Welch Allyn KleenSpec User Manual

KleenSpec
®
790 Series
Cordless Illumination System
for the 590 Series Vaginal Speculum
REF
Directions for Use
79910
© 2012 Welch Allyn. All rights are reserved. To support the intended use of the product described in this publication, the purchaser of the product is permitted to copy this publication, for internal distribution only,from the media provided by Welch Allyn.
Caution: Federal US law restricts sale of the device identified in this manual to, or on the order of, a licensed physician.
Welch Allyn assumes no responsibility for any injury, or for any illegal or improper use of the product, that may result from failure to use this product in accordance with the instructions, cautions, warnings, or indications for use published in this manual.
Welch Allyn and KleenSpec are registered trademarks of Welch Allyn. CaviWipe is a registered trademark of Metrex Research Corporation.
Caution: Changes or modifications not expressly approved by Welch Allyn could void the purchaser's authority to operate the equipment.
Patents pending.
Mat. Number: 718913 , Ver: A
Welch Allyn 4341 State Street Road, PO Box 220 Skaneateles Falls, NY 13153-0220
www.welchallyn.com
Printed in USA
Contents
Introduction ............................1
Charging the illuminator .................10
Using the illuminator for a pelvic exam.....11
Using the illuminator as light source .......11
Maintenance ...........................12
iii
Intended use .................................1
Contraindications ..............................1
User profile ..................................2
Symbols .....................................2
Using the system safely ........................4
Utilisation sécuritaire du système .................7
Cleaning and disinfecting .......................12
Inspecting the system .........................13
Recycling the system components ...............13
Troubleshooting ........................14
Contacts ..............................14
Appendix..............................15
Specifications................................15
Part numbers ................................17
Limited warranty .............................18
Service policy ................................19
EMC guidance and manufacturer’s declarations .....20
iv
Directions for use Cordless illumination system 1

Introduction

The Welch Allyn KleenSpec®790 series cordless illumination system can be used with or without the KleenSpec 590 series disposable vaginal speculum. The cordless illumination system consists of the following items.
Cordless illuminator
Charging station (base and power cord)

Intended use

When used with the vaginal speculum, the cordless illuminator provides illumination during pelvic examinations and other gynecological procedures, such as pap smears, dilation and curettage (D&C) biopsy, and electrosurgery.
The cordless illuminator can be used independent of the speculum as a general-purpose light source.

Contraindications

The 790 series illuminator (either by itself or in conjunction with a 590 series speculum) is not intended to be used for eye examinations.
The 790 series illuminator (either by itself or in conjunction with a 590 series speculum) is not intended to be used for diagnosis.
Auxiliary power cord (optional)
2 Welch Allyn

User profile

The intended users of the illuminator are medical professionals who are qualified to perform a pelvic examination.

Symbols

The symbols illustrated on the following pages may appear in this manual or on the illuminator, charging station, auxiliary power cord, packaging, or shipping container.
Table 1. Documentation symbols
WARNING
Caution In this manual, this symbol and “Caution”
In this manual, this symbol and “WARNING” indicate conditions or practices that could lead to illness, injury, or death.
indicate conditions or practices that could damage the equipment or other property.
On the product, this symbol means consult operating instructions.
Table 2. Shipping, storing, and environment symbols
This end up Keep dry
95%
15%
-20C
Fragile Relative humidity limit
+49C
Temperature limits
Directions for use Cordless illumination system 3
Table 3. Certification symbols
Meets essential requirements of European Medical Device Directive 93/42/EEC.
EC REP
Regulatory Affairs Representative Welch Allyn Limited Navan Business Park Dublin Road, Navan County Meath, Republic of Ireland
CERTIFIED TO: CAN/CSA STD C22.2 NO. 601.1 CONFORMS TO: IEC 60601-1, IEC60601-1-2, UL 60601-1
EMC framework of Australia
N344
Table 4. Operation symbols
Caution, hot surface
The speculum is a Type BF applied part
Alternating current
Direct current
Double insulation
IPXØ
Not protected from the ingress of liquid.
Recycle the product separately from other disposables. See “Recycling
the system components” on page 13.
Power plug
4 Welch Allyn

Using the system safely

Before using the cordless illumination system, you must read and understand the following safety-related information.

General warnings

The following warning statements apply to use of the cordless illumination system in general. Warning statements that apply specifically to particular procedures, such as cleaning, appear in the corresponding sections of the manual.
Warning statements indicate conditions or practices that could lead to illness, injury, or death.
WARNING To avoid the risk of electric shock, this equipment must only be connected to a supply mains with protective earth.
WARNING No modification to this equipment is allowed.
WARNING The illuminator may generate heat after prolonged
use.
Operation time limits: Illuminator with speculum
10 minutes on, 5 minutes off between uses. Illuminator alone 2 minutes on, 5 minutes off between uses.
Illuminator could be harmful to skin if touched after
being on for extended times.
To avoid fire, do not place illuminator onto a
flammable surface if LED is on.
Illuminator is not suitable for use in the presence of a
flammable anesthetic mixture with air or oxygen or nitrous oxide. An explosion may result.
Directions for use Cordless illumination system 5
WARNING To prevent shock hazards, the power outlet must be properly grounded. Grounding reliability is achieved only when connected to hospital-use or hospital-grade receptacles.
WARNING The illuminator is not intended to be used for eye examinations.
WARNING The illuminator is not intended to be used for diagnosis.
WARNING The illuminator is intended to be used by trained medical professionals.
WARNING This product comes with a power supply and/or power cord that is intended for use only with this product. The power supply and/or power cord have not been tested and approved for use with other products that may have the same power connectors. If you cannot locate the original power supply and/or power cord, please contact Welch Allyn to obtain replacement parts.

General cautions

The following caution statements apply to use of the cordless illumination system in general. Caution statements that apply specifically to particular procedures, such as pelvic examinations, appear in the corresponding sections of the manual.
Caution statements indicate conditions or practices that could damage the equipment or other property.
Caution Disconnect the AC power cord to remove power to the device.
Caution The cordless illuminator could be a biohazard after use. Follow good hospital practices. See “Cleaning and
disinfecting” on page 12.
Caution Cordless illuminator, charging station, and auxiliary power cord are not protected from the ingress of liquid.
Caution There are no user-serviceable parts inside the device. In case of a malfunction, call Welch Allyn and precisely describe the problem.
6 Welch Allyn
Caution This device complies with current required standards for electromagnetic interference and should not present problems to other equipment or be affected by other devices. As a precaution, avoid using this device in close proximity to other equipment.
Caution Use only Welch Allyn-approved parts. See “Part
numbers” on page 17.
Directions for use Cordless illumination system 7

Utilisation sécuritaire du système

Avant d'utiliser le système d’éclairage sans fil, vous devez bien lire et comprendre les renseignements de sécurité suivants.

Avertissements généraux

Les avertissements suivants s’appliquent à l'utilisation du système d’éclairage sans fil en général. Les avertissements qui s'appliquent à des procédures particulières, comme le nettoyage, sont abordés dans les sections correspondantes de ce manuel.
Ces avertissements indiquent des conditions ou des pratiques qui pourraient causer des maladies, des blessures ou des décès.
AVERTISSEMENT Pour éviter les risques de choc électrique, cet équipement doit seulement être branché sur une alimentation secteur dotée d'une mise à la masse. AVERTISSEMENT Aucune modification de cet équipement n'est autorisée.
AVERTISSEMENT Le système d’éclairage peut dégager de la chaleur après une utilisation prolongée.
Les durées d'opération maximales : Illuminateur avec
spéculum 10 minutes de marche, 5 minutes d’arrêt entre les utilisations. Illuminateur seul 2 minutes de marche,  5 minutes d’arrêt entre les utilisations.
Le système d’éclairage peut causer des brûlures cutanées
si vous y touchez après une utilisation prolongée.
Pour éviter les risques d'incendie, ne déposez pas le
système d’éclairage sur une surface inflammable lorsque la DEL est allumée.
Le système d’éclairage ne devrait pas être utilisé en
présence d'un mélange d'anesthésique inflammable et de l’air ou encore d'oxygène ou d'oxyde nitreux. Cela pourrait causer une explosion.
8 Welch Allyn
AVERTISSEMENT La prise électrique doit être correctement mise à la terre afin d’éviter les chocs électriques. Vous ne pouvez être certain de la mise à la terre que si l'appareil est branché dans une prise utilisée par les hôpitaux ou de qualité hospitalière.
AVERTISSEMENT Le système d’éclairage ne devrait pas être utilisé pour les examens des yeux.
AVERTISSEMENT Le système d’éclairage ne devrait pas être utilisé pour effectuer un diagnostic.
AVERTISSEMENT Le système d’éclairage est conçu pour être utilisé par des professionnels médicaux formés.
AVERTISSEMENT Un bloc d'alimentation et/ou un cordon d'alimentation, destinés à une utilisation avec ce produit seulement, sont fournis avec le produit. Le bloc d'alimentation et/ou le cordon d'alimentation n'ont pas été testés et approuvés pour une utilisation avec d'autres produits qui pourraient être munis des mêmes connecteurs d'alimentation. Si vous n'avez plus le bloc d'alimentation et/ou le cordon d'alimentation originaux, veuillez communiquer avec Welch Allyn pour obtenir des pièces de remplacement.
Directions for use Cordless illumination system 9

Mises en garde générales

Les mises ne garde suivantes s’appliquent à l'utilisation du système d’éclairage sans fil en général. Les énoncés de mise en garde qui s'appliquent à des procédures particulières, comme les examens pelviens, sont abordées dans les sections correspondantes de ce manuel.
Les énoncés de mise en garde indiquent des conditions ou des pratiques qui pourraient endommager les équipements ou la propriété.
Mise en garde Débrancher le cordon d'alimentation c.a. pour mettre l'appareil hors tension.
Mise en garde Le système d’éclairage sans fil pourrait être considéré à risque biologique après utilisation. Conformez­vous aux bonnes pratiques hospitalières. Voir “Cleaning and
disinfecting” à la page 12.
Mise en garde Le système d’éclairage sans fil, la borne de recharge et le cordon d'alimentation auxiliaire ne sont pas protégés contre les infiltrations de liquide.
Mise en garde Cet appareil ne contient aucune pièce à entretenir. En cas de mauvais fonctionnement, appelez Welch Allyn et décrivez le problème de manière précise.
Mise en garde Cet appareil est conforme aux normes actuelles obligatoires sur les interférences électromagnétiques. Il ne devrait présenter aucun problème pour les autres équipements et ne devrait pas être affecté par d'autres appareils. Par mesure de précaution, évitez d'utiliser cet appareil à proximité d'autres équipements.
Mise en garde N'utilisez que des pièces approuvées par Welch Allyn. Voir “Part numbers” à la page 17.
10 Welch Allyn

Charging the illuminator

Charge the illuminator before using it the first time, or whenever its low-charge indicator turns on.
LED
Power switch
Low-charge indicator:
On = battery low
To charge the illuminator
1. If LED is lit, turn power switch off.
2. Connect charging base to AC power.
3. Place the illuminator (either direction) into the charging station. Full charging takes 6 to 8 hours.
Status indicator:
Off = no AC power Green = AC power Amber (illuminator inserted) = charging. Full charging takes 6 to 8 hours.
4. Remove the charged illuminator when you are ready to use it.
It is safe to leave the illuminator in the charging station after it is charged.
Directions for use Cordless illumination system 11

Using the illuminator for a pelvic exam

1. Fully insert the illuminator into a 590 series speculum (a). The LED automatically turns on.
2. Conduct the examination.
3. If the LED turns off, do not remove the speculum. Do either of these things:
Leave the illuminator in place, and
use an external light source.
or
Connect the auxiliary power cord
to AC power. Hold the speculum in place, and remove the illuminator. Connect the auxiliary power cord to illuminator (b). Reinsert the illuminator.
a
b
4. When finished, remove the speculum, and then remove the illuminator. The LED automatically turns off.

Using the illuminator as light source

To use the illuminator as a light source, slide the power switch to turn it on (all the way down) and off (all the way up).
Note
For operation time limits, see
“Intermittent operation” on page 16.
Power switch
12 Welch Allyn

Maintenance

Cleaning and disinfecting

Use CaviWipe®wipes for both cleaning and disinfecting.*
*EPA-approved intermediate-level disinfectant appropriately labeled as effective against TB mycobacterium tuberculosis (per AAMI TIR-12, 2004, section 5.2.3)
Caution Use only the cleaning and disinfection methods described in this manual. Clean the illuminator prior to disinfection.
To clean and disinfect the illuminator
1. After using the illuminator, unplug any power cords.
2. Clean the illuminator.
a. Following the wipe manufacturer's instructions, wipe the entire
unit to completely remove any gross debris.
b. Gently wipe the gold contacts on the bottom with 70% alcohol.
If debris builds up on these contacts, the charging or auxiliary power functions might be affected.
3. Disinfect the illuminator.
a. Following the wipe manufacturer's instructions, thoroughly wet the
surface of the entire illuminator.
b. Using sterile gauze soaked with 70% isopropyl alcohol, thoroughly
wet the surface of the entire illuminator. The treated surface must remain wet for five minutes. If needed, reapply the solution to keep the surface continuously wet during this time.
c. Allow the surface to dry.
To clean the charging base or auxiliary power cord
1. After using the charging base or auxiliary power cord, unplug the power cord.
2. Clean the component. Following the wipe manufacturer's instructions, wipe the surface to completely remove any gross debris.
Directions for use Cordless illumination system 13

Inspecting the system

Examine all components of the cordless illumination system monthly. Components include the illuminator, charging station, and auxiliary power cord.
If any component is worn or damaged, replace it with the Welch Allyn-approved part. For ordering information, see “Part
numbers” on page 17.

Recycling the system components

Within the EU
Do not dispose of this product as unsorted municipal waste. Prepare it for reuse or separate collection as specified by Directive 2002/96/EC of the European Parliament and the Council of the European Union on Waste Electronic and Electrical Equipment (WEEE).
If this product is contaminated, this directive does not apply.
For more specific disposal information, see
www.welchallyn.com/weee, or contact Welch Allyn.
Outside the EU
When the product reaches its end of life, recycle the components according to national, state, and local regulations, or return the product to Welch Allyn.
14 Welch Allyn

Troubleshooting

If you try these suggestions and still have problems, contact Welch Allyn.
Problem Probable causes & solutions
LED does not light. Battery is low. Charge the illuminator.
Illuminator is not seated correctly in speculum. Insert illuminator fully into speculum.
LED does not light when the auxiliary power cord is attached.
LED does not automatically turn off when removing the illuminator from the speculum.
Illuminator does not charge in the charging base.

Contacts

For information about any Welch Allyn product, call the nearest Welch Allyn representative:
USA 1 800 535 6663
Canada 1 800 561 8797 China + 86 216 327 9631 European Call Center + 353 46 906 7790 France + 3315 569 5849 Germany + 49 747 792 7186 Japan + 81 42 703 6084 Latin America + 1 305 669 9003 Netherlands + 31 202 061 360 Singapore + 656 419 8100 South Africa + 2711 777 7555 United Kingdom + 44 207 365 6780 Sweden + 46 85 853 6551
Auxiliary power cord is not plugged in. Plug the auxiliary power cord into an AC outlet.
Auxiliary power cord is not seated properly. Plug the cord into the illuminator properly.
Power switch is damaged or blocked. Clean the switch area. If the problem continues, replace the illuminator.
Charging base is not plugged in. Connect charging base to AC power.
Illuminator is not seated properly in charging base. Reinsert the illuminator into the charging base, and verify that the status indicator glows amber.
Australia + 6129 638 3000
+ 1 315 685 4560
Directions for use Cordless illumination system 15

Appendix

Specifications

Item Specification
Auxiliary power cord transformer
Classification
Input Output Category
Charging station transformer
Classification
Input
Output Category
Power cord length
Auxiliary power cord Charging station
U.S. & Canada Class II and internally powered, type BF applied part 120V~60Hz
5.5 VDC, 300 mA Not AP/APG equipment
U.S. & Canada Class I and internally powered 120V~60Hz
9.8 VDC, 150 mA Not AP/APG equipment
305 cm (120 in) 92 cm (36 in)
International Class I and internally powered, type BF applied part 100 - 240 VAC 200-100 mA
5.5 VDC, 300 mA Not AP/APG equipment
International Class I, Class II and internally powered 230 V (240 V in Australia) ~ 50 Hz
7.5 – 9.2 VDC , 200 mA Not AP/APG equipment
Weight
Illuminator Charging station (base and power cord) Auxiliary power cord
Battery cell
Capacity Voltage Chemistry
Protection against ingress of water, per IEC 60529 (illuminator, auxiliary power cord, charging station)
Operating conditions
Temperature Relative humidity Altitude
48 g (0.1 lb) 142 g (0.31 lb)
290 g (0.64 lb)
420 mAh
3.7 V Li-ion (single lithium ion)
IPXØ
+10° C to +35° C (+50° F to +95° F) 30% to 75% noncondensing 700 to 1060 hPa
16 Welch Allyn
Item (continued) Specification (continued)
Shipping and storage conditions
Temperature Relative humidity Altitude
Intermittent operation
Illuminator with speculum Illuminator alone
-20° C to +49° C (-4° F to +120° F) 15% to 95% noncondensing 500 hPa to 1060 hPa
10 minutes on, 5 minutes off between uses2 minutes on, 5 minutes off between uses
Safety, EMC, and regulatory compliance
LED radiation output < 6.67 mW at 400 – 700 nm wavelengths
CAN/CSA C22.2 No. 601.1-M90, including amendments 1 & 2 EN/IEC 60601-1 EN/IEC 60601-1-2 EN/IEC 60825-1 UL 60601-1 AS/NZS CISPR11
Directions for use Cordless illumination system 17

Part numbers

Item Order Number
Cordless illuminator system with charging station
Cordless illuminator system with charging station and auxiliary power cord
Cordless illuminator U.S. & Canada: 79900
Charging station (base and power cord)
Auxiliary power cord U.S. & Canada: 75910
* Specify country and voltage.
U.S. & Canada: 79910 International: 79915*
U.S. & Canada: 79920 International: 79925*
U.S. & Canada: 73910 International: 73915*
International: 75915*
REF
18 Welch Allyn

Limited warranty

Welch Allyn warrants that the KleenSpec vaginal speculum cordless illumination system products meet the labeled specifications of the Products and will be free from defects in materials and workmanship that occur within one year after the date of purchase.
The date of purchase is the date specified in our records, if you purchased the Product directly from us or 30 days after the date on which the Product was sold to the dealer from whom you bought the Product, as documented in our records.
This warranty does not cover damage caused by: 1) handling during shipping, 2) use or maintenance contrary to labeled instructions, 3) alteration or repair by anyone not authorized by Welch Allyn, and 4) accidents.
If a Product or accessory covered by this warranty is determined to be defective because of defective materials, components, or workmanship, and the warranty claim is made within the warranty period described above, Welch Allyn will, at its discretion, repair or replace the defective Product or accessory free of charge. If your Product requires repairs covered by this warranty, upon your request Welch Allyn will loan to you, at no cost, a substitute Product for use until your repaired Product is returned.
You must obtain a return authorization from Welch Allyn to return your Product before you send it to Welch Allyn’s designated service center for repair. Contact Welch Allyn Technical Support.
THIS WARRANTY IS IN LIEU OF ALL OTHER WARRANTIES, EXPRESS OR IMPLIED, INCLUDING BUT NOT LIMITED TO THE IMPLIED WARRANTIES OF MERCHANTABILTY AND FITNESS FOR A PARTICULAR PURPOSE. WELCH ALLYN'S OBLIGATION UNDER THIS WARRANTY IS LIMITED TO REPAIR OR REPLACEMENT OF PRODUCTS CONTAINING A DEFECT. WELCH ALLYN IS NOT RESPONSIBLE FOR ANY INDIRECTOR CONSEQUENTIAL DAMAGES RESULTING FROM A PRODUCT DEFECT COVERED BY THE WARRANTY.
Directions for use Cordless illumination system 19

Service policy

All repairs on products under warranty must be performed or approved by Welch Allyn. Unauthorized repairs will void the warranty. In addition, whether or not covered under warranty, any product repair shall exclusively be performed by Welch Allyn certified service personnel.
If the product fails to function properly — or if you need assistance, service, or spare parts — contact Welch Allyn.
Before contacting Welch Allyn, try to duplicate the problem, and check all accessories to ensure that they are not causing the problem. When calling, please be prepared to provide:
• Product name and model number and complete description of the problem.
• Serial number of your product (if applicable).
• Complete name, address and phone number of your facility.
• For out-of-warranty repairs or spare parts orders, a purchase order (or credit card)
number.
• For parts orders, the required spare or replacement part numbers.
If your product requires warranty, extended warranty, or non-warranty repair service, please call Welch Allyn. A representative will assist you troubleshooting the problem and will make every effort to solve it over the phone, avoiding potential unnecessary returns.
In case a return cannot be avoided, the representative will record all necessary information and will provide a Return Material Authorization (RMA) number, as well as the appropriate return address. An RMA number must be obtained prior to any return.
If you have to return goods for service, follow these recommended packing instructions:
• Remove all hoses, cables, sensors, power cords, and ancillary products (as
appropriate) before packing, unless you suspect they are associated with the problem.
• Wherever possible use the original shipping carton and packing materials.
• Include a packing list and the Welch Allyn Return Material Authorization (RMA)
number.
It is recommended that all returned goods be insured. Claims for loss or damage to the product must be initiated by the sender.
20 Welch Allyn

EMC guidance and manufacturer’s declarations

Table 1. Emissions for all EQUIPMENT and SYSTEMS
The 790 series cordless illumination system is intended for use in the electromagnetic environment specified below. The customer or user of the 790 series cordless illumination system should assure that it is used in such an environment.
Emissions test Compliance Electromagnetic environment - guidance
RF emissions CISPR 11
RF emissions CISPR 11
Harmonic emissions IEC 61000-3-2
Voltage fluctuations/ flicker emissions IEC 61000-3-3
Group 1 The790 series cordless illumination system uses RF energy
Class B The 790 series cordless illumination system is suitable for
Class A
Complies
only for its internal function. Therefore, its RF emissions are very low and are not likely to cause any interference in nearby electronic equipment.
use in all establishments including domestic and those directly connected to the public low-voltage power supply network that supplies buildings used for domestic purposes.
Directions for use Cordless illumination system 21
Table 2. Immunity for all EQUIPMENT and SYSTEMS
The 790 series cordless illumination system is intended for use in the electromagnetic environment specified below. The customer or user of the 790 series cordless illumination system should assure that it is used in such an environment.
Immunity test IEC 60601
Electrostatic discharge (ESD) IEC 61000-4-2
Electrical fast transient (EFT)/ burst IEC 61000-4-4
Surge IEC 61000-4-5
Voltage dips, short interruptions, and voltage variations on power supply input lines. IEC 61000-4-11
Power frequency (50/60Hz) magnetic field IEC 61000-4-8
test level
± 6 kV contact ±8kVair
±2 kV mains ±1 kV I/O
±1 kV differential ±2 kV common
>95% dip in
0.5 cycle
60% dip in 5 cycles
30% dip for 25 cycles
>95% dip in 5 seconds
3 A/m 3 A/m Power frequency magnetic fields
Compliance level
± 6 kV contact ±8kVair
±2 kV mains Not applicable
±1 kV differential ±2 kV common
>95% dip in
0.5 cycle
60% dip in 5 cycles
30% dip for 25 cycles
>95% dip in 5 seconds
Electromagnetic environment ­guidance
Floors should be wood, concrete, or ceramic tile. If floors are covered with synthetic material, the relative humidity should be at least 30%.
Mains power quality should be that of a typical commercial or hospital environment.
Mains power quality should be that of a typical commercial or hospital environment.
Mains power quality should be that of a typical commercial or hospital environment. If the user of the 790 series cordless illumination system requires continued operation during power mains interruptions, it is recommended that the 790 series cordless illumination system be powered from an uninterruptible power supply or battery.
should be at levels characteristic of a typical location in a typical commercial or hospital environment.
22 Welch Allyn
Table 3. Electromagnetic immunity
The 790 series cordless illumination system is intended for use in the electromagnetic environment specified below. The customer or user of the 790 series cordless illumination system should assure that it is used in such an environment.
Immunity test
Conducted RF, IEC 61000-4-6
Radiated RF, IEC 61000-4-3
IEC 60601 test level
3 Vrms 150 kHz to 80 MHz
3 V/m 80 MHz to
2.5 GHz
Compliance
Electromagnetic environment - guidance
level
Portable and mobile RF communications equipment should be used no closer to any part of the 790 series cordless illumination system, including cables, than the recommended separation distance calculated from the equation applicable to the frequency of the transmitter.
Recommended separation distance
3 Vrms d = (1.17)
3 V/m d = (1.17) 80 MHz to 800 MHz
d = (2.33) 800 MHz to 2.5 GHz
P
P
P
where P is the maximum output power rating of the transmitter in watts according to the transmitter manufacturer and d is the recommended separation distance in meters.
Field strengths from fixed RF transmitters, as determined by an electromagnetic site survey, be less than the compliance level in each frequency
b
range.
Interference may occur in the vicinity of equipment marked with the following symbol:
a
should
Directions for use Cordless illumination system 23
Table 3. Electromagnetic immunity (continued)
Note 1: At 80 MHz and 800 MHz, the higher frequency range applies.
Note 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects, and people.
a
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field strength in the location in which the 790 series cordless illumination system is used exceeds the applicable RF compliance level above, the 790 series cordless illumination system should be observed to verify normal operation. If abnormal performance is observed, additional measures may be necessary, such as reorienting or relocating the 790 series cordless illumination system.
b
Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.
24 Welch Allyn
Table 4. Recommended separation distances between portable and mobile RF communications equipment and the 790 series cordless illumination system
The 790 series cordless illumination system is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer or user of the 790 series cordless illumination system can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and the 790 series cordless illumination system as recommended below, according to the maximum output power of the communications equipment.
Rated max. output power of transmitter
(W)
0.01 0.117 0.117 0.233
0.1 0.37 0.37 0.74
1 1.17 1.17 2.33
10 3.70 3.70 7.37
100 11.70 11.70 23.30
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in meters (m) can be estimated using the equation applicable to the frequency of the transmitter, where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer.
Note 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
Note 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects, and people.
Separation distance according to frequency of transmitter (m)
150 kHz to 80 MHz
d = (1.17)
P P P
80 MHz to 800 MHz
d = (1.17)
800 MHz to 2.5 GHz
d = (2.33)
4341 State Street Road, PO Box 220, Skaneateles Falls, NY 13153-0220 USA
1 800 535 6663, + 1 315 685 4560 www.welchallyn.com
Mat. Number: 718913, Ver: A