
NO. 49003
HALOGEN EADLIGHT
with
Thank you for purchasing this modern, convenient halogen headlight.
This manual provides specifications and instructions for use and maintenance, enabling you to fully enjoy the benefits offered by this headlight
over a long period of time.
Wall
Mounted Power Supply
Welch
State Street Road
Skaneateles Falls, NY
Allyn
13153-0220
U.S.A.

MOUNTING
Fasten mounting plate to wall or other surface with
screws supplied. The two holes in the mounting plate
should be vertical and flush with the wall. The wide
end of the plate should be down.
Slide power source into mounting plate from above,
being sure tabs on unit engage tabs on mounting
plate. Plug into any 120 V., 50-60 Hz AC* grounded
outlet and unit is ready to use.
l
*220-250 V., 50-60 Hz also available.
CAUTION: NOT FOR USE IN THE
PRESENCE OF FLAMMABLE
ANESTHETICS.

OPERATION
“QUICK DETACH” connector/switch
11 Ft. Cord to Headlight
CAUTION: NOT FOR USE
IN THE PRESENCE OF
FLAMMABLE ANESTHETICS.
m
_
C---
Knurled Ring--Green-- Ring-Green-
controls spot size

OPERATING INSTRUCTIONS
The headlight is attached to the power source cord by a “Quick
Detach” connector/switch provided on the headband. This connector will accept only the 490 Series Headlights.
Correct headband size is obtained by use of headband adjustment screw.
For convenience, two on-off switches are furnished. The main
power switch is on the power source and is provided with a pilot
light. This may be left on when the headlight is not in use. A
secondary switch is incorporated on the “Quick Detach” connector at the back of the headband. To operate, simply push connector
all the way down for “on,” back up to the first click for “off.” This
all
switch operates headlight only. Pull
headband.
Top of power source has a flange to provide convenience for hang-
ing headlight when not in use.
The diameter of the light spot can be regulated from
at distance of 14” by means of a green knurled ring on the lamp
housing. This ring controls an iris diaphragm. The intensity of
light is uniform at 400 foot candles minimum over entire spot area
up to 3” diameter.
The halogen (regenerative cycle) lamp in this unit will provide
uniform intensity of illumination for the life of the lamp-no gradual
dimming. A sophisticated optical system eliminates filament image,
shadows or streaks. The natural tissue colors are preserved.
the way off to detach from
3
/4" to 61/2"

SPECIFICATIONS
Model No. 49003
Input . . . . . . . . . . . . . . .
Output
Lamp
Fuse
Power Cord
Headlight Cord
Weight (approximate).
Dimensions (Power Supply) H.
.
.
. . . .
.
.
. . . .
120 V., 50-60 Hz,
7.2 V.
.
.
.
~3% at transformer-6.3 V. AC, 2.1 amp. at lamp
Welch
Allyn
.3
amp.
Slo-Blow--Factory
No. 18 AWG, 3 wire grounded cord, 8 feet long
11 feet long-will accept only No. 490 Series W.A. Headlights
Headlight-6 oz. Power Source w/Mounting Plate-l lb. 10 oz.
=
4.39”
D. = 2.30”
W. = 2.34”
.19
amp. (220-250 V., 50-60 Hz also available)
No. 02600 (Average life 100 hrs.)
Replacement Required
Circuit
Diagram
/ 7
Grounded

LAMP REPLACEMENT
___-
--r-1
CAUTION:
H
ALLOW LAMP TO COOL BEFORE CHANGING.
Unscrew serrated section “A” counterclockwise from section "B"
counterclockwise, and pull straight out. Replace with WA. No 02600 Lamp. Replace
by aligning lamp key wiih slot, depress fully, and turn clockwise until locked.
.
Depress lamp, turn
IMPORTANT
For best performance of halogen lamps, it is necessary to remove any grease or
fingerprints from the glass with a clean cloth moistened with alcohol.
Failure to follow cleaning procedure will result in permanent stains on the glass and
reduction of light output.
Fuse-Factory Replacement Necessary.
Ground Continuity Check-The resistance measured between the back plate of the
power source and the ground pin of the plug should be virtually zero.
For repairs, send complete unit (headlight and power source) to:
WELCH ALLYN, INC., Technical Services, 4341 State St. Rd. Skaneateles Falls, NY 13153-0220
Printed
Part
No. 490081-2
in U.S.A.