Welch Allyn Kreativ Klean Air II User manual

Klean Air® II
Kreativ Menu
Manuals
Instructions for Use
Please read this manual thoroughly
prior to first use
Copyright Notice:
reproduced without written permission from Kreativ, Inc.
Manufactured by
Kreativ, Inc. (USA) 7420 Carroll Road San Diego, CA 92121-2334
European Regulatory Manager
Welch Allyn LTD.
Navan, Co. Meath
Republic of Ireland
Ph: 353-46-79060
990070 Rev. C

TABLE OF CONTENTS

1. INTRODUCTION AND GENERAL INFORMATION 3
2. SAFETY 6 WARNINGS 6 CAUTIONS 6 SAFETY MARKINGS ON LABELS 7
3. CONTROLS AND CONNECTORS 8 ITEM DESCRIPTIONS 8 FUNCTIONAL DESCRIPTION 9
4. UNPACKING AND SETUP 10 GENERAL INFORMATION 10 UNPACKING AND SETUP 10 ELECTRICAL REQUIREMENTS 10
5. OPERATION 11 FILTER PANEL INDICATORS 11 UNIT SHUTDOWN 12
6. SERVICING 13
7. TROUBLESHOOTING AND MAINTENANCE 14
8. RECYCLING 15 RECYCLING INSTRUCTIONS FOR DEVICE, ACCESSORIES AND PACKAGING 15 ENVIRONMENTALLY RELEVANT MATERIAL 15
9. WARRANTY, RETURNS AND ADJUSTMENTS 16 RETURNS AND ADJUSTMENTS 16 LIMITED WARRANTY 16 STORAGE AND SHIPMENT 17
Page 2 990070 Rev. C

1. Introduction and General Information

1

Introduction

The Kreativ Model Klean Air II is a highly efficient, compact, state-of-the-art instrument designed for air filtration. It is designed to be used with the dental office high velocity evacuation (HVE) to eliminate overspray from air abrasion procedures.

Purpose of this Manual

This manual should be thoroughly studied before using the equipment. Please call your nearest authorized Welch Allyn / Kreativ service representative if you need additional information regarding installation, operation, or service of the unit.
Chapter
Information essential to the installation, operation, and routine maintenance of the Klean Air II device manufactured by Kreativ, Inc., is included in this manual.

General Information and Features

A combination pre/charcoal filter serves two distinct purposes. The pre-filter captures aluminum oxide particles, old composites and other large particles. The charcoal removes most dental office odors and gasses.
A third-stage filter eliminates the biological aerosol created by a high speed drill. This HEPA filter, similar to the type used in a hospital operating room, removes
99.97% of airborne biological contaminants down to 0.3 microns in size.
Features
High Power Operation
Low Power Operation
Easy to read LED Display
Visual Filter status indicator
Quick detachable hose for easy access
Doors for easy filter replacement
Convenient handle
Mobility
Page 3 990070 Rev. C

Specifications

Power Requirements: 100-127 volt, ~, 60 Hz., 1.75A
220-240 volt, ~, 50 Hz., 650 mA
IEC 320 socket with removable cord is located at back; verify unit has correct
Air Flow: 80 cfm (low) or 120 cfm (high)
Approximate Size:13” x 13” x 27” (33 x 33 x 68.6 cm)
Approximate Weight: 45 lbs. (20.4 kg)
Control Panel: located on door of instrument for easy viewing
Filters: HEPA DOP 99.97% / 99.99% and activated carbon
Filter Replacement: user access through door
Hose: 3” dia. (7.6 cm), 3’ (91 cm). length with expanded
Power Switch: located adjacent to power inlet
voltage before use
(2.26 or 3.39 m3/min)
pre-filter
intake
Environment Limitations: Operational Requirements: Temperature: 50° to 95° F (+10° to 35° C)
Humidity: 30 % to 75 % Pressure range: 10.15 to15.37 psia (70 to 106 kPa)
Transport and Storage Requirements:
Temperature: 14° to 122° F (-10° to 50° C) Humidity: 10 % to 90 % Pressure range: 7.25 to 15.37 psia (50 to 106 kPa)
Equipment Classification: the Kreativ Model Klean Air II system is an EC Class I device requiring an electrical earth ground. This device is classified as ordinary with regard to protection against water ingress.
Electrical Fuses: Littelfuse, SlowBlow
Model Fuse Mfr. Mfr. P/N Kreativ P/N Fuse Rating
Klean Air II
Littelfuse Type 218-02.5 600109 2.5A, 250Vac Domestic Klean Air II
Littelfuse Type 218-01.6 600096 1.6A, 250Vac European
Page 4 990070 Rev. C
UL Canadian Classification Marking:
Oral Vacuum, Model 8000-XX
(where XX can be any alphanumeric character)
WITH RESPECT TO ELECTRIC SHOCK, FIRE, MECHANICAL
AND OTHER SPECIFIED HAZARDS ONLY
IN ACCORDANCE WITH CAN/CSA C22.2 NO. 601.1
21ZJ
Page 5 990070 Rev. C

2. Safety

2

Warnings

Make sure that the Kreativ Klean Air II is connected to a properly grounded electrical outlet.
Use protective rubber gloves when changing the filters and properly dispose of them when you are finished.
Be sure to wipe down the patient end of the hose with disinfectant and replace the Intake Liner prior to each patient use.
Chapter
Always replace fuses with fuses of the proper rating as indicated in the Specifications section of the previous chapter.

Cautions

Do not attempt to service the unit beyond those items described in the mainten-ance section. Call your nearest authorized Welch Allyn / Kreativ service representative for instructions.
Do not substitute non-Kreativ, Inc., materials or accessories. Use of materials or accessories supplied by other manufacturers may result in damage to the unit and will void the factory warranty.
Sound clinical judgment and good common sense are essential to the successful use of the unit in order to provide safe and efficient patient care.
Always follow “standard of care” protocols in performing dental operative procedures.