We, Grason-Stadler, Inc. warrant that this product is free from defects in material and
workmanship and, when properly installed and used, will perform in accordance with
applicable specifications.
meet this standard, it will be repaired, or at our option, replaced at no charge except
for transportation costs, when returned to an authorized Grason-Stadler service facility.
If field service is requested, there will be no charge for labor or material, however, there
will be a charge for travel expense at the service center’s current rate.
If within one year after original shipment it is found not to
NOTE
Changes in the product not approved in writing
Grason-Stadler shall void this warranty. Grason-Stadler
shall not be liable for any indirect, special or
consequential damages, even if notice has been given
in advance of the possibility of such damages.
THIS WARRANTY IS IN LIEU OF ALL OTHER
WARRANTIES, EXPRESS OR IMPLIED, INCLUDING BUT
NOT LIMITED TO, ANY IMPLIED WARRANTY OF
MERCHANTABILITY OF FITNESS FOR A PARTICULAR
PURPOSE.
by
GSI
Grason-Stadler
A Welch Allyn Company
Grason-Stadler, Inc.
1 Westchester Drive
Milford, NH 03055-3056
Telephone: 603-672-0470
Fax: 603-672-0487
Service
Manual 1717-0110,,
Printed January, 1991
Printed April, 1991
Printed January, 1992
Printed January, 1994
i
(#1)
(#2)
(#3)
(#4)
Rev.
4
Table of Contents
Warranty
Table of
Warning
...........................
.L.
..............................
Key to Symbols Used on Schematics
Foreword
...........................
SECTION 1- PRODUCT SPECIFICATIONS
1.1
1.2
1.3
1.4
1.5
1.6
1.7
1.8
1.9
1.10
1.11
1.12
1.13
1.14
1.15
1.16
1.17
1.18
Catalog Listings
Standards
.....................
Test Stimulus
..................
..................
Output Hearing Level Controi
Signal to Noise Ratio
Tone Switch
Stimulus
Transducers
Front Panel
....................
......................
................
Controls/Rear
Power and Power Line
Battery Pack Voltage
...............
................
................
Environmental Conditions
Mechanical Dimensions
Material Used in
Leakage Test
Manufacture
....................
Headband Static Force
..............
...............
Earphone Cushion Attenuation
Accessories
....................
..............
............
Panel Connectors
..............
............
............
.....
: : 1
i
iii
vii
viii
ix
1
1
1
2
2
3"
3
3
5
5
5
6
6
6
6
6
7
-
SECTION 2
FUNCTIONAL DESCRIPTION
Figure 2-l:
2.1
2.2
2.3
2.4
2.5
SECTION 3
3.1
3.2
3.3
3.4
Front Panel Controls and Indicators
Rear Panel Connectors and Controls
General Operation of Controls
Selection of the Diagnostic Mode
Hardware Diagnostic Test (DO)
Pushbutton Diagnostic Test
..................................
................................
...................
.................
...................
................
.............
............
(D1)
.......
..........
...........
Display Diagnostic Test (D2)
A/D Converter Diagnostic Test
Attenuator Maximum Output Level Calibration
(D3) 1
(D4) .
: : 1 : 1 1 : :
16
: :
17
18
19
..
19
1
19
19
20
21
21
21
2";
22
22
22
23
23
23
23
23
23
24
24
25
25
25
27
27
28
29
31
33
34
..
38
SECTION
5
Figure 5-l:
5.1
Disassembly Instructions
5.2
5.3
SECTION
Accessories and Replacement Parts
6
6.1
Diagnostic Mode (DO)
6.2
6.3
-
DISASSEMBLY
Disassembly
Parts Numbers
-
TROUBLESHOOTING
....................
..............
..............
..........
Troubleshooting Using the Hardware
............
Figure 6-l:
Diagnostic
Functions
Output Related
Figure 6-2:
Troubleshooting
Block Diagram of Amplifier
Signal Flow
Left or Right Amplifier Troubleshooting Symptoms'
Mode
Controls and
.
..................
Symptoms
...
.......
iv
42
:3
44
45
46
48
48
49
-
FOREWORD
GENERAL SERVICE INFORMATION
Operating,
can be found in the Instruction Manual (1717-0100).
and/or bench testing of GSI 17 Audiometers should only be
performed by trained personnel. The following instructions are
provided primarily for use by persons who are skilled in the
repair of electronic equipment.
CMOS HANDLING PRECAUTIONS
Many of the integrated circuits on the P.C. boards are CMOS and
NMOS type.
Failure to observe the following precautions whenever a circuit
board or an integrated-circuit package is handled can result in
damage to the GSI 17.
a.
b.
C.
Place instrument and parts on a grounded, conductive work
surface.
Ground yourself (with a strap having about 1 M ohm
resistance).
Ground the frame of any test instrument or soldering iron to
be used.
check-out procedures,
CAUTION
trouble-shooting hints, etc.,
Repair
d.
-
If any circuit boards are
enclose them in conductive (anti-static) envelopes.
to be stored or transported,
ix
SECTION 1
._-
Product Specifications
1.1 CATALOG LISTINGS
1717-9700
1717-9710 GSI 17 Audiometer,
1717-9705 GSI 17 Audiometer, AC Power (Export)
1717-9715 GSI 17 Audiometer,
1.2 STANDARDS
This unit is a single channel,
It
is equipped with pure
conduction transducers (TDH39 Earphones).
ANSI S3.6 and IEC 645 Type 4 Audiometer Standards.
The GSI 17 is designed to meet current revisions of the following
standards and specifications for audiometers:
GSI 17 Audiometer, AC Power (USA)
AC Power and Battery (USA)
AC Power and Battery (Export)
pure tone screening audiometer.
tone stimulus signals, and air
This instrument meets
ANSI S3.6
IEC 645
IS0 389
IEC 601
1.3 TEST STIMULUS
PURE TONE SPECIFICATIONS
A pure tone is the only stimulus source for this audiometer.
1717-2010
Battery Pack includes GSI supplied NiCad Battery.
May also be used with six
(6),
Size C, Alkaline
Batteries (not included)
Replacement NiCad Battery
8410-0060
P
R3
R5
io
Figure 2-1:
CONNECTORS, CONTROLS, INDICATORS
8
SECTION
2
-.
This
Connectors,
Figure
Controls,and Indicators.
2.1 THE FRONT PANEL CONTROLS AND INDICATORS (FIGURE 2-1)
section of
2-l for the layout and positioning of
F1
-
F2 - Range
F3
-
F4
-
Present Bar for stimulus presentation.
Functional Description
the Service Manual is
Controls,
Power switch and indicators for ON
increase
standard
operation is indicated by the
LCD.
HL control for setting stimulus
Level indicated on LCD.
and Indicators of the GSI 17.
extension pushbutton allows
the
stimulus intensity 10
maximum
HL at
used to
any
frequency
"+"
being lit on
intensity
describe
Refer to
all
(1)
Connectors,
and OFF (0).
operator to
dB
above
-
level.
the
the
its
the
F5
F6 - Subject response
F7 - Frequency
F8 - Routing
F9 -
F10 -
Loudspeaker
-
presentation indicated by illuminated LED.
that the test
button.
control for setting stimulus
Frequency indicated in window adjacent to control.
switch for routing the stimulus signal to
the
indicated by illustration of
Tone
type.
FM
---
Low battery indicator to alert the
the
available
depending
are used).
left or
type
= warble tone
= steady tone
= pulsed tone
batteries
switch for setting
(recharge or
on whether
in operation as such.
indicator LED shows the operator
subject
right
have a
has pressed the handswitch
earphone.
limited
replace
NiCad
or
Left or
subject.
the
alkaline
stimulus
operator
operating
batteries
Actual
frequency.
right
soon,
batteries
tone
when
time
9
2.2 REAR PANEL CONNECTORS AND INDICATORS
R1 -
Power input
position
and rear panel indication of input
jack with front panel
indication of
power
specifications.
-
Attention,consult accompanying documents.
R2
-
Earphone output jacks with front panel
R3
indication
of left and right phone provided by illustration of
subject.
R4
Subject response switch input jack with front panel
-
indication of position.
R5
-
Calibration switch indication.
2.3 GENERALOPERATION OF CONTROLS
All
initialization is complete,
in
depends on the HL selection. When the Extended Range is
controls
this
document and for the Extended
are
valid at
except for special cases as indicated
all
times
Range
once
the
Pushbutton
power up
which
invalid
there will be no indication given to the operator other than
absence of the
"+"
LCD segment.
the
All polled
state.
All operations of all controls are
controls are checked every 12
msec
for a
debounced
change in
for 12
msec
before processing.
The
Extended
manner.
and
hold manner (except in the Calibration and Diagnostic modes
The Present Bar and Response Switch operate
Range pushbutton operates by a push ON,
by a
push OFF
press
where the Present Bar is push ON, push OFF).
When a press and hold type control is operated all other controls
are still active.
When the operation of a control is being processed the processing
of
all
other controls is delayed until the first
operation is
completed.
2.4POWER UP INITIALIZATION
When
the power switch is set to the ON position
the
instrument
will go through the following initialization process:
All
microprocessor
internal
and
external
hardware
components of
system will be initialized for the
the
required
type of operation.
All
LEDs
and LCD segments will be turned on to indicate that
the power up initialization is in progress.
The RAM will be tested using a read after write verification
process.
initialized
will be displayed on the HL display and the instrument
halt.
After each location is tested its value
to 0.
If an error is detected, an
will be
error
code
will
The
generate a
stored in the EPROM.
using any of the RAM.
will be displayed on the HL display and the instrument
halt.
The operating mode of the instrument will be determined by
the position of the Cal/Normal Switch and the four Cal
Option Dip Switches. Refer to Sections 3 for the
Calibration and Diagnostic Mode initialization.
remaining initialization process will be performed for the
Normal Mode.
The
initialized to 0 HL.
be
analog hardware
selections.
After
all controls active.
2.5 SLEEP MODE
EPROM will be tested by reading the complete
current
initialized to their currently selected positions.
initialization,
EPROM to
checksum which will be compared to a
This test will be performed without
If an error is detected an error code
HL selection
All other controls and displays
will
be initialized
the instrument will remain idle
and
the
HL display
according to
checksum
will be
will
The
will
All
these
with
The GSI 17 has a mode which is designed to lengthen battery life.
When
operation
Sleep
only when the unit
available when connected to mains (ac) power.
If a period
operated (front panel, internal,
"sleep",
Prior to entering the Sleep Mode,
indicate 3 dashes ' - - -
mode status.
7 Volt) which will disable much of the analog circuitry.
the
disconnected,
a STOP state which will halt its operation.
The
either the Present Bar or power switch are toggled.
the
unit
Sleep
unit has been left on, under
for
Mode.
thus drawing minirnal current.
enters
more than 5 minutes, the GSI 17 will
It is important to emphasize that this mode
is operating with batteries, and is not
of 5 minutes has elapsed
external) the unit will go to
the display will be updated to
'I
to advise the operator of its
It will also disable the voltage converter
the
the channel will turn OFF and the CPU
Mode is exited, and normal
sleep
mode,
battery power,
since
all
transducers
operation
any control was
resumed,
without
enter the
exists
sleep
(+7
to
When
will be
will
enter
when
-
SECTION 3
-
3.1 EARPHONE CORDS
With extended use,earphone cords tend to fray internally at
junctions with
fraying
associated
cord is flexed.
To
check
control (F7) to 1000 or 2000 Hz, set HL control (F3) at a
comfortable audible level,
earphone cord
intermittent
scratchy
the
conditions signifies that the cord should be replaced.
Repeat the test for the other earphone.
3.2 HUMAND RANDOM NOISE
will ultimately either decrease the signal level in
earphone or cause signals to be intermittent as
for
signal,
sound superimposed over the signal that coincides
flexing of the cord.
Routine Maintenance
both earphone and audiometer
either condition, set
press the Present Bar (F4)
next to plug at
abrupt
changes in
The presence of any of
the
both ends,
audiometer
signal
the
connectors. This
the
the
frequency
and
listening
level, or a
these
flex
for
with
three
This test can be made during the check for attenuator noise.
With the instrument set on 1000 Hz,
0 to 60 dB and listen for low-frequency hum (60 to 120 Hz) and
random noise (hiss or low rushing
levels.
permissible.
audible.
naive subjects and affect the accuracy of the audiogram.
Schedule the audiometer for immediate service if any of these
symptoms are detected.
3.3
DISTORTION AND FREQUENCY SHIFT
This check is most easily performed by listening to the output of
the
frequencies at
dBHL for normal ears).
Listen for rattling, rasping or distortion in the tones
presented.
change when the Frequency Selector (F5) is moved to a new
position.
other,
should be replaced. In any case,the audiometer should be
immediately scheduled for maintenance.
GSI 17
Some audible random noise at levels above 60 dB is
Below 60
Any of these noises can be confused with the signal by
through
a loud,
Listen also to ascertain that signal frequencies
If distortion is heard in one earphone but not in the
the chances are high that the earphone is at fault and
dB
however, only the signal should be
the
earphones
but not uncomfortable, level (70 to 80
move the HL control (F2) from
sound) at all attenuator
while
presenting
all 11
13
3.4 SPECIAL MESSAGES
The
GSI 17 is a
microprocessor-controlled
performs a self check each time the instrument is turned on
self check does not occur when the instrument is "awakened"
sleep mode).
the front panel LCD if any error in the instrument
on
is detected.
Certain messages will be displayed to the
These messages are described below.
instrument
operator
operation
which
(the
from
3.4.1 CAL
When a
calibration
"CAL" will be displayed to the operator.
function at
invalid results
displayed as long as the "problem" transducer and frequency
selected.
changing either
transducer or frequency is
error
this
can be
(i.e.
frequency with this transducer, so
right earphone at 2000
recorded.
The word
If the calibration error is an
the
frequency or
the
selected
which
Hz),
has a
the
word
The audiometer will not
that no
"CAL"
will be
are
isolated
transducer
situation,
(i.e. left
earphone or 3000 Hz) will restore normal instrument function.
As is
service technician should be contacted immediately.
make
the
case with any instrument
malfunction, a
note of the combination of transducer and
certified
Remember to
frequency which
resulted in the "CAL" message.
3.4.2 Exx
When an error code consisting of an
"E"
and a 2-digit number (xx,
= number) appears on the audiometer's display, a system error has
been detected.
not
will
permit the instrument to operate.
remain on
instrument will
The GSI 17 will enter a "lockout" mode which will
the
display
shut
The specific
for
several
seconds,
itself down completely.
error
then
Should an
code
the
Exx
appear on the LCD,take the following steps:
a.
Power down,
then power up again.
This error could be
only a temporary failure and may never appear again.
However,
should the Exx message
reappear,
proceed to
steps b. and c.
b.
C.
Record the numbers that appear on the display.
Contact your certified service representative and give
them the specific numbers you have recorded.
4.1 QUICK
REFERENCE
CALIBRATION
--X
a.
b.
C.
d.
Select Calibration Mode using Cal/Norm Switch.
Select Frequency and Reference HL for calibration.
Select Earphone Routing (L or R).
Turn Tone ON using Present Bar.
NOTE:
Display will update to the ANSI
proper calibration
as necessary.
e.
f.
Adjust Output Level to the required SPL using the HL Dial.
Store data using the
LCD Segment will illuminate to indicate storage of data.
g.
h
i.
NOTE:
REPLACED,
r?
w
Repeat for all frequencies.
Repeat for opposite earphone.
. .
Return Cal Switch to NORM position.
If A CAL ERROR occurs or the MICROPROCESSOR HAS BEEN
the following steps should be performed:
Short
JP1
pins 1 and 2 then power-up for 2-3 seconds.
Power-down and remove jumper.
6.6).
SPL required for
subtract or add microphone
-
+10 dB
Range Extender Pushbutton
(See Troubleshooting, Section
corrections
(+)
b.
C
(2
6
e.
Load Default Data. (See Dip Switch 4, this section).
Battery low/shut down calibration.
(See Diagnostic Mode
(D3), this section).
Attenuator maximum output level calibration.
Diagnostic Mode
(D4),
this section).
Proceed with Transducer Calibration.
17
(See
NORMAL/CAL MODE SWITCH
\
-
NORM/CAL
-
LR403
DIP SWITCH S2
(SQUARE PORT)
1
POSITION l/POSITION 4
Figure 4-2:
ACCESS TO CALIBRATION CONTROLS
R404
LINEARITY ADJUST
DISTORTION ADJUST
---t
18
-
-
Freq
(Hz)
Table 4-l:
125 250 500
CALIBRATION REFERENCE THRESHOLDS
750
1000
1500 2000 3000
4000
6000
8000
dB45.0 25.5 11.5 8.0
Standard Reference Threshold Levels re: 20
TDH39-P
9-A coupler.
earphones as measured on the National Bureau of Standards
Reference ANSI
7.0 6.5
S3.6-1989,
IS0 389-1975 Standards.
9.0
uPa
10.0
for
9.5
15.5
Telephonics
4.2CALIBRATION MODE OPERATION
The
Calibration Mode is
calibration.
Normal
Mode
produce
the required SPL for the selected HL.
to determine the correct
This
used to
calibration information is used
perform
attenuator
the
transducer
during
position to
The
Calibration
the
Mode consists of two modes.
The Custom SPL Selection Mode allows the selection of
SPL values
Calibration
for
alternate
transducers.
Mode is used to calibrate the sound
The
Transducer
pressure
customized
SPL
level
(SPL) of each stimulus for each transducer.
4.2
.1
CAL/NORMAL SWITCH
This switch is used to select the operating mode of the
instrument.
switch to the CAL position while Dip Switch
position.
The Calibration Mode is entered by setting this
S1
is in the OFF
The Calibration Mode is exited by setting this switch
to the NORMAL position.
13.0
4.2.2CALIBRATION MODE DIP SWITCHES
(See
"Calibration Mode Dip Switch In-Depth Description" later in
this section.)
These
Mode.
Dip Switches select various options of
They are read to determine the selected options when
Calibration Mode is entered.
the Calibration
the
79
Switch
#
Function
OFF (down)
ON
(up)
(up)
s4
Default Data
OFF
Load
Loads the default Calibration
data into EEPROM when the
Calibration Mode is entered.
s3
Custom SPL Mode
GSI
Custom
Enables the selection of user
specified SPL values during
the transducer calibration
procedure when set to the
Custom position.
s2
S1
Not used
Diagnostic Mode
N/A
OFF
N/A
Selected
Selects the Calibration Mode
when the Cal/Normal Switch
is set to the Cal position
if this switch is in the OFF
position.
The status of Dip Switches #4 and #1 are read only when the
Calibration Mode is entered.
Dip Switch #3 is read and a change
in its position is processed if the Calibration Mode was entered
with this switch in the ON position.
Then the switch cay be
returned to the OFF position without exiting the Calibration
Mode.
-’
4.3CALIBRATION MODE ENTRY
The Calibration Mode can be entered via the Cal/Normal Switch at
any time except during the power up initialization phase or when
a stimulus is being presented.
In these cases the Cal Mode will
be entered when the initialization is complete or the stimulus
turned off.
The indication that the instrument is in the
Calibration Mode will be the non integer value in the format of
xx.x on the HL Display.
During the Calibration Mode initialization process all
controls
will be inactive.
When the Calibration Mode is entered,
low battery segment will be turned off.
all LCD segments except the
If Dip Switch S4 is
set
to the ON position the process of loading the default calibration
data
except
default data and checksum are being loaded.
will be started and the LCD display will have all
segments
the low battery segment turned on to indicate that the
When completed,
the
display will reinitialize to all segments except the low battery
segment will be off.
-
20
id'
The EEPROM will then be tested by reading the complete EEPROM to
generate
in
the
flashed on
a checksum which will be compared to a checksum
EEPROM.
the HL display and the initialization
If an error is found the word
"CAL
"
process will
stored
will be
continue.
Dip Switch #3 will be checked to determine which CAL mode is to
be initialized.
If in the OFF position the transducer SPL
calibration mode is selected,and if in the ON position, the
Custom SPL selection mode will be selected.
4.4 MODE INITIALIZATION
When
the
transducer
SPL Calibration
Mode is
selected the
instrument will be initialized as follows:
Tone Type = Current selection
,=
Frequency
Current selection
Channel = OFF
Calibration HL = 10 dB below the nonextended HL limit
for.
the current transducer and frequency
Transducer SPL = Calibration HL + Current RTL
+10 dB
Pad = Out
Transducer Routing = Current selection
HL Display =Current custom SPL level stored in EEPROM
Response LED = OFF
Stimulus
On LED = OFF
Low Battery Segment = Current state
"+"
Segment = OFF
4.5 EARPHONE SPL CALIBRATION MODE
This Calibration Mode is used to calibrate the SPL for each
stimulus for each transducer.
This is done by adjusting the
selected transducer output SPL level to the specified value
displayed on the HL display.
This SPL value is the sum of the
calibration HL level and the Reference Threshold Level (RTL) for
the current transducer and frequency.
standard
RTL value or
a custom RTL
The RTL may be either the
value
depending on
the
position of Dip Switch #3 when the Calibration Mode was entered.
\_
The transducer
SPL calibration data is
stored in
EEPROM by
transducer and stimulus.
4.5
.1
TONE TYPE SWITCH
This switch
operates as in the Normal Mode to select
Type of the stimulus.
4.5.2 ROUTING SWITCH
This
switch
operates as
in the Normal
Mode to
transducer to be calibrated. All transducers must be
and they can be calibrated in any sequence.
21
the
select
Tone
the
calibrated
4.5.3 FREQUENCY SELECTOR
This control
frequency
operates as
in the Normal
Mode to
select
to be calibrated. All frequencies must be
the
calibrated
and they can be calibrated in any sequence.
4.5.4
The
operates in
operated
PRESENT BAR
Present Bar controls the presentation of the
a push ON/push OFF manner.
If the Present
while the stimulus is off, the stimulus will be
stimulus.
Bar is
turned
It
on and remain on until the Present Bar is operated a second time.
The
stimulus will actually be turned ON or OFF when the
Present
Bar is pressed.
4.5.5
1.
HL DIAL
This control performs the dual function of selecting the HL
level at which the calibration is being performed and of
adjusting the output level of the transducer.
2.
While the stimulus is off,
Calibration HL level (Cal HL).
up by 5
dB
for each clockwise step of the selector and down
by 5 dB for each counter clockwise step.
the HL selector will select the
The level will be adjusted
For each change in
the Calibration HL level the transducer output level
(attenuator position) will be recalculated using the current
calibration data stored in the EEPROM.
the calibration HL will be from the nonextended
of
range) maximum HL limit to 20 dB below the limit.
the selector to select levels outside of this range
of
invalid.
When this occurs the level will default to the
The allowable range
(+10 dB
Operation
are
limit and the HL display will be flashed.
3.
While the stimulus is on,the HL selector (dial) will adjust
the transducer output level to allow the SPL level to be
adjusted to the required SPL level on the sound level meter.
The output level will be adjusted up by 0.5 dB for each
clockwise step of the selector and down by 0.5 dB for each
counter clockwise step. The output level is changed
according to the HL selector by directly setting the
attenuator position. Operation of the HL dial to select
positions outside of the instrument range will be invalid.
The attenuator position will be adjusted within the limits
and the displayed SPL value on the HL display will be
flashed.
4.5.6 EXTENDED RANGE PUSHBUTTON
This pushbutton is used to store the transducer SPL
data
stored,
Response LED will be turned on for 2 seconds.
data can
into
the
not be stored successfully, the word
EEPROM.
After the calibration data is
EEPROM checksum will be updated
and
If the calibration
calibration
successfully
the
"CAL"
Patient
"
will be
flashed on the HL display and the previously displayed SPL value
22
Loading...
+ 74 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.