Welch Allyn VITAL SIGNS MONITOR Service manual

Service Manual
Welch Allyn
Vital Signs Monitor
Medical Division
4341 State Street Road, P.O. Box 220
Skaneateles Falls, NY 13153-0220
PN 95P360E
Copyright 1996
Welch Allyn Vital Signs Monitor Service Manual

SECTION 1:

General Information
Revision 4.0
SERVICE MANUAL
Welch Allyn
Vital Signs Monitor
Copyright 1996 Welch Allyn Co.
4341 State Street Road Service Manual pn95P360E Skaneateles Falls, NY 13153-0220 Rev. 4, 9/98
Welch Allyn Vital Signs Monitor Service Manual
Revision 4.0
*Note: All drawings enclosed are for reference only. Consult the system or the factory for all current drawings.
Revision History
Revision Sections Date Author Description
1.00 All 4/18/96 JDB/RJS Introduction of manual.
2.00 All 7/9/96 JDB Revision made due to problems found in training and changes to the repair software.
3.00 All 5/28/97 JDB Revisions made for support of Nellcor SpO module.
4.00 All 10/1/98 JDB Added schematics, board layouts and bill of materials, miscellaneous corrections.
2
Welch Allyn Vital Signs Monitor Service Manual
Revision 4.0

Table Of Contents

SECTION 1: General Information Sub-Section: Page:
To Service Personnel................................................ 1.1 1
Limited Warranty/Warranty Assistance ..................... 1.2 3-4
Basic System Description.......................................... 1.3 5
Basic System Operation Index .................................. 1.4 7
Basic System Operation............................................ 1.4 8-14
Block Diagram of Unit Operation............................... 1.5 16-17
Specifications ............................................................ 1.6 19-23
Serial numbering system defined.............................. 1.7 25
Firmware Identification .............................................. 1.8 27
SECTION 2: Service Set Up
Service Manual Intent & Scope ................................. 2.1 1
Test Equipment Bench Layout .................................. Fig 2.2.1 3-4
Required Tools/Equipment List ................................. 2.2.2 5
Replacement Parts List ............................................. 2.3 7-9
Maintenance/Service Support ................................... 2.4 11
SECTION 3: Problem Diagnosis
Overview ................................................................... 3.1 1
Diagnostic Process/Returned units ........................... 3.2 3-5
When and how to calibrate........................................ 3.3 7-8
Loading Unit Software............................................... 3.4 9
Welch Allyn Monitor Self Diagnostic Fault Codes ..... 3.5 11-13
“Complaint/Cause/Corrective Action” ........................ 3.6 15-19
SECTION 4: Major Component Removal/Replacement
(IN ORDER OF REMOVAL ACCESS)
Front Housing............................................................ 4.1 1-2
Main Fuse F1 on Main PCB ...................................... 4.2 3-4
Main Printed Circuit Board (PCB).............................. 4.3 5-6
DC TO DC Converter (SpO
Pressure Transducer on Main PCB........................... 4.5 9-10
Display PCB .............................................................. 4.6 11-12
Key Pad (front housing)............................................. 4.7 13-14
Temperature PCB ..................................................... 4.8 15-16
Specific Oxygen Board (Nonin) ................................. 4.9 17-19
Specific Oxygen Board (Nellcor) ............................... 4.10 21-22
Pump ......................................................................... 4.11 23-24
Pressure Switch MPL-503......................................... 4.12 25-26
Valve (Pneutronics) ................................................... 4.13 27-28
Printer/Printer PCB/keypad ....................................... 4.14 29-30
Battery ....................................................................... 4.15 31-32
units) ............................ 4.4 7-8
2
Welch Allyn Vital Signs Monitor Service Manual
Revision 4.0
SECTION 5: Test Procedures Sub-Section: Page:
Calibration test.......................................................... 5.1 1-2
Current tests............................................................. 5.2 3
Noise test.................................................................. 5.3 5
Button Test ............................................................... 5.4 7
Print quality............................................................... 5.5 9
Pneumatic test.......................................................... 5.6 11-12
Hardware Fail-safe Tests ......................................... 5.7 13
option testing................................................... 5.8 15
SpO
2
Temperature option testing....................................... 5.9 17
Check List for CBPM Service Work.......................... 5.10 19-20
SECTION 6: Drawings, Schematics, Board Layouts, Bill of Materials
SECTION 7: Repair Test Specifications
Specifications ........................................................... 7.1 1-5
CAUTION: PRIOR TO DOING REPAIR WORK, USE GROUNDING MAT AND ANTI-ST A TIC WRIST STRAP T O MINIMIZE RISK OF DAMA GE T O SENSITIVE ELECTRONIC COMPONENTS .
Welch Allyn Vital Signs Monitor Service Manual
Revision 4.0

1.1 To Service Personnel:

Read and understand the Welch Allyn Vital Signs Monitor Operator Manual and this Service Manual. The information contained in both of these publications is subject to change without notice and should not be construed as a commitment by Welch Allyn, Inc.
Welch Allyn assumes no responsibility for any errors that may appear in this manual. If the product and/or its operation varies significantly from any description herein, please contact the WELCH ALLYN TECHNICAL SERVICE DEPARTMENT: 1-800-450-9275 (828) 684-4895 Fax (828)687-9323
Welch Allyn, Inc. 95 Old Shoals Road Arden, North Carolina 28704
This product has been designed to provide a high degree of safety and reliability. However, we can not guarantee against: deterioration of components due to aging, normal wear, tampering, and abuse.
Section 1: General Information
All service and repairs must be performed by authorized Welch Allyn personnel or agents, using approved Welch Allyn replacement parts and approved process materials. Failure to so will invalidate the product warranty. Please refer to the product warranty for specific coverage.
Welch Allyn, Inc. 95 Old Shoals Road Arden, North Carolina 28704 USA 1 800 450-9275
CAUTION WHEN USING ADHESIVES:
ALWAYS WEAR SAFETY GLASSES AND PROVIDE ADEQUATE VENTILATION WHEN USING ADHESIVES, THEIR ACCELERATORS AND RTV SEALANTS. READ, UNDERSTAND, AND COMPLY WITH SAFETY RECOMMENDATIONS IN THE MATERIAL SAFETY DATA SHEET MSDS) FOR THESE MATERIALS. SUPER-GLUES (CA ADHESIVES) CAN CAUSE CORNEAL DAMAGE IF ACCIDENTLY SQUIRTED IN THE EYES, AND CAN SPONTANEOUSLY COMBUST IF SPILLAGE IS WIPED UP WITH CLOTH OR PAPER TISSUES. READ THE WARNINGS AND CAUTIONS PRINTED ON THE BOTTLES OR TUBES OF THESE GLUES!
1
Welch Allyn Vital Signs Monitor Service Manual
Revision 4.0

1.2 Limited Warranty

Welch Allyn warrants the Welch Allyn Vital Signs Monitor, when new, to be free of defects in material and workmanship and to perform in accordance with manufacturer’s specifications for a period of two years from the date of purchase from Welch Allyn or its authorized distributors or agents. Welch Allyn will either repair or replace any components found to be defective or at variance from manufacturer’s specifications within this time at no cost to the customer. It shall be the purchaser’s responsibility to return the instrument to Welch Allyn or an authorized distributor, agent or service representative. This warranty does not include breakage or failure due to tampering, misuse, neglect, accidents, modifi­cation or shipping. This warranty is also void if the instrument is not used in accordance with manufacturer’s recommendations or if repaired by other than Welch Allyn or an authorized agent. Purchase date determines warranty re­quirements. No other express warranty is given.
Section 1: General Information
To receive service assistance or to ask questions regarding this warranty, please call or write:
Welch Allyn Technical Services Dept. 95 Old Shoals Road Arden, North Carolina 28704 USA 1 (800) 450-9275 or (828) 684-4895 Fax: (828) 687-9323

Service Policy

All repairs on products under warranty must be performed or approved by a Welch Allyn Service Center. should be repaired by qualified electronics personnel or a Welch Allyn Service Center.
Unauthorized repairs will void the warranty
. Products out of warranty

Technical Assistance

If you have an equipment problem that you cannot resolve, call the Welch Allyn Service Center nearest you for assistance. Technical service support is available to you by tele­phone on normal business days, from 8:30AM until 4:30PM Eastern Standard Time at the phone numbers listed on the next page.
3
Section 1: General Information
Before returning a product for repair you must obtain authorization from Welch Allyn. An RGA (Return Goods Authorization) number will be given to you by our service personnel. Be sure to note this number on the outside of your shipping box. Returns without an RGA number will not be accepted for delivery.
Welch Allyn Vital Signs Monitor Service Manual
Revision
Welch Allyn Service Centers
For Service or Repair
4.0
USA Customers Welch Allyn, Inc.
Technical Service Centers 95 Old Shoals Arden, NC 28704-9739 USA Phone: 828-684-4895 Fax: 828-687-9323
Welch Allyn
Technical Service Center Zollerstrasse 2-4 Road D-7241 7 Jungingen GERMANY Phone: [49]-7477-9271-73 Fax: [49]-7477-9271-93
Welch Allyn Ltd.
21-09 Golden Beach Road Singapore 199589 REPUBLIC OF SINGAPORE Phone: Fax: [65]-291-5780
[65]-29
Road
GmbH
Mile to 6001
l-0882
CANADA Customers Welch Allyn Canada Limited
Technical Service Centers
160 Matheson Blvd., East Mississauga, Ontario Ph: 905-890-0004 or 1 800-561-8797 Fax: 905-890-0008
INTERNATIONAL Customers
Welch Allyn Australia Pty. Ltd.
Technical Service Center Ground Floor, 18-20 Lane Cove AUSTRALIA Phone: [61]-29-4183-155 Fax: [61]-29-4183-650
NSW 2066
L4Z1
Orian
V4 CANADA
Road
MDI International
Technical Service Centers 7324 SW 48th Street Miami, FL 33155 USA Phone: 305-669-9591 Fax: 305-669-l 971
4
LATIN AMERICA Customers
Welch Allyn
\N!!
Welch Allyn Vital Signs Monitor Service Manual
Revision 4.0
Section 1: General Information

1.3 Basic System Description

Introduction
The Welch Allyn Vital Signs Monitor is designed to non-invasively and automatically measure systolic and diastolic pressure, pulse rate, tempera­ture and oxygen saturation (SpO sure, pulse, temperature and read displays, and may be printed via an integrated thermal printer, as desired.
The rechargeable battery and wide variety of mounting accessories make the Clinical Monitor convenient for many locations. The operator may choose any combination of simultaneous measurement modalities. This flexibility, combined with features such as programmable alarms and auto­matic BP cycles, makes the Clinical Monitor ideal for a wide variety of pa­tient monitoring needs.
) for adult and pediatric patients. All pres-
2
SpO2 values are displayed on large, easy-to-
Refer to the Operator Manual PART NUMBER 5200-85 (SPECIFY LAN­GUAGE) Section 1 for complete information on Indications For Use, Special Features, Supplies and Accessories
5
Welch Allyn Vital Signs Monitor Service Manual
Revision 4.0
Section 1: General Information

1.4 Basic System Operation

INDEX
1.5.1 Blood Pressure Operating Modes a. Max/Min Blood Pressure and Pulse Ranges b. Blood Pressure Manual Mode c. Blood Pressure Auto Mode
1.5.2 Temperature Operating Modes a. Max/Min Temperature Ranges b. Temperature Normal Mode c. Temperature Monitor Mode
1.5.3 SpO
Operating Mode
2
a. Max/Min SpO b. SpO
Monitor Mode
2
1.5.4 Pulse Rate Feature Max/Min Pulse Rate Ranges
Ranges
2
7
Section 1: General Information
Welch Allyn Vital Signs Monitor Service Manual
Revision 4.0

1.4.1 Blood Pressure Operating Modes

When a blood pressure measurement cycle is initiated, the cuff will automatically be inflated to the operator selected pressure level.
The cuff will immediately begin to deflate in a stepped fashion and will determine systolic pressure and diastolic pressure from the pulses sensed by the cuff at various pressure levels. This is the oscillometric method of non-invasive blood pressure monitoring.
Blood pressure measurements may be initiated manually, or automatically at time inter­vals determined by the user.
At the completion of a measurement cycle the systolic and diastolic pressures are dis­played. If the Monitor is in Automatic Mode, the measured values are kept on display until the next BP measurement is initiated. When not in Automatic Mode, the measured val­ues are displayed for two minutes, after which time the display screen is blanked. The most recent BP measurement may be recalled by pressing the REVIEW button.
When in Automatic Mode, if the unit is unable to determine the systolic or diastolic value, the measurement will be automatically repeated once.
1.4.1.a. Max/Min Blood Pressure Ranges
The maximum and minimum ranges of blood pressure are detailed below:
Measurement Maximum Minimum
Systolic Pressure 250 mmHg 60 mmHg Diastolic Pressure 160 mmHg 30 mmHg
8
Welch Allyn Vital Signs Monitor Service Manual
Revision 4.0
Section 1: General Information
1.4.1.b. Blood Pressure Manual Mode
In the Manual Mode, a single blood pressure determination is made only when the START button is pushed. The manual mode is the default mode of operation for blood pressure determinations.
A measurement cycle may be canceled at any time by pressing the CANCEL button. This action immediately initiates a rapid cuff deflation.
The blood pressure measurement data will appear on the display immediately following the measurement. The display will blank after two minutes. If the display is blanked, pressing the REVIEW button on the front panel will recall the measurement. Up to 99 prior measurements are available for review or printing.
In Blood Pressure Manual Mode, the following alarm limits may be activated; SYSTOLIC HIGH LIMIT, SYSTOLIC LOW LIMIT, DIASTOLIC HIGH LIMIT, DIASTOLIC LOW LIMIT, PULSE RATE HIGH LIMIT and PULSE RATE LOW LIMIT. Blood pressure determina­tions which activate alarms are indicated by flashing displays and a repetitive audible tone. If an alarm limit is violated, subsequent blood pressure determinations may be made only after the alarm condition is reset by pressing any button on the Monitor’s display.
1.4.1.c. Blood Pressure Automatic Mode
The Automatic Blood Pressure Mode is entered by pressing the AUTO button. Pressing the AUTO button displays a choice of 10 cycle interval times as follows: “st” (STAT mode) 3 min, 4 min, 5 min, 10 min, 15 min, 30 min, 45 min, 60 min and 90 min. These choices represent the time interval from the beginning of one cycle to the beginning of the next automatic cycle. STAT mode allows the monitor to take continuous blood pressure mea­surements for 15 minutes. In addition, the operator may choose to disable the Auto Mode by choosing “—” (two dashes).
A measurement cycle may be canceled at any time by pressing the CANCEL button. This action immediately initiates a rapid cuff deflation.
The blood pressure measurement data will appear on the display immediately following the measurement and will remain displayed until the next measurement cycle is initiated.
9
Section 1: General Information
Welch Allyn Vital Signs Monitor Service Manual
Revision 4.0
1.4.1.c. Blood Pressure Automatic Mode continued
In Blood Pressure Auto Mode, the following alarm limits may be activated; SYSTOLIC HIGH LIMIT, SYSTOLIC LOW LIMIT, DIASTOLIC HIGH LIMIT, DIASTOLIC LOW LIMIT, PULSE RATE HIGH LIMIT and PULSE RATE LOW LIMIT. Blood pressure determina­tions which activate alarms are indicated by flashing displays and a repetitive audible tone. Any alarm limit violation sure determinations. The alarm may be reset by pressing any button on the monitor’s display.
must be reset to continue automatically timed blood pres-
1.4.2. Temperature Operating Modes
Thermistor Thermometry measurements are made with the DIATEK/Welch Allyn SureTemp4TM thermometer. Oral and rectal probes utilize single-use disposable probe covers which limit cross-contamination. Oral or rectal temperatures are taken using ‘Normal’ or ‘Monitor’ operating modes. Auxiliary temperatures may be taken using the oral probe in the monitor operating mode.
In the normal mode the thermometer’s microprocessor “predicts” body temperature in about 4 seconds for oral temperatures and in about 15 seconds for rectal temperatures. The monitor mode displays the patient’s actual temperature after 3 minutes and will continue to display an updated temperature as long as the probe remains in place.
Temperature readings may be displayed in Fahrenheit or Celsius scales.
1.4.2.a. Max/Min T emperature Ranges
Readings Normal Mode Monitor Mode
Temperature Max 108.0F Max 108.0F
Max 42.2C Max 42.2C Min 84.0F Min 84.0F
Min 28.9C Min 28.9C
10
Welch Allyn Vital Signs Monitor Service Manual
Revision 4.0
Section 1: General Information
1.4.2.b. T emperature Normal Mode
In Normal Mode, the device will measure temperature at discrete intervals and then calculate the rate of change according to a known algorithm. This allows the thermom­eter to predict the end point that the thermistor would reach if it were left in the mouth until it reached mouth temperature. This predictive an accurate oral temperature reading in approximately 4 seconds.
Normal mode is the default operating mode for temperature determinations.
Operator selectable patient alarm limits are not available in Temperature Normal
Mode. However, temperatures which are outside of the operating range of the
device will be noted on the temperature display.
feature allows the thermometer to arrive at
1.4.2.c. Temperature Monitor Mode
Continuous Monitor Mode operation is normally used for longer term monitoring and when difficult situations prevent accurate temperatures from being taken in the normal mode. The probe must be in contact with tissue for at least three (3) minutes for accurate temperature measurement. Monitor mode temperatures may not be identical to predicted ‘Normal’ temperatures because of ambient temperature influence and other factors. The trend in temperature is the important standard to be observed when in the ‘Monitor’ mode.
Operator selectable patient alarm limits are not available in Temperature Normal
Mode. However, temperatures which are outside of the operating range of the
device will be noted on the temperature display.
11
Section 1: General Information
Welch Allyn Vital Signs Monitor Service Manual
Revision 4.0

1.4.3 SpO Operating Mode

2
The Welch Allyn Vital Signs Monitor incorporates either the Nonin® pulse oximetry system
®
or the Nellcor (% SpO
) by measuring the absorption of red and infrared light passed through the tis-
2
pulse oximetry system which determines arterial oxyhemoglobin saturation
sues. Changes in absorption caused by pulsation of blood in the vascular bed are used to determine arterial saturation and pulse rate.
Oxygen saturation percent is calculated with each pulse detected, and thus the monitor display is continually updated. The pulse signal bar graph is an indicator of the strength and quality of the detected pulses.
When SpO pulse rate measurement from the SpO
is measured, the patient’s pulse rate is also measured and displayed. A
2
determination overrides a pulse rate measure-
2
ment derived from a blood pressure measurement. In SpO2 monitoring mode operator selectable alarm limits for low SpO2 % may be acti-
vated. A condition which violates the SpO
low limit alarm is indicated by a flashing
2
display and repetitive audible tone. Should a patient alarm condition for or pulse rate occur, the Monitor will indicate an alarm condition (flashing & beeping) while continuing to monitor and display the patient’s current SpO
%. The alarm will automatically reset when
2
the patient’s condition returns to within the preset alarm parameters. Should a patient alarm condition for SpO
or pulse rate occur, the operator may invoke
2
“Silence Mode” by pressing the SILENCE button. This will silence the audible tone (display will continue to flash), while the practitioner attends to the patient and the moni­tor. Silence mode resets automatically after 30 seconds, or when the patient’s condition returns to within the preset alarm parameters.
Removal of the SpO
sensor from the patient will initiate an alarm. To reset the sensor
2
alarm, press any button on the monitor’s display. SpO
is generally measured via pulses detected using a finger sensor. However, for
2
certain situations SpO
may be measured at alternate sites including the earlobe, fore-
2
head and toes. Special sensors must be employed in these situations.
12
Welch Allyn Vital Signs Monitor Service Manual
Revision 4.0
Section 1: General Information
1.4.3.a. Max/Min SpO
Ranges
2
The SpO2 sensor is designed to detect oxygen saturation as follows.
Maximum Minimum
SpO
2
1.4.3.b. SpO
Monitor Mode
2
99% 40%
The SpO2 monitor performs most accurately with the finger clip sensor, which may be used on all fingers except the thumb. The finger clip sensor is recommended for spot checks or short term (<30 minutes) continuous monitoring.
The device determines arterial oxyhemoglobin saturation (SpO
%) by measuring the
2
absorption of red and infrared light passed through the tissue. Oxygen saturation and pulse rate are displayed on the LED digital display. On each detected pulse, the pulse signal bar graph flashes. The intensity of this signal is a simple visual indicator of wave­form signal strength, and can identify situations where the pulsatile nature of the tissue may not be adequate for an accurate SpO
reading. The update interval of the bar graph
2
should correspond to the patient’s pulse rate. This is an indication of the quality of the SpO2 signal.

1.4.4 Pulse Rate Feature

The Welch Allyn Vital Signs Monitor is capable of determining pulse rate as an adjunct to the blood pressure measurement and the SpO
The pulse rate, in beats per minute, will be determined primarily from the SpO methodology. In the case where SpO
is not available, or is disabled, the pulse rate display
2
will be driven from data collected as part of the blood pressure measurement method. There are two operator selectable alarm limits for the pulse rate. They are PULSE RATE
HIGH LIMIT and PULSE RATE LOW LIMIT. Pulse rates which activate alarm limits are indicated by a flashing display and a repetitive audible tone.
Should a pulse rate alarm occur when the pulse rate measurement is derived from the blood pressure measurement, no subsequent blood pressure or pulse rate measurements may be made until the alarm is reset. The alarm may be reset by pressing any button on the monitor’s display.
measurement.
2
measurement
2
13
Section 1: General Information
Welch Allyn Vital Signs Monitor Service Manual
Revision 4.0
Should a pulse rate alarm occur when the pulse rate measurement is derived from the SpO
measurement, the monitor will indicate an alarm condition (flashing & beeping)
2
while continuing to monitor and display the patient’s current SpO
% and pulse rate. The
2
alarm will automatically reset when the patient’s condition returns to within the preset alarm parameters.
Should a patient alarm condition for pulse rate occur during SpO
monitoring, the opera-
2
tor may invoke “Silence Mode” by pressing the SILENCE button. This will silence the audible tone (display will continue to flash), while the practitioner attends to the patient and Monitor. Silence mode resets automatically after 30 seconds, or when the patient’s condition returns to within the preset alarm parameters.
1.4.4.a. Max/Min Pulse Rate Ranges
The maximum and minimum pulse rate ranges are as follows:
Maximum Minimum
Pulse Rate 200 bpm 40 bpm
14
Section 1: General Information
Welch Allyn Vital Signs Monitor Service Manual
Figure 1.5 Block Diagram of Unit Operation
Revision 4.0
Off
Yes
Yes
Reset
Display?
Yes
Yes
Start Taking
Reading
Is Cancel Pressed?
No
Is Reading
Finished?
No
Display Last
Reading
Any button than
review select?
No
Is 10 sec. since last
action?
No
Is Review
Pressed?
Yes
Display
Next RDG?
Is Power Button
No
Pressed?
No
Power Start
Review
Display
Record Setting
Reset
Yes
Silence
No
Is Alarm
Activated?
Yes
Silence Alarm
for 30 Seconds
Yes
Yes
Is Start
Button Pressed?
No
➤➤
Post
Ready State
Auto
Display Current
Timer Setting
Any button than
auto select?
Is 10 sec.
since last
action?
Is Auto
Pressed?
Display
Next Setting?
No
No
➤➤
Yes
No
Yes
Is Cancel
Button Pressed?
No
Select
Reset
Display
Record Setting
Is Adjust
Pressed?
No
Is Select Pressed?
Yes
Record Setting
for Displayed
Alarm
Display Next Alarm
@ Current Setting
Yes
F2
Feed
Print
Mode
Display Current
Setting of 1st Alarm
Any button than
Yes
Yes
Yes
No
M2
adjust select?
No
Is 10 sec. since last
action?
No
Display
Next
Setting?
D
C
F1
P1
P2
M1
16
Welch Allyn Vital Signs Monitor Service Manual
Revision 4.0
Section 1: General Information
Figure 1.5 Block Diagram of Unit Operation
M2
BP module
active, SpO
module inactive
M1
Both BP and SpO
2
SpO2 module
active, BP
module inactive
C
Post
Show Versions
Print produces version report
Start closes valve
Cancel opens valve
Auto runs pump while pressed
module active
Review advances to next box
Power returns to “O” at any time
Cancel resets changes
Calibration
P1
Print
Report
F2
D
Post
Print produces version report
Auto runs pump while pressed
Auto runs pump while pressed
F1
Advance Paper
One line
Review advances to next box
Power returns to “O” at any time
Cancel resets changes
Show Versions
Calibration
Start closes valve
Cancel opens valve
Battery Voltage
Start closes valve
Cancel opens valve
P2
2
Battery Voltage
Start closes valve
Cancel opens valve
Auto runs pump while pressed
Temp Module Setup
Adjust changes setup
Set Date Adjust changes flashing display Select makes next display flash
Set Time
Adjust changes flashing display
Select makes next display flash
SpO2 perfusion signal - systolic display, display perfusion status
Temp Module Setup
Adjust changes setup
Adjust changes flashing display
Select makes next display flash
Adjust changes flashing display
Select makes next display flash
A/D Channel Noise
Select changes A/D channels
RS232/Service Port Test
Start begins cycling
Cancel halts cycling
Set Date
Set Time
Life Test
17
Welch Allyn Vital Signs Monitor Service Manual
Revision 4.0
Section 1: General Information

1.6 Specifications

Index:
1.6.1 Performance Specifications
1.6.2 Technical Specifications a. Mechanical Specifications b. Electrical Specifications c. Environmental Specifications
1.6.3 Printer Specifications
1.6.1 Performance Specifications
The performance specifications of the Clinical Vital Signs Monitor are as follows:
CUFF PRESSURE RANGE
0 mmHg - 300 mmHg
INITIAL CUFF INFLATION
120 mmHg, 140 mmHg, 160 mmHg, 200 mmHg, 240 mmHg or 280 mmHg depending on pressure preset level.
SYSTOLIC DETERMINATION
Maximum: 250 mmHg Minimum: 60 mmHg
DIASTOLIC DETERMINATION
Maximum: 160 mmHg Minimum: 30 mmHg
BLOOD PRESSURE ACCURACY
Blood pressure accuracy meets or exceeds SP10-1992 AAMI standards for non-invasive blood pressure accuracy (AAMI standard: ± 5 mmHg mean error; 8 mmHg standard deviation). Blood pressure accuracy is validated for pressure measurement using the upper arm only.
BLOOD PRESSURE DETERMINATION TIME
20 seconds to 45 seconds typical, 140 seconds maximum.
19
Section 1: General Information
Welch Allyn Vital Signs Monitor Service Manual
PULSE RATE DETERMINATION
Maximum: 200 bpm Minimum: 40 bpm
PULSE RATE ACCURACY
SpO
Module Heart Rate ±3.0%
2
Blood Pressure Algorithm Heart Rate ±5.0%
OVERPRESSURE CUTOFF
295 mmHg to 330 mmHg
TEMPERATURE DETERMINATION
Normal Mode Maximum 108.0F (42.2C)
Minimum 84.0F (28.9C)
Monitor Mode Maximum 108.0F (42.2C)
Minimum 84.0F (28.9C)
Revision 4.0
TEMPERATURE ACCURACY
Temperature accuracy meets ASTM E112-86: “Standard Specification for Electronic Thermometer for Intermittent Determination of Patient Temperature.”
TEMPERATURE DETERMINATION TIME
Normal Mode: ORAL: 4 seconds typical, 15 seconds maximum Monitor Mode: ORAL: 3 minutes.
OXYGEN SATURATION RANGE (SpO2%) 40-99% oxygen saturation
SpO2 ACCURACY
±3% in the range of 70-99% oxygen saturation (1 Standard Deviation)
BATTERY CHARGING
To at least 90% capacity in 12 hours. Unit will operate and charge battery simultaneously when connected to power source.
20
Loading...
+ 134 hidden pages