Welch Allyn VLX-10 User manual

VideoPath
®
VLX-10
Light Source
Operating Manual
LAMP
PICTURE
BRIGHTNESS
STANDBY
POWER
AIR
Light Source Menu
Service Manual
Thank you for purchasing the Welch Allyn VideoPath VLX-10 Light Source. The operating and maintenance instructions found in this manual
should be followed to ensure many years of reliable service. Please read these instructions thoroughly before attempting to use your new VLX-10 Light Source.
IMPORTANT: The material outlined in this manual should be reviewed and understood prior to operation of the equipment.
CAUTION: Federal Law restricts this device to sale by/to or on the order of a physician or other appropriately licensed medical professional.
WARNING: The user of this equipment should be thoroughly trained in the medical procedures appropriate to the instrumentation. Furthermore, time should be taken to read and understand these instructions before performing any procedures. Instructions for other equipment used in conjunction (e.g., suction machines) should also be read and understood. Failure to do so may result in injury to the patient and/or damage to the instrument.
Table of Contents
Conventions
General Precautions..........................................................................................2
Symbols.............................................................................................................3
Specifications
....................................................................................................4
Components
.......................................................................................................5
VLX-10 Light Source .........................................................................................5
Optional Accessories.........................................................................................5
Nomenclature
Light Source Front Panel...................................................................................6
Light Source Bottom View.................................................................................7
Water Bottle/Cleaning Bottle .............................................................................7
Light Source Back View.....................................................................................8
Monitor..........................................................................................................9-10
Connection Diagrams
....................................................................................11
Preparation For Use
General Precautions........................................................................................12
Setting Up the VLX-10 Light Source...............................................................13
Connecting the VideoEndoscope and Water Bottle...................................13-14
Inspection
Powering Up the Light Source.........................................................................14
Lamp Ignition...................................................................................................15
Inspection of Illumination.................................................................................15
Inspection of Air/Water Feed...........................................................................16
Operation
Power Switch...................................................................................................17
Lamp Ignition...................................................................................................17
Illumination Brightness ....................................................................................17
Air Pump .........................................................................................................17
Maintenance and Storage
After Each Procedure......................................................................................18
Water Bottle Cleaning......................................................................................18
Enzymatic Solutions....................................................................................18
Water Bottle Sterilization.................................................................................19
ETO Sterilization..............................................................................................19
Cold Sterilization..............................................................................................20
Sterilizing/Disinfecting Solutions......................................................................20
Bottle Storage..................................................................................................20
Replacing Lamps........................................................................................21-22
Light Source Storage .......................................................................................23
Troubleshooting
..............................................................................................24
VLX-10 Light Source 1
2 VideoPath

Conventions

WARNING:Alerts the user to possible serious injury, death or other
adverse reaction associated with the use or misuse of the device.
CAUTION: Indicates a potentially hazardous situation which, if not avoided, may result in minor or moderate injury. It also alerts against unsafe practices.
NOTE:
Provides supplemental information to the text and indicates a potentially hazardous situation, which, if not avoided, may result in property damage. Additionally it highlights important information on the use of this equipment.
General Precautions
• For safety, this light source should only be coupled to a grounded 110-120V AC outlet (hospital grade receptacle).
• The light source should not be operated in the presence of flammable or explosive gases or chemicals or installed in areas where these materials are commonly utilized.
• To avoid overheating, the light source should be positioned on a hard, level surface and the back of the light source should not be positioned closer than 6" to any wall.
• Keep all liquids away from electrical equipment to avoid the possibility of operator shock and instrument damage.
• Occasionally inspect the power cord for signs of cuts, abrasions or dents.
• The light source should never be stored or operated in areas where it could get wet or exposed to any environmental conditions such as extreme temperature, humidity, direct sunlight, dust, etc.
• All service to the light source must be performed by Welch Allyn or an authorized repair center.
• The light source and connected endoscope should not be used with electrosurgical generators. The VideoSigmoidoscope system has not been designed for therapeutic procedures.
• Light source use should adhere to the operating conditions found in this manual. Otherwise, instrument damage may occur and/or operator/patient safety compromised.
VLX-10 Light Source 3
Conventions
(Continued)

Symbols

Off: (Power: disconnects from the mains)
On: (Power: connects to the mains)
Off: (Power for air switch)
On: (Power for air switch)
Attention: Consult user’s manual for additional information
Dangerous voltage
Protective earth (ground)
Risk of Explosion
Caution High Temperatures
Type B Equipment
Warning: Risk of fire, replace fuses as marked
Caution: High Intensity Light
POWER
POWER
AIR
AIR

4 VideoPath

Specifications

Item Specification Technical Data
Power Requirements Voltage 115VAC
Frequency 50/60Hz Current 2.0 Amps Max.
Dimensions H x W x D 7 x 17 x16-1/2
(18 x 43 x 42cm)
Weight 15.0 lbs. (6.8 kg)
Operating Environment Ambient Temperature +10° C (50° F) to
+40°C (104° F) Relative Humidity 30-75% Atmospheric Pressure 700 hPa to 1060 hPa
Illumination Lamp Type Metal Halide Arc
Lamp Life (average) 750 hrs @ 1 hr per start Brightness Adjustment Automatic Shutter Cooling Method Forced Air
Air Feeding Pump Type Linear Motor, Free Piston
Capacity 6.5 PSIG Maximum Flow Rate .177 c.f.m. @ 2.84 PSIG Control Power switch activated
Water Feed System Water Compression Pressurized by pumped air
Watter Bottle Capacity 260 ml
Compatible Endoscopes Welch Allyn Color CCD only
VideoEndoscopes Other Manufacturer’s All models with use
Fiberscopes of video adapter
module (not presently available)
Color System Color mosaic CCD Video Outputs Composite (NTSC) and
Y/C (SVHS)
Output Devices VCR
Monitor Thermal or Video Printer

Compliance Designed in Accordance UL 2601

Degree of explosion proofing Use in potentially flammable
surroundings prohibited
Classification Electro medical equipment Class I Equipment,
3 prong plug Degree of protection against Type B, use on heart is electrical shock prohibited

Fuse T2.0A-250V Lamp Welch Allyn Part No. 04060 Lamp Voltage 60 Volts

Components

VLX-10 Light Source Includes:

40500 Light Source

45510 Water Bottle

761076-0 Power Cord

45512 S-VHS Video Cable

45520 Cleaning/Disinfection Bottle

Optional Accessories

04060 Lamp Replacement Assembly

VLX-10 Light Source 5

POWER AIR
!
STANDBY
PICTURE
BRIGHTNESS
LAMP
IGNITION
6 VideoPath

Nomenclature

Light Source Front Panel

Front Panel

Name
Function
1. Main Power Switch Primary power control for the light source.
If not activated, other controls will not function.
2. Air Pump Switch Activates flow of air at air output port in endoscope connector port.
3. Lamp Ignition Ignites and activates the lamps.
4. Lamp Standby Indicator When LED blinks, lamps cannot be ignited.
When LED remains on, lamps have warmed up/cooled down and can be ignited.
5. Picture Brightness Control Pressing up or down arrows adjusts image on monitor to desired brightness level (level indicated by illuminated green LED)
6. Endoscope Connector Port Accepts endoscope connector terminal to transmit light and electrical signals through the endoscope.
7. Water Bottle Water bottle couples to endoscope connector terminal’s air and water input ports and light source air output port.
8. Air Output Port Couples to and pressurizes the water bottle which provides air and water to the endoscope.
POWER AIR
!
STANDBY
LAMP
IGNITION
PICTURE
BRIGHTNESS
1. Main Power Switch
2. Air Pump Switch
3. Lamp Ignition
4. Lamp Standby Indicator
7. Water Bottle
5. Picture Brightness Control
8. Air Output Port
6. Endoscope Connector Port
VLX-10 Light Source 7
Nomenclature
(Continued)
Light Source Bottom View
Light Source Bottom View
Name Function
9. Lamp Access Door Opens for lamp replacement.
(Note: If door is not properly closed, interlock power switch will deactivate and the unit will not power up.)
10. Ventilation Holes Allows air to enter unit to cool internal components.

Water Bottle/Cleaning Bottle

Name Function
11. Air Input Connector Couples to light source
air output port.
12. Air Output Connector Couples to endoscope air input port.
13. Water Output Connector Couples to endoscope water input port.
14. Fill Line Indicates maximum water level to which bottle should be filled for normal use.
Water Bottle
NOTE:
Cleaning bottle is identical to water bottle with the exception that it allows fluid to be introduced into the air input port of the endo­scope. The bottle can be easily identified via the red CAUTION label on the front of the bottle.
14.Fill Line
11. Air Input Connector
12.Air Output Connector
13.Water Output Connector
9. Lamp Access Door
10.Ventilation Holes