Welch Allyn Euro-Med Rose Coated Electrosurgical Instructionumentation, Directions For Use and Sterilization User Manual

I. Turning the INSTRUMENT On

On/Off Power Switch

Video Colposcope
Quick Reference
Guide

Focus
Zoom
Green Filter
Flip the power switch on the handle of the colposcope to the “ON” posi­tion by pressing |. The video colpo­scope will then automatically execute an internal check of the optical sys­tem by cycling from low to high mag­nification ending in the low magnification setting. NOTE: DUR­ING THIS PROCESS, WHICH WILL TAKE ABOUT 30 SECONDS, you will not have the ability to use the controls.

II. Re-starting the System

The colposcope must be turned off for 15 seconds prior to re­starting the system. If the system is turned off for less the 15 seconds, the lamp will not re-ignite. If this occurs, turn the system off and wait for 15 seconds prior to turning the sys­tem on.

III. Focusing

To focus the video colposcope, adjust either the exam table or instrument to the desired height. Position the video colposcope so that the cervical image is centered on the monitor. At this point the image will most likely be blurry and out of focus. To bring the instrument into focus, move the entire colposcope by slowly advancing or withdrawing the instrument, relative to the patient, until the image comes into focus.
Once focus at low magnification is achieved, depress the ‘ZOOM +’ control until maximum magnification is obtained. The monitor will again most likely show a blurry image. To bring the image into focus, very slowly advance or withdraw the instrument, relative to the patient, until the image is sharp and in focus. Once this is achieved at high magnifica­tion, depress the ‘ZOOM -’ control and return to a lower mag-
nification setting. This procedure will allow the image to stay
Focus
Zoom
Mag Index
in focus throughout the full range of magnification during the entire procedure. You will only lose focus if either the colpo­scope or patient is moved.

IV. Fine Focus

Use of the fine ‘focus’< > controls is recommended ONLY for visualization at high magnification where the depth of field is 5 mm, or for inspecting the endocervical canal or vaginal fornices. Once focus is achieved (Section III), at high magnifi­cation, simply depress the ‘focus’ < > until the area of visual­ization is in focus.
Note: To “reset” the focus back to its original setting when the colposcope is first turned on, press and hold one of the ‘zoom +’ controls for 4 seconds after maximum magnification has been achieved.

V. Ta king a B io p s y

Once you have selected the appropriate magnification set­ting, pull the video colposcope away from the patient until the image is just out of focus. Press the fine focus forward control ‘focus >‘ until the image is in focus. This process will maximize your working distance for biopsy. Insert the forceps in the usual manner, then guide the instrument on the video monitor.

VI. Using the Green Filter

To use the electronic green filter, press the green filter button located on the left side of the video colposcope. The LED will remain lit while the green filter is on. To deactivate the filter, press the green filter button again.

VII. Mag Index Control

To activate the Magnification Index control, press the blue button located on the right side of the colposcope. To deactivate the Magnification Index control, press the blue Mag Index control button again.

VIII. Polarization Filter Control

The polarization filter can be rotated to reduce the glare from the image. The filter is completely engaged (minimum glare) when the control is fully rotated to the left. The filter can be disen­gaged by fully rotating the Polarization filter control to the right.