Welch Allyn GS 777 User Manual

Intended use
This product is intended to supply power to Welch Allyn 3.5V instruments.
Symbol descriptions
General warnings and cautions
WARNING Read and understand the instructions before using this
product. Failure to understand the operating requirements for this product can result in injury to you or the patient and can damage the product.
WARNING Electric shock hazard. Do not simultaneously touch the patient and the center metal contact of the handle.
WARNING Do not modify this equipment. Any modification of this equipment can lead to patient injury. Any modification of this equipment voids the product warranty.
Caution As a precaution, avoid using this product in close proximity to other equipment. The product complies with current required standards for electromagnetic interference and should not present problems to other equipment or be affected by other devices.
Caution Only authorized Welch Allyn Service Centers can perform service on this product.
Caution To make sure that the transformer can be unplugged from the AC power source, never block access to the wall socket.
Caution Rx only (for professional use).
Setup
WARNING Welch Allyn is not responsible for the integrity of any
mounting installation. Welch Allyn recommends that customers contact their Biomedical Engineering Department or maintenance service to ensure professional installation for safety and reliability of any mounting accessory.
Attach Welch Allyn 3.5V instrument heads of your choice to the handles. See the directions for use for each instrument head.
Using the handles
Ensure that the transformer is plugged in.
1. Lift the handle from the cradle.
The instrument powers on. Only one instrument can be powered at a time.
2. Adjust the light output by turning the rheostat on the handle.
• Turning the rheostat clockwise increases the light output.
• Turning the rheostat counterclockwise decreases the light output.
Note Decreasing the light output to the minimum level does not power down the instrument.
3. Follow the directions for use for the instrument head when examining a patient.
Caution To prevent damage, do not overstretch the cords on the handles.
Caution To minimize the external housing temperature of the instrument head, on-time must not exceed 2 minutes, and off-time must be a minimum of 10 minutes.
4. Return the handle to the cradle.
The 3.5V instrument powers down.
Cleaning
WARNING Before you clean the transformer, unplug it from the wall
socket.
WARNING Electric shock hazard. DO NOT autoclave the transformer or accessories.
WARNING Liquids can damage electronics inside the transformer. Prevent liquids from spilling on or dripping into the transformer. If liquids are spilled on or drip into the transformer:
1. Disconnect the power plug.
2. Remove the transformer from the wall.
3. Dry off excess liquid from the transformer.
4. Mount the transformer on the wall.
5. Verify that the transformer functions normally before using it.
Clean the transformer on a routine basis according to your facility's protocols and standards or local regulations.
To clean the transformer (excluding the instrument heads), use any of the following:
• A clean cloth slightly dampened with 70 percent isopropyl alcohol or 10
percent chlorine bleach solution. Wipe dry with a clean cloth.
• Sani-Cloth
®
Plus.
• CaviWipes
(see caution below).
Caution Do not use CaviWipes to clean the blue handle cradles. This cleaning agent produces bubbles and liquid during the cleaning process that can enter openings in the cradles and limit handle performance for as long as 30 minutes after cleaning.
To clean the instrument heads, follow the instructions provided in their directions for use.
Specifications
Environment
Operating:
Transport/storage:
Additional technical information
www.welchallyn.com/GreenSeries777WallTransformer
Approvals
IEC 60601-1, UL 60601-1, IEC 60601-1-2, CAN/CSA C22.2 No 601.1-M90, AS/NZS3200
Warranty
Five-year limited warranty.
Guidance and manufacturer’s declaration
The guidance and manufacturer’s declaration information is available at www.welchallyn.com/GreenSeries777WallTransformer.
AC input power
DC input power
DC output power
Separate Waste Electronic and Electrical Equipment (WEEE) from other disposables. Prepare this product for reuse or separate collection as specified by the WEEE Directive 2002/96/EC or local law.
The product does not contain hazardous substances above threshold values defined in China Standard SJ/T 11363 –
2006. The product is green and environmental and can be recycled after use.
Illuminated: Power is present. Not illuminated: No power.
Type BF applied part
GS 777 Wall Transformer
Manual part no. 718210, 80017279 Ver. E © 2012 Welch Allyn
777XX
1
Caution Host USB to be used to power Welch Allyn approved devices only.
2
4
1/8 in (3 mm)
5
6
6
Temperature limitation Humidity limitation Atmospheric pressure limitation
Temperature limitation Humidity limitation Atmospheric pressure limitation
Safety Class II, Intermittent Operation
Electrical Power supply:
100–240V, 50–60 Hz, 0.18A max 5V, 1A
Transformer:
5V, 1A 5V, 0.5A
Welch Allyn, Inc. 4341 State Street Road Skaneateles Falls, NY 13153-0220 USA www.welchallyn.com
Regulatory Affairs Representative Welch Allyn Limited Navan Business Park Dublin Road Navan, County Meath Republic of Ireland
www.welchallyn.com/weee
59°F (15°C)
104°F (40°C)
700 hPa
1060 hPa
15%
95%
-4°F (-20°C)
120°F (50°C)
700 hPa
1060 hPa
15%
95%
3
1 3/8 in
(35 mm)
5 in
(127 mm)