The Welch Allyn 13500 Wireless LED Insight Binocular Indirect Ophthalmoscope (Insight
BIO) is a battery-powered device containing illumination and viewing optics intended for
use by medical professionals to examine the cornea, aqueous, lens, vitreous, and the
retina of the eye.
1
General operation and schematics
For illumination, the Welch Allyn Insight BIO uses a Light-Emitting Diode (LED). The
following optical schematic highlights the traditional principles that govern how the
device operates.
2
OverviewWelch Allyn® 13500 Insight™ Wireless LED Binocular Indirect Ophthalmoscope
Unpacking the ophthalmoscope
Open the package and verify that all components listed below are included. If any
component is missing, contact Welch Allyn immediately.
ItemQuantity
Insight BIO1
Insight battery2
Insight battery charger (USA, EU, UK plug)1
Insight 0 diopter eyepiece lens2
Insight directions for use1
Insight wall mount1
Accessories (available by order)
Part numberDescription
13500SInsight BIO kit complete with soft case
75360Insight BIO rechargeable battery
13510Insight BIO battery charger (USA, EU, UK plug)
13550Insight BIO teaching mirror
13560Insight BIO wall mount kit
13520Insight BIO 2.0D eyepiece lenses (qty 2)
13530Insight BIO 0.0D eyepiece lenses (qty 2)
13570Insight BIO soft case
13540Insight headband pad kit (all three pads)
13546Insight BIO eye cups (qty 2)
Symbols and definitions
Documentation symbols
WARNING The warning statements in this manual identify conditions or practices that could
lead to illness, injury, or death. Warning statements appear with a gray background in a black
and white document but have a yellow background in color hard copies.
3
CAUTION The caution statements in this manual identify conditions or practices that could
result in damage to the equipment or other property, or loss of data.
Consult directions for use (DFU).
A copy of the DFU is available on this website.
A printed copy of the DFU can be ordered from Welch Allyn for delivery within 7 days.
Symbols and definitionsWelch Allyn® 13500 Insight™ Wireless LED Binocular Indirect Ophthalmoscope
Temperature limitSerial Number
Product IdentifierGlobal Trade Item Number
Reorder number
Do not dispose of in trash, for
batteries only
Rechargeable battery
ManufacturerCE Mark for Class I equipment
By prescription or order of
physician
Do not dispose of in trash, for
devices
Authorized Representative in the
European Community
Warnings and cautions
All users of the Insight BIO must read and understand all safety information presented in
this manual before using the device.
WARNING The Welch Allyn 13500 Insight BIO is classified as a Group 2
instrument according to EN/ISO 15004-2:2007. The classification was
performed together with a Ø54mm/20 Diopter Ophthalmoscopy loupe.
5
WARNING The light emitted from this instrument is potentially hazardous.
The longer the duration of exposure, the greater the risk of ocular damage.
Exposure to light from this instrument when operated at maximum
intensity will exceed the safety guideline after 10 minutes.
WARNING Because prolonged intense light exposure can damage the
retina, the use of the Insight Binocular Indirect Ophthalmoscope for ocular
examination should not be unnecessarily prolonged, and the brightness
setting should not exceed what is needed to provide clear visualization of
the target structures. This device should be used with filters that eliminate
UV radiation (˂400 nm) and, whenever possible, filters that eliminate UV
short wavelength blue light (˂420 nm).
WARNING The retinal exposure dose for a photochemical hazard is a
product of the radiance and the exposure time. If the value of radiance
were reduced in half, twice the time would be needed to reach the
maximum exposure limit. While no accurate optical radiation hazards have
been identified for indirect ophthalmoscopes, it is recommended that the
intensity of light directed into the patient’s eye be limited to the minimum
level which is necessary for diagnosis. Infants and persons with aphakia or
other eye disease/disorders may be at greater risk. The risk may also
increase if the person being examined has had any exposure with the same
instrument or any other ophthalmic instrument using a visible light source
during the previous 24 hours. This will apply particularly if the eye has been
exposed to retinal photography.
CAUTION Charge the Insight BIO only with the provided charger. Using
any other charger could damage the instrument.
CAUTION Do not autoclave, immerse, excessively wet, or use germicidal
cleaners other than those identified. Doing so will damage the instrument.
CAUTION United States Federal law restricts this device to sale,
distribution, or use by or on the order of a physician or licensed healthcare
professional.
6
Warnings and cautionsWelch Allyn® 13500 Insight™ Wireless LED Binocular Indirect Ophthalmoscope
Controls, indicators, and connectors
Instrument top, right side view
7
No. FeatureDescription
1Headband adjustment knobsLoosen or tighten fit of the headband.
2BinocularProvides illumination and viewing optics to examine the fundus
(shown in OFF position).
3Battery packProvides power to the binocular. Rechargeable with battery
charger.