Welch Allyn LB-24 User Manual

®
Light Source
Operating Manual
REF LB-24
®
CONTENTS
System Symbol Descriptions . . . . . . . . . . . . . . . . . . . . .1
System Description and Operation. . . . . . . . . . . . . . . . .3
Lamp Replacement. . . . . . . . . . . . . . . . . . . . . . . . . . . . .3
Cleaning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6
Maintenance and Repair. . . . . . . . . . . . . . . . . . . . . . . . .6
Technical Specifications . . . . . . . . . . . . . . . . . . . . . . . . .6
Environment. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7
Electrical Ratings . . . . . . . . . . . . . . . . . . . . . . . . . . .7
Dimensions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7
Illumination System . . . . . . . . . . . . . . . . . . . . . . . . .8
Patents. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .8
Approvals . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .8
Warranty. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .9
Agency Compliance Statements . . . . . . . . . . . . . . . . . .9
FCC Class A Compliance Statement . . . . . . . . . . . .9
Canadian Notice . . . . . . . . . . . . . . . . . . . . . . . . . . .10
iv CONTENTS Solarc Light Source
Operating Manual 1

System Symbol Descriptions

Caution: Consult accompanying documentation
Caution Hazard Warning
Lamp On Lamp Off
Light; Lighting; Illumination Intensity
Storage Humidity
MAX
Alternating Current Direct Current
Manufacturer
Do not dispose of this product as unsorted municipal waste. Prepare this product for reuse or separate collection as specified by Directive 2002/ 96/EC of the European Parliament and the Council of the European Union on Waste Electronic and Electrical Equipment (WEEE). If this product contaminated, this directive does not apply. For more specific disposal information contact Customer Service at +44 207 365 6780.
Caution: Hot Surface
Bright Light
Transport Temperature
is
2 Solarc Light Source

Warnings and Precautions

WARNING RISK OF SHOCK. Do not operate the Solarc Light Source
without a lamp installed. Disconnect power before attempting to change lamp.
WARNING Only qualified personnel should make electrical inspections and repair of the S
WARNING The high intensity light at the front of the Solarc Light
Source and at the tip of the fiber optic bundle will create high temperatures and bright light. To minimize the risk of injury, avoid direct viewing or contact.
WARNING To prevent temporary blinding and contact with heated parts, always plug the fiber optic bundle into the S before turning the power on.
WARNING Do not use the Solarc Light Source directly in medical
applications.
Caution Any changes or modifications made to this device that are not expressly approved by manufacturer may void the user’s authority to operate the equipment.
olarc Light Source.
olarc Light Source
Caution Do not switch the S succession. Doing so will dramatically shorten lamp life. Allow lamp to cool down for 10 to 15 minutes between each use.
Caution PROVIDE ADEQUATE VENTILATION TO PREVENT OVERHEATING. Do not drape this light source. Provide a 3-inch (7.62 cm) distance between the S
Caution DO NOT use or store liquids above or on the S Source.
Caution Do not operate device without the cover in place. Operate in the horizontal position only.
Caution Since the S have a spare lamp available for replacement. USE CAUTION BECAUSE LAMP MAY BE HOT. Allow lamp to cool before replacing.
olarc Light Source ON and OFF in rapid
olarc Light Source and any solid objects.
olarc Light
olarc Light Source uses a custom lamp, always
Operating Manual 3

System Description and Operation

The Solarc Light Source is used to supply high-quality , w hite light for video and optical illumination systems.
1. Plug the power supply into a grounded AC receptacle. The
power switch should be in the OFF position . If your So
option, turn it to the minimum position (Figure 1). Plug the fiber optic bundle into the light port and conne ct t he oppo sit e end to the equipment being used.
2. Turn the power switch to the ON position .
If your So option, set the light intensity to the desired level (Figure 1).
3. Turn unit OFF when not in use.
larc Light Source has the light intensity control
larc Light Source has the light intensity control
Caution Do not restart the Solarc Light Source immediately after shutting it
off - it may fail to light. If this occurs, turn the power off and wait 10 to 15 seconds before turning the unit back on.
Figure 1. Front and Back View
Maximum
Light Control
Minimum
Power Switch

Lamp Replacement

WARNING RISK OF SHOCK. Do not operate the Solarc
Light Source without a lamp installed. Disconnect power before attempting to change lamp.
Power Input
Fan
4 Solarc Light Source
Caution LAMP MAY BE HOT. Allow lamp to cool before
handling.
Recommended lamp cooling time is a minimum of 10 minutes.
Caution Do not operate device without the cover in place. Operate in the horizontal position only.
To Replace the Lamp
1. Turn the So
larc Light Source of f ( ) and unplug the power
cord from both the wall outlet and the rear of the unit.
2. Remove the five screws from the So
larc Light Source using a #2 philips head screwdriver. Pivot the cover to expose the lamp.
Figure 2. Removing the Top Cover
Pivot
Removable Screws (two on opposite side)
3. Flip up the retaining latch. Pull the lamp base back and up against the retaining spring to remove th e lamp.
4. Remove the lamp, with harness attached, and then disconnect the lamp connector.
Operating Manual 5
Figure 3. Inside the Light Source
Retaining Latch
Retaining Spring
Lamp Base
Lamp Connector
32
5. Replace with Welch Allyn replacement lamp AL - 1824 only. DO NOT touch the lamp or the interior reflective surface of the lamp. Skin oils on the lamp will cause premature lamp failure. Only hold the outside of the reflector or the base of the lamp. Remove any grease or fingerprints with a clean Q­tip moistened with alcohol. Do not leave any lint on the lamp.
6. Connect the new lamp to the connecto r and insert lamp to properly seat the alignment pin located on the bottom of the lamp into the slot located on the lamp bracket.
Figure 4. Bottom View of the Lamp Slot
Lamp Slot.
7. Close the retaining latch and replace the cover. Secure the five screws on the So
larc Light Source. Reconnect t he power
supply.
Caution Do not operate the device without the cover in place.
6 Solarc Light Source
Dispose of the lamp according to local, state, and federal laws.

Cleaning

1. Turn the Solarc Light Source off ( ) and unplug the power cord from both the wall outlet and the rear of the unit.
2. Wipe the external surfaces clean with a clot h dampen ed with mild soap and water. DO NOT IMMERSE.
3. Wipe the power cords clean with a cloth dampe ned with mild soap and water. DO NOT IMMERSE. DO NOT RECONNECT
WET.
4. DO NOT plug the power source into a wall outlet until it is thoroughly dry.

Maintenance and Repair

WARNING Only qualified personnel should make electrical inspections and repair of the S
olarc Light Source.
For troubleshooting or repair information, contact Customer Service at +1 315 685 4347.

Technical Specifications

Supplier Name: Address:
Model or Type: S
Welch Allyn Lighting Products Division 4619 Jordan Road PO Box 187 Skaneateles Falls, NY 13153-0187
olarc Light Source LB-24
Operating Manual 7

Environment

Operating:
Temperature: Humidity:
Storage:
Temperature: Humidity: Atmospheric Pressure:
Mode of Operation: Continuous
41° F to 104° F (5° C to 40° C) 0 to 95% rh
-10° F to 140° F (-20° C to 49° C) 30 to 75% rh
700 hpa to 1060 hpa
Safety System Classification:
System Pollution Degree: 2
Installation Category: II

Electrical Ratings

Power Supply Rating Input: Output:
Recommended Power Supply:
Battery: Welch Allyn recommends using a UR (or other recognized

Dimensions

Length: 5.8 in. (14.73 cm)
Width: 3.5 in. (8.89 cm)
Height: 2.8 in. (7.11 cm)
Class II
100 - 240 V~, 50/60 Hz, 1.6 A max 12 V , 3.4 A
North American: Welch Allyn Part # LBM-PS-00 International: Welch Allyn Part # LBM-PS-01
testing laboratory) recognized battery rated at 12V/10Ah or equivalent.
8 Solarc Light Source

Illumination System

Lamp: Welch Allyn Solarc Lamp # AL - 1824
Color Temperature: 5500 K
Lamp: 60 Volts / 24 Watts
Lamp Life: 24 Watt: 500 hours
Dimming System: Manual Shutter (Light Intensity Control)

Patents

Welc h Allyn Solar c lamps are manuf act ured unde r one or mo re of the following patents: 5,083,059; 5,117,154; 5,138,228; 5,144,201; 5,539,273, 5,420,477.

Approvals

The CE mark on this product indicates that it has been tested to and conforms with the provisions noted within the 93/42/EEC Medical Device Directive.
3045357
European Regulatory Manager Welch Allyn Ltd. Navan Business Park Dublin Road Navan, County Meath, Republic of Ireland Tel.: 353 46 90 67700 Fax: 353 46 90 67755
CERTIFIED TO: CAN/CSA STD C22.2 NO. 61010-1
CONFORMS TO: UL STD 61010-1
EN 61010-1 IEC 61010-1
EMC Framework of Australia
Operating Manual 9

Warranty

Welch Allyn warrants the Solarc Light Source, when new, to be free of defects in mat erial and workmanship and to perform in accordance with the manufacturer’s specifications when subject to normal use and service for a period of one year from the date of purchase from Welch Allyn or an authorize d agent. W elc h Allyn will either repair or replace any components found to be defective or at variance from the manufacturer’s specifications within this time at no cost to the customer. It shall be the purchaser’s responsibility to return the instrument to the authorized distributor, agent, or service representative.
This warranty does not cover the So due to tampering, misuse, neglect, accidents, improper installation, modification, shipping, or to improper maintenance, service, and cleaning procedures. This warranty is also vo id if the instrument is not used in accordance with the manufacturer’s recommendations or if required service is performed by anyone other than Welch Allyn or an authorized agent. The purc hase date determines warranty requirements. No other express or implied warranty is given.
larc lamp, breakage, or failure

Agency Compliance Statements

This device complies with part 15 of the FCC Rules. Operation is subject to the following two conditions: (1) this device may not cause harmful interference, and (2) this device must accept any interference received, including interference that may cause undesired operation.

FCC Class A Compliance Statement

This equipment has been tested and found to comply with the limits for a Class A digital device, pursuant to part 15 of the FCC Rules. These limits are designed to pro vide reasonable protection against harmful interference when the equipment is operated in a commercial environment. This e quipment generates, us es, and
10 Solarc Light Source
Note
can radiate radio frequency energy and, if not installed and used in accordance with the instruction manual, may cause harmful interference to radi o communications. Operation of this equipment in a residential area is likely to cause harmful interference, in which case the user will be required to correct the interference at his own expense.
Caution Any changes or modifications made to this device that are not expressly approved may void the user’s authority to operate the equipment.
To maintain compliance with FCC Rules and Regulations, cables connected to this device must be shielded cables, in which the cable shield wire(s) have been grounded (tied) to the connector shell.

Canadian Notice

This equipment does not e xceed the Class A limits for radio noise emissions as described in the Radio Interference Regulations of the Canadian Department of Communications.
Le present appareil numerique n’emet pas de bruits radioelectriques depassant le s lim ites applicables aux appareils numeriques de la classe A prescrites dans le Reglement sur le brouillage radioelectrique edicte par le ministere des Communications du Canada.
Material No. 708364 Rev. B
Welch Allyn Lighting Products Division
4619 Jordan Road
PO Box 187
Skaneateles Falls, NY 13153-0187
Tel: +1 315 685 4347
Fax: +1 315 685 2854
www.walamp.com
Loading...