Welch Allyn EXPENDABLE ILLUMINATOR User Manual

EXPENDABLE ILLUMINATOR
Mat’l No. 718532 Rev. A
4341 State Street Road Skaneateles Falls, NY 13153 USA Phone: (800) 535-6663 (US Only) or (315) 685-4560 • Fax: (315) 685-3361 www.welchallyn.com
EXPENDABLE ILLUMINATOR
Mat’l No. 718532 Rev. A
4341 State Street Road Skaneateles Falls, NY 13153 USA Phone: (800) 535-6663 (US Only) or (315) 685-4560 • Fax: (315) 685-3361 www.welchallyn.com
This illuminator is used in conjunction with 58600 and 58601 Disposable Vaginal Specula, Vaginal Speculum Light Handle (78800), and Transformer (73410)
To prepare for use:
1. Place illuminator on light handle.
2. Place illuminator and light handle into slot on 58600/58601 vaginal speculum handle.
3. Slide illuminator assembly up until it locks into place.
This illuminator is used in conjunction with 58600 and 58601 Disposable Vaginal Specula, Vaginal Speculum Light Handle (78800), and Transformer (73410)
To prepare for use:
1. Place illuminator on light handle.
2. Place illuminator and light handle into slot on 58600/58601 vaginal speculum handle.
3. Slide illuminator assembly up until it locks into place.
4. Plug in transformer.
4. Plug in transformer.
To remove illuminator:
To remove illuminator:
1. Place thumb at base of illuminator assembly and pull down.
2. Pull illuminator off of light handle.
Caution Lamp becomes hot during operation.
Cleaning and Sterilization
The 78600 Expendable Illuminator withstands over 50 cycles of steam autoclaving (270° F, 27 p.s.i., 6 min. cycle). After each time the unit shows degradation of light transmission and eventually needs replacement.
The 78600 Expendable Illuminator may be soaked in the following solutions:
Cidex/Cidex 7
Sporicidin
70% Isopropyl Alcohol
1. Place thumb at base of illuminator assembly and pull down.
2. Pull illuminator off of light handle.
Caution Lamp becomes hot during operation.
Cleaning and Sterilization
The 78600 Expendable Illuminator withstands over 50 cycles of steam autoclaving (270° F, 27 p.s.i., 6 min. cycle). After each time the unit shows degradation of light transmission and eventually needs replacement.
The 78600 Expendable Illuminator may be soaked in the following solutions:
Cidex/Cidex 7
Sporicidin
70% Isopropyl Alcohol
Note
Do not use iodine-based soaks (e.g. Betadine, Wescodyne) as these discolor the illuminator and adversely affect light output.
Note
Do not use iodine-based soaks (e.g. Betadine, Wescodyne) as these discolor the illuminator and adversely affect light output.
Loading...