
cannot be changed.
Setting of a user-dened unit or a user-dened measuring range:
In the menu item SET FACTOR, a multiplier between 0.001 and 100 can be set for
a user-dened measuring range. The dened factor is multiplied by the measured
pressure value in psi. The result is displayed.
Example: 40 psi correspond to 1.000 g of a product in a tank. The weight of the product
should appear on the display.
If 25 is dened as the multiplicator, the pressure of 40 psi corresponds to the value
1.000 (40 x 25). „CUST“ is displayed on the instrument as a technical unit.
LOCK
" and press the ENTER key. On the digital indicator
20
WIKA operating instructions calibration system, model BCS10
14051046.01 10/2012 GB/D/F/E
6. Commissioning, operation
6.8 Battery replacement
1.Turn the orice plate of the digital pressure gauge approx. 90° to the right and
remove it.
2.Remove the front panel of the digital pressure gauge. This will give you access to
the battery compartment.
3.Replace the batteries. Use 1.5 V AA alkaline batteries only.
4.Close the instrument again. Ensure that the front is properly aligned.
6.9 Measurement setup and adjustment
WARNING!
Only connect or disconnect test and calibration installations once the
system has been depressurised!
1)Turn the instrument on and o
2)Perform a zero-point adjustment
1
3)Fine-adjustment valve
8
4)Pressure relief valve
2
5)Switching valve pressure/vacuum
3
4
generation
6)Pump handles
7)Adjustable knurled nut for
9
5
setting the pump performance
(overpressure protection)
GB
6
7
14051046.01 10/2012 GB/D/F/E
WIKA operating instructions calibration system, model BCS10
8)Adjustable overpressure valve
9)Metal exible hose for connecting
the test item, with quick-acting
coupling on both sides, length 2 m
21
6. Commissioning, operation
GB
Select an appropriate adapter and
screw it together with the test item.
Connect the test item via the quickacting coupling to the metal exible
hose.
Connect the other side of the metal exible hose via
the quick-acting coupling to the calibration system.
Check that the switching for pressure or vacuum
generation (arrow) is set to pressure.
The calibration system is suitable for pressure
generation only.
22
WARNING!
The construction of the calibration system is not designed for vacuum
generation. Always set the switching valve to pressure generation.
Never adjust the switching valve when the system is under pressure,
switch only when the release valve is open.
Close the pressure relief valve.
Turn the pressure relief valve clockwise until the valve
is closed. No hard stop can be felt.
14051046.01 10/2012 GB/D/F/E
WIKA operating instructions calibration system, model BCS10
6. Commissioning, operation
Switch on the calibration system.
The measurement setup shall be now as in the gure.
Actuate the pump handles several times until the
desired pressure is slightly exceeded.
To achieve the maximum pumping capacity, the spring above the knurled
nut must be relaxed.
For test items with low measuring ranges, the stroke path can be reduced
by turning the knurled nut. This will reduce the pressure increase per
stroke and minimize the risk of the overpressure.
GB
Set the desirede pressure via the ne-adjustment
valve (arrow).
Normally, the test is carried out when the pressure falls
because the gas density measuring instruments are
generally specied and adjusted in such a way.
To reduce the pressure, turn the ne-adjustment valve
counterclockwise.
Further reduction of the pressure is possible by means
of the pressure relief valve.
To increase the pressure, turn the ne-adjustment
valve clockwise.
Use the digital pressure gauge to read the set pressure.
14051046.01 10/2012 GB/D/F/E
WIKA operating instructions calibration system, model BCS10
23
6. Commissioning, operation
GB
To improve the readability of the display, the backlighting can be activated
by means of the ENTER key.
Compare and record the displayed value of the test
item with the displayed value of the digital pressure
gauge.
Open the pressure relief valve after calibration to
depressurize the system.
The test item can be separated from the calibration
system.
24
The calibration system can be dismount now in the
reverse order.
WARNING!
Only disconnect once the system has been depressurised!
14051046.01 10/2012 GB/D/F/E
WIKA operating instructions calibration system, model BCS10
7. Maintenance and cleaning / 8. Faults
7. Maintenance and cleaning
7.1 Maintenance
Repairs must only be carried out by the manufacturer.
7.2 Cleaning
CAUTION!
■
Clean the instrument with a moist cloth.
■
Electrical connections must not come into contact with moisture.
■
Residual media on the calibration system can result in a risk to
persons, environment and equipment. Take sucient precautionary
measures.
For information on returning the instrument see chapter 9.2 "Return".
7.3 Recalibration
DKD/DAkkS certicate - Ocial certicates:
We recommend that the calibration system is regularly recalibrated by the manufacturer,
with time intervals of approx. 12 months. Every factory recalibration includes, additionally, an extensive free-of-charge check of all system parameters with respect to their
compliance with the specication. The basic settings will be corrected if necessary.
8. Faults
8.1 Digital pressure gauge fault
Display indicationCausesMeasures
OL
-OL
No display or undenable
characters
Low battery voltage
functioning is only guaranteed for a
short period of time
Reading above or below the
measuring range
Battery at
Power supply unit is set incorrectly
or the polarity is revered
System error
Instrument faulty
Replace alkaline batteries
Adjust the line pressure to the
admissible range
Replace alkaline batteries
Set the power supply unit
correctly and ensure correct
polarity
Switch o the instrument, wait
for a short period of time and
switch on again
GB
14051046.01 10/2012 GB/D/F/E
WIKA operating instructions calibration system, model BCS10
Send to the manufacturer for
repair.
25
8. Faults / 9. Dismounting, return and disposal
CAUTION!
If faults cannot be eliminated by means of the measures listed above,
shut down the calibration system immediately, and ensure that pressure
and/or signal are no longer present, and secure the instrument from being
put back into operation inadvertently.
GB
8.2 Test pump fault
ProblemCausesMeasures
The test pump is leakingPressure relief valve is open
In this case, contact the manufacturer.
If a return is needed, please follow the instructions given in chapter 9.2
"Return".
Close the pressure relief valve
The pump handles are sti
to operate
The pressure and vacuum switching is not switched correctly
(The switching is in the middle
position)
The pump has not been used for a
long time
Turn the shift lever to pressure
generation.
Actuate the pump until it
functions smoothly
CAUTION!
If faults cannot be eliminated by means of the measures listed above,
shut down the calibration system immediately, and ensure that pressure
and/or signal are no longer present, and secure the instrument from being
put back into operation inadvertently.
In this case, contact the manufacturer.
If a return is needed, please follow the instructions given in chapter 9.2
"Return".
9. Dismounting, return and disposal
WARNING!
Residual media on submersible calibration system can result in a risk to
persons, the environment and equipment.
Take sucient precautionary measures.
9.1 Dismounting
Only dismount measurement setups once the system has been depressurized!
26
WIKA operating instructions calibration system, model BCS10
14051046.01 10/2012 GB/D/F/E
9. Dismounting, return and disposal / 10. Accessories
9.2 Return
WARNING!
Strictly observe the following when shipping the instrument:
All instruments delivered to WIKA must be free from any kind of hazardous substances (acids, bases, solutions, etc.).
When returning the instrument, use the original packaging or a suitable transport
package.
To avoid damage:
1. Wrap the calibration system in an antistatic plastic lm.
2. Place the calibration system, along with shock-absorbent material, in the packaging.
Place shock-absorbent material evenly on all sides of the transport packaging.
3. If possible, place a bag containing a desiccant inside the packaging.
4. Label the shipment as carriage of a highly sensitive measuring instrument.
Information on returns can be found under the heading "Service" on our
local website.
9.3 Disposal
Incorrect disposal can put the environment at risk.
Dispose of instrument components and packaging materials in an environmentally
compatible way and in accordance with the country-specic waste disposal regulations.
GB
This marking on the instruments indicates that they must not be disposed
of in domestic waste. The disposal is carried out by return to the manufacturer or by the corresponding municipal authorities (see EU directive
2002/96/EC).
10. Accessories
For use of the calibration system model BCS10 with the CPGlog data-logger evaluation
software see instruction manual "Data-logger software" at www.wika.de.
The order number of the CPGlog data-logger is given in the data sheet SP 60.08.
14051046.01 10/2012 GB/D/F/E
WIKA operating instructions calibration system, model BCS10
27
Appendix 1: Declaration of conformity for model BCS10
GB
28
WIKA operating instructions calibration system, model BCS10
14051046.01 10/2012 GB/D/F/E
Inhalt
Inhalt
1.Allgemeines30
2.Sicherheit31
3.Technische Daten35
4.Aufbau und Funktion37
5.Transport, Verpackung und Lagerung39
6.Inbetriebnahme, Betrieb40
7.Wartung und Reinigung51
8.Störungen51
9.Demontage, Rücksendung und Entsorgung52
D
10.Zubehör53
Anlage 1: EG-Konformitätserklärung Typ BCS1054
Konformitätserklärungen nden Sie online unter www.wika.de.
14051046.01 10/2012 GB/D/F/E
29WIKA Betriebsanleitung Kalibriersystem, Typ BCS10
1. Allgemeines
1. Allgemeines
■
Das in der Betriebsanleitung beschriebene Kalibriersystem wird nach dem aktuellen
Stand der Technik konstruiert und gefertigt.
Alle Komponenten unterliegen während der Fertigung strengen Qualitäts- und
Umweltkriterien. Unsere Managementsysteme sind nach ISO 9001 und ISO 14001
zertiziert.
■
Diese Betriebsanleitung gibt wichtige Hinweise zum Umgang mit dem Gerät. Voraussetzung für sicheres Arbeiten ist die Einhaltung aller angegebenen Sicherheitshin-
D
weise und Handlungsanweisungen.
■
Die für den Einsatzbereich des Gerätes geltenden örtlichen Unfallverhütungsvorschriften und allgemeinen Sicherheitsbestimmungen einhalten.
■
Die Betriebsanleitung ist Produktbestandteil und muss in unmittelbarer Nähe des
Gerätes für das Fachpersonal jederzeit zugänglich aufbewahrt werden.
■
Das Fachpersonal muss die Betriebsanleitung vor Beginn aller Arbeiten sorgfältig
durchgelesen und verstanden haben.
■
Die Haftung des Herstellers erlischt bei Schäden durch bestimmungswidrige Verwendung, Nichtbeachten dieser Betriebsanleitung, Einsatz ungenügend qualizierten
Fachpersonals sowie eigenmächtiger Veränderung am Gerät.
■
Es gelten die allgemeinen Geschäftsbedingungen in den Verkaufsunterlagen.
■
Technische Änderungen vorbehalten.
■
Werkskalibrierungen / DKD/DAkkS-Kalibrierungen erfolgen nach internationalen
Normen.