Welch Allyn H-BP100SBP User Manual

Page 1
Welch Allyn Home

Blood Pressure Monitor
Directions for use
901042 Automated Blood Pressure System
Software version A01
Page 2
ii Welch Allyn Home Blood Pressure Monitor
© 2019 Welch Allyn. All rights are reserved. To support the intended use of the product described in this publication. The purchaser of the product is permitted to copy this publication, for internal distribution only, from the media provided by Welch Allyn. No other use, reproduction, or distribution of this publication, or any part of it, is permitted without written permission from Welch Allyn. Welch Allyn assumes no responsibility for any injury to anyone, or for any illegal or improper use of the product, that may result from failure to use this product in accordance with the instructions, cautions, warnings, or statement of intended use published in this manual.
SureBP® Technology is a registered trademark of Welch Allyn.
The Bluetooth® word mark and logos are registered trademarks owned by Bluetooth SIG, Inc. and any use of such marks by Welch Allyn is under license.
Software in this product is © 2019 Welch Allyn or its vendors. All rights are reserved. The software is protected by United States of America copyright laws and international treaty provisions applicable worldwide. Under such laws, the licensee is entitled to use the copy of the software incorporated with this instrument as intended in the operation of the product in which it is embedded. The software may not be copied, decompiled, reverse engineered, disassembled, or otherwise reduced to human-perceivable form. This is not a sale of the software or any copy of the software; all right, title, and ownership of the software remain with Welch Allyn or its vendors.
For information about any Welch Allyn product, contact Welch Allyn Technical Support:
:
749521 DIR 80025286 Ver. B Revised: 2019-08
Distributed by Welch Allyn, Inc. 4341 State Street Road Skaneateles Falls, NY 13153 USA
hillrom.com
This manual applies to the 901042 Automated Blood Pressure System
hillrom.com
Page 3
Directions for use iii
Contents
Introduction . . . . . . . . . . . . . . . . . . . . . . . . 1
Indications for use. . . . . . . . . . . . . . . . . . . 2
Contraindications . . . . . . . . . . . . . . . . . . . 2
Symbols. . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
Documentation symbols . . . . . . . . . . . . . . . .3
Power symbols . . . . . . . . . . . . . . . . . . . . . . . . 3
Shipping, storing, and environment symbols . .4
Blood pressure device and cuff symbols . . . . . . 5
Miscellaneous symbols. . . . . . . . . . . . . . . . . . . . . .6
About warnings and cautions. . . . . . . . . 8
Warnings and cautions . . . . . . . . . . . . . . . . . 9
Contents list. . . . . . . . . . . . . . . . . . . . . . . . . . 14
Controls and indicators . . . . . . . . . . . . . . . . . . . . 15
Power options . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
Screen elements . . . . . . . . . . . . . . . . . . . 18
Measuring blood pressure . . . . . . . . . . . 21
Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
Set up your equipment . . . . . . . . . . . . . . . . 22
Pair a smartphone, a tablet, or a communication hub with the monitor . . . 23
Position the blood pressure cuff . . . . . . . .25
Start the measurement . . . . . . . . . . . . . . . .26
Some points to remember about your
blood pressure device: . . . . . . . . . . . . . . . . 29
Blood pressure facts, pointers, and best
practices . . . . . . . . . . . . . . . . . . . . . . . . . .30
Helpful definitions . . . . . . . . . . . . . . . . . . . .30
Irregular heartbeat detector . . . . . . . . . . . 31
Maintenance . . . . . . . . . . . . . . . . . . . . . . . 35
Page 4
iv Welch Allyn Home Blood Pressure Monitor
Maintain the monitor and cuff . . . . . . . . . . 35
Troubleshooting . . . . . . . . . . . . . . . . . . . 36
Problems and error messages . . . . . . . . . .36
Technical alarms . . . . . . . . . . . . . . . . . . . . . .36
Physiological alarms . . . . . . . . . . . . . . . . . . . . . . .39
Specifications. . . . . . . . . . . . . . . . . . . . . 40
General radio compliance . . . . . . . . . . . . . .42
Federal Communication Commission (FCC)
Interference Statement . . . . . . . . . . . . . . . .43
FCC Radiation Exposure Statement . . . . .44
Industry Canada (IC) compliance . . . . . . .44
Complied standards list . . . . . . . . . . . . . . . . . . . .47
Warranty . . . . . . . . . . . . . . . . . . . . . . . . . .50
Approved accessories . . . . . . . . . . . . . . 52
Emissions and immunity information . 53
Page 5
Directions for use 1
Introduction
The Welch Allyn Home™ blood pressure monitor is indicated for adult use in the home or domestic setting only.
The Welch Allyn Home blood pressure monitor features blood pressure measurement via SureBP pulse rate measurement, and data management using Bluetooth technology.
SureBP Technology is an advanced blood pressure measurement technology
that obtains a blood pressure reading during the inflation cycle (as the cuff pumps up). This technology allows for a faster, more comfortable reading. If on inflation, SureBP Technology is unable to determine a blood pressure due to excessive patient movement or any arrhythmia, it will inflate the cuff to a slightly higher pressure, then measure the blood pressure while deflating the cuff, providing an accurate reading with less chance of a reading failure.
The design provides you with two years of reliable service. Readings taken by the monitor are equivalent to those obtained
®
Technology,
®
Smart
Page 6
2
Welch Allyn Home Blood Pressure Monitor
by a trained observer using the cuff and stethoscope auscultation method.
This Directions for use contains important safety and care information and provides step by step instructions for using the monitor. Read the manual thoroughly before using the monitor.
Indications for use
The Welch Allyn Home blood pressure monitor is a digital monitor intended for use in measuring blood pressure and heart rate in adult patient populations with arm circumferences between 15 cm to 54 cm (approximately 6 to 21 inches).
Contraindications
This device is contraindicated for any person who is connected to a wearable or implantable electronic device or instrument, such as a pacemaker or defibrillator.
This blood pressure monitor is not intended to be a diagnostic device. Contact your physician if hypertensive values are indicated.
Page 7
Symbols
Documentation symbols
Directions for use 3
War ning
manual identify conditions or practices that could lead to illness, injury, or death.
Caution
manual identify conditions or practices that could result in damage to the equipment or other property, or loss of data.
Mandatory - Consult Directions for Use
: The warning statements in this
: The caution statements in this
Power symbols
Direct current (DC)
Stand-By
Page 8
4
Welch Allyn Home Blood Pressure Monitor
Shipping, storing, and environment symbols
Humidity limitation
Separate collection of Electrical and Electronic Equipment. Do not dispose as unsorted municipal waste.
Temperature limit
Atmospheric pressure limitation
Recyclable
Stacking limit by number
Page 9
Directions for use 5
Blood pressure device and cuff symbols
SureBP technology logo
Artery marker
Range
Index edge
Limb circumference (Minimum/Maximum)
Lot code
Not made with natural rubber latex
Page 10
6
Welch Allyn Home Blood Pressure Monitor
Miscellaneous symbols
Type BF applied part
Serial Number
Product Identifier
Manufacturer
Reorder Number
Non-ionizing electromagnetic radiation
Global Trade Item Number
Class II equipment
Page 11
Directions for use 7
IP22 Ingress protection: the device is
protected against solid foreign objects of 12.5mm and greater and against vertically falling water drops when ENCLOSURE is tilted up to 15°
Bluetooth
Page 12
8
Welch Allyn Home Blood Pressure Monitor
About warnings and cautions
Caution statements can appear on the Welch Allyn Home blood pressure device, the packaging, the shipping container, or in this Directions for use.
The Welch Allyn Home blood pressure device is safe for adults when used in accordance with the instructions and caution statements presented in this Directions for use.
Before using the device, you must familiarize yourself with all cautions, with the steps to power up the device, and with the sections of this Directions for use that pertain to your use of the device. In addition to reviewing the general cautions presented in the next section, you must also review the more specific cautions that appear throughout the manual in conjunction with setup/ startup, operation, and maintenance tasks.
Failure to understand and observe
any warning statement in this manual could lead to injury or illness.
Failure to understand and observe
any caution statement in this manual could lead to damage to the
Page 13
Directions for use 9
equipment or other property, or loss of measurement data.
Warnings and cautions
WARNING Injury risk. The device is not suitable for measuring the blood pressure of neonatal infants or children.
WARNING Injury risk. The device is not suitable for use on pregnant women, users with implanted electrical devices, users with preeclampsia, premature ventricular beats, atrial fibrillation, peripheral arterial disease, and users undergoing intravascular therapy or arterio­venous shunt or people who received a mastectomy. Consult your doctor prior to using the unit if you suffer from any of these illnesses or conditions.
WARNING Injury risk. Do not take any therapeutic measures on the basis of a self measurement. Never alter the dose of a medicine prescribed by a doctor. Consult your doctor if you have any question about your blood pressure.
WARNING Injury risk. Keep the unit out of reach of infants, children, or pets, since inhalation or swallowing of small parts is dangerous or even fatal.
Page 14
10
Welch Allyn Home Blood Pressure Monitor
WARNING Injury risk. The device is not intended to be a self­measurement station used in a public place or center. Use the device only in your home or temporary residence.
WARNING Injury risk. Do not burn batteries. Batteries may leak or explode.
WARNING Injury risk. If you experience discomfort during a measurement, such as pain in the arm or other complaints, press the power button immediately to release the air from the cuff. Loosen and remove the cuff from your arm.
WARNING Injury risk. On the rare occasion of a fault causing the cuff to remain fully inflated during measurement, open the cuff immediately. Prolonged high pressure applied to the arm (cuff pressure >300mmHg or constant pressure >15mmHg for more than 3 minutes) might lead to bruising and discolored skin.
WARNING Injury risk. This unit is not suitable for continuous monitoring during medical emergencies or operations.
WARNING Injury risk. Taking blood pressure measurements too frequently could disrupt blood circulation and cause injuries.
Page 15
Directions for use 11
WARNING Injury risk. Do not kink the connection tube during use. The cuff pressure might continuously increase, which could prevent blood flow and result in injury.
WARNING Injury risk. Never apply the cuff over damaged skin. This can cause further injury.
WARNING Injury risk. If applying to someone else, do not touch output of the batteries/adapter and the user simultaneously.
WARNING Injury risk. Do not use the unit if you are allergic to polyester or synthetic materials.
WARNING Injury risk. Excessive tube lengths could cause strangulation if you don't manage them properly.
WARNING Injury risk. Do not connect the air tube to other medical equipment. This could cause air to be pumped into intravascular systems or high pressure, which could lead to serious injuries.
WARNING Injury risk. Dispose of accessories, detachable parts, and the device according to the local guidelines.
Page 16
12
Welch Allyn Home Blood Pressure Monitor
WARNING Injury risk. Do not service or perform any maintenance while using the device.
WARNING Injury risk. Use only accessories approved by the manufacturer. Using unapproved accessories might cause damage to the unit and injure users.
WARNING Injury risk. No modification to this equipment is allowed. Modifying the equipment could damage the unit or endanger the user.
Page 17
Directions for use 13
Caution This device is intended for non-invasive measuring and monitoring of arterial blood pressure. It is not intended for use on extremities other than arm or for functions other than obtaining a blood pressure measurement.
Caution Do not wrap the cuff on the same arm to which another monitoring device is applied. One or both devices could temporarily stop functioning if you try to use them on the same arm at the same time.
Caution To avoid measurement errors, avoid taking blood pressure measurements near a strong electromagnetic field radiated interference signal or electrical fast transient/burst signal.
Caution The materials of the cuff have been tested and found to comply with requirements of ISO 10993-5:2009 and ISO 10993­10:2010. It will not cause any potential sensitization or irritation reaction.
Caution Before use, make sure the device functions safely and is in proper working condition.
Caution Use the device in the environment described in this directions for use. Otherwise, you will compromise the device’s performance and reduce its lifetime.
Page 18
14
Welch Allyn Home Blood Pressure Monitor
Caution Do not attempt to repair the unit yourself if it malfunctions. Only have repairs carried out by authorized service centers.
Caution Report any unexpected operation or events to the manufacturer.
Caution Use a soft cloth to clean the entire unit. Do not use any abrasive or volatile cleaners. See the cleaning instructions presented later in this Directions for use.
Contents list
The following items are in the device box:
Blood Pressure Monitor
D-Ring Standard Wide Cuff (22cm
to 42cm) with air tube attached
(4) AA alkaline batteries
Directions for use
Quick reference guide
Page 19
Controls and indicators
Monitor front
No. Feature Description
1D-Ring
blood pressure cuff
2Power
button
3LCD
Display
Apply to upper arm to take a blood pressure measurement
Powers on the monitor, starts and stops a blood pressure measurement, and — when pressed for 2 seconds — initiates a
Bluetooth
Displays blood pressure reading and other pertinent information regarding the reading
Directions for use 15
pairing
Page 20
16
Welch Allyn Home Blood Pressure Monitor
Monitor back
No. Feature Description
1 Direct current
power connection
2 Battery
compartment (behind cover)
When used with an accessory power cord (not included), connects the monitor to a power outlet
Houses 4 AA alkaline batteries
Page 21
Directions for use 17
Power options
Caution To get optimal performance and protect your monitor, use only the correct batteries or the Welch Allyn­approved power adapter.
The monitor is powered by one of two sources:
4 AA alkaline batteries
AC adapter (6v 1A) (not
included)
Page 22
18
Welch Allyn Home Blood Pressure Monitor
Screen elements
The liquid crystal display (LCD) displays the following: systolic blood pressure (mmHg), diastolic blood pressure (mmHg), pulse rate (bpm), heart beat (regular or irregular) while acquiring blood pressure measurements, excessive motion alert, Bluetooth connectivity status or data transmission error, alarm priority, and battery charge level.
Page 23
Directions for use 19
Symbol Description
Systolic blood pressure
mmHg = measurement unit
of the blood pressure
Diastolic blood pressure
mmHg = measurement unit
of the blood pressure
Pulse in beats per minute
Irregular heartbeat
Monitor is detecting an
irregular heartbeat during
measurement
Heart beat
Monitor is detecting a
heartbeat during
measurement
Battery indicator
Indicates the current battery
charge
Battery indicator
Indicates the current battery
charge
Page 24
20
Welch Allyn Home Blood Pressure Monitor
Symbol Description
Bluetooth
or data transmission error
Motion indicator Motion may result in an inaccurate measurement.
Reading out of range Either SYS > 260mmHg or DIA > 220mmHg. The symbol may appear in either the SYS or DIA area of the screen. Alarm priority = Low (an! appears near the top of the screen)
Reading out of range Either SYS < 50mmHg or DIA <25mmHg. The symbol may appear in either the SYS or DIA area of the screen. Alarm priority = Low (an! appears near the top of the screen)
connectivity status
Page 25
Directions for use 21
Measuring blood pressure
Overview
You are the intended operator of the device. You can measure your blood pressure and then save and send measurement data to a smartphone or tablet. You can also maintain your device and its accessories, and you can change the batteries under normal circumstances.
To measure your blood pressure, you need the following:
Blood pressure monitor with
batteries inserted
Blood pressure cuff ("D-ring" style
for easy, one-handed adjustment)
Blood pressure air tube
To save and transfer blood pressure measurements, you also need a smartphone or tablet with Bluetooth wireless connectivity and the application or software to receive your data, such as the Welch Allyn Home application or another healthcare provider application.
Page 26
22
Welch Allyn Home Blood Pressure Monitor
Set up your equipment
Insert or replace the batteries
WARNING Injury risk. Do not burn batteries. Batteries may leak or explode.
Caution Remove the batteries if the device is not used regularly.
Caution Dispose of old batteries by following your local recycling guidelines.
If you are not using AC power, you must install 4 AA alkaline batteries before using the device.
Replace the batteries when any of the following occurs:
The battery charge indicator
indicates a low charge
The display dims
The display does not light up
1. Slide off the battery cover.
Page 27
Directions for use 23
2. Install the batteries by matching the
polarity as shown in the diagram.
3. Replace the cover.
Pair a smartphone, a tablet, or a communication hub with the monitor
Advanced Bluetooth 4.0 technology enables your smartphone, your tablet, or a communication hub to receive your personal health information from your blood pressure monitor. You have two alternatives to pair the monitor with these devices:
If you are using the Welch Allyn
Home application, follow the pairing
Page 28
24
Welch Allyn Home Blood Pressure Monitor
instructions presented in the application.
Contact your Remote Monitoring
Solution provider or use the instructions that office provided to set up these communications devices.
With either alternative, you complete the same steps on the monitor:
1. Confirm that smartphone, tablet, or
communication hub is powered on and in discoverable mode (ready for pairing). Bluetooth wireless connectivity is only available for devices which support Bluetooth 4.0 technology.
2. Press and hold the Power button on
the blood pressure monitor for 2 seconds to power on the device and activate Bluetooth pairing. The Bluetooth symbol flashes while discovering the monitor.
If successful, a steady symbol surrounded by a dotted line appears on the LCD screen. If unsuccessful, the
appears.
Page 29
Directions for use 25
Note
Position the blood pressure cuff
To achieve an accurate blood pressure reading, follow these steps to position the blood pressure cuff properly.
1. Remove any sleeves covering your
upper arm.
The preferred site for taking a blood pressure measurement is your bare upper arm.
2. Place the cuff loosely over your arm
so that the blood pressure tube extends down over your hand. It should appear on your arm as shown.
3.
Open the cuff to resize it as needed and slide it over your bare upper arm. The bottom edge of the cuff should be 2 to 3 centimeters (0.8 to 1.2 inches) above the bend in your arm,
Page 30
26
Welch Allyn Home Blood Pressure Monitor
and the tube and artery marker should appear on your arm as shown
4. Keeping the artery marker and the
tube in place, wrap the cuff snugly around your arm, but not too tight. If you can insert just two fingers between the cuff and your arm, the tightness is acceptable.
.
Start the measurement
1. Prepare yourself to take your blood
pressure.
Page 31
Directions for use 27
a. Sit comfortably in a chair that
supports your back.
b. Keep legs uncrossed.
c. Place feet flat on the floor.
d. Support your arm on the flat
surface.
e. Ensure the cuff is level with your
heart (right atrium).
f. Rest for 5 minutes after sitting
down and positioning the blood pressure cuff.
2. Verify that the cuff is positioned
correctly with the tube running down the middle of your forearm.
3. Press the Power button to power up
the monitor and begin the blood pressure measurement.
The cuff inflates and tightens around your arm until it completes the measurement. A vertical stack of segments appears onscreen to show increasing pressure in the cuff. The heart icon in the upper-left corner blinks to indicate your pulse rate until the measurement is complete.
Page 32
28
Note
Welch Allyn Home Blood Pressure Monitor
If you press the Power button to stop the measurement, dotted lines replace the SYS and DIA numbers onscreen.
When the inflation cycle is complete, systolic and diastolic measurements as well as your pulse rate appear onscreen.
4. Record your measurement results.
Either record these measurements manually along with the time of measurement, or transfer them electronically to your connected device.
Caution The device that receives transferred measurements displays the most recent record first as record 1. Each new measurement becomes record 1, and all other records move down the list (for example, 2 becomes 3, and so on). You can store a maximum of 99 records. When you pass that limit, the oldest record drops from the list.
Page 33
Directions for use 29
Note
Note
For a meaningful comparison, try to take measurements under consistent conditions. For example, take daily measurements at approximately the same time, on the same upper arm, or as directed by your physician.
Some points to remember about your blood pressure device:
After 10 seconds of inactivity, the
device powers down.
If you press and release the Power
button, the device powers down.
After successful transfer, the device
powers off the Bluetooth radio, and the Bluetooth icon (and rectangle) disappear.
Remove and attach a blood pressure cuff
The air tube remains connected to the blood pressure monitor at all times, but you can remove and attach blood pressure cuffs as follows.
Page 34
30
Welch Allyn Home Blood Pressure Monitor
1. To remove a connected cuff, gently
pull the air tube connector away from the cuff connector.
2. To attach a cuff, gently press the
cuff connector into the air tube.
Blood pressure facts, pointers, and best practices
Helpful definitions
Systolic pressure When ventricles
contract and pump blood out of your heart, the point at which blood pressure reaches its maximum value in the cycle.
Diastolic pressure When the ventricles
relax, the point at which blood pressure reaches its minimum value in the cycle.
Pulse rate A measurement of the heart
rate, or the number of times the heart beats per minute. As the heart pushes blood through the arteries, the arteries
Page 35
Directions for use 31
expand and contract with the flow of the blood.
Irregular heartbeat detector
An irregular heartbeat (IHB) is detected when a heartbeat rhythm varies while the unit is measuring the systolic and diastolic blood pressure. During each measurement, the monitor records the heartbeat intervals and calculates the average. If any average is larger than or equal to 25%, the irregular heartbeat symbol appears on the display when the measurement results appear.
Caution The appearance of the IHB icon indicates that a pulse irregularity consistent with an irregular heartbeat was detected during measurement. Usually this is NOT a cause for concern. However, if the symbol appears often, we recommend you seek medical advice. The device does not replace a cardiac examination but detects pulse irregularities at an early stage.
Page 36
32
Welch Allyn Home Blood Pressure Monitor
Why does my blood pressure fluctuate throughout the day?
Individual blood pressure varies
multiple times every day because of weather, emotion, and activity.
The way you position your cuff and
your measurement posture can affect blood pressure. Try to take blood pressure measurements under similar conditions.
Sometimes blood pressure on one
arm is different from blood pressure on your other arm. Use the same arm for blood pressure measurements.
Medication might affect your blood
pressure.
The "white coat effect" might cause
your blood pressure to be higher in a hospital or doctor's office.
Page 37
Directions for use 33
Factors that could cause inaccurate measurements
Cuff position is poor
Cuff is too loose or too tight
Cuff is applied over clothing rather
than on a bare upper arm
Taking a measurement within 1 hour
of eating or drinking
Taking a measurement within 20
minutes of taking a bath
Being in a cold environment
Taking a measurement immediately
after smoking or drinking caffeinated beverages
Taking a measurement while talking
or moving your fingers
Taking a measurement when you
need to urinate
Taking measurements in rapid
succession
Factors that promote accurate blood pressure measurements
Proper cuff position on the upper
arm
Proper cuff tightness
No clothing under the cuff or
restricting circulation above the cuff
Proper posture: feet flat on the floor,
back straight, legs uncrossed, arm resting on a flat surface, upper arm at heart level
Relaxing or sitting still for a few
minutes after positioning the cuff
Page 38
34
Welch Allyn Home Blood Pressure Monitor
and before taking a blood pressure measurement
Sitting still during measurement
Not talking during measurement
Waiting at least 3 minutes between
measurements
Page 39
Directions for use 35
Maintenance
Maintain the monitor and cuff
The monitor does not require calibration.
To get the best performance from your monitor, follow the maintenance steps below.
Store the unit in a dry place away
from direct sunlight.
Avoid shaking and dropping the
monitor.
Avoid operating the monitor in dusty
and unstable temperature environments.
Cleaning
Caution Use a soft cloth to clean the entire unit. Do not use any abrasive or volatile cleaners.
1. Clean the monitor only when
necessary with one of the following compatible cleaning agents:
70% isopropyl alcohol
10% chlorine bleach/90% water solution (standard bleach wipe)
2. If necessary, you may wipe the cuff with a soft, damp cloth.
Page 40
36
Welch Allyn Home Blood Pressure Monitor
Troubleshooting
This section includes a list of error messages and frequently asked questions for problems you may encounter with your blood pressure monitor. If the device is not operating as you think it should, check here before arranging for service.
Problems and error messages
The device presents technical alarms and low-priority physiological alarms. Technical alarms occur when there is a device-related error. Physiological alarms occur when blood pressure measurements fall outside of set alarm limits.
Technical alarms
Problem Symptom Root
No power Display will
not light up
cause
Batteries are drained.
Batteries are inserted incorrectly.
AC ad apter is inserted incorrectly
Solution
Replace with new batteries
Insert the batteries correctly
Insert the AC ad apter tightly
Page 41
Directions for use 37
Problem Symptom Root
cause
Low batteries
Error messages
The display indicates the “BAT­LO” message, pauses for 3 seconds. The battery icon shows empty (does not flash).
E 01 shows The cuff is
E 03 shows There is too
E 10 or E 11 shows
Batteries are low.
not secure, the cuff is too tight, or a leak is detected.
much pressure in the cuff
The monitor detected motion while measuring
Solution
Replace with new batteries.
Readjust the cuff, have the patient relax for a moment, and then measure again. Inspect the connection tube for leaks or kinks.
Refasten the cuff and then measure again
Readjust the cuff, relax for a moment, and then measure again
Page 42
38
Welch Allyn Home Blood Pressure Monitor
Problem Symptom Root
cause
Visit
hillrom.com
EE or E 19 shows
Hardware error
for further assistance.
Solution
Measure again. If the problem persists, contact the retailer or Wel ch Allyn customer service at
hillrom.com
for further assistance.
Page 43
Physiological alarms
Symptom Root cause Solution
Out of range. Either SYS >260mmHg or DIA >220mmHg. The symbol may appear in either the SYS or DIA area of the screen.
Out of range. Either SYS <50mmHg or DIA <25mmHg. The symbol may appear in either the SYS or DIA area of the screen.
Directions for use 39
Press and hold the
Power
button to reestablish
Bluetooth
connectivity. Measure again. If the problem persists, contact your physician.
Alarm priority = Low
Press and hold the
Power
button to reestablish
Bluetooth
connectivity. Measure again. If the problem persists, contact your physician.
Alarm priority = Low
Page 44
40
Welch Allyn Home Blood Pressure Monitor
Specifications
Item Specification
Power supply: Battery powered mode:
Power supply: AC adapter powered mode:
Power supply model number
Display mode Digital LCD
Measurement model Oscillometric testing
Measurement range Rated cuff pressure:
Accuracy Pressure:
6VDC 4*AA batteries
Input: 100-240V, 50­60Hz, 400mA; Output: 6V, 1A
UE08WCP­060100SPA
V. A. 6 8 mm * 90 m m
mode
0mmHg to 300mmHg (0kPa­40kPa) Measurement pressure: SYS: 50 to 260mmHg DIA: 25 to 220mmHg Pulse value: (40 to
199)beat/minute
±0.4kPa (3mmHg) Pulse value:±4%
Page 45
Directions for use 41
Item Specification
Normal working conditions
Storage and transportation condition
Tem pe ra tu re : 5 °C to 40°C Relative humidity
85% RH Atmospheric pressure: 86kPa to 106kPa
Tem pe ra tu re :- 20 °C to 60°C Relative Humidity: 10% RH–93% RH Atmospheric Pressure: 50kPa to 106 kPa
Circumferen ce of the upper arm
Part Number
D-Ring Extra Small Cuff BPACC-01
Cuff size
15–24cm
D-Ring Standard Wide Cuff BPACC­02
D-Ring Extra Large Cuff BPACC­03
22–42cm
40–54cm
Page 46
42
Welch Allyn Home Blood Pressure Monitor
Item Specification
Net Weight Approx. 283g
(Excluding the dry cells)
External dimensions Approx. 94mm
*142mm*66mm
Degree of protection Type BF applied part
Protection against ingress of water
IP22
Software Version Version A01
General radio compliance
Item Specification
Bluetooth
No.
Radio Frequency (RF) Range
Output Power 4.0 dBm
Supply Voltage 1.8V to 3.6V
Antenna Gain 0.0 dBi
Transmitting Distance
Module
AW5182 2
2402 MHz to 2480 MHz
10 meters (30 feet)
Page 47
Directions for use 43
Federal Communication Commission (FCC) Interference Statement
This device complies with part 15 of the FCC Rules. Operation is subject to the following two conditions.
This device may not cause harmful interference.
This device must accept any interference received, including interference that may cause undesired operation.
This equipment (FCC ID: OU9TMB1591­A) has been tested and found to comply with the limits for a Class B digital device, pursuant to Part 15 of the FCC Rules. These limits are designed to provide reasonable protection against harmful interference in a residential installation. This equipment generates, uses, and can radiate radio frequency energy and, if not installed and used in accordance with the instructions, may cause harmful interference to radio communications. However, there is no guarantee that interference will not occur in a particular installation. If this equipment does cause harmful interference to radio or television reception, which can be determined by
Page 48
44
Welch Allyn Home Blood Pressure Monitor
turning the equipment off and on, the user is encouraged to try to correct the interference by one of the following measures.
1. Reorient or relocate the receiving antenna.
2. Increase the separation between the equipment and receiver.
3. Consult the dealer or an experienced radio/TV technician for help.
Any changes or modifications not expressly approved by the party responsible for compliance could void the user's authority to operate this equipment.
FCC Radiation Exposure Statement
This equipment complies with FCC radiation exposure limits set forth for an uncontrolled environment. This transmitter must not be co-located or operating in conjunction with any other antenna or transmitter.
Industry Canada (IC) compliance
To ensure compliance with FCC and Industry Canada RF exposure requirements, this device must be
Page 49
Directions for use 45
installed in a location where the antennas of the device will have a minimum distance of at least 20 cm from all persons. Using higher gain antennas and types of antennas not certified for use with this product is not allowed. The device shall not be co-located with another transmitter.
Pour s'assurer de la conformité des exigences aux expositions RF de la FCC et de l'Industry Canada, cet appareil doit être installé dans un emplacement où les antennes de l'appareil sont au moins à 20 cm de distance de toute personne. L'utilisation d'antennes et de types d'antennes à gain supérieur non garantis avec ce produit est interdite. L'appareil ne doit pas être installé à proximité d'un autre émetteur.
This device complies with Industry Canada licence-exempt RSS standard(s). Operation is subject to the following two conditions: (1) this device may not cause interference, and (2) this device must accept any interference, including interference that may cause undesired operation of the device.
Cet appareil est conforme aux normes RSS exemptes de licence de l'Industry
Page 50
46
Welch Allyn Home Blood Pressure Monitor
Canada. Son fonctionnement est soumis aux deux conditions suivantes: (1) l'appareil ne doit pas produire de brouillage, et (2) l'appareil doit accepter tout brouillage radioélectrique subi, même si le brouillage est susceptible d'en compromettre le fonctionnement.
This radio transmitter (IC: 12725A­TMB1591A) has been approved by Industry Canada to operate with the antenna listed in the specification table.
Ce transmetteur de radio (IC: 12725A­TMB1591A) a été approuvé par l'Industry Canada pour fonctionner avec l'antenne répertoriée dans la table des spécifications.
This Class B digital apparatus complies with Canadian ICES-003.
Cet appareil numérique de la classe B est conforme à la norme NMB-003 du Canada.
Page 51
Directions for use 47
Complied standards list
Item Standard
Risk management
Labeling ISO/EN 15223-1 Medical
User manual EN 1041 Medical
General Requirements for Safety
ISO/EN 14971 Medical devices — Application of risk management to medical devices
devices. Symbols to be used with medical device labels, labeling and information to be supplied. General requirements
equipment manufacturers to provide information
IEC 60601-1+A1 Medical electrical equipment ­Part 1: General requirements for basic safety and essential performance
IEC 60601-1-11 Medical electrical equipment ­Part 1-11: General requirements for basic safety and essential performance - Collateral standard: Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment
Page 52
48
Welch Allyn Home Blood Pressure Monitor
Item Standard
Electromagnetic compatibility
IEC/EN 60601-1-2 Medical electrical equipment ­Part 1-2: General requirements for basic safety and essential performance - Collateral standard: Electromagnetic compatibility ­Requirements and tests
Performance requirements and clinical investigation
IEC 80601-2-30 Medical electrical equipment- Part 2-30: Particular requirements for the basic safety and essential performance of automated non-invasive sphygmomanometers
ISO81060-2:2013 Non­invasive sphygmomanometers — Part 2: Clinical validation of automated measurement type
Software life­cycle processes
IEC/EN 62304+AC: Medical device software ­Software life cycle processes
Page 53
Directions for use 49
Item Standard
Usability IEC 62366 Medical
Small-bore connectors
devices - Application of usability engineering to medical devices (IEC
62366)
IEC 60601-1-6 Medical electrical equipment ­Part 1 -6: General requirements for basic safety and essential performance - collateral standard: Usability
IEC 80369-5 Small-Bore Connectors for Liquids and Gases in Healthcare Applications - Part 5: Connectors for Limb Cuff Inflation Applications
Page 54
50
Welch Allyn Home Blood Pressure Monitor
Warranty
Welch Allyn will warranty the blood pressure monitor to be free of defects in material and workmanship and to perform in accordance with manufacturer specifications for the period of two years from the date of purchase from Welch Allyn or its authorized distributors or agents.
The warranty period shall start on the date of purchase. The date of purchase is: 1) the invoiced ship date if the device was purchased directly from Welch Allyn, 2) the date specified during product registration, 3) the date of purchase of the product from a Welch Allyn authorized distributor as documented from a receipt from said distributor.
This warranty does not cover damage caused by: 1) handling during shipping,
2) use or maintenance contrary to
labeled instructions, 3) alteration or repair by anyone not authorized by Welch Allyn, and 4) accidents.
Page 55
Directions for use 51
The product warranty is also subject to the following terms and limitations.
Accessories are not covered by the warranty.
Shipping cost to return a device to a Welch Allyn service center is not included.
A service notification number must be obtained from Welch Allyn prior to returning any products or accessories to Welch Allyn's designated service centers for repair. To obtain a service notification number, contact Welch Allyn Technical Support at technicalsupport@hillrom.com.
Page 56
52
Welch Allyn Home Blood Pressure Monitor
Approved accessories
Item Description
BPACC-01 D-Ring Extra Small Cuff
BPACC-02 D-Ring Standard Wide
BPACC-03 D-Ring Extra Large Cuff
107041 ProBP 2000 adapter
(15–24cm)
Cuff (22-42cm)
(40–54cm)
set. This adapter is an alternate power source for the blood pressure monitoring device.
Page 57
Electromagnetic emissions
Note
Caution Wireless communications equipment, such as wireless home network
devices, mobile phones, cordless telephones and their base stations, walkie-talkies
can affect this equipment and should be kept at least a distance d=3,3m away
from the equipment.
Directions for use 53
energy only for its internal function. Therefore, its RF
emissions are very low and are not likely to cause any
interference in nearby electronic equipment.
Compliance Electromagnetic environment - guidance
As indicated in Table 6 of IEC 60601-1-2:2007 for ME Equipment, a typical cell
phone with a maximum output power of 2 W yields d=3,3m at an IMMUNITY LEVEL
of 3V/m).
Group 1 The Welch Allyn Home Blood Pressure Monitor uses RF
The Welch Allyn Home Blood Pressure Monitor H-BP100SBP is intended for use in the
electromagnetic environment specified below. The customer or user of the Welch Allyn
Home Blood Pressure Monitor should assure that it is used in such an environment.
Emissions
test
RF emissions
CISPR 11
Emissions and immunity information
Page 58
54
Welch Allyn Home Blood Pressure Monitor
suitable for use in all establishments, including domestic
establishments and those directly connected to the
public low-voltage power supply network that supplies
buildings used for domestic purposes.
Electromagnetic emissions
Class B The Welch Allyn Home Blood Pressure Monitor is
Class A
Complies
RF emissions
CISPR 11
Harmonic
emissions
IEC 61000-3-2
Voltage
fluctuations/
flicker
emissions
IEC 61000-3-3
Page 59
Electromagnetic immunity
Electromagnetic environment -
Compliance
IEC 60601
guidance
level
test level
Floors should be wood, concrete or
ceramic tile. If floors are covered with
synthetic material, the relative
humidity should be at least 30%.
±8 kV contact
± 15kV air
±8 kV contact
± 15kV air
Directions for use 55
Mains power quality should be that of
a typical commercial or hospital
±2 kV for
±2 kV for
environment.
power supply
lines
power supply
lines
Mains power quality should be that of
± 1 kV
±1 kV for input/
output lines
± 1 kV line(s) to
a typical commercial or hospital
differential
line(s)
environment.
mode
± 2kV line(s) to
earth
The Welch Allyn Home Blood Pressure Monitor is intended for use in the electromagnetic
environment specified below. The customer or the user of the Welch Allyn Home Blood
Pressure Monitor should assure that it is used in such an environment.
Immunity
test
Electrostatic
discharge
(ESD)
IEC 61000-4-2
Electrical fast
transient/burst
IEC 61000-4-4
Surge
IEC 61000-4-5
Page 60
56
Welch Allyn Home Blood Pressure Monitor
Mains power quality should be that of
a typical commercial or hospital
environment. If the user of the Welch
Allyn Home Blood Pressure Monitor
requires continued operation during
power mains interruptions, it is
recommended that the Welch Allyn
Home Blood Pressure Monitor be
powered from an uninterruptible
power supply or a battery.
)
)
T
T
Power frequency magnetic fields
should be at levels characteristic of a
typical location in a typical
commercial or hospital environment.
T
< 5% U
Electromagnetic immunity
T
< 5% U
Voltage dips,
short
(>95% dip in
(>95% dip in
interruptions
T
T
) for 0.5
T
U
cycle
) for 0.5
T
U
cycle
and voltage
variations on
40% U
40% U
power supply
input lines
for 5 cycles
70% U
(60% dip in U
)
T
T
T
70% U
for 5 cycles
(60% dip in U
IEC 61000-4-11
T
) for 5
for 25 cycles
(30% dip in U
)
T
for 25 cycles
(30% dip in U
T
(>95% dip in
U
seconds
3A/m and
< 5% U
T
< 5% U
(>95% dip in
30 A/m
) for 5
T
3A/m and
30 A/m
U
seconds
Power
frequency
(50Hz)
magnetic field
IEC 61000-4-8
Note: UT is the AC mains voltage prior to application of the test level.
Page 61
Directions for use 57
P
Electromagnetic environment -
guidance
Portable and mobile RF
communications equipment should
be used no closer to any part of the
Welch Allyn Home Blood Pressure
Monitor, including cables, than the
recommended separation distance
calculated from the equation
applicable to the frequency of the
transmitter.
Compliance
level
= 1,2
Recommended separation
distance
d
3 Vrms and
6Vrms
Electromagnetic immunity
IEC 60601
The Welch Allyn Home Blood Pressure Monitor is intended for use in the electromagnetic
test level
environment specified below. The customer or the user of the Welch Allyn Home Blood
Pressure Monitor should assure that it is used in such an environment.
Immunity
test
3 Vrms and
6Vrms
150 kHz to
Conducted RF
IEC 61000-4-6
80 MHz
Page 62
58
PPP
Welch Allyn Home Blood Pressure Monitor
800 MHz to 2,5 GHz
80 to 800 MHz
= 1,2
= 2,3
d
d
800 MHz to 2,6 GHz
= 2,3
d
where P is the maximum output power
should be less than the
a
the transmitter in watts (W)
is the recommended separation
d
and
rating of
. Interference may occur in the
b
compliance level in each frequency
range
distance in meters (m). Field
strengths from fixed RF transmitters,
as determined by an electromagnetic
site survey
vicinity of equipment marked with the
following symbol:
3 V/m
10 V/m
Electromagnetic immunity
10 V/m
80 MHz to
3 V/m
80 MHz to
2,5 GHz
Radiated RF
IEC 61000-4-3
2,6 GHz
Page 63
Electromagnetic immunity
Directions for use 59
Note 1: At 80 MHz and 800 MHz, the higher frequency range applies.
Note 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones
absorption and reflection from structures, objects, and people.
a
and land mobile radios, amateur radio, AM and FM radio broadcast, and TV broadcast cannot be
predicted theoretically with accuracy. To assess the electromagnetic environment due to fixed RF
transmitters, an electromagnetic site survey should be considered. If the measured field strength in the
location in which the Welch Allyn Home Blood Pressure Monitor is used exceeds the applicable RF
compliance level above, the Welch Allyn Home Blood Pressure Monitor should be observed to verify
Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.
normal operation. If abnormal performance is observed,additional measures may be necessary, such as
reorienting or relocating the Welch Allyn Home Blood Pressure Monitor.
b
Page 64
60
PPP
Welch Allyn Home Blood Pressure Monitor
= 2,3
800 MHz to
2,6 GHz
d
= 1,2
80 MHz to
800 MHz
d
= 1,2
Separation distance according to frequency of transmitter (m)
150 kHz to
80 MHz
d
Recommended separation distances between portable and mobile RF
communications equipment and the Welch Allyn Home Blood Pressure
Monitor
The Welch Allyn Home Blood Pressure Monitor is intended for use in an electromagnetic
environment in which radiated RF disturbances are controlled. The customer or user of the
Welch Allyn Home Blood Pressure Monitor can help prevent electromagnetic interference
by maintaining a minimum distance between portable and mobile RF communications
equipment (transmitters) and the Welch Allyn Home Blood Pressure Monitor as
recommended below, according to the maximum output power of the communications
equipment.
Rated maximum
output power of
transmitter
(W)
0.01 0.12 0.12 0.23
0.1 0.38 0.38 0.73
1 1.2 1.2 2.3
10 3.8 3.8 7.3
100121223
Page 65
is the maximum output power rating of the
P
in meters (m) can be estimated using the equation applicable to the
d
Directions for use 61
Recommended separation distances between portable and mobile RF
communications equipment and the Welch Allyn Home Blood Pressure
Monitor
transmitter in watts (W) according to the transmitter manufacturer.
frequency of the transmitter, where
For transmitters rated at a maximum output power not listed above, the recommended
separation distance
Note 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency range
applies.
Note 2: These guidelines may not apply in all situations. Electromagnetic propagation is
affected by absorption and reflection from structures, objects, and people.
Page 66
Page 67
Page 68
Welch Allyn® Home
Blood Pressure Monitor
Model: H-BP100SBP
749521 DIR 80025286 Ver. B