Caution: Federal US law restricts sale of the device identified in this manual to, or
on the order of, a licensed physician.
Welch Allyn assumes no responsibility for any injury, or for any illegal or improper
use of the product, that may result from failure to use this product in accordance
with the instructions, cautions, warnings, or indications for use published in this
manual.
Welch Allyn and KleenSpec are registered trademarks of Welch Allyn. CaviWipe is
a registered trademark of Metrex Research Corporation.
Caution: Changes or modifications not expressly approved by Welch Allyn could
void the purchaser's authority to operate the equipment.
Patents pending.
Mat. Number: 718913 , Ver: A
Welch Allyn
4341 State Street Road, PO Box 220
Skaneateles Falls, NY 13153-0220
www.welchallyn.com
Printed in USA
Contents
Introduction ............................1
Charging the illuminator .................10
Using the illuminator for a pelvic exam.....11
Using the illuminator as light source .......11
Maintenance ...........................12
iii
Intended use .................................1
Contraindications ..............................1
User profile ..................................2
Symbols .....................................2
Using the system safely ........................4
Utilisation sécuritaire du système .................7
Cleaning and disinfecting .......................12
Inspecting the system .........................13
Recycling the system components ...............13
Troubleshooting ........................14
Contacts ..............................14
Appendix..............................15
Specifications................................15
Part numbers ................................17
Limited warranty .............................18
Service policy ................................19
EMC guidance and manufacturer’s declarations .....20
iv
Directions for useCordless illumination system1
Introduction
The Welch Allyn KleenSpec®790 series cordless illumination
system can be used with or without the KleenSpec 590 series
disposable vaginal speculum. The cordless illumination system
consists of the following items.
Cordless
illuminator
Charging station
(base and power cord)
Intended use
When used with the vaginal speculum, the cordless illuminator
provides illumination during pelvic examinations and other
gynecological procedures, such as pap smears, dilation and
curettage (D&C) biopsy, and electrosurgery.
The cordless illuminator can be used independent of the
speculum as a general-purpose light source.
Contraindications
The 790 series illuminator (either by itself or in conjunction with a
590 series speculum) is not intended to be used for eye
examinations.
The 790 series illuminator (either by itself or in conjunction with a
590 series speculum) is not intended to be used for diagnosis.
Auxiliary power cord
(optional)
2Welch Allyn
User profile
The intended users of the illuminator are medical professionals
who are qualified to perform a pelvic examination.
Symbols
The symbols illustrated on the following pages may appear in this
manual or on the illuminator, charging station, auxiliary power
cord, packaging, or shipping container.
Table 1. Documentation symbols
WARNING
CautionIn this manual, this symbol and “Caution”
In this manual, this symbol and “WARNING”
indicate conditions or practices that could lead to
illness, injury, or death.
indicate conditions or practices that could
damage the equipment or other property.
On the product, this symbol means consult
operating instructions.
Table 2. Shipping, storing, and environment symbols
This end upKeep dry
95%
15%
-20C
FragileRelative humidity limit
+49C
Temperature limits
Directions for useCordless illumination system3
Table 3. Certification symbols
Meets essential requirements of European Medical Device Directive
93/42/EEC.
EC REP
Regulatory Affairs Representative
Welch Allyn Limited
Navan Business Park
Dublin Road, Navan
County Meath, Republic of Ireland
CERTIFIED TO: CAN/CSA STD C22.2 NO. 601.1
CONFORMS TO: IEC 60601-1, IEC60601-1-2, UL 60601-1
EMC framework of Australia
N344
Table 4. Operation symbols
Caution, hot surface
The speculum is a Type BF applied part
Alternating current
Direct current
Double insulation
IPXØ
Not protected from the ingress of liquid.
Recycle the product separately from other disposables. See “Recycling
the system components” on page 13.
Power plug
4Welch Allyn
Using the system safely
Before using the cordless illumination system, you must read and
understand the following safety-related information.
General warnings
The following warning statements apply to use of the cordless
illumination system in general. Warning statements that apply
specifically to particular procedures, such as cleaning, appear in
the corresponding sections of the manual.
Warning statements indicate conditions or practices that could
lead to illness, injury, or death.
WARNING To avoid the risk of electric shock, this equipment
must only be connected to a supply mains with protective
earth.
WARNING No modification to this equipment is allowed.
WARNING The illuminator may generate heat after prolonged
use.
•Operation time limits: Illuminator with speculum
10 minutes on, 5 minutes off between uses.
Illuminator alone 2 minutes on, 5 minutes off
between uses.
•Illuminator could be harmful to skin if touched after
being on for extended times.
•To avoid fire, do not place illuminator onto a
flammable surface if LED is on.
•Illuminator is not suitable for use in the presence of a
flammable anesthetic mixture with air or oxygen or
nitrous oxide. An explosion may result.
Directions for useCordless illumination system5
WARNING To prevent shock hazards, the power outlet must
be properly grounded. Grounding reliability is achieved only
when connected to hospital-use or hospital-grade receptacles.
WARNING The illuminator is not intended to be used for eye
examinations.
WARNING The illuminator is not intended to be used for
diagnosis.
WARNING The illuminator is intended to be used by trained
medical professionals.
WARNING This product comes with a power supply and/or
power cord that is intended for use only with this product.
The power supply and/or power cord have not been tested and
approved for use with other products that may have the same
power connectors. If you cannot locate the original power
supply and/or power cord, please contact Welch Allyn to obtain
replacement parts.
General cautions
The following caution statements apply to use of the cordless
illumination system in general. Caution statements that apply
specifically to particular procedures, such as pelvic examinations,
appear in the corresponding sections of the manual.
Caution statements indicate conditions or practices that could
damage the equipment or other property.
Caution Disconnect the AC power cord to remove power to
the device.
Caution The cordless illuminator could be a biohazard after
use. Follow good hospital practices. See “Cleaning and
disinfecting” on page 12.
Caution Cordless illuminator, charging station, and auxiliary
power cord are not protected from the ingress of liquid.
Caution There are no user-serviceable parts inside the device.
In case of a malfunction, call Welch Allyn and precisely
describe the problem.
6Welch Allyn
Caution This device complies with current required standards
for electromagnetic interference and should not present
problems to other equipment or be affected by other devices.
As a precaution, avoid using this device in close proximity to
other equipment.
Caution Use only Welch Allyn-approved parts. See “Part
numbers” on page 17.
Directions for useCordless illumination system7
Utilisation sécuritaire du système
Avant d'utiliser le système d’éclairage sans fil, vous devez bien lire et
comprendre les renseignements de sécurité suivants.
Avertissements généraux
Les avertissements suivants s’appliquent à l'utilisation du système
d’éclairage sans fil en général. Les avertissements qui s'appliquent à des
procédures particulières, comme le nettoyage, sont abordés dans les
sections correspondantes de ce manuel.
Ces avertissements indiquent des conditions ou des pratiques qui
pourraient causer des maladies, des blessures ou des décès.
AVERTISSEMENT Pour éviter les risques de choc électrique,
cet équipement doit seulement être branché sur une
alimentation secteur dotée d'une mise à la masse.
AVERTISSEMENT Aucune modification de cet équipement
n'est autorisée.
AVERTISSEMENT Le système d’éclairage peut dégager de la
chaleur après une utilisation prolongée.
•Les durées d'opération maximales : Illuminateur avec
spéculum 10 minutes de marche, 5 minutes d’arrêt
entre les utilisations. Illuminateur seul 2 minutes de
marche, 5 minutes d’arrêt entre les utilisations.
•Le système d’éclairage peut causer des brûlures cutanées
si vous y touchez après une utilisation prolongée.
•Pour éviter les risques d'incendie, ne déposez pas le
système d’éclairage sur une surface inflammable lorsque
la DEL est allumée.
•Le système d’éclairage ne devrait pas être utilisé en
présence d'un mélange d'anesthésique inflammable et de
l’air ou encore d'oxygène ou d'oxyde nitreux. Cela pourrait
causer une explosion.
8Welch Allyn
AVERTISSEMENT La prise électrique doit être correctement
mise à la terre afin d’éviter les chocs électriques. Vous ne
pouvez être certain de la mise à la terre que si l'appareil est
branché dans une prise utilisée par les hôpitaux ou de qualité
hospitalière.
AVERTISSEMENT Le système d’éclairage ne devrait pas être
utilisé pour les examens des yeux.
AVERTISSEMENT Le système d’éclairage ne devrait pas être
utilisé pour effectuer un diagnostic.
AVERTISSEMENT Le système d’éclairage est conçu pour
être utilisé par des professionnels médicaux formés.
AVERTISSEMENT Un bloc d'alimentation et/ou un cordon
d'alimentation, destinés à une utilisation avec ce produit
seulement, sont fournis avec le produit. Le bloc d'alimentation
et/ou le cordon d'alimentation n'ont pas été testés et
approuvés pour une utilisation avec d'autres produits qui
pourraient être munis des mêmes connecteurs d'alimentation.
Si vous n'avez plus le bloc d'alimentation et/ou le cordon
d'alimentation originaux, veuillez communiquer avec Welch
Allyn pour obtenir des pièces de remplacement.
Directions for useCordless illumination system9
Mises en garde générales
Les mises ne garde suivantes s’appliquent à l'utilisation du
système d’éclairage sans fil en général. Les énoncés de mise en
garde qui s'appliquent à des procédures particulières, comme
les examens pelviens, sont abordées dans les sections
correspondantes de ce manuel.
Les énoncés de mise en garde indiquent des conditions ou des
pratiques qui pourraient endommager les équipements ou la
propriété.
Mise en garde Débrancher le cordon d'alimentation c.a. pour
mettre l'appareil hors tension.
Mise en garde Le système d’éclairage sans fil pourrait être
considéré à risque biologique après utilisation. Conformezvous aux bonnes pratiques hospitalières. Voir “Cleaning and
disinfecting” à la page 12.
Mise en garde Le système d’éclairage sans fil, la borne de
recharge et le cordon d'alimentation auxiliaire ne sont pas
protégés contre les infiltrations de liquide.
Mise en garde Cet appareil ne contient aucune pièce à
entretenir. En cas de mauvais fonctionnement, appelez Welch
Allyn et décrivez le problème de manière précise.
Mise en garde Cet appareil est conforme aux normes
actuelles obligatoires sur les interférences
électromagnétiques. Il ne devrait présenter aucun problème
pour les autres équipements et ne devrait pas être affecté par
d'autres appareils. Par mesure de précaution, évitez d'utiliser
cet appareil à proximité d'autres équipements.
Mise en garde N'utilisez que des pièces approuvées par
Welch Allyn. Voir “Part numbers” à la page 17.
10Welch Allyn
Charging the illuminator
Charge the illuminator before using it the first time, or whenever
its low-charge indicator turns on.
LED
Power switch
Low-charge indicator:
On = battery low
To charge the illuminator
1. If LED is lit, turn power switch off.
2. Connect charging base to AC power.
3. Place the illuminator (either direction) into the charging
station. Full charging takes 6 to 8 hours.
Status indicator:
Off = no AC power
Green = AC power
Amber (illuminator inserted) = charging.
Full charging takes 6 to 8 hours.
4. Remove the charged illuminator when you are ready to use it.
It is safe to leave the illuminator in the charging station after it
is charged.
Directions for useCordless illumination system11
Using the illuminator for a pelvic exam
1. Fully insert the illuminator into a 590
series speculum (a). The LED
automatically turns on.
2. Conduct the examination.
3. If the LED turns off, do not remove the
speculum. Do either of these things:
•Leave the illuminator in place, and
use an external light source.
or
•Connect the auxiliary power cord
to AC power. Hold the speculum in
place, and remove the illuminator.
Connect the auxiliary power cord to
illuminator (b). Reinsert the
illuminator.
a
b
4. When finished, remove the speculum,
and then remove the illuminator. The
LED automatically turns off.
Using the illuminator as light source
To use the illuminator as a light source,
slide the power switch to turn it on (all the
way down) and off (all the way up).
Note
For operation time limits, see
“Intermittent operation” on page 16.
Power
switch
12Welch Allyn
Maintenance
Cleaning and disinfecting
Use CaviWipe®wipes for both cleaning and disinfecting.*
*EPA-approved intermediate-level disinfectant appropriately labeled as effective against
TB mycobacterium tuberculosis (per AAMI TIR-12, 2004, section 5.2.3)
Caution Use only the cleaning and disinfection methods
described in this manual. Clean the illuminator prior to
disinfection.
To clean and disinfect the illuminator
1.After using the illuminator, unplug any power cords.
2.Clean the illuminator.
a.Following the wipe manufacturer's instructions, wipe the entire
unit to completely remove any gross debris.
b.Gently wipe the gold contacts on the bottom with 70% alcohol.
If debris builds up on these contacts, the charging or auxiliary
power functions might be affected.
3.Disinfect the illuminator.
a.Following the wipe manufacturer's instructions, thoroughly wet the
surface of the entire illuminator.
b.Using sterile gauze soaked with 70% isopropyl alcohol, thoroughly
wet the surface of the entire illuminator. The treated surface must
remain wet for five minutes. If needed, reapply the solution to keep
the surface continuously wet during this time.
c.Allow the surface to dry.
To clean the charging base or auxiliary power cord
1.After using the charging base or auxiliary power cord, unplug the power
cord.
2.Clean the component. Following the wipe manufacturer's instructions,
wipe the surface to completely remove any gross debris.
Directions for useCordless illumination system13
Inspecting the system
Examine all components of the cordless illumination system
monthly. Components include the illuminator, charging station,
and auxiliary power cord.
If any component is worn or damaged, replace it with the Welch
Allyn-approved part. For ordering information, see “Part
numbers” on page 17.
Recycling the system components
Within the EU
Do not dispose of this product as unsorted municipal waste.
Prepare it for reuse or separate collection as specified by
Directive 2002/96/EC of the European Parliament and the
Council of the European Union on Waste Electronic and
Electrical Equipment (WEEE).
If this product is contaminated, this directive does not apply.
For more specific disposal information, see
www.welchallyn.com/weee, or contact Welch Allyn.
Outside the EU
When the product reaches its end of life, recycle the components
according to national, state, and local regulations, or return the
product to Welch Allyn.
14Welch Allyn
Troubleshooting
If you try these suggestions and still have problems, contact
Welch Allyn.
ProblemProbable causes & solutions
LED does not light.Battery is low. Charge the illuminator.
Illuminator is not seated correctly in speculum. Insert
illuminator fully into speculum.
LED does not light when the
auxiliary power cord is
attached.
LED does not automatically
turn off when removing the
illuminator from the
speculum.
Illuminator does not charge
in the charging base.
Contacts
For information about any Welch Allyn product, call the nearest
Welch Allyn representative:
Auxiliary power cord is not plugged in. Plug the
auxiliary power cord into an AC outlet.
Auxiliary power cord is not seated properly. Plug the
cord into the illuminator properly.
Power switch is damaged or blocked. Clean the switch
area. If the problem continues, replace the illuminator.
Charging base is not plugged in. Connect charging
base to AC power.
Illuminator is not seated properly in charging base.
Reinsert the illuminator into the charging base, and
verify that the status indicator glows amber.
Australia+ 6129 638 3000
+ 1 315 685 4560
Directions for useCordless illumination system15
Appendix
Specifications
ItemSpecification
Auxiliary power cord transformer
Classification
Input
Output
Category
Charging station transformer
Classification
Input
Output
Category
Power cord length
Auxiliary power cord
Charging station
U.S. & Canada
Class II and internally
powered, type BF applied
part
120V~60Hz
5.5 VDC, 300 mA
Not AP/APG equipment
U.S. & Canada
Class I and internally
powered
120V~60Hz
9.8 VDC, 150 mA
Not AP/APG equipment
305 cm (120 in)
92 cm (36 in)
International
Class I and internally
powered, type BF applied
part
100 - 240 VAC 200-100 mA
5.5 VDC, 300 mA
Not AP/APG equipment
International
Class I, Class II and internally
powered
230 V (240 V in Australia) ~
50 Hz
7.5 – 9.2 VDC , 200 mA
Not AP/APG equipment
Weight
Illuminator
Charging station
(base and power cord)
Auxiliary power cord
Battery cell
Capacity
Voltage
Chemistry
Protection against ingress of
water, per IEC 60529 (illuminator,
auxiliary power cord, charging
station)
Operating conditions
Temperature
Relative humidity
Altitude
48 g (0.1 lb)
142 g (0.31 lb)
290 g (0.64 lb)
420 mAh
3.7 V
Li-ion (single lithium ion)
IPXØ
+10° C to +35° C (+50° F to +95° F)
30% to 75% noncondensing
700 to 1060 hPa
16Welch Allyn
Item (continued)Specification (continued)
Shipping and storage conditions
Temperature
Relative humidity
Altitude
Intermittent operation
Illuminator with speculum
Illuminator alone
-20° C to +49° C (-4° F to +120° F)
15% to 95% noncondensing
500 hPa to 1060 hPa
10 minutes on, 5 minutes off between uses
2 minutes on, 5 minutes off between uses
Safety, EMC, and regulatory
compliance
LED radiation output< 6.67 mW at 400 – 700 nm wavelengths
CAN/CSA C22.2 No. 601.1-M90, including amendments 1 &
2
EN/IEC 60601-1
EN/IEC 60601-1-2
EN/IEC 60825-1
UL 60601-1
AS/NZS CISPR11
Directions for useCordless illumination system17
Part numbers
ItemOrder Number
Cordless illuminator system
with charging station
Cordless illuminator system
with charging station and
auxiliary power cord
Cordless illuminatorU.S. & Canada: 79900
Charging station
(base and power cord)
Auxiliary power cordU.S. & Canada: 75910
* Specify country and voltage.
U.S. & Canada: 79910
International:79915*
U.S. & Canada: 79920
International:79925*
U.S. & Canada: 73910
International:73915*
International:75915*
REF
18Welch Allyn
Limited warranty
Welch Allyn warrants that the KleenSpec vaginal speculum cordless illumination system
products meet the labeled specifications of the Products and will be free from defects in
materials and workmanship that occur within one year after the date of purchase.
The date of purchase is the date specified in our records, if you purchased the Product
directly from us or 30 days after the date on which the Product was sold to the dealer
from whom you bought the Product, as documented in our records.
This warranty does not cover damage caused by: 1) handling during shipping, 2) use or
maintenance contrary to labeled instructions, 3) alteration or repair by anyone not
authorized by Welch Allyn, and 4) accidents.
If a Product or accessory covered by this warranty is determined to be defective because
of defective materials, components, or workmanship, and the warranty claim is made
within the warranty period described above, Welch Allyn will, at its discretion, repair or
replace the defective Product or accessory free of charge. If your Product requires repairs
covered by this warranty, upon your request Welch Allyn will loan to you, at no cost, a
substitute Product for use until your repaired Product is returned.
You must obtain a return authorization from Welch Allyn to return your Product before
you send it to Welch Allyn’s designated service center for repair. Contact Welch Allyn
Technical Support.
THIS WARRANTY IS IN LIEU OF ALL OTHER WARRANTIES, EXPRESS OR IMPLIED,
INCLUDING BUT NOT LIMITED TO THE IMPLIED WARRANTIES OF MERCHANTABILTY
AND FITNESS FOR A PARTICULAR PURPOSE. WELCH ALLYN'S OBLIGATION UNDER THIS
WARRANTY IS LIMITED TO REPAIR OR REPLACEMENT OF PRODUCTS CONTAINING A
DEFECT. WELCH ALLYN IS NOT RESPONSIBLE FOR ANY INDIRECTOR CONSEQUENTIAL
DAMAGES RESULTING FROM A PRODUCT DEFECT COVERED BY THE WARRANTY.
Directions for useCordless illumination system19
Service policy
All repairs on products under warranty must be performed or approved by Welch Allyn.
Unauthorized repairs will void the warranty. In addition, whether or not covered under
warranty, any product repair shall exclusively be performed by Welch Allyn certified
service personnel.
If the product fails to function properly — or if you need assistance, service, or spare
parts — contact Welch Allyn.
Before contacting Welch Allyn, try to duplicate the problem, and check all accessories to
ensure that they are not causing the problem. When calling, please be prepared to
provide:
• Product name and model number and complete description of the problem.
• Serial number of your product (if applicable).
• Complete name, address and phone number of your facility.
• For out-of-warranty repairs or spare parts orders, a purchase order (or credit card)
number.
• For parts orders, the required spare or replacement part numbers.
If your product requires warranty, extended warranty, or non-warranty repair service,
please call Welch Allyn. A representative will assist you troubleshooting the problem and
will make every effort to solve it over the phone, avoiding potential unnecessary returns.
In case a return cannot be avoided, the representative will record all necessary
information and will provide a Return Material Authorization (RMA) number, as well as
the appropriate return address. An RMA number must be obtained prior to any return.
If you have to return goods for service, follow these recommended packing instructions:
• Remove all hoses, cables, sensors, power cords, and ancillary products (as
appropriate) before packing, unless you suspect they are associated with the
problem.
• Wherever possible use the original shipping carton and packing materials.
• Include a packing list and the Welch Allyn Return Material Authorization (RMA)
number.
It is recommended that all returned goods be insured. Claims for loss or damage to the
product must be initiated by the sender.
20Welch Allyn
EMC guidance and manufacturer’s declarations
Table 1. Emissions for all EQUIPMENT and SYSTEMS
The 790 series cordless illumination system is intended for use in the electromagnetic
environment specified below. The customer or user of the 790 series cordless illumination system
should assure that it is used in such an environment.
Voltage fluctuations/
flicker emissions
IEC 61000-3-3
Group 1The790 series cordless illumination system uses RF energy
Class BThe 790 series cordless illumination system is suitable for
Class A
Complies
only for its internal function. Therefore, its RF emissions
are very low and are not likely to cause any interference in
nearby electronic equipment.
use in all establishments including domestic and those
directly connected to the public low-voltage power supply
network that supplies buildings used for domestic
purposes.
Directions for useCordless illumination system21
Table 2. Immunity for all EQUIPMENT and SYSTEMS
The 790 series cordless illumination system is intended for use in the electromagnetic environment
specified below. The customer or user of the 790 series cordless illumination system should assure
that it is used in such an environment.
Immunity testIEC 60601
Electrostatic
discharge (ESD)
IEC 61000-4-2
Electrical fast
transient (EFT)/
burst
IEC 61000-4-4
Surge
IEC 61000-4-5
Voltage dips, short
interruptions, and
voltage variations
on power supply
input lines.
IEC 61000-4-11
Power frequency
(50/60Hz)
magnetic field
IEC 61000-4-8
test level
± 6 kV contact
±8kVair
±2 kV mains
±1 kV I/O
±1 kV differential
±2 kV common
>95% dip in
0.5 cycle
60% dip in
5 cycles
30% dip for
25 cycles
>95% dip in
5 seconds
3 A/m3 A/mPower frequency magnetic fields
Compliance
level
± 6 kV contact
±8kVair
±2 kV mains
Not applicable
±1 kV differential
±2 kV common
>95% dip in
0.5 cycle
60% dip in
5 cycles
30% dip for
25 cycles
>95% dip in
5 seconds
Electromagnetic environment guidance
Floors should be wood, concrete, or
ceramic tile. If floors are covered with
synthetic material, the relative
humidity should be at least 30%.
Mains power quality should be that of
a typical commercial or hospital
environment.
Mains power quality should be that of
a typical commercial or hospital
environment.
Mains power quality should be that of
a typical commercial or hospital
environment. If the user of the 790
series cordless illumination system
requires continued operation during
power mains interruptions, it is
recommended that the 790 series
cordless illumination system be
powered from an uninterruptible
power supply or battery.
should be at levels characteristic of a
typical location in a typical
commercial or hospital environment.
22Welch Allyn
Table 3. Electromagnetic immunity
The 790 series cordless illumination system is intended for use in the electromagnetic environment
specified below. The customer or user of the 790 series cordless illumination system should assure
that it is used in such an environment.
Immunity
test
Conducted
RF, IEC
61000-4-6
Radiated
RF, IEC
61000-4-3
IEC 60601
test level
3 Vrms
150 kHz to
80 MHz
3 V/m
80 MHz to
2.5 GHz
Compliance
Electromagnetic environment - guidance
level
Portable and mobile RF communications equipment
should be used no closer to any part of the 790 series
cordless illumination system, including cables, than the
recommended separation distance calculated from the
equation applicable to the frequency of the transmitter.
Recommended separation distance
3 Vrmsd = (1.17)
3 V/md = (1.17)80 MHz to 800 MHz
d = (2.33)800 MHz to 2.5 GHz
P
P
P
where P is the maximum output power rating of the
transmitter in watts according to the transmitter
manufacturer and d is the recommended separation
distance in meters.
Field strengths from fixed RF transmitters, as
determined by an electromagnetic site survey,
be less than the compliance level in each frequency
b
range.
Interference may occur in the vicinity of equipment
marked with the following symbol:
a
should
Directions for useCordless illumination system23
Table 3. Electromagnetic immunity (continued)
Note 1: At 80 MHz and 800 MHz, the higher frequency range applies.
Note 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected
by absorption and reflection from structures, objects, and people.
a
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless)
telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast
cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to
fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field
strength in the location in which the 790 series cordless illumination system is used exceeds the
applicable RF compliance level above, the 790 series cordless illumination system should be
observed to verify normal operation. If abnormal performance is observed, additional measures
may be necessary, such as reorienting or relocating the 790 series cordless illumination system.
b
Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.
24Welch Allyn
Table 4. Recommended separation distances between portable and mobile
RF communications equipment and the 790 series cordless illumination system
The 790 series cordless illumination system is intended for use in an electromagnetic environment
in which radiated RF disturbances are controlled. The customer or user of the 790 series cordless
illumination system can help prevent electromagnetic interference by maintaining a minimum
distance between portable and mobile RF communications equipment (transmitters) and the 790
series cordless illumination system as recommended below, according to the maximum output
power of the communications equipment.
Rated max.
output power
of transmitter
(W)
0.010.1170.1170.233
0.10.370.370.74
11.171.172.33
103.703.707.37
10011.7011.7023.30
For transmitters rated at a maximum output power not listed above, the recommended separation
distance d in meters (m) can be estimated using the equation applicable to the frequency of the
transmitter, where P is the maximum output power rating of the transmitter in watts (W) according
to the transmitter manufacturer.
Note 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
Note 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected
by absorption and reflection from structures, objects, and people.
Separation distance according to frequency of transmitter (m)
150 kHz to 80 MHz
d = (1.17)
PPP
80 MHz to 800 MHz
d = (1.17)
800 MHz to 2.5 GHz
d = (2.33)
4341 State Street Road, PO Box 220, Skaneateles Falls, NY 13153-0220 USA