HALOGEN
EXAM LIGHT II
No. 48600 No. 48610
No. 48625 No. 48635
Thank you for purchasing the Welch Allyn
Halogen Exam Light II. We believe it is the
finest product of its kind available to you today.
By following the care and use guidelines found
in this booklet, you should be rewarded with
years of dependable, trouble-free service.
WARNING
The user of this equipment should be
thoroughly trained in the medical procedures appropriate to the instrumentation.
Furthermore, time should be taken to read
and understand these instructions before
performing any procedures. Instructions
for other equipment used in conjunction
with the Exam Light II (e.g., electrosurgical
generators, suction machines, etc.) should
also be read and understood. Failure to
do so may result in injury to the patient
and/or damage to the instrument.
MOUNTING
W
ALL MOUNTING
• Choose appropriate mounting location.
• Using mounting plate as template, mark drilling holes as needed for your wall type:
Dry wall, paneling, thin plywood or plaster
walls less than 1 inch (2.5 cm) thick.
Drill four .31 inch (.79 cm) holes, insert hollow
metal wall anchors making sure barbs are extended.
Concrete block or thick plaster walls
greater than 1 inch (2.5 cm) thick.
Drill four .25 inch (.64 cm) holes, insert plastic anchors.
Metal panel or wood walls greater
than 1 inch (2.5 cm) thick.
Drill four .31 inch (.79 cm) holes.
• Install mounting plate with appropriate screws. When tightened, there will be a gap
between the screw heads and surface of the mounting plate.
• Mount unit by fitting key hole slots on back of unit over screw heads.
• Push unit against wall while pushing down to lock screw heads into small end of key
hole slots.
TABLE MOUNTING
NOTE: Some table manufacturers supply table mounting hardware for mounting the
Exam Light II. Check with your table supplier for availability.
• Determine side of table for mounting the Exam Light II.
• Trace the outline of the mounting plate and mark the hole locations on a
1.00 inch (2.5 cm) thick piece of wood.
• Cut the piece of wood to the outline size and drill four .25 inch (.64 cm) holes.
• Using the mounting plate as template, mark the hole locations on the table and
drill four .25 inch (.64 cm) holes.
NOTE: Be sure mounting location is at least 10 inches (25.4 cm) below table cushion,
and that mounting will not obstruct table interior.
• Using four 10-32 x 2 inch (5.08 cm) machine screws with lock washers and nuts,
install the block of wood and the mounting plate to the table as shown in diagram.
• Mount unit as with wall mounting.
Wood Block
Mounting Plate
2 Inch (5.08 cm)
Long Screws
2
MOUNTING
M
OBILE MOUNTING
• Mount unit by fitting key hole slots on back of unit over screw heads.
• Push unit against mounting plate while pushing down to lock screw heads into
small end of key hole slots.
SET UP
• Carefully remove plastic plug from light pipe holder on top of unit.
• Insert fiber optic light pipe into light pipe holder until it is secure.
• Plug power cord into base of unit and into AC grounded outlet.
• Position the power switch located on the top of the unit to ON.
Plug
NOTE: Please refer to use of light pipe section
prior to using unit.
Max.
180°
To ensure long unit life, high light output and continued position rigidity, we
recommend that you do not acutely
bend the Light Pipe. For example: Middle
Bend—not more than 180°. End Bends—
not more than 90° each.
WARNING:
Not for use in the presence of flammable anesthetic.
Do not use Model 48200 (Standard model) Light Pipe for neonate transillumination.
Even though the distal end of the light pipe does not feel warm to the touch, it can
burn neonates. Erythema may result from contact of more than 15 seconds. Use
Model 48210 Neonate Transilluminator for safe transillumination.
Light
Pipe
Max. 90°
3