The Welch Allyn Home™ blood pressure
monitor is indicated for adult use in the
home or domestic setting only.
The Welch Allyn Home blood pressure
monitor features blood pressure
measurement via SureBP
pulse rate measurement, and data
management using Bluetooth
technology.
SureBP Technology is an
advanced blood pressure
measurement technology
that obtains a blood pressure
reading during the inflation cycle (as the
cuff pumps up). This technology allows
for a faster, more comfortable reading. If
on inflation, SureBP Technology is unable
to determine a blood pressure due to
excessive patient movement or any
arrhythmia, it will inflate the cuff to a
slightly higher pressure, then measure
the blood pressure while deflating the
cuff, providing an accurate reading with
less chance of a reading failure.
The design provides you with two years
of reliable service. Readings taken by the
monitor are equivalent to those obtained
®
Technology,
®
Smart
Page 6
2
Welch Allyn Home Blood Pressure Monitor
by a trained observer using the cuff and
stethoscope auscultation method.
This Directions for use contains
important safety and care information
and provides step by step instructions
for using the monitor. Read the manual
thoroughly before using the monitor.
Indications for use
The Welch Allyn Home blood pressure
monitor is a digital monitor intended for
use in measuring blood pressure and
heart rate in adult patient populations
with arm circumferences between 15 cm
to 54 cm (approximately 6 to 21 inches).
Contraindications
This device is contraindicated for any
person who is connected to a wearable
or implantable electronic device or
instrument, such as a pacemaker or
defibrillator.
This blood pressure monitor is not
intended to be a diagnostic device.
Contact your physician if hypertensive
values are indicated.
Page 7
Symbols
Documentation symbols
Directions for use 3
War ning
manual identify conditions or practices
that could lead to illness, injury, or death.
Caution
manual identify conditions or practices
that could result in damage to the
equipment or other property, or loss of
data.
Mandatory - Consult Directions for Use
: The warning statements in this
: The caution statements in this
Power symbols
Direct current (DC)
Stand-By
Page 8
4
Welch Allyn Home Blood Pressure Monitor
Shipping, storing, and environment
symbols
Humidity limitation
Separate collection of Electrical and
Electronic Equipment. Do not dispose
as unsorted municipal waste.
Temperature limit
Atmospheric pressure limitation
Recyclable
Stacking limit by number
Page 9
Directions for use 5
Blood pressure device and cuff
symbols
SureBP technology logo
Artery marker
Range
Index edge
Limb circumference
(Minimum/Maximum)
Lot code
Not made with natural
rubber latex
Page 10
6
Welch Allyn Home Blood Pressure Monitor
Miscellaneous symbols
Type BF applied part
Serial Number
Product Identifier
Manufacturer
Reorder Number
Non-ionizing electromagnetic
radiation
Global Trade Item Number
Class II equipment
Page 11
Directions for use 7
IP22Ingress protection: the device is
protected against solid foreign
objects of 12.5mm and greater
and against vertically falling
water drops when ENCLOSURE
is tilted up to 15°
Bluetooth
Page 12
8
Welch Allyn Home Blood Pressure Monitor
About warnings and cautions
Caution statements can appear on the
Welch Allyn Home blood pressure
device, the packaging, the shipping
container, or in this Directions for use.
The Welch Allyn Home blood pressure
device is safe for adults when used in
accordance with the instructions and
caution statements presented in this
Directions for use.
Before using the device, you must
familiarize yourself with all cautions, with
the steps to power up the device, and
with the sections of this Directions for use that pertain to your use of the
device. In addition to reviewing the
general cautions presented in the next
section, you must also review the more
specific cautions that appear throughout
the manual in conjunction with setup/
startup, operation, and maintenance
tasks.
•Failure to understand and observe
any warning statement in this
manual could lead to injury or illness.
•Failure to understand and observe
any caution statement in this manual
could lead to damage to the
Page 13
Directions for use 9
equipment or other property, or loss
of measurement data.
Warnings and cautions
WARNING Injury risk. The device is
not suitable for measuring the blood
pressure of neonatal infants or
children.
WARNING Injury risk. The device is
not suitable for use on pregnant
women, users with implanted
electrical devices, users with
preeclampsia, premature ventricular
beats, atrial fibrillation, peripheral
arterial disease, and users undergoing
intravascular therapy or arteriovenous shunt or people who received
a mastectomy. Consult your doctor
prior to using the unit if you suffer
from any of these illnesses or
conditions.
WARNING Injury risk. Do not take
any therapeutic measures on the basis
of a self measurement. Never alter the
dose of a medicine prescribed by a
doctor. Consult your doctor if you
have any question about your blood
pressure.
WARNING Injury risk. Keep the unit
out of reach of infants, children, or
pets, since inhalation or swallowing of
small parts is dangerous or even fatal.
Page 14
10
Welch Allyn Home Blood Pressure Monitor
WARNING Injury risk. The device is
not intended to be a selfmeasurement station used in a public
place or center. Use the device only in
your home or temporary residence.
WARNING Injury risk. Do not burn
batteries. Batteries may leak or
explode.
WARNING Injury risk. If you
experience discomfort during a
measurement, such as pain in the arm
or other complaints, press the power
button immediately to release the air
from the cuff. Loosen and remove the
cuff from your arm.
WARNING Injury risk. On the rare
occasion of a fault causing the cuff to
remain fully inflated during
measurement, open the cuff
immediately. Prolonged high pressure
applied to the arm (cuff pressure
>300mmHg or constant pressure
>15mmHg for more than 3 minutes)
might lead to bruising and discolored
skin.
WARNING Injury risk. This unit is not
suitable for continuous monitoring
during medical emergencies or
operations.
WARNING Injury risk. Taking blood
pressure measurements too
frequently could disrupt blood
circulation and cause injuries.
Page 15
Directions for use 11
WARNING Injury risk. Do not
kink the connection tube during
use. The cuff pressure might
continuously increase, which
could prevent blood flow and
result in injury.
WARNING Injury risk. Never
apply the cuff over damaged skin.
This can cause further injury.
WARNING Injury risk. If applying
to someone else, do not touch
output of the batteries/adapter
and the user simultaneously.
WARNING Injury risk. Do not use
the unit if you are allergic to
polyester or synthetic materials.
WARNING Injury risk. Excessive
tube lengths could cause
strangulation if you don't manage
them properly.
WARNING Injury risk. Do not
connect the air tube to other
medical equipment. This could
cause air to be pumped into
intravascular systems or high
pressure, which could lead to
serious injuries.
WARNING Injury risk. Dispose of
accessories, detachable parts, and
the device according to the local
guidelines.
Page 16
12
Welch Allyn Home Blood Pressure Monitor
WARNING Injury risk. Do not
service or perform any
maintenance while using the
device.
WARNING Injury risk. Use only
accessories approved by the
manufacturer. Using unapproved
accessories might cause damage
to the unit and injure users.
WARNING Injury risk. No
modification to this equipment is
allowed. Modifying the equipment
could damage the unit or
endanger the user.
Page 17
Directions for use 13
Caution This device is intended for
non-invasive measuring and
monitoring of arterial blood
pressure. It is not intended for use
on extremities other than arm or for
functions other than obtaining a
blood pressure measurement.
Caution Do not wrap the cuff on
the same arm to which another
monitoring device is applied. One or
both devices could temporarily stop
functioning if you try to use them
on the same arm at the same time.
Caution To avoid measurement
errors, avoid taking blood pressure
measurements near a strong
electromagnetic field radiated
interference signal or electrical fast
transient/burst signal.
Caution The materials of the cuff
have been tested and found to
comply with requirements of ISO
10993-5:2009 and ISO 1099310:2010. It will not cause any
potential sensitization or irritation
reaction.
Caution Before use, make sure the
device functions safely and is in
proper working condition.
Caution Use the device in the
environment described in this
directions for use. Otherwise, you
will compromise the device’s
performance and reduce its lifetime.
Page 18
14
Welch Allyn Home Blood Pressure Monitor
Caution Do not attempt to repair
the unit yourself if it malfunctions.
Only have repairs carried out by
authorized service centers.
Caution Report any unexpected
operation or events to the
manufacturer.
Caution Use a soft cloth to clean
the entire unit. Do not use any
abrasive or volatile cleaners. See
the cleaning instructions
presented later in this Directions for use.
Contents list
The following items are in the device
box:
•Blood Pressure Monitor
•D-Ring Standard Wide Cuff (22cm
to 42cm) with air tube attached
•(4) AA alkaline batteries
•Directions for use
•Quick reference guide
Page 19
Controls and indicators
Monitor front
No. FeatureDescription
1D-Ring
blood
pressure
cuff
2Power
button
3LCD
Display
Apply to upper arm to
take a blood pressure
measurement
Powers on the monitor,
starts and stops a blood
pressure measurement,
and — when pressed for 2
seconds — initiates a
Bluetooth
Displays blood pressure
reading and other
pertinent information
regarding the reading
Directions for use 15
pairing
Page 20
16
Welch Allyn Home Blood Pressure Monitor
Monitor back
No. FeatureDescription
1Direct current
power
connection
2 Battery
compartment
(behind
cover)
When used with an
accessory power
cord (not included),
connects the
monitor to a power
outlet
Houses 4 AA
alkaline batteries
Page 21
Directions for use 17
Power options
Caution To get optimal
performance and protect your
monitor, use only the correct
batteries or the Welch Allynapproved power adapter.
The monitor is powered by one of two
sources:
•4 AA alkaline batteries
•AC adapter (6v 1A) (not
included)
Page 22
18
Welch Allyn Home Blood Pressure Monitor
Screen elements
The liquid crystal display (LCD) displays
the following: systolic blood pressure
(mmHg), diastolic blood pressure
(mmHg), pulse rate (bpm), heart beat
(regular or irregular) while acquiring
blood pressure measurements, excessive
motion alert, Bluetooth connectivity
status or data transmission error, alarm
priority, and battery charge level.
Page 23
Directions for use 19
Symbol Description
Systolic blood pressure
mmHg = measurement unit
of the blood pressure
Diastolic blood pressure
mmHg = measurement unit
of the blood pressure
Pulse in beats per minute
Irregular heartbeat
Monitor is detecting an
irregular heartbeat during
measurement
Heart beat
Monitor is detecting a
heartbeat during
measurement
Battery indicator
Indicates the current battery
charge
Battery indicator
Indicates the current battery
charge
Page 24
20
Welch Allyn Home Blood Pressure Monitor
Symbol Description
Bluetooth
or data transmission error
Motion indicator
Motion may result in an
inaccurate measurement.
Reading out of range
Either SYS > 260mmHg or
DIA > 220mmHg. The symbol
may appear in either the SYS
or DIA area of the screen.
Alarm priority = Low (an!
appears near the top of the
screen)
Reading out of range
Either SYS < 50mmHg or DIA
<25mmHg. The symbol may
appear in either the SYS or
DIA area of the screen.
Alarm priority = Low (an!
appears near the top of the
screen)
connectivity status
Page 25
Directions for use 21
Measuring blood pressure
Overview
You are the intended operator of the
device. You can measure your blood
pressure and then save and send
measurement data to a smartphone or
tablet. You can also maintain your device
and its accessories, and you can change
the batteries under normal
circumstances.
To measure your blood pressure, you
need the following:
•Blood pressure monitor with
batteries inserted
•Blood pressure cuff ("D-ring" style
for easy, one-handed adjustment)
•Blood pressure air tube
To save and transfer blood pressure
measurements, you also need a
smartphone or tablet with Bluetooth
wireless connectivity and the application
or software to receive your data, such as
the Welch Allyn Home application or
another healthcare provider application.
Page 26
22
Welch Allyn Home Blood Pressure Monitor
Set up your equipment
Insert or replace the batteries
WARNING Injury risk. Do not
burn batteries. Batteries may leak
or explode.
Caution Remove the batteries if
the device is not used regularly.
Caution Dispose of old batteries
by following your local recycling
guidelines.
If you are not using AC power, you must
install 4 AA alkaline batteries before
using the device.
Replace the batteries when any of the
following occurs:
•The battery charge indicator
indicates a low charge
•The display dims
•The display does not light up
1.Slide off the battery cover.
Page 27
Directions for use 23
2.Install the batteries by matching the
polarity as shown in the diagram.
3.Replace the cover.
Pair a smartphone, a tablet, or a
communication hub with the monitor
Advanced Bluetooth 4.0 technology
enables your smartphone, your tablet, or
a communication hub to receive your
personal health information from your
blood pressure monitor. You have two
alternatives to pair the monitor with
these devices:
•If you are using the Welch Allyn
Home application, follow the pairing
Page 28
24
Welch Allyn Home Blood Pressure Monitor
instructions presented in the
application.
•Contact your Remote Monitoring
Solution provider or use the
instructions that office provided to
set up these communications
devices.
With either alternative, you complete the
same steps on the monitor:
1.Confirm that smartphone, tablet, or
communication hub is powered on
and in discoverable mode (ready for
pairing). Bluetooth wireless
connectivity is only available for
devices which support Bluetooth 4.0
technology.
2.Press and hold the Power button on
the blood pressure monitor for 2
seconds to power on the device and
activate Bluetooth pairing. The Bluetooth symbol flashes while
discovering the monitor.
If successful, a steady symbol
surrounded by a dotted line appears on
the LCD screen. If unsuccessful, the
appears.
Page 29
Directions for use 25
Note
Position the blood pressure cuff
To achieve an accurate blood pressure
reading, follow these steps to position
the blood pressure cuff properly.
1.Remove any sleeves covering your
upper arm.
The preferred site for taking a
blood pressure measurement is
your bare upper arm.
2.Place the cuff loosely over your arm
so that the blood pressure tube
extends down over your hand. It
should appear on your arm as shown.
3.
Open the cuff to resize it as needed
and slide it over your bare upper arm.
The bottom edge of the cuff should
be 2 to 3 centimeters (0.8 to 1.2
inches) above the bend in your arm,
Page 30
26
Welch Allyn Home Blood Pressure Monitor
and the tube and artery marker
should appear on your arm as shown
4. Keeping the artery marker and the
tube in place, wrap the cuff snugly
around your arm, but not too tight. If
you can insert just two fingers
between the cuff and your arm, the
tightness is acceptable.
.
Start the measurement
1.Prepare yourself to take your blood
pressure.
Page 31
Directions for use 27
a.Sit comfortably in a chair that
supports your back.
b. Keep legs uncrossed.
c.Place feet flat on the floor.
d. Support your arm on the flat
surface.
e.Ensure the cuff is level with your
heart (right atrium).
f.Rest for 5 minutes after sitting
down and positioning the blood
pressure cuff.
2.Verify that the cuff is positioned
correctly with the tube running
down the middle of your forearm.
3.Press the Power button to power up
the monitor and begin the blood
pressure measurement.
The cuff inflates and tightens around
your arm until it completes the
measurement. A vertical stack of
segments appears onscreen to show
increasing pressure in the cuff. The
heart icon in the upper-left corner
blinks to indicate your pulse rate
until the measurement is complete.
Page 32
28
Note
Welch Allyn Home Blood Pressure Monitor
If you press the Power button to
stop the measurement, dotted
lines replace the SYS and DIA
numbers onscreen.
When the inflation cycle is complete,
systolic and diastolic measurements
as well as your pulse rate appear
onscreen.
4. Record your measurement results.
Either record these measurements
manually along with the time of
measurement, or transfer them
electronically to your connected
device.
Caution The device that receives
transferred measurements
displays the most recent record
first as record 1. Each new
measurement becomes record 1,
and all other records move down
the list (for example, 2 becomes 3,
and so on). You can store a
maximum of 99 records. When
you pass that limit, the oldest
record drops from the list.
Page 33
Directions for use 29
Note
Note
For a meaningful comparison,
try to take measurements under
consistent conditions. For
example, take daily
measurements at approximately
the same time, on the same
upper arm, or as directed by
your physician.
Some points to remember about your
blood pressure device:
•After 10 seconds of inactivity, the
device powers down.
•If you press and release the Power
button, the device powers down.
•After successful transfer, the device
powers off the Bluetooth radio, and
the Bluetooth icon (and rectangle)
disappear.
Remove and attach a blood pressure
cuff
The air tube remains connected to
the blood pressure monitor at all
times, but you can remove and
attach blood pressure cuffs as
follows.
Page 34
30
Welch Allyn Home Blood Pressure Monitor
1.To remove a connected cuff, gently
pull the air tube connector away
from the cuff connector.
2.To attach a cuff, gently press the
cuff connector into the air tube.
Blood pressure facts, pointers, and
best practices
Helpful definitions
Systolic pressure When ventricles
contract and pump blood out of your
heart, the point at which blood pressure
reaches its maximum value in the cycle.
Diastolic pressure When the ventricles
relax, the point at which blood pressure
reaches its minimum value in the cycle.
Pulse rate A measurement of the heart
rate, or the number of times the heart
beats per minute. As the heart pushes
blood through the arteries, the arteries
Page 35
Directions for use 31
expand and contract with the flow of the
blood.
Irregular heartbeat detector
An irregular heartbeat (IHB) is detected
when a heartbeat rhythm varies while
the unit is measuring the systolic and
diastolic blood pressure. During each
measurement, the monitor records the
heartbeat intervals and calculates the
average. If any average is larger than or
equal to 25%, the irregular heartbeat
symbol appears on the display when the
measurement results appear.
Caution The appearance of the
IHB icon indicates that a pulse
irregularity consistent with an
irregular heartbeat was detected
during measurement. Usually this
is NOT a cause for concern.
However, if the symbol appears
often, we recommend you seek
medical advice. The device does
not replace a cardiac examination
but detects pulse irregularities at
an early stage.
Page 36
32
Welch Allyn Home Blood Pressure Monitor
Why does my blood pressure fluctuate
throughout the day?
•Individual blood pressure varies
multiple times every day because of
weather, emotion, and activity.
•The way you position your cuff and
your measurement posture can
affect blood pressure. Try to take
blood pressure measurements under
similar conditions.
•Sometimes blood pressure on one
arm is different from blood pressure
on your other arm. Use the same
arm for blood pressure
measurements.
•Medication might affect your blood
pressure.
•The "white coat effect" might cause
your blood pressure to be higher in a
hospital or doctor's office.
Page 37
Directions for use 33
Factors that could cause inaccurate
measurements
•Cuff position is poor
•Cuff is too loose or too tight
•Cuff is applied over clothing rather
than on a bare upper arm
•Taking a measurement within 1 hour
of eating or drinking
•Taking a measurement within 20
minutes of taking a bath
•Being in a cold environment
•Taking a measurement immediately
after smoking or drinking
caffeinated beverages
•Taking a measurement while talking
or moving your fingers
•Taking a measurement when you
need to urinate
•Taking measurements in rapid
succession
Factors that promote accurate blood
pressure measurements
•Proper cuff position on the upper
arm
•Proper cuff tightness
•No clothing under the cuff or
restricting circulation above the cuff
•Proper posture: feet flat on the floor,
back straight, legs uncrossed, arm
resting on a flat surface, upper arm
at heart level
•Relaxing or sitting still for a few
minutes after positioning the cuff
Page 38
34
Welch Allyn Home Blood Pressure Monitor
and before taking a blood pressure
measurement
•Sitting still during measurement
•Not talking during measurement
•Waiting at least 3 minutes between
measurements
Page 39
Directions for use 35
Maintenance
Maintain the monitor and cuff
The monitor does not require calibration.
To get the best performance from your
monitor, follow the maintenance steps
below.
•Store the unit in a dry place away
from direct sunlight.
•Avoid shaking and dropping the
monitor.
•Avoid operating the monitor in dusty
and unstable temperature
environments.
Cleaning
Caution Use a soft cloth to clean
the entire unit. Do not use any
abrasive or volatile cleaners.
1.Clean the monitor only when
necessary with one of the following
compatible cleaning agents:
•70% isopropyl alcohol
•10% chlorine bleach/90% water
solution (standard bleach wipe)
2.If necessary, you may wipe the cuff
with a soft, damp cloth.
Page 40
36
Welch Allyn Home Blood Pressure Monitor
Troubleshooting
This section includes a list of error
messages and frequently asked
questions for problems you may
encounter with your blood pressure
monitor. If the device is not operating as
you think it should, check here before
arranging for service.
Problems and error messages
The device presents technical alarms
and low-priority physiological alarms.
Technical alarms occur when there is a
device-related error. Physiological alarms
occur when blood pressure
measurements fall outside of set alarm
limits.
Technical alarms
Problem Symptom Root
No powerDisplay will
not light up
cause
Batteries are
drained.
Batteries are
inserted
incorrectly.
AC ad apter
is inserted
incorrectly
Solution
Replace
with new
batteries
Insert the
batteries
correctly
Insert the
AC ad apter
tightly
Page 41
Directions for use 37
Problem Symptom Root
cause
Low
batteries
Error
messages
The display
indicates
the “BATLO”
message,
pauses for
3 seconds.
The battery
icon shows
empty
(does not
flash).
E 01 shows The cuff is
E 03 shows There is too
E 10 or E 11
shows
Batteries are
low.
not secure,
the cuff is
too tight, or
a leak is
detected.
much
pressure in
the cuff
The monitor
detected
motion while
measuring
Solution
Replace
with new
batteries.
Readjust
the cuff,
have the
patient
relax for a
moment,
and then
measure
again.
Inspect the
connection
tube for
leaks or
kinks.
Refasten
the cuff and
then
measure
again
Readjust
the cuff,
relax for a
moment,
and then
measure
again
Page 42
38
Welch Allyn Home Blood Pressure Monitor
Problem Symptom Root
cause
Visit
hillrom.com
EE or E 19
shows
Hardware
error
for further assistance.
Solution
Measure
again. If
the
problem
persists,
contact
the
retailer
or
Wel ch
Allyn
customer
service
at
hillrom.com
for
further
assistance.
Page 43
Physiological alarms
Symptom Root causeSolution
Out of range.
Either SYS
>260mmHg or
DIA
>220mmHg.
The symbol
may appear in
either the SYS
or DIA area of
the screen.
Out of range.
Either SYS
<50mmHg or
DIA <25mmHg.
The symbol
may appear in
either the SYS
or DIA area of
the screen.
Directions for use 39
Press and
hold the
Power
button to
reestablish
Bluetooth
connectivity.
Measure
again. If the
problem
persists,
contact your
physician.
Alarm
priority =
Low
Press and
hold the
Power
button to
reestablish
Bluetooth
connectivity.
Measure
again. If the
problem
persists,
contact your
physician.
Alarm
priority =
Low
Page 44
40
Welch Allyn Home Blood Pressure Monitor
Specifications
ItemSpecification
Power supply:
Battery powered
mode:
Power supply: AC
adapter powered
mode:
Power supply model
number
Display modeDigital LCD
Measurement model Oscillometric testing
Measurement range Rated cuff pressure:
AccuracyPressure:
6VDC 4*AA batteries
Input: 100-240V, 5060Hz, 400mA;
Output: 6V, 1A
UE08WCP060100SPA
V. A. 6 8 mm * 90 m m
mode
0mmHg to
300mmHg (0kPa40kPa)
Measurement
pressure:
SYS: 50 to 260mmHg
DIA: 25 to 220mmHg
Pulse value: (40 to
199)beat/minute
±0.4kPa (3mmHg)
Pulse value:±4%
Page 45
Directions for use 41
ItemSpecification
Normal working
conditions
Storage and
transportation
condition
Tem pe ra tu re : 5 °C to
40°C
Relative humidity
85% RH
Atmospheric
pressure: 86kPa to
106kPa
Tem pe ra tu re :- 20 °C to
60°C
Relative Humidity:
10% RH–93% RH
Atmospheric
Pressure: 50kPa to
106 kPa
Circumferen ce of the
upper arm
Part
Number
D-Ring
Extra
Small Cuff
BPACC-01
Cuff size
15–24cm
D-Ring
Standard
Wide Cuff
BPACC02
D-Ring
Extra
Large
Cuff
BPACC03
22–42cm
40–54cm
Page 46
42
Welch Allyn Home Blood Pressure Monitor
ItemSpecification
Net WeightApprox. 283g
(Excluding the dry
cells)
External dimensions Approx. 94mm
*142mm*66mm
Degree of protection Type BF applied part
Protection against
ingress of water
IP22
Software VersionVersion A01
General radio compliance
ItemSpecification
Bluetooth
No.
Radio Frequency
(RF) Range
Output Power4.0 dBm
Supply Voltage1.8V to 3.6V
Antenna Gain0.0 dBi
Transmitting
Distance
Module
AW5182 2
2402 MHz to 2480 MHz
10 meters (30 feet)
Page 47
Directions for use 43
Federal Communication Commission
(FCC) Interference Statement
This device complies with part 15 of the
FCC Rules. Operation is subject to the
following two conditions.
•This device may not cause harmful
interference.
•This device must accept any
interference received, including
interference that may cause
undesired operation.
This equipment (FCC ID: OU9TMB1591A) has been tested and found to comply
with the limits for a Class B digital
device, pursuant to Part 15 of the FCC
Rules. These limits are designed to
provide reasonable protection against
harmful interference in a residential
installation. This equipment generates,
uses, and can radiate radio frequency
energy and, if not installed and used in
accordance with the instructions, may
cause harmful interference to radio
communications. However, there is no
guarantee that interference will not
occur in a particular installation. If this
equipment does cause harmful
interference to radio or television
reception, which can be determined by
Page 48
44
Welch Allyn Home Blood Pressure Monitor
turning the equipment off and on, the
user is encouraged to try to correct the
interference by one of the following
measures.
1.Reorient or relocate the receiving
antenna.
2.Increase the separation between the
equipment and receiver.
3.Consult the dealer or an experienced
radio/TV technician for help.
Any changes or modifications not
expressly approved by the party
responsible for compliance could
void the user's authority to operate
this equipment.
FCC Radiation Exposure Statement
This equipment complies with FCC
radiation exposure limits set forth for an
uncontrolled environment. This
transmitter must not be co-located or
operating in conjunction with any other
antenna or transmitter.
Industry Canada (IC) compliance
To ensure compliance with FCC and
Industry Canada RF exposure
requirements, this device must be
Page 49
Directions for use 45
installed in a location where the
antennas of the device will have a
minimum distance of at least 20 cm from
all persons. Using higher gain antennas
and types of antennas not certified for
use with this product is not allowed. The
device shall not be co-located with
another transmitter.
Pour s'assurer de la conformité des
exigences aux expositions RF de la FCC
et de l'Industry Canada, cet appareil doit
être installé dans un emplacement où les
antennes de l'appareil sont au moins à
20 cm de distance de toute personne.
L'utilisation d'antennes et de types
d'antennes à gain supérieur non garantis
avec ce produit est interdite. L'appareil
ne doit pas être installé à proximité d'un
autre émetteur.
This device complies with Industry
Canada licence-exempt RSS standard(s).
Operation is subject to the following two
conditions: (1) this device may not cause
interference, and (2) this device must
accept any interference, including
interference that may cause undesired
operation of the device.
Cet appareil est conforme aux normes
RSS exemptes de licence de l'Industry
Page 50
46
Welch Allyn Home Blood Pressure Monitor
Canada. Son fonctionnement est soumis
aux deux conditions suivantes: (1)
l'appareil ne doit pas produire de
brouillage, et (2) l'appareil doit accepter
tout brouillage radioélectrique subi,
même si le brouillage est susceptible
d'en compromettre le fonctionnement.
This radio transmitter (IC: 12725ATMB1591A) has been approved by
Industry Canada to operate with the
antenna listed in the specification table.
Ce transmetteur de radio (IC: 12725ATMB1591A) a été approuvé par l'Industry
Canada pour fonctionner avec l'antenne
répertoriée dans la table des
spécifications.
This Class B digital apparatus complies
with Canadian ICES-003.
Cet appareil numérique de la classe B est
conforme à la norme NMB-003 du
Canada.
Page 51
Directions for use 47
Complied standards list
ItemStandard
Risk
management
LabelingISO/EN 15223-1 Medical
User manualEN 1041 Medical
General
Requirements
for Safety
ISO/EN 14971 Medical
devices — Application of
risk management to
medical devices
devices. Symbols to be
used with medical device
labels, labeling and
information to be
supplied. General
requirements
equipment manufacturers
to provide information
IEC 60601-1+A1 Medical
electrical equipment Part 1: General
requirements for basic
safety and essential
performance
IEC 60601-1-11 Medical
electrical equipment Part 1-11: General
requirements for basic
safety and essential
performance - Collateral
standard: Requirements
for medical electrical
equipment and medical
electrical systems used in
the home healthcare
environment
Page 52
48
Welch Allyn Home Blood Pressure Monitor
ItemStandard
Electromagnetic
compatibility
IEC/EN 60601-1-2 Medical
electrical equipment Part 1-2: General
requirements for basic
safety and essential
performance - Collateral
standard: Electromagnetic
compatibility Requirements and tests
Performance
requirements
and clinical
investigation
IEC 80601-2-30 Medical
electrical equipment- Part
2-30: Particular
requirements for the basic
safety and essential
performance of
automated non-invasive
sphygmomanometers
ISO81060-2:2013 Noninvasive
sphygmomanometers —
Part 2: Clinical validation
of automated
measurement type
Software lifecycle processes
IEC/EN 62304+AC:
Medical device software Software life cycle
processes
Page 53
Directions for use 49
ItemStandard
UsabilityIEC 62366 Medical
Small-bore
connectors
devices - Application of
usability engineering to
medical devices (IEC
62366)
IEC 60601-1-6 Medical
electrical equipment Part 1 -6: General
requirements for basic
safety and essential
performance - collateral
standard: Usability
IEC 80369-5 Small-Bore
Connectors for Liquids
and Gases in Healthcare
Applications - Part 5:
Connectors for Limb Cuff
Inflation Applications
Page 54
50
Welch Allyn Home Blood Pressure Monitor
Warranty
Welch Allyn will warranty the blood
pressure monitor to be free of defects in
material and workmanship and to
perform in accordance with
manufacturer specifications for the
period of two years from the date of
purchase from Welch Allyn or its
authorized distributors or agents.
The warranty period shall start on the
date of purchase. The date of purchase
is: 1) the invoiced ship date if the device
was purchased directly from Welch
Allyn, 2) the date specified during
product registration, 3) the date of
purchase of the product from a Welch
Allyn authorized distributor as
documented from a receipt from said
distributor.
This warranty does not cover damage
caused by: 1) handling during shipping,
2) use or maintenance contrary to
labeled instructions, 3) alteration or
repair by anyone not authorized by
Welch Allyn, and 4) accidents.
Page 55
Directions for use 51
The product warranty is also subject to
the following terms and limitations.
•Accessories are not covered by the
warranty.
•Shipping cost to return a device to a
Welch Allyn service center is not
included.
•A service notification number must
be obtained from Welch Allyn prior
to returning any products or
accessories to Welch Allyn's
designated service centers for repair.
To obtain a service notification
number, contact Welch Allyn
Technical Support at
technicalsupport@hillrom.com.
Page 56
52
Welch Allyn Home Blood Pressure Monitor
Approved accessories
ItemDescription
BPACC-01D-Ring Extra Small Cuff
BPACC-02D-Ring Standard Wide
BPACC-03D-Ring Extra Large Cuff
107041ProBP 2000 adapter
(15–24cm)
Cuff (22-42cm)
(40–54cm)
set. This adapter is an
alternate power source
for the blood pressure
monitoring device.
Page 57
Electromagnetic emissions
Note
Caution Wireless communications equipment, such as wireless home network
devices, mobile phones, cordless telephones and their base stations, walkie-talkies
can affect this equipment and should be kept at least a distance d=3,3m away
from the equipment.
Directions for use 53
energy only for its internal function. Therefore, its RF
emissions are very low and are not likely to cause any
interference in nearby electronic equipment.
ComplianceElectromagnetic environment - guidance
As indicated in Table 6 of IEC 60601-1-2:2007 for ME Equipment, a typical cell
phone with a maximum output power of 2 W yields d=3,3m at an IMMUNITY LEVEL
of 3V/m).
Group 1The Welch Allyn Home Blood Pressure Monitor uses RF
The Welch Allyn Home Blood Pressure Monitor H-BP100SBP is intended for use in the
electromagnetic environment specified below. The customer or user of the Welch Allyn
Home Blood Pressure Monitor should assure that it is used in such an environment.
Emissions
test
RF emissions
CISPR 11
Emissions and immunity information
Page 58
54
Welch Allyn Home Blood Pressure Monitor
suitable for use in all establishments, including domestic
establishments and those directly connected to the
public low-voltage power supply network that supplies
buildings used for domestic purposes.
Electromagnetic emissions
Class BThe Welch Allyn Home Blood Pressure Monitor is
Class A
Complies
RF emissions
CISPR 11
Harmonic
emissions
IEC 61000-3-2
Voltage
fluctuations/
flicker
emissions
IEC 61000-3-3
Page 59
Electromagnetic immunity
Electromagnetic environment -
Compliance
IEC 60601
guidance
level
test level
Floors should be wood, concrete or
ceramic tile. If floors are covered with
synthetic material, the relative
humidity should be at least 30%.
±8 kV contact
± 15kV air
±8 kV contact
± 15kV air
Directions for use 55
Mains power quality should be that of
a typical commercial or hospital
±2 kV for
±2 kV for
environment.
power supply
lines
power supply
lines
Mains power quality should be that of
± 1 kV
±1 kV for input/
output lines
± 1 kV line(s) to
a typical commercial or hospital
differential
line(s)
environment.
mode
± 2kV line(s) to
earth
The Welch Allyn Home Blood Pressure Monitor is intended for use in the electromagnetic
environment specified below. The customer or the user of the Welch Allyn Home Blood
Pressure Monitor should assure that it is used in such an environment.
Immunity
test
Electrostatic
discharge
(ESD)
IEC 61000-4-2
Electrical fast
transient/burst
IEC 61000-4-4
Surge
IEC 61000-4-5
Page 60
56
Welch Allyn Home Blood Pressure Monitor
Mains power quality should be that of
a typical commercial or hospital
environment. If the user of the Welch
Allyn Home Blood Pressure Monitor
requires continued operation during
power mains interruptions, it is
recommended that the Welch Allyn
Home Blood Pressure Monitor be
powered from an uninterruptible
power supply or a battery.
)
)
T
T
Power frequency magnetic fields
should be at levels characteristic of a
typical location in a typical
commercial or hospital environment.
T
< 5% U
Electromagnetic immunity
T
< 5% U
Voltage dips,
short
(>95% dip in
(>95% dip in
interruptions
T
T
) for 0.5
T
U
cycle
) for 0.5
T
U
cycle
and voltage
variations on
40% U
40% U
power supply
input lines
for 5 cycles
70% U
(60% dip in U
)
T
T
T
70% U
for 5 cycles
(60% dip in U
IEC 61000-4-11
T
) for 5
for 25 cycles
(30% dip in U
)
T
for 25 cycles
(30% dip in U
T
(>95% dip in
U
seconds
3A/m and
< 5% U
T
< 5% U
(>95% dip in
30 A/m
) for 5
T
3A/m and
30 A/m
U
seconds
Power
frequency
(50Hz)
magnetic field
IEC 61000-4-8
Note: UT is the AC mains voltage prior to application of the test level.
Page 61
Directions for use 57
P
Electromagnetic environment -
guidance
Portable and mobile RF
communications equipment should
be used no closer to any part of the
Welch Allyn Home Blood Pressure
Monitor, including cables, than the
recommended separation distance
calculated from the equation
applicable to the frequency of the
transmitter.
Compliance
level
= 1,2
Recommended separation
distance
d
3 Vrms and
6Vrms
Electromagnetic immunity
IEC 60601
The Welch Allyn Home Blood Pressure Monitor is intended for use in the electromagnetic
test level
environment specified below. The customer or the user of the Welch Allyn Home Blood
Pressure Monitor should assure that it is used in such an environment.
Immunity
test
3 Vrms and
6Vrms
150 kHz to
Conducted RF
IEC 61000-4-6
80 MHz
Page 62
58
PPP
Welch Allyn Home Blood Pressure Monitor
800 MHz to 2,5 GHz
80 to 800 MHz
= 1,2
= 2,3
d
d
800 MHz to 2,6 GHz
= 2,3
d
where P is the maximum output power
should be less than the
a
the transmitter in watts (W)
is the recommended separation
d
and
rating of
. Interference may occur in the
b
compliance level in each frequency
range
distance in meters (m). Field
strengths from fixed RF transmitters,
as determined by an electromagnetic
site survey
vicinity of equipment marked with the
following symbol:
3 V/m
10 V/m
Electromagnetic immunity
10 V/m
80 MHz to
3 V/m
80 MHz to
2,5 GHz
Radiated RF
IEC 61000-4-3
2,6 GHz
Page 63
Electromagnetic immunity
Directions for use 59
Note 1: At 80 MHz and 800 MHz, the higher frequency range applies.
Note 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones
absorption and reflection from structures, objects, and people.
a
and land mobile radios, amateur radio, AM and FM radio broadcast, and TV broadcast cannot be
predicted theoretically with accuracy. To assess the electromagnetic environment due to fixed RF
transmitters, an electromagnetic site survey should be considered. If the measured field strength in the
location in which the Welch Allyn Home Blood Pressure Monitor is used exceeds the applicable RF
compliance level above, the Welch Allyn Home Blood Pressure Monitor should be observed to verify
Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.
normal operation. If abnormal performance is observed,additional measures may be necessary, such as
reorienting or relocating the Welch Allyn Home Blood Pressure Monitor.
b
Page 64
60
PPP
Welch Allyn Home Blood Pressure Monitor
= 2,3
800 MHz to
2,6 GHz
d
= 1,2
80 MHz to
800 MHz
d
= 1,2
Separation distance according to frequency of transmitter (m)
150 kHz to
80 MHz
d
Recommended separation distances between portable and mobile RF
communications equipment and the Welch Allyn Home Blood Pressure
Monitor
The Welch Allyn Home Blood Pressure Monitor is intended for use in an electromagnetic
environment in which radiated RF disturbances are controlled. The customer or user of the
Welch Allyn Home Blood Pressure Monitor can help prevent electromagnetic interference
by maintaining a minimum distance between portable and mobile RF communications
equipment (transmitters) and the Welch Allyn Home Blood Pressure Monitor as
recommended below, according to the maximum output power of the communications
equipment.
Rated maximum
output power of
transmitter
(W)
0.010.120.120.23
0.10.380.380.73
11.21.22.3
103.83.87.3
100121223
Page 65
is the maximum output power rating of the
P
in meters (m) can be estimated using the equation applicable to the
d
Directions for use 61
Recommended separation distances between portable and mobile RF
communications equipment and the Welch Allyn Home Blood Pressure
Monitor
transmitter in watts (W) according to the transmitter manufacturer.
frequency of the transmitter, where
For transmitters rated at a maximum output power not listed above, the recommended
separation distance
Note 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency range
applies.
Note 2: These guidelines may not apply in all situations. Electromagnetic propagation is
affected by absorption and reflection from structures, objects, and people.