Welch Allyn Service Utility CD (requires separate Serial Cable)
TESTER, CALIBRATION, 9600 PLUS
BP TEST VOLUME REPAIR FIXTURE 113670
Neonatal cuff hose, 96-inch
Disposable One-Piece Cuffs, Neo #1, 1-Tube (10/Box)
Battery substitution connector
Tycos air release valve and small bulb
Only Required for Devices Equipped With Masimo SpO2
Only Required For Devices Equipped with Nellcor SpO2
CABLE, SP02 EXTENSION, NELCOR
NELLCOR PORTABLE OXIMETRY TESTER (MP506)
TOOL_FLUKE, MULTIMETER, 187
TOOL BLOOD PRESSURE SIMULATOR
MISC TOOL, CPU W/ WIN XP O/S
SC TOOL, FLAT HEAD SCREWDRIVER
MISC TOOL, CORDLESS SCREWDRIVER
MISC TOOL, NEEDLE NOSE PLIERS, SMALL
MISC TOOL, NEEDLE NOSE PLIERS, MEDIUM
MISC TOOL, USB TO SERIAL CABLE
MISC TOOL, RS232 NETWORK CABLE
Indicates items only available from Welch Allyn or a Welch Allyn distributor.
Service and Maintenance Toolset
Vital Signs Monitor 300 Series
The following tools are required to perform service and maintenance, as well as functional verification and calibration
on the Welch Allyn Vital Signs Monitor 300 Series. Please note that these tools must also be readily available to
complete Welch Allyn Technical Service Training for this device.
Required Service Tools and Test Equipment*
Required Miscellaneous Off-the-Shelf Tools
The following tools are not available from Welch Allyn, but will be used in service and maintenance activities.
Power supply Variable, 0-8 VDC, 0.75 amperes (minimum), with
voltage and current indicators (for 1mA current measurement)
Syringe, 60 mL, Slip tip, LuerBD (Becton, Dickinson) 309654 or
equivalent
AC withstand voltage (hi-pot) tester Associated Research 3605 or
equivalent and Connectors
Required Miscellaneous Off-the-Shelf Tools (continued)
To place an order for your required tools, contact your Welch Allyn representative or call 1.800.535.6663. For more information on Welch Allyn
Technical Service Training, visit www.welchallyn.com/service
*The listed tools meet the performance requirements necessary to perform service and maintenance, as well as functional verification and
calibration as described in the Welch Allyn Vital Signs Monitor 300 Series™ Service Manual. For complete details on test procedures and
parameters, please review the service manual.